ZOLL SurePower Battery Pack: troubleshooting and error codes
The ServiceForge knowledge base holds 12 documented symptoms for the ZOLL SurePower Battery Pack, with 12 ranked causes, 9 diagnostic checks with test points and expected values, and 2 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Battery capacity below service threshold patient safety
Capacity check lights only a single LED, indicating battery capacity is at or near 60% of original capacity
Ranked causes
- 1. Battery capacity degradation to 60% or below original capacity (common)
Check battery capacity
- Test point
- Battery Display button (held for at least 5 seconds)
- Expected
- Five LEDs light to indicate capacity equal to or close to new battery; fewer LEDs indicate decreased capacity
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery casing cracked or damaged patient safety
Physical damage visible on battery exterior casing
Ranked causes
- 1. Physical damage to battery casing (unknown)
Inspect battery casing for cracks or damage
- Test point
- Battery exterior casing
- Expected
- No cracks or visible damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery fault condition prevents defibrillator operation patient safety
Battery indicates fault condition or operating state that prevents effective use of the defibrillator
Ranked causes
- 1. Battery malfunction or fault state (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery fault detected patient safety
Fault LED lights for approximately 10 seconds after pressing Display button; runtime indicator does not display runtime for that battery
Ranked causes
- 1. Battery fault condition (unknown)
Press Display button
- Test point
- Battery Display button
- Expected
- Fault LED lights for approximately 10 seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery LEDs do not work
LEDs do not illuminate when Display button is pressed
Ranked causes
- 1. LED failure (unknown)
Press Display button to test LEDs
- Test point
- Battery Display button and LED indicators
- Expected
- LEDs illuminate
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery over temperature during charge (Fault Code 13) caution
Battery is at or above maximum temperature for charging (140°F/60°C)
Ranked causes
- 1. Battery temperature at or above 140°F/60°C during charging (occasional)
Press and hold Display button for 15 seconds to read fault code
- Test point
- Battery Display button
- Expected
- Fault LED blinks 13 times to indicate Fault Code 13 (over temperature during charge)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery performance degraded after drop or physical damage patient safety
SurePower Battery Pack dropped or physically damaged, may expose users to safety hazard and affect reliable operation
Ranked causes
- 1. Battery physically dropped or damaged (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery requires recalibration
Recalibration LED lights for approximately 10 seconds after pressing Display button; runtime indicator does not display runtime for that battery
Ranked causes
- 1. Battery needs recalibration (unknown)
Press Display button
- Test point
- Battery Display button
- Expected
- Recalibration LED lights for approximately 10 seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery runtime indicator LEDs not illuminating
When Display button is pressed, LEDs do not light or fail to show expected runtime indication
Ranked causes
- 1. Display button not pressed or released correctly (common)
Verify Display button responds to press
- Test point
- Display button on top of SurePower Battery Pack
- Expected
- All LEDs blink when button pressed and released
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery runtime indicator shows low or blinking LEDs patient safety
Left-most LED blinks or fewer than expected LEDs illuminate indicating insufficient runtime
Ranked causes
- 1. Battery charge depleted below 15 minutes runtime (common)
Press Display button and observe LED pattern
- Test point
- Runtime indicator LEDs on SurePower Battery Pack
- Expected
- If left-most LED blinks, battery runtime is less than 15 minutes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery under temperature during charge (Fault Code 14) caution
Battery is at or below minimum temperature for charging (32°F/0°C)
Ranked causes
- 1. Battery temperature at or below 32°F/0°C during charging (occasional)
Press and hold Display button for 15 seconds to read fault code
- Test point
- Battery Display button
- Expected
- Fault LED blinks 14 times to indicate Fault Code 14 (under temperature during charge)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display indicators inconsistent between battery and charger/defibrillator patient safety
Display indicators on SurePower Battery Pack do not match indicators on SurePower Charger Station or ZOLL defibrillator, indicating potential fault condition
Ranked causes
- 1. Battery fault condition (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
13 | Battery Over Temperature During Charge | Battery temperature reaches or exceeds 140°F/60°C during charging | warning |
14 | Battery Under Temperature During Charge | Battery temperature drops to or below 32°F/0°C during charging | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.