ZOLL R: troubleshooting and error codes
The ServiceForge knowledge base holds 150 documented symptoms for the ZOLL R, with 290 ranked causes, 307 diagnostic checks with test points and expected values, and 214 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Excessive artifact with paddles as ECG source
Paddles used as ECG source show excessive noise or artifact on display
Ranked causes
- 1. PADDLES not selected as ECG source (common)
- 2. Paddles not pressed firmly on patient (common)
- 3. Insufficient gel on paddles (common)
- 4. Paddle surface contaminated or dirty (common)
- 5. Debris between adult and pediatric paddle plate (occasional)
- 6. Paddle cable damaged (occasional)
- 7. Poor electrode contact; use ECG electrodes instead (common)
Verify PADDLES selected as ECG source
- Test point
- ECG source selector
- Expected
- PADDLES mode indicated on display
3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Energy does not discharge when SHOCK button pressed high voltage
SHOCK button activated but no therapeutic energy delivered to patient
Ranked causes
- 1. Charge timeout - over 60/120 seconds elapsed in Manual mode (common)
- 2. Device in Sync/Remote Sync mode with no QRS detected (occasional)
- 3. Energy selection changed during charge or after ready (occasional)
- 4. Device not fully charged when SHOCK pressed (common)
- 5. SHOCK button not held down long enough (occasional)
- 6. Pads or paddles not making good contact (common)
Check if charge timeout occurred
- Test point
- Manual mode timer and device display
- Expected
- Less than 60/120 seconds since charge ready
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive artifact when using paddles as ECG source
ECG signal shows high noise and artifact when defibrillation paddles are used for ECG acquisition
Ranked causes
- 1. Paddles not selected or in manual mode (common)
- 2. Insufficient paddle pressure on patient skin (common)
- 3. Inadequate conductive gel on paddles (common)
- 4. Dirty or contaminated paddle surfaces (occasional)
- 5. Damaged paddle cable (occasional)
- 6. Electrodes provide better signal than paddles (common)
Confirm paddles mode selected
- Test point
- Mode display or selector
- Expected
- PADDLES displayed as active ECG source
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK PADS or ATTACH PADS message for defibrillation patient safety
Unit displays message indicating therapy electrodes are not properly connected or are defective
Ranked causes
- 1. Therapy electrodes not connected to OneStep cable (common)
- 2. Dried out electrode gel (common)
- 3. Poor electrode-to-patient contact (common)
- 4. Defective OneStep cable (occasional)
- 5. Consider using paddles if electrode issues persist (occasional)
Disconnect and reconnect OneStep cable and electrodes
- Test point
- Cable and electrode connectors
- Expected
- Message clears after reconnection
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No 30J TEST OK message in manual self-test high voltage
Manual defibrillator self-test at 30 joules does not produce expected TEST OK message
Ranked causes
- 1. Energy not set to 30 joules (common)
- 2. OneStep electrode package not sealed (test with electrodes) (common)
- 3. OneStep cable not fully inserted into electrode connector (common)
- 4. OneStep cable not fully inserted into test port (test port method) (common)
- 5. Paddles not firmly pressed during discharge (paddle test) (common)
Verify energy set to 30 joules
- Test point
- Energy level selector
- Expected
- 30 joules selected
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ventricular capture beat after stimulus marker patient safety
Pacing stimulus present on ECG but no cardiac response or capture beat visible
Ranked causes
- 1. Patient's pulse already present (common)
- 2. Pacing output current too low (common)
- 3. Electrode-to-patient contact poor (common)
- 4. Suboptimal electrode placement (common)
- 5. Wrong ECG lead selected (occasional)
Assess patient's perfusion status
- Test point
- Patient vitals and pulse
- Expected
- Patient has adequate perfusion; capture needed
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ventricular capture beat after stimulus marker on ECG patient safety
Pacing stimulus marker present but no cardiac response (capture beat) appears on ECG
Ranked causes
- 1. No cardiac output or patient in cardiac arrest (common)
- 2. Output current insufficient for capture (common)
- 3. Poor electrode contact with patient (common)
- 4. Suboptimal electrode placement for pacing (occasional)
- 5. Suboptimal ECG lead configuration (occasional)
Check patient pulse and cardiac status
- Test point
- Patient assessment
- Expected
- Patient has palpable pulse or cardiac output evident
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Noisy ECG, artifact, wandering baseline
ECG display shows excessive noise, artifact or baseline drift affecting signal quality
Ranked causes
- 1. Poor electrode adhesion or skin preparation (common)
- 2. Electrode or cable movement creating artifact (common)
- 3. Patient movement (common)
- 4. Excessive radio frequency interference (occasional)
- 5. Inappropriate filter settings (occasional)
Inspect electrode adhesion and skin contact
- Test point
- Electrode pads and patient skin
- Expected
- Electrodes fully adhered with good skin contact
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sync marker absent or inconsistent with QRS patient safety
Synchronization marker not appearing or not aligned with QRS complex
Ranked causes
- 1. Device not in Sync mode (common)
- 2. Device in Remote Sync mode instead of Sync (occasional)
- 3. Wrong ECG lead selected (common)
- 4. Suboptimal electrode placement (common)
- 5. Recorder paper too narrow (rare)
Verify Sync mode enabled
- Test point
- Mode display
- Expected
- Sync mode active (not Remote Sync)
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sync marker absent or inconsistent with QRS waveform patient safety
Synchronization marker not appearing or not aligned with QRS complex on display and recorder
Ranked causes
- 1. Device not in Sync mode (common)
- 2. Device in Remote Sync mode when Sync mode intended (occasional)
- 3. Suboptimal ECG lead selection (common)
- 4. Poor electrode contact or placement (common)
- 5. Recorder paper too narrow (rare)
Verify Sync mode is selected
- Test point
- Mode selector or display
- Expected
- SYNC mode displayed; Sync marker visible on QRS
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to deliver SHOCK in Sync mode high voltage
Cannot deliver energy to patient while in Sync (synchronous cardioversion) mode
Ranked causes
- 1. SYNC selection not displayed (common)
- 2. No ECG signal displayed (common)
- 3. Sync markers not present on QRS (common)
- 4. Wrong electrode placement (common)
- 5. SHOCK button not held long enough (common)
Verify SYNC mode is active
- Test point
- Monitor display
- Expected
- SYNC XXXJ SEL displayed
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to SHOCK when in Sync mode high voltage
Defibrillator will not deliver synchronized shock despite Sync mode active
Ranked causes
- 1. No SYNC XXXJ SEL display indicating sync not ready (common)
- 2. No ECG signals displayed (common)
- 3. No Sync markers (arrows) above R waves (common)
- 4. SHOCK button not held down long enough (occasional)
- 5. Suboptimal electrode placement (occasional)
Verify SYNC XXXJ SEL displayed
- Test point
- Monitor display status line
- Expected
- SYNC XXXJ SEL visible on display
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery indicator alternating green and yellow
Battery status indicator light cycles between green and yellow colors
Ranked causes
- 1. Battery not installed or properly seated (common)
- 2. Battery fault condition detected (common)
- 3. Battery depleted or defective (common)
- 4. Charger or charging system fault (occasional)
Verify battery is installed
- Test point
- Battery compartment
- Expected
- Battery pack properly installed and seated
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK PADS message displayed patient safety
Message indicates therapy electrode connection or integrity problem
Ranked causes
- 1. Therapy electrode not connected to OneStep cable (common)
- 2. Electrode gel is dry (common)
- 3. Electrode-to-patient contact is poor (common)
- 4. OneStep cable is defective (occasional)
Verify electrode to OneStep cable connection
- Test point
- Therapy electrode connector to OneStep cable
- Expected
- Firm connection between electrode and cable
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK PADS message for pacing patient safety
Unit displays CHECK PADS message when pacing mode is selected
Ranked causes
- 1. Therapy electrodes not connected to OneStep cable (common)
- 2. Electrode gel dried out or aged beyond specification (common)
- 3. Poor electrode-to-patient contact (common)
- 4. Defective OneStep cable (occasional)
Verify therapy electrode connection to OneStep cable
- Test point
- OneStep cable connector to electrodes
- Expected
- Electrodes firmly connected to cable
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK PADS or ATTACH PADS message patient safety
Message indicating hands-free therapy electrode connection problem
Ranked causes
- 1. Pads not properly connected or seated (common)
- 2. Electrode contact or gel condition poor (common)
- 3. OneStep cable is defective (occasional)
- 4. Pad/electrode set is defective (occasional)
Disconnect and reconnect OneStep cable
- Test point
- OneStep cable connection
- Expected
- Cable firmly seated on both electrodes and device
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG LEAD OFF message displayed patient safety
Unit displays message indicating ECG cable or electrodes are not connected or making contact
Ranked causes
- 1. ECG cable or electrodes not connected or making poor contact (common)
- 2. Dried out or degraded electrodes (common)
- 3. Cable change requiring reset (occasional)
- 4. Defective cable (occasional)
Verify ECG cable and electrode connections
- Test point
- Cable connectors and patient electrode pads
- Expected
- Cable firmly connected to unit and patient, electrodes in contact with skin
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LOW BATTERY message appears with AC power connected
LOW BATTERY message displayed even though unit is plugged into AC mains
Ranked causes
- 1. AC Power Indicator not illuminated (common)
- 2. Battery depleted or defective (common)
- 3. Charger connection intermittent (occasional)
- 4. AC mains not functional (occasional)
Verify AC Power Indicator illuminated
- Test point
- AC Power Indicator light
- Expected
- AC Power Indicator lit
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No 30J TEST OK message during manual defibrillator self-test high voltage
Defibrillator self-test at 30 joules fails to display TEST OK message
Ranked causes
- 1. Device not set to 30 joules (common)
- 2. OneStep electrode not properly connected or package sealed (common)
- 3. Test Port connector not firmly inserted (occasional)
- 4. Paddles not firmly pressed during paddle test (occasional)
Verify device set to 30 joules
- Test point
- Energy selector display
- Expected
- 30 joules displayed
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No apparent energy delivery to patient high voltage
No visible patient reaction when defibrillator shock is delivered, or energy delivery uncertain
Ranked causes
- 1. Some patients will not show physical reaction to energy (common)
- 2. Device malfunction or insufficient energy output (occasional)
- 3. Poor pad contact or CHECK PADS message (common)
- 4. Poor pad placement or adhesion with hands-free pads (occasional)
Review discharge stripchart for energy confirmation
- Test point
- Stripchart printout showing Joules delivered
- Expected
- Energy delivery recorded (e.g., 200J delivered)
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
POOR LEAD CONTACT message displayed patient safety
Unit indicates ECG electrodes or pacing cable not making adequate contact with patient
Ranked causes
- 1. Electrode or cable not connected or making poor skin contact (common)
- 2. Dried out or degraded electrodes (common)
- 3. Cable type change requiring reset (occasional)
- 4. Defective cable (occasional)
Verify cable and electrode connections
- Test point
- Cable connectors and electrode pads
- Expected
- Cable firmly connected, electrodes in good contact with patient skin
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Standby pacing patient gets paced intermittently patient safety
Patient on standby pacing mode is being paced sporadically instead of continuously when needed
Ranked causes
- 1. ECG lead connection or placement problem (common)
- 2. ECG cable damaged (occasional)
- 3. Patient's R-to-R interval varying (common)
- 4. Pacing rate not set appropriately (common)
Verify ECG/pacing cable connection
- Test point
- Cable connectors
- Expected
- Cable firmly seated and connected
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to deliver SHOCK in Remote Sync mode high voltage
Cannot deliver energy to patient while in Remote Sync (external synchronization) mode
Ranked causes
- 1. REMOTE SYNC not properly displayed (common)
- 2. Remote device not properly connected (common)
- 3. Remote device not sending proper Sync markers (common)
- 4. SHOCK button not held long enough (common)
Verify REMOTE SYNC mode active
- Test point
- Monitor display
- Expected
- REMOTE SYNC shown in place of ECG; REMOTE SYNC XXXJ SEL displayed
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to SHOCK when in Remote Sync mode high voltage
Defibrillator will not deliver energy in Remote Sync mode despite remote device connected
Ranked causes
- 1. REMOTE SYNC not displayed or not set up (common)
- 2. Remote device not properly connected or not conforming to specs (occasional)
- 3. Sync markers not present or not synchronized with R waves on remote (occasional)
- 4. SHOCK button not held down properly (occasional)
Verify REMOTE SYNC displayed
- Test point
- Device display status
- Expected
- REMOTE SYNC shown and REMOTE SYNC XXXJ SEL visible
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery indicator alternately illuminating green and yellow patient safety
Battery status indicator shows alternating green and yellow lights indicating abnormal charging state
Ranked causes
- 1. Battery pack not installed or not recognized (common)
- 2. Battery pack discharged or defective (common)
- 3. Charging circuit malfunction (occasional)
Verify battery pack installed
- Test point
- Battery compartment
- Expected
- Battery pack present and properly seated
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery LED not showing proper status caution
Battery LED is flashing or not displaying solid green or amber, indicating battery fault or seating issue
Ranked causes
- 1. Battery not properly seated in well (common)
- 2. Battery displaying fault indicator (occasional)
- 3. Damaged or dirty battery pins and contacts (occasional)
Check battery LED status
- Test point
- Battery LED on unit
- Expected
- Solid green or solid amber LED
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Impedance after board replacement patient safety
After replacing digital board, analog board, or PD engine, impedance calibration is required to ensure proper defibrillation delivery
Ranked causes
- 1. Digital board replaced (unknown)
- 2. Analog board replaced (unknown)
- 3. PD engine replaced (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK RECORDER message displayed
Unit displays message indicating recorder malfunction or paper issue
Ranked causes
- 1. Recorder out of paper (common)
- 2. Recorder door open (common)
- 3. Paper loaded upside down or incorrectly (occasional)
Check paper level in recorder
- Test point
- Recorder paper cartridge or tray
- Expected
- Paper present and not empty
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code Readiness indicator shows red X when unit is powered on patient safety
Red X displayed on Code Readiness indicator while defibrillator is running
Ranked causes
- 1. Code readiness fault during operation (common)
- 2. Defibrillator self-test failure (occasional)
- 3. Persistent code readiness fault (rare)
Power cycle unit
- Test point
- Power control
- Expected
- Unit powers off and back on
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Defibrillation charging or delivery not working patient safety
CHARGE button does not charge unit to selected energy; SHOCK button does not deliver energy; energy selection not incrementing correctly
Ranked causes
- 1. Energy selection circuit malfunction (occasional)
- 2. Charge circuit malfunction (occasional)
- 3. SHOCK button not making contact or delivery circuit failure (rare)
Test energy selection decrements correctly
- Test point
- Display energy value and ENERGY SELECT buttons
- Expected
- Energy decrements from selected value to 1J when down arrow held; increments sequentially when up arrow pressed
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light or poor quality printing
Recorded output is too faint or print quality is degraded
Ranked causes
- 1. Wrong paper type installed (common)
- 2. Paper installed backwards (common)
- 3. Recorder print head requires cleaning (occasional)
Verify paper type
- Test point
- Recorder paper
- Expected
- Correct manufacturer-approved paper type
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light or poor quality printing from recorder
Recorder output is faint or low contrast; printout is difficult to read
Ranked causes
- 1. Incorrect or deteriorated paper type (common)
- 2. Paper installed upside down (common)
- 3. Print head requires cleaning (occasional)
Verify correct paper type installed
- Test point
- Paper specification and packaging
- Expected
- Correct paper type for device installed
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LOW BATTERY message while plugged into AC mains patient safety
Unit displays low battery warning even though device is connected to AC power
Ranked causes
- 1. AC Power Indicator not illuminated; AC not connected properly (common)
- 2. Battery pack discharged or defective (common)
- 3. AC power cord or wall outlet not working (occasional)
Verify AC Power Indicator illuminated
- Test point
- AC Power Indicator light
- Expected
- AC Power Indicator lit (green or appropriate color)
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient on standby pacing gets paced intermittently patient safety
Device unexpectedly delivers pacing pulses when in standby mode or at inappropriate times
Ranked causes
- 1. ECG lead wire disconnected causing asynchronous pacing (common)
- 2. Damaged ECG cable (occasional)
- 3. Patient R-to-R interval varying close to pacing rate (occasional)
Verify ECG electrode connection and placement
- Test point
- Electrode pads and cable connectors
- Expected
- All electrodes firmly connected; good skin contact
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply verification after board or supply replacement high voltage
Power supply testing required after replacing digital board, AC power supply, or battery interconnect board to verify proper operation
Ranked causes
- 1. Digital board replaced (unknown)
- 2. AC power supply replaced (unknown)
- 3. Battery interconnect board replaced (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 Monitor not displaying reading patient safety
Monitor shows dashed line or no SpO2 value after simulator connection
Ranked causes
- 1. Simulator not enabled or in wrong mode (common)
- 2. Cable or sensor connection loose or damaged (common)
- 3. Display update delay (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not turn on or unexpectedly shuts off patient safety
Defibrillator fails to power on or powers off without user action
Ranked causes
- 1. Battery pack not properly installed or discharged (common)
- 2. Unit not connected to AC power (common)
- 3. Battery pack fully discharged (occasional)
Verify battery pack installation
- Test point
- Battery compartment
- Expected
- Battery pack fully inserted and locked
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wi-Fi card does not transmit Full Disclosure File patient safety
R Series fails to send Full Disclosure File via Wi-Fi; error message displayed or transmission unsuccessful
Ranked causes
- 1. Wi-Fi card not configured correctly for network connection (common)
- 2. Wi-Fi access point not available or out of range (occasional)
- 3. Defective Wi-Fi card (occasional)
Turn R Series to On or Monitor mode
- Test point
- Device power mode
- Expected
- Device in On or Monitor mode
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wi-Fi card fails to transmit device history
Device History transmission to ZOLL server unsuccessful; error message displayed instead of 'Defib History has sent'
Ranked causes
- 1. Wi-Fi card not configured correctly for network connection (common)
- 2. Wi-Fi access point unavailable (occasional)
- 3. Defective Wi-Fi card (occasional)
Verify Wi-Fi card configuration matches intended transmission type
- Test point
- R Series Print Config menu
- Expected
- Configuration matches Full Disclosure File, Device History, or both as intended
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery and AC Power indicators not illuminated when plugged in
Neither Battery nor AC Power indicator lights up when device is connected to AC mains
Ranked causes
- 1. AC power connection problem (common)
- 2. AC mains not functional (occasional)
Disconnect and reconnect AC power
- Test point
- AC power connection
- Expected
- Device unplugged and replugged; indicators illuminate
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code Readiness indicator displays red X while powered on patient safety
Code readiness LED shows red X symbol when defibrillator unit is powered on and running.
Ranked causes
- 1. Code readiness fault active during operation (unknown)
- 2. Unit continues to fail after troubleshooting (occasional)
Power cycle unit
- Test point
- Power switch
- Expected
- Unit boots up, readiness indicator becomes green or shows status message
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code Readiness indicator displays red X while unit is on patient safety
Code Readiness indicator shows red X symbol when defibrillator is powered on
Ranked causes
- 1. Internal fault detected by Code Readiness system (common)
- 2. Defibrillator output circuit fault (occasional)
Power cycle unit
- Test point
- Code Readiness indicator
- Expected
- Indicator clears and unit returns to normal operation
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Defibrillator will not charge high voltage
Defibrillator energy level does not increment on display when charging is initiated
Ranked causes
- 1. SHOCK button stuck in ON position (common)
- 2. Battery pack fully discharged (common)
Check SHOCK button is not stuck
- Test point
- SHOCK button on paddles and front panel
- Expected
- Button releases properly; not stuck ON
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heart rate display reads 0 despite pacing capture patient safety
ECG shows paced beats on display but heart rate counter shows 0
Ranked causes
- 1. Patient actually has no pulse (common)
- 2. Wrong ECG lead selected (occasional)
Assess patient's actual perfusion status
- Test point
- Patient pulse and vital signs
- Expected
- Patient has palpable pulse
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heart rate display reads 0 with proper pacing capture patient safety
Heart rate monitor shows zero while pacing is functioning and capture beats visible on ECG
Ranked causes
- 1. Patient cardiac output absent or very weak (common)
- 2. Suboptimal ECG lead configuration not detecting intrinsic beats (occasional)
Assess patient cardiac status
- Test point
- Patient pulse and perfusion
- Expected
- Patient has cardiac output; pacing is functional
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inconsistent QRS beep or heart rate patient safety
QRS detection beep is irregular or heart rate display fluctuates abnormally
Ranked causes
- 1. Wrong lead selected (common)
- 2. Suboptimal electrode placement (common)
Change ECG lead
- Test point
- Lead selector
- Expected
- QRS beep becomes consistent
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inconsistent QRS beep or heart rate display patient safety
QRS beep sound or heart rate measurement is inconsistent or unreliable
Ranked causes
- 1. Suboptimal ECG lead selection (common)
- 2. Poor electrode contact or placement (common)
Change ECG lead selection
- Test point
- Lead selector
- Expected
- Consistent QRS beep and stable heart rate display
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Neither Battery nor AC Power indicator illuminated with device plugged in patient safety
Both battery and AC power status indicators are dark even with device connected to wall outlet
Ranked causes
- 1. AC connection issue or internal fault (common)
- 2. AC wall outlet or mains power failure (occasional)
Power cycle AC connection
- Test point
- AC plug and connection
- Expected
- After disconnecting and reconnecting, indicators illuminate
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No stimulus marker present on ECG trace patient safety
Pacing stimulus marker does not appear on ECG display
Ranked causes
- 1. Unit not in PACER mode (common)
- 2. PACER RATE set lower than patient heart rate (common)
Verify PACER mode is active
- Test point
- Mode indicator
- Expected
- Unit display shows PACER mode active
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No stimulus marker present on ECG trace during pacing patient safety
Pacing stimulus marker not visible on ECG display during pacing operation
Ranked causes
- 1. Device not in PACER mode (common)
- 2. Pacing rate not set above patient heart rate (common)
Verify unit in PACER mode
- Test point
- Mode selector or display
- Expected
- PACER mode selected and active
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NOISY ECG RETRY ANALYSIS message patient safety
Unit displays message indicating ECG signal is too noisy for automatic analysis and user should retry
Ranked causes
- 1. Poor electrode adhesion or electrode condition (common)
- 2. Patient or external movement causing artifact (common)
Check electrode application and adhesion
- Test point
- Electrode pads on patient skin
- Expected
- Electrodes properly applied and firmly adhered
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pads expired or need replacement patient safety
OneStep electrode expiration date has passed or pad condition is poor
Ranked causes
- 1. OneStep electrodes expired (common)
- 2. Electrode gel dried or aged (common)
Inspect pad expiration date
- Test point
- OneStep electrode package
- Expected
- Current date is before expiration date
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Plethysmogram waveform not displayed
Monitor does not show plethysmogram waveform during SpO2 monitoring
Ranked causes
- 1. Sensor not properly positioned or making contact (common)
- 2. Display issue or configuration error (occasional)
Verify Mode Selector is set to MONITOR
- Test point
- Mode Selector position
- Expected
- MONITOR
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor ECG signal amplitude with normal calibration
ECG signal is weak or small but calibration pulse is normal
Ranked causes
- 1. Wrong ECG lead selected (common)
- 2. Suboptimal electrode placement (common)
Try alternate ECG lead
- Test point
- Lead selection
- Expected
- Signal amplitude increases
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor ECG signal amplitude with normal calibration pulse
ECG trace shows low amplitude signal although calibration pulse appears normal
Ranked causes
- 1. Suboptimal ECG lead selection for patient cardiac signal (common)
- 2. Poor electrode contact or placement (common)
Select alternate ECG lead
- Test point
- Lead selection control
- Expected
- ECG signal amplitude increases
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power LED status incorrect caution
AC Power LED not illuminated when AC connected; Battery LED not showing correct charge state
Ranked causes
- 1. AC power supply failure or LED failure (occasional)
- 2. Battery not present or charging fault (common)
Verify AC Power LED when AC connected and unit off
- Test point
- AC Power LED
- Expected
- Illuminated (steady)
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulse rate discrepancy patient safety
Displayed pulse rate does not match manually measured pulse
Ranked causes
- 1. Sensor motion or poor contact (common)
- 2. Sensor placement on different finger or hand (occasional)
Measure operator pulse manually
- Test point
- Radial or carotid artery
- Expected
- Stable pulse rate in beats per minute
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SELECT PADS message displayed patient safety
Unit displays message indicating hands-free pads must be attached for automatic analysis mode
Ranked causes
- 1. ECG analysis requires hands-free electrodes in Advisory mode (common)
- 2. Manual mode with paddles selected (occasional)
Attach hands-free therapy electrodes if Advisory mode needed
- Test point
- Therapy electrode connectors
- Expected
- Hands-free electrodes connected and ready for ECG analysis
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SET CLOCK message displayed
Time and date information message appears on monitor
Ranked causes
- 1. Time and date need to be set (common)
- 2. Internal lithium battery needs replacement (occasional)
Set clock date and time
- Test point
- Clock setting interface
- Expected
- Current date and time entered
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 sensor not detected patient safety
Unit displays 'CHECK SPO2 SENSOR' message and emits two beeps when sensor is removed or patient cable is unplugged
Ranked causes
- 1. SpO2 sensor disconnected from finger (common)
- 2. Patient cable unplugged from unit (common)
Verify SpO2 sensor is attached to finger
- Test point
- Sensor contact with patient finger
- Expected
- Sensor firmly attached to finger
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Strip chart recording incorrect or missing data
- Alarm suspend function not working
- Alarm tone sounds when unit powered on but alarm not enabled
- Bedside monitor ECG becomes erratic during pacing
- Bedside/Central Station/Telemetry ECG display erratic during pacing
- Calibration: Charge time verification at 200J
- Calibration: NIBP transducer
- Calibration: Power Supply Test
- Calibration: Print output
- Calibration: Shock energy delivery
- Calibration: SpO2 Monitor
- Charge time to 200 J exceeds 10 seconds
- CHECK CPR PUCK message displayed
- CHECK PADS alarm displays when pacing
- Code Readiness indicator displays red X while unit off
- Code Readiness indicator shows red X when unit is off
- CPR Remaining Timer enabled but fails to decrement
- DEFIB MAINT. REQ. message displayed
- Defibrillator discharge test failure
- Defibrillator does not discharge at pads
- Defibrillator does not display ready state at 30J
- ECG lead off alarm not triggered
- ECG LEAD OFF detection not functioning on all leads
- ECG LEAD OFF message does not display when lead disconnected
- ECG lead selection not working
- ECG signal not displaying
- ECG waveform size not adjusting
- Full preventive maintenance after any component replacement
- Heart rate alarm not sounding on lead disconnect
- Heart rate outside expected range
- Internal discharge does not occur
- LOW BATTERY message does not display at low threshold
- No ECG trace or dashed line displayed
- No ECG trace or dashed line on display
- Pace stimulus marker spacing incorrect on recorder output
- Pace stimulus markers not displayed on ECG
- Pacer output amplitude out of specification
- Pacer output not detected
- Pacer output verification during preventive maintenance
- Pacer pulse width out of specification
- Pacer rate inaccurate
- Persistent Red X in RFU window
- Recorder makes stuttering sound when activated
- Remote ECG signal input test
- Remote synchronization cardioversion test
- REPLACE BATTERY message does not display at critical voltage
- SELECT 30J FOR TEST displayed
- SELECT ASYNC PACE displayed
- SELECT DEFIB MODE displayed
- SELECT LIMB LEADS displayed
- SELECT PADS ECG displayed
- SET CLOCK displayed
- SET PACE MA displayed
- SHOCK ADVISED displayed
- SpO2 alarm functionality verification
- SPO2 AMBIENT LIGHT displayed
- SPO2 COMM ERROR displayed
- SPO2 PULSE SEARCH displayed
- STAND CLEAR displayed
- Strip chart paper not advancing
- Summary Report will not print
- Summary Report will not print when Report/Print Chart activated
- SUMMARY TIMEOUT displayed
- SYNC DEFIB DISABLED displayed
- System current consumption exceeds specification
- SYSTEM FAULT 210 displayed
- SYSTEM FAULT 211 displayed
- SYSTEM FAULT 212 displayed (Pace)
- SYSTEM FAULT 212 displayed (User Interface)
- SYSTEM FAULT 214 displayed
- SYSTEM FAULT 215 displayed
- SYSTEM FAULT 36 displayed
- SYSTEM FAULT 37 displayed
- SYSTEM FAULT 38 displayed
- TEST FAILED 30J displayed
- TRANSFERRING DATA displayed
- Unit does not charge to specified energy level
- Unit does not power on
- Unit does not turn on at minimum voltage
- Unit does not turn on at nominal voltage
- Unit off-mode current consumption exceeds specification
- USE PADDLE DISCHG displayed
- USE PADS TO PACE displayed
- USER SETUP REQ. displayed
- VF ALARMS OFF displayed
- X FAULT XX message displayed
- ZERO CO2 ADAPTER displayed
- ZERO DONE displayed
- ZERO FAILED displayed
- ZEROING CO2 ADAPTER displayed
90 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
200J MAX BIPHASIC | User attempted to set defibrillation energy above 200J on biphasic unit | User input requesting energy level greater than 200J on biphasic defibrillator | info |
30 J TEST OK | 30J defibrillator self-test passed | Unit successfully completed 30J defib self-test | info |
30 J TEST OK | Unit successfully passed 30J defibrillation self-test | Completion of 30J self-test procedure | info |
30J TEST OK | Defibrillator successfully discharged 30J energy with measured parameters within specification | SHOCK button held after unit reached DEFIB 30J READY state | info |
50 J MAX | Energy setting limited to 50J maximum for internal paddles | User input requesting energy level greater than 50J with internal paddles | info |
50 J MAX | Internal paddle defibrillation energy limited to 50J maximum | User attempts to set energy <50J for internal paddles | info |
ALARM SET | Status message indicating alarm configuration in progress | User selects alarm settings | info |
ALARM SET | Alarm settings are being configured | User is adjusting alarm parameters on the unit | info |
ANALYSIS HALTED | ECG analysis stopped during processing | User interaction or charging error during ECG analysis | warning |
ASYNC PACE ONLY | PD module cannot detect synchronization pulses | System attempting sync paced defibrillation with absent sync pulses | unknown |
ATTACH PADS | No defibrillation pads attached to patient in AED auto mode | AED power-up without pads | critical |
ATTACH PADS | Therapy pads/electrodes not detected on system startup | AED power-up in Auto Defib mode without pads attached | unknown |
AUDIO QUEUE FULL | Audio output queue cannot accept additional voice prompts | Multiple voice prompts queued simultaneously | warning |
BATTERY COMMS ERROR | Battery not communicating with host computer | Initialization or battery status check | error |
BATTERY FAULT | Battery hardware fault detected | Battery self-test or voltage monitoring | critical |
BATTERY FAULT | General battery failure detected | Battery self-test or operation | error |
BATTERY ID FAULT | Battery identification chip failure or mismatch | Battery ID verification during power-up | error |
BATT HIGH CURRENT | Battery not charged and current draw exceeds 1.6 A | Battery discharge with excessive current | error |
BATT HIGH CURRENT | Battery charging current exceeds 1.6 A when battery not charged | Charging operation when battery in low state | error |
BATT HIGH VOLTAGE | Charger output voltage is too high | AC charger or battery monitoring during charge | error |
BATT LOW VOLTAGE | Charger output voltage is too low | AC charger or battery voltage monitoring | error |
CAL. BARO. PRESSURE | Barometric pressure sensor reading is out of acceptable range | Barometric pressure measurement during operation | error |
CALIBRATE NIBP | NIBP (Non-Invasive Blood Pressure) calibration incomplete or failed | NIBP system initialization | unknown |
CANNOT CHARGE | Defibrillator charging circuit failure | Charge button pressed | critical |
CANNOT CHARGE | Unit cannot perform charging when charge button pressed | User presses charge button but charging cannot proceed | error |
CF TRANSFER FAILED | CompactFlash data transfer error for Summary/DVCK/ALOG files | Data transfer to or from CompactFlash card | error |
CHANGE LEADS | Defib Sync mode active with heart rate below 20 BPM | Pacing in Sync mode when heart rate drops below 20 BPM | unknown |
CHARGE FAILED | Defibrillator failed to reach requested energy level | Charge cycle failure | critical |
CHARGE FAILED | Requested charge operation did not complete | User initiates charge but charging circuits fail | error |
CHECK CO2 ADAPTER | CO2 airway adapter removed, occluded, or zeroing performed incorrectly | CO2 monitoring when adapter in abnormal state | unknown |
CHECK CO2 SENSOR | EtCO2 sensor unplugged or defective | CO2 monitoring initialization or during measurement | unknown |
CHECK CUFF/HOSE | Blood pressure cuff or hose connection/function problem | NIBP measurement attempt | warning |
CHECK ECG CABLE | ECG cable identifier chip error or invalid cable | ECG cable connection detection | error |
CHECK ECG CABLE | Invalid ECG cable ID detected | ECG system initialization | unknown |
CHECK ELECTRODE | Cannot read electrode ID chip or checksum failed | Electrode verification at system startup | unknown |
CHECK ELECTRODE | Defibrillation electrode identification or verification failure | Electrode connection detection | error |
CHECK PADS | Pads have poor contact or short circuit | Impedance measurement or defib delivery preparation | unknown |
CHECK PADS | Defibrillation pads contact or connection problem | Pad impedance measurement or defib pad detection | critical |
CHECK PADS/POOR PAD CONTACT | Pad contact detection circuit detected missing or poorly seated pads | Defibrillator selector switched to DEFIB mode with leads set to PADS and no pads detected | warning |
CHECK PATIENT | Background ECG analysis detects shockable rhythm | Continuous background ECG monitoring | unknown |
CHECK PULSE | No-shock advised or alternate message configuration | ECG analysis after shock or with auto-analyze enabled | unknown |
CHECK PULSE | No shock advised based on ECG analysis; check for patient pulse | ECG analysis completion | info |
CHECK RECORDER | Paper tray empty, paper jam, or recorder door open | Recorder subsystem status check | unknown |
CHECK SPO2 SENSOR | Invalid SpO2 reading due to sensor or application issue | Inappropriate sensor site, poor application, no patient cable plugged in, or motion artifact | warning |
CHECK SPO2 SENSOR | SpO2 sensor not detected or disconnected | Sensor removed from finger or patient cable unplugged from unit | warning |
CHECK SPO2 SITE | Insufficient perfusion at sensor site | Blood flow to sensor location is inadequate for measurement | warning |
CLOCK BATTERY FAULT | Real-time clock coin battery failure | RTC boot-up or operation | error |
CLOCK FAULT 11 | Real Time Clock oscillator failure | RTC initialization during power-up | critical |
CLOCK FAULT 12 | RTC backup power supply failure; oscillator not running at startup | RTC power-up | error |
CLOCK FAULT 13 | Real Time Clock time unit out of range | RTC data validation at startup | error |
CO2 COMM ERROR | No or invalid communication from EtCO2 module | EtCO2 module communication attempt | error |
CO2 DEVICE NOT READY | CO2 in airway adapter when attempting zero operation | User attempts to zero EtCO2 sensor | unknown |
CO2 IN LINE: WAIT | Adapter zero attempted with CO2 still in adapter | User initiates zero sequence while CO2 present | unknown |
CO2 IN LINE: WAIT | CO2 detected in adapter during zeroing attempt | Adapter zero attempted with CO2 present | warning |
CO2 MODULE NOT VALID | Sidestream CO2 sensor connected (not supported) | CO2 module identification during startup | error |
CO2 OUT OF RANGE | Calculated CO2 value exceeds 150 mmHg | CO2 measurement during breathing | unknown |
CO2 UNIT ERROR | EtCO2 sensor or module hardware error detected | Sensor/module self-test or operation | error |
CO2 WARM UP | Mainstream CO2 sensor warming up after connection | CO2 sensor initialization | info |
CPR FAULT 8 | ECG processor not receiving CPR measurement data | CPR compression detection attempt | error |
DATA TRANSFERRED | Data transfer to external device completed successfully | End of data transfer operation | info |
DEFIB 30J READY | Defibrillator charged to 30J energy level and ready to discharge | Unit successfully charges after CHARGE button pressed at 30J energy setting | info |
DEFIB DISABLED | Defibrillation function disabled due to configuration or fault | Defibrillator disable via configuration or simultaneous faults | critical |
DEFIB FAULT 76 | PD Defib module failure during power-on self-test | POST execution | critical |
DEFIB FAULT 77 | PD module defibrillation failure while running | Defibrillation attempt during operation | critical |
DEFIB FAULT 78 | PD Defib functional safety error detected during operation | Defib operation | critical |
DEFIB FAULT 79 | PD module Defib/Pace failure during power-on self-test | POST execution | critical |
DEFIB FAULT 80 | Undefined error received from PD module | Communication with PD module | critical |
DEFIB FAULT 94 | PD module reset 3 times without being requested | PD module watchdog timeout or crash condition | critical |
DEFIB FAULT 95 | PD module not communicating with host processor | Communication timeout | critical |
DEFIB FAULT 96 | PD module reported error on discharge | Defibrillation delivery attempt | critical |
DEFIB MAINT. REQ. | Defibrillator maintenance required after 5000+ discharges at 200J | Discharge counter reaches 5000 discharges at 200J | warning |
DEFIB NOT CHARGED | Discharge button pressed but unit not charged | User presses discharge in Defib mode without charge | warning |
DEFIB NOT CHARGED | Defibrillator not charged when discharge attempted | Discharge button pressed in Defib mode but unit is not charged | warning |
DEFIB OVERUSE | Unit overheated from excessive shocks | More than 50 shocks delivered in less than 20 minutes | warning |
DEFIB OVERUSE | More than 50 defibrillator shocks delivered in less than 20 minutes | Shock delivery rate exceeds threshold | warning |
DEFIB PAD SHORT | Pad impedance indicates short circuit | Measured impedance between high voltage leads of MFC abnormal | critical |
DEFIB PAD SHORT | Abnormally low impedance measured between high voltage leads | Impedance measurement before/during defibrillation | error |
DEFIB PAD SHORT | Pad connector short circuit detected during pad presence check | Universal cable connected to test port with unit in DEFIB mode and leads set to PADS | warning |
DISK FORMAT REQ. | Problem accessing DOC (data) file on storage disk | File access error during data retrieval | error |
DISK FORMAT REQ. | Disk format error detected | Any problem with DOC file access occurred | error |
DISK FORMAT REQ. | Error detected during DOC file access operation | Disk file system access during operation | error |
ECG DISABLED | Persistent critical hardware failure on ECG module | ECG system initialization and diagnostic check | critical |
ECG DISABLED | ECG monitoring disabled due to hardware failure | Persistent critical hardware failure on ECG module | critical |
ECG FAULT 200 | No ECG applicable software available | ECG software loading during startup | error |
ECG FAULT 3 | ECG processor reset failure | ECG processor initialization | error |
ECG FAULT 4 | Excessive number of missed samples from ECG processor | ECG signal sampling during monitoring | error |
ECG FAULT 5 | ECG processor power-up initialization failure | ECG processor startup | error |
ECG FAULT 6 | ECG processor hardware failure | ECG processor self-test | critical |
ECG FAULT 7 | ECG processor is not responding to commands | Communication with ECG processor | error |
ECG LEAD OFF | One or more ECG leads disconnected with non-MFE leads selected | ECG input validation during monitoring | unknown |
ECG LEAD OFF | One or more ECG leads not connected when non-MFE leads selected | Lead connection detection with ECG input | warning |
ECG RA LEAD OFF | ECG right arm (RA) lead disconnected | ECG lead continuity check | unknown |
ECG TOO LARGE | Number of large ECG intervals exceeds threshold | ECG signal analysis during monitoring | warning |
ECG TOO LARGE | Number of ECG intervals exceeding amplitude threshold | ECG analysis | warning |
ECG V LEAD OFF | ECG precordial (V) lead disconnected | ECG lead continuity check | unknown |
ENABLE ETCO2 | EtCO2 feature not enabled | User attempts to zero EtCO2 sensor while feature disabled | info |
ENABLE ETCO2 | User attempts to zero EtCO2 sensor while feature is disabled | User initiates zero operation for disabled CO2 feature | warning |
ENERGY INCREMENTED | Defib energy automatically increased | After shock 1 or 2 delivered with auto-escalation configured | info |
ERASING REPORT | Unit is erasing the selected report data | info | |
ERASING REPORT | Report data erasure in progress | User selected report data to erase | info |
FULL DISCLS STOPPED | Full disclosure data recording stopped | Full disclosure data exceeds storage capacity (4 hours or 32MB) | warning |
IF NO PULSE | CPR period message | CPR period specified by AHA protocol reached | info |
IF NO PULSE | Message displayed upon entered CPR period specified by AHA protocol | info | |
INSERT CARD | No memory card detected | No card detected during manual or semi-automatic modes | error |
INSERT CARD | No card detected during manual or semi-automatic modes | CF Card not detected | warning |
INT. DUMP OVERLOAD | Excessive internal discharges in short time | More than 15 internal discharges in 5 minutes | warning |
INVALID GAS TEMP: RETRY | Operator attempted to set a gas temperature outside the Capnostat's operating range | warning | |
LOW BATTERY | Low battery | Battery charge below operational threshold | warning |
LOW BATTERY | Battery voltage low | Battery voltage drops below acceptable level | critical |
NIBP ARTIFACT | Unit is unable to detect systolic, diastolic or mean blood pressure due to excessive motion or vibration | Patient movement or vibration during measurement | warning |
NIBP ARTIFACT | Blood pressure measurement failed due to motion artifact | Excessive motion or vibration detected during measurement | warning |
NIBP COMM ERR 259 | NIBP communication error - framing/parity/fifo error | Communication framing, parity, or FIFO error detected | error |
NIBP COMM ERR 260 | NIBP communication error - invalid packets | Invalid packets received from NIBP module | error |
NIBP COMM ERR 261 | NIBP device not responding to action request | NIBP device not taking requested action | error |
NIBP COMM ERR 262 | NIBP device no reply to command | No response from NIBP device to command | error |
NIBP COMM ERR 262 | No reply from device | NIBP module not responding | error |
NIBP FAULT 263 | NIBP module fault 90 received | Fault 90 reported by NIBP module | critical |
NIBP FAULT 264 | NIBP module fault 91 received | Fault 91 reported by NIBP module | critical |
NIBP FAULT 265 | NIBP module fault 97 received | Fault 97 reported by NIBP module | critical |
NIBP FAULT 266 | NIBP module fault 98 received | Fault 98 reported by NIBP module | critical |
NIBP FAULT 267 | NIBP module fault 99 received | Fault 99 reported by NIBP module | critical |
NIBP FAULT 268 | NIBP device no response after power up | NIBP module not responding after power-up | critical |
NIBP MEAS ABORTED | Blood pressure measurement was aborted | Cuff inflation pressure set too high, inflation too fast, unable to find systolic after 180s, defibrillator charging, or user abort | error |
NIBP NOT READY | NIBP module not ready for measurement | Defibrillator charging or NIBP module performing power-up self-test | info |
NIBP OUT OF RANGE | Blood pressure reading out of physiological range | Data from NIBP module is out of range | warning |
NIBP OUT OF RANGE | Data from the NIBP module is out of range | NIBP measurement outside valid range | error |
NIBP SIGNAL WEAK | Weak or absent oscillometric signal | Weak or no oscillometric signal detected | warning |
NO DATA TO TRANSFER | Report data transfer error | Summary Report/Device Check/Activity Log data transfer error | error |
NOISY ECG | Excessive noise in ECG signal | Number of noisy analysis intervals exceeds threshold | warning |
NOISY ECG | Number of noisy analysis intervals exceeds threshold | Excessive ECG noise during analysis | warning |
NO QRS DETECT | Unit is in sync mode and heart rate is below 20 BPM or QRS amplitude is too low for proper synchronization | Low heart rate or low ECG amplitude in sync mode | warning |
NO QRS DETECT | QRS signal not detected | In sync mode with heart rate < 20 BPM or QRS amplitude too low | warning |
NO SHOCK ADV. | No shock advised message | Analysis finds non-shockable rhythm | info |
NO SHOCK ADV. | No shock advised. Analysis finds non-shockable rhythm | ECG analysis detects non-shockable rhythm | info |
OPEN AIR DISCHARGE | Measured Defib Impedance was greater than 1000 ohms and no energy was delivered | Defibrillation attempted with pads not in contact or open circuit | warning |
OPEN AIR DISCHARGE | Discharge impedance indicates open circuit | Measured defib impedance > 1000 ohms with no energy delivered | critical |
PACER DISABLED | Pacing function disabled | Pacing disabled simultaneously with other pace faults | error |
PACER FAULT 115 | PD pacing failure during power-on self-test | PD pace failed during POST | critical |
PACER FAULT 117 | PD pacing functional safety failure during operation | PD pace functional safety fault while running | critical |
PACER FAULT 121 | PD pacing failure during operation | PD pace failed while running | critical |
PACER FAULT 122 | PD module Defib/Pace failure during power-on self-test | PD module defib/pace failure during POST | critical |
PACER WARNING 124 | PD pacing warning during operation | PD pace warning while running | warning |
PADDLE FAULT | Cannot detect paddle type | Cannot detect type of accessory attached to universal cable | error |
PADDLE FAULT | Cannot detect type of accessory attached to the universal cable | Paddle type detection failure | error |
PEDIATRIC PADS IN USE | The one of the Pediatric ID is missing | Pediatric identification missing during pediatric mode | warning |
PEDIATRIC PADS IN USE | Pediatric electrode pads active | One of the Pediatric ID is missing | warning |
PERFORM CPR | Message displayed upon entered CPR period specified by AHA protocol | info | |
PERFORM CPR | CPR in progress message | CPR period specified by AHA protocol entered | info |
POOR LEAD CONTACT | ECG lead electrode poor or absent connection | One or more ECG leads poorly connected or not connected to patient | warning |
POOR PAD CONTACT | Electrode impedance exceeds threshold | Defibrillation pads not adequately coupled to patient skin | warning |
POOR PAD CONTACT | Defibrillation pad impedance high | Electrode impedance exceeds threshold | warning |
PRESS ANALYZE | Alternate message for Check Patient message if this configuration option is selected | info | |
PRESS CHARGE | User prompt to charge defibrillator | Discharge button pressed before charge or advisory after SHOCK ADVISED | info |
PRESS CHARGE | Discharge button pressed in Manual Defib mode but unit is not charged, or advisory message following SHOCK ADVISED | Button pressed when unit not ready or auto-charge disabled | info |
PRESS SHOCK | Deliver defibrillation prompt | Unit fully charged and ready to deliver shock in AED auto mode | info |
PRESS SHOCK | Prompt issued in AED auto defib mode when defib is charged and ready | info | |
READING LOG DATA | Summary report printing in progress | User requested log data printout | info |
RECORDER FAULT 142 | Strip chart system error | Printer system malfunction detected | error |
RECORDER FAULT 143 | Printer failed power-up self-test | Strip chart failed echo test during POST | error |
RECORDER FAULT 147 | Printer printhead overheated | Printhead temperature exceeded safe operating limit | error |
RELEASE BUTTONS | Simultaneous external paddle button presses detected before unit reached full defib charge | Both paddle buttons pressed before ready state | warning |
RELEASE SHOCK | Discharge switch interference detected | Discharge switch closed when pressing charge button or pressed before ready state | warning |
RELEASE SHOCK | Discharge switch closed when pressing charge button, or discharge button pressed before defib reached ready state | Discharge button activation during charging | warning |
REMOVE SYNC | Exit synchronization mode | Analyze button pressed or no heart rate detected in Defib Sync mode | info |
REMOVE SYNC | Analyze button pressed, or no heart rate detected while in Defib Sync mode | Analyze pressed in sync mode or no HR detected | info |
REPEAT NIBP MEAS | Blood pressure measurement retry needed | Maximum inflation attempts exceeded or 180-second time limit exceeded | warning |
REPLACE BATTERY | Battery voltage is less than absolute minimum | Battery voltage critically low | critical |
REPLACE BATTERY | Battery completely discharged | Battery voltage less than absolute minimum; shutdown imminent | critical |
REPORT FULL | Summary report memory full | Summary report storage capacity reached | warning |
RETRY ANALYSIS | Rerun rhythm analysis | Noisy ECG detected during analysis | warning |
RETRY ANALYSIS | Advisory message in conjunction with noisy ECG | ECG analysis halted due to noise | info |
RPT NO DISK | Internal disk not formatted | Internal disk not formatted and cannot be formatted | critical |
SELECT 30J FOR TEST | Self test attempt at energy other than 30J | Attempt to run self test at energy other than 30J | warning |
SELECT 30J FOR TEST | System requests operator reduce energy level to 30J for safe internal discharge testing | Operator selects 100J energy level and presses CHARGE button | info |
SELECT ASYNC PACE | PD module cannot detect sync pulses; async pacing required | Continuous sync mode operation when sync pulses unavailable | warning |
SELECT DEFIB MODE | Analyze button pressed in non-defib mode | Analyze button pressed in pace or monitor mode | warning |
SELECT LIMB LEADS | Paddles or augmented ECG leads selected when continuous analysis active or started | Lead selection error during analysis | info |
SELECT LIMB LEADS | Wrong lead type selected during analysis | Paddles or augmented ECG leads selected when continuous analysis active or started | warning |
SELECT PADS | ECG lead (non-MFE Pads) selected when ANALYZE pressed | Lead type mismatch during analysis | info |
SELECT PADS ECG | Non-MFE pads selected during analysis | ECG lead (non-MFE Pads) selected when ANALYZE pressed | warning |
SET CLOCK | Real time clock failure with invalid date or time | RTC detects invalid date/time on power-up or operation | warning |
SET PACE MA | Multiple copy errors in pace module software or memory | Copy errors on first pace entry or after 10+ minutes in other mode | error |
SHOCK ADVISED | Analysis finds a shockable rhythm | ECG analysis detects shockable rhythm | info |
SHOCK ADVISED | Shockable rhythm detected during analysis | Analysis detects VF or VT rhythm | critical |
SPO2 AMBIENT LIGHT | Excessive ambient light detected | High ambient light at sensor site interfering with SpO2 measurement | warning |
SPO2 COMM ERROR | No transmissions from SpO2 unit received or communication error | SpO2 module communication failure | error |
SPO2 COMM ERROR | No communication from SpO2 module | System cannot receive transmissions from SpO2 unit | error |
SPO2 PULSE SEARCH | Sensor cannot detect a pulse | No pulse signal detected at sensor site during normal operation | warning |
SPO2 PULSE SEARCH | SpO2 module pulse search in progress | SpO2 module actively searching for patient pulse signal | info |
STAND CLEAR | Patient rhythm is being analyzed | Analysis started or auto-analysis initiated per rescue protocol | info |
STAND CLEAR | Patient rhythm analysis in progress; keep clear of patient | Analyze button pressed or automatic analysis started during rescue | warning |
SUMMARY TIMEOUT | Summary report generation timeout error | System fails to generate or transmit summary report within timeout period | error |
SYNC DEFIB DISABLED | Sync mode enabled when Analyze pressed in defib mode | Sync mode active when Analyze button pressed during defib operation | warning |
SYSTEM FAULT 210 | ECG control task not activated for 500 ms | Digital board ECG task fails to activate within 500 ms cycle | critical |
SYSTEM FAULT 211 | Defib task not activated for 500 ms | Digital board defib task fails to activate within 500 ms cycle | critical |
SYSTEM FAULT 212 | Pace task not activated for 500 ms; or User Interface task not activated for 500 ms | Digital board task fails to activate within 500 ms cycle | critical |
SYSTEM FAULT 213 | System task have not been activated for 500 ms – User Interface | UI task timeout | critical |
SYSTEM FAULT 214 | Display task not activated for 500 ms | Digital board display task fails to activate within 500 ms cycle | critical |
SYSTEM FAULT 215 | Reserved task not activated for 500 ms (not used) | Digital board reserved task fails to activate within 500 ms cycle | critical |
SYSTEM FAULT 36 | Filtered sum of all supply voltages is out of range | Voltage measurement from filtered sum exceeds acceptable limits | error |
SYSTEM FAULT 37 | 1/2 scale reference voltage is out of range | Reference voltage measurement exceeds acceptable tolerance | error |
SYSTEM FAULT 38 | Failure to shutdown after shutdown order written to RTC | System fails to power down when shutdown command is sent to Real Time Clock | critical |
TEST FAILED | 30J defib self-test failed | Self-test procedure executed and detected failure in 30J defibrillation circuit | critical |
TRANSFERRING DATA | Data transfer in progress (informational) | System initiating or executing data transfer to external device or memory | info |
USE PADDLE DISCHG | Defibrillator is charged and front-panel discharge button pressed when external paddles or internal spoons with discharge buttons are connected | Discharge button pressed while paddles/spoons connected and unit charged | warning |
USE PADS TO PACE | MFE accessory other than Pads detected in Pace mode | Non-Pads accessory connected while in Pace mode | warning |
USER SETUP REQ. | Both copies of configuration data are bad, or software with older configuration version was loaded | Configuration memory corruption or old software version installation | error |
VF ALARMS OFF | VF alarms disabled in incompatible mode or lead selection | Alarms enabled in Pace Mode or Continuous Analysis with non-standard lead, or Alarm button pressed with Heart Rate alarm disabled | warning |
VF ALARMS OFF | Alarms enabled in Pace Mode or when Continuous Analysis active or started in manual mode but current lead is not Pads/Lead I/II/III; or Alarm button pressed but Heart Rate alarm is disabled | VF alarm function disabled in incompatible mode or alarm disabled state | warning |
ZERO CO2 ADAPTER | Negative CO2 detected. May be caused by sensor zeroed with CO2 in airway or optical blockage | CO2 sensor reading below acceptable baseline during measurement | warning |
ZERO CO2 ADAPTER | Negative CO2 detected during capnography monitoring | CO2 sensor detects negative (impossible) CO2 value | warning |
ZERO DONE | Sensor/adapter zero operation completed successfully (informational) | CO2 adapter zeroing procedure completes normally | info |
ZERO FAILED | The zero operation did not complete successfully | Zero calibration procedure failed to complete or detect valid baseline | error |
ZEROING CO2 ADAPTER | CO2 adapter zeroing procedure in progress (informational) | User initiates CO2 adapter zero or system auto-starts procedure | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI