ZOLL R: troubleshooting and error codes

The ServiceForge knowledge base holds 150 documented symptoms for the ZOLL R, with 290 ranked causes, 307 diagnostic checks with test points and expected values, and 214 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Excessive artifact with paddles as ECG source

Paddles used as ECG source show excessive noise or artifact on display

Ranked causes

  1. 1. PADDLES not selected as ECG source (common)
  2. 2. Paddles not pressed firmly on patient (common)
  3. 3. Insufficient gel on paddles (common)
  4. 4. Paddle surface contaminated or dirty (common)
  5. 5. Debris between adult and pediatric paddle plate (occasional)
  6. 6. Paddle cable damaged (occasional)
  7. 7. Poor electrode contact; use ECG electrodes instead (common)
First check

Verify PADDLES selected as ECG source

Test point
ECG source selector
Expected
PADDLES mode indicated on display

3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Energy does not discharge when SHOCK button pressed high voltage

SHOCK button activated but no therapeutic energy delivered to patient

Ranked causes

  1. 1. Charge timeout - over 60/120 seconds elapsed in Manual mode (common)
  2. 2. Device in Sync/Remote Sync mode with no QRS detected (occasional)
  3. 3. Energy selection changed during charge or after ready (occasional)
  4. 4. Device not fully charged when SHOCK pressed (common)
  5. 5. SHOCK button not held down long enough (occasional)
  6. 6. Pads or paddles not making good contact (common)
First check

Check if charge timeout occurred

Test point
Manual mode timer and device display
Expected
Less than 60/120 seconds since charge ready

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Excessive artifact when using paddles as ECG source

ECG signal shows high noise and artifact when defibrillation paddles are used for ECG acquisition

Ranked causes

  1. 1. Paddles not selected or in manual mode (common)
  2. 2. Insufficient paddle pressure on patient skin (common)
  3. 3. Inadequate conductive gel on paddles (common)
  4. 4. Dirty or contaminated paddle surfaces (occasional)
  5. 5. Damaged paddle cable (occasional)
  6. 6. Electrodes provide better signal than paddles (common)
First check

Confirm paddles mode selected

Test point
Mode display or selector
Expected
PADDLES displayed as active ECG source

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CHECK PADS or ATTACH PADS message for defibrillation patient safety

Unit displays message indicating therapy electrodes are not properly connected or are defective

Ranked causes

  1. 1. Therapy electrodes not connected to OneStep cable (common)
  2. 2. Dried out electrode gel (common)
  3. 3. Poor electrode-to-patient contact (common)
  4. 4. Defective OneStep cable (occasional)
  5. 5. Consider using paddles if electrode issues persist (occasional)
First check

Disconnect and reconnect OneStep cable and electrodes

Test point
Cable and electrode connectors
Expected
Message clears after reconnection

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No 30J TEST OK message in manual self-test high voltage

Manual defibrillator self-test at 30 joules does not produce expected TEST OK message

Ranked causes

  1. 1. Energy not set to 30 joules (common)
  2. 2. OneStep electrode package not sealed (test with electrodes) (common)
  3. 3. OneStep cable not fully inserted into electrode connector (common)
  4. 4. OneStep cable not fully inserted into test port (test port method) (common)
  5. 5. Paddles not firmly pressed during discharge (paddle test) (common)
First check

Verify energy set to 30 joules

Test point
Energy level selector
Expected
30 joules selected

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No ventricular capture beat after stimulus marker patient safety

Pacing stimulus present on ECG but no cardiac response or capture beat visible

Ranked causes

  1. 1. Patient's pulse already present (common)
  2. 2. Pacing output current too low (common)
  3. 3. Electrode-to-patient contact poor (common)
  4. 4. Suboptimal electrode placement (common)
  5. 5. Wrong ECG lead selected (occasional)
First check

Assess patient's perfusion status

Test point
Patient vitals and pulse
Expected
Patient has adequate perfusion; capture needed

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No ventricular capture beat after stimulus marker on ECG patient safety

Pacing stimulus marker present but no cardiac response (capture beat) appears on ECG

Ranked causes

  1. 1. No cardiac output or patient in cardiac arrest (common)
  2. 2. Output current insufficient for capture (common)
  3. 3. Poor electrode contact with patient (common)
  4. 4. Suboptimal electrode placement for pacing (occasional)
  5. 5. Suboptimal ECG lead configuration (occasional)
First check

Check patient pulse and cardiac status

Test point
Patient assessment
Expected
Patient has palpable pulse or cardiac output evident

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Noisy ECG, artifact, wandering baseline

ECG display shows excessive noise, artifact or baseline drift affecting signal quality

Ranked causes

  1. 1. Poor electrode adhesion or skin preparation (common)
  2. 2. Electrode or cable movement creating artifact (common)
  3. 3. Patient movement (common)
  4. 4. Excessive radio frequency interference (occasional)
  5. 5. Inappropriate filter settings (occasional)
First check

Inspect electrode adhesion and skin contact

Test point
Electrode pads and patient skin
Expected
Electrodes fully adhered with good skin contact

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sync marker absent or inconsistent with QRS patient safety

Synchronization marker not appearing or not aligned with QRS complex

Ranked causes

  1. 1. Device not in Sync mode (common)
  2. 2. Device in Remote Sync mode instead of Sync (occasional)
  3. 3. Wrong ECG lead selected (common)
  4. 4. Suboptimal electrode placement (common)
  5. 5. Recorder paper too narrow (rare)
First check

Verify Sync mode enabled

Test point
Mode display
Expected
Sync mode active (not Remote Sync)

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sync marker absent or inconsistent with QRS waveform patient safety

Synchronization marker not appearing or not aligned with QRS complex on display and recorder

Ranked causes

  1. 1. Device not in Sync mode (common)
  2. 2. Device in Remote Sync mode when Sync mode intended (occasional)
  3. 3. Suboptimal ECG lead selection (common)
  4. 4. Poor electrode contact or placement (common)
  5. 5. Recorder paper too narrow (rare)
First check

Verify Sync mode is selected

Test point
Mode selector or display
Expected
SYNC mode displayed; Sync marker visible on QRS

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to deliver SHOCK in Sync mode high voltage

Cannot deliver energy to patient while in Sync (synchronous cardioversion) mode

Ranked causes

  1. 1. SYNC selection not displayed (common)
  2. 2. No ECG signal displayed (common)
  3. 3. Sync markers not present on QRS (common)
  4. 4. Wrong electrode placement (common)
  5. 5. SHOCK button not held long enough (common)
First check

Verify SYNC mode is active

Test point
Monitor display
Expected
SYNC XXXJ SEL displayed

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to SHOCK when in Sync mode high voltage

Defibrillator will not deliver synchronized shock despite Sync mode active

Ranked causes

  1. 1. No SYNC XXXJ SEL display indicating sync not ready (common)
  2. 2. No ECG signals displayed (common)
  3. 3. No Sync markers (arrows) above R waves (common)
  4. 4. SHOCK button not held down long enough (occasional)
  5. 5. Suboptimal electrode placement (occasional)
First check

Verify SYNC XXXJ SEL displayed

Test point
Monitor display status line
Expected
SYNC XXXJ SEL visible on display

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery indicator alternating green and yellow

Battery status indicator light cycles between green and yellow colors

Ranked causes

  1. 1. Battery not installed or properly seated (common)
  2. 2. Battery fault condition detected (common)
  3. 3. Battery depleted or defective (common)
  4. 4. Charger or charging system fault (occasional)
First check

Verify battery is installed

Test point
Battery compartment
Expected
Battery pack properly installed and seated

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CHECK PADS message displayed patient safety

Message indicates therapy electrode connection or integrity problem

Ranked causes

  1. 1. Therapy electrode not connected to OneStep cable (common)
  2. 2. Electrode gel is dry (common)
  3. 3. Electrode-to-patient contact is poor (common)
  4. 4. OneStep cable is defective (occasional)
First check

Verify electrode to OneStep cable connection

Test point
Therapy electrode connector to OneStep cable
Expected
Firm connection between electrode and cable

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CHECK PADS message for pacing patient safety

Unit displays CHECK PADS message when pacing mode is selected

Ranked causes

  1. 1. Therapy electrodes not connected to OneStep cable (common)
  2. 2. Electrode gel dried out or aged beyond specification (common)
  3. 3. Poor electrode-to-patient contact (common)
  4. 4. Defective OneStep cable (occasional)
First check

Verify therapy electrode connection to OneStep cable

Test point
OneStep cable connector to electrodes
Expected
Electrodes firmly connected to cable

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CHECK PADS or ATTACH PADS message patient safety

Message indicating hands-free therapy electrode connection problem

Ranked causes

  1. 1. Pads not properly connected or seated (common)
  2. 2. Electrode contact or gel condition poor (common)
  3. 3. OneStep cable is defective (occasional)
  4. 4. Pad/electrode set is defective (occasional)
First check

Disconnect and reconnect OneStep cable

Test point
OneStep cable connection
Expected
Cable firmly seated on both electrodes and device

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG LEAD OFF message displayed patient safety

Unit displays message indicating ECG cable or electrodes are not connected or making contact

Ranked causes

  1. 1. ECG cable or electrodes not connected or making poor contact (common)
  2. 2. Dried out or degraded electrodes (common)
  3. 3. Cable change requiring reset (occasional)
  4. 4. Defective cable (occasional)
First check

Verify ECG cable and electrode connections

Test point
Cable connectors and patient electrode pads
Expected
Cable firmly connected to unit and patient, electrodes in contact with skin

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

LOW BATTERY message appears with AC power connected

LOW BATTERY message displayed even though unit is plugged into AC mains

Ranked causes

  1. 1. AC Power Indicator not illuminated (common)
  2. 2. Battery depleted or defective (common)
  3. 3. Charger connection intermittent (occasional)
  4. 4. AC mains not functional (occasional)
First check

Verify AC Power Indicator illuminated

Test point
AC Power Indicator light
Expected
AC Power Indicator lit

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No 30J TEST OK message during manual defibrillator self-test high voltage

Defibrillator self-test at 30 joules fails to display TEST OK message

Ranked causes

  1. 1. Device not set to 30 joules (common)
  2. 2. OneStep electrode not properly connected or package sealed (common)
  3. 3. Test Port connector not firmly inserted (occasional)
  4. 4. Paddles not firmly pressed during paddle test (occasional)
First check

Verify device set to 30 joules

Test point
Energy selector display
Expected
30 joules displayed

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No apparent energy delivery to patient high voltage

No visible patient reaction when defibrillator shock is delivered, or energy delivery uncertain

Ranked causes

  1. 1. Some patients will not show physical reaction to energy (common)
  2. 2. Device malfunction or insufficient energy output (occasional)
  3. 3. Poor pad contact or CHECK PADS message (common)
  4. 4. Poor pad placement or adhesion with hands-free pads (occasional)
First check

Review discharge stripchart for energy confirmation

Test point
Stripchart printout showing Joules delivered
Expected
Energy delivery recorded (e.g., 200J delivered)

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

POOR LEAD CONTACT message displayed patient safety

Unit indicates ECG electrodes or pacing cable not making adequate contact with patient

Ranked causes

  1. 1. Electrode or cable not connected or making poor skin contact (common)
  2. 2. Dried out or degraded electrodes (common)
  3. 3. Cable type change requiring reset (occasional)
  4. 4. Defective cable (occasional)
First check

Verify cable and electrode connections

Test point
Cable connectors and electrode pads
Expected
Cable firmly connected, electrodes in good contact with patient skin

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Standby pacing patient gets paced intermittently patient safety

Patient on standby pacing mode is being paced sporadically instead of continuously when needed

Ranked causes

  1. 1. ECG lead connection or placement problem (common)
  2. 2. ECG cable damaged (occasional)
  3. 3. Patient's R-to-R interval varying (common)
  4. 4. Pacing rate not set appropriately (common)
First check

Verify ECG/pacing cable connection

Test point
Cable connectors
Expected
Cable firmly seated and connected

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to deliver SHOCK in Remote Sync mode high voltage

Cannot deliver energy to patient while in Remote Sync (external synchronization) mode

Ranked causes

  1. 1. REMOTE SYNC not properly displayed (common)
  2. 2. Remote device not properly connected (common)
  3. 3. Remote device not sending proper Sync markers (common)
  4. 4. SHOCK button not held long enough (common)
First check

Verify REMOTE SYNC mode active

Test point
Monitor display
Expected
REMOTE SYNC shown in place of ECG; REMOTE SYNC XXXJ SEL displayed

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to SHOCK when in Remote Sync mode high voltage

Defibrillator will not deliver energy in Remote Sync mode despite remote device connected

Ranked causes

  1. 1. REMOTE SYNC not displayed or not set up (common)
  2. 2. Remote device not properly connected or not conforming to specs (occasional)
  3. 3. Sync markers not present or not synchronized with R waves on remote (occasional)
  4. 4. SHOCK button not held down properly (occasional)
First check

Verify REMOTE SYNC displayed

Test point
Device display status
Expected
REMOTE SYNC shown and REMOTE SYNC XXXJ SEL visible

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery indicator alternately illuminating green and yellow patient safety

Battery status indicator shows alternating green and yellow lights indicating abnormal charging state

Ranked causes

  1. 1. Battery pack not installed or not recognized (common)
  2. 2. Battery pack discharged or defective (common)
  3. 3. Charging circuit malfunction (occasional)
First check

Verify battery pack installed

Test point
Battery compartment
Expected
Battery pack present and properly seated

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery LED not showing proper status caution

Battery LED is flashing or not displaying solid green or amber, indicating battery fault or seating issue

Ranked causes

  1. 1. Battery not properly seated in well (common)
  2. 2. Battery displaying fault indicator (occasional)
  3. 3. Damaged or dirty battery pins and contacts (occasional)
First check

Check battery LED status

Test point
Battery LED on unit
Expected
Solid green or solid amber LED

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Impedance after board replacement patient safety

After replacing digital board, analog board, or PD engine, impedance calibration is required to ensure proper defibrillation delivery

Ranked causes

  1. 1. Digital board replaced (unknown)
  2. 2. Analog board replaced (unknown)
  3. 3. PD engine replaced (unknown)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CHECK RECORDER message displayed

Unit displays message indicating recorder malfunction or paper issue

Ranked causes

  1. 1. Recorder out of paper (common)
  2. 2. Recorder door open (common)
  3. 3. Paper loaded upside down or incorrectly (occasional)
First check

Check paper level in recorder

Test point
Recorder paper cartridge or tray
Expected
Paper present and not empty

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Code Readiness indicator shows red X when unit is powered on patient safety

Red X displayed on Code Readiness indicator while defibrillator is running

Ranked causes

  1. 1. Code readiness fault during operation (common)
  2. 2. Defibrillator self-test failure (occasional)
  3. 3. Persistent code readiness fault (rare)
First check

Power cycle unit

Test point
Power control
Expected
Unit powers off and back on

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Defibrillation charging or delivery not working patient safety

CHARGE button does not charge unit to selected energy; SHOCK button does not deliver energy; energy selection not incrementing correctly

Ranked causes

  1. 1. Energy selection circuit malfunction (occasional)
  2. 2. Charge circuit malfunction (occasional)
  3. 3. SHOCK button not making contact or delivery circuit failure (rare)
First check

Test energy selection decrements correctly

Test point
Display energy value and ENERGY SELECT buttons
Expected
Energy decrements from selected value to 1J when down arrow held; increments sequentially when up arrow pressed

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Light or poor quality printing

Recorded output is too faint or print quality is degraded

Ranked causes

  1. 1. Wrong paper type installed (common)
  2. 2. Paper installed backwards (common)
  3. 3. Recorder print head requires cleaning (occasional)
First check

Verify paper type

Test point
Recorder paper
Expected
Correct manufacturer-approved paper type

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Light or poor quality printing from recorder

Recorder output is faint or low contrast; printout is difficult to read

Ranked causes

  1. 1. Incorrect or deteriorated paper type (common)
  2. 2. Paper installed upside down (common)
  3. 3. Print head requires cleaning (occasional)
First check

Verify correct paper type installed

Test point
Paper specification and packaging
Expected
Correct paper type for device installed

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

LOW BATTERY message while plugged into AC mains patient safety

Unit displays low battery warning even though device is connected to AC power

Ranked causes

  1. 1. AC Power Indicator not illuminated; AC not connected properly (common)
  2. 2. Battery pack discharged or defective (common)
  3. 3. AC power cord or wall outlet not working (occasional)
First check

Verify AC Power Indicator illuminated

Test point
AC Power Indicator light
Expected
AC Power Indicator lit (green or appropriate color)

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Patient on standby pacing gets paced intermittently patient safety

Device unexpectedly delivers pacing pulses when in standby mode or at inappropriate times

Ranked causes

  1. 1. ECG lead wire disconnected causing asynchronous pacing (common)
  2. 2. Damaged ECG cable (occasional)
  3. 3. Patient R-to-R interval varying close to pacing rate (occasional)
First check

Verify ECG electrode connection and placement

Test point
Electrode pads and cable connectors
Expected
All electrodes firmly connected; good skin contact

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power supply verification after board or supply replacement high voltage

Power supply testing required after replacing digital board, AC power supply, or battery interconnect board to verify proper operation

Ranked causes

  1. 1. Digital board replaced (unknown)
  2. 2. AC power supply replaced (unknown)
  3. 3. Battery interconnect board replaced (unknown)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 Monitor not displaying reading patient safety

Monitor shows dashed line or no SpO2 value after simulator connection

Ranked causes

  1. 1. Simulator not enabled or in wrong mode (common)
  2. 2. Cable or sensor connection loose or damaged (common)
  3. 3. Display update delay (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not turn on or unexpectedly shuts off patient safety

Defibrillator fails to power on or powers off without user action

Ranked causes

  1. 1. Battery pack not properly installed or discharged (common)
  2. 2. Unit not connected to AC power (common)
  3. 3. Battery pack fully discharged (occasional)
First check

Verify battery pack installation

Test point
Battery compartment
Expected
Battery pack fully inserted and locked

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wi-Fi card does not transmit Full Disclosure File patient safety

R Series fails to send Full Disclosure File via Wi-Fi; error message displayed or transmission unsuccessful

Ranked causes

  1. 1. Wi-Fi card not configured correctly for network connection (common)
  2. 2. Wi-Fi access point not available or out of range (occasional)
  3. 3. Defective Wi-Fi card (occasional)
First check

Turn R Series to On or Monitor mode

Test point
Device power mode
Expected
Device in On or Monitor mode

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wi-Fi card fails to transmit device history

Device History transmission to ZOLL server unsuccessful; error message displayed instead of 'Defib History has sent'

Ranked causes

  1. 1. Wi-Fi card not configured correctly for network connection (common)
  2. 2. Wi-Fi access point unavailable (occasional)
  3. 3. Defective Wi-Fi card (occasional)
First check

Verify Wi-Fi card configuration matches intended transmission type

Test point
R Series Print Config menu
Expected
Configuration matches Full Disclosure File, Device History, or both as intended

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery and AC Power indicators not illuminated when plugged in

Neither Battery nor AC Power indicator lights up when device is connected to AC mains

Ranked causes

  1. 1. AC power connection problem (common)
  2. 2. AC mains not functional (occasional)
First check

Disconnect and reconnect AC power

Test point
AC power connection
Expected
Device unplugged and replugged; indicators illuminate

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Code Readiness indicator displays red X while powered on patient safety

Code readiness LED shows red X symbol when defibrillator unit is powered on and running.

Ranked causes

  1. 1. Code readiness fault active during operation (unknown)
  2. 2. Unit continues to fail after troubleshooting (occasional)
First check

Power cycle unit

Test point
Power switch
Expected
Unit boots up, readiness indicator becomes green or shows status message

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Code Readiness indicator displays red X while unit is on patient safety

Code Readiness indicator shows red X symbol when defibrillator is powered on

Ranked causes

  1. 1. Internal fault detected by Code Readiness system (common)
  2. 2. Defibrillator output circuit fault (occasional)
First check

Power cycle unit

Test point
Code Readiness indicator
Expected
Indicator clears and unit returns to normal operation

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Defibrillator will not charge high voltage

Defibrillator energy level does not increment on display when charging is initiated

Ranked causes

  1. 1. SHOCK button stuck in ON position (common)
  2. 2. Battery pack fully discharged (common)
First check

Check SHOCK button is not stuck

Test point
SHOCK button on paddles and front panel
Expected
Button releases properly; not stuck ON

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Heart rate display reads 0 despite pacing capture patient safety

ECG shows paced beats on display but heart rate counter shows 0

Ranked causes

  1. 1. Patient actually has no pulse (common)
  2. 2. Wrong ECG lead selected (occasional)
First check

Assess patient's actual perfusion status

Test point
Patient pulse and vital signs
Expected
Patient has palpable pulse

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Heart rate display reads 0 with proper pacing capture patient safety

Heart rate monitor shows zero while pacing is functioning and capture beats visible on ECG

Ranked causes

  1. 1. Patient cardiac output absent or very weak (common)
  2. 2. Suboptimal ECG lead configuration not detecting intrinsic beats (occasional)
First check

Assess patient cardiac status

Test point
Patient pulse and perfusion
Expected
Patient has cardiac output; pacing is functional

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inconsistent QRS beep or heart rate patient safety

QRS detection beep is irregular or heart rate display fluctuates abnormally

Ranked causes

  1. 1. Wrong lead selected (common)
  2. 2. Suboptimal electrode placement (common)
First check

Change ECG lead

Test point
Lead selector
Expected
QRS beep becomes consistent

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inconsistent QRS beep or heart rate display patient safety

QRS beep sound or heart rate measurement is inconsistent or unreliable

Ranked causes

  1. 1. Suboptimal ECG lead selection (common)
  2. 2. Poor electrode contact or placement (common)
First check

Change ECG lead selection

Test point
Lead selector
Expected
Consistent QRS beep and stable heart rate display

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Neither Battery nor AC Power indicator illuminated with device plugged in patient safety

Both battery and AC power status indicators are dark even with device connected to wall outlet

Ranked causes

  1. 1. AC connection issue or internal fault (common)
  2. 2. AC wall outlet or mains power failure (occasional)
First check

Power cycle AC connection

Test point
AC plug and connection
Expected
After disconnecting and reconnecting, indicators illuminate

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No stimulus marker present on ECG trace patient safety

Pacing stimulus marker does not appear on ECG display

Ranked causes

  1. 1. Unit not in PACER mode (common)
  2. 2. PACER RATE set lower than patient heart rate (common)
First check

Verify PACER mode is active

Test point
Mode indicator
Expected
Unit display shows PACER mode active

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No stimulus marker present on ECG trace during pacing patient safety

Pacing stimulus marker not visible on ECG display during pacing operation

Ranked causes

  1. 1. Device not in PACER mode (common)
  2. 2. Pacing rate not set above patient heart rate (common)
First check

Verify unit in PACER mode

Test point
Mode selector or display
Expected
PACER mode selected and active

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NOISY ECG RETRY ANALYSIS message patient safety

Unit displays message indicating ECG signal is too noisy for automatic analysis and user should retry

Ranked causes

  1. 1. Poor electrode adhesion or electrode condition (common)
  2. 2. Patient or external movement causing artifact (common)
First check

Check electrode application and adhesion

Test point
Electrode pads on patient skin
Expected
Electrodes properly applied and firmly adhered

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pads expired or need replacement patient safety

OneStep electrode expiration date has passed or pad condition is poor

Ranked causes

  1. 1. OneStep electrodes expired (common)
  2. 2. Electrode gel dried or aged (common)
First check

Inspect pad expiration date

Test point
OneStep electrode package
Expected
Current date is before expiration date

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Plethysmogram waveform not displayed

Monitor does not show plethysmogram waveform during SpO2 monitoring

Ranked causes

  1. 1. Sensor not properly positioned or making contact (common)
  2. 2. Display issue or configuration error (occasional)
First check

Verify Mode Selector is set to MONITOR

Test point
Mode Selector position
Expected
MONITOR

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Poor ECG signal amplitude with normal calibration

ECG signal is weak or small but calibration pulse is normal

Ranked causes

  1. 1. Wrong ECG lead selected (common)
  2. 2. Suboptimal electrode placement (common)
First check

Try alternate ECG lead

Test point
Lead selection
Expected
Signal amplitude increases

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Poor ECG signal amplitude with normal calibration pulse

ECG trace shows low amplitude signal although calibration pulse appears normal

Ranked causes

  1. 1. Suboptimal ECG lead selection for patient cardiac signal (common)
  2. 2. Poor electrode contact or placement (common)
First check

Select alternate ECG lead

Test point
Lead selection control
Expected
ECG signal amplitude increases

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power LED status incorrect caution

AC Power LED not illuminated when AC connected; Battery LED not showing correct charge state

Ranked causes

  1. 1. AC power supply failure or LED failure (occasional)
  2. 2. Battery not present or charging fault (common)
First check

Verify AC Power LED when AC connected and unit off

Test point
AC Power LED
Expected
Illuminated (steady)

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pulse rate discrepancy patient safety

Displayed pulse rate does not match manually measured pulse

Ranked causes

  1. 1. Sensor motion or poor contact (common)
  2. 2. Sensor placement on different finger or hand (occasional)
First check

Measure operator pulse manually

Test point
Radial or carotid artery
Expected
Stable pulse rate in beats per minute

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SELECT PADS message displayed patient safety

Unit displays message indicating hands-free pads must be attached for automatic analysis mode

Ranked causes

  1. 1. ECG analysis requires hands-free electrodes in Advisory mode (common)
  2. 2. Manual mode with paddles selected (occasional)
First check

Attach hands-free therapy electrodes if Advisory mode needed

Test point
Therapy electrode connectors
Expected
Hands-free electrodes connected and ready for ECG analysis

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SET CLOCK message displayed

Time and date information message appears on monitor

Ranked causes

  1. 1. Time and date need to be set (common)
  2. 2. Internal lithium battery needs replacement (occasional)
First check

Set clock date and time

Test point
Clock setting interface
Expected
Current date and time entered

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 sensor not detected patient safety

Unit displays 'CHECK SPO2 SENSOR' message and emits two beeps when sensor is removed or patient cable is unplugged

Ranked causes

  1. 1. SpO2 sensor disconnected from finger (common)
  2. 2. Patient cable unplugged from unit (common)
First check

Verify SpO2 sensor is attached to finger

Test point
Sensor contact with patient finger
Expected
Sensor firmly attached to finger

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Strip chart recording incorrect or missing data
  • Alarm suspend function not working
  • Alarm tone sounds when unit powered on but alarm not enabled
  • Bedside monitor ECG becomes erratic during pacing
  • Bedside/Central Station/Telemetry ECG display erratic during pacing
  • Calibration: Charge time verification at 200J
  • Calibration: NIBP transducer
  • Calibration: Power Supply Test
  • Calibration: Print output
  • Calibration: Shock energy delivery
  • Calibration: SpO2 Monitor
  • Charge time to 200 J exceeds 10 seconds
  • CHECK CPR PUCK message displayed
  • CHECK PADS alarm displays when pacing
  • Code Readiness indicator displays red X while unit off
  • Code Readiness indicator shows red X when unit is off
  • CPR Remaining Timer enabled but fails to decrement
  • DEFIB MAINT. REQ. message displayed
  • Defibrillator discharge test failure
  • Defibrillator does not discharge at pads
  • Defibrillator does not display ready state at 30J
  • ECG lead off alarm not triggered
  • ECG LEAD OFF detection not functioning on all leads
  • ECG LEAD OFF message does not display when lead disconnected
  • ECG lead selection not working
  • ECG signal not displaying
  • ECG waveform size not adjusting
  • Full preventive maintenance after any component replacement
  • Heart rate alarm not sounding on lead disconnect
  • Heart rate outside expected range
  • Internal discharge does not occur
  • LOW BATTERY message does not display at low threshold
  • No ECG trace or dashed line displayed
  • No ECG trace or dashed line on display
  • Pace stimulus marker spacing incorrect on recorder output
  • Pace stimulus markers not displayed on ECG
  • Pacer output amplitude out of specification
  • Pacer output not detected
  • Pacer output verification during preventive maintenance
  • Pacer pulse width out of specification
  • Pacer rate inaccurate
  • Persistent Red X in RFU window
  • Recorder makes stuttering sound when activated
  • Remote ECG signal input test
  • Remote synchronization cardioversion test
  • REPLACE BATTERY message does not display at critical voltage
  • SELECT 30J FOR TEST displayed
  • SELECT ASYNC PACE displayed
  • SELECT DEFIB MODE displayed
  • SELECT LIMB LEADS displayed
  • SELECT PADS ECG displayed
  • SET CLOCK displayed
  • SET PACE MA displayed
  • SHOCK ADVISED displayed
  • SpO2 alarm functionality verification
  • SPO2 AMBIENT LIGHT displayed
  • SPO2 COMM ERROR displayed
  • SPO2 PULSE SEARCH displayed
  • STAND CLEAR displayed
  • Strip chart paper not advancing
  • Summary Report will not print
  • Summary Report will not print when Report/Print Chart activated
  • SUMMARY TIMEOUT displayed
  • SYNC DEFIB DISABLED displayed
  • System current consumption exceeds specification
  • SYSTEM FAULT 210 displayed
  • SYSTEM FAULT 211 displayed
  • SYSTEM FAULT 212 displayed (Pace)
  • SYSTEM FAULT 212 displayed (User Interface)
  • SYSTEM FAULT 214 displayed
  • SYSTEM FAULT 215 displayed
  • SYSTEM FAULT 36 displayed
  • SYSTEM FAULT 37 displayed
  • SYSTEM FAULT 38 displayed
  • TEST FAILED 30J displayed
  • TRANSFERRING DATA displayed
  • Unit does not charge to specified energy level
  • Unit does not power on
  • Unit does not turn on at minimum voltage
  • Unit does not turn on at nominal voltage
  • Unit off-mode current consumption exceeds specification
  • USE PADDLE DISCHG displayed
  • USE PADS TO PACE displayed
  • USER SETUP REQ. displayed
  • VF ALARMS OFF displayed
  • X FAULT XX message displayed
  • ZERO CO2 ADAPTER displayed
  • ZERO DONE displayed
  • ZERO FAILED displayed
  • ZEROING CO2 ADAPTER displayed

90 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
200J MAX BIPHASICUser attempted to set defibrillation energy above 200J on biphasic unitUser input requesting energy level greater than 200J on biphasic defibrillatorinfo
30 J TEST OK30J defibrillator self-test passedUnit successfully completed 30J defib self-testinfo
30 J TEST OKUnit successfully passed 30J defibrillation self-testCompletion of 30J self-test procedureinfo
30J TEST OKDefibrillator successfully discharged 30J energy with measured parameters within specificationSHOCK button held after unit reached DEFIB 30J READY stateinfo
50 J MAXEnergy setting limited to 50J maximum for internal paddlesUser input requesting energy level greater than 50J with internal paddlesinfo
50 J MAXInternal paddle defibrillation energy limited to 50J maximumUser attempts to set energy <50J for internal paddlesinfo
ALARM SETStatus message indicating alarm configuration in progressUser selects alarm settingsinfo
ALARM SETAlarm settings are being configuredUser is adjusting alarm parameters on the unitinfo
ANALYSIS HALTEDECG analysis stopped during processingUser interaction or charging error during ECG analysiswarning
ASYNC PACE ONLYPD module cannot detect synchronization pulsesSystem attempting sync paced defibrillation with absent sync pulsesunknown
ATTACH PADSNo defibrillation pads attached to patient in AED auto modeAED power-up without padscritical
ATTACH PADSTherapy pads/electrodes not detected on system startupAED power-up in Auto Defib mode without pads attachedunknown
AUDIO QUEUE FULLAudio output queue cannot accept additional voice promptsMultiple voice prompts queued simultaneouslywarning
BATTERY COMMS ERRORBattery not communicating with host computerInitialization or battery status checkerror
BATTERY FAULTBattery hardware fault detectedBattery self-test or voltage monitoringcritical
BATTERY FAULTGeneral battery failure detectedBattery self-test or operationerror
BATTERY ID FAULTBattery identification chip failure or mismatchBattery ID verification during power-uperror
BATT HIGH CURRENTBattery not charged and current draw exceeds 1.6 ABattery discharge with excessive currenterror
BATT HIGH CURRENTBattery charging current exceeds 1.6 A when battery not chargedCharging operation when battery in low stateerror
BATT HIGH VOLTAGECharger output voltage is too highAC charger or battery monitoring during chargeerror
BATT LOW VOLTAGECharger output voltage is too lowAC charger or battery voltage monitoringerror
CAL. BARO. PRESSUREBarometric pressure sensor reading is out of acceptable rangeBarometric pressure measurement during operationerror
CALIBRATE NIBPNIBP (Non-Invasive Blood Pressure) calibration incomplete or failedNIBP system initializationunknown
CANNOT CHARGEDefibrillator charging circuit failureCharge button pressedcritical
CANNOT CHARGEUnit cannot perform charging when charge button pressedUser presses charge button but charging cannot proceederror
CF TRANSFER FAILEDCompactFlash data transfer error for Summary/DVCK/ALOG filesData transfer to or from CompactFlash carderror
CHANGE LEADSDefib Sync mode active with heart rate below 20 BPMPacing in Sync mode when heart rate drops below 20 BPMunknown
CHARGE FAILEDDefibrillator failed to reach requested energy levelCharge cycle failurecritical
CHARGE FAILEDRequested charge operation did not completeUser initiates charge but charging circuits failerror
CHECK CO2 ADAPTERCO2 airway adapter removed, occluded, or zeroing performed incorrectlyCO2 monitoring when adapter in abnormal stateunknown
CHECK CO2 SENSOREtCO2 sensor unplugged or defectiveCO2 monitoring initialization or during measurementunknown
CHECK CUFF/HOSEBlood pressure cuff or hose connection/function problemNIBP measurement attemptwarning
CHECK ECG CABLEECG cable identifier chip error or invalid cableECG cable connection detectionerror
CHECK ECG CABLEInvalid ECG cable ID detectedECG system initializationunknown
CHECK ELECTRODECannot read electrode ID chip or checksum failedElectrode verification at system startupunknown
CHECK ELECTRODEDefibrillation electrode identification or verification failureElectrode connection detectionerror
CHECK PADSPads have poor contact or short circuitImpedance measurement or defib delivery preparationunknown
CHECK PADSDefibrillation pads contact or connection problemPad impedance measurement or defib pad detectioncritical
CHECK PADS/POOR PAD CONTACTPad contact detection circuit detected missing or poorly seated padsDefibrillator selector switched to DEFIB mode with leads set to PADS and no pads detectedwarning
CHECK PATIENTBackground ECG analysis detects shockable rhythmContinuous background ECG monitoringunknown
CHECK PULSENo-shock advised or alternate message configurationECG analysis after shock or with auto-analyze enabledunknown
CHECK PULSENo shock advised based on ECG analysis; check for patient pulseECG analysis completioninfo
CHECK RECORDERPaper tray empty, paper jam, or recorder door openRecorder subsystem status checkunknown
CHECK SPO2 SENSORInvalid SpO2 reading due to sensor or application issueInappropriate sensor site, poor application, no patient cable plugged in, or motion artifactwarning
CHECK SPO2 SENSORSpO2 sensor not detected or disconnectedSensor removed from finger or patient cable unplugged from unitwarning
CHECK SPO2 SITEInsufficient perfusion at sensor siteBlood flow to sensor location is inadequate for measurementwarning
CLOCK BATTERY FAULTReal-time clock coin battery failureRTC boot-up or operationerror
CLOCK FAULT 11Real Time Clock oscillator failureRTC initialization during power-upcritical
CLOCK FAULT 12RTC backup power supply failure; oscillator not running at startupRTC power-uperror
CLOCK FAULT 13Real Time Clock time unit out of rangeRTC data validation at startuperror
CO2 COMM ERRORNo or invalid communication from EtCO2 moduleEtCO2 module communication attempterror
CO2 DEVICE NOT READYCO2 in airway adapter when attempting zero operationUser attempts to zero EtCO2 sensorunknown
CO2 IN LINE: WAITAdapter zero attempted with CO2 still in adapterUser initiates zero sequence while CO2 presentunknown
CO2 IN LINE: WAITCO2 detected in adapter during zeroing attemptAdapter zero attempted with CO2 presentwarning
CO2 MODULE NOT VALIDSidestream CO2 sensor connected (not supported)CO2 module identification during startuperror
CO2 OUT OF RANGECalculated CO2 value exceeds 150 mmHgCO2 measurement during breathingunknown
CO2 UNIT ERROREtCO2 sensor or module hardware error detectedSensor/module self-test or operationerror
CO2 WARM UPMainstream CO2 sensor warming up after connectionCO2 sensor initializationinfo
CPR FAULT 8ECG processor not receiving CPR measurement dataCPR compression detection attempterror
DATA TRANSFERREDData transfer to external device completed successfullyEnd of data transfer operationinfo
DEFIB 30J READYDefibrillator charged to 30J energy level and ready to dischargeUnit successfully charges after CHARGE button pressed at 30J energy settinginfo
DEFIB DISABLEDDefibrillation function disabled due to configuration or faultDefibrillator disable via configuration or simultaneous faultscritical
DEFIB FAULT 76PD Defib module failure during power-on self-testPOST executioncritical
DEFIB FAULT 77PD module defibrillation failure while runningDefibrillation attempt during operationcritical
DEFIB FAULT 78PD Defib functional safety error detected during operationDefib operationcritical
DEFIB FAULT 79PD module Defib/Pace failure during power-on self-testPOST executioncritical
DEFIB FAULT 80Undefined error received from PD moduleCommunication with PD modulecritical
DEFIB FAULT 94PD module reset 3 times without being requestedPD module watchdog timeout or crash conditioncritical
DEFIB FAULT 95PD module not communicating with host processorCommunication timeoutcritical
DEFIB FAULT 96PD module reported error on dischargeDefibrillation delivery attemptcritical
DEFIB MAINT. REQ.Defibrillator maintenance required after 5000+ discharges at 200JDischarge counter reaches 5000 discharges at 200Jwarning
DEFIB NOT CHARGEDDischarge button pressed but unit not chargedUser presses discharge in Defib mode without chargewarning
DEFIB NOT CHARGEDDefibrillator not charged when discharge attemptedDischarge button pressed in Defib mode but unit is not chargedwarning
DEFIB OVERUSEUnit overheated from excessive shocksMore than 50 shocks delivered in less than 20 minuteswarning
DEFIB OVERUSEMore than 50 defibrillator shocks delivered in less than 20 minutesShock delivery rate exceeds thresholdwarning
DEFIB PAD SHORTPad impedance indicates short circuitMeasured impedance between high voltage leads of MFC abnormalcritical
DEFIB PAD SHORTAbnormally low impedance measured between high voltage leadsImpedance measurement before/during defibrillationerror
DEFIB PAD SHORTPad connector short circuit detected during pad presence checkUniversal cable connected to test port with unit in DEFIB mode and leads set to PADSwarning
DISK FORMAT REQ.Problem accessing DOC (data) file on storage diskFile access error during data retrievalerror
DISK FORMAT REQ.Disk format error detectedAny problem with DOC file access occurrederror
DISK FORMAT REQ.Error detected during DOC file access operationDisk file system access during operationerror
ECG DISABLEDPersistent critical hardware failure on ECG moduleECG system initialization and diagnostic checkcritical
ECG DISABLEDECG monitoring disabled due to hardware failurePersistent critical hardware failure on ECG modulecritical
ECG FAULT 200No ECG applicable software availableECG software loading during startuperror
ECG FAULT 3ECG processor reset failureECG processor initializationerror
ECG FAULT 4Excessive number of missed samples from ECG processorECG signal sampling during monitoringerror
ECG FAULT 5ECG processor power-up initialization failureECG processor startuperror
ECG FAULT 6ECG processor hardware failureECG processor self-testcritical
ECG FAULT 7ECG processor is not responding to commandsCommunication with ECG processorerror
ECG LEAD OFFOne or more ECG leads disconnected with non-MFE leads selectedECG input validation during monitoringunknown
ECG LEAD OFFOne or more ECG leads not connected when non-MFE leads selectedLead connection detection with ECG inputwarning
ECG RA LEAD OFFECG right arm (RA) lead disconnectedECG lead continuity checkunknown
ECG TOO LARGENumber of large ECG intervals exceeds thresholdECG signal analysis during monitoringwarning
ECG TOO LARGENumber of ECG intervals exceeding amplitude thresholdECG analysiswarning
ECG V LEAD OFFECG precordial (V) lead disconnectedECG lead continuity checkunknown
ENABLE ETCO2EtCO2 feature not enabledUser attempts to zero EtCO2 sensor while feature disabledinfo
ENABLE ETCO2User attempts to zero EtCO2 sensor while feature is disabledUser initiates zero operation for disabled CO2 featurewarning
ENERGY INCREMENTEDDefib energy automatically increasedAfter shock 1 or 2 delivered with auto-escalation configuredinfo
ERASING REPORTUnit is erasing the selected report datainfo
ERASING REPORTReport data erasure in progressUser selected report data to eraseinfo
FULL DISCLS STOPPEDFull disclosure data recording stoppedFull disclosure data exceeds storage capacity (4 hours or 32MB)warning
IF NO PULSECPR period messageCPR period specified by AHA protocol reachedinfo
IF NO PULSEMessage displayed upon entered CPR period specified by AHA protocolinfo
INSERT CARDNo memory card detectedNo card detected during manual or semi-automatic modeserror
INSERT CARDNo card detected during manual or semi-automatic modesCF Card not detectedwarning
INT. DUMP OVERLOADExcessive internal discharges in short timeMore than 15 internal discharges in 5 minuteswarning
INVALID GAS TEMP: RETRYOperator attempted to set a gas temperature outside the Capnostat's operating rangewarning
LOW BATTERYLow batteryBattery charge below operational thresholdwarning
LOW BATTERYBattery voltage lowBattery voltage drops below acceptable levelcritical
NIBP ARTIFACTUnit is unable to detect systolic, diastolic or mean blood pressure due to excessive motion or vibrationPatient movement or vibration during measurementwarning
NIBP ARTIFACTBlood pressure measurement failed due to motion artifactExcessive motion or vibration detected during measurementwarning
NIBP COMM ERR 259NIBP communication error - framing/parity/fifo errorCommunication framing, parity, or FIFO error detectederror
NIBP COMM ERR 260NIBP communication error - invalid packetsInvalid packets received from NIBP moduleerror
NIBP COMM ERR 261NIBP device not responding to action requestNIBP device not taking requested actionerror
NIBP COMM ERR 262NIBP device no reply to commandNo response from NIBP device to commanderror
NIBP COMM ERR 262No reply from deviceNIBP module not respondingerror
NIBP FAULT 263NIBP module fault 90 receivedFault 90 reported by NIBP modulecritical
NIBP FAULT 264NIBP module fault 91 receivedFault 91 reported by NIBP modulecritical
NIBP FAULT 265NIBP module fault 97 receivedFault 97 reported by NIBP modulecritical
NIBP FAULT 266NIBP module fault 98 receivedFault 98 reported by NIBP modulecritical
NIBP FAULT 267NIBP module fault 99 receivedFault 99 reported by NIBP modulecritical
NIBP FAULT 268NIBP device no response after power upNIBP module not responding after power-upcritical
NIBP MEAS ABORTEDBlood pressure measurement was abortedCuff inflation pressure set too high, inflation too fast, unable to find systolic after 180s, defibrillator charging, or user aborterror
NIBP NOT READYNIBP module not ready for measurementDefibrillator charging or NIBP module performing power-up self-testinfo
NIBP OUT OF RANGEBlood pressure reading out of physiological rangeData from NIBP module is out of rangewarning
NIBP OUT OF RANGEData from the NIBP module is out of rangeNIBP measurement outside valid rangeerror
NIBP SIGNAL WEAKWeak or absent oscillometric signalWeak or no oscillometric signal detectedwarning
NO DATA TO TRANSFERReport data transfer errorSummary Report/Device Check/Activity Log data transfer errorerror
NOISY ECGExcessive noise in ECG signalNumber of noisy analysis intervals exceeds thresholdwarning
NOISY ECGNumber of noisy analysis intervals exceeds thresholdExcessive ECG noise during analysiswarning
NO QRS DETECTUnit is in sync mode and heart rate is below 20 BPM or QRS amplitude is too low for proper synchronizationLow heart rate or low ECG amplitude in sync modewarning
NO QRS DETECTQRS signal not detectedIn sync mode with heart rate < 20 BPM or QRS amplitude too lowwarning
NO SHOCK ADV.No shock advised messageAnalysis finds non-shockable rhythminfo
NO SHOCK ADV.No shock advised. Analysis finds non-shockable rhythmECG analysis detects non-shockable rhythminfo
OPEN AIR DISCHARGEMeasured Defib Impedance was greater than 1000 ohms and no energy was deliveredDefibrillation attempted with pads not in contact or open circuitwarning
OPEN AIR DISCHARGEDischarge impedance indicates open circuitMeasured defib impedance > 1000 ohms with no energy deliveredcritical
PACER DISABLEDPacing function disabledPacing disabled simultaneously with other pace faultserror
PACER FAULT 115PD pacing failure during power-on self-testPD pace failed during POSTcritical
PACER FAULT 117PD pacing functional safety failure during operationPD pace functional safety fault while runningcritical
PACER FAULT 121PD pacing failure during operationPD pace failed while runningcritical
PACER FAULT 122PD module Defib/Pace failure during power-on self-testPD module defib/pace failure during POSTcritical
PACER WARNING 124PD pacing warning during operationPD pace warning while runningwarning
PADDLE FAULTCannot detect paddle typeCannot detect type of accessory attached to universal cableerror
PADDLE FAULTCannot detect type of accessory attached to the universal cablePaddle type detection failureerror
PEDIATRIC PADS IN USEThe one of the Pediatric ID is missingPediatric identification missing during pediatric modewarning
PEDIATRIC PADS IN USEPediatric electrode pads activeOne of the Pediatric ID is missingwarning
PERFORM CPRMessage displayed upon entered CPR period specified by AHA protocolinfo
PERFORM CPRCPR in progress messageCPR period specified by AHA protocol enteredinfo
POOR LEAD CONTACTECG lead electrode poor or absent connectionOne or more ECG leads poorly connected or not connected to patientwarning
POOR PAD CONTACTElectrode impedance exceeds thresholdDefibrillation pads not adequately coupled to patient skinwarning
POOR PAD CONTACTDefibrillation pad impedance highElectrode impedance exceeds thresholdwarning
PRESS ANALYZEAlternate message for Check Patient message if this configuration option is selectedinfo
PRESS CHARGEUser prompt to charge defibrillatorDischarge button pressed before charge or advisory after SHOCK ADVISEDinfo
PRESS CHARGEDischarge button pressed in Manual Defib mode but unit is not charged, or advisory message following SHOCK ADVISEDButton pressed when unit not ready or auto-charge disabledinfo
PRESS SHOCKDeliver defibrillation promptUnit fully charged and ready to deliver shock in AED auto modeinfo
PRESS SHOCKPrompt issued in AED auto defib mode when defib is charged and readyinfo
READING LOG DATASummary report printing in progressUser requested log data printoutinfo
RECORDER FAULT 142Strip chart system errorPrinter system malfunction detectederror
RECORDER FAULT 143Printer failed power-up self-testStrip chart failed echo test during POSTerror
RECORDER FAULT 147Printer printhead overheatedPrinthead temperature exceeded safe operating limiterror
RELEASE BUTTONSSimultaneous external paddle button presses detected before unit reached full defib chargeBoth paddle buttons pressed before ready statewarning
RELEASE SHOCKDischarge switch interference detectedDischarge switch closed when pressing charge button or pressed before ready statewarning
RELEASE SHOCKDischarge switch closed when pressing charge button, or discharge button pressed before defib reached ready stateDischarge button activation during chargingwarning
REMOVE SYNCExit synchronization modeAnalyze button pressed or no heart rate detected in Defib Sync modeinfo
REMOVE SYNCAnalyze button pressed, or no heart rate detected while in Defib Sync modeAnalyze pressed in sync mode or no HR detectedinfo
REPEAT NIBP MEASBlood pressure measurement retry neededMaximum inflation attempts exceeded or 180-second time limit exceededwarning
REPLACE BATTERYBattery voltage is less than absolute minimumBattery voltage critically lowcritical
REPLACE BATTERYBattery completely dischargedBattery voltage less than absolute minimum; shutdown imminentcritical
REPORT FULLSummary report memory fullSummary report storage capacity reachedwarning
RETRY ANALYSISRerun rhythm analysisNoisy ECG detected during analysiswarning
RETRY ANALYSISAdvisory message in conjunction with noisy ECGECG analysis halted due to noiseinfo
RPT NO DISKInternal disk not formattedInternal disk not formatted and cannot be formattedcritical
SELECT 30J FOR TESTSelf test attempt at energy other than 30JAttempt to run self test at energy other than 30Jwarning
SELECT 30J FOR TESTSystem requests operator reduce energy level to 30J for safe internal discharge testingOperator selects 100J energy level and presses CHARGE buttoninfo
SELECT ASYNC PACEPD module cannot detect sync pulses; async pacing requiredContinuous sync mode operation when sync pulses unavailablewarning
SELECT DEFIB MODEAnalyze button pressed in non-defib modeAnalyze button pressed in pace or monitor modewarning
SELECT LIMB LEADSPaddles or augmented ECG leads selected when continuous analysis active or startedLead selection error during analysisinfo
SELECT LIMB LEADSWrong lead type selected during analysisPaddles or augmented ECG leads selected when continuous analysis active or startedwarning
SELECT PADSECG lead (non-MFE Pads) selected when ANALYZE pressedLead type mismatch during analysisinfo
SELECT PADS ECGNon-MFE pads selected during analysisECG lead (non-MFE Pads) selected when ANALYZE pressedwarning
SET CLOCKReal time clock failure with invalid date or timeRTC detects invalid date/time on power-up or operationwarning
SET PACE MAMultiple copy errors in pace module software or memoryCopy errors on first pace entry or after 10+ minutes in other modeerror
SHOCK ADVISEDAnalysis finds a shockable rhythmECG analysis detects shockable rhythminfo
SHOCK ADVISEDShockable rhythm detected during analysisAnalysis detects VF or VT rhythmcritical
SPO2 AMBIENT LIGHTExcessive ambient light detectedHigh ambient light at sensor site interfering with SpO2 measurementwarning
SPO2 COMM ERRORNo transmissions from SpO2 unit received or communication errorSpO2 module communication failureerror
SPO2 COMM ERRORNo communication from SpO2 moduleSystem cannot receive transmissions from SpO2 uniterror
SPO2 PULSE SEARCHSensor cannot detect a pulseNo pulse signal detected at sensor site during normal operationwarning
SPO2 PULSE SEARCHSpO2 module pulse search in progressSpO2 module actively searching for patient pulse signalinfo
STAND CLEARPatient rhythm is being analyzedAnalysis started or auto-analysis initiated per rescue protocolinfo
STAND CLEARPatient rhythm analysis in progress; keep clear of patientAnalyze button pressed or automatic analysis started during rescuewarning
SUMMARY TIMEOUTSummary report generation timeout errorSystem fails to generate or transmit summary report within timeout perioderror
SYNC DEFIB DISABLEDSync mode enabled when Analyze pressed in defib modeSync mode active when Analyze button pressed during defib operationwarning
SYSTEM FAULT 210ECG control task not activated for 500 msDigital board ECG task fails to activate within 500 ms cyclecritical
SYSTEM FAULT 211Defib task not activated for 500 msDigital board defib task fails to activate within 500 ms cyclecritical
SYSTEM FAULT 212Pace task not activated for 500 ms; or User Interface task not activated for 500 msDigital board task fails to activate within 500 ms cyclecritical
SYSTEM FAULT 213System task have not been activated for 500 ms – User InterfaceUI task timeoutcritical
SYSTEM FAULT 214Display task not activated for 500 msDigital board display task fails to activate within 500 ms cyclecritical
SYSTEM FAULT 215Reserved task not activated for 500 ms (not used)Digital board reserved task fails to activate within 500 ms cyclecritical
SYSTEM FAULT 36Filtered sum of all supply voltages is out of rangeVoltage measurement from filtered sum exceeds acceptable limitserror
SYSTEM FAULT 371/2 scale reference voltage is out of rangeReference voltage measurement exceeds acceptable toleranceerror
SYSTEM FAULT 38Failure to shutdown after shutdown order written to RTCSystem fails to power down when shutdown command is sent to Real Time Clockcritical
TEST FAILED30J defib self-test failedSelf-test procedure executed and detected failure in 30J defibrillation circuitcritical
TRANSFERRING DATAData transfer in progress (informational)System initiating or executing data transfer to external device or memoryinfo
USE PADDLE DISCHGDefibrillator is charged and front-panel discharge button pressed when external paddles or internal spoons with discharge buttons are connectedDischarge button pressed while paddles/spoons connected and unit chargedwarning
USE PADS TO PACEMFE accessory other than Pads detected in Pace modeNon-Pads accessory connected while in Pace modewarning
USER SETUP REQ.Both copies of configuration data are bad, or software with older configuration version was loadedConfiguration memory corruption or old software version installationerror
VF ALARMS OFFVF alarms disabled in incompatible mode or lead selectionAlarms enabled in Pace Mode or Continuous Analysis with non-standard lead, or Alarm button pressed with Heart Rate alarm disabledwarning
VF ALARMS OFFAlarms enabled in Pace Mode or when Continuous Analysis active or started in manual mode but current lead is not Pads/Lead I/II/III; or Alarm button pressed but Heart Rate alarm is disabledVF alarm function disabled in incompatible mode or alarm disabled statewarning
ZERO CO2 ADAPTERNegative CO2 detected. May be caused by sensor zeroed with CO2 in airway or optical blockageCO2 sensor reading below acceptable baseline during measurementwarning
ZERO CO2 ADAPTERNegative CO2 detected during capnography monitoringCO2 sensor detects negative (impossible) CO2 valuewarning
ZERO DONESensor/adapter zero operation completed successfully (informational)CO2 adapter zeroing procedure completes normallyinfo
ZERO FAILEDThe zero operation did not complete successfullyZero calibration procedure failed to complete or detect valid baselineerror
ZEROING CO2 ADAPTERCO2 adapter zeroing procedure in progress (informational)User initiates CO2 adapter zero or system auto-starts procedureinfo

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.