ZOLL E: troubleshooting and error codes
The ServiceForge knowledge base holds 17 documented symptoms for the ZOLL E, with 54 ranked causes, 49 diagnostic checks with test points and expected values, and 128 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Displayed heart rate not accurate (artifact present) patient safety
Heart rate display inaccurate with ECG artifact visible, signal contamination from movement or interference
Ranked causes
- 1. ECG cable or electrode movement causing artifact (common)
- 2. Patient movement or muscle artifact (common)
- 3. Excessive radio frequency interference (RFI) (occasional)
- 4. Poor electrode-to-patient connection (common)
- 5. ECG cable connectivity issue (occasional)
- 6. Contamination on electrode snaps or broken springs (occasional)
- 7. Intermittent ECG patient cable or connector wiring (occasional)
- 8. Defective ECG input connector (rare)
- 9. Faulty ECG connector-to-system board cable (rare)
- 10. Defective system board (rare)
Inspect ECG cable routing and electrode connections
- Test point
- ECG cable path and electrode attachment
- Expected
- Cables routed away from movement, electrodes secure
6 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flash or arcing under defibrillator pad patient safety
Electrical arcing or flashing visible under defibrillator pad during pad application or operation, risk of skin burn
Ranked causes
- 1. Increased skin conductivity from alcohol or betadine preparation (common)
- 2. Gel leakage from pad (gel droop) (common)
- 3. Poor pad coupling to patient's skin (common)
- 4. Dried out gel on defibrillator pad (common)
- 5. Excessive patient hair (common)
- 6. Expired defibrillator pads (occasional)
- 7. Chest compression applied through pads (occasional)
- 8. Front pad placed before back pad positioning (occasional)
- 9. Wet gel pads stored improperly (occasional)
Verify skin preparation method
- Test point
- Patient skin preparation method
- Expected
- No alcohol or betadine used
5 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG LEAD OFF message displays patient safety
Monitor shows ECG LEAD OFF message for 3, 5, or 12-lead cable, indicating loss of signal from electrodes or cable connection issue
Ranked causes
- 1. Poor electrode contact due to skin preparation or dry gel (common)
- 2. Expired ECG electrodes (occasional)
- 3. Loose or disconnected ECG leads (common)
- 4. ECG cable corrosion, broken connector pins, or intermittent connections (occasional)
- 5. Faulty ECG input connector or pins (occasional)
- 6. Defective ECG cable connection to system board (occasional)
- 7. Damaged or degraded system board ECG shielding (rare)
- 8. Defective system board (rare)
Verify electrode site preparation and gel condition
- Test point
- Electrode contact area on patient skin
- Expected
- Site clean, skin lightly abraded, gel moist
6 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK PADS/POOR PAD CONTACT message displays patient safety
Monitor displays CHECK PADS or POOR PAD CONTACT message indicating defibrillator pad connectivity or condition issue
Ranked causes
- 1. Loose or improperly inserted PADS connector (common)
- 2. Damaged defibrillator pads, wires, or connector (common)
- 3. Dried out or expired defibrillator pads (common)
- 4. Poor pad skin contact due to hair or moisture (common)
- 5. Universal cable shorted or open circuit (occasional)
- 6. Faulty High Voltage Module (occasional)
- 7. Defective system board (rare)
Verify PADS connector is properly seated in cable
- Test point
- PADS connector to universal cable connection
- Expected
- Connector firmly inserted with no gaps
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Displayed heart rate not accurate (no artifact) patient safety
Heart rate display on monitor does not match actual patient heart rate; no ECG artifact visible
Ranked causes
- 1. Incorrect lead selection (common)
- 2. Heart rate not flashing synchronously with QRS (common)
- 3. Non-optimal ECG waveform size (occasional)
- 4. Poorly positioned ECG electrodes (common)
Verify heart rate flashes synchronously with QRS
- Test point
- Monitor ECG display
- Expected
- Heart rate indicator flashes with each QRS
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
EtCO2 monitor not displaying waveform patient safety
CO2 waveform does not appear on display after warm-up and zeroing procedure
Ranked causes
- 1. CO2 cable not properly connected to yellow connector (common)
- 2. Monitor not in correct operating mode (occasional)
- 3. CO2 module not warmed up or zeroed (common)
Verify CO2 cable connection to yellow connector
- Test point
- Yellow connector at back of E Series
- Expected
- CAPNOSTAT 5 cable connected with airway adapter or cannula installed
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP calibration fails patient safety
E Series displays FAIL reading instead of PASS during low or high point calibration
Ranked causes
- 1. NIBP Analyzer pressure setting incorrect (common)
- 2. NIBP Analyzer connection to E Series loose or faulty (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
V LEADS OFF message displays patient safety
Monitor shows V LEADS OFF message indicating V lead connection failure or improper termination
Ranked causes
- 1. V lead terminator not attached to unused V lead connector (common)
- 2. V1 lead metal snap in contact with patient skin (occasional)
Check if V leads are in use
- Test point
- V lead connector
- Expected
- Either V leads connected or terminator plug attached
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Strip chart pulses
Verify calibration signal output on strip chart recorder and video display
Ranked causes
- 1. Calibration signal verification required (unknown)
Press RECORDER button
- Test point
- RECORDER button
- Expected
- Recorder activated
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Electronic interference
Radio frequency interference (RFI) or other electronic noise affecting defibrillator operation
Ranked causes
- 1. Excessive radio frequency interference from nearby equipment (occasional)
Identify and mitigate RFI sources
- Test point
- Surrounding electrical equipment
- Expected
- No strong RFI sources nearby or turned off
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
False "Pacer Pulses Seen" message patient safety
Defibrillator displays pacer pulse detection when patient does not have internal pacemaker
Ranked causes
- 1. High-frequency artifact triggering pacer detection (common)
Disable pacer detection feature
- Test point
- Enable Pacer Detection setting
- Expected
- Feature disabled
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery detection
Verify LOW BATTERY message displays and clears at correct voltage thresholds
Ranked causes
- 1. Low battery detection circuit malfunction (unknown)
Verify no low battery alert at safe voltage
- Test point
- Battery well terminals
- Expected
- No LOW BATTERY message V
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Off-mode leakage current test
Measure standby current draw when unit is powered off
Ranked causes
- 1. Excessive leakage current in OFF mode (unknown)
Measure voltage across leakage current shunt resistor
- Test point
- 1kΩ resistor terminals
- Expected
- Less than 450 mV, < 450 mV
- Tools
- DMM set to DCV
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply voltage threshold test high voltage
Verify unit power-on and power-off thresholds at specific voltage levels
Ranked causes
- 1. Power supply voltage out of specification (unknown)
Verify unit does not power on at low voltage
- Test point
- Battery well terminals
- Expected
- Unit should not turn on
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System current (ON mode) test high voltage
Measure current draw when unit is powered on at nominal voltage
Ranked causes
- 1. Excessive current draw in ON mode (unknown)
Measure voltage across current shunt resistor
- Test point
- 0.1Ω resistor terminals
- Expected
- 116 mV or less, ≤ 116 mV
- Tools
- Fluke 75 multimeter or equivalent
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit shutdown voltage test
Verify unit shuts down automatically at low voltage
Ranked causes
- 1. Shutdown circuit malfunction or incorrect threshold (unknown)
Verify automatic shutdown at low voltage
- Test point
- Battery well terminals
- Expected
- Unit should shut off within 30 seconds V
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wandering baseline patient safety
ECG baseline wanders or drifts on display, indicating signal quality issue similar to heart rate inaccuracy
Ranked causes
- 1. ECG signal quality issue (see Displayed HR not accurate) (common)
Follow artifact removal procedures from HR inaccuracy troubleshooting
- Test point
- ECG cable and electrodes
- Expected
- Baseline stabilized
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
200J MAX BIPHASIC | User attempted to set defibrillation energy >200J on Biphasic Unit | User tries to set energy above 200J limit | info |
ANALYSIS HALTED | ECG analysis halted due to user interaction or charging error | Lead/size change, Analyze button pressed again, Impedance fault, or Charging error in auto defib mode | warning |
AUDIO FAULT 136 | Audio DSP hardware error | Audio DSP hardware malfunction | error |
AUDIO NOT RECORDING | Audio is not recording | Recording system malfunction | error |
AUDIO QUEUE FULL | Audio output queue is full | Too many voice prompts queued simultaneously | warning |
BATT HIGH CURRENT | Battery current out of acceptable range | Battery charged with current >0.1A or battery not charged with current >1.6A | error |
BATT HIGH VOLTAGE | Battery voltage exceeds safe limit | Battery voltage >15.5V | error |
BATT LOW CURRENT | Battery current below acceptable range | Battery not charged with current <0.35A | error |
BATT LOW VOLTAGE | Battery voltage insufficient for operation | Battery voltage <9.5V | error |
BATT OVERCHARGE | Battery charger has been on too long | Charger remains on for >4 hours | error |
CABLE FAULT | Incorrect A/D reading for paddle ID in auto defib mode | Paddle ID detection failure | error |
CARD FULL | Memory card is full | Memory card storage capacity exceeded | warning |
CHECK CO2 ADAPTER | Airway adapter is removed, occluded, or zeroing needs to be performed | Adapter missing/occluded or zeroing failed | unknown |
CHECK CO2 SENSOR | EtCO2 Sensor is unplugged or defective | Sensor not detected or sensor failure | unknown |
CHECK PADS | Defibrillation pads have poor contact or are shorted | POOR PAD CONTACT or DEFIB PAD SHORT message | unknown |
CHECK PADS | Defibrillator pad connection or condition issue preventing proper pad function or testing | Loose pad connector, damaged pads or cable, dry/expired pads, poor pad-to-skin contact | error |
CHECK PATIENT | Background ECG analysis detects shockable rhythm | Shockable rhythm detected on monitor | unknown |
CHECK PULSE | Alternate message for NO SHOCK ADVISED | No shockable rhythm detected or after third shock | unknown |
CHECK RECORDER | Paper recorder fault - paper tray empty, jammed, or door open | Paper tray empty, paper jam detected, or recorder door opened | unknown |
CHECK SPO2 SENSOR | SpO2 sensor signal inadequate | SpO2 measurement fails or signal lost | unknown |
CLOCK FAULT 11 | Real time clock oscillator failure | Clock oscillator malfunction at startup | error |
CLOCK FAULT 12 | Real time clock back-up power supply failure | Oscillator stopped at power-up but running when main power applied | error |
CLOCK FAULT 13 | Set time units out of range | Seconds, minutes, year, or other time unit invalid | error |
CONFIRM MANUAL MODE | User confirmation prompt for manual mode entry | Manual mode selection attempted | info |
DEFIB DISABLED | Defibrillator function is disabled | Defib disabled configuration or fault | error |
DEFIB FAULT 108 | VMON voltage is less than the target energy during charging | Voltage monitor reads below target during charge | unknown |
DEFIB FAULT 109 | Defib capacitor voltage exceeds selected energy during charging or ready state | Capacitor voltage higher than programmed energy level | unknown |
DEFIB FAULT 111 | Defib capacitor voltage has exceeded absolute maximum acceptable voltage | Capacitor voltage exceeds absolute maximum rating | unknown |
DEFIB FAULT 195 | Current higher than expected during biphasic bridge test or discharge | Excessive current detected during bridge test or after discharge | unknown |
DEFIB FAULT 196 | Biphasic module not operating properly while charging | Biphasic circuit malfunction during charge cycle | unknown |
DEFIB FAULT 71 | Excessive internal defibrillator dumps detected | More than 50 internal dumps in less than 20 minutes | unknown |
DEFIB FAULT 72 | Defibrillator charging inhibited due to voltage out of range | Charging voltage out of acceptable range | unknown |
DEFIB FAULT 76 | Defibrillator capacitor voltage too high for selected energy | Capacitor overvoltage condition | unknown |
DEFIB FAULT 77 | Defibrillator capacitor voltage exceeds absolute rated maximum | Capacitor voltage > absolute maximum rating | unknown |
DEFIB FAULT 78 | Unable to charge defibrillator capacitor | Capacitor charging failure | unknown |
DEFIB FAULT 79 | Defibrillator charging too slowly | Charging time exceeds expected limit | unknown |
DEFIB FAULT 80 | Multiple defibrillator faults detected in rapid succession | 4 defibrillator faults within 20 second period | unknown |
DEFIB FAULT 81 | Discharge switch in undefined state | Discharge switch malfunction | unknown |
DEFIB FAULT 84 | Upper discharge transistor shorted | Upper transistor short circuit detected via A/D channel | unknown |
DEFIB FAULT 85 | Lower discharge transistor shorted | Lower transistor short circuit detected via A/D channel | unknown |
DEFIB FAULT 86 | One discharge switch closed during power up test | Discharge switch closure detected during power-up self test | unknown |
DEFIB FAULT 87 | Both discharge switches closed during power up test | Both switches closed simultaneously during power-up self test | unknown |
DEFIB FAULT 94 | Processor fault causing safety monitor port to be non-functional | Safety monitor processor failure | unknown |
DEFIB FAULT 95 | Safe or shutdown line is not functional | Safe/shutdown control line malfunction | unknown |
DEFIB NOT CHARGED | Discharge button pressed but unit not charged | Discharge button activated when capacitor voltage insufficient | warning |
DEFIB PAD SHORT | Measured impedance between high voltage leads of MFC is shorted | Low impedance detected between HV output leads | unknown |
DEFIB PAD SHORT | Test message indicating defibrillator pad cable is functioning correctly when connected to shorting plug | Universal cable connected to test shorting plug when pads not applied to patient | info |
DISABLE SYNC | Sync mode active when analyze pressed in defibrillator mode | Sync mode enabled during defibrillator analyze | warning |
DISCHARGE FAULT | Defib capacitor voltage is not decreasing after discharge | Capacitor voltage remains high after discharge attempt | unknown |
ECG FAULT 4 | Communication fault between ECG processor and main processor | ECG/main processor communication link failure | error |
ECG FAULT 5 | ECU RAM test failure or ROM checksum test failure | Memory self-test failure at startup | error |
ECG LEAD OFF | ECG electrode or cable connection has been lost for one or more leads | Electrode detachment, broken connection, dry gel, or cable fault | warning |
ECG LEAD OFF | One or more ECG leads are not properly connected when leads are selected | ECG lead detection failure | unknown |
ECG TOO LARGE | ECG signal too large for accurate shockable rhythm analysis | ECG signal amplitude exceeds analysis range | unknown |
ECG V1 LEAD OFF | Chest lead V1 is not properly attached to patient | V1 lead connection lost | unknown |
ECG V2 LEAD OFF | Chest lead V2 is not properly attached to patient | V2 lead connection lost | unknown |
ECG V3 LEAD OFF | Chest lead V3 is not properly attached to patient | V3 lead connection lost | unknown |
ECG V4 LEAD OFF | Chest lead V4 is not properly attached to patient | V4 lead connection lost | unknown |
ECG V5 LEAD OFF | Chest lead V5 is not properly attached to patient | V5 lead connection lost | unknown |
ECG V6 LEAD OFF | Chest lead V6 is not properly attached to patient | V6 lead connection lost | unknown |
ECG V LEAD OFF | V LEAD cable is not properly attached | V lead cable connector loose or disconnected | unknown |
ENTER ACCESS CODE | Manual mode access code is required to enter manual mode | Manual mode selection attempted on AED | info |
ERASING REPORT | Summary report is being erased from memory | Report erase operation in progress | info |
FAX BUSY | FAX line at receiving FAX is busy | Remote fax machine line busy | info |
FAX DIALING | Preparation for sending fax in progress | Fax transmission initiated | info |
FAX DONE | Fax transmission completed successfully | Fax transmission completed | info |
FAX PREPARING | System preparing fax for transmission | Fax being prepared for send | info |
FAX SENDING | Fax transmission in progress | Fax transmission active | info |
INSERT CARD | No memory card installed in unit | Unit in manual or semi-automated modes without memory card | warning |
LOW BATTERY | Battery voltage insufficient for continued operation | Battery voltage drops below operating minimum | warning |
NOISY ECG | Excessive ECG signal noise detected | Number of noisy analysis intervals exceeds threshold | unknown |
NO QRS DETECT | Unable to detect QRS complex for synchronization | Unit in sync mode with heart rate <20 BPM or QRS amplitude too low | unknown |
NO SHOCK ADV | No shock recommended by analysis | ECG analysis finds non-shockable rhythm | info |
OPEN AIR DISCHARGE | Capacitor voltage too high after discharge attempt | Full energy discharge did not occur, capacitor remains charged | unknown |
PACER DISABLED | Pacing function is disabled | Pacing disabled by configuration or fault | error |
PACER FAULT 115 | Flyback pulse width control circuit not under processor control | Flyback circuit control failure | unknown |
PACER FAULT 116 | Failure to detect XPACE_ON signal | Pace enable signal not detected | unknown |
PACER FAULT 117 | Pace relay is stuck in closed position | Pace relay mechanical failure | unknown |
PACER FAULT 121 | Pace pulse width out of acceptable range | During pacing, pulse width <30ms or >50ms | unknown |
PACER FAULT 122 | Pace current deviation from selected value | Pace current >15mA above and below selected value | unknown |
PACER FAULT 123 | Measured pace rate exceeds selected rate | Pacing rate faster than programmed | unknown |
PACER FAULT 126 | Pace current below selected value threshold | Pace current >15mA below selected value | unknown |
PADDLE FAULT | Cannot detect accessory type on universal cable | Paddle/accessory type detection failure | error |
PERFORM CPR | Operator advisory in AED auto defibrillator mode | AED mode operation | info |
POOR LEAD CONTACT | One or more ECG leads are poorly or not connected | ECG lead detection failure (user configurable) | unknown |
POOR PAD CONTACT | Defibrillator pads have inadequate contact with patient's skin or electrode connection is unreliable | Loose pad connector, damaged pads, poor pad adhesion to skin, cable connectivity issue | error |
POOR PAD CONTACT | Electrode impedance exceeds acceptable threshold | High impedance between defibrillation pads | unknown |
PRESS ANALYZE | Prompt to press analyze button | Operator confirmation needed for rhythm analysis | info |
PRESS CHARGE | Advisory to charge defibrillator | Shock advised, charging authorized | info |
PRESS SHOCK | Prompt to deliver shock in AED mode | Defibrillator charged and ready to deliver shock | info |
RECORDER FAULT 142 | Strip chart recorder system error | Strip chart system malfunction | error |
RECORDER FAULT 143 | Strip chart power-up echo test failed | Recorder communication failure at startup | error |
RECORDER FAULT 147 | Strip chart printhead over temperature condition | Printhead temperature exceeds safe limit | error |
RELEASE BUTTONS | Simultaneous external paddle button presses detected | Both paddle discharge buttons pressed before unit reaches full charge | warning |
RELEASE SHOCK | Discharge switch closed during charge button press | Discharge activated before unit reaches ready state or during charging | warning |
REPLACE BATTERY | Battery voltage at absolute minimum threshold | Battery voltage drops below absolute minimum | critical |
REPLACE CARD | Write errors to memory card during operation | Memory card write failure in manual or semi-automated modes | error |
REPORT FULL | Summary report memory is full | Summary report storage capacity exceeded | warning |
REPORT HALTED | Summary report stops printing unexpectedly | Report print failure or interruption | error |
SELECT 30J FOR TEST | Self test must be run at 30J energy setting | Attempt to run self test at energy other than 30J | warning |
SELECT DEFIB MODE | Analyze button pressed while in wrong operating mode | Analyze pressed in pacing or monitoring mode | warning |
SELECT LEADS | 12 Lead monitor determined user selected MFE leads | MFE (monitoring) leads selected in 12-lead mode | warning |
SELECT LIMB LEADS | Limb leads must be selected for continuous analysis | Paddles or augmented leads selected during continuous analysis | warning |
SELECT PADS | Pads must be selected when using analyze function | Leads I, II, or III selected when analyze pressed | warning |
SET CLOCK | Real time clock date/time is invalid | Invalid date or time values detected | error |
SET PACE mA | Pace current setting error or memory corruption | Multiple copy errors in pace current memory | error |
SET PACE RATE | Pace rate setting error or memory corruption | Multiple copies of pace rate setting do not match | error |
SHOCK ADVISED | Analysis detects shockable rhythm requiring defibrillation | ECG analysis identifies ventricular fibrillation or pulseless VT | unknown |
SpO2 AMBIENT LIGHT | Ambient light is too bright for SpO2 sensor operation | Excessive ambient light at sensor location | unknown |
SpO2 COMM ERR | No communication received from SpO2 module | SpO2 module communication link failure | error |
SpO2 PULSE SEARCH | SpO2 sensor actively searching for pulse signal | Pulse waveform not yet detected | info |
STAND CLEAR | Patient rhythm analysis in progress (AED mode only) | Single analysis mode active with defib idle | info |
SYSTEM FAULT 1 | Watchdog timer failure | Watchdog circuit not responding | error |
SYSTEM FAULT 36 | PS_MON (power supply monitor) out of acceptable range | Power supply voltage monitoring circuit fault | error |
SYSTEM FAULT 37 | Disable pacing, defibrillation, and MFE monitoring functions | Critical system fault detected | critical |
SYSTEM FAULT 38 | Failure to shutdown after shutdown order received | System shutdown sequence failure | critical |
SYSTEM FAULT 6 | No communication from ECG unit (ECU) for 4 seconds | ECU communication link timeout | error |
SYSTEM FAULT 7 | A/D converter not performing conversions in timely manner | Analog-to-digital converter performance degraded | error |
TEST FAILED | Unit failed 30J defibrillator self test | MCU ipeak test (defib peak current) failure at 30J | error |
TEST OK | Unit passed 30J defibrillator self test | MCU ipeak test (defib peak current) passes at 30J | info |
USE PADDLE DISCHG | Front panel discharge button pressed with paddles connected | Front panel discharge activated while external paddles or internal spoons connected | warning |
USE PADS | Must use pads instead of paddles in AED mode | Attempt to defib with paddles in auto defib (AED) mode | warning |
USE PADS TO PACE | External paddles detected when pacing mode attempted | Paddles connected when pacing mode activated | warning |
USER SETUP REQ | Configuration and calibration data missing or corrupted | Both copies of stored calibration/config data invalid or never programmed | error |
VF ALARMS OFF | Ventricular fibrillation alarms disabled | VF alarms disabled in pacing mode or when paddles selected as leads | info |
V LEADS OFF | V leads (precordial leads) have become disconnected or improperly terminated | V lead terminator unplugged when V leads not in use, or V1 lead metal snap contacting patient skin | warning |
VX LEADS OFF | Chest lead V (X = lead number) not properly attached | V lead connection lost or not properly attached to patient | unknown |
ZERO CO2 ADAPTER | New EtCO2 airway adapter requires zeroing calibration | New EtCO2 adapter detected | unknown |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI