Zonare z.one: troubleshooting and error codes

The ServiceForge knowledge base holds 90 documented symptoms for the Zonare z.one, with 124 ranked causes, 160 diagnostic checks with test points and expected values, and 3 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

No or poor B-mode image

B-mode ultrasound image absent, very dim, noisy, or lacking grayscale detail; display settings may be incorrect or transducer contact inadequate

Ranked causes

  1. 1. LCD display brightness and/or contrast set far from factory defaults (common)
  2. 2. B-mode gain control set incorrectly (common)
  3. 3. Acoustic power (A Output) set too low (common)
  4. 4. User presets maladjusted from factory settings (occasional)
  5. 5. Inadequate acoustic coupling gel or poor transducer contact (common)
  6. 6. Defective transducer causing poor image quality (occasional)
First check

Use grayscale test pattern to assess display quality

Test point
Diagnostic menu grayscale test pattern on LCD display
Expected
All grayscale shades visible from black to white with distinct gradations

6 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No or poor Color Mode image or PW Doppler

Color Doppler or Pulsed Wave Doppler mode imaging absent, very dim, noisy, or missing blood flow signals

Ranked causes

  1. 1. LCD display brightness and/or contrast set far from factory defaults (common)
  2. 2. Doppler control settings (gain, filter, scale) incorrect (common)
  3. 3. Scanning angle not optimized for blood flow measurement (common)
  4. 4. User presets maladjusted from factory settings (occasional)
  5. 5. Inadequate acoustic coupling gel or poor transducer contact (common)
  6. 6. Defective transducer causing poor Doppler signal (occasional)
First check

Use grayscale test pattern to verify display quality

Test point
Diagnostic menu grayscale test pattern on LCD display
Expected
All grayscale shades visible with good contrast

6 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SuperCart with docked Scan Engine fails to power on

System does not power on when AC power connected and Scan Engine docked to SuperCart

Ranked causes

  1. 1. AC power not connected to wall outlet (common)
  2. 2. Circuit breaker in off position (common)
  3. 3. Scanner link indicator lamp not flashing (Scan Engine not properly docked) (common)
  4. 4. SuperCart LCD monitor display powered off (occasional)
  5. 5. Problem with SuperCart or Scan Engine component failure (occasional)
First check

Verify AC power connection

Test point
AC power cord and wall outlet
Expected
AC power cord connected to powered wall outlet

4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Z.ONE will not communicate with target DICOM Store or Print device

DICOM connectivity failed; network communication between z.one and target DICOM device (storage or printer) not established; no handshaking

Ranked causes

  1. 1. Network hardware connection not active or cabling issue (common)
  2. 2. DICOM General configuration parameters incorrect or mismatched (common)
  3. 3. DICOM Network Store or Printer configuration parameters incorrect or mismatched (common)
  4. 4. DICOM handshaking verification test unsuccessful (occasional)
  5. 5. TCP/IP network connectivity issue between z.one and target device (occasional)
First check

Verify network hardware connection active

Test point
Networking ICON on lower/right of main imaging screen
Expected
Networking ICON displayed without X overlay, indicating active connection

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery does not charge in Scan Engine when docked in SuperCart

Battery charge level not increasing when Scan Engine docked and SuperCart powered; no charging indication

Ranked causes

  1. 1. AC power not present at SuperCart (common)
  2. 2. Circuit breaker at rear of SuperCart in OFF position (common)
  3. 3. DC voltages not present on Main/IO board (occasional)
  4. 4. Battery discharge level at 0% preventing charging of depleted battery (occasional)
First check

Verify AC power present at SuperCart

Test point
AC power cord and wall outlet
Expected
AC power cord connected to powered outlet; SuperCart receiving power

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Transducer not recognized by system (no B-mode imaging)

System does not detect transducer; no B-mode imaging available; transducer not appearing in system software

Ranked causes

  1. 1. Transducer connection intermittent or loose (common)
  2. 2. Transducer defective or incompatible (occasional)
  3. 3. Scan Engine software not recognizing transducer after power-on (occasional)
  4. 4. Scan Engine in SuperCart vs stand-alone issue (occasional)
First check

Disconnect and firmly reconnect transducer

Test point
Transducer connector on Scan Engine
Expected
Transducer recognized in system; B-mode imaging available

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Color or B/W printer fails to print

Printer powered on and connected but does not print images or data from SuperCart

Ranked causes

  1. 1. Printer out of paper (common)
  2. 2. Video cable not properly connected (common)
  3. 3. Incorrect video format selected (NTSC/PAL mismatch) (occasional)
First check

Check printer paper supply

Test point
Printer paper tray or magazine
Expected
Paper loaded and available in printer

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

External printer/recorder peripheral fails to power on

Peripheral device connected to SuperCart does not power on; no indication lights or operation

Ranked causes

  1. 1. AC power not connected or not reaching outlet box inside backpack (common)
  2. 2. AC power outlet box not powered (common)
  3. 3. Peripheral AC power cord defective or connection loose (occasional)
First check

Verify AC power cord connections at outlet box

Test point
AC power outlet box connections inside SuperCart backpack storage
Expected
AC power cord fully connected and seated at outlet box

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD display monitor on SuperCart has no video

SuperCart LCD display powered on but no video output; internal LEDs may be illuminated but display is blank

Ranked causes

  1. 1. LCD display powered off (common)
  2. 2. Software handshaking between SuperCart and Scan Engine interrupted (occasional)
  3. 3. Loose or disconnected cable connections on monitor back (occasional)
First check

Verify LCD display is powered on

Test point
SuperCart display; internal LED illumination visible through vent slots
Expected
Internal LEDs illuminated showing display is powered

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scan Engine undocked fails to power on

Scan Engine will not power on when operated as stand-alone unit without SuperCart docking

Ranked causes

  1. 1. Battery charge insufficient (less than 25%) (common)
  2. 2. Transducer connected causing power drain or conflict (occasional)
  3. 3. Scan Engine hardware failure or battery issue (occasional)
First check

Check battery charge level

Test point
Battery charge status button on Scan Engine
Expected
Charge greater than 25%

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display absent or partial

Total lack of OLED display or missing/random segments in display windows

Ranked causes

  1. 1. Missing or misaligned nylon insulating washer on P-Clamp (common)
  2. 2. Metal P-Clamp shorting OLED board traces or vias (common)
First check

Inspect P-Clamp assembly on back of OLED board

Test point
Metal P-Clamp retaining interface cable
Expected
Nylon insulating washer present and properly positioned

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display not operating normally

OLED display windows not functioning correctly, possibly after system power-up

Ranked causes

  1. 1. Missing nylon insulating washer under metal P-CLAMP on cable (common)
  2. 2. OLED board electronics damaged (rare)
First check

Inspect cable strapping for metal or nylon clamp type

Test point
Cable strapping clamp
Expected
Determine if metal P-CLAMP or nylon clamp is used

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Barco LCD display blank on SuperCart after docking Scan Engine

Video output missing on LCD display, no video initialization occurs on Cart Main Board after docking Scan Engine or powering on system with docked Scan Engine

Ranked causes

  1. 1. Inconsistent software initialization during handshaking between Scan Engine and SuperCart during docking or power-on (occasional)
First check

Inspect four green LEDs on right/rear of Rear I/O panel

Test point
Rear I/O panel, four green LEDs
Expected
Top 3 LEDs solidly illuminated; bottom LED (SCAN LINK) illuminated and flashing

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cable strap hardware assembly verification

Inspection procedure for OLED display cable strap insulating washer installation to prevent intermittent failures

Ranked causes

  1. 1. Manufacturing assembly defect in SmartCart production run (occasional)
First check

Visually inspect P-Clamp washer assembly behind OLED panel

Test point
Back of OLED board, metal P-Clamp, insulating washer location
Expected
Nylon insulating washer is present and correctly positioned to isolate P-Clamp from board traces

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Full Autoset display geometry and gain

Automatic calibration cycle using bar pattern to adjust image geometry, phase, and analog gain on SuperCart display

Ranked causes

  1. 1. Initial setup or recalibration required (common)
First check

Verify bar pattern displays on SuperCart

Test point
SuperCart display
Expected
Bar pattern visible

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: OLED display assembly check

Manufacturing assembly verification for SmartCart OLED display

Ranked causes

  1. 1. Assembly error: missing or mispositioned insulating washer (occasional)
First check

Inspect nylon insulating washer installation on P-Clamp

Test point
metal P-Clamp at OLED board interface
Expected
Washer present, correctly positioned, no metal-to-trace contact

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Scanner touchscreen

Touchscreen calibration procedure for Scan Engine to ensure accurate stylus contact detection at calibration points

Ranked causes

  1. 1. Touchscreen requires calibration for consistent operation (common)
First check

Remove Scan Engine from SuperCart

Test point
Scan Engine mounting
Expected
Scan Engine separated from SuperCart

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Touchscreen

Touchscreen calibration procedure for Scan Engine to ensure consistent operator contact response at calibration points

Ranked causes

  1. 1. Touchscreen requires calibration for accurate stylus contact detection (common)
First check

Remove Scan Engine from SuperCart

Test point
Scan Engine
Expected
Unit physically separated from cart

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Compatibility issue between early Main Board hardware and FPGA upgrade process

SmartCart Main Boards manufactured before Rev 3F version unable to complete FPGA programming with pre-4.3.88 software

Ranked causes

  1. 1. Early Main Board revisions incompatible with pre-4.3.88 FPGA upgrade firmware (occasional)
First check

Confirm Main Board hardware revision

Test point
Main Board USB/ATA Bridge device
Expected
Pre-Rev 3F or Rev 3F and later

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failed to open install file during USB software installation

Software installation from USB stick fails with inability to locate Install.zpk file when using rear USB ports on SmartCart

Ranked causes

  1. 1. USB port communication issue with rear SmartCart ports (unknown)
First check

Verify USB memory stick contains desired SW version installer files

Test point
USB Memory Stick
Expected
Install.zpk file present

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

FPGA firmware upgrade fails on SmartCart

Firmware upgrade routines cannot find USB/ATA Bridge Device and upgrade process fails during FPGA programming attempt.

Ranked causes

  1. 1. USB/ATA Bridge Device not programmed with ZONARE internal ID (common)
First check

Verify scanner system software version

Test point
Scanner system
Expected
4.3.88 or above

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

FPGA firmware upgrade fails on SmartCart Main Board

Firmware upgrade routines cannot find the USB/ATA Bridge device during FPGA programming process

Ranked causes

  1. 1. Cypress USB/ATA Bridge Device lacks internal ZONARE ID (common)
First check

Verify current system software version is below 4.3.88

Test point
Scanner system software
Expected
Pre-4.3.88 version

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Function (F1-F4) keys do not operate

Function keys on SuperCart or Scan Engine do not respond to user input; key presses have no effect

Ranked causes

  1. 1. Function keys not configured or misconfigured in user settings (common)
First check

Access function key configuration menu

Test point
On-screen menu path TOOLS | SYSTEM SETUP | FUNCTION KEYS
Expected
Configuration menu opens allowing function key reassignment

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Hard drive low-level formatting required

SmartCart with 80-120GB HDD requires low-level formatting procedure using command-line interface

Ranked causes

  1. 1. HDD requires low-level formatting for SmartCart deployment (unknown)
First check

Obtain service password from Zonare Technical Support

Test point
Service Support contact
Expected
Valid service password received

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

HDD low-level formatting required

SmartCart 80-120GB HDD needs low-level formatting and verification before use; accessed via System Console

Ranked causes

  1. 1. HDD initialization required for SmartCart operation (common)
First check

Obtain service password

Test point
Zonare Technical Support
Expected
Service password provided

7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Main Board FPGA firmware upgrade fails

FPGA programming procedure fails to complete on SmartCart systems during firmware upgrade process

Ranked causes

  1. 1. Cypress USB/ATA Bridge Device lacks ZONARE internal ID name (occasional)
First check

Verify Main Board hardware version is pre-Rev 3F USB/ATA Bridge device

Test point
Main Board
Expected
Pre-Rev 3F USB/ATA Bridge device version

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Main Board FPGA programming fails during upgrade

Firmware upgrade routines cannot locate USB/ATA Bridge device; FPGA programming process fails to complete

Ranked causes

  1. 1. Cypress USB/ATA Bridge device not programmed with ZONARE internal ID (common)
First check

Verify system software version is 4.3.88 or above

Test point
System software version display
Expected
4.3.88 or above

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Main Board FPGA programming fails on older SmartCart

Firmware upgrade routines cannot find the USB/ATA Bridge device during FPGA programming process

Ranked causes

  1. 1. Cypress USB/ATA Bridge Device not programmed with ZONARE internal ID (common)
First check

Verify system software version

Test point
System software version
Expected
4.3.88 or above

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No Doppler audio from SuperCart speakers after docking Scan Engine

Audio output missing from SuperCart speakers during operation, no audio initialization occurs on Cart Main Board; audio functions normally from undocked Scan Engine speakers

Ranked causes

  1. 1. Inconsistent software initialization during handshaking between Scan Engine and SuperCart during docking or power-on (occasional)
First check

Inspect four green LEDs on right/rear of Rear I/O panel

Test point
Rear I/O panel, four green LEDs
Expected
Top 3 LEDs solidly illuminated; bottom LED (SCAN LINK) illuminated and flashing

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nylon insulating washer incorrectly positioned on OLED clamp assembly patient safety

Nylon insulating washer installed only on top of P-CLAMP instead of between clamp and PC board

Ranked causes

  1. 1. Incorrect installation procedure during assembly or service (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nylon insulating washer installed incorrectly on OLED clamp patient safety

Insulating washer placed only on top of metal P-CLAMP instead of between clamp and PC board

Ranked causes

  1. 1. Incorrect assembly procedure during installation (unknown)
First check

Verify washer position under P-CLAMP

Test point
Metal P-CLAMP retention point
Expected
Nylon insulating washer positioned between clamp and PC board

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display assembly installed without proper insulation caution

Nylon insulating washer missing or incorrectly positioned between metal P-CLAMP and PCB, potentially causing electrical shorts or ground loops on User Interface assembly.

Ranked causes

  1. 1. Nylon insulating washer installed only on top of clamp instead of under clamp (common)
First check

Verify nylon insulating washer position

Test point
P-CLAMP assembly on PC board
Expected
Washer positioned under clamp between clamp and PCB

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display completely absent

User Controls LED display panel shows no display output on SmartCart systems

Ranked causes

  1. 1. Missing or mispositioned nylon insulating washer on metal P-Clamp (common)
First check

Visually inspect metal P-Clamp on back of OLED board for presence of nylon insulating washer

Test point
Metal P-Clamp assembly behind OLED board
Expected
Nylon insulating washer present and properly positioned

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display completely blank

Total lack of display on User Controls LED display panel (OLED) assembly on SmartCart

Ranked causes

  1. 1. Missing or mispositioned insulating washer on metal P-Clamp (common)
First check

Inspect P-Clamp on back of OLED board for presence and correct position of nylon insulating washer

Test point
Metal P-Clamp cable retention point
Expected
Nylon insulating washer present and correctly installed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display completely dark

Total lack of OLED display output on SmartCart assembly

Ranked causes

  1. 1. Missing or mispositioned nylon insulating washer on metal P-Clamp (common)
First check

Verify nylon insulating washer presence and position on P-Clamp

Test point
Metal P-Clamp on OLED board back
Expected
Nylon washer present and correctly positioned between P-Clamp and circuit board traces

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display completely non-functional

Total lack of OLED (User Controls LED display panel) output on SmartCart

Ranked causes

  1. 1. Missing or mispositioned nylon insulating washer on metal P-Clamp (common)
First check

Inspect P-Clamp for nylon insulating washer presence and position

Test point
Metal P-Clamp on back of OLED board
Expected
Nylon insulating washer present and correctly positioned

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display has missing or random segments

Some segments missing or appearing randomly in windows of OLED display on SmartCart

Ranked causes

  1. 1. Metal P-Clamp short-circuiting OLED board traces or vias due to missing washer (occasional)
First check

Visually inspect metal P-Clamp for nylon insulating washer presence and position

Test point
P-Clamp on back of OLED board
Expected
Nylon insulating washer present and correctly positioned to prevent metal contact with board traces

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display not functioning

Total lack of OLED (User Controls LED display panel) display output on SmartCart system

Ranked causes

  1. 1. Missing or incorrectly positioned nylon insulating washer on metal P-Clamp (occasional)
First check

Inspect P-Clamp assembly on back of OLED board

Test point
Metal P-Clamp interface cable retention area
Expected
Nylon insulating washer present and correctly seated

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display not illuminating

Total lack of OLED (User Controls LED display panel) illumination on SmartCart

Ranked causes

  1. 1. Missing or mispositioned nylon insulating washer on metal P-Clamp (common)
First check

Visually inspect P-Clamp and insulating washer on OLED board assembly

Test point
P-Clamp retaining interface cable at OLED board back
Expected
Nylon insulating washer present and correctly positioned

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display not showing

Total lack of OLED (User Controls LED display panel) output on SmartCart

Ranked causes

  1. 1. Missing or incorrectly positioned nylon insulating washer on metal P-Clamp (common)
First check

Inspect P-Clamp assembly on back of OLED board for presence and correct positioning of nylon insulating washer

Test point
Metal P-Clamp, back of OLED board
Expected
Nylon insulating washer present and covering metal contact areas

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display not visible

Total lack of OLED (User Controls LED display panel) output on SmartCart

Ranked causes

  1. 1. Missing or mispositioned nylon insulating washer on metal P-Clamp (common)
First check

Inspect P-Clamp on back of OLED board for nylon insulating washer presence and position

Test point
Metal P-Clamp cable retention point
Expected
Nylon insulating washer present and correctly positioned

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display partial or random segment loss

Some missing or random segments visible in OLED display windows, intermittent or incomplete display output

Ranked causes

  1. 1. Missing or mispositioned nylon insulating washer causing trace/via shorting (common)
First check

Inspect metal P-Clamp contact with OLED board back surface

Test point
Interface cable P-Clamp and OLED board traces/vias
Expected
No metal-to-trace contact; insulating washer properly positioned

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display segments missing or intermittent

Some missing or random segments in windows of OLED display on SmartCart

Ranked causes

  1. 1. Metal P-Clamp contacting OLED board traces or vias due to missing insulator (common)
First check

Verify nylon insulating washer installed on P-Clamp

Test point
Metal P-Clamp at OLED board interface cable point
Expected
Nylon insulating washer present and in correct position

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display segments missing or random

Some missing or random segments visible in windows of OLED display output

Ranked causes

  1. 1. Partial shorting of traces or vias on OLED board backside due to metal P-Clamp contact (occasional)
First check

Inspect metal P-Clamp for contact with OLED board

Test point
Back of OLED board under P-Clamp
Expected
No metal contact; nylon insulating washer fully seated between clamp and board

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display shows missing or random segments

Some segments missing or random segments appear in windows of OLED display on SmartCart

Ranked causes

  1. 1. Missing or mispositioned nylon insulating washer causing shorts on OLED board (common)
First check

Visually inspect metal P-Clamp on back of OLED board for presence and correct position of nylon insulating washer

Test point
Metal P-Clamp assembly behind OLED board
Expected
Nylon insulating washer present and properly positioned to prevent metal contact with traces

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display shows no image

Total lack of OLED (User Controls LED display panel) output on SmartCart miniCart assembly

Ranked causes

  1. 1. Missing or mispositioned insulating washer on P-Clamp causes metal contact with OLED board traces (occasional)
First check

Visually inspect P-Clamp on back of OLED board for presence and correct positioning of nylon insulating washer

Test point
P-Clamp metal contact area on OLED board back
Expected
Nylon insulating washer present and positioned to prevent metal-to-trace contact

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display still malfunctioning after washer installation

OLED display continues to have problems even after nylon insulating washer has been installed and reassembled

Ranked causes

  1. 1. OLED board electronics damaged from previous electrical contact (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED display totally absent

User Controls LED display panel shows no display output on SmartCart OLED assembly

Ranked causes

  1. 1. Missing or mispositioned nylon insulating washer on metal P-Clamp (common)
First check

Inspect back of OLED board for nylon insulating washer on metal P-Clamp

Test point
Metal P-Clamp cable retention point
Expected
Nylon insulating washer present and properly positioned

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OLED insulating washer incorrectly positioned on clamp high voltage

Nylon insulating washer installed only on top of metal P-CLAMP instead of between clamp and PC board, causing potential electrical isolation failure

Ranked causes

  1. 1. Washer installed on top of clamp instead of underneath (common)
First check

Verify z.oneUltra powered off

Test point
Power switch
Expected
OFF

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power button becomes inoperative

Main Power ON/OFF switch on Scan Engine stops responding to button press after excessive or improper pressure

Ranked causes

  1. 1. Rubber power button cover (elastomer) displaced from alignment with switch mechanism (common)
First check

Verify power button responds to depression

Test point
Main Power ON/OFF switch on Scan Engine
Expected
Switch activates when pressed 1/16 inch with flat fingertip surface

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power ON/OFF switch becomes inoperative

Scan Engine main Power ON/OFF switch stops responding to button presses due to improper pressing technique or excessive force

Ranked causes

  1. 1. Rubber power button cover (elastomer) displaced from alignment with switch (common)
First check

Verify correct finger positioning on power button

Test point
Scan Engine Power ON/OFF button
Expected
Finger positioned perpendicular to button travel direction using flat surface area

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power switch becomes inoperative

Main Power ON/OFF switch on Scan Engine no longer functions after improper button pressing

Ranked causes

  1. 1. Rubber power button cover displaced from alignment due to excessive hard or pinpoint pressing (common)
First check

Inspect power button cover alignment

Test point
Scan Engine Power ON/OFF button
Expected
Button cover properly aligned with switch mechanism below

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power switch becomes inoperative after use

Main Power ON/OFF switch on Scan Engine stops responding to button presses due to misalignment of rubber button cover (elastomer) with internal switch mechanism.

Ranked causes

  1. 1. Rubber button cover displaced due to excessive or off-center pressing (common)
First check

Inspect rubber button cover alignment with internal switch mechanism below

Test point
Power ON/OFF button on Scan Engine
Expected
Rubber cover centered and flush with switch housing

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Prevent power switch operational problems caution

Proper technique for pressing the Power ON/OFF button to avoid damage or misalignment of the button mechanism

Ranked causes

  1. 1. Excessive force or pinpoint pressing with fingertip (common)
First check

Verify button depression depth

Test point
Power ON/OFF button on Scan Engine
Expected
Button activates when depressed 1/16 inch or less

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scan Engine undocked fails to properly boot

Scan Engine powers on but does not complete boot sequence; system hangs or freezes during startup

Ranked causes

  1. 1. System software files corrupted or missing (rare)
First check

Boot from USB Memory Stick with system software

Test point
USB port at rear of Scan Engine; boot selection menu
Expected
System boots successfully from USB Memory Stick

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

SmartCart cart serial programming via HDD formatting

HDD formatting also programs the Cart Serial Number during low-level format operation

Ranked causes

  1. 1. SmartCart requires HDD format to program Serial Number (unknown)

SmartCart Main Board hardware incompatibility with pre-4.3.88 software

Early revision Main Boards (pre-Rev 3F) cannot be programmed for FPGA upgrade with software versions before 4.3.88

Ranked causes

  1. 1. Pre-Rev 3F Main Board USB/ATA Bridge lacks ZONARE identifier (common)
First check

Identify Main Board revision

Test point
Main Board hardware
Expected
Pre-Rev 3F or Rev 3F

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Software installation fails from USB port

Attempt to install software from USB stick fails with error 'Failed to Open Install File', unable to locate 'Install.zpk' file

Ranked causes

  1. 1. USB port connection or file access issue at SmartCart rear ports (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Software installation fails from USB port on SmartCart

Failure to locate Install.zpk file when attempting software installation from USB stick using rear USB ports on SmartCart

Ranked causes

  1. 1. USB port detection or file access issue (occasional)

Software installation fails from USB stick

USB port at rear of SmartCart fails to locate Install.zpk file during software installation attempt

Ranked causes

  1. 1. USB port intermittent connection or recognition issue at rear SmartCart port (unknown)

More symptoms

  • Software installation fails from USB to SmartCart
  • Software installation fails to locate Install.zpk file from USB
  • Software installation fails via USB port
  • Software installation from USB fails on SmartCart rear port
  • Software installation from USB fails to locate Install file
  • Software installation from USB fails to locate Install.zpk file
  • Software installation from USB fails to open install file
  • Some missing or random segments in OLED display
  • SuperCart instability during relocation
  • SuperCart instability or tipping during movement
  • SuperCart rear wheel orientation unsafe
  • SuperCart rear wheels incorrectly oriented
  • SuperCart rear wheels oriented incorrectly
  • SuperCart tips during relocation
  • SuperCart tips or is unstable during movement
  • SuperCart video display blank after docking Scan Engine
  • Total lack of OLED display
  • Touchscreen calibration contact technique failure
  • Touchscreen calibration fails or produces inaccurate response
  • Touchscreen calibration points contact failure
  • Touchscreen calibration points not responding
  • Touchscreen calibration technique error
  • Transducer case or face cracked or damaged
  • Transducer connector pins bent, broken, or missing
  • Transducer files lost from internal CF card
  • Transducer support files not available
  • Unable to modify existing User Preset (shown as grayed out on screen)
  • USB port software install failures
  • USB software installation fails on Smartcart rear ports
  • User Interface removal required

30 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
Calibration failed messageTouchscreen calibration process did not complete successfullyImproper stylus contact technique or excessive debounce during target point contactwarning
Failed to Open Install FileSoftware installation cannot locate or access Install.zpk fileAttempt to perform software installation from USB stick at SmartCart rear USB portserror
Transducer Not SupportedOne or more transducer types are not available on the scanner due to missing transducer support files.User performs FORMAT operation on C drive on scanner with 1.9GB partition configuration, erasing transducer files.error

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI