FR FieldReady Academy
Cheat sheet

CBET Memorization Sheet

Print this and keep it beside you. The CBET exam rewards recall of normal values, safety limits, device settings, units, standards, and formulas across every kind of equipment in a hospital. Every number on this sheet is fair game.

Download the PDF Start the free course

01The exam itself

FactValue
Certifying bodyAAMI Credentials Institute (ACI)
Questions165 multiple choice; 150 scored, 15 unscored pretest items mixed in
Time and rules3 hours, closed book, simple calculator allowed, no phones
PassingCriterion-referenced cut score (modified Angoff), about 116 of 165 correct, roughly 70%
Healthcare Technology and Function30%
Healthcare Technology Problem Solving30%
Healthcare Information Technology17%
Public Safety in the Healthcare Facility10%
Anatomy and Physiology7%
Fundamentals of Electricity and Electronics6%
Function and problem solving are 60% of the exam. For every device: what it measures or delivers, its settings and normal values, its alarms, its common faults, and the first check.

02Anatomy and physiology

Vital sign (adult, resting)Normal rangeMeasured by
Heart rate60 to 100 bpm (below 60 bradycardia, above 100 tachycardia)ECG, pulse oximeter, NIBP
Respiratory rate12 to 20 breaths per minuteImpedance pneumography, capnography
Blood pressureAbout 120/80 mmHg; MAP = diastolic + (systolic minus diastolic) / 3, about 70 to 100NIBP, arterial line
Oxygen saturation (SpO2)95 to 100%Pulse oximeter
End-tidal CO2 (EtCO2)35 to 45 mmHgCapnograph
Body temperature37 C (98.6 F); fever above about 38 CThermistor, thermocouple, infrared
Fetal heart rate110 to 160 bpmDoppler ultrasound
Tidal volumeAbout 500 mL at rest (6 to 8 mL/kg ideal body weight on a ventilator)Ventilator, spirometer
Cardiac outputAbout 5 L/min = heart rate x stroke volume (about 70 mL)Derived
SystemKey partsWhat the technician's devices do with it
CirculatoryHeart (right atrium, right ventricle, left atrium, left ventricle), tricuspid, pulmonary, mitral, aortic valves, arteries, veins, capillariesECG records electrical activity; NIBP and arterial lines measure pressure; defibrillators and pacemakers control rhythm; balloon pumps assist
Blood flowBody, vena cava, RA, tricuspid, RV, pulmonary valve, pulmonary artery, lungs, pulmonary veins, LA, mitral, LV, aortic valve, aorta, bodyRight side pumps to lungs, left side to body
Conduction pathSA node, AV node, bundle of His, bundle branches, Purkinje fibersP wave atrial depolarization, QRS ventricular depolarization, T wave ventricular repolarization
RespiratoryTrachea, bronchi, bronchioles, alveoli, diaphragmVentilators move gas; oximeters measure oxygen carried; capnographs measure CO2 exhaled; spirometers measure volumes
NervousBrain (cerebrum, cerebellum, brainstem), spinal cord, peripheral and autonomic nervesEEG records brain activity; EMG records muscle; stimulators test conduction; sympathetic raises heart rate, parasympathetic lowers it
Musculoskeletal206 bones, joints, skeletal (voluntary), smooth (involuntary), and cardiac musclePhysical therapy stimulators, traction, continuous passive motion
GastrointestinalEsophagus, stomach, small intestine (absorption), large intestine, liver, pancreas, gallbladderFeeding pumps deliver enteral nutrition; suction removes; endoscopes view
EndocrinePancreas (insulin, glucagon), thyroid, adrenals (epinephrine), pituitaryGlucose meters, insulin pumps
RenalKidneys (nephrons filter blood), ureters, bladderDialysis replaces filtration
SkinEpidermis, dermis, subcutaneous layer; barrier, temperature control, sensationElectrode contact and skin impedance, burns, pressure injuries
BloodPlasma, red cells (hemoglobin carries oxygen), white cells (immunity), platelets (clotting)Analyzers count and measure; oximetry depends on hemoglobin

Directional terms and word parts

  • Anterior front, posterior back; superior above, inferior below; medial toward the midline, lateral away; proximal near the trunk, distal far; supine face up, prone face down; planes: sagittal (left and right), coronal (front and back), transverse (top and bottom)
  • Systole contraction, diastole relaxation. Hypertension high pressure, hypotension low. Hypoxia low oxygen, hypercapnia high CO2, apnea no breathing
  • Prefixes: brady- slow, tachy- fast, hyper- high, hypo- low, a-/an- without, dys- difficult, peri- around, endo- inside
  • Roots: cardio heart, pulmo/pneumo lung, nephro/reno kidney, hepat liver, neuro nerve, gastro stomach, derm skin, osteo bone, hemo/hemat blood, myo muscle, encephal brain
  • Suffixes: -itis inflammation, -ectomy removal, -ostomy opening, -oscopy viewing, -graphy recording, -gram the record, -algia pain, -emia blood condition

ECG timing (normal adult)

  • PR interval 0.12 to 0.20 s; QRS under 0.12 s; QT under about 0.44 s
  • Paper speed 25 mm/s; 1 mV = 10 mm; small box 0.04 s (1 mm), large box 0.20 s (5 mm)
  • Heart rate = 1500 / small boxes between R waves, or 300 / large boxes
  • Ventricular fibrillation: chaotic, no QRS, shockable. Asystole: flat line, not shockable. Atrial fibrillation: irregular, no P waves
Anatomy detailFacts
Skin layersEpidermis (stratum basale at its base) over dermis over hypodermis (deepest); sebaceous glands oil, sweat glands cool, vitamin D made in skin
Muscle typesSkeletal striated voluntary; cardiac striated involuntary; smooth non-striated involuntary (organs, vessels, uterus, bladder)
Skeleton206 bones: 80 axial (skull, 33 vertebrae, ribs, sternum with manubrium and xiphoid, ossicles, hyoid) and 126 appendicular (scapula, clavicle, pelvis, humerus, radius, ulna, femur, patella, tibia, fibula)
BrainForebrain: cerebrum (frontal plans, parietal senses, occipital sees, temporal hears), thalamus relays, hypothalamus drives the pituitary. Midbrain relays vision and hearing. Hindbrain: cerebellum coordinates, pons relays, medulla runs breathing and heart rate. Brain stem = midbrain + pons + medulla
NervesSomatic voluntary; autonomic involuntary (sympathetic fight or flight, parasympathetic rest and digest); neurons signal by action potentials across synapses
Digestion in orderMouth (amylase, lipase), esophagus (epiglottis, lower esophageal sphincter), stomach (pH 1.5 to 3.5, pyloric sphincter releases chyme), duodenum, jejunum, ileum, large intestine (reabsorbs water); liver makes bile, glycogen, urea; gallbladder stores bile; pancreas enzymes and insulin
Lymph and immunityLymph filtered in nodes (neck, armpits, groin); tonsils, thymus (T cells), spleen (cleans blood), Peyer's patches; the immune system avoids, detects, and eliminates infection
EndocrineHypothalamus and pituitary control; thyroid metabolism; parathyroid (posterior thyroid lobes) raises calcium; adrenal cortex cortisol, medulla epinephrine; islets insulin and glucagon; pineal rhythms; negative feedback stabilizes, positive amplifies; the pons secretes nothing
KidneyCortex, medulla, pelvis; nephron: glomerulus in Bowman's capsule (blood pressure drives filtration), tubules reabsorb and secrete; ureter, bladder, urethra
ReproductionTestes, epididymis, vas deferens, prostate; FSH sperm, LH testosterone. Ovaries (estrogen, progesterone), fallopian tube (fertilization), uterus (implantation)
Blood termsHemoglobin carries oxygen; hematocrit is the red-cell fraction (about 45%); about 5 liters total; lymphocytes, neutrophils, monocytes, eosinophils, basophils

03Electricity and electronics

FormulaStatement
Ohm's lawV = I x R; I = V / R; R = V / I
PowerP = V x I = I squared x R = V squared / R (watts)
Series resistanceR total = R1 + R2 + ... ; same current through each; voltages add
Parallel resistance1 / R total = 1/R1 + 1/R2 + ... ; two resistors: (R1 x R2) / (R1 + R2); same voltage across each; currents add
KirchhoffCurrents into a node equal currents out; voltages around a loop sum to zero
AC valuesRMS = 0.707 x peak; peak = 1.414 x RMS; peak-to-peak = 2 x peak; US mains 120 V RMS at 60 Hz (many countries 230 V at 50 Hz)
TransformerV secondary / V primary = N secondary / N primary; step-down lowers voltage and raises current; isolation transformer 1:1
Capacitive reactanceXc = 1 / (2 x pi x f x C); capacitors pass AC, block DC
Inductive reactanceXL = 2 x pi x f x L; inductors pass DC, oppose changes in current
RC time constanttau = R x C; capacitor reaches 63% of full charge in one tau, about 99% in five
Frequency and periodf = 1 / T; 60 Hz has a period of 16.7 ms
Stored energyE = one half x C x V squared (joules); how a defibrillator capacitor stores its charge
EnergyE = P x t; 60 W for 8 h = 0.48 kWh; kWh is energy, kW is power
Voltage dividerV across R1 = V source x R1 / R total (series); 120 V across 60 and 100 ohms puts 45 V on the 60
Current dividerI through a branch = I total x (1/R branch) / (sum of 1/R); the smallest resistance takes the most current
Kirchhoff's lawsKCL: current into a node equals current out (nodal analysis). KVL: voltages around a loop sum to zero (mesh analysis)
Internal resistance(open-circuit voltage minus loaded voltage) / load current; 1.58 V to 1.46 V into 80 ohms = 6.58 ohms; rises with age
Wheatstone bridgeBalanced (no meter current) when R1 / R2 = R3 / R4; strain gauges and thermistors sit in bridges
Complex impedanceZ = R + jX; 40 mH at 60 Hz is about 15 ohms of reactance; an ideal 30 mH inductor at 60 Hz is 11.3j ohms
AC powerS squared = P squared + Q squared; P true (W), Q reactive (VAR), S apparent (VA); 180 W and 240 VAR give 300 VA; power factor = P / S
Op-amp gainInverting gain = R feedback / R input (gain 9 with 5 k in needs 45 k feedback); non-inverting = 1 + Rf / Rin
Transformer ratioV primary / V secondary = N primary / N secondary; 240 V to 6,000 V is a 0.04 ratio
C-rateCharge or discharge current relative to capacity; 60 kWh at 0.5 C = 30 kW
  • Resistor color code digits: black 0, brown 1, red 2, orange 3, yellow 4, green 5, blue 6, violet 7, gray 8, white 9; tolerance gold 5%, silver 10%
  • US cord colors: black (or red) hot, white neutral, green or bare ground. IEC cords: brown live, blue neutral, green-yellow ground
  • Schematic symbols: zigzag resistor, parallel lines capacitor, coil inductor, triangle-with-bar diode, ground, fuse, switch, battery; signals flow left to right; compare measured voltages to the values printed at test points
  • Passive devices (resistors, capacitors, inductors, transformers) cannot add energy; active devices (diodes, transistors, ICs, op-amps) control or amplify
  • Diode conducts one way, forward drop about 0.7 V silicon; zener regulates; LED lights; rectifiers: half-wave, full-wave, bridge; capacitor filters ripple; regulator holds output steady
  • Linear power supply: heavy transformer, low noise, inefficient. Switching supply: light, efficient, more noise, needs a load
  • Transistor: NPN or PNP, base controls collector-emitter; used as a switch or amplifier. SCR and triac switch AC power
  • Op-amp rules: huge gain, inputs draw no current, feedback makes the inputs equal; instrumentation amplifier rejects common-mode noise (high CMRR) for ECG front ends; isolation amplifiers and optocouplers keep patient circuits floating
  • Filters: low-pass keeps slow signals, high-pass removes baseline drift, band-pass keeps a range, notch removes 60 Hz
  • DMM measures volts, amps (in series), ohms (power off); oscilloscope shows voltage against time, 10x probe divides by ten; function generator provides test signals
  • Passive: resistor, capacitor, inductor, memristor, thermistor (no gain). Active: transistor, op-amp, IC, SCR. Transducer converts energy: sensor to electrical, actuator from electrical; piezoelectric crystals do both
  • US distribution: step-up to transmit, step-down to use; split-phase 120/240 V at 60 Hz with a grounded center neutral; NEC is NFPA 70 (building wiring), NESC is IEEE (utility lines)
  • Battery specifications: voltage per cell, capacity (Ah), energy density (sets weight), internal resistance (rises with age, sag under load), shelf life, discharge curve, cycle life, C-rate
  • Regulators: linear drops excess as heat (quiet, inefficient); switching chops at a duty cycle (efficient, noisy); shorted switching element passes full input voltage, open element gives zero; test regulator input and output pins to ground
  • Filters by what they pass: low-pass, high-pass, band-pass, band-stop (notch is a narrow band-stop)
Power storage and conditioningFacts
Cell voltagesLead-acid 2.0 V, nickel-cadmium 1.2 V, nickel-metal hydride 1.2 V, lithium-ion 3.6 to 3.7 V, alkaline 1.5 V
Battery behaviorNiCd memory effect (condition with full cycles); Li-ion no memory but thermal runaway risk and must not over-discharge; sealed lead-acid sulfates if left discharged; capacity in amp-hours falls with age
UPS typesStandby (switches on loss), line-interactive (regulates voltage), online double-conversion (always on inverter, best protection)
Power problemsSag or brownout (low voltage), surge (high), transient spike, noise, blackout; conditioners regulate and filter
Hospital essential electrical system (NFPA 99 and NFPA 70)Generator restores power within 10 seconds; branches: life safety, critical (patient care, red outlets), equipment; UPS bridges the gap for critical devices
Isolated power and LIMUngrounded secondary of an isolation transformer; line isolation monitor alarms at 5 mA total hazard current; used in wet procedure locations
GFCITrips at 4 to 6 mA of imbalance between hot and neutral in about 25 ms; wet locations

04Electrical safety

Current through the body (60 Hz, hand to hand)Effect
1 mAThreshold of perception (tingle)
5 mAAccepted maximum harmless current
10 to 20 mALet-go threshold: muscles contract and the person cannot release
50 mAPain, possible fainting, respiratory arrest
100 to 300 mAVentricular fibrillation likely
6 A and aboveSustained heart contraction, burns, respiratory paralysis
MicroshockAs little as 10 microamps delivered directly to the heart (through a catheter or pacing wire) can cause fibrillation
IEC 60601-1 limitNormal conditionSingle fault condition
Earth (ground) leakage5 mA10 mA
Touch (enclosure, chassis) current100 microamps500 microamps
Patient leakage, type B and BF, AC100 microamps500 microamps
Patient leakage, type CF, AC10 microamps50 microamps
Patient leakage, DC (all types)10 microamps50 microamps
Mains voltage on the applied partBF: 5 mA; CF: 50 microamps(single fault test)
NFPA 99 and facility factsValue
Ground (chassis to plug ground pin) resistance0.5 ohm maximum for patient-care equipment
Chassis (touch) leakage, cord-connected patient-care equipment500 microamps in current editions (older editions used 300 or 100 microamps; know the limit your facility cites)
Patient care vicinityAbout 6 feet (1.8 m) beyond the bed or chair and 7.5 feet (2.3 m) above the floor
Line isolation monitor alarm5 mA total hazard current
GFCI trip4 to 6 mA (nominal 5 mA)
Hospital-grade receptacleGreen dot; test retention force, polarity, ground continuity
Relocatable power tapsOnly where allowed, rated, and secured; never daisy-chained; no extension cords for patient care equipment
Relocatable power tap loadNo more than 75% of the tap's rating (an AWG 10 tap rated 35 A carries at most 26 A); never daisy-chained
ProhibitedThree-to-two prong (cheater) adapters; tinned stranded wire ends under screw terminals; cords without strain relief; isolated-ground receptacles in the patient care vicinity
Patient room receptaclesAt least 8 per general bed (more in critical care), non-locking, tamper-resistant; no minimum in bathrooms; OR receptacles single, duplex, or quadruplex with insulated grounds, locking or not
Household electronicsAllowed only outside the patient care vicinity unless double insulated; inspected before patient use; worn units removed
Gas cylinder storageUnwrapped, secured upright, minus 7 to 52 C; oxidizers 20 ft from flammables (5 ft with sprinklers, none in a compliant cabinet); no smoking within 20 ft; not in tight enclosures; door warning for other positive-pressure gases
Medical gas systemsDOT/TC/ASME cylinders; gas-specific outlets, no adapters; redundant supply; alarms; labeled zone shutoff valves; inventory and inspection
Telecommunications roomsNo data outlet more than 90 m from its room; at least one per floor; each serves at most 20,000 sq ft; interior rooms in disaster areas
Nurse callCarries device alarms, staff emergencies, code calls, patient requests; audible and visual annunciation; may forward to pagers
Touch current and clean agentsTouch current: chassis leakage reachable by a person via a path other than ground. Clean agents: nonconductive, residue-free extinguishants (nitrogen, CO2, noble gases, halocarbons)

Classes, types, symbols, and testing

  • Class I: protective earth (three-prong ground). Class II: double insulation, no ground (square-in-square symbol). Internally powered: battery only
  • Applied part type B: body, not for direct cardiac use (stick figure). BF: body floating, isolated (figure in a box). CF: cardiac floating, safe for direct heart connection (heart in a box)
  • Defibrillation-proof applied part: the symbol with paddles; withstands a defibrillator discharge
  • Macroshock: current through the skin; needs milliamps. Microshock: current bypassing the skin to the heart; microamps
  • Safety analyzer sequence: visual inspection and cord check, ground resistance, chassis leakage (normal, reversed polarity, open ground, open neutral), patient lead leakage, lead-to-lead, mains on applied part
  • Leakage rises with open ground and reversed polarity; a rising leakage trend on PM predicts insulation failure; test at incoming inspection, after repair, and per PM schedule

05Infection control, hazards, fire, gas, radiation, MRI, and standards

Infection control

  • Standard (universal) precautions: treat all blood and body fluids as infectious; hand hygiene before and after every patient contact and before touching equipment
  • Transmission-based precautions: contact (gown, gloves), droplet (surgical mask), airborne (N95 respirator, negative-pressure room)
  • PPE donning: gown, mask, eye protection, gloves. Doffing: gloves, eye protection, gown, mask (clean hands between)
  • Cleaning removes soil; disinfection kills most microbes; sterilization kills all including spores
  • Spaulding: critical items (enter sterile tissue) sterilize; semi-critical (touch mucous membranes) high-level disinfect; non-critical (intact skin) low-level disinfect
  • Steam autoclave: 121 C (250 F) at 15 psi for 15 to 30 minutes, or 132 to 135 C flash cycles; biological indicators verify; ethylene oxide and hydrogen peroxide plasma for heat-sensitive items
  • Clean and disinfect equipment before servicing it; sharps in puncture-resistant containers; OSHA bloodborne pathogens standard covers exposure control; report needlesticks immediately
  • Universal precautions (1980s): blood and blood-contaminated fluids only, not sweat, tears, nasal secretions, or urine; bleach cleanup; bagged waste. Standard precautions (CDC 1996): all body fluids plus body substance isolation; transmission-based precautions supplement them
  • Regulated waste: liquid or free-flowing blood, sealed for a licensed medical waste service; dried blood on dressings sealed in a plastic bag with regular trash; broken glass with tools, never hands
  • Lab practice: annual biological risk assessment sets the biosafety level; no mouth pipetting; no food, drink, or contact lenses; retractable needles; decontaminate before disposal; eyewash station; hands-free sinks; clean utility room separate from dirty utility; lab coats are not gowns

Hazard communication and signage

  • Safety data sheet (SDS): 16 sections including identification, hazards, first aid, fire fighting, handling and storage, exposure controls and PPE; must be accessible to staff
  • GHS pictograms: flame (flammable), flame over circle (oxidizer), skull (acute toxicity), corrosion, exclamation mark (irritant), health hazard (silhouette), gas cylinder, environment
  • NFPA 704 diamond: blue health, red flammability, yellow reactivity, white special (W with a line means water reactive, OX oxidizer); scale 0 to 4
  • Laser signs state the class and wavelength; classes 1 (safe) through 4 (fire and eye hazard); wavelength-specific eyewear
  • Ionizing radiation: trefoil symbol; time, distance, shielding; ALARA; inverse square law; lead aprons about 0.5 mm; dosimeters
  • MRI: zones I (public) through IV (magnet room); the magnet is always on; ferromagnetic objects become projectiles; quench vents helium; MR Safe, MR Conditional, MR Unsafe labels
  • Patient precaution signs at the door: contact, droplet, airborne, fall risk, latex allergy, no blood pressure on a limb
  • HazCom (OSHA 29 CFR 1910.1200) requires labels, SDS access, training, and records; it adopted the UN Globally Harmonized System (GHS) for classification, pictograms, signal words, and the 16-section SDS
  • SDS order: identification, hazards, composition, first aid, fire fighting, accidental release, handling and storage, exposure controls and PPE, physical and chemical properties, stability and reactivity, toxicology, ecology, disposal, transport, regulatory, other
  • Toxicity and limits: LD50 lethal dose for 50%, LC50 lethal concentration; PEL legal limit (OSHA); TLV recommended (ACGIH); REL recommended (NIOSH); STEL 15-minute ceiling; carcinogen lists from NTP, IARC, OSHA
  • GHS label: identity, manufacturer contact, pictograms, signal word (Danger or Warning only), hazard and precautionary statements; secondary containers need the same; pyrophoric ignites in air within 5 minutes; oxidizer pictogram exists, reducer does not

Fire and medical gas

  • Fire triangle: heat, fuel, oxygen. RACE: rescue, alarm, confine, extinguish. PASS: pull, aim, squeeze, sweep
  • Extinguisher classes: A ordinary combustibles, B flammable liquids, C electrical, D metals, K cooking oils
  • Oxygen-enriched atmospheres burn faster; no oil or grease on oxygen fittings; cylinders secured upright, capped, stored below 125 F, full separated from empty
  • US medical gas cylinder colors: oxygen green, nitrous oxide blue, carbon dioxide gray, nitrogen black, medical air yellow, helium brown; pin index and DISS fittings prevent wrong connections
  • Full E cylinder of oxygen: about 2,000 to 2,200 psi, about 660 liters; wall outlets about 50 psi; zone valves shut off gas to an area
Standards, regulators, and accreditationWhat to know
CMSCenters for Medicare and Medicaid Services; sets Conditions of Participation hospitals must meet to be paid; can survey directly
Deemed statusAccrediting organizations survey on CMS's behalf: The Joint Commission, DNV Healthcare, ACHC (formerly HFAP), CIHQ
The Joint CommissionEnvironment of Care and Emergency Management standards cover equipment inventory, maintenance strategies, testing, and documentation
FDARegulates devices; medical device reporting: user facilities report a device-related death to FDA and the manufacturer within 10 work days, a serious injury to the manufacturer within 10 work days; recalls Class I (serious harm), II (temporary harm), III (unlikely harm)
OSHAWorker safety: hazard communication, bloodborne pathogens, PPE, electrical safety in the workplace
ANSI/AAMI EQ56Recommended practice for a medical equipment management program
ANSI/AAMI EQ89Guidance for scheduled maintenance and performance verification
ANSI/AAMI EQ93Medical equipment management vocabulary and terminology
ANSI/AAMI EQ103Alternative equipment maintenance (AEM) programs: how to justify and document maintenance strategies that differ from the manufacturer's
AEM rules (CMS)Allowed with documented risk assessment by qualified staff; not for imaging or radiologic equipment, lasers, or where the manufacturer or law requires otherwise; new equipment follows the manufacturer until history supports a change
CMMSComputerized maintenance management system: inventory and asset tags, work orders, PM schedules, equipment history, parts, reports (PM completion rate, mean time between failures, cost); every activity documented there
After an incidentRemove the device with all accessories and disposables attached, do not clear logs or settings, tag and sequester, document, report; root cause analysis asks why repeatedly
Joint Commission programsAccredits and certifies by peer survey; publishes National Patient Safety Goals yearly; links quality and risk management; does not set financial benefits
DNVAccredits hospitals under NIAHO, integrating CMS Conditions of Participation with ISO 9001; annual surveys
ISO 9000 / 9001Quality management standards; plan-do-check-act; risk-based thinking; customer focus, leadership, process approach
CMS programsMedicare, Medicaid, CHIP; regulates human laboratory testing; publishes yearly core measure sets (clinical quality measures); fights fraud; does not accredit
CAPCollege of American Pathologists: nonprofit accreditor of clinical laboratories under CMS authority; inspections every two years plus unannounced
SMDASafe Medical Devices Act (1990, amended 1992): user facilities report device-related deaths to FDA and manufacturer, serious injuries to the manufacturer; annual summaries

06Physiological monitoring

ECG factValue
Einthoven's triangleLead I = LA minus RA; Lead II = LL minus RA; Lead III = LL minus LA; Einthoven's law: Lead II = Lead I + Lead III
Augmented and chest leadsaVR, aVL, aVF from the limb electrodes; V1 to V6 across the chest; 12 leads from 10 electrodes
Electrode colors (US, AHA)RA white, LA black, RL green (ground), LL red, V brown ("white on right, smoke over fire")
BandwidthMonitoring mode 0.5 to 40 Hz; diagnostic mode 0.05 to 150 Hz
Interference60 Hz: bad electrode contact, unshielded leads, nearby equipment; wandering baseline: respiration or motion; spikes: muscle tremor or loose lead
Lead-offHigh impedance detected on an electrode; check gel, skin prep, cable
Simulator testPatient simulator provides rate, amplitude, and arrhythmias to verify the monitor
Rhythms on the stripSinus tachycardia fast regular; atrial (supraventricular) tachycardia rapid consistent narrow complexes; atrial fibrillation irregular with quivering waves, no P; ventricular tachycardia rapid WIDE QRS; ventricular fibrillation chaotic; asystole flat (check leads)
Lead views and countsaVF's positive pole is the left leg; 12 leads from 10 electrodes (4 limb, 6 chest); IEC colors RA red, LA yellow, LL green, RL black differ from AHA; thermal print head tested by printing letters or diagonals; paper speeds 10, 12.5, 25, 50 mm/s test the drive
Blood pressure factValue
Korotkoff soundsAuscultatory method: cuff above systolic, deflate; first sound (phase 1) is systolic, sounds muffle then disappear (phase 5) at diastolic; automated devices use oscillometry instead
NIBP methodOscillometric: cuff inflates above systolic, deflates in steps, MAP is the point of maximum oscillation, systolic and diastolic derived
Cuff sizingBladder width about 40% of arm circumference, length about 80%; too small reads high, too large reads low
Invasive lineTransducer zeroed to atmosphere and leveled at the phlebostatic axis (fourth intercostal space, midaxillary line)
Dicrotic notchThe small dip on the downslope of the arterial waveform marking aortic valve closure; its absence suggests overdamping
Height error13.6 cm of water equals 10 mmHg; every 13.6 cm the transducer is off level changes the reading about 10 mmHg
DampingFast-flush (square wave) test: overdamped rounds the wave and reads low systolic; underdamped overshoots and reads high
Transducer sensitivity5 microvolts per volt per mmHg (industry standard)
SphygmomanometerCuff, bulb, valve, manometer (mercury, aneroid, electronic), stethoscope; gauge not at zero: zeroing screw; unstable: movement or placement; reads high: cuff below the heart; reads low: cuff above the heart or an air leak (most common); invasive lines read continuously even when hypotensive
Oximetry and capnographyValue
Pulse oximeter wavelengthsRed about 650 to 660 nm, infrared about 805 to 940 nm; oxyhemoglobin absorbs more infrared; ratio of pulsatile absorbances gives SpO2; high perfusion helps, low perfusion hurts
ErrorsCarboxyhemoglobin reads falsely high; methemoglobin trends toward 85%; motion, low perfusion, nail polish, ambient light, dyes
TestingSpO2 simulator or optical tester; check probe LED and detector, cable, and site
CapnographyInfrared absorption by CO2 (about 4.26 micrometers); mainstream sensor at the airway, sidestream samples through tubing (delay, water trap)
CapnogramBaseline near zero, rapid rise, alveolar plateau, EtCO2 at end of plateau, drop on inspiration; loss of waveform means disconnect, apnea, or esophageal intubation

Other monitoring

  • Temperature: thermistor (resistance falls as temperature rises, NTC), thermocouple (voltage from two metals), infrared tympanic; YSI 400 and 700 series probe curves are not interchangeable
  • Respiration on a monitor: impedance pneumography through the ECG electrodes; capnography is more reliable
  • Alarm priorities (IEC 60601-1-8): high red flashing, medium yellow flashing, low cyan or yellow steady; alarm fatigue is a patient safety problem; alarm management programs set limits per unit
  • Telemetry: WMTS bands 608 to 614 MHz, 1395 to 1400 MHz, 1427 to 1432 MHz; dropouts from antenna coverage, low battery, interference; central station shows all beds
  • Fetal monitor: ultrasound Doppler for fetal heart rate (110 to 160 bpm), tocodynamometer (TOCO) for contractions by pressure on the abdomen; internal scalp electrode and intrauterine pressure catheter are the direct methods
  • Respiration methods: chest impedance through ECG electrodes (most common), respiratory belts, airflow sampling (thermistor or capnograph), magnetometry; a fingertip oximeter that shuts off mid-use has a low battery
  • EEG 10-20 landmarks: nasion (bridge of the nose) to inion (back of the skull); beta above 13 Hz awake and active; differential amplifier removes common signals
  • EEG bands: delta under 4 Hz, theta 4 to 8, alpha 8 to 13, beta above 13; electrodes placed by the 10-20 system; EMG measures muscle electrical activity; evoked potentials time the response to a stimulus

07Test equipment and units of measure

Test equipmentVerifies
Electrical safety analyzerGround resistance, chassis and patient lead leakage under normal and fault conditions
Patient (physiological) simulatorECG rate, amplitude, arrhythmias, respiration, invasive pressure, temperature
SpO2 simulatorSaturation and pulse readings through the probe
NIBP simulatorCuff pressures and oscillometric readings; leak test
Defibrillator analyzerDelivered energy into 50 ohms, charge time, synchronization, pacing output
ESU analyzerOutput power into a load, leakage, return electrode monitoring
Gas flow and ventilator analyzerVolume, flow, pressure, oxygen concentration
Infusion pump analyzerFlow rate accuracy, occlusion pressure
DVM or DMM and metersVoltage, current, resistance; oxygen analyzer, pressure meter, tachometer, light meter, sound level meter
Cable tracer or tonerFinds the far end of a network cable
What each analyzer does not doVentilator analyzer: flow, pressure, volume, O2, breath detection, no CO2 concentration. ESU analyzer: RMS power, crest factor, peak-to-peak voltage, RF leakage, REM thresholds, no handpiece continuity or chassis leakage (multimeter and safety analyzer). Defibrillator analyzer: energy, charge time, sync, not internal impedance. Simulator: no current. DVM: voltage only, draws a little current; DMM continuity = low resistance, unpowered
UnitConversion or meaning
mmHgBlood and gas pressure; 760 mmHg = 1 atmosphere; 1 mmHg = 1.36 cm H2O
cm H2OVentilator and airway pressures; 1 cm H2O = 0.74 mmHg
psiGas cylinders and pipelines; 14.7 psi = 1 atmosphere; 1 psi = 51.7 mmHg; 1 psi = 6.9 kPa
bar1 bar = 100 kPa = 14.5 psi = 750 mmHg, about 1 atmosphere
kPa1 kPa = 7.5 mmHg; 101.3 kPa = 1 atmosphere
Fahrenheit and CelsiusF = C x 1.8 + 32; C = (F minus 32) / 1.8; 37 C = 98.6 F; 0 C = 32 F; 100 C = 212 F
JoulesEnergy: 1 J = 1 watt-second; defibrillator output
Lumens and luxLumens: total light output of a lamp; lux: lumens per square meter at the surface (surgical lights are rated in lux at 1 m)
DecibelsSound level and audiometer hearing threshold (dB HL); 10 dB is a tenfold power ratio
Weight and length1 kg = 2.2 lb; 1 inch = 2.54 cm; 1 L = 1000 mL; 1 mL = 1 cc
Prefixeskilo 10 to the 3, milli 10 to the minus 3, micro 10 to the minus 6, nano 10 to the minus 9

08Diagnostic, infusion, therapeutic, and laboratory equipment

Diagnostic deviceFacts
Otoscope, ophthalmoscopeBattery or wall-powered light and optics for ear and eye; check lamp, battery, charging contacts, specula
AudiometerHearing thresholds in dB HL at 250 to 8000 Hz; calibrate with a sound level meter and coupler; quiet room or booth
SpirometerFVC, FEV1, FEV1/FVC ratio (normal about 0.7 to 0.8); calibrate with a 3 L syringe; filters and mouthpieces
Stress test systemTreadmill or bike with 12-lead ECG and NIBP; emergency stop; treadmill speed and grade calibration; defibrillator nearby
ScalesCalibrate with certified test weights; zero with the bed or chair empty; load cells
UltrasoundPiezoelectric transducer sends and receives; 2 to 15 MHz; higher frequency, better resolution, less penetration; Doppler measures flow; gel removes air; transducer crystals crack if dropped
Otoscope partsHandle, head, lamp with collar, magnifying lens, specula, insufflation port; no microphone; loose head cuts light off, worn fibers dim it
OphthalmoscopeViews the fundus; lens disc, filters, mirror, condensing lens; clean with lens wipes, alcohol discolors
AudiogramThreshold in dB versus frequency; air conduction (headphones) versus bone conduction (mastoid vibrator) separates sensorineural from conductive loss; creates tones, does not measure frequencies
Uroflow meterFlow rate and volume of voiding (bladder emptying); funnel, beaker, transducer; wireless within about 3 m
Tympanic thermometerInfrared from the eardrum onto a thermopile; probe covers; 30 minutes to settle; ambient 10 to 40 C; torn cover reads high, cracked lens or obstruction reads low
EndoscopeFiber-optic or camera, cool light source, insufflation, irrigation, suction; leak test before immersion; no wireless in flexible scopes
ImagingX-ray: electrons from a tungsten filament strike a tungsten anode; radiopaque bone is white. Fluoroscopy: live X-ray through an image intensifier. CT: rotating tube and detectors, computed slices. MRI: magnet aligns hydrogen, RF, realignment signal, no ionizing radiation, pacemakers contraindicated. PET: FDG tracer, positron annihilation, two gamma photons. Ultrasound Doppler: toward the probe = higher pitch
Infusion deviceFacts
Large-volume pumpPeristaltic; accuracy typically within 5%; rate in mL/hr; free-flow protection; occlusion and air-in-line alarms; drug library
Syringe pumpPrecise small volumes; syringe size must be recognized; start-up delay and siphoning risks
PCA pumpPatient button, bolus dose, lockout interval, hourly limit; locked to prevent tampering
Feeding (enteral) pumpEnteral-only connectors on purpose; lower accuracy acceptable
Contrast injectorProgrammed volume and flow for imaging; pressure limit; air detection; heated syringe
Rate mathmL/hr = total mL / hours; drops per minute = (mL/hr x drop factor) / 60; percent error = (delivered minus set) / set x 100
Enteral versus parenteralTPN by vein through a standard pump; enteral by rotary peristaltic head on wide tubing; low flow = empty bag, occlusion, drip chamber seated wrong or coated; no set = seating, wrong set, or bright light on the detector
Syringe pump settingsMust know the syringe diameter and brand; small volume, fine control; runs past empty = sensor, size setting, or multi-phase program, not low voltage
PCA settingsBolus, lockout interval, maximum dose per hour or 4 hours, optional background; logged; locked; ignores requests = check settings first
Infusion modesContinuous; intermittent (high-rate periods separated by low-rate intervals); bolus; secondary or piggyback; pressure infusers squeeze a bag and are not pumps
Therapeutic deviceFacts
Infant incubator and warmerAir and skin temperature servo control, humidity, oxygen; radiant warmer with skin probe; over-temperature alarms; phototherapy blue light about 425 to 475 nm
Patient temperature managementHypo- and hyperthermia units circulate water through blankets or pads; water temperature limits about 4 to 42 C; check flow, leaks, temperature sensor
Aspiration (suction)Regulators in mmHg, continuous or intermittent; adult range about 80 to 120 mmHg; check filter, collection canister, tubing, vacuum source
Sequential compression deviceInflates leg sleeves in sequence to prevent clots; check pressures and cycle timing, hoses, sleeves
Physical therapy equipmentTENS and muscle stimulators, therapeutic ultrasound (1 to 3 MHz), diathermy, traction, continuous passive motion; output and timer verification
Radiant warmer versus incubatorRadiant: open bed, overhead heater, skin sensor servo, easy access, drafts and dehydration. Incubator: enclosed, air, humidity, oxygen control. Fluctuating temperature = sensor attachment
Ultrasound therapy1 to 3 MHz deep heat for pain and blood flow; needs gel and motion; head too cold is not detected; overheating shutdown = too little gel; slow warming = low drive voltage
Hypo/hyperthermia flowHeating needs flow of about 1 L/min or more, correct limits, and a good element; poor heating and cooling with no leaks = insufficient flow
Aspirator partsVacuum pump, gauge or regulator, tubing, collection container with overflow float, hydrophobic bacterial filter; pulmonary, gastric, surgical; fading suction = leak or filter; sudden stop = fuse or overheated pump
SCD and bilirubinSCD inflates distal (ankle) to proximal (thigh), holds seconds, rests about a minute; high pressure = sleeve too tight. Bilirubin light 360 to 520 nm, blue-green most effective, brighter, closer, more skin; bulb, ballast, fan filter
Laboratory deviceFacts
CentrifugeSpeed in RPM, force in RCF (g); balance opposite loads; lid interlock; verify speed with a tachometer and timer
IncubatorTemperature (often 37 C) and sometimes CO2 and humidity; verify with a reference thermometer; door seals
Rocker and shakerSpeed and tilt; motor and belt
Refrigerator and freezerBlood and vaccine storage 2 to 8 C; freezers minus 20 or minus 80 C; continuous temperature logging and alarms; door gaskets and defrost
MicroscopeLight source, objectives, focus; cleaning optics
Water bathGentle heating under 100 C with stirring; temperature control and uniformity via thermostat and heater; incubates microbial samples; not heating = setpoint or limit below ambient, or low water
AnalyzersReagents, calibration, quality control samples, fluidics and clogs, temperature
Cryostat and microtomeCryostat cuts frozen sections in a cold chamber (about minus 20 C); microtome cuts thin sections; blade safety, temperature, advance mechanism
Blood gas analyzerClark electrode pO2 (platinum cathode, silver anode, O2 membrane, current); glass electrode pH (voltage); Severinghaus electrode pCO2 (pH electrode behind a CO2 membrane in bicarbonate); too cold = thermistor; barcode = window
SpectrophotometerAbsorbance at selected wavelengths; tungsten lamp for visible, deuterium for UV; read a blank first; quartz cuvettes for UV; intermittent at all wavelengths = power cable
Hematocrit centrifugePacks red cells in capillary tubes at about 10,000 to 15,000 g; RCF rises with speed squared and radius
Microtome detailParaffin-embedded blocks; thickness depends on blade clamping, sharpness, clearance angle, cool block, slow steady stroke, no wax; cryostat frost = drafts or breath, splinter = too cold

09Perioperative and life-support devices

Electrosurgery factValue
FrequencyAbout 300 kHz to 3 MHz, above the frequency that stimulates nerves and muscle
ModesCut: continuous sine wave, vaporizes. Coag: interrupted or damped bursts, dries and seals. Blend mixes them
MonopolarCurrent flows from the active electrode through the patient to a large return (dispersive) electrode; return electrode contact quality monitoring prevents burns
BipolarCurrent flows between the two tips of the forceps; no return pad needed
BurnsPoor return pad contact, pad on bony or hairy site, alternate paths through ECG electrodes or metal; capacitive coupling in laparoscopy
Video integrationCameras, light sources, monitors, recording and routing systems; check cable, connector, light cable fibers, white balance
Waveform and faultsCut = continuous sine, 100% duty cycle, monopolar; coag = pulsed or damped, low duty cycle, higher peak; on but no output = blown high-voltage fuse; will not turn on = cord, fuse, low-voltage supply, switch; footswitch dead = footswitch board then control board

Other perioperative equipment

  • Pneumatic tourniquet: cuff pressure above systolic (commonly 250 mmHg arm, 300 mmHg leg or adjusted to the patient), time limit alarms (about 2 hours), leak test, calibrated gauge
  • Fluid and blood warmer: keeps fluids near body temperature; over-temperature cutoff around 41 to 42 C; check heater, sensor, alarms
  • Sterilizer: steam 121 C at 15 psi or 132 C flash; verify temperature, pressure, time, gasket, chamber; biological indicator; ethylene oxide and hydrogen peroxide for heat-sensitive items
  • OR table: hydraulics or motors, controls, brakes, weight rating; surgical lights: lux at 1 m, color temperature, handles, bulbs or LEDs; surgical microscope: optics, illumination, balance, focus motors
  • Sterilization methods: steam (fastest; dry heat for items that cannot be wet), ethylene oxide gas (heat- and moisture-sensitive, about 10 h plus aeration), liquid chemical (small hard-to-reach lumens, rinse), ionizing gamma or beta (disposables in sealed packages); no steam = water pump, low fluid pressure, breaker, heater
  • Surgical lasers: CO2 far infrared absorbed by water for skin and soft tissue; Nd:YAG near infrared penetrates (hair, tattoo, coagulation); argon blue-green for pigment and eye; excimer UV for corneas; aiming beam and articulated arm; aiming only = footswitch
  • Blood warmer: countercurrent warm water around the blood line; pressurized-flow units add a compressor (loud = unseated tubing or failing compressor; hot = blocked vents); prevents transfusion hypothermia, not for lab samples or infants
  • Heart-lung machine: venous reservoir, roller or centrifugal pump (gentle on cells), membrane oxygenator (older bubble), heat exchanger for hypothermia, filters, arterial return; heparin thins blood; cardioplegia stops the heart; backup pumps
  • Hemodialysis order: access, arterial monitor, blood pump, heparin pump, dialyzer (countercurrent dialysate across a semipermeable membrane), venous monitor, air trap and detector (last), return
Defibrillation and pacingValue
Adult energyMonophasic 360 J; biphasic 120 to 200 J (device specific); pediatric energies lower and weight based
Synchronized cardioversionShock timed to the R wave for an organized rhythm the clinician chooses to cardiovert; unsynchronized for VF and pulseless VT
TestingDefibrillator analyzer with a 50 ohm load; check delivered energy, charge time, synchronization, pacing output; daily user checks
AEDAnalyzes rhythm, advises shock only for VF and pulseless VT
External pacingDemand mode paces only when intrinsic rate falls below the set rate; fixed mode paces regardless; output in mA; capture confirmed by QRS after each spike
Intra-aortic balloon pumpBalloon in the descending aorta inflates in diastole (raises coronary perfusion) and deflates just before systole (lowers afterload); timed to the ECG or arterial waveform; helium gas; alarms for timing, leak, and gas loss
Pacing routesTranscutaneous: chest pads with gel, tens of mA, painful, fastest. Transvenous: lead through a vein (subclavian) into the right ventricle, low output. Epicardial: wires on the heart surface at surgery, no subclavian leads. External generators run on AC or battery
Ventilator and anesthesiaValue
Tidal volume and minute volume6 to 8 mL/kg ideal body weight; minute volume = tidal volume x rate
Typical settingsRate 12 to 20; I:E ratio 1:2; PEEP 5 cm H2O; FiO2 0.21 to 1.0
ModesVolume control delivers a set volume; pressure control a set pressure; assist-control, SIMV, pressure support, CPAP; BiPAP has separate inspiratory and expiratory pressures
AlarmsHigh pressure: obstruction, kink, coughing, secretions. Low pressure or low volume: disconnect or leak. Apnea: no breath detected
Oxygen analyzer calibrationRoom air 21% and 100% oxygen; galvanic cells wear out; concentrators deliver about 90 to 95%
Anesthesia machinePipeline (50 psi) and cylinder supply, flowmeters, agent-specific vaporizers (sevoflurane yellow, isoflurane purple, desflurane blue), hypoxic guard keeps oxygen at least 21 to 25%, oxygen flush, CO2 absorber, scavenging removes waste gas, pin index and DISS prevent misconnection; leak test before use
Anesthesia gas pathSupply (pipeline 50 psi, cylinders), regulators, flow controls and flowmeters, vaporizer (first section after the agent's flow control), common gas outlet, breathing circuit with CO2 absorber, ventilator, patient; scavenging removes waste

10Problem solving and troubleshooting

The method, in order

  • 1. Confirm the complaint: talk to the user, reproduce the problem, rule out use error (wrong settings, wrong accessory, wrong technique), and check the right disposables.
  • 2. Check the simple things: power, plug, battery, fuse, cables, connectors, settings, sensors.
  • 3. Isolate: which subsystem (power, sensor, signal path, processor, display, mechanical); half-split the path with a simulator and test equipment; localized device or system-wide?
  • 4. Form and test one hypothesis at a time; swap known-good parts; use the schematic.
  • 5. Repair or replace; follow the service manual.
  • 6. Verify: full performance test and electrical safety test after any repair, before return to service.
  • 7. Document in the CMMS: symptom, cause, action, parts, test results, time. Educate the user if it was a use error.
SymptomFirst suspects
Device dead, no powerOutlet, cord, plug, fuse, power switch, battery, power supply output
Runs on AC but not batteryBattery aged or not charging; charger circuit; battery connector
Intermittent reboot or errorsFailing power supply capacitors, loose connector, overheating
ECG shows 60 Hz noiseElectrode contact, gel, skin prep, lead cable shield, nearby equipment, ground
Wandering baselinePatient movement or respiration, loose electrode, high-pass filter setting
Intermittent signalCable flex failure near connectors, loose connector, cracked solder
NIBP reads highCuff too small, cuff too low, patient movement, hose leak causing slow deflation
Arterial line reads low and roundedOverdamped: air bubble, clot, kinked tubing, loose connection
SpO2 no readingProbe placement, perfusion, ambient light, probe LED failure, cable
Ventilator high-pressure alarmKinked circuit, secretions, patient coughing, water in tubing
Ventilator low-pressure or volume alarmDisconnect, leak, cuff leak, circuit crack
Infusion pump occlusion alarmKinked tubing, closed clamp, clotted line, set loaded wrong
Infusion pump air-in-line alarmAir in set, sensor dirty, set not seated
Defibrillator low delivered energyCapacitor aging, relay, paddles or pads, battery
ESU patient burnReturn electrode contact, alternate path, damaged cable
Telemetry dropouts on one patientTransmitter battery, leads, distance; many patients: antenna system or receiver, a system-wide problem
Monitor not sending to the central station or EMROne bed: cable, port, network config on that monitor. Every bed: switch, server, middleware, network outage
Centrifuge vibrationUnbalanced load, worn rotor or bearing
Refrigerator temperature alarmDoor seal, defrost cycle, compressor, probe, overloading
EtCO2 no value or occlusionCannula and tubing, filter present, moisture trap (occlusion test), internal leak (occlude upstream of filter and trap); occlusion: kink, clogged filter, full trap
Respiration rate missing, ECG fineInput printed circuit board; flat or short trace: raise sensitivity, then electrodes, leads, cable
Blood gas analyzer will not runCable and outlet first; 'too cold' after warming = thermistor; no battery run = charger; barcode dead = window
Centrifuge rotor will not spinLid not closed firmly (interlock), then connections, board, motor
Microtome sections wrong thicknessBlade clamp, dull edge, clearance angle; compressed = dull, warm block, too fast, wax
Tympanic thermometer reads lowObstruction or cracked lens (high = torn cover); rest 30 minutes; ambient 10 to 40 C
Enteral pump low flow versus no setLow flow: empty bag, occlusion, drip chamber. No set: seating, wrong set, bright light
PCA ignores requestsSettings (lockout, maximum) before pendant or motor; 'check settings' = door locked before setup
Aspirator suction fadesLeak (tubing, container, vacuum line) or clogging filter; sudden stop = fuse or overheated pump; none = disconnected, blocked filter, full container
ESU on but no outputHigh-voltage supply fuse
Tourniquet pump runs constantlyLeak in tubing, cuff, or valves, or a bad pressure transducer; will not inflate = pump disconnected
Autoclave no steamWater pump, low fluid pressure, breaker or overload, burned-out heater, low water; over-pressure = pressure switch
Bilirubin light dark or dimToo hot to relight; bulb then ballast; dirty lens; overheating = fan filter, fans, thermistor
Device went to battery while plugged inOutlet and power cable continuity first, then the internal power circuit; regulator: measure input and output pins to ground

Priorities, communication, maintenance

  • Priority order: life support and high-risk devices first, then patient risk, then urgency and downtime and clinical impact; loaners for critical downtime
  • Device error versus use error: a device that performs to specification but was used wrongly needs education and possibly a design or labeling report, not a repair
  • In-service and cross-training: teach proper use, care, cleaning, and alarm response; document attendance; communicate clearly with clinicians, staff, and manufacturers in writing and verbally
  • Preventive maintenance: scheduled or risk-based; AEM programs allowed with documented justification (not imaging, lasers, or where the manufacturer requires); PM after repair; performance verification against tolerances
  • Calibration traceable to national standards; test equipment has its own calibration schedule
  • Incoming inspection before first use: physical, functional, safety test, inventory entry, manuals on file; lifecycle ends with data wiped, decontamination, and disposition
  • Criticality = importance to the mission + likelihood of failure or PM demand + potential harm; repair time is not a factor; redundancy lowers priority but does not prevent all downtime; checklists guide user inspections
  • Root cause analysis: what happened, why, how to prevent; keep asking why (five whys); ultrasound blurry = crystal then gel; oximeter erratic = ambient light; one sphygmomanometer = technique then cracked tubing (a leak reads low)
  • Local versus network: one room missing = that room; several = shared configuration or network; false alarms on many monitors = alarm limits; check physical links first; search repair records for fleet patterns; ping, cable tester, multimeter help, a safety analyzer does not
  • Switching supply shorted element = full input voltage out, open = zero; UPS short runtime = battery, no backup = inverter, switched over while plugged in = outlet or cable; desktop PSUs die of fan failure and heat

11Healthcare information technology

Networking factValue
IPv4 addressFour numbers 0 to 255; subnet mask separates network and host; default gateway is the exit router; same subnet required to talk without a router
MAC address48-bit hardware address in 12 hex digits, burned into the network card; switches forward by MAC
Private ranges10.0.0.0/8; 172.16.0.0 to 172.31.255.255; 192.168.0.0/16
DHCP and staticDHCP assigns addresses automatically; medical devices often use static addresses or DHCP reservations so they are always reachable
DNSNames to addresses; ping by address works but by name fails means DNS
APIPA169.254.x.x means DHCP failed
PortsHTTP 80, HTTPS 443, SSH 22, DICOM commonly 104 or 11112, HL7 commonly 2575 (site specific)
DevicesSwitch connects devices in a LAN by MAC; router connects networks by IP; access point connects Wi-Fi clients; access point controller manages many APs; KVM switch shares one keyboard, video, and mouse among servers; virtual servers run many operating systems on one host; LAN is local, WAN spans sites, cloud is provider-hosted
CablingCat 5e 1 Gbps, Cat 6A 10 Gbps, 100 m limit; fiber for longer runs; link light off means Layer 1; cable tracer finds the far end
Wi-Fi2.4 GHz longer range, channels 1, 6, 11; 5 GHz more channels, less interference; WPA2 or WPA3; enterprise mode uses 802.1X with RADIUS
Telemetry and RTLSWMTS bands for patient telemetry; real-time location systems track equipment with RF, infrared, or ultrasound tags
OSI layers1 physical (media, hubs), 2 data link (frames, MAC, switches), 3 network (IP, routers), 4 transport (TCP/UDP, ports), 5 session, 6 presentation (format conversion, encryption), 7 application (APIs); no conversion layer
TopologiesBus and ring fail on any break; star (central switch) survives one bad cable; hybrid; mesh partial or full; point-to-multipoint (central node also processes); physical versus logical
Binary masks255.255.255.240 = ...11110000 (/28, 16 addresses); 248 = 11111000; 252 = 11111100; ones must be contiguous; subnetting separates departments and limits broadcasts
Hub versus switch, PPPHub repeats to every port at layer 1, half duplex, collisions; switch keeps a source address table and forwards at layer 2, full duplex; PPP links two routers over serial or dial with authentication
Command or toolUse
pingIs the address reachable, and how fast; ping the gateway first
ipconfig (Windows), ifconfig or ip (Linux)Show the device's address, mask, gateway; release and renew DHCP
tracert or tracerouteEvery hop to a destination; where the path breaks
nslookupTest name resolution
arp -aIP to MAC table; duplicates
netstatOpen connections and listening ports
Cable tracer and testerFind the far end; check pinout and continuity
TDRTime-domain reflectometer: pulse reflection shows whether a cable is open and how far to the break
ping (ICMP)Echo request and reply, IPv4 and IPv6; reports round-trip ms, packet loss, min, max, average, deviation; not a cable resistance test; power cycling is not a test
Fiber inspection microscopeChecks connector ends for dirt and damage; clean with isopropyl alcohol and lint-free wipes or cut and repolish; light source and power meter measure loss
InteroperabilityMeaning
HL7 version 2Text messages between systems: ADT (admit, discharge, transfer), ORM (orders), ORU (results, including device data)
HL7 FHIRModern web-based standard using resources and APIs
DICOMStandard for medical images and their transfer; modality worklist pulls patient data to the device; PACS stores and serves images
EMR and EHRElectronic medical record (within one organization) and electronic health record (across); devices feed vitals in directly or through middleware
MiddlewareTranslates and routes between devices and the EMR; device gateways; where most integration faults are diagnosed
Time synchronizationNTP keeps device clocks aligned so records and alarms line up
PACS, RIS, LIS, LIMSPACS stores and transports images with patient data (redundant servers); RIS schedules, tracks, and reports radiology; LIS tracks specimens and analyzer results; LIMS is the research and high-volume version
DICOM and HL7 detailDICOM is a 22-part standard for image format and transfer referencing IEC definitions; HL7 standards: version 2 messaging, version 3 XML, Clinical Document Architecture; HL7 does not set leakage limits or breach rules

Protected data and cybersecurity

  • HIPAA: privacy rule (who may see PHI), security rule (administrative, physical, technical safeguards), breach notification; PHI is any identifiable health information; minimum necessary
  • NIST Cybersecurity Framework functions: govern, identify, protect, detect, respond, recover
  • HITRUST: a certifiable framework combining many requirements; HHS cybersecurity performance goals: essential and enhanced practices for hospitals (MFA, patching, backups, segmentation, training)
  • MDS2: manufacturer disclosure statement for medical device security; SBOM: software bill of materials
  • Controls: unique accounts and MFA, least privilege, encryption at rest and in transit, patching on the vendor's validated schedule, network segmentation for devices that cannot be patched, antivirus where the vendor allows, physical security of workstations, screen locks
  • Backups: 3 copies, 2 media, 1 offsite; test restores; ransomware response: isolate, report, restore
  • Incident: disconnect the device from the network, do not power off if forensics are needed, report to IT security and the vendor, document
  • HITECH (2009): EHR incentives and penalties, meaningful use, breach notification to individuals and HHS, business associates bound, no sale of PHI, patient access; sets no spending guidelines
  • HIPAA titles: I insurance portability; II administrative simplification with the privacy and security rules, fraud control, and fines; III medical savings accounts; IV enforcement by the Office for Civil Rights; V revenue offsets
  • DMCA (1998): bans circumvention tools, raises penalties, limits provider liability; exemptions for security testing and personal implant data; does not ban copying devices as such
  • FDA MDDS: systems that store, transfer, reformat, or display device data without controlling devices; reclassified Class III to Class I in 2011; a vital signs monitor yes, a lobby TV no
  • IEC 80001-1 (2010, IEC and ISO): risk management for IT networks with medical devices; excludes drugs; applies to any organization's network; defines no acceptable risk level; not from the FDA
  • MDS2 contents: automatic logoff, audit controls, authorization, configurable security, security updates, de-identification, backup and recovery, emergency (break-glass) access, data integrity, malware protection, node and user authentication, physical locks, third-party component plans, remote servicing; not structural integrity

Computer hardware and software troubleshooting

  • Hardware: power supply (no power, random reboots), hard drive (slow, errors, clicking; SMART warnings; replace and restore image), memory (crashes), peripherals (drivers, cables, ports), connectors and cables (reseat, swap)
  • Software: Windows, Office, Linux, and Unix updates and patches only on the vendor's validated schedule for medical devices; drivers after updates; event logs; restore from image; user accounts and permissions
  • Order of checks: power and cables, link lights, address and gateway, ping the gateway, ping the server, name resolution, application or service, then escalate to IT or the vendor

12Acronyms

Every acronym used in this course, 288 in all.

AcronymStands forMeaning
802.11IEEE 802.11The family of Wi-Fi standards.
802.1XIEEE 802.1XPort-based network access control: a device must authenticate before the switch port or Wi-Fi passes traffic.
AAAA cellA common single-cell battery size, about 1.5 V.
AAMIAssociation for the Advancement of Medical InstrumentationThe standards body for medical equipment and the sponsor of the CBET credential.
ABCClass A, B, and C fire extinguisherA multipurpose extinguisher for ordinary combustibles, flammable liquids, and electrical fires.
ACAlternating currentMains power that reverses direction, 60 Hz in the US and 50 Hz elsewhere.
ACGIHAmerican Conference of Governmental Industrial HygienistsThe organization that publishes threshold limit values.
ACHCAccreditation Commission for Health CareAn accrediting body that absorbed the HFAP hospital program.
ACIAAMI Credentials InstituteThe body within AAMI that administers the CBET and related certifications.
ACKAcknowledgment (HL7 message)The reply an HL7 interface sends to confirm it received a message, carrying an accept or error code.
ADCAnalog-to-digital converterTurns a continuous voltage into digital samples.
ADTAdmit, discharge, transferThe HL7 messages that tell every system who is in which bed.
AEApplication entityA DICOM node, identified by its AE title, IP address, and port.
AEDAutomated external defibrillatorA defibrillator that analyzes the rhythm and advises a shock only for VF or pulseless VT.
AEMAlternative equipment maintenanceA CMS-allowed program that adjusts maintenance from manufacturer recommendations based on risk and history.
AFAtrial fibrillationAn irregular rhythm with no P waves.
AHAAmerican Heart AssociationThe organization behind CPR guidelines and the US ECG electrode color code.
ALARAAs low as reasonably achievableThe radiation safety principle: minimize dose with time, distance, and shielding.
ANSIAmerican National Standards InstituteThe US body that accredits standards developers such as AAMI.
APAccess pointThe Wi-Fi radio that bridges wireless clients onto the wired network.
APIApplication programming interfaceThe defined way one program talks to another, such as the web endpoints a cloud service exposes.
APIPAAutomatic Private IP AddressingThe 169.254.x.x address a host gives itself when it cannot reach a DHCP server.
ASMEAmerican Society of Mechanical EngineersThe standards body for pressure vessels and boilers.
AVAtrioventricular nodeThe relay between atria and ventricles that delays the impulse so the ventricles fill.
AWGAmerican Wire GaugeThe wire size scale; smaller numbers are thicker wire with higher ampacity.
BAABusiness associate agreementThe HIPAA contract a vendor signs before handling protected health information.
BFBody floating (applied part)An applied part that touches the patient but not the heart, isolated from ground; stick-figure-in-a-box symbol.
BIOSBasic Input/Output SystemThe original PC firmware: 16-bit, text setup, MBR disks up to 2 TB.
BiPAPBilevel positive airway pressureTwo pressure levels, higher on inhalation and lower on exhalation, for non-invasive ventilation.
BPBlood pressureThe pressure in the arteries, systolic over diastolic, in mmHg.
C-ECHODICOM echo serviceThe DICOM ping that verifies two application entities can communicate.
CAPCollege of American PathologistsThe organization that accredits clinical laboratories.
CBETCertified Biomedical Equipment TechnicianThe AAMI credential for biomedical equipment technicians.
CDAClinical Document ArchitectureHL7's XML standard for clinical documents such as discharge summaries.
CDCCenters for Disease Control and PreventionThe US public health agency that sets infection control guidance such as standard precautions.
CFCardiac floating (applied part)An applied part that may connect directly to the heart, with the strictest leakage limits; heart-in-a-box symbol.
CFRCode of Federal RegulationsThe codified US federal rules; 29 CFR 1910 holds OSHA standards, 21 CFR the FDA's.
CHIPChildren's Health Insurance ProgramFederal and state health coverage for children, run by CMS.
CIAConfidentiality, integrity, and availabilityThe three goals of security: keep data secret, keep it correct, keep it reachable.
CIHQCenter for Improvement in Healthcare QualityA newer CMS-approved hospital accreditor.
CMMSComputerized maintenance management systemThe database that tracks equipment inventory, work orders, and preventive maintenance.
CMOSComplementary metal-oxide semiconductorThe small battery-backed memory that keeps firmware settings and the clock.
CMRRCommon-mode rejection ratioHow well an amplifier cancels signals present on both inputs, such as 60 Hz interference.
CMSCenters for Medicare and Medicaid ServicesThe federal agency whose Conditions of Participation hospitals must meet to be paid.
COCarbon monoxideA toxic gas that binds hemoglobin and makes pulse oximeters read falsely high.
CO2Carbon dioxideThe gas exhaled as the waste product of metabolism; measured by capnography, and the lasing medium of CO2 surgical lasers.
COPDChronic obstructive pulmonary diseaseLong-term airway obstruction, usually from smoking, with a low FEV1/FVC ratio.
CPAPContinuous positive airway pressureA constant pressure applied to the airway of a spontaneously breathing patient.
CPRCardiopulmonary resuscitationChest compressions and breaths to keep blood moving during cardiac arrest.
CPUCentral processing unitThe processor that executes instructions.
CTComputed tomographyCross-sectional X-ray imaging from a rotating tube and detector array.
DACDigital-to-analog converterTurns digital values back into a voltage to drive displays, speakers, or outputs.
dBDecibelA logarithmic ratio unit for signal levels, gain, and sound.
DBPDiastolic blood pressureThe arterial pressure between beats, while the heart fills.
DCDirect currentCurrent in one direction: batteries and power supply outputs.
DHCPDynamic Host Configuration ProtocolHands out IP addresses, masks, gateways, and DNS servers automatically (UDP 67 and 68).
DICOMDigital Imaging and Communications in MedicineThe standard for medical image files and their network exchange, by convention on port 104.
DISSDiameter Index Safety SystemGas-specific threaded fittings that make it impossible to connect the wrong gas hose.
dLDeciliterOne tenth of a liter; lab values such as glucose in mg per dL.
DMCADigital Millennium Copyright ActThe 1998 copyright law that bans circumventing digital protections.
DMMDigital multimeterThe handheld meter for voltage, current, resistance, and continuity.
DNADeoxyribonucleic acidThe genetic material; sterilization kills microbes by destroying it.
DNSDomain Name SystemTranslates names such as www.example.com into IP addresses (UDP 53, TCP 53 for zone transfers).
DNVDNV HealthcareAn accrediting organization that surveys hospitals annually under its NIAHO program.
DOTDepartment of TransportationThe US agency whose rules govern compressed gas cylinders.
DVMDigital voltmeterA meter for voltage; in practice the same instrument as a DMM.
ECGElectrocardiogramThe recording of the heart's electrical activity: P wave, QRS complex, T wave.
ECRIECRI InstituteThe independent organization that tests devices and publishes hazard alerts and the annual top-ten hazards list.
EDIElectronic data interchangeStandardized electronic exchange of business documents such as claims and payments.
EEGElectroencephalogramThe recording of the brain's electrical activity from scalp electrodes, in microvolts.
EHRElectronic health recordThe patient's longitudinal digital record, shared across providers.
EMElectromagneticElectric and magnetic fields; EM interference disrupts monitors and telemetry.
EMGElectromyogramThe recording of muscle electrical activity.
EMIElectromagnetic interferenceNoise from motors, power lines, fluorescent lights, and radios that corrupts signals on copper cable.
EMRElectronic medical recordThe hospital's clinical record system: orders, notes, results, medications.
EQAAMI EQ seriesAAMI's equipment management standards: EQ56, EQ89, EQ93, EQ103.
EQ103ANSI/AAMI EQ103Guidance for alternative equipment maintenance programs.
EQ56ANSI/AAMI EQ56The recommended practice for a medical equipment management program.
EQ89ANSI/AAMI EQ89Guidance for scheduled maintenance and performance verification.
EQ93ANSI/AAMI EQ93The vocabulary of medical equipment management.
ESDElectrostatic dischargeStatic electricity that can destroy chips; prevented with a wrist strap, mat, and antistatic bags.
ESUElectrosurgical unitThe generator that cuts and coagulates tissue with high-frequency current above about 100 kHz.
EtCO2End-tidal carbon dioxideThe CO2 concentration at the end of exhalation, normally 35 to 45 mmHg, measured by capnography.
ETOEthylene oxideA toxic gas used to sterilize heat-sensitive devices.
FDAFood and Drug AdministrationThe US regulator of medical devices; classes I to III, recalls, and medical device reporting.
FDGFluorodeoxyglucoseThe glucose-like radioactive tracer used in PET scans.
FEV1Forced expiratory volume in one secondThe volume exhaled in the first second of a forced breath out.
FHIRFast Healthcare Interoperability ResourcesHL7's web-style standard using REST APIs and JSON resources.
FHRFetal heart rateThe baby's heart rate, measured by ultrasound Doppler on a fetal monitor.
FiO2Fraction of inspired oxygenThe oxygen concentration being delivered, from 0.21 (room air) to 1.0.
FSHFollicle-stimulating hormoneThe pituitary hormone that drives sperm production and egg development.
FVCForced vital capacityThe total volume a patient can exhale forcefully after a full breath in.
GbpsGigabits per secondBillions of bits per second; gigabit Ethernet is 1 Gbps.
GFCIGround-fault circuit interrupterA receptacle or breaker that trips when current leaking to ground exceeds about 5 mA.
GHSGlobally Harmonized SystemThe international standard for chemical hazard labels and safety data sheets.
GHzGigahertzBillions of cycles per second; CPU clock speed and Wi-Fi band frequency.
GIGastrointestinalThe digestive tract.
H2OWaterThe pressure unit centimeters of water (cmH2O) for airway and ventilator pressures.
HDDHard disk driveSpinning magnetic storage: cheap per terabyte, slower, and fragile.
HFAPHealthcare Facilities Accreditation ProgramThe former osteopathic hospital accreditation program, now part of ACHC.
HHSDepartment of Health and Human ServicesThe US department that enforces HIPAA through its Office for Civil Rights.
HIPAAHealth Insurance Portability and Accountability ActThe US law protecting health information; Title II holds the privacy and security rules.
HITECHHealth Information Technology for Economic and Clinical Health ActThe 2009 law that funded EHR adoption and strengthened HIPAA enforcement and breach notification.
HITRUSTHealth Information Trust AllianceA certifiable security framework that combines HIPAA, NIST, and other requirements for healthcare.
HIVHuman immunodeficiency virusA bloodborne virus; one of the pathogens the OSHA bloodborne pathogen standard addresses.
HLHearing levelThe decibel scale audiometers use, referenced to normal hearing thresholds.
HL7Health Level SevenThe messaging standard that connects clinical systems: version 2 pipes, version 3 XML, FHIR web APIs.
HRHeart rateBeats per minute, derived from the ECG or pulse.
HTTPHypertext Transfer ProtocolWeb traffic in the clear on TCP 80.
HTTPSHypertext Transfer Protocol SecureHTTP inside TLS on TCP 443.
HVHigh voltageThe kilovolt sections of X-ray generators, defibrillators, and ESUs.
HVACHeating, ventilation, and air conditioningThe building's climate systems, which keep server rooms cool and are themselves networked controls.
HzHertzCycles per second; display refresh rate, AC mains frequency, and signal frequency.
IaaSInfrastructure as a serviceThe provider supplies virtual machines, storage, and networks; you manage the operating system and up.
IABPIntra-aortic balloon pumpA balloon in the aorta that inflates in diastole and deflates in systole to support the heart.
IARCInternational Agency for Research on CancerThe WHO body that classifies carcinogens.
IBPInvasive blood pressurePressure measured directly through an arterial line and a transducer.
ICIntegrated circuitA chip: many transistors on one piece of silicon.
ICMPInternet Control Message ProtocolThe protocol behind ping and traceroute.
ICUIntensive care unitThe critical care area with the highest density of monitors, ventilators, and pumps.
IDIdentificationA patient identifier or a device identifier.
IECInternational Electrotechnical CommissionThe international standards body; IEC 60601-1 is the medical electrical equipment safety standard.
IEEEInstitute of Electrical and Electronics EngineersThe engineering body behind Ethernet (802.3), Wi-Fi (802.11), and many device standards.
IPInternet ProtocolThe Layer 3 protocol that addresses and routes packets between networks.
IPv4Internet Protocol version 432-bit addresses written as four decimal octets, such as 192.168.1.10.
IPv6Internet Protocol version 6128-bit addresses in eight hex groups; no broadcast, built-in autoconfiguration.
ISOInternational Organization for Standardization; also an ISO disc imageThe standards body, and the .iso file format that holds a complete disc image.
ITInformation technologyComputers, networks, and the people who run them.
IUPCIntrauterine pressure catheterA catheter that measures contraction pressure directly.
IVIntravenousInto a vein; IV lines, pumps, and fluids.
KCLKirchhoff's current lawCurrent into a node equals current out.
kHzKilohertzOne thousand cycles per second.
kVKilovoltOne thousand volts; X-ray tube voltage and defibrillator charge levels.
KVLKirchhoff's voltage lawThe voltages around any closed loop sum to zero.
KVMKeyboard, video, mouse (switch)One keyboard, monitor, and mouse shared among several computers.
KVOKeep vein openA very low infusion rate that keeps an IV line from clotting.
kWKilowattOne thousand watts of power.
kWhKilowatt-hourThe energy unit on the electric bill: one kilowatt for one hour.
LALeft arm electrode (also left atrium)The black ECG electrode on the left arm; in heart anatomy, the left atrium.
LANLocal area networkOne site's network: a building, floor, or campus that one organization owns and runs.
LC50Lethal concentration 50 percentThe airborne concentration lethal to half of test subjects.
LCDLiquid crystal displayThe flat panel that shapes light from a backlight into an image; TN, IPS, and VA are the panel types.
LD50Lethal dose 50 percentThe dose that kills half of test subjects.
LEDLight-emitting diodeA small light source: the status lights on a port, the lamp in cheap fiber transmitters, and modern display backlights.
LHLuteinizing hormoneThe pituitary hormone that drives testosterone and triggers ovulation.
LIMLine isolation monitorContinuously measures how much current a fault could drive from an isolated power system to ground and alarms at 5 mA.
LIMSLaboratory information management systemSample and workflow tracking software for laboratories.
LISLaboratory information systemThe software that manages lab orders, analyzer results, and reporting.
LLLeft leg electrodeThe red ECG electrode on the left leg.
LVLeft ventricleThe chamber that pumps blood to the body at systemic pressure.
LVPLarge-volume pumpAn infusion pump for bags of fluid, usually peristaltic, accurate to about 5 percent.
mAMilliampereOne thousandth of an ampere; the unit for shock thresholds and leakage currents.
MACMedia access control (address)The 48-bit hardware address burned into a network interface, used at Layer 2.
MAPMean arterial pressureThe average arterial pressure: diastolic plus one third of the pulse pressure.
MDDSMedical device data systemSoftware or hardware that transfers, stores, or displays device data without changing it.
MDS2Manufacturer Disclosure Statement for Medical Device SecurityThe standard form describing a device's security capabilities.
MFAMulti-factor authenticationTwo or more different kinds of proof of identity.
MHzMegahertzMillions of cycles per second; memory speed, radio channels, and older CPU clocks.
mLMilliliterOne thousandth of a liter; tidal volumes, drug doses, and infusion rates in mL per hour.
mmHgMillimeters of mercuryThe pressure unit for blood pressure, gas pressures, and vacuum.
MRMagnetic resonanceThe MRI environment; equipment is labeled MR Safe, MR Conditional, or MR Unsafe.
MRIMagnetic resonance imagingImaging with a strong magnet and radio pulses; no ionizing radiation, but the magnet is always on.
MRNMedical record numberThe hospital's unique identifier for a patient.
msMillisecondOne thousandth of a second; latency, pulse widths, and ECG timing.
MSHMessage header segmentThe first segment of every HL7 version 2 message.
mVMillivoltOne thousandth of a volt; ECG signals are about 1 mV.
NATNetwork address translationRewrites private inside addresses to a public address at the router so many hosts can share it.
NECNational Electrical CodeThe US wiring code, NFPA 70.
NESCNational Electrical Safety CodeThe utility-side safety code for power lines and substations.
NFPANational Fire Protection AssociationThe standards body for fire and electrical safety: NFPA 70 (NEC), NFPA 99 (health care facilities), NFPA 101 (life safety).
NIAHONational Integrated Accreditation for Healthcare OrganizationsDNV's hospital accreditation program.
NIBPNon-invasive blood pressureAutomatic cuff measurement by the oscillometric method.
NIOSHNational Institute for Occupational Safety and HealthThe CDC's research arm for workplace safety, which publishes RELs.
NISTNational Institute of Standards and TechnologyThe US standards body whose cybersecurity framework hospitals use.
NPNNegative-positive-negative transistorA bipolar transistor that conducts when its base is pulled positive.
NPSGNational Patient Safety GoalsThe Joint Commission's annual list of safety priorities, including alarm management.
NTCNegative temperature coefficient (thermistor)A thermistor whose resistance falls as temperature rises; the sensor in patient temperature probes.
NTPNational Toxicology ProgramThe US program that publishes the Report on Carcinogens, one of the carcinogen listings cited on a safety data sheet.
NTPNetwork Time ProtocolThe protocol (UDP 123) that keeps device clocks synchronized so logs, alarms, and records line up.
OBRObservation request segmentThe HL7 segment describing the order a result answers.
OBXObservation segmentThe HL7 segment holding one observation and its value.
OCROffice for Civil RightsThe HHS office that enforces HIPAA privacy and security rules.
OROperating roomThe surgical suite, with isolated power, ESUs, anesthesia machines, and lasers.
ORMOrder messageThe HL7 version 2 message that carries an order.
ORUObservation resultThe HL7 message that carries results, including vital signs from monitors.
OSHAOccupational Safety and Health AdministrationThe US agency that sets workplace safety rules.
OSIOpen Systems Interconnection modelThe seven-layer model for describing where a network function happens.
OXOxidizer (NFPA 704 symbol)The special hazard code in the white diamond for oxidizing materials.
PaaSPlatform as a serviceThe provider runs the platform (OS, runtime, database); you deploy code.
PACSPicture archiving and communication systemThe server that stores and distributes medical images using DICOM.
PASSPull, aim, squeeze, sweepHow to use a fire extinguisher.
PCPersonal computerA desktop or laptop for one user.
PCAPatient-controlled analgesiaAn infusion pump that lets the patient self-dose pain medication within programmed limits.
PCBPrinted circuit boardThe board that holds and connects a device's components.
PDCAPlan, do, check, actThe continuous improvement cycle.
PEEPPositive end-expiratory pressurePressure the ventilator keeps in the lungs at the end of exhalation so alveoli stay open.
PELPermissible exposure limitOSHA's legal maximum exposure to a substance, averaged over a shift.
PETPositron emission tomographyNuclear imaging of metabolic activity using a positron-emitting tracer such as FDG.
pHPotential of hydrogenThe acidity scale; blood is normally 7.35 to 7.45.
PHIProtected health informationAny health information tied to an identifiable person.
PIDPatient identification segmentThe HL7 segment carrying patient name, ID, and demographics.
PMPreventive maintenanceScheduled inspection and testing of equipment before it fails.
PNPPositive-negative-positive transistorA bipolar transistor that conducts when its base is pulled negative.
PoEPower over EthernetDelivers DC power to devices over the data cable: 802.3af 15.4 W, 802.3at 30 W, 802.3bt 60 to 90 W.
PPPulse pressureSystolic minus diastolic pressure.
PPEPersonal protective equipmentGloves, gowns, masks, eye protection, and lead aprons.
PPPPoint-to-Point ProtocolThe Layer 2 protocol for serial and dial-up links that carries IP with authentication.
PRPR intervalThe ECG interval from the start of the P wave to the start of the QRS, 0.12 to 0.20 seconds; the AV node delay.
psiPounds per square inchThe pressure unit for cylinders, sterilizers, and pneumatic systems.
PSUPower supply unitConverts wall AC to the DC voltages a computer needs: 12 V, 5 V, 3.3 V.
PSVPressure support ventilationThe ventilator boosts each spontaneous breath with a set pressure.
PTPhysical therapyThe rehabilitation department, whose equipment includes TENS units, stimulators, and therapeutic ultrasound.
QCQuality controlRunning known samples to verify an analyzer is reading correctly.
QRSQRS complexThe sharp ECG spike of ventricular depolarization; normally under 0.12 seconds wide.
QTQT intervalThe ECG interval from the start of the QRS to the end of the T wave, under about 0.44 seconds.
RARight arm electrode (also right atrium)The white ECG electrode on the right arm; in heart anatomy, the right atrium.
RACERescue, alarm, confine, extinguish or evacuateThe order of actions when a fire is discovered.
RADIUSRemote Authentication Dial-In User ServiceThe AAA server behind Wi-Fi Enterprise, VPNs, and 802.1X; UDP 1812 and 1813.
RAIDRedundant array of independent disksCombining disks for speed, capacity, or fault tolerance: 0 stripe, 1 mirror, 5 stripe with parity, 6 double parity, 10 mirrored stripes.
RAMRandom-access memoryThe working memory that holds running programs and data; lost at power off.
RCResistor-capacitor (circuit)A resistor and capacitor together, whose time constant is R times C.
RCARoot cause analysisA structured investigation of what happened, why, and how to prevent it.
RCFRelative centrifugal forceThe g-force a centrifuge applies, which depends on speed and rotor radius.
RELRecommended exposure limitNIOSH's recommended exposure limit.
REMRoentgen equivalent manA unit of radiation dose equivalent, 100 rem equals 1 sievert.
RFRadio frequencyRadio signals; the medium of Wi-Fi, cellular, and Bluetooth, and a source of interference and, in medicine, surgical energy.
RISRadiology information systemScheduling, tracking, and reporting for imaging departments.
RJ45Registered jack 45The eight-pin plug on Ethernet twisted-pair cables (properly an 8P8C connector).
RLRight leg electrodeThe green ECG electrode on the right leg, the reference or driven ground.
RMSRoot mean squareThe effective value of an AC voltage or current, the DC equivalent for heating; 120 V mains is RMS.
RPMRevolutions per minuteRotation speed: hard drive platters at 5400 or 7200, fans, and laboratory centrifuges.
RS-232Recommended Standard 232 serialThe classic serial port: slow, simple, and still used for console access and legacy devices.
RTLSReal-time location systemTags and receivers that track equipment location inside the hospital.
RVRight ventricleThe chamber that pumps blood to the lungs.
SASinoatrial nodeThe heart's natural pacemaker in the right atrium.
SaaSSoftware as a serviceComplete applications delivered over the web, such as email and office suites.
SATSource address tableThe switch's table of which MAC address lives on which port.
SBOMSoftware bill of materialsA list of the software components inside a device.
SBPSystolic blood pressureThe peak arterial pressure during ventricular contraction.
SCDSequential compression deviceInflatable leg sleeves that inflate in sequence to prevent blood clots.
SCRSilicon-controlled rectifierA thyristor that switches large currents on when triggered and stays on until current stops.
SDSSafety data sheetThe document describing a chemical's hazards, handling, and first aid.
SIMVSynchronized intermittent mandatory ventilationA ventilator mode that gives a set number of breaths synchronized with the patient's own efforts.
SMARTSelf-Monitoring, Analysis, and Reporting TechnologyHard drive self-diagnostics that warn before failure.
SMBServer Message BlockWindows file and printer sharing on TCP 445.
SMDASafe Medical Devices ActThe 1990 law requiring hospitals to report device-related deaths and serious injuries to the FDA and manufacturer.
SMPSSwitch-mode power supplyA power supply that switches at high frequency to regulate voltage efficiently.
SpO2Peripheral oxygen saturationThe percentage of hemoglobin carrying oxygen, measured by a pulse oximeter.
SSDSolid-state driveFlash storage with no moving parts: fast, silent, and shock resistant.
SSHSecure ShellEncrypted remote command-line access and file transfer on TCP 22; replaces Telnet and FTP.
STST segmentThe ECG segment between the QRS and the T wave, where injury and ischemia show as elevation or depression.
STELShort-term exposure limitThe maximum average exposure over a 15-minute period.
SVTSupraventricular tachycardiaA fast, regular rhythm originating above the ventricles, with narrow QRS.
TCTransport CanadaThe Canadian counterpart of the DOT for cylinder rules.
TCPTransmission Control ProtocolReliable, ordered, connection-based transport: three-way handshake, acknowledgments, retransmission.
TDRTime-domain reflectometerA cable tester that finds the distance to a break from signal reflections.
TENSTranscutaneous electrical nerve stimulationLow-level electrical pulses through skin electrodes for pain relief.
TLSTransport Layer SecurityThe encryption layer under HTTPS and most secure protocols; use 1.2 or 1.3.
TLVThreshold limit valueThe ACGIH recommended exposure limit.
TOCOTocodynamometerThe pressure sensor on a fetal monitor that records uterine contractions.
TPNTotal parenteral nutritionComplete nutrition delivered into a vein through an infusion pump.
TRTelecommunications roomThe wiring closet on each floor that serves nearby devices within the 90 m horizontal cable limit.
TVTelevisionA display or broadcast receiver; smart TVs are IoT devices.
UDPUser Datagram ProtocolConnectionless, fast transport with no delivery guarantee; used for voice, video, DNS, DHCP.
UEFIUnified Extensible Firmware InterfaceModern firmware that replaces BIOS: GPT disks, Secure Boot, mouse-driven setup.
UNUnited NationsThe body whose Globally Harmonized System is the format hazard communication now uses.
UPSUninterruptible power supplyA battery unit that keeps equipment running through a power cut and conditions the power.
USUnited StatesThe country; used for US standards, the 60 Hz mains frequency, and the AHA electrode color code.
USBUniversal Serial BusThe standard port for peripherals, storage, and charging.
uVMicrovoltOne millionth of a volt; EEG signals are in microvolts.
UVUltravioletLight beyond violet, used for disinfection and in spectrophotometers, and absorbed by glass and plastic.
VAVolt-ampereThe unit of apparent power, which a supply must deliver regardless of power factor.
VARVolt-ampere reactiveThe unit of reactive power, which shuttles between source and load without doing work.
VFVentricular fibrillationChaotic ventricular activity with no output; the rhythm defibrillators treat.
VLANVirtual local area networkA logical LAN on a switch: ports in different VLANs are separate broadcast domains.
VMVirtual machineA complete computer running as software on a hypervisor.
VPNVirtual private networkAn encrypted tunnel across an untrusted network such as the internet.
VTVentricular tachycardiaA fast rhythm from the ventricles with wide QRS complexes; pulseless VT is shockable.
VTTidal volumeThe volume of one breath, set on a ventilator at about 6 to 8 mL per kilogram of ideal body weight.
WANWide area networkConnects LANs across cities or countries, usually over links leased from a provider.
WEPWired Equivalent PrivacyThe original Wi-Fi encryption, broken in minutes; never use it.
Wi-FiWireless fidelityThe trade name for IEEE 802.11 wireless networking.
WIPOWorld Intellectual Property OrganizationThe UN agency for international copyright and patent treaties.
WLANWireless local area networkA LAN delivered over Wi-Fi instead of cables.
WMTSWireless Medical Telemetry ServiceThe FCC-protected bands for patient telemetry: 608 to 614, 1395 to 1400, and 1427 to 1432 MHz.
WPA2Wi-Fi Protected Access 2Wi-Fi security using AES with CCMP; still common and acceptable.
WPA3Wi-Fi Protected Access 3The current Wi-Fi security standard: SAE in Personal mode, stronger encryption, protected management frames.
XMLExtensible Markup LanguageA text format for structured data; HL7 version 3 and CDA use it.
YAGYttrium aluminum garnet (laser)The crystal in Nd:YAG lasers, which emit near-infrared light that penetrates tissue.
YSIYellow Springs Instruments (temperature probe standard)The company whose YSI 400 and YSI 700 thermistor curves became the standards for temperature probes.