FR FieldReady Academy
Defibrillators · E Series family

ZOLL E: troubleshooting and error codes

The ServiceForge knowledge base holds 17 documented symptoms for the ZOLL E, with 54 ranked causes, 49 diagnostic checks with test points and expected values, and 128 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Displayed heart rate not accurate (artifact present) patient safety

Heart rate display inaccurate with ECG artifact visible, signal contamination from movement or interference

Ranked causes

  1. 1. ECG cable or electrode movement causing artifact (common)
  2. 2. Patient movement or muscle artifact (common)
  3. 3. Excessive radio frequency interference (RFI) (occasional)
  4. 4. Poor electrode-to-patient connection (common)
  5. 5. ECG cable connectivity issue (occasional)
  6. 6. Contamination on electrode snaps or broken springs (occasional)
  7. 7. Intermittent ECG patient cable or connector wiring (occasional)
  8. 8. Defective ECG input connector (rare)
  9. 9. Faulty ECG connector-to-system board cable (rare)
  10. 10. Defective system board (rare)
First check

Inspect ECG cable routing and electrode connections

Test point
ECG cable path and electrode attachment
Expected
Cables routed away from movement, electrodes secure

6 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flash or arcing under defibrillator pad patient safety

Electrical arcing or flashing visible under defibrillator pad during pad application or operation, risk of skin burn

Ranked causes

  1. 1. Increased skin conductivity from alcohol or betadine preparation (common)
  2. 2. Gel leakage from pad (gel droop) (common)
  3. 3. Poor pad coupling to patient's skin (common)
  4. 4. Dried out gel on defibrillator pad (common)
  5. 5. Excessive patient hair (common)
  6. 6. Expired defibrillator pads (occasional)
  7. 7. Chest compression applied through pads (occasional)
  8. 8. Front pad placed before back pad positioning (occasional)
  9. 9. Wet gel pads stored improperly (occasional)
First check

Verify skin preparation method

Test point
Patient skin preparation method
Expected
No alcohol or betadine used

5 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG LEAD OFF message displays patient safety

Monitor shows ECG LEAD OFF message for 3, 5, or 12-lead cable, indicating loss of signal from electrodes or cable connection issue

Ranked causes

  1. 1. Poor electrode contact due to skin preparation or dry gel (common)
  2. 2. Expired ECG electrodes (occasional)
  3. 3. Loose or disconnected ECG leads (common)
  4. 4. ECG cable corrosion, broken connector pins, or intermittent connections (occasional)
  5. 5. Faulty ECG input connector or pins (occasional)
  6. 6. Defective ECG cable connection to system board (occasional)
  7. 7. Damaged or degraded system board ECG shielding (rare)
  8. 8. Defective system board (rare)
First check

Verify electrode site preparation and gel condition

Test point
Electrode contact area on patient skin
Expected
Site clean, skin lightly abraded, gel moist

6 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CHECK PADS/POOR PAD CONTACT message displays patient safety

Monitor displays CHECK PADS or POOR PAD CONTACT message indicating defibrillator pad connectivity or condition issue

Ranked causes

  1. 1. Loose or improperly inserted PADS connector (common)
  2. 2. Damaged defibrillator pads, wires, or connector (common)
  3. 3. Dried out or expired defibrillator pads (common)
  4. 4. Poor pad skin contact due to hair or moisture (common)
  5. 5. Universal cable shorted or open circuit (occasional)
  6. 6. Faulty High Voltage Module (occasional)
  7. 7. Defective system board (rare)
First check

Verify PADS connector is properly seated in cable

Test point
PADS connector to universal cable connection
Expected
Connector firmly inserted with no gaps

4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Displayed heart rate not accurate (no artifact) patient safety

Heart rate display on monitor does not match actual patient heart rate; no ECG artifact visible

Ranked causes

  1. 1. Incorrect lead selection (common)
  2. 2. Heart rate not flashing synchronously with QRS (common)
  3. 3. Non-optimal ECG waveform size (occasional)
  4. 4. Poorly positioned ECG electrodes (common)
First check

Verify heart rate flashes synchronously with QRS

Test point
Monitor ECG display
Expected
Heart rate indicator flashes with each QRS

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

EtCO2 monitor not displaying waveform patient safety

CO2 waveform does not appear on display after warm-up and zeroing procedure

Ranked causes

  1. 1. CO2 cable not properly connected to yellow connector (common)
  2. 2. Monitor not in correct operating mode (occasional)
  3. 3. CO2 module not warmed up or zeroed (common)
First check

Verify CO2 cable connection to yellow connector

Test point
Yellow connector at back of E Series
Expected
CAPNOSTAT 5 cable connected with airway adapter or cannula installed

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP calibration fails patient safety

E Series displays FAIL reading instead of PASS during low or high point calibration

Ranked causes

  1. 1. NIBP Analyzer pressure setting incorrect (common)
  2. 2. NIBP Analyzer connection to E Series loose or faulty (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

V LEADS OFF message displays patient safety

Monitor shows V LEADS OFF message indicating V lead connection failure or improper termination

Ranked causes

  1. 1. V lead terminator not attached to unused V lead connector (common)
  2. 2. V1 lead metal snap in contact with patient skin (occasional)
First check

Check if V leads are in use

Test point
V lead connector
Expected
Either V leads connected or terminator plug attached

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Strip chart pulses

Verify calibration signal output on strip chart recorder and video display

Ranked causes

  1. 1. Calibration signal verification required (unknown)
First check

Press RECORDER button

Test point
RECORDER button
Expected
Recorder activated

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Electronic interference

Radio frequency interference (RFI) or other electronic noise affecting defibrillator operation

Ranked causes

  1. 1. Excessive radio frequency interference from nearby equipment (occasional)
First check

Identify and mitigate RFI sources

Test point
Surrounding electrical equipment
Expected
No strong RFI sources nearby or turned off

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

False "Pacer Pulses Seen" message patient safety

Defibrillator displays pacer pulse detection when patient does not have internal pacemaker

Ranked causes

  1. 1. High-frequency artifact triggering pacer detection (common)
First check

Disable pacer detection feature

Test point
Enable Pacer Detection setting
Expected
Feature disabled

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low battery detection

Verify LOW BATTERY message displays and clears at correct voltage thresholds

Ranked causes

  1. 1. Low battery detection circuit malfunction (unknown)
First check

Verify no low battery alert at safe voltage

Test point
Battery well terminals
Expected
No LOW BATTERY message V

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Off-mode leakage current test

Measure standby current draw when unit is powered off

Ranked causes

  1. 1. Excessive leakage current in OFF mode (unknown)
First check

Measure voltage across leakage current shunt resistor

Test point
1kΩ resistor terminals
Expected
Less than 450 mV, < 450 mV
Tools
DMM set to DCV

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power supply voltage threshold test high voltage

Verify unit power-on and power-off thresholds at specific voltage levels

Ranked causes

  1. 1. Power supply voltage out of specification (unknown)
First check

Verify unit does not power on at low voltage

Test point
Battery well terminals
Expected
Unit should not turn on

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System current (ON mode) test high voltage

Measure current draw when unit is powered on at nominal voltage

Ranked causes

  1. 1. Excessive current draw in ON mode (unknown)
First check

Measure voltage across current shunt resistor

Test point
0.1Ω resistor terminals
Expected
116 mV or less, ≤ 116 mV
Tools
Fluke 75 multimeter or equivalent

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit shutdown voltage test

Verify unit shuts down automatically at low voltage

Ranked causes

  1. 1. Shutdown circuit malfunction or incorrect threshold (unknown)
First check

Verify automatic shutdown at low voltage

Test point
Battery well terminals
Expected
Unit should shut off within 30 seconds V

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wandering baseline patient safety

ECG baseline wanders or drifts on display, indicating signal quality issue similar to heart rate inaccuracy

Ranked causes

  1. 1. ECG signal quality issue (see Displayed HR not accurate) (common)
First check

Follow artifact removal procedures from HR inaccuracy troubleshooting

Test point
ECG cable and electrodes
Expected
Baseline stabilized

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
200J MAX BIPHASICUser attempted to set defibrillation energy >200J on Biphasic UnitUser tries to set energy above 200J limitinfo
ANALYSIS HALTEDECG analysis halted due to user interaction or charging errorLead/size change, Analyze button pressed again, Impedance fault, or Charging error in auto defib modewarning
AUDIO FAULT 136Audio DSP hardware errorAudio DSP hardware malfunctionerror
AUDIO NOT RECORDINGAudio is not recordingRecording system malfunctionerror
AUDIO QUEUE FULLAudio output queue is fullToo many voice prompts queued simultaneouslywarning
BATT HIGH CURRENTBattery current out of acceptable rangeBattery charged with current >0.1A or battery not charged with current >1.6Aerror
BATT HIGH VOLTAGEBattery voltage exceeds safe limitBattery voltage >15.5Verror
BATT LOW CURRENTBattery current below acceptable rangeBattery not charged with current <0.35Aerror
BATT LOW VOLTAGEBattery voltage insufficient for operationBattery voltage <9.5Verror
BATT OVERCHARGEBattery charger has been on too longCharger remains on for >4 hourserror
CABLE FAULTIncorrect A/D reading for paddle ID in auto defib modePaddle ID detection failureerror
CARD FULLMemory card is fullMemory card storage capacity exceededwarning
CHECK CO2 ADAPTERAirway adapter is removed, occluded, or zeroing needs to be performedAdapter missing/occluded or zeroing failedunknown
CHECK CO2 SENSOREtCO2 Sensor is unplugged or defectiveSensor not detected or sensor failureunknown
CHECK PADSDefibrillation pads have poor contact or are shortedPOOR PAD CONTACT or DEFIB PAD SHORT messageunknown
CHECK PADSDefibrillator pad connection or condition issue preventing proper pad function or testingLoose pad connector, damaged pads or cable, dry/expired pads, poor pad-to-skin contacterror
CHECK PATIENTBackground ECG analysis detects shockable rhythmShockable rhythm detected on monitorunknown
CHECK PULSEAlternate message for NO SHOCK ADVISEDNo shockable rhythm detected or after third shockunknown
CHECK RECORDERPaper recorder fault - paper tray empty, jammed, or door openPaper tray empty, paper jam detected, or recorder door openedunknown
CHECK SPO2 SENSORSpO2 sensor signal inadequateSpO2 measurement fails or signal lostunknown
CLOCK FAULT 11Real time clock oscillator failureClock oscillator malfunction at startuperror
CLOCK FAULT 12Real time clock back-up power supply failureOscillator stopped at power-up but running when main power appliederror
CLOCK FAULT 13Set time units out of rangeSeconds, minutes, year, or other time unit invaliderror
CONFIRM MANUAL MODEUser confirmation prompt for manual mode entryManual mode selection attemptedinfo
DEFIB DISABLEDDefibrillator function is disabledDefib disabled configuration or faulterror
DEFIB FAULT 108VMON voltage is less than the target energy during chargingVoltage monitor reads below target during chargeunknown
DEFIB FAULT 109Defib capacitor voltage exceeds selected energy during charging or ready stateCapacitor voltage higher than programmed energy levelunknown
DEFIB FAULT 111Defib capacitor voltage has exceeded absolute maximum acceptable voltageCapacitor voltage exceeds absolute maximum ratingunknown
DEFIB FAULT 195Current higher than expected during biphasic bridge test or dischargeExcessive current detected during bridge test or after dischargeunknown
DEFIB FAULT 196Biphasic module not operating properly while chargingBiphasic circuit malfunction during charge cycleunknown
DEFIB FAULT 71Excessive internal defibrillator dumps detectedMore than 50 internal dumps in less than 20 minutesunknown
DEFIB FAULT 72Defibrillator charging inhibited due to voltage out of rangeCharging voltage out of acceptable rangeunknown
DEFIB FAULT 76Defibrillator capacitor voltage too high for selected energyCapacitor overvoltage conditionunknown
DEFIB FAULT 77Defibrillator capacitor voltage exceeds absolute rated maximumCapacitor voltage > absolute maximum ratingunknown
DEFIB FAULT 78Unable to charge defibrillator capacitorCapacitor charging failureunknown
DEFIB FAULT 79Defibrillator charging too slowlyCharging time exceeds expected limitunknown
DEFIB FAULT 80Multiple defibrillator faults detected in rapid succession4 defibrillator faults within 20 second periodunknown
DEFIB FAULT 81Discharge switch in undefined stateDischarge switch malfunctionunknown
DEFIB FAULT 84Upper discharge transistor shortedUpper transistor short circuit detected via A/D channelunknown
DEFIB FAULT 85Lower discharge transistor shortedLower transistor short circuit detected via A/D channelunknown
DEFIB FAULT 86One discharge switch closed during power up testDischarge switch closure detected during power-up self testunknown
DEFIB FAULT 87Both discharge switches closed during power up testBoth switches closed simultaneously during power-up self testunknown
DEFIB FAULT 94Processor fault causing safety monitor port to be non-functionalSafety monitor processor failureunknown
DEFIB FAULT 95Safe or shutdown line is not functionalSafe/shutdown control line malfunctionunknown
DEFIB NOT CHARGEDDischarge button pressed but unit not chargedDischarge button activated when capacitor voltage insufficientwarning
DEFIB PAD SHORTMeasured impedance between high voltage leads of MFC is shortedLow impedance detected between HV output leadsunknown
DEFIB PAD SHORTTest message indicating defibrillator pad cable is functioning correctly when connected to shorting plugUniversal cable connected to test shorting plug when pads not applied to patientinfo
DISABLE SYNCSync mode active when analyze pressed in defibrillator modeSync mode enabled during defibrillator analyzewarning
DISCHARGE FAULTDefib capacitor voltage is not decreasing after dischargeCapacitor voltage remains high after discharge attemptunknown
ECG FAULT 4Communication fault between ECG processor and main processorECG/main processor communication link failureerror
ECG FAULT 5ECU RAM test failure or ROM checksum test failureMemory self-test failure at startuperror
ECG LEAD OFFECG electrode or cable connection has been lost for one or more leadsElectrode detachment, broken connection, dry gel, or cable faultwarning
ECG LEAD OFFOne or more ECG leads are not properly connected when leads are selectedECG lead detection failureunknown
ECG TOO LARGEECG signal too large for accurate shockable rhythm analysisECG signal amplitude exceeds analysis rangeunknown
ECG V1 LEAD OFFChest lead V1 is not properly attached to patientV1 lead connection lostunknown
ECG V2 LEAD OFFChest lead V2 is not properly attached to patientV2 lead connection lostunknown
ECG V3 LEAD OFFChest lead V3 is not properly attached to patientV3 lead connection lostunknown
ECG V4 LEAD OFFChest lead V4 is not properly attached to patientV4 lead connection lostunknown
ECG V5 LEAD OFFChest lead V5 is not properly attached to patientV5 lead connection lostunknown
ECG V6 LEAD OFFChest lead V6 is not properly attached to patientV6 lead connection lostunknown
ECG V LEAD OFFV LEAD cable is not properly attachedV lead cable connector loose or disconnectedunknown
ENTER ACCESS CODEManual mode access code is required to enter manual modeManual mode selection attempted on AEDinfo
ERASING REPORTSummary report is being erased from memoryReport erase operation in progressinfo
FAX BUSYFAX line at receiving FAX is busyRemote fax machine line busyinfo
FAX DIALINGPreparation for sending fax in progressFax transmission initiatedinfo
FAX DONEFax transmission completed successfullyFax transmission completedinfo
FAX PREPARINGSystem preparing fax for transmissionFax being prepared for sendinfo
FAX SENDINGFax transmission in progressFax transmission activeinfo
INSERT CARDNo memory card installed in unitUnit in manual or semi-automated modes without memory cardwarning
LOW BATTERYBattery voltage insufficient for continued operationBattery voltage drops below operating minimumwarning
NOISY ECGExcessive ECG signal noise detectedNumber of noisy analysis intervals exceeds thresholdunknown
NO QRS DETECTUnable to detect QRS complex for synchronizationUnit in sync mode with heart rate <20 BPM or QRS amplitude too lowunknown
NO SHOCK ADVNo shock recommended by analysisECG analysis finds non-shockable rhythminfo
OPEN AIR DISCHARGECapacitor voltage too high after discharge attemptFull energy discharge did not occur, capacitor remains chargedunknown
PACER DISABLEDPacing function is disabledPacing disabled by configuration or faulterror
PACER FAULT 115Flyback pulse width control circuit not under processor controlFlyback circuit control failureunknown
PACER FAULT 116Failure to detect XPACE_ON signalPace enable signal not detectedunknown
PACER FAULT 117Pace relay is stuck in closed positionPace relay mechanical failureunknown
PACER FAULT 121Pace pulse width out of acceptable rangeDuring pacing, pulse width <30ms or >50msunknown
PACER FAULT 122Pace current deviation from selected valuePace current >15mA above and below selected valueunknown
PACER FAULT 123Measured pace rate exceeds selected ratePacing rate faster than programmedunknown
PACER FAULT 126Pace current below selected value thresholdPace current >15mA below selected valueunknown
PADDLE FAULTCannot detect accessory type on universal cablePaddle/accessory type detection failureerror
PERFORM CPROperator advisory in AED auto defibrillator modeAED mode operationinfo
POOR LEAD CONTACTOne or more ECG leads are poorly or not connectedECG lead detection failure (user configurable)unknown
POOR PAD CONTACTDefibrillator pads have inadequate contact with patient's skin or electrode connection is unreliableLoose pad connector, damaged pads, poor pad adhesion to skin, cable connectivity issueerror
POOR PAD CONTACTElectrode impedance exceeds acceptable thresholdHigh impedance between defibrillation padsunknown
PRESS ANALYZEPrompt to press analyze buttonOperator confirmation needed for rhythm analysisinfo
PRESS CHARGEAdvisory to charge defibrillatorShock advised, charging authorizedinfo
PRESS SHOCKPrompt to deliver shock in AED modeDefibrillator charged and ready to deliver shockinfo
RECORDER FAULT 142Strip chart recorder system errorStrip chart system malfunctionerror
RECORDER FAULT 143Strip chart power-up echo test failedRecorder communication failure at startuperror
RECORDER FAULT 147Strip chart printhead over temperature conditionPrinthead temperature exceeds safe limiterror
RELEASE BUTTONSSimultaneous external paddle button presses detectedBoth paddle discharge buttons pressed before unit reaches full chargewarning
RELEASE SHOCKDischarge switch closed during charge button pressDischarge activated before unit reaches ready state or during chargingwarning
REPLACE BATTERYBattery voltage at absolute minimum thresholdBattery voltage drops below absolute minimumcritical
REPLACE CARDWrite errors to memory card during operationMemory card write failure in manual or semi-automated modeserror
REPORT FULLSummary report memory is fullSummary report storage capacity exceededwarning
REPORT HALTEDSummary report stops printing unexpectedlyReport print failure or interruptionerror
SELECT 30J FOR TESTSelf test must be run at 30J energy settingAttempt to run self test at energy other than 30Jwarning
SELECT DEFIB MODEAnalyze button pressed while in wrong operating modeAnalyze pressed in pacing or monitoring modewarning
SELECT LEADS12 Lead monitor determined user selected MFE leadsMFE (monitoring) leads selected in 12-lead modewarning
SELECT LIMB LEADSLimb leads must be selected for continuous analysisPaddles or augmented leads selected during continuous analysiswarning
SELECT PADSPads must be selected when using analyze functionLeads I, II, or III selected when analyze pressedwarning
SET CLOCKReal time clock date/time is invalidInvalid date or time values detectederror
SET PACE mAPace current setting error or memory corruptionMultiple copy errors in pace current memoryerror
SET PACE RATEPace rate setting error or memory corruptionMultiple copies of pace rate setting do not matcherror
SHOCK ADVISEDAnalysis detects shockable rhythm requiring defibrillationECG analysis identifies ventricular fibrillation or pulseless VTunknown
SpO2 AMBIENT LIGHTAmbient light is too bright for SpO2 sensor operationExcessive ambient light at sensor locationunknown
SpO2 COMM ERRNo communication received from SpO2 moduleSpO2 module communication link failureerror
SpO2 PULSE SEARCHSpO2 sensor actively searching for pulse signalPulse waveform not yet detectedinfo
STAND CLEARPatient rhythm analysis in progress (AED mode only)Single analysis mode active with defib idleinfo
SYSTEM FAULT 1Watchdog timer failureWatchdog circuit not respondingerror
SYSTEM FAULT 36PS_MON (power supply monitor) out of acceptable rangePower supply voltage monitoring circuit faulterror
SYSTEM FAULT 37Disable pacing, defibrillation, and MFE monitoring functionsCritical system fault detectedcritical
SYSTEM FAULT 38Failure to shutdown after shutdown order receivedSystem shutdown sequence failurecritical
SYSTEM FAULT 6No communication from ECG unit (ECU) for 4 secondsECU communication link timeouterror
SYSTEM FAULT 7A/D converter not performing conversions in timely mannerAnalog-to-digital converter performance degradederror
TEST FAILEDUnit failed 30J defibrillator self testMCU ipeak test (defib peak current) failure at 30Jerror
TEST OKUnit passed 30J defibrillator self testMCU ipeak test (defib peak current) passes at 30Jinfo
USE PADDLE DISCHGFront panel discharge button pressed with paddles connectedFront panel discharge activated while external paddles or internal spoons connectedwarning
USE PADSMust use pads instead of paddles in AED modeAttempt to defib with paddles in auto defib (AED) modewarning
USE PADS TO PACEExternal paddles detected when pacing mode attemptedPaddles connected when pacing mode activatedwarning
USER SETUP REQConfiguration and calibration data missing or corruptedBoth copies of stored calibration/config data invalid or never programmederror
VF ALARMS OFFVentricular fibrillation alarms disabledVF alarms disabled in pacing mode or when paddles selected as leadsinfo
V LEADS OFFV leads (precordial leads) have become disconnected or improperly terminatedV lead terminator unplugged when V leads not in use, or V1 lead metal snap contacting patient skinwarning
VX LEADS OFFChest lead V (X = lead number) not properly attachedV lead connection lost or not properly attached to patientunknown
ZERO CO2 ADAPTERNew EtCO2 airway adapter requires zeroing calibrationNew EtCO2 adapter detectedunknown

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.