ZOLL AED Plus: troubleshooting and error codes
The ServiceForge knowledge base holds 10 documented symptoms for the ZOLL AED Plus, with 14 ranked causes, 15 diagnostic checks with test points and expected values, and 6 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Status indicator shows red X patient safety
Status indicator LED displays red X instead of normal green check
Ranked causes
- 1. Electrode cable not properly connected (common)
- 2. Defective electrodes (common)
- 3. Internal unit failure (occasional)
Verify electrode cable connection
- Test point
- Patient cable connector
- Expected
- Cable properly seated
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pediatric electrode application incorrect patient safety
Improper placement or application technique of pedi•padz II electrodes may result in poor contact or treatment failure
Ranked causes
- 1. Patient skin not prepared (wet or clothed) (common)
- 2. Electrode not properly adhered or air trapped (common)
Verify chest is dry and clear of clothing
- Test point
- Patient's chest
- Expected
- All clothing removed, chest surface dry
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit beeps while turned off
Device produces audible beep tone when powered off
Ranked causes
- 1. Electrode cable disconnected or loose (common)
- 2. Battery low (occasional)
Verify electrode cable connection
- Test point
- Patient cable connector
- Expected
- Cable properly seated
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrode adhesion issues during application patient safety
Air bubbles trapped under electrodes or electrodes not properly adhered to patient's skin
Ranked causes
- 1. Improper electrode placement technique (common)
Patient chest prepared before electrode application
- Test point
- Patient's chest
- Expected
- All clothing removed and chest dry
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Indication criteria not met for AED use patient safety
AED Plus applied to patient without confirmed cardiac arrest signs (consciousness, normal breathing, or pulse present)
Ranked causes
- 1. Patient does not meet all three indication criteria (unknown)
Confirm patient unconsciousness
- Test point
- Patient responsiveness
- Expected
- Patient does not respond to voice or physical stimulation
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pediatric mode not activated patient safety
AED Plus does not automatically switch to pediatric mode or display 'PEDIATRIC PADS' prompt when pedi•padz II electrodes connected
Ranked causes
- 1. Electrode cable not fully connected (occasional)
Unit mode indication after cable connection
- Test point
- AED Plus display and speaker
- Expected
- Display and voice prompt show 'PEDIATRIC PADS'
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit not equipped for pediatric patients patient safety
AED Plus unit without special symbol cannot treat pediatric patients and will not work with pedi•padz II pediatric electrodes
Ranked causes
- 1. Unit lacks pediatric upgrade (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit not equipped for pediatric treatment patient safety
AED Plus unit without pediatric symbol will not work with pedi•padz II electrodes and cannot treat pediatric patients
Ranked causes
- 1. Unit lacking pediatric upgrade (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit switches to incorrect mode for electrode type patient safety
AED Plus may apply wrong cardiac rhythm detection and shock energy selection when adult or pediatric electrodes are connected
Ranked causes
- 1. Automatic mode switching based on electrode cable type (unknown)
Verify correct mode prompt when connecting electrodes
- Test point
- Voice and display output
- Expected
- ADULT PADS (adult electrodes) or PEDIATRIC PADS (pediatric electrodes)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wrong electrode type for pediatric patient age/weight patient safety
Adult electrodes applied to child under 8 years or weighing less than 55 lbs instead of pediatric electrodes
Ranked causes
- 1. Pediatric criteria not recognized or assessed (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
ANALYSIS HALTED KEEP PATIENT STILL | Patient motion detected during ECG analysis | Excessive patient movement detected during rhythm analysis | warning |
CHANGE BATTERIES | Battery voltage low or depleted | Battery voltage drops below operational threshold | warning |
NO SHOCK DELIVERED (NO TREATMENT DELIVERED) | Shock or treatment not delivered when button activated | Treatment button pressed but no shock delivered | critical |
PLUG IN CABLE | Electrode cable not connected or loose | Cable disconnected from patient cable connector | error |
RELEASE SHOCK BUTTON (RELEASE TREATMENT BUTTON) | Treatment button stuck or held during charging | Treatment button not released during charge cycle | error |
UNIT FAILED | Self-test failed | Status indicator displays unit failure message | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI