ZOLL 731 Z Vent: troubleshooting and error codes

The ServiceForge knowledge base holds 54 documented symptoms for the ZOLL 731 Z Vent, with 89 ranked causes, 74 diagnostic checks with test points and expected values, and 176 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Breath performance parameters out of tolerance patient safety

Tidal volume, PEEP, PIP, BPM, FIO2 or I-Time fail to meet specified tolerances during calibration tests

Ranked causes

  1. 1. Test setup error or apparatus contamination (common)
  2. 2. Test lung defective or O2 sensor contaminated (common)
  3. 3. O2 sensor calibration drift or sensor failure (occasional)
  4. 4. Anti-asphyxia valve damaged or contaminated (occasional)
  5. 5. Ventilator requires recalibration (occasional)
First check

Measure tidal volume at VT=700, FIO2=21, BPM=20, Itime=1000, PEEP=0

Test point
Flow sensor
Expected
700 ml ± (5ml + 8% of setting), ± (5ml + 8%)

5 more checks with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Compressor system check failure

System reports failure when comparing measured flow values to published specifications or when checking intake pressure, RPM, or airway pressure.

Ranked causes

  1. 1. Loose intake assembly (common)
  2. 2. Dirty or clogged disk filter (common)
  3. 3. Anti-Asphyxia valve failure or leak during calibration (occasional)
  4. 4. System leak or high flow measurement (occasional)
  5. 5. Dirty filters (common)
First check

Check airflow tolerance

Test point
measured flow
Expected
within ± (0.1 LPM + 5% of setting)

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Complete power failure patient safety

Loss of power from both internal battery and external source during operation

Ranked causes

  1. 1. External power not applied (common)
  2. 2. Battery disconnected or not charging (common)
  3. 3. PIM board connection issues (occasional)
  4. 4. PIM board needs replacement (occasional)
First check

Apply external power to device

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device does not communicate with Remote Calibration System

USB connection fails, device serial numbers cannot be read, test load cannot be installed

Ranked causes

  1. 1. USB connection problem or hardware issue (common)
  2. 2. FTDI component failure on CPU board (occasional)
  3. 3. Multiple RCS applications running (occasional)
  4. 4. SPM board communication failure (occasional)
First check

Verify USB cable for wear and degradation

Test point
USB connector panel
Expected
No visible damage or wear
Tools
Visual inspection

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ventilator system test failure or misalignment with specifications patient safety

System Tests sequence fails to confirm ventilator operating within expected tolerances or calibration tables do not test properly; limited to three retry attempts before fault recorded

Ranked causes

  1. 1. Incorrect test setup configuration (common)
  2. 2. Compressor not exercised or long-term storage effects (occasional)
  3. 3. USB connection between ventilator and computer insecure or disconnected (occasional)
  4. 4. Incorrect power-on timing relative to RCS prompts (common)
First check

Verify test setup configuration

Test point
Ventilator connections and RCS interface
Expected
Setup matches detailed diagrams and instructions in maintenance chapter

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Compressor flow calibration failure

Remote Calibration System reports failed calibration for compressor flow during discrete tests or system check.

Ranked causes

  1. 1. Compressor flow gain too low (common)
  2. 2. Compressor flow gain too high (occasional)
  3. 3. Dirty disk filter or restricted intake (common)
First check

Review compressor flow offset range

Test point
CF_Offset
Expected
26823 - 35474

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Compressor system calibration failure

Remote Calibration System reports failed calibration for intake pressure or compressor flow during discrete tests.

Ranked causes

  1. 1. Intake pressure gain too low (common)
  2. 2. Intake pressure gain too high (common)
  3. 3. Damaged fresh gas intake filters (occasional)
First check

Review intake pressure calibration range

Test point
IP_Offset
Expected
22588 - 41168

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 Flow Calibration failure

Oxygen flow calibration fails with out-of-range offset or gain values

Ranked causes

  1. 1. Gain calibration error - Gain Too Low (common)
  2. 2. Gain calibration error - Gain Too High (occasional)
  3. 3. SPM/Vent assembly defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

5V bus compromised

Self-check failure of the 5V power bus

Ranked causes

  1. 1. PIM to SPM board connection damaged (common)
  2. 2. PIM board defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Communication error SC1175 occurs during service operation

Alarm SC1175 (Internal COMM) appears intermittently during RCS service procedures; expected during service but not operational use

Ranked causes

  1. 1. Software assert in RCS or intermittent hardware condition (common)
  2. 2. SPM board hardware problem (rare)
First check

Power cycle device

Test point
Device power
Expected
Alarm clears after reset
Tools
Power switch

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Compressor bypass protection alarm not triggered patient safety

Fresh gas intake blockage does not trigger intake blocked alarm or alarms incorrectly indicate partial blockage

Ranked causes

  1. 1. Loose or incomplete intake assembly blockage (common)
  2. 2. Partial blockage or incorrect test procedure (occasional)
First check

Verify intake blocked alarm 1030 triggers during fresh gas intake occlusion

Test point
Intake assembly
Expected
Alarm 1030 triggered

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

External battery fault indicator

Possible faulty wiring of the external power connector

Ranked causes

  1. 1. External power connector wiring fault (common)
  2. 2. External power cable defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Internal communication battery fault (no external power) patient safety

Communication failure to device main battery without external power

Ranked causes

  1. 1. Battery connection to PIM board loose or faulty (common)
  2. 2. Battery or PIM defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Internal communication battery fault (with external power) patient safety

Communication failure to device main battery while operating on external power

Ranked causes

  1. 1. Battery connection to PIM board loose or faulty (common)
  2. 2. Battery or PIM defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Internal communication power components fault

Communication error with the PIM (power interface module)

Ranked causes

  1. 1. PIM to SPM or CPU connection fault (common)
  2. 2. PIM or CPU connector cable defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD is dark or GUI is blank during service

Screen shows vertical line only or remains OFF during RCS programming, expected behavior during reset state

Ranked causes

  1. 1. Device in reset or intermediate state (common)
  2. 2. LCD or display hardware failure (occasional)
First check

Observe LCD during startup

Test point
LCD screen
Expected
Normal display after power cycle or vertical line during reset
Tools
Visual inspection

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

RTC battery low or date/PM due alarm resets caution

Alarm 3110 indicates low RTC battery; date and PM due settings reset frequently after battery replacement

Ranked causes

  1. 1. RTC battery voltage low (common)
  2. 2. RTC battery retention clip bent during installation (occasional)
First check

Verify RTC battery is properly seated

Test point
RTC battery slot
Expected
Battery seated firmly with retention clip secured
Tools
Visual inspection

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 readings not displayed or sensor LEDs not illuminated patient safety

Pulse oximeter sensor fails to illuminate and no SpO2 readings appear on ventilator display

Ranked causes

  1. 1. Defective SpO2 sensor or cable (common)
  2. 2. Masimo board failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

System test failure during service

Device test fails during RCS System Tests; requires troubleshooting before calibration can proceed.

Ranked causes

  1. 1. Setup error or equipment configuration issue (common)
  2. 2. Identified failure in ventilator or calibration (occasional)
First check

Physical Inspection

Test point
Device exterior
Expected
No damage or defects; room temperature 16–28 °C/60–82 °F

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

TSI calibration out of specification

TSI barometric pressure, temperature, or calibration date exceeds valid operating range during test station operation

Ranked causes

  1. 1. Environmental conditions outside specification (temperature or pressure) (common)
  2. 2. TSI calibration date expired (common)
First check

Verify TSI barometric pressure reading

Test point
TSI sensor
Expected
69kPa < reading < 102kPa, 69–102

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Button Check failure

One or more buttons on device not functioning correctly

Ranked causes

  1. 1. Button hardware defective (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Buzzer Check failure patient safety

Device buzzer not producing audible alarm sound

Ranked causes

  1. 1. Buzzer hardware defective (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: 731 Series Ventilator patient safety

Service calibration procedure using RCS 9.1 software; requires Calibrations sequence to pass, followed by System Tests and forensic memory erasure

Ranked causes

  1. 1. Calibration setup via RCS 9.1 (common)
First check

Run Calibrations sequence from Service Actions list

Test point
RCS 9.1 software
Expected
PASS status with no failures

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Erase Forensic Memory

Clear ventilator forensic memory after system tests and calibrations

Ranked causes

  1. 1. Forensic memory requires erasing as part of service procedure (common)
First check

Run Erase Forensic Memory from Service Actions list

Test point
RCS 9.1 interface
Expected
Erase Forensic Memory completes successfully

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: O2 Kickstart

Near zero flow target calibration for fine flow control after opening valve

Ranked causes

  1. 1. Oxygen Delivery System restriction or gas flow issue (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Remote Calibration System (RCS) preventive maintenance patient safety

The RCS performs testing and calibration of the ventilator through discrete tests, either individually for troubleshooting or as part of larger sequences (Calibrations, System Tests). TSI Flow Module requires yearly calibration before ventilator calibration.

Ranked causes

  1. 1. Preventive maintenance due or PM/Calibration Due alarm (#3120) triggered (common)
First check

Verify TSI Flow Module calibration date

Test point
TSI Flow Module
Expected
Last calibration date within one year

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: System Tests patient safety

Run System Tests sequence from Service Actions after successful Calibrations to verify ventilator operation

Ranked causes

  1. 1. System Tests must follow successful Calibrations (common)
First check

Run System Tests sequence

Test point
Service Actions menu
Expected
System Tests completes without failures

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Write serial numbers to ventilator

Update serial numbers for new parts or recover corrupted component data using RCS software

Ranked causes

  1. 1. New serialized parts installed or data corruption recovery required (unknown)
First check

RCS software launched and ventilator communication established

Test point
RCS user interface Read Vent button
Expected
Communication established between RCS application and ventilator

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Comm RAM and Bootloader Check failure

EMV COMM interface, CPU RAM, bootloader, or SPM communication not functioning

Ranked causes

  1. 1. Front panel assembly defective (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

EMV Self Check failure

RCS detects active alarms on EMV indicating power system, communications, or pulse oximeter problems

Ranked causes

  1. 1. Power system, communications, or pulse oximeter fault (unknown)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Encoder Check failure

Rotary encoder not functioning in one or both directions

Ranked causes

  1. 1. Rotary encoder hardware defective (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exhalation valve timing out of specification

Exhalation valve rise or fall delay exceeds maximum allowed time at reference pressures

Ranked causes

  1. 1. SPM assembly or board fault (common)
First check

Measure autocal rise settling time at 4 cmH2O

Test point
Autocal valve output
Expected
< 30ms, < 30ms

6 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Input protection circuit failure

Damaged fuse on the PIM board

Ranked causes

  1. 1. PIM board fuse damaged (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD Contrast Set failure

LCD contrast cannot be properly adjusted; contrast control not functioning

Ranked causes

  1. 1. Ribbon cable from LCD to button board not seated correctly (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD Visual Inspection failure

LCD display defects detected; stuck pixels, blemishes, backlight or contrast problems

Ranked causes

  1. 1. LCD defective with stuck pixels, blemishes, backlight or contrast issues (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

LED Check failure

One or more device LEDs not functioning; green, yellow, red, or IR LED defective

Ranked causes

  1. 1. LED hardware defective (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Manifold pressure leak detected caution

Airway pressure slope during simulated expiratory phase exceeds specification indicating system leak

Ranked causes

  1. 1. Leak in manifold connections or tubing (common)
First check

Measure airway pressure slope at 5 cmH2O

Test point
Airway pressure sensor
Expected
< 0.5 cmH2O per second cmH2O/s, < 0.5 cmH2O per second

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Masimo communication error alarm

Masimo communication thread out or cable/board connection damaged; codes 2301 or 3301 displayed

Ranked causes

  1. 1. Masimo cable or board connection issue (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Masimo module failure alarm during self-check patient safety

Internal OEM Masimo board indicates failure during system self-check; codes 2300 or 3300 displayed

Ranked causes

  1. 1. CPU/SPO2 Stack or Front Panel Assembly failure (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 filter resistance out of specification

Sintered bronze filter pressure drop measured outside acceptable range at 100 LPM

Ranked causes

  1. 1. Sintered bronze filter contaminated or clogged (common)
First check

Measure filter resistance at 100 LPM flow

Test point
O2 inlet filter
Expected
0.01–0.06 psi/LPM, 0.01–0.06 psi/LPM

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 regulator damaged or not functioning properly caution

O2 regulators may be damaged from improper use, affecting oxygen supply to test station

Ranked causes

  1. 1. Oxygen regulator not closed or over-tightened (common)
First check

Verify O2 regulator is not over-tightened

Test point
O2 regulator handle
Expected
Hand-tight only, not over-tightened

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 supply hose damaged or O-rings compromised caution

O2 hoses show signs of damage, debris, or missing/damaged O-rings affecting oxygen delivery

Ranked causes

  1. 1. O-rings missing or damaged in O2 hose fittings (occasional)
First check

Inspect O2 hose O-rings and connections

Test point
O2 data cable and O2 cable connections
Expected
O-rings in place, no visible damage or debris

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 supply pressure out of specification caution

O2 input pressure measured outside acceptable limits at reference pressures

Ranked causes

  1. 1. O2 supply pressure incorrect (common)
First check

Measure O2 input pressure at 90 psi reference

Test point
O2 inlet
Expected
+/- 1 psi, +/- 1 psi

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 system leak rate excessive caution

O2 leak rate exceeds maximum allowed specification at reference pressure

Ranked causes

  1. 1. Leaks in O2 tubing or fittings (common)
First check

Measure O2 leak rate at 55 psi

Test point
O2 system output
Expected
<= 2000 ml/hour, <= 2000 ml/hour

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 valve leakage excessive caution

Reported O2 leak through valve exceeds maximum allowed specification

Ranked causes

  1. 1. O2 valve damaged or tubing kinked (common)
First check

Measure reported O2 leak through valve

Test point
O2 valve outlet
Expected
<= 0.3 LPM, <= 0.3 LPM

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power Switch Check failure

Power switch not functioning correctly; unit does not shut off or turn on as expected

Ranked causes

  1. 1. Power switch functional issue (common)
First check

Unit fully shuts off with no external power

Test point
Power switch Off position with no external power
Expected
Off and No External Power = Pass

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Reported O2 supply pressure out of limits caution

O2 supply pressure reported by system outside acceptable limits at zero and 70 psi reference

Ranked causes

  1. 1. O2 supply tubing kinked or damaged (common)
First check

Measure reported O2 supply pressure at zero reference

Test point
O2 system sensor
Expected
< 2 psi

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

RTC date and time check failure

RCS verification of date and time in device versus PC failed; RTC battery or connector may be defective

Ranked causes

  1. 1. RTC battery depleted or faulty battery connector (common)
First check

RCS verifies date and time in device matches PC

Test point
SPM real-time clock module
Expected
Date and time match PC value

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

RTC date and time set failure

RCS unable to set date and time in ventilator; may indicate RTC battery depletion

Ranked causes

  1. 1. RTC battery depleted (common)
First check

RCS attempts to set date and time

Test point
SPM real-time clock module
Expected
Date and time written to device

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Self-Check Failure: O2 Valve Stuck Open

System detects oxygen valve stuck in open position

Ranked causes

  1. 1. O2 Valve Stuck Open (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SPM blink LED and buzzer failure

SPM unable to drive red LEDs and buzzer; operator confirmation not received within 15 seconds during test

Ranked causes

  1. 1. SPM board faulty or cable connections failed (common)
First check

Verify red LEDs and buzzer activation

Test point
SPM output
Expected
Red LEDs and buzzer driven within 15 seconds

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SPM communication failure

RCS test confirms communication signals (RX & TX), bootload and toggle are not functional

Ranked causes

  1. 1. SPM board faulty or connections to SPM failed (common)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Total Power Failure Alarm Check failure patient safety

Power failure circuit not functioning; total power loss alarm does not trigger or display correctly

Ranked causes

  1. 1. Power failure circuit defective; alarm does not trigger on total power loss (occasional)
First check

Verify Total Power Failure alarm activation

Test point
Power failure circuit alarm output
Expected
120 second Total Power Failure alarm triggers (fast beep and red alarm LED blink pattern) and lasts for at least 10 seconds.

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

User prompt not visible on screen

Expected UI prompt does not appear to the operator during normal operation

Ranked causes

  1. 1. Prompt hidden behind application window (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
1001Self-Check Failure or Self Check FaultCompressor control fault detectederror
1001Compressor failureCompressor fails to operate or fails to provide required flow for breath delivery when high pressure O2 unavailablecritical
1001Compressor does not respond to control signalsAt power up or during operation; blocked output tubing, kinked tubing, loose connector, diffuser or flow screen issuecritical
1002Compressor flow signal high, low, or stuckCompressor internal communication failure; faulty compressor flow transducer or tubing issuecritical
1002Compressor communication faultCommunication lost between compressor controller and Smart Pneumatic Module when high pressure O2 unavailablecritical
1002Self Check FailurePGA or ADC faults for Compressor System detectederror
1003Low compressor speed at startupFlow from first breath is ±20% of expected flow for tidal volume at start uperror
1003Fails power-on self check breathSensor/transducers/calibration file failure at startupcritical
1010O2 valve fails in open positionO2 valve failure causing continuous flow during expiratory phase, preventing ventilationcritical
1011Loss of O2 flow controlO2 flow path failure or fault preventing O2 flow controlcritical
1011O2 valve flow delivery failureSignal to O2 valve not delivering required flow rate and compressor unavailablecritical
1011Control fault indicating insufficient O2 flow rateSelf-check detects inadequate oxygen flowerror
1012PGA or ADC faults detected for O2 SystemSelf-check detects signal processing electronics failureerror
1012O2 valve communication failureCommunication lost between O2 valve and SPM when compressor unavailable for ventilationcritical
1012O2 flow sensor signal to SPM failsInternal communication failure between O2 flow sensor and SPMcritical
1020Low oxygen supply pressure (< 35 psig)O2 supply pressure falls below minimum thresholderror
1030Gas Intake FailureInsufficient intake flow detectederror
1030Fresh gas or emergency air inlet blockedFresh Gas/Emergency Air Inlet blocked preventing compressor from delivering sufficient flowcritical
1030Insufficient flow or blocked fresh gas intakeGas intake failure or fault preventing fresh gas intakecritical
1041High oxygen supply pressure (> 87 psig)O2 supply pressure exceeds maximum thresholdwarning
1051Autocal valve malfunctionAfter 10 breaths, valve does not open, opens part way, or readings too noisyerror
1051Run-time calibration failureSelf check failure during runtime or startupcritical
1052Airway pressure signal stuck high, low, or unchangingSelf check failure during power-on or runtimecritical
1052Airway pressure sensor communication lossCommunication lost between airway pressure sensor and SPMerror
1052Self-check failure: airway pressure transducerTransducer not behaving as expected during self-checkerror
1060Exhalation system valve failurePIP fails to return to baseline pressure for 3 consecutive breathscritical
1060PEEP higher than limit for 3 consecutive breathsExhalation system failure with excessive PEEPcritical
1061Airway pressure excessive for extended timeExhalation system failure with airway pressure >70 cmH2O for 1.5s or >35 cmH2O/PIP high limit for 5scritical
11725V self check failureSelf check failure during power-upcritical
11725V bus compromisedSelf-check failure of the 5V buserror
1173Communication failurePIM to CPU communication fault at power-upcritical
1174Unable to zero/null transducer offsetSelf check failure unable to null one or more transducer offsetscritical
1174Self Check FailureFlow transducers behaving outside specificationserror
1174PGA offset control calibration failureDevice unable to calibrate transducers at startup; PGA offset control failscritical
1174Flow transducers are not behaving as expectedSelf-check detects flow sensor anomalyerror
1175Internal communication bus failureInternal communication bus and host unable to communicate with sub-assembliescritical
1175Loss of I2C communication for 4 consecutive attemptsSelf check failure losing I2C bus communicationcritical
1176Calibration table or header CRC errorSelf check failure with corrupted calibration tablecritical
1176Corrupted calibration file detectedCalibration file checksum failure or file corruption during writewarning
1420Complete power failurePower lost from both internal battery and external source during operationcritical
1430Drained batteryInternal battery power drops below amount required for ventilation without external powererror
1471User interface module communication failureDevice unable to communicate with User Interface Module and interface controlserror
1471Cannot communicate with IO expander for buttons and LEDsSelf check failure on IO expander communicationerror
1472High rate of corrupted packets from SPMSelf check detects excessive SPM packet corruptionerror
1472Smart pneumatic module communication failureDevice unable to communicate with Smart Pneumatic Moduleerror
1473SPM data timeoutNo valid data sent from SPM within 1 seconderror
1474Cyclic redundancy check failureCRC checking fails between device and SPMerror
1474CPU and PIM communication gaps detectedSelf check detects communication gaps between CPU and PIMerror
1475Cannot communicate with LCD contrast control circuitSelf check failure on LCD contrast control communicationerror
1475Contrast control communication failureDevice lost communication with contrast controlerror
1480CPU and SPM software revisions incompatibleSelf check detects version mismatch between CPU and SPMerror
2001Compressor does not respond to control signals (medium priority)Medium priority version of code 1001error
2001Self Check FaultCompressor control fault detectederror
2002Self Check FaultPGA or ADC faults for Compressor System detectederror
2002Compressor flow signal high, low, or stuck (medium priority)Medium priority version of code 1002error
2010O2 Valve Stuck OpenAlarm code variant of 1010error
2011Self Check FaultCommunication between O2 valve and SPM fails and compressor is available to provide ventilationwarning
2011Control fault indicating insufficient O2 flow rateAlarm code variant of 1011error
2011Loss of O2 flow control (medium priority)Medium priority version of 1011error
2012O2 flow sensor signal to SPM fails (medium priority)Medium priority version of 1012error
2012Self Check FaultSignal to O2 valve outside calibration range for required flow rate and compressor availablewarning
2020O2 supply pressure less than 35 psiLow O2 supply fault conditionerror
2030Insufficient flow or blocked fresh gas intake (medium priority)Medium priority version of 1030error
2030Gas Intake FaultFresh Gas/Emergency Air Inlet blocked; compressor unable to deliver breath within ±10% of settings; high pressure O2 availableerror
2041High oxygen supply pressure (> 87 psig)Alarm code variant of 1041warning
2053Self Check FaultExpiratory time < 170 ms for 3 consecutive breathswarning
2062Exhalation System FaultAirway pressure PIP measured at end of expiration > 5 cm H2O above baseline pressure PEEPerror
2062PEEP measured exceeds setting plus 5 cmH2OExhalation system fault with elevated PEEPerror
2070Airway Pressure HighAirway pressure PIP greater than high airway pressure limit for 2 consecutive breathswarning
2071Low Airway PressureAirway pressure PIP less than low airway pressure limit for 2 consecutive breathswarning
2072High Tidal VolumeDuring pressure targeted ventilation, delivered tidal volume exceeds user defined limit for 2 consecutive breathswarning
2073Low Tidal VolumeDuring pressure targeted ventilation, delivered tidal volume does not reach user-defined limit for 2 consecutive breathswarning
2074High Breath RateActual breathing rate (set rate plus spontaneous patient rate) exceeds high alarm limitinfo
2075Low Breath Rate/ApneaActual breathing rate (set rate plus spontaneous patient rate) less than low alarm limitwarning
2076ApneaSpontaneous breathing rate less than low alarm limit in noninvasive ventilation, CPAP and BL modeswarning
2090PEEP LeakAirway pressure drops below PEEP setting by 2 cm H2O during expiratory phasewarning
2095Insufficient FlowPressure target not reached during inspiratory period during pressure targeted ventilationwarning
2100Patient DisconnectAirway pressure fails to exceed PEEP setting by approximately 7 cm H2Oerror
2170Spont. Breath PIP HighAirway pressure PIP exceeds High PIP Limit Setting during 2 consecutive spontaneous breathswarning
2171Spont. Breath PIP LowAirway pressure PIP exceeds Low PIP Limit Setting during 2 consecutive spontaneous breathswarning
2172Spont. Breath Vt HighHigh VT Limit exceeded during 2 consecutive spontaneous breathswarning
2173Spont. Breath Tt LowLow VT Limit Setting not achieved during 2 consecutive spontaneous breathswarning
2300Self-Check: Pulse Ox - Module Failure (Internal OEM Board)Masimo board indicates failure during power-up self-checkcritical
2300Self Check FaultPulse oximeter module fails while in useerror
2300Masimo board indicates a failureSelf check detects board failureerror
2301Communication from Masimo board has timed outSelf check timeout on board communicationerror
2301Self Check FaultCommunication between pulse oximeter module and device failserror
2301Internal COMM: Masimo communication thread out or damagedMasimo communication link failure or thread timeouterror
2314Pulse Ox Sensor Off PatientOperating sensor loses patient signal due to disconnection, misalignment or poor perfusionwarning
2401SpO2 LowSpO2 value drops below Low SpO2 Limit (default 94%)warning
2410Heart Rate HighHeart rate greater than High Heart Rate Limit (default 120 beats/min)warning
2411Heart Rate LowHeart rate less than Low Heart Rate Limit (default 40 beats/min)warning
2421Self Check FaultFailure of input protection circuit and device is able to operateerror
2421Input protection circuit failureDamaged fuse on the PIM boarderror
2423Self Check FaultInternal power circuit failed and external power connected but cannot be usederror
2423Power system failureSelf check detects power faulterror
2430Nearly Drained BatteryDevice detects ≤5 minutes of battery operation remaining and external power not connectedwarning
2450Battery Discharge FaultBattery temperature reaches 70°C (158°F), 5°C from maximum operating temperatureerror
2450Battery pack internal temperature too highBattery internal temperature ≥ 70 °Cwarning
2455Battery communication fault while powered by batterySystem powered by battery AND battery communication failserror
2455Internal COMM battery faultCommunication failure to device main battery without external powererror
3001Self Check FaultCompressor fails to operate or provide required flow within ±10% of settings; high pressure O2 available; FIO2 set to 100%warning
3001Compressor does not respond to control signals (low priority)Low priority version of code 1001warning
3002Self Check FaultCommunication between compressor controller and SPM lost; high pressure O2 available; FIO2 set to 100%warning
3003Compressor flow signal high, low, or stuck (low priority)Low priority version of code 1002warning
3011Loss of O2 flow control (low priority)Low priority version of 1011warning
3011Self Check FaultSignal to O2 valve outside calibration range; compressor available; user set FIO2 to 21%warning
3012Self Check FaultCommunication between O2 valve and SPM lost; compressor available; FIO2 set to 21%warning
3012O2 flow sensor signal to SPM fails (low priority)Low priority version of 1012warning
3030Insufficient flow or blocked fresh gas intake (low priority)Low priority version of 1030warning
3030Gas Intake FaultFresh Gas/Emergency Air Inlet blocked; compressor unable to deliver breaths within ±10%; high pressure O2 available; FIO2 set to 100%error
3031Intake RestrictedFresh Gas/Emergency Air Inlet blocked but still capable of delivering breaths within ±10% of settingswarning
3031Insufficient flow from fresh gas intake (intake restricted)Fresh gas intake restricted or partially occludedwarning
3032Fresh gas intake pressure sensor signal stuck high or lowSensor and transducer fault on fresh gas/intake pressurewarning
3032Self Check FaultInlet pressure transducer malfunction detectederror
3041High O2 Supply FaultHigh pressure O2 supply ≥75 psig (517 kPa) and < 80 psig (552 kPa)warning
3041O2 input pressure too high (low priority)Low priority version of 1041warning
3073Tubing Compliance FaultTubing compliance correction greater than set tidal volumewarning
3091AutoPEEPExhaled flow continues throughout expiratory periodwarning
3092Inspiratory DemandEnd-inspiratory pressure < -1.0 cm H2O for 3 consecutive breathswarning
3110RTC battery drained at startupReal-time clock alarm at power-up with date/time reset to defaultwarning
3120PM DuePreselected number of days elapsed since last calibration at startupinfo
3120Calibration due or overdueCalibration interval exceeded or RTC battery issuewarning
3120PM Due - Preventive Maintenance requiredDevice reaches one year anniversary since last serviceinfo
#3120PM/Calibration Due alarmPreventive maintenance interval expired or calibration overduewarning
3121Power Cycle NeedDevice running continuously for 30 dayswarning
3130Ambient pressure sensor reads near zero or full scaleSelf check fault on barometric pressure sensorwarning
3131Excessive altitude detectedAmbient pressure <0.36 ATM indicating altitude greater than 25,000 feetwarning
3132Low altitude detected (below -2000 feet)Ambient pressure >1.1 ATM indicating altitude below -2000 feetwarning
3132Low AltitudeAmbient pressure transducer detects altitude < -2,000 feet below sea level (610 meters, 15.8 psig or 103 kPa)warning
3140Ambient Temperature FaultAmbient temperature exceeds normal operating range > 131 °F (55 °C)warning
3141Ambient Temperature FaultAmbient temperature falls below normal operating range < 14 °F (-10 °C)warning
3143Temperature sensor faultSelf check fault with temperature sensor reading beyond thresholdwarning
3172Self-check fault: airway pressure transducer cannot zeroTransducer unable to zero during initializationerror
3300Alarm priority for Pulse Ox - Module FailureSame condition as 2300 with alarm priority indicationcritical
3300Masimo board indicates a failureSelf check detects board faultcritical
3300Pulse Oximeter Module Self Check FaultPulse oximeter module failure with user acknowledging conditionerror
3301Communication from Masimo board has timed outSelf check timeout on board communicationcritical
3301Pulse Oximeter Communication Self Check FaultCommunication between pulse oximeter module and device failserror
3301Alarm priority for Internal COMM error (dependent on sensor usage)Same as 2301 with alarm priority based on Masimo sensor active statuserror
3310Pulse Ox Sensor Not ConnectedPulse oximeter detects no SpO2 sensor after successful operationwarning
3311Defective pulse oximeter sensorPulse oximeter cannot identify or sensor has failederror
3312Pulse SearchPulse oximeter is searching for pulse signalinfo
3313Pulse Ox Signal InterferenceOutside signal or energy source prevents accurate readingwarning
3315Ambient Light FaultExcessive ambient light on SpO2 sensor or inadequate tissue coveringwarning
3316Invalid pulse oximeter sensorPulse oximeter does not recognize connected sensorwarning
3317Low SpO2 PerfusionAmplitude of arterial pulsation is weakwarning
3318Low SpO2 PerfusionQuality of input signal is low due to excessive motion or artifactwarning
3421External Power Low/DisconnectExternal power drops below ~11 VDC or is disconnectederror
3422Battery pack not detectedDevice unable to detect batteryerror
3423Battery Charging FaultBattery charging circuit failserror
3430Low BatteryDevice detects < 30 minutes battery operation remaining with no external powerwarning
3431Low BatteryExternal power connected to device with internal battery drained to low statuswarning
3441External power input voltage too highInput voltage exceeds 33 VDC after fuseerror
3442External power input voltage too lowInput voltage below 11.5 VDC after fuseerror
3444External power voltage negativeVoltage in at less than -1 VDCerror
3450Battery Discharge FaultBattery temperature reaches 70 °C (158 °F) with external power connectedwarning
3451Battery discharge fault at high temperatureBattery internal temperature ≥ 75 °Ccritical
3452Battery charging fault due to high temperatureExternal power present AND battery temperature ≥ 45 °Cwarning
3453Battery charging fault due to low temperatureExternal power present AND battery temperature ≤ 0 °Cwarning
3455Battery communication fault while powered by batterySystem powered by battery AND battery communication failscritical
3455Battery FaultDevice unable to communicate with internal battery with external power connectederror
3455Internal COMM battery fault (external power active)Communication failure to device main battery while operating on external powererror
3470Internal COMM power components faultCommunication error with the PIMerror
3470PIM communication failureToo many consecutive NACKS for LTC1760, LTC4260, or AN1112error
3480SPM serial number mismatch between CPU and SPMSelf check fault on SPM serial number verificationcritical
3480Calibration Not PerformedDevice software detects lack of calibration with installed SPMwarning
EXT BATTExternal battery indicatorPossible faulty wiring of the external power connectorwarning
SC 1001High priority alarm indicating occlusion conditionGas Output Port occluded or equivalent condition detectedwarning
SC 1030High priority alarm indicating occlusion or system faultGas output blockage or calibration resistor occlusion detectedwarning
SC1175Internal COMM I2C fault - cannot communicate with PGA or RTCI2C bus communication error detected during operationwarning
SC1471Internal COMM CPU cannot communicate with UIM adapter boardCPU-UIM communication loss detected at startuperror
SC1472CPU-SPM communication failure due to CRC errorsCPU detects excessive CRC errors in SPM communicationerror
SC1473Internal COMM CPU-SPM communication failureCPU monitors SPM and detects no responseerror
SC1474CPU-SPM communication failure due to timeoutCPU SPM communication timeout during operationerror
SC1475Contrast fault - LCD not visibleLCD visibility check fails during startup self-checkerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.