ZOLL M: troubleshooting and error codes
The ServiceForge knowledge base holds 60 documented symptoms for the ZOLL M, with 126 ranked causes, 144 diagnostic checks with test points and expected values, and 310 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Displayed HR not accurate; artifact present.
Heart rate value shown on display does not match actual patient heart rate; visible signal artifact or noise present on screen
Ranked causes
- 1. ECG cable or electrode movement causing artifact (common)
- 2. Patient motion during monitoring (common)
- 3. Radio frequency interference (RFI) (occasional)
- 4. Poor electrode to patient skin contact (common)
- 5. Inadequate skin preparation (common)
- 6. Cable proximity to electrical equipment or RFI sources (occasional)
- 7. Loose ECG cable connection in unit (occasional)
- 8. Defective ECG cable (occasional)
- 9. Electrode contamination or broken snaps (occasional)
- 10. Intermittent patient cable or connector wiring issue (rare)
- 11. Defective ECG input connector (rare)
- 12. Defective ECG connector cable to system board (rare)
- 13. Defective system board (rare)
Inspect and reposition ECG cables
- Test point
- Cable routing and electrode connections
- Expected
- Cables do not pull on electrodes; no excessive swinging observed
13 more checks with pass and fail actions, 13 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK PADS/POOR PAD CONTACT message displays patient safety
Message indicates defibrillator pad contact or connection problem
Ranked causes
- 1. PADS connector not fully inserted into universal cable (common)
- 2. Damaged defibrillator pads, wires or connector (common)
- 3. Dried out or expired defibrillator pads (common)
- 4. Poor pad contact with patient skin (common)
- 5. Cable defect or open circuit (occasional)
- 6. High Voltage Module cable connection faulty (occasional)
- 7. Defective High Voltage Module (occasional)
- 8. Defective system board (rare)
Reseat PADS connector
- Test point
- PADS connector to universal cable
- Expected
- Connector fully and firmly seated
9 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG LEAD OFF message displays
Message appears on display indicating loss of ECG signal on one or more leads (3, 5, or 12 lead cable)
Ranked causes
- 1. Dry or expired electrode gel (common)
- 2. Poor electrode site preparation (common)
- 3. Leads not attached (common)
- 4. ECG cable corrosion or broken connector pins (occasional)
- 5. Intermittent cable connection (occasional)
- 6. Faulty ECG input connector or pins (occasional)
- 7. ECG shielding issue on system board (rare)
- 8. Defective system board (rare)
Verify electrode site preparation
- Test point
- Patient skin at electrode sites
- Expected
- Clean skin, lightly abraded, hair clipped if necessary
10 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flash or arcing under defibrillator pad patient safety
Visible electrical discharge or arcing beneath defibrillator pad during application or use
Ranked causes
- 1. Use of alcohol or betadine skin preparation (common)
- 2. Gel droop or leakage from pad (common)
- 3. Pads not properly coupling to patient skin or cable (common)
- 4. Dried out gel on defibrillator pad (common)
- 6. Expired defibrillator pads (occasional)
- 7. Chest compression applied through pads (occasional)
- 8. Front pad lifted during patient repositioning (occasional)
Verify skin preparation method
- Test point
- Patient skin at pad site
- Expected
- Prepared with non-alcohol, non-betadine agent
8 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heart rate not accurate (artifact present) patient safety
Displayed heart rate inaccurate due to electrical artifact on ECG waveform; may include muscle noise or cable movement
Ranked causes
- 1. ECG electrode or cable movement artifact (common)
- 2. Patient muscle artifact (common)
- 3. Poor electrode contact (common)
- 4. Radio frequency interference from nearby equipment (occasional)
- 5. ECG cable damage or intermittent connection (occasional)
Ensure ECG cables do not pull on electrodes
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check pads or poor pad contact patient safety
CHECK PADS/POOR PAD CONTACT message displayed; defibrillator pads not making adequate contact
Ranked causes
- 1. Poor pad-to-patient contact (common)
- 2. Poor electrical connection between pads and cable (occasional)
- 3. Damaged defibrillator pads or cable (occasional)
- 4. High voltage module or cable failure (occasional)
Connect universal cable to test shorting plug
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Displayed HR not accurate. No artifact present.
Heart rate value shown on display does not match actual patient heart rate; no visible signal artifact on screen
Ranked causes
- 1. Wrong lead selected (common)
- 2. ECG display size issue (common)
- 3. Electrode placement issue (common)
- 4. Heart rate not flashing with QRS complex (occasional)
Verify HR flashing synchronously with QRS
- Test point
- Display: HR value and QRS waveform
- Expected
- HR indicator flashes once per each QRS complex visible
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG lead off patient safety
ECG LEAD OFF message displayed during patient monitoring; ECG waveform not displayed
Ranked causes
- 1. Poor electrode contact (common)
- 2. Dry or expired electrodes (common)
- 3. ECG cable damage or poor connection (occasional)
- 4. ECG input connector or system board failure (occasional)
Try to reproduce problem using ECG simulator
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electronic interference
ECG waveform displays excessive noise or 50/60 Hz interference pattern
Ranked causes
- 1. Excessive radio frequency interference from nearby equipment (common)
- 2. Patient cables in proximity to electrical equipment (common)
- 3. M Series unit placed near RFI source (occasional)
- 4. Active RFI sources operating in environment (common)
Identify sources of RFI in environment
- Test point
- Surrounding electrical equipment and devices
- Expected
- RFI sources identified
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK PADS/POOR PAD CONTACT message displayed patient safety
Unit displays CHECK PADS or POOR PAD CONTACT message indicating defibrillator pad contact or connection issue
Ranked causes
- 1. Poor pad contact or expired pads (common)
- 2. Cable or High Voltage Module defect (occasional)
- 3. High Voltage Module or system board failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Defibrillator pad contact problem patient safety
CHECK PADS/POOR PAD CONTACT message displayed
Ranked causes
- 1. Poor pad contact or defective pads (common)
- 2. Defective cable or connector (occasional)
- 3. High Voltage Module defect (rare)
Test pad cable with shorting plug
- Test point
- PADS connector
- Expected
- DEFIB PAD SHORT message displays
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Displayed heart rate not accurate with artifact present patient safety
Displayed heart rate inaccurate due to ECG artifact from electrode/cable movement or patient motion
Ranked causes
- 1. ECG artifact from electrode or cable movement (common)
- 2. Radio frequency interference (occasional)
- 3. ECG cable or connector wiring intermittent fault (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG LEAD OFF message displayed patient safety
Unit displays ECG LEAD OFF message for 3, 5, or 12 lead cable indicating loss of ECG signal on one or more leads
Ranked causes
- 1. Electrode preparation issue (common)
- 2. ECG cable or connector issue (occasional)
- 3. System board or internal connection failure (rare)
Verify electrode expiration date and condition
- Expected
- Electrodes not expired and gel not dry
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heart rate display inaccurate with artifact patient safety
Displayed heart rate inaccurate with ECG artifact present
Ranked causes
- 1. ECG cable movement or patient motion (common)
- 2. Radio frequency interference (occasional)
- 3. Poor electrode or cable connection (common)
Ensure ECG cable fits snugly
- Test point
- ECG cable connector
- Expected
- Cable seated properly in unit
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG lead disconnection or poor contact patient safety
One or more ECG leads not properly attached to patient
Ranked causes
- 1. Lead cable disconnection (common)
- 2. Patient electrode failure or poor skin contact (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heart rate not accurate (no artifact) patient safety
Displayed heart rate does not correlate with actual patient cardiac rhythm; no electrical artifact visible on waveform
Ranked causes
- 1. Lead selection not optimal for this patient (common)
- 2. Electrode repositioning needed (common)
Verify heart rate flashes with each QRS on display
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Noisy ECG signal patient safety
Excessive electrical noise in ECG waveform exceeds analysis threshold
Ranked causes
- 1. Patient movement or muscle artifact (common)
- 2. Poor lead contact (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor lead contact patient safety
One or more ECG leads poorly connected or not connected to patient
Ranked causes
- 1. Electrode attachment failure (common)
- 2. Unit malfunction (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder paper or print mechanism failure caution
Strip chart recorder unable to print or experiencing thermal issues
Ranked causes
- 1. Paper feed or printhead issue (common)
- 2. Printhead overtemperature (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
V leads off patient safety
V LEADS OFF message displayed on monitor; V lead waveforms not displayed
Ranked causes
- 1. V lead terminator plug not attached (common)
- 2. V1 lead touching patient skin (occasional)
Verify V lead terminator is attached
- Test point
- V lead connector
- Expected
- terminator plug in place
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
V LEADS OFF message displayed patient safety
Unit displays V LEADS OFF message indicating V lead detection failure, possibly from V1 lead wire contact with patient skin
Ranked causes
- 1. V lead terminator not connected when not in use (common)
- 2. V1 lead wire metal snap contact with skin (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
V LEADS OFF message displays
Message appears indicating loss of V lead signal
Ranked causes
- 1. V lead terminator plug not attached when V leads not in use (common)
- 2. V1 lead wire metal snap in contact with patient skin (occasional)
Verify V lead terminator plug attachment
- Test point
- V lead connector
- Expected
- Terminator plug attached when V leads not in use
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
A/D converter timing failure caution
Analog-to-digital converter not performing conversions in timely manner
Ranked causes
- 1. A/D converter failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm suspend function does not work
Alarm suspend button does not mute alarm tone or suspend alarm for specified duration
Ranked causes
- 1. Alarm suspend button not functioning (rare)
Press ALARM SUSPEND button during active alarm
- Test point
- ALARM SUSPEND button
- Expected
- Alarm tone stops, alarm symbol displays with X through it, heart symbol continues flashing
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chest lead disconnection (V1-V6) patient safety
One or more chest leads (V1 through V6) not properly attached
Ranked causes
- 1. Chest lead cable disconnection (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Displayed heart rate not accurate without artifact patient safety
Displayed heart rate does not match actual patient heart rate with clean ECG waveform showing no artifact
Ranked causes
- 1. Display filter or size settings inappropriate (common)
Verify heart rate flashes with each QRS on display
- Expected
- Heart rate indicator synchronizes with QRS complex
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG LEAD OFF alarm does not trigger patient safety
No lead-off alarm tone or visual indication when ECG lead is disconnected from analyzer
Ranked causes
- 1. ECG lead disconnected or loose connection (common)
Remove ECG lead from analyzer and observe alarm response
- Test point
- QED 6 Analyzer ECG input
- Expected
- Alarm symbol flashes, heart symbol stops flashing, ECG LEAD OFF alarm tone sounds
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG processor communication failure caution
Communication lost between ECG processor and main processor
Ranked causes
- 1. ECG processor fault or communication failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG signal too large patient safety
ECG signal amplitude exceeds expected range for accurate rhythm analysis
Ranked causes
- 1. High amplitude ECG signal (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG signal too small patient safety
ECG signal amplitude too low for accurate rhythm analysis
Ranked causes
- 1. Low amplitude ECG signal (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electronic interference detected caution
System detects excessive radio frequency interference affecting measurements
Ranked causes
- 1. Nearby RFI source (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
EtCO2 communication error caution
No or invalid communication from EtCO2 module
Ranked causes
- 1. Module disconnection or failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
EtCO2 sensor warming up
EtCO2 sensor requires extended warm-up period
Ranked causes
- 1. Sensor cable defect (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heart rate alarm does not function patient safety
Alarm does not sound when heart rate exceeds threshold; alarm symbol and tone absent during high heart rate condition
Ranked causes
- 1. Alarm not enabled in monitor menu (common)
Verify alarm state is set to ENABLE
- Test point
- ECG HR alarm menu
- Expected
- ENABLE appears on display
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heart rate display inaccurate without artifact patient safety
Displayed heart rate does not match actual QRS rate, no artifact visible
Ranked causes
- 1. Suboptimal lead selection or electrode placement (common)
Verify HR matches QRS
- Test point
- ECG display
- Expected
- Heart rate flashes with each QRS on display
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
High voltage capacitor requires discharge before removal high voltage
Unit contains lethal voltages in the high voltage capacitor assembly that must be safely discharged prior to service work
Ranked causes
- 1. High voltage capacitor charged during normal operation (common)
Open high voltage capacitor plastic isolator
- Test point
- Capacitor assembly isolator cover
- Expected
- Mylar tape lifted to expose capacitor
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP monitor zeroing and measurement accuracy verification
Verification procedure to ensure proper pressure channel zeroing and accurate blood pressure measurements on M Series CCT unit with IBP option.
Ranked causes
- 1. Inaccurate IBP channel calibration or zeroing (unknown)
Set up M Series CCT unit with IBP option and fully charged battery
- Test point
- M Series unit
- Expected
- Unit ready for testing
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP noise level verification
Verify absence of excessive noise in blood pressure circuitry using patient simulator and oscilloscope
Ranked causes
- 1. Routine maintenance to ensure signal quality of IBP channels (common)
Power on M Series CCT unit with battery
- Test point
- Unit power state
- Expected
- Unit powers on and displays ready
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP noise test
Verification procedure to ensure there is no excessive noise in the blood pressure circuitry on M Series CCT unit.
Ranked causes
- 1. Excessive noise in blood pressure circuitry (unknown)
Set up M Series CCT unit with IBP option and fully charged battery
- Test point
- M Series unit
- Expected
- Unit ready for testing
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP pressure reading accuracy verification
Verify that blood pressure measurements are accurate and channel zeroing functions properly using patient simulator
Ranked causes
- 1. Need periodic accuracy verification of IBP channels (common)
Power on M Series CCT unit with fully-charged battery
- Test point
- Unit power state
- Expected
- Unit powers on
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP static pressure measurement accuracy
Verify accurate measurement of both positive and negative static pressure values using transducer simulator
Ranked causes
- 1. Need verification of static pressure calibration accuracy (common)
Power on M Series CCT unit with battery
- Test point
- Unit power state
- Expected
- Unit powers on and ready
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP static pressure test
Verification procedure to ensure accuracy of both positive and negative static pressure measurements on M Series CCT unit IBP channels.
Ranked causes
- 1. Inaccurate static pressure measurement calibration (unknown)
Set up M Series CCT unit with IBP option and fully charged battery
- Test point
- M Series unit
- Expected
- Unit ready for testing
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery voltage caution
Battery voltage below operating minimum
Ranked causes
- 1. Battery depletion (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Maintenance schedule verification
Confirm NIBP maintenance procedures completed per schedule: Monitor Test every 6 months, Volume Leak Test annually or per 10,000 readings, Transducer Calibration annually or per 10,000 readings
Ranked causes
- 1. Maintenance procedures not performed on schedule (occasional)
Verify NIBP Monitor Test completed within 6 months
- Test point
- Maintenance file and checklist records
- Expected
- Pass/Fail marked on checklist within last 6 months
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Maintenance test setup and checklist documentation
Preparation and documentation procedures for conducting quarterly IBP maintenance tests on M Series CCT unit.
Ranked causes
- 1. Need for periodic IBP function verification (common)
Refer to Operator's Shift Checklist for routine verification
- Test point
- Operator documentation
- Expected
- Shift checklist completed at beginning of each shift
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP Monitor Test verification patient safety
Verifies accuracy of systolic, diastolic, mean blood pressure measurements and pulse rate calculation using Bio-Tek BP Pump NIBP Monitor Analyzer simulator
Ranked causes
- 1. Transducers or measurement system out of calibration (common)
Connect NIBP Analyzer simulator hose to NIBP connector
- Test point
- NIBP connector on M Series unit
- Expected
- Hose firmly connected
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP Transducer Calibration patient safety
Two-point calibration procedure to ensure accurate pressure measurements; optional at 6 months, recommended annually or every 10,000 readings
Ranked causes
- 1. Transducer calibration drift from factory settings (occasional)
Connect analyzer simulator hose and configure for pressure simulation
- Test point
- NIBP connector and analyzer MODE/SELECT buttons
- Expected
- Analyzer in Pressure Simulator mode, hose connected
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No QRS detection in sync mode patient safety
Unit cannot detect heart rate or QRS complex in synchronization mode
Ranked causes
- 1. Low heart rate or signal amplitude (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Open air discharge high voltage
Defibrillation capacitor voltage remains high after discharge attempt
Ranked causes
- 1. Failed discharge (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacer circuit failure (multiple types) high voltage
Pacing circuit unable to deliver therapy or verify output
Ranked causes
- 1. High voltage module failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor electrode pad contact patient safety
Defibrillation pad impedance exceeds acceptable threshold
Ranked causes
- 1. Pad coupling failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Real time clock failure
Unit real time clock contains invalid date or time
Ranked causes
- 1. Clock data corruption (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self test failure high voltage
30J self test failed during defib peak current measurement
Ranked causes
- 1. Defib circuit component failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 ambient light interference
Ambient light too bright for SpO2 sensor operation
Ranked causes
- 1. Excessive ambient light (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 communication error caution
No communication from SpO2 measurement module
Ranked causes
- 1. SpO2 module disconnection or failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 sensor fault
SpO2 sensor is defective or unrecognized
Ranked causes
- 1. Sensor failure or incompatibility (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System fault caution
Processor or memory integrity error detected
Ranked causes
- 1. ROM or RAM failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature measurement accuracy verification
Verify that temperature channels T1 and T2 on M Series CCT unit with Temperature option display accurate readings when tested with calibrated probe simulator
Ranked causes
- 1. Temperature sensor or channel calibration drift requiring verification (unknown)
Set Temperature Probe Simulator to room temperature (~68°F)
- Test point
- T1 channel display
- Expected
- 67.9 °F
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test setup for IBP maintenance procedures
Prepare M Series CCT unit with proper configurations for IBP testing
Ranked causes
- 1. Initial setup required before performing any IBP maintenance test (common)
Power on M Series CCT unit
- Test point
- Unit power state
- Expected
- Unit displays boot screen and main menu
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wandering baseline
ECG baseline drifts or moves slowly across display instead of remaining stable
Ranked causes
- 1. Electrode placement or contact issues (90% of cases) (common)
Inspect electrode sites and contact
- Test point
- Electrode placement and skin contact
- Expected
- Electrodes in firm full-surface contact with clean skin
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
200J MAX BIPHASIC | User attempted to set defibrillation energy exceeding 200J on biphasic unit | User attempts to select energy level above 200J during defibrillation | info |
50J MAX | Selected energy below minimum for internal paddles | User attempts to set energy below 50J with internal paddles | info |
ADJUST ECG | Unit unable to synchronize on detected heart rate | Sync mode active with heart rate below 20 BPM or QRS size set too small | warning |
ANALYSIS HALTED | ECG analysis halted due to user interaction such as lead/size change, analyze button pressed again, impedance fault, or charging error detected in auto defib mode. | User interaction interrupted analysis or charging error occurred | info |
ANALYSIS HALTED | ECG analysis interrupted and stopped | User interaction during analysis; lead change; size change; analyze button pressed; impedance fault; charging error | info |
ANALYSIS RESTARTED | ECG analysis restarted after automatic size adjustment | ECG signal detected out of range; automatic adjustment completed | info |
ANALYSIS RESTARTED | Device detected ECG signals out of range, automatically adjusted ECG size and is restarting its shockable rhythm analysis sequence. | ECG signal out of range (too large or too small) | info |
AUDIO FAULT 136 | Audio DSP hardware failure | Audio digital signal processor hardware malfunction at power-up or during operation | error |
AUDIO NOT RECORDING | Audio recording system not functional | Audio recording attempted but system non-functional | warning |
AUDIO QUEUE FULL | Audio output queue buffer full; cannot accept additional voice prompts | Voice prompt queue reaches capacity during operation | info |
AUDIO QUEUE FULL | Audio output queue is full. Additional voice prompts can't be queued at this time. | Too many audio prompts queued | info |
BAD CARD | Cannot read from card | When reading from memory card during configuration | error |
BATT HIGH CURRENT | Battery is charged and battery current is >.1 A or: Battery is not charged and battery current is > 1.6 A. | Battery current exceeds limits | error |
BATT HIGH CURRENT | Battery current exceeds safe operating limits | Charged battery current >0.1A or uncharged battery current >1.6A | error |
BATT HIGH VOLTAGE | Battery voltage exceeds safe operating level | Battery voltage measures above 15.5 volts | error |
BATT LOW CURRENT | Battery is not charged and battery current is <.35 A. | Battery current drops below minimum | error |
BATT LOW CURRENT | Battery charging current below minimum acceptable level | Uncharged battery with charging current below 0.35A | error |
BATT LOW VOLTAGE | Battery voltage below minimum safe operating level | Battery voltage measures below 9.5 volts | critical |
BATT OVERCHARGE | Charger on for > 4 hours. | Battery charged for extended period | warning |
BATT OVERCHARGE | Battery charger has been on longer than maximum safe duration | Charger on for more than 4 hours continuously | error |
BATT OVERCHARGE | Battery charging exceeded safe limits | Charging circuit malfunction or charger fault | error |
BATT OVERCHARGE | Battery overcharge condition detected | Charging system malfunction or battery fault | warning |
BRIDGE SHORT | Biphasic bridge circuit showing higher than expected current | During biphasic bridge test or immediately after discharge | error |
BRIDGE SHORT | Current higher than expected was detected during Biphasic bridge test or immediately following a discharge. | Excessive current during bridge test or discharge | error |
BRIDGE TEST FAILED | Biphasic module not operating properly while charging. | Biphasic module fails charging test | error |
BRIDGE TEST FAILED | Defibrillator output stage bridge test failed | During self-test or defibrillation initialization | error |
CABLE FAULT | ECG or patient cable connection fault detected | Intermittent or broken cable detected during monitoring | warning |
CABLE FAULT | Cable integrity failure | Cable continuity or insulation test failure | error |
CABLE FAULT | Incorrect or invalid paddle identification signal on cable | Auto defib mode; paddle ID analog-to-digital reading invalid | error |
CANNOT CHARGE | Defibrillator capacitor will not charge when charge button pressed | Charge button activated but capacitor voltage does not rise | critical |
CANNOT CHARGE | Unit unable to charge | Charging operation initiated but charging circuit failure detected | error |
CANNOT CHARGE | Battery charging not possible | Charger unable to initiate or maintain charging | error |
CANNOT CHARGE | Charger unable to initiate or maintain charging | When charger is connected and charging button pressed | error |
CANNOT CHARGE 45 | System unable to charge defibrillator | Charging request initiated | error |
CARD EMPTY | No data on card | When reading from memory card that contains no configuration data | warning |
CARD FULL | Memory card storage capacity reached | Memory card filled to capacity during recording | warning |
CARD READ FAILED | No response from card reader | When reading from memory card during configuration | error |
CHECK CO2 ADAPTER | Airway adapter is removed, occluded or adapter zeroing needs to be performed or was performed incorrectly. | Adapter missing, blocked, or calibration error | warning |
CHECK CO2 SENSOR | EtCO2 Sensor is unplugged or defective | Sensor not detected or sensor failure | warning |
CHECK MEMORY CARD | No memory card detected during manual or semi-automated mode | Manual or semi-automated mode active without memory card installed | warning |
CHECK PADS | Electrode pads not properly applied or have high impedance | Displayed with POOR PAD CONTACT or DEFIB PAD SHORT | unknown |
CHECK PADS | Pad contact issue detected with poor contact or short circuit | Poor pad contact or defib pad short detected | warning |
CHECK PADS | Message displayed in conjunction with POOR PAD CONTACT or DEFIB PAD SHORT | Pad contact or connection fault detected | error |
CHECK PADS/POOR PAD CONTACT | Defibrillator electrode pads have inadequate contact with patient | Poor pad-to-skin contact, pad damage, expired pads, connector issue | error |
CHECK PATIENT | Shockable cardiac rhythm detected during background analysis | Background ECG analysis detects VF or pulseless VT | unknown |
CHECK PATIENT | Background ECG analysis detects shockable rhythm | Shockable rhythm detected during background analysis | warning |
CHECK PULSE | Alternate message for NO SHOCK ADVISED or message after third shock when auto analyze 3 times enabled | No shockable rhythm detected or third shock delivered | info |
CHECK PULSE | Alternate message for NO SHOCK ADVISED message. Message also shown after delivering third shock when auto analyze 3 times option is enabled. | Non-shockable rhythm detected | info |
CHECK RECORDER | Paper tray empty, paper jam detected, or recorder door opened | Recorder self-test detects paper out; mechanical jam; door opened | error |
CHECK RECORDER | Recorder unit requires inspection or maintenance | Recorder malfunction detected or periodic maintenance due | warning |
CHECK RECORDER | Recorder module fault detected | Recorder circuit malfunction or component failure | warning |
CHECK RECORDER | Recorder module malfunction detected | During recording initialization or operation | warning |
CHECK RECORDER 47 | Recorder malfunction detected | Recorder operation attempted | warning |
CHECK SPO2 SENSOR | Pulse oximetry sensor not optimally positioned | SpO2 sensor dislodged or suboptimal placement | info |
CHECK SPO2 SENSOR | Reposition SpO2 sensor on patient. | SpO2 sensor signal lost or poor | warning |
CHECK SPO2 SITE | Pulse oximetry signal absent or insufficient from monitored finger or toe | SpO2 sensor placed on low or no perfusion site | warning |
CLOCK FAULT 11 | Real time clock oscillator failure | Clock oscillator malfunction during power-up or operation | error |
CLOCK FAULT 12 | Real-time clock memory or circuit failure 12 | During system startup or time verification | error |
CLOCK FAULT 12 | Real-time clock fault condition 12 | Clock circuit or timing synchronization failure | warning |
CLOCK FAULT 12 47 | unknown | ||
CLOCK FAULT 13 | Clock circuit fault 13 | System clock error detected | warning |
CLOCK FAULT 13 | Real-time clock fault condition 13 | Clock circuit or timing synchronization failure | warning |
CLOCK FAULT 13 47 | unknown | ||
CO2 COMM ERROR | No or invalid communication from EtCO2 module | Communication failure between CO2 module and main processor | error |
CO2 SENSOR WARM UP | EtCO2 sensor currently warming to operational temperature | Sensor powered on; warm-up cycle in progress | info |
CONFIG DATA NOT COMPATIBLE | Internal software is unable to read card data | When loading configuration from memory card | error |
CONFIRM MANUAL MODE | System requesting confirmation to enter manual mode | Manual mode activation initiated | info |
CONFIRM MANUAL MODE | Displayed when manual mode is entered. Alerts user to confirm that manual mode is desired. | Manual mode selected | info |
CONFIRM MANUAL MODE | Manual mode confirmation required | System entering manual operation mode | info |
CONFIRM MANUAL MODE 47 | unknown | ||
DEFIB DISABLED | Defibrillator function disabled | Defibrillator intentionally or unintentionally disabled | warning |
DEFIB DISABLED | User prompt issued simultaneously with other faults if defib is disabled. | Defibrillation disabled by configuration or fault | error |
DEFIB DISABLED 47 | unknown | ||
DEFIB FAULT 108 | Defibrillator fault code 108 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 108 | VMON voltage is less than the target energy during charging | Capacitor voltage monitoring shows charging below target | error |
DEFIB FAULT 108 49 | unknown | ||
DEFIB FAULT 109 | Defibrillator fault code 109 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 109 | Defib capacitor voltage is greater than selected energy when defibrillator is charging or ready. | Capacitor voltage exceeds selected energy level | error |
DEFIB FAULT 109 49 | unknown | ||
DEFIB FAULT 111 | Defibrillator fault code 111 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 111 | Capacitor voltage exceeded absolute maximum safe voltage | Defib capacitor voltage exceeds absolute maximum acceptable limit | critical |
DEFIB FAULT 111 49 | unknown | ||
DEFIB FAULT 71 | Defibrillator fault code 71 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 71 | More than 50 internal dumps occurred in less than 20 minutes. | Excessive number of defib energy discharges | error |
DEFIB FAULT 71 | Excessive internal capacitor discharge events | More than 50 internal dump events occur within 20 minutes | critical |
DEFIB FAULT 71 48 | unknown | ||
DEFIB FAULT 72 | Defibrillator fault code 72 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 72 | General defibrillator malfunction | Unspecified defib system error during operation | critical |
DEFIB FAULT 72 48 | unknown | ||
DEFIB FAULT 76 | Defibrillator fault code 76 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 76 | Defibrillator capacitor voltage too high for selected energy level | Capacitor charging above target for selected energy | error |
DEFIB FAULT 76 48 | unknown | ||
DEFIB FAULT 77 | Defibrillator fault code 77 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 77 | Defibrillator capacitor voltage exceeds absolute maximum rating | Capacitor voltage charge exceeds rated maximum safe voltage | critical |
DEFIB FAULT 77 48 | unknown | ||
DEFIB FAULT 78 | Defibrillation fault 78 | Defibrillation energy delivery failure | error |
DEFIB FAULT 78 | Defibrillator fault code 78 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 78 48 | unknown | ||
DEFIB FAULT 79 | Defibrillator fault code 79 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 79 48 | unknown | ||
DEFIB FAULT 80 | Defibrillator fault code 80 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 80 48 | unknown | ||
DEFIB FAULT 81 | Defibrillator fault code 81 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 81 | Discharge switch in undefined or invalid state | Discharge switch state not recognized during operation | critical |
DEFIB FAULT 81 48 | unknown | ||
DEFIB FAULT 84 | Defibrillator fault code 84 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 84 | Upper discharge transistor shorted (measured via applicable A/D channel). | Upper discharge transistor short circuit | error |
DEFIB FAULT 84 48 | unknown | ||
DEFIB FAULT 85 | Defibrillator fault code 85 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 85 | Lower discharge transistor shorted (measured via applicable A/D channel). | Lower discharge transistor short circuit | error |
DEFIB FAULT 85 48 | unknown | ||
DEFIB FAULT 86 | Defibrillator fault code 86 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 86 | One discharge switch closed during power-up self-test | Power-up self-test detects single discharge switch closed | critical |
DEFIB FAULT 86 49 | unknown | ||
DEFIB FAULT 87 | Defibrillator fault code 87 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 87 | Both discharge switches closed during power up test | Both discharge switches improperly closed at power-up | critical |
DEFIB FAULT 87 49 | unknown | ||
DEFIB FAULT 94 | Defibrillator fault code 94 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 94 | Processor fault rendering safety monitor port non-functional | Main processor detects safety monitor port failure | critical |
DEFIB FAULT 94 49 | unknown | ||
DEFIB FAULT 95 | Defibrillator fault code 95 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 95 | Safe or shutdown line not functional | Safety line circuit failure | critical |
DEFIB FAULT 95 49 | unknown | ||
DEFIB FAULT 96 | Defibrillator fault code 96 | High voltage module or defibrillator circuit malfunction | error |
DEFIB FAULT 96 | XPATREL or XPAT_ENABLE is faulted or discharge transistor shorted | Power delivery control circuit failure or transistor short | critical |
DEFIB FAULT 96 49 | unknown | ||
DEFIB NOT CHARGED | Defibrillator capacitor not charged to required level | Charging cycle incomplete or capacitor discharge detected | error |
DEFIB NOT CHARGED | Defibrillator capacitor not charged | Charging circuit failure or charging timeout | error |
DEFIB NOT CHARGED 49 | unknown | ||
DEFIB PAD SHORT | Defibrillator pad or cable short circuit detected | Continuity measured between defibrillator output and ground | error |
DEFIB PAD SHORT | Impedance between high voltage paddle leads measured as short circuit | Measured impedance between HV leads indicates short | unknown |
DEFIB PAD SHORT | Cable is functioning and connected to pads properly | Normal message when universal cable is connected to shorting test plug | info |
DEFIB PAD SHORT 50 | unknown | ||
DISABLE SYNC | Synchronization mode disabled or cannot be enabled | Sync mode disable request or sync circuit malfunction | warning |
DISABLE SYNC | Sync mode active when analyze pressed during defibrillation | Analyze pressed in defib mode while sync mode enabled | info |
DISABLE SYNC 50 | unknown | ||
DISCHARGE FAULT | Defibrillation discharge did not complete properly | After shock delivery attempt | error |
DISCHARGE FAULT | Defibrillator discharge circuit malfunction | Discharge not detected after defibrillation command | critical |
DISCHARGE FAULT | Defibrillator capacitor voltage not decreasing after discharge attempt | Discharge commanded but capacitor voltage remains high | critical |
DISCHARGE FAULT 50 | unknown | ||
ECG FAULT 4 | Communication fault between ECG processor and main processor | Inter-processor communication failure detected | error |
ECG FAULT 5 | ECU RAM test failure or ROM checksum test failure | Memory self-test failure in ECG processing unit | error |
ECG FAULT 5 | ECG circuit or signal processing fault type 5 | ECG analog front end or signal path malfunction | error |
ECG FAULT 5 | ECG processing fault 5 | During ECG signal processing | error |
ECG FAULT 5 50 | unknown | ||
ECG LEAD OFF | ECG lead disconnection detected | One or more ECG leads disconnected or loose | warning |
ECG LEAD OFF | ECG lead has lost contact with patient or cable is disconnected | Electrode not in contact with patient, cable disconnected, or cable failure | warning |
ECG LEAD OFF | One or more ECG leads are not properly connected when leads are selected as input. | ECG lead disconnected from patient | warning |
ECG LEAD OFF 50 | unknown | ||
ECG TOO LARGE | ECG signal exceeds maximum input range | Excessive ECG signal amplitude detected | warning |
ECG TOO LARGE | ECG signal amplitude exceeds maximum input range | ECG signal saturation detected | warning |
ECG TOO LARGE | ECG signal amplitude too large for accurate analysis | QRS complex exceeds upper limit for algorithm | warning |
ECG TOO LARGE 50 | unknown | ||
ECG TOO SMALL | ECG signal amplitude too low | ECG signal below acceptable range | warning |
ECG TOO SMALL | ECG signal amplitude too small for accurate analysis | QRS complex below minimum for algorithm | warning |
ECG V1 LEAD OFF | Chest lead V1 not properly attached to patient | V1 ECG lead signal absent or not detected | unknown |
ECG V2 LEAD OFF | Chest lead V2 not properly attached to patient | V2 ECG lead signal absent or not detected | unknown |
ECG V2 LEAD OFF | ECG V2 precordial lead disconnected or poor contact | V2 lead signal lost or below threshold | warning |
ECG V2 LEAD OFF | V2 (chest) ECG lead is disconnected or has poor contact | During ECG monitoring | warning |
ECG V2 LEAD OFF 51 | unknown | ||
ECG V3 LEAD OFF | V3 (chest) ECG lead is disconnected or has poor contact | During ECG monitoring | warning |
ECG V3 LEAD OFF | ECG V3 precordial lead disconnected or poor contact | V3 lead signal lost or below threshold | warning |
ECG V3 LEAD OFF | Chest lead V3 not properly attached to patient | V3 ECG lead signal absent or not detected | unknown |
ECG V3 LEAD OFF 51 | unknown | ||
ECG V4 LEAD OFF | ECG V4 precordial lead disconnected or poor contact | V4 lead signal lost or below threshold | warning |
ECG V4 LEAD OFF | V4 (chest) ECG lead is disconnected or has poor contact | During ECG monitoring | warning |
ECG V4 LEAD OFF | Chest lead V4 not properly attached to patient | V4 ECG lead signal absent or not detected | unknown |
ECG V4 LEAD OFF 51 | unknown | ||
ECG V5 LEAD OFF | Chest lead V5 not properly attached to patient | V5 ECG lead signal absent or not detected | unknown |
ECG V6 LEAD OFF | Chest lead V6 not properly attached to patient | V6 ECG lead signal absent or not detected | unknown |
ECG V6 LEAD OFF | ECG V6 precordial lead disconnected or poor contact | V6 lead signal lost or below threshold | warning |
ECG V6 LEAD OFF | V6 (chest) ECG lead is disconnected or has poor contact | During ECG monitoring | warning |
ECG V6 LEAD OFF 51 | unknown | ||
ECU CRC FAULT | ECU data checksum or CRC error detected | Software or firmware data corruption detected | error |
ECU CRC FAULT | Invalid ECG samples detected over one second period | Cyclic redundancy check failure on ECG data | error |
ECU CRC FAULT 51 | unknown | ||
ENTER ACCESS CODE | System requires access code entry for operation | At startup or when accessing service mode | info |
ENTER ACCESS CODE | System requesting access code or PIN entry | Protected function access attempted | info |
ENTER ACCESS CODE 51 | unknown | ||
ERASING REPORT | System erasing stored report data | Report deletion initiated by user or system | info |
ERASING REPORT | Report or data file erase operation in progress | System is actively erasing stored reports or patient data | info |
ERASING REPORT | Report data being erased | User initiated report deletion | info |
ERASING REPORT 51 | unknown | ||
EtCO2 COM ERROR | Capnography module communication error | Serial or network communication failure between main unit and EtCO2 module | error |
EtCO2 COM ERROR | End-tidal CO2 module communication error | EtCO2 module interface communication failure | error |
EtCO2 COM ERROR 51 | unknown | ||
FAX DIALING | Fax transmission is dialing destination number | During automated fax transmission | info |
FAX DIALING | Fax transmission dialing in progress | System initiating fax connection to remote number | info |
FAX DIALING | Fax modem dialing in progress | Fax transmission initiated | info |
FAX DIALING 51 | unknown | ||
FAX DONE | Fax transmission completed | Fax send operation has finished successfully or with error | info |
FAX DONE 51 | unknown | ||
FAX PREPARING | Fax transmission preparation in progress | System preparing data for fax transmission | info |
FAX PREPARING | System is preparing fax data for transmission | During fax setup phase | info |
FAX PREPARING 51 | unknown | ||
FAX SENDING | Fax transmission in progress | Device is actively transmitting fax data over communication line | info |
FAX SENDING | Fax is actively transmitting data | During fax document transmission | info |
FAX SENDING 52 | unknown | ||
INSERT CARD | System requesting PCMCIA card insertion | Card reader detected empty slot | info |
INSERT CARD | Memory card insertion required | System requires PCMCIA or memory card to be inserted in card slot | info |
INSERT CARD | Memory card not installed during manual/semi-automated modes | No card detected at startup or operation | warning |
INSERT CARD 52 | unknown | ||
LOW BATTERY | Internal battery charge level is critically low | When battery voltage drops below minimum threshold | warning |
LOW BATTERY | Battery voltage below minimum operating level | Battery voltage drops below threshold | warning |
LOW BATTERY | Low battery. | Battery charge level low | warning |
LOW BATTERY 52 | unknown | ||
NO CARD INSERTED | Card is not installed in PCMCIA slot | When attempting to read from memory card without card present | warning |
NOISY ECG | ECG signal quality degraded by noise or artifact | Continuous during monitoring when noise detected | warning |
NOISY ECG | Excessive noise in ECG signal exceeding acceptable threshold | Number of noisy analysis intervals exceeds threshold | warning |
NOISY ECG | Number of noisy analysis intervals exceeds threshold. | Excessive noise in ECG signal | warning |
NOISY ECG 52 | unknown | ||
NO QRS DETECT | No QRS complex detected | ECG signal absent or QRS detection failed | warning |
NO QRS DETECT | QRS complex not detected in ECG signal | During ECG analysis or during pacing operation | warning |
NO QRS DETECT | No QRS complex detected in ECG signal | ECG signal processing fails to identify QRS complex | warning |
NO QRS DETECT | Unit is in sync mode and heart rate is < 20 BPM or QRS amplitude is too low for proper synchronization. | Sync mode with low heart rate or low QRS | warning |
NO QRS DETECT 52 | unknown | ||
NO SHOCK AD | Defibrillation not advised by analysis algorithm | After ECG rhythm analysis | info |
NO SHOCK AD | No shock advisable or no shockable rhythm detected | Defibrillation analysis determines rhythm not shockable | info |
NO SHOCK AD 52 | unknown | ||
NO SHOCK ADV | No defibrillation advised; non-shockable rhythm analysis result | ECG analysis complete indicating non-shockable rhythm | info |
OPEN AIR DISCHARGE | Cap voltage too high after discharge attempt, e.g., full energy discharge did not occur. | Discharge without proper load (pads not connected) | critical |
PACER DISABLED | Pacing function disabled via configuration or fault | Pacing capability disabled | critical |
PACER DISABLED | User prompt issued simultaneously with other pace faults if pacing is disabled. | Pacing disabled by configuration or fault | error |
PACER FAULT 115 | Flyback pulse width control circuit is not under proper control of the processor and gate array. | Pacing control circuit failure | error |
PACER FAULT 116 | XPACE_ON detection failure | Unable to detect XPACE_ON signal | critical |
PACER FAULT 116 | Failure to detect XPACE_ON signal | Pace enable signal not detected | critical |
PACER FAULT 117 | Pace relay is stuck closed. | Pacing relay malfunction | error |
PACER FAULT 121 | Pace pulse width outside acceptable range during pacing | Pace pulse width less than 30ms or greater than 50ms | error |
PACER FAULT 121 | During pace, the pace pulse width <30ms or >50ms. | Pacing pulse width out of range | error |
PACER FAULT 122 | Pace current is more than 15mA above and below selected value. | Pacing current deviation exceeds limit | error |
PACER FAULT 123 | Measured pacing rate too fast compared to selected rate | Actual pacing rate exceeds programmed rate | error |
PACER FAULT 126 | Issued in conjunction with message 122. Pace current is more than 15mA and below selected value. | Pacing current below selected value | error |
PADDLE FAULT | Cannot detect paddle type | Accessory type unrecognized | error |
PADDLE FAULT | Cannot detect type of accessory paddle attached to universal cable | Paddle identification failure during initialization | critical |
PERFORM CPR | Advisory message in AED auto defib mode. | No shock advised; CPR needed | info |
PERFORM CPR | Advisory to perform cardiopulmonary resuscitation | Issued in auto defib mode during cardiac arrest | unknown |
PLACE ON ZERO CELL | EtCO2 sensor cable plugged into unit for the first time. Zeroing error or probe drift error detected. | First sensor connection or calibration error | warning |
POOR LEAD CONTACT | One or more ECG leads are poorly connected or not connected to patient. (User configurable.) | ECG lead connection loss or degradation | warning |
POOR PAD CONTACT | Defibrillation electrode impedance exceeds acceptable threshold | Measured impedance between paddle/pads exceeds threshold | unknown |
POOR PAD CONTACT | Defibrillation pad impedance high | Pad electrode impedance exceeds threshold | warning |
PRESS ANALYZE | Alternate message for check patient prompt | User prompt to initiate ECG analysis | info |
PRESS CHARGE | Advisory message; defibrillation advised; press charge button | ECG analysis complete indicating shockable rhythm | unknown |
PRESS SHOCK | Prompt to deliver shock; defibrillator charged and ready | Issued in AED auto defib mode when capacitor fully charged | unknown |
RECORDER FAULT 142 | Paper chart recorder system error | Recorder hardware or paper feed malfunction | error |
RECORDER FAULT 143 | Strip chart failed power-up echo test. Communications error. | Recorder communication failure at boot | error |
RECORDER FAULT 147 | Strip chart printhead over temperature. | Print head temperature excessive | error |
RELEASE BUTTONS | Simultaneous external paddle button presses before full charge | Both paddle buttons pressed during charging | info |
RELEASE SHOCK | Discharge switch(es) closed when pressing charge button. Discharge button pressed before defib reached ready state. | Shock button pressed before charging complete | warning |
RELEASE SHOCK | Discharge button pressed before defib ready or switch closed prematurely | Premature shock attempt or stuck discharge switch | warning |
REPLACE BATTERY | Battery critically low | Battery voltage below absolute minimum | critical |
REPLACE BATTERY | Battery voltage below absolute minimum; shutdown imminent | Battery voltage critically low at power-up | critical |
REPLACE CARD | Memory card write error during manual or semi-automated recording | Attempt to write to memory card during manual or semi-automated mode | warning |
REPLACE EtCO2 SENSOR | EtCO2 sensor defective | Warm-up exceeds 5 minutes | error |
REPORT FULL | Summary report memory full | Report buffer exceeded capacity | warning |
REPORT HALTED | Summary report stops printing unexpectedly | Report printing interrupted during operation | error |
RESERVED 1 | Watchdog timer not functional | Watchdog timer failure detected | critical |
RETRY ANALYSIS | Advisory in conjunction with noisy ECG signal | ECG acquisition with excessive noise | warning |
RETRY ANALYSIS | Retry rhythm analysis | Initial rhythm analysis inconclusive or failed | warning |
RETRY ANALYSIS | ECG analysis failed; system requesting new analysis attempt | After failed rhythm analysis | warning |
RETRY ANALYSIS | Analysis should be retried | Previous analysis failed, incomplete, or inconclusive result | info |
RETRY ANALYSIS | Retry ECG analysis or rhythm interpretation | Initial analysis inconclusive or error; device recommends re-analysis | warning |
RETRY ANALYSIS 56 | unknown | ||
SELECT 30J FOR TEST | Incorrect energy level selected for self-test | Attempt to run self-test at energy other than 30J | info |
SELECT 30J FOR TEST | Self test requested at incorrect energy level | Self-test button pressed at energy other than 30J | warning |
SELECT DEFIB MODE | Analyze button pressed in wrong mode | Analyze button pressed in pace or monitor mode | unknown |
SELECT LIMB LEADS | Invalid lead selection during active analysis | Paddles or augmented leads selected during continuous analysis | unknown |
SELECT LIMB LEADS | Paddles selected when continuous ECG analysis active | Paddle or augmented ECG leads selected during continuous analysis | warning |
SELECT PADS | Incorrect lead selection for analyze function | Lead I, II or III selected when analyze button pressed | info |
SET CLOCK | Real-time clock failure with invalid date or time | RTC malfunction detected or invalid date/time | critical |
SET PACE mA | Multiple copy errors in pace current settings | Pace current software or memory error detected | error |
SET PACE mA | Pacing current setting required or software/memory error | Pacing mode requires current setting or multiple copy errors detected | warning |
SET PACE RATE | Multiple copies of pace rate don't match | Pace rate parameter mismatch in memory | error |
SET PACE RATE | Pacing rate setting required or software error | Pacing mode requires rate setting or multiple pace rate copies do not match | warning |
SHOCK ADVISED | Shockable rhythm detected during analysis | Automatic analysis identifies ventricular fibrillation or pulseless VT | critical |
SpO2 AMBIENT LIGHT | Ambient light intensity too high for accurate SpO2 measurement | SpO2 sensor exposed to excessive ambient light | warning |
SpO2 COMM ERR | No communications received from SpO2 module | SpO2 module communication link lost | error |
SpO2 COMM ERR | SpO2 module communication failure | No transmissions from SpO2 unit received | error |
SpO2 PULSE SEARCH | SpO2 pulse search in progress | SpO2 sensor acquiring pulse signal | info |
SpO2 SENSOR FAULT 1 | Defective SpO2 sensor detected | SpO2 sensor hardware failure | error |
SpO2 SENSOR FAULT 5 | SpO2 sensor unrecognized | Unrecognized sensor type detected | error |
STAND CLEAR | Auto defib mode active with patient rhythm being analyzed | Single analysis mode active in auto defib mode with no defib in progress | info |
STAND CLEAR | Auto defib mode - analysis in progress | Auto defib mode analysis active | info |
SYSTEM FAULT 2 | MCU ROM checksum or RAM test failure | Microcontroller self-test failure detected | critical |
SYSTEM FAULT 36 | P1MON out of range - pace/defib disabled | P1MON measurement out of valid range | error |
SYSTEM FAULT 36 | P1MON count out of range with pace/defib disabled | P1MON less than 412 or greater than 612 A/D counts | critical |
SYSTEM FAULT 37 | Pace/defib and MFE monitoring disabled | Pacing and monitoring system failure | critical |
SYSTEM FAULT 38 | Shutdown failure | Unit fails to shut down on command | critical |
SYSTEM FAULT 38 | Failure to shutdown after shutdown order | System shutdown command failed | critical |
SYSTEM FAULT 5 | ECU RAM or ROM checksum test failure | ECU self-test failure detected | critical |
SYSTEM FAULT 6 | No communications from ECU for 4 seconds | ECU communication timeout | critical |
SYSTEM FAULT 6 | No ECU communications for 4 seconds | ECU communication lost | critical |
SYSTEM FAULT 7 | A/D converter not performing timely conversions | Analog-to-digital converter performance failure | critical |
SYSTEM FAULT 7 | A/D converter failure | Converter not performing timely | critical |
TEST FAILED | Unit failed 30J self-test for defib peak current | MCU defib ipeak test failure during 30J self-test | critical |
TEST FAILED | 30J self-test failed | MCU ipeak test (defib peak current) test failed | error |
TEST OK | Unit passed 30J self-test for defib peak current | MCU defib ipeak test successful | info |
TEST OK | 30J self-test passed | MCU ipeak test (defib peak current) test passed | info |
USE PADDLE DISCHG | Front panel discharge button pressed with incompatible paddles | Discharge button pressed with external paddles or internal spoons with discharge buttons | warning |
USE PADS | Attempt to defib with paddles in auto defib (AED) mode | Paddles selected instead of pads in AED mode | warning |
USE PADS TO PACE | External paddles detected in pace mode | Paddles connected during pacing mode | unknown |
USE ROOM AIR ADAPTER | EtCO2 sensor zeroing with adapter incorrectly placed | Adapter zeroing started with CO2 in adapter or adapter on REF/0 cell | unknown |
USER SETUP REQ | Cal/config data missing or corrupt | Both cal/config copies invalid | warning |
USER SETUP REQ | Both copies of calibration/configuration data corrupted or never programmed | Cal/config data checksum failure or missing initialization | warning |
VF ALARMS OFF | VF alarms disabled in pace mode or with paddles selected | Inappropriate mode for VF alarm operation | warning |
VF ALARMS OFF | Ventricular fibrillation alarms disabled | VF alarms disabled in pace mode or with paddles selected | warning |
V LEADS OFF | V lead connector not terminated or V1 lead wire contacting patient | Missing V lead terminator plug or V1 wire touching patient skin | warning |
VX LEADS OFF | V lead not properly attached to patient | V lead disconnection or poor contact detected | warning |
WRONG CARD TYPE | Card is the wrong type | When inserting wrong type of memory card in PCMCIA slot | warning |
ZERO CO2 ADAPTER | New EtCO2 airway adapter requires zero calibration | New adapter detected without calibration | warning |
ZERO CO2 SENSOR | New EtCO2 sensor requires zero calibration | New sensor detected without calibration | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI