ZOLL 731 Z Vent: troubleshooting and error codes
The ServiceForge knowledge base holds 54 documented symptoms for the ZOLL 731 Z Vent, with 89 ranked causes, 74 diagnostic checks with test points and expected values, and 176 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Breath performance parameters out of tolerance patient safety
Tidal volume, PEEP, PIP, BPM, FIO2 or I-Time fail to meet specified tolerances during calibration tests
Ranked causes
- 1. Test setup error or apparatus contamination (common)
- 2. Test lung defective or O2 sensor contaminated (common)
- 3. O2 sensor calibration drift or sensor failure (occasional)
- 4. Anti-asphyxia valve damaged or contaminated (occasional)
- 5. Ventilator requires recalibration (occasional)
Measure tidal volume at VT=700, FIO2=21, BPM=20, Itime=1000, PEEP=0
- Test point
- Flow sensor
- Expected
- 700 ml ± (5ml + 8% of setting), ± (5ml + 8%)
5 more checks with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor system check failure
System reports failure when comparing measured flow values to published specifications or when checking intake pressure, RPM, or airway pressure.
Ranked causes
- 1. Loose intake assembly (common)
- 2. Dirty or clogged disk filter (common)
- 3. Anti-Asphyxia valve failure or leak during calibration (occasional)
- 4. System leak or high flow measurement (occasional)
- 5. Dirty filters (common)
Check airflow tolerance
- Test point
- measured flow
- Expected
- within ± (0.1 LPM + 5% of setting)
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Complete power failure patient safety
Loss of power from both internal battery and external source during operation
Ranked causes
- 1. External power not applied (common)
- 2. Battery disconnected or not charging (common)
- 3. PIM board connection issues (occasional)
- 4. PIM board needs replacement (occasional)
Apply external power to device
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device does not communicate with Remote Calibration System
USB connection fails, device serial numbers cannot be read, test load cannot be installed
Ranked causes
- 1. USB connection problem or hardware issue (common)
- 2. FTDI component failure on CPU board (occasional)
- 3. Multiple RCS applications running (occasional)
- 4. SPM board communication failure (occasional)
Verify USB cable for wear and degradation
- Test point
- USB connector panel
- Expected
- No visible damage or wear
- Tools
- Visual inspection
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator system test failure or misalignment with specifications patient safety
System Tests sequence fails to confirm ventilator operating within expected tolerances or calibration tables do not test properly; limited to three retry attempts before fault recorded
Ranked causes
- 1. Incorrect test setup configuration (common)
- 2. Compressor not exercised or long-term storage effects (occasional)
- 3. USB connection between ventilator and computer insecure or disconnected (occasional)
- 4. Incorrect power-on timing relative to RCS prompts (common)
Verify test setup configuration
- Test point
- Ventilator connections and RCS interface
- Expected
- Setup matches detailed diagrams and instructions in maintenance chapter
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor flow calibration failure
Remote Calibration System reports failed calibration for compressor flow during discrete tests or system check.
Ranked causes
- 1. Compressor flow gain too low (common)
- 2. Compressor flow gain too high (occasional)
- 3. Dirty disk filter or restricted intake (common)
Review compressor flow offset range
- Test point
- CF_Offset
- Expected
- 26823 - 35474
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor system calibration failure
Remote Calibration System reports failed calibration for intake pressure or compressor flow during discrete tests.
Ranked causes
- 1. Intake pressure gain too low (common)
- 2. Intake pressure gain too high (common)
- 3. Damaged fresh gas intake filters (occasional)
Review intake pressure calibration range
- Test point
- IP_Offset
- Expected
- 22588 - 41168
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 Flow Calibration failure
Oxygen flow calibration fails with out-of-range offset or gain values
Ranked causes
- 1. Gain calibration error - Gain Too Low (common)
- 2. Gain calibration error - Gain Too High (occasional)
- 3. SPM/Vent assembly defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
5V bus compromised
Self-check failure of the 5V power bus
Ranked causes
- 1. PIM to SPM board connection damaged (common)
- 2. PIM board defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication error SC1175 occurs during service operation
Alarm SC1175 (Internal COMM) appears intermittently during RCS service procedures; expected during service but not operational use
Ranked causes
- 1. Software assert in RCS or intermittent hardware condition (common)
- 2. SPM board hardware problem (rare)
Power cycle device
- Test point
- Device power
- Expected
- Alarm clears after reset
- Tools
- Power switch
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor bypass protection alarm not triggered patient safety
Fresh gas intake blockage does not trigger intake blocked alarm or alarms incorrectly indicate partial blockage
Ranked causes
- 1. Loose or incomplete intake assembly blockage (common)
- 2. Partial blockage or incorrect test procedure (occasional)
Verify intake blocked alarm 1030 triggers during fresh gas intake occlusion
- Test point
- Intake assembly
- Expected
- Alarm 1030 triggered
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External battery fault indicator
Possible faulty wiring of the external power connector
Ranked causes
- 1. External power connector wiring fault (common)
- 2. External power cable defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal communication battery fault (no external power) patient safety
Communication failure to device main battery without external power
Ranked causes
- 1. Battery connection to PIM board loose or faulty (common)
- 2. Battery or PIM defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal communication battery fault (with external power) patient safety
Communication failure to device main battery while operating on external power
Ranked causes
- 1. Battery connection to PIM board loose or faulty (common)
- 2. Battery or PIM defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal communication power components fault
Communication error with the PIM (power interface module)
Ranked causes
- 1. PIM to SPM or CPU connection fault (common)
- 2. PIM or CPU connector cable defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD is dark or GUI is blank during service
Screen shows vertical line only or remains OFF during RCS programming, expected behavior during reset state
Ranked causes
- 1. Device in reset or intermediate state (common)
- 2. LCD or display hardware failure (occasional)
Observe LCD during startup
- Test point
- LCD screen
- Expected
- Normal display after power cycle or vertical line during reset
- Tools
- Visual inspection
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
RTC battery low or date/PM due alarm resets caution
Alarm 3110 indicates low RTC battery; date and PM due settings reset frequently after battery replacement
Ranked causes
- 1. RTC battery voltage low (common)
- 2. RTC battery retention clip bent during installation (occasional)
Verify RTC battery is properly seated
- Test point
- RTC battery slot
- Expected
- Battery seated firmly with retention clip secured
- Tools
- Visual inspection
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 readings not displayed or sensor LEDs not illuminated patient safety
Pulse oximeter sensor fails to illuminate and no SpO2 readings appear on ventilator display
Ranked causes
- 1. Defective SpO2 sensor or cable (common)
- 2. Masimo board failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System test failure during service
Device test fails during RCS System Tests; requires troubleshooting before calibration can proceed.
Ranked causes
- 1. Setup error or equipment configuration issue (common)
- 2. Identified failure in ventilator or calibration (occasional)
Physical Inspection
- Test point
- Device exterior
- Expected
- No damage or defects; room temperature 16–28 °C/60–82 °F
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
TSI calibration out of specification
TSI barometric pressure, temperature, or calibration date exceeds valid operating range during test station operation
Ranked causes
- 1. Environmental conditions outside specification (temperature or pressure) (common)
- 2. TSI calibration date expired (common)
Verify TSI barometric pressure reading
- Test point
- TSI sensor
- Expected
- 69kPa < reading < 102kPa, 69–102
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Button Check failure
One or more buttons on device not functioning correctly
Ranked causes
- 1. Button hardware defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Buzzer Check failure patient safety
Device buzzer not producing audible alarm sound
Ranked causes
- 1. Buzzer hardware defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: 731 Series Ventilator patient safety
Service calibration procedure using RCS 9.1 software; requires Calibrations sequence to pass, followed by System Tests and forensic memory erasure
Ranked causes
- 1. Calibration setup via RCS 9.1 (common)
Run Calibrations sequence from Service Actions list
- Test point
- RCS 9.1 software
- Expected
- PASS status with no failures
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Erase Forensic Memory
Clear ventilator forensic memory after system tests and calibrations
Ranked causes
- 1. Forensic memory requires erasing as part of service procedure (common)
Run Erase Forensic Memory from Service Actions list
- Test point
- RCS 9.1 interface
- Expected
- Erase Forensic Memory completes successfully
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: O2 Kickstart
Near zero flow target calibration for fine flow control after opening valve
Ranked causes
- 1. Oxygen Delivery System restriction or gas flow issue (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Remote Calibration System (RCS) preventive maintenance patient safety
The RCS performs testing and calibration of the ventilator through discrete tests, either individually for troubleshooting or as part of larger sequences (Calibrations, System Tests). TSI Flow Module requires yearly calibration before ventilator calibration.
Ranked causes
- 1. Preventive maintenance due or PM/Calibration Due alarm (#3120) triggered (common)
Verify TSI Flow Module calibration date
- Test point
- TSI Flow Module
- Expected
- Last calibration date within one year
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: System Tests patient safety
Run System Tests sequence from Service Actions after successful Calibrations to verify ventilator operation
Ranked causes
- 1. System Tests must follow successful Calibrations (common)
Run System Tests sequence
- Test point
- Service Actions menu
- Expected
- System Tests completes without failures
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Write serial numbers to ventilator
Update serial numbers for new parts or recover corrupted component data using RCS software
Ranked causes
- 1. New serialized parts installed or data corruption recovery required (unknown)
RCS software launched and ventilator communication established
- Test point
- RCS user interface Read Vent button
- Expected
- Communication established between RCS application and ventilator
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Comm RAM and Bootloader Check failure
EMV COMM interface, CPU RAM, bootloader, or SPM communication not functioning
Ranked causes
- 1. Front panel assembly defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
EMV Self Check failure
RCS detects active alarms on EMV indicating power system, communications, or pulse oximeter problems
Ranked causes
- 1. Power system, communications, or pulse oximeter fault (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Encoder Check failure
Rotary encoder not functioning in one or both directions
Ranked causes
- 1. Rotary encoder hardware defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation valve timing out of specification
Exhalation valve rise or fall delay exceeds maximum allowed time at reference pressures
Ranked causes
- 1. SPM assembly or board fault (common)
Measure autocal rise settling time at 4 cmH2O
- Test point
- Autocal valve output
- Expected
- < 30ms, < 30ms
6 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Input protection circuit failure
Damaged fuse on the PIM board
Ranked causes
- 1. PIM board fuse damaged (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD Contrast Set failure
LCD contrast cannot be properly adjusted; contrast control not functioning
Ranked causes
- 1. Ribbon cable from LCD to button board not seated correctly (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD Visual Inspection failure
LCD display defects detected; stuck pixels, blemishes, backlight or contrast problems
Ranked causes
- 1. LCD defective with stuck pixels, blemishes, backlight or contrast issues (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LED Check failure
One or more device LEDs not functioning; green, yellow, red, or IR LED defective
Ranked causes
- 1. LED hardware defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manifold pressure leak detected caution
Airway pressure slope during simulated expiratory phase exceeds specification indicating system leak
Ranked causes
- 1. Leak in manifold connections or tubing (common)
Measure airway pressure slope at 5 cmH2O
- Test point
- Airway pressure sensor
- Expected
- < 0.5 cmH2O per second cmH2O/s, < 0.5 cmH2O per second
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Masimo communication error alarm
Masimo communication thread out or cable/board connection damaged; codes 2301 or 3301 displayed
Ranked causes
- 1. Masimo cable or board connection issue (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Masimo module failure alarm during self-check patient safety
Internal OEM Masimo board indicates failure during system self-check; codes 2300 or 3300 displayed
Ranked causes
- 1. CPU/SPO2 Stack or Front Panel Assembly failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 filter resistance out of specification
Sintered bronze filter pressure drop measured outside acceptable range at 100 LPM
Ranked causes
- 1. Sintered bronze filter contaminated or clogged (common)
Measure filter resistance at 100 LPM flow
- Test point
- O2 inlet filter
- Expected
- 0.01–0.06 psi/LPM, 0.01–0.06 psi/LPM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 regulator damaged or not functioning properly caution
O2 regulators may be damaged from improper use, affecting oxygen supply to test station
Ranked causes
- 1. Oxygen regulator not closed or over-tightened (common)
Verify O2 regulator is not over-tightened
- Test point
- O2 regulator handle
- Expected
- Hand-tight only, not over-tightened
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 supply hose damaged or O-rings compromised caution
O2 hoses show signs of damage, debris, or missing/damaged O-rings affecting oxygen delivery
Ranked causes
- 1. O-rings missing or damaged in O2 hose fittings (occasional)
Inspect O2 hose O-rings and connections
- Test point
- O2 data cable and O2 cable connections
- Expected
- O-rings in place, no visible damage or debris
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 supply pressure out of specification caution
O2 input pressure measured outside acceptable limits at reference pressures
Ranked causes
- 1. O2 supply pressure incorrect (common)
Measure O2 input pressure at 90 psi reference
- Test point
- O2 inlet
- Expected
- +/- 1 psi, +/- 1 psi
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 system leak rate excessive caution
O2 leak rate exceeds maximum allowed specification at reference pressure
Ranked causes
- 1. Leaks in O2 tubing or fittings (common)
Measure O2 leak rate at 55 psi
- Test point
- O2 system output
- Expected
- <= 2000 ml/hour, <= 2000 ml/hour
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 valve leakage excessive caution
Reported O2 leak through valve exceeds maximum allowed specification
Ranked causes
- 1. O2 valve damaged or tubing kinked (common)
Measure reported O2 leak through valve
- Test point
- O2 valve outlet
- Expected
- <= 0.3 LPM, <= 0.3 LPM
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power Switch Check failure
Power switch not functioning correctly; unit does not shut off or turn on as expected
Ranked causes
- 1. Power switch functional issue (common)
Unit fully shuts off with no external power
- Test point
- Power switch Off position with no external power
- Expected
- Off and No External Power = Pass
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reported O2 supply pressure out of limits caution
O2 supply pressure reported by system outside acceptable limits at zero and 70 psi reference
Ranked causes
- 1. O2 supply tubing kinked or damaged (common)
Measure reported O2 supply pressure at zero reference
- Test point
- O2 system sensor
- Expected
- < 2 psi
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RTC date and time check failure
RCS verification of date and time in device versus PC failed; RTC battery or connector may be defective
Ranked causes
- 1. RTC battery depleted or faulty battery connector (common)
RCS verifies date and time in device matches PC
- Test point
- SPM real-time clock module
- Expected
- Date and time match PC value
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
RTC date and time set failure
RCS unable to set date and time in ventilator; may indicate RTC battery depletion
Ranked causes
- 1. RTC battery depleted (common)
RCS attempts to set date and time
- Test point
- SPM real-time clock module
- Expected
- Date and time written to device
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self-Check Failure: O2 Valve Stuck Open
System detects oxygen valve stuck in open position
Ranked causes
- 1. O2 Valve Stuck Open (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SPM blink LED and buzzer failure
SPM unable to drive red LEDs and buzzer; operator confirmation not received within 15 seconds during test
Ranked causes
- 1. SPM board faulty or cable connections failed (common)
Verify red LEDs and buzzer activation
- Test point
- SPM output
- Expected
- Red LEDs and buzzer driven within 15 seconds
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SPM communication failure
RCS test confirms communication signals (RX & TX), bootload and toggle are not functional
Ranked causes
- 1. SPM board faulty or connections to SPM failed (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Total Power Failure Alarm Check failure patient safety
Power failure circuit not functioning; total power loss alarm does not trigger or display correctly
Ranked causes
- 1. Power failure circuit defective; alarm does not trigger on total power loss (occasional)
Verify Total Power Failure alarm activation
- Test point
- Power failure circuit alarm output
- Expected
- 120 second Total Power Failure alarm triggers (fast beep and red alarm LED blink pattern) and lasts for at least 10 seconds.
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
User prompt not visible on screen
Expected UI prompt does not appear to the operator during normal operation
Ranked causes
- 1. Prompt hidden behind application window (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1001 | Self-Check Failure or Self Check Fault | Compressor control fault detected | error |
1001 | Compressor failure | Compressor fails to operate or fails to provide required flow for breath delivery when high pressure O2 unavailable | critical |
1001 | Compressor does not respond to control signals | At power up or during operation; blocked output tubing, kinked tubing, loose connector, diffuser or flow screen issue | critical |
1002 | Compressor flow signal high, low, or stuck | Compressor internal communication failure; faulty compressor flow transducer or tubing issue | critical |
1002 | Compressor communication fault | Communication lost between compressor controller and Smart Pneumatic Module when high pressure O2 unavailable | critical |
1002 | Self Check Failure | PGA or ADC faults for Compressor System detected | error |
1003 | Low compressor speed at startup | Flow from first breath is ±20% of expected flow for tidal volume at start up | error |
1003 | Fails power-on self check breath | Sensor/transducers/calibration file failure at startup | critical |
1010 | O2 valve fails in open position | O2 valve failure causing continuous flow during expiratory phase, preventing ventilation | critical |
1011 | Loss of O2 flow control | O2 flow path failure or fault preventing O2 flow control | critical |
1011 | O2 valve flow delivery failure | Signal to O2 valve not delivering required flow rate and compressor unavailable | critical |
1011 | Control fault indicating insufficient O2 flow rate | Self-check detects inadequate oxygen flow | error |
1012 | PGA or ADC faults detected for O2 System | Self-check detects signal processing electronics failure | error |
1012 | O2 valve communication failure | Communication lost between O2 valve and SPM when compressor unavailable for ventilation | critical |
1012 | O2 flow sensor signal to SPM fails | Internal communication failure between O2 flow sensor and SPM | critical |
1020 | Low oxygen supply pressure (< 35 psig) | O2 supply pressure falls below minimum threshold | error |
1030 | Gas Intake Failure | Insufficient intake flow detected | error |
1030 | Fresh gas or emergency air inlet blocked | Fresh Gas/Emergency Air Inlet blocked preventing compressor from delivering sufficient flow | critical |
1030 | Insufficient flow or blocked fresh gas intake | Gas intake failure or fault preventing fresh gas intake | critical |
1041 | High oxygen supply pressure (> 87 psig) | O2 supply pressure exceeds maximum threshold | warning |
1051 | Autocal valve malfunction | After 10 breaths, valve does not open, opens part way, or readings too noisy | error |
1051 | Run-time calibration failure | Self check failure during runtime or startup | critical |
1052 | Airway pressure signal stuck high, low, or unchanging | Self check failure during power-on or runtime | critical |
1052 | Airway pressure sensor communication loss | Communication lost between airway pressure sensor and SPM | error |
1052 | Self-check failure: airway pressure transducer | Transducer not behaving as expected during self-check | error |
1060 | Exhalation system valve failure | PIP fails to return to baseline pressure for 3 consecutive breaths | critical |
1060 | PEEP higher than limit for 3 consecutive breaths | Exhalation system failure with excessive PEEP | critical |
1061 | Airway pressure excessive for extended time | Exhalation system failure with airway pressure >70 cmH2O for 1.5s or >35 cmH2O/PIP high limit for 5s | critical |
1172 | 5V self check failure | Self check failure during power-up | critical |
1172 | 5V bus compromised | Self-check failure of the 5V bus | error |
1173 | Communication failure | PIM to CPU communication fault at power-up | critical |
1174 | Unable to zero/null transducer offset | Self check failure unable to null one or more transducer offsets | critical |
1174 | Self Check Failure | Flow transducers behaving outside specifications | error |
1174 | PGA offset control calibration failure | Device unable to calibrate transducers at startup; PGA offset control fails | critical |
1174 | Flow transducers are not behaving as expected | Self-check detects flow sensor anomaly | error |
1175 | Internal communication bus failure | Internal communication bus and host unable to communicate with sub-assemblies | critical |
1175 | Loss of I2C communication for 4 consecutive attempts | Self check failure losing I2C bus communication | critical |
1176 | Calibration table or header CRC error | Self check failure with corrupted calibration table | critical |
1176 | Corrupted calibration file detected | Calibration file checksum failure or file corruption during write | warning |
1420 | Complete power failure | Power lost from both internal battery and external source during operation | critical |
1430 | Drained battery | Internal battery power drops below amount required for ventilation without external power | error |
1471 | User interface module communication failure | Device unable to communicate with User Interface Module and interface controls | error |
1471 | Cannot communicate with IO expander for buttons and LEDs | Self check failure on IO expander communication | error |
1472 | High rate of corrupted packets from SPM | Self check detects excessive SPM packet corruption | error |
1472 | Smart pneumatic module communication failure | Device unable to communicate with Smart Pneumatic Module | error |
1473 | SPM data timeout | No valid data sent from SPM within 1 second | error |
1474 | Cyclic redundancy check failure | CRC checking fails between device and SPM | error |
1474 | CPU and PIM communication gaps detected | Self check detects communication gaps between CPU and PIM | error |
1475 | Cannot communicate with LCD contrast control circuit | Self check failure on LCD contrast control communication | error |
1475 | Contrast control communication failure | Device lost communication with contrast control | error |
1480 | CPU and SPM software revisions incompatible | Self check detects version mismatch between CPU and SPM | error |
2001 | Compressor does not respond to control signals (medium priority) | Medium priority version of code 1001 | error |
2001 | Self Check Fault | Compressor control fault detected | error |
2002 | Self Check Fault | PGA or ADC faults for Compressor System detected | error |
2002 | Compressor flow signal high, low, or stuck (medium priority) | Medium priority version of code 1002 | error |
2010 | O2 Valve Stuck Open | Alarm code variant of 1010 | error |
2011 | Self Check Fault | Communication between O2 valve and SPM fails and compressor is available to provide ventilation | warning |
2011 | Control fault indicating insufficient O2 flow rate | Alarm code variant of 1011 | error |
2011 | Loss of O2 flow control (medium priority) | Medium priority version of 1011 | error |
2012 | O2 flow sensor signal to SPM fails (medium priority) | Medium priority version of 1012 | error |
2012 | Self Check Fault | Signal to O2 valve outside calibration range for required flow rate and compressor available | warning |
2020 | O2 supply pressure less than 35 psi | Low O2 supply fault condition | error |
2030 | Insufficient flow or blocked fresh gas intake (medium priority) | Medium priority version of 1030 | error |
2030 | Gas Intake Fault | Fresh Gas/Emergency Air Inlet blocked; compressor unable to deliver breath within ±10% of settings; high pressure O2 available | error |
2041 | High oxygen supply pressure (> 87 psig) | Alarm code variant of 1041 | warning |
2053 | Self Check Fault | Expiratory time < 170 ms for 3 consecutive breaths | warning |
2062 | Exhalation System Fault | Airway pressure PIP measured at end of expiration > 5 cm H2O above baseline pressure PEEP | error |
2062 | PEEP measured exceeds setting plus 5 cmH2O | Exhalation system fault with elevated PEEP | error |
2070 | Airway Pressure High | Airway pressure PIP greater than high airway pressure limit for 2 consecutive breaths | warning |
2071 | Low Airway Pressure | Airway pressure PIP less than low airway pressure limit for 2 consecutive breaths | warning |
2072 | High Tidal Volume | During pressure targeted ventilation, delivered tidal volume exceeds user defined limit for 2 consecutive breaths | warning |
2073 | Low Tidal Volume | During pressure targeted ventilation, delivered tidal volume does not reach user-defined limit for 2 consecutive breaths | warning |
2074 | High Breath Rate | Actual breathing rate (set rate plus spontaneous patient rate) exceeds high alarm limit | info |
2075 | Low Breath Rate/Apnea | Actual breathing rate (set rate plus spontaneous patient rate) less than low alarm limit | warning |
2076 | Apnea | Spontaneous breathing rate less than low alarm limit in noninvasive ventilation, CPAP and BL modes | warning |
2090 | PEEP Leak | Airway pressure drops below PEEP setting by 2 cm H2O during expiratory phase | warning |
2095 | Insufficient Flow | Pressure target not reached during inspiratory period during pressure targeted ventilation | warning |
2100 | Patient Disconnect | Airway pressure fails to exceed PEEP setting by approximately 7 cm H2O | error |
2170 | Spont. Breath PIP High | Airway pressure PIP exceeds High PIP Limit Setting during 2 consecutive spontaneous breaths | warning |
2171 | Spont. Breath PIP Low | Airway pressure PIP exceeds Low PIP Limit Setting during 2 consecutive spontaneous breaths | warning |
2172 | Spont. Breath Vt High | High VT Limit exceeded during 2 consecutive spontaneous breaths | warning |
2173 | Spont. Breath Tt Low | Low VT Limit Setting not achieved during 2 consecutive spontaneous breaths | warning |
2300 | Self-Check: Pulse Ox - Module Failure (Internal OEM Board) | Masimo board indicates failure during power-up self-check | critical |
2300 | Self Check Fault | Pulse oximeter module fails while in use | error |
2300 | Masimo board indicates a failure | Self check detects board failure | error |
2301 | Communication from Masimo board has timed out | Self check timeout on board communication | error |
2301 | Self Check Fault | Communication between pulse oximeter module and device fails | error |
2301 | Internal COMM: Masimo communication thread out or damaged | Masimo communication link failure or thread timeout | error |
2314 | Pulse Ox Sensor Off Patient | Operating sensor loses patient signal due to disconnection, misalignment or poor perfusion | warning |
2401 | SpO2 Low | SpO2 value drops below Low SpO2 Limit (default 94%) | warning |
2410 | Heart Rate High | Heart rate greater than High Heart Rate Limit (default 120 beats/min) | warning |
2411 | Heart Rate Low | Heart rate less than Low Heart Rate Limit (default 40 beats/min) | warning |
2421 | Self Check Fault | Failure of input protection circuit and device is able to operate | error |
2421 | Input protection circuit failure | Damaged fuse on the PIM board | error |
2423 | Self Check Fault | Internal power circuit failed and external power connected but cannot be used | error |
2423 | Power system failure | Self check detects power fault | error |
2430 | Nearly Drained Battery | Device detects ≤5 minutes of battery operation remaining and external power not connected | warning |
2450 | Battery Discharge Fault | Battery temperature reaches 70°C (158°F), 5°C from maximum operating temperature | error |
2450 | Battery pack internal temperature too high | Battery internal temperature ≥ 70 °C | warning |
2455 | Battery communication fault while powered by battery | System powered by battery AND battery communication fails | error |
2455 | Internal COMM battery fault | Communication failure to device main battery without external power | error |
3001 | Self Check Fault | Compressor fails to operate or provide required flow within ±10% of settings; high pressure O2 available; FIO2 set to 100% | warning |
3001 | Compressor does not respond to control signals (low priority) | Low priority version of code 1001 | warning |
3002 | Self Check Fault | Communication between compressor controller and SPM lost; high pressure O2 available; FIO2 set to 100% | warning |
3003 | Compressor flow signal high, low, or stuck (low priority) | Low priority version of code 1002 | warning |
3011 | Loss of O2 flow control (low priority) | Low priority version of 1011 | warning |
3011 | Self Check Fault | Signal to O2 valve outside calibration range; compressor available; user set FIO2 to 21% | warning |
3012 | Self Check Fault | Communication between O2 valve and SPM lost; compressor available; FIO2 set to 21% | warning |
3012 | O2 flow sensor signal to SPM fails (low priority) | Low priority version of 1012 | warning |
3030 | Insufficient flow or blocked fresh gas intake (low priority) | Low priority version of 1030 | warning |
3030 | Gas Intake Fault | Fresh Gas/Emergency Air Inlet blocked; compressor unable to deliver breaths within ±10%; high pressure O2 available; FIO2 set to 100% | error |
3031 | Intake Restricted | Fresh Gas/Emergency Air Inlet blocked but still capable of delivering breaths within ±10% of settings | warning |
3031 | Insufficient flow from fresh gas intake (intake restricted) | Fresh gas intake restricted or partially occluded | warning |
3032 | Fresh gas intake pressure sensor signal stuck high or low | Sensor and transducer fault on fresh gas/intake pressure | warning |
3032 | Self Check Fault | Inlet pressure transducer malfunction detected | error |
3041 | High O2 Supply Fault | High pressure O2 supply ≥75 psig (517 kPa) and < 80 psig (552 kPa) | warning |
3041 | O2 input pressure too high (low priority) | Low priority version of 1041 | warning |
3073 | Tubing Compliance Fault | Tubing compliance correction greater than set tidal volume | warning |
3091 | AutoPEEP | Exhaled flow continues throughout expiratory period | warning |
3092 | Inspiratory Demand | End-inspiratory pressure < -1.0 cm H2O for 3 consecutive breaths | warning |
3110 | RTC battery drained at startup | Real-time clock alarm at power-up with date/time reset to default | warning |
3120 | PM Due | Preselected number of days elapsed since last calibration at startup | info |
3120 | Calibration due or overdue | Calibration interval exceeded or RTC battery issue | warning |
3120 | PM Due - Preventive Maintenance required | Device reaches one year anniversary since last service | info |
#3120 | PM/Calibration Due alarm | Preventive maintenance interval expired or calibration overdue | warning |
3121 | Power Cycle Need | Device running continuously for 30 days | warning |
3130 | Ambient pressure sensor reads near zero or full scale | Self check fault on barometric pressure sensor | warning |
3131 | Excessive altitude detected | Ambient pressure <0.36 ATM indicating altitude greater than 25,000 feet | warning |
3132 | Low altitude detected (below -2000 feet) | Ambient pressure >1.1 ATM indicating altitude below -2000 feet | warning |
3132 | Low Altitude | Ambient pressure transducer detects altitude < -2,000 feet below sea level (610 meters, 15.8 psig or 103 kPa) | warning |
3140 | Ambient Temperature Fault | Ambient temperature exceeds normal operating range > 131 °F (55 °C) | warning |
3141 | Ambient Temperature Fault | Ambient temperature falls below normal operating range < 14 °F (-10 °C) | warning |
3143 | Temperature sensor fault | Self check fault with temperature sensor reading beyond threshold | warning |
3172 | Self-check fault: airway pressure transducer cannot zero | Transducer unable to zero during initialization | error |
3300 | Alarm priority for Pulse Ox - Module Failure | Same condition as 2300 with alarm priority indication | critical |
3300 | Masimo board indicates a failure | Self check detects board fault | critical |
3300 | Pulse Oximeter Module Self Check Fault | Pulse oximeter module failure with user acknowledging condition | error |
3301 | Communication from Masimo board has timed out | Self check timeout on board communication | critical |
3301 | Pulse Oximeter Communication Self Check Fault | Communication between pulse oximeter module and device fails | error |
3301 | Alarm priority for Internal COMM error (dependent on sensor usage) | Same as 2301 with alarm priority based on Masimo sensor active status | error |
3310 | Pulse Ox Sensor Not Connected | Pulse oximeter detects no SpO2 sensor after successful operation | warning |
3311 | Defective pulse oximeter sensor | Pulse oximeter cannot identify or sensor has failed | error |
3312 | Pulse Search | Pulse oximeter is searching for pulse signal | info |
3313 | Pulse Ox Signal Interference | Outside signal or energy source prevents accurate reading | warning |
3315 | Ambient Light Fault | Excessive ambient light on SpO2 sensor or inadequate tissue covering | warning |
3316 | Invalid pulse oximeter sensor | Pulse oximeter does not recognize connected sensor | warning |
3317 | Low SpO2 Perfusion | Amplitude of arterial pulsation is weak | warning |
3318 | Low SpO2 Perfusion | Quality of input signal is low due to excessive motion or artifact | warning |
3421 | External Power Low/Disconnect | External power drops below ~11 VDC or is disconnected | error |
3422 | Battery pack not detected | Device unable to detect battery | error |
3423 | Battery Charging Fault | Battery charging circuit fails | error |
3430 | Low Battery | Device detects < 30 minutes battery operation remaining with no external power | warning |
3431 | Low Battery | External power connected to device with internal battery drained to low status | warning |
3441 | External power input voltage too high | Input voltage exceeds 33 VDC after fuse | error |
3442 | External power input voltage too low | Input voltage below 11.5 VDC after fuse | error |
3444 | External power voltage negative | Voltage in at less than -1 VDC | error |
3450 | Battery Discharge Fault | Battery temperature reaches 70 °C (158 °F) with external power connected | warning |
3451 | Battery discharge fault at high temperature | Battery internal temperature ≥ 75 °C | critical |
3452 | Battery charging fault due to high temperature | External power present AND battery temperature ≥ 45 °C | warning |
3453 | Battery charging fault due to low temperature | External power present AND battery temperature ≤ 0 °C | warning |
3455 | Battery communication fault while powered by battery | System powered by battery AND battery communication fails | critical |
3455 | Battery Fault | Device unable to communicate with internal battery with external power connected | error |
3455 | Internal COMM battery fault (external power active) | Communication failure to device main battery while operating on external power | error |
3470 | Internal COMM power components fault | Communication error with the PIM | error |
3470 | PIM communication failure | Too many consecutive NACKS for LTC1760, LTC4260, or AN1112 | error |
3480 | SPM serial number mismatch between CPU and SPM | Self check fault on SPM serial number verification | critical |
3480 | Calibration Not Performed | Device software detects lack of calibration with installed SPM | warning |
EXT BATT | External battery indicator | Possible faulty wiring of the external power connector | warning |
SC 1001 | High priority alarm indicating occlusion condition | Gas Output Port occluded or equivalent condition detected | warning |
SC 1030 | High priority alarm indicating occlusion or system fault | Gas output blockage or calibration resistor occlusion detected | warning |
SC1175 | Internal COMM I2C fault - cannot communicate with PGA or RTC | I2C bus communication error detected during operation | warning |
SC1471 | Internal COMM CPU cannot communicate with UIM adapter board | CPU-UIM communication loss detected at startup | error |
SC1472 | CPU-SPM communication failure due to CRC errors | CPU detects excessive CRC errors in SPM communication | error |
SC1473 | Internal COMM CPU-SPM communication failure | CPU monitors SPM and detects no response | error |
SC1474 | CPU-SPM communication failure due to timeout | CPU SPM communication timeout during operation | error |
SC1475 | Contrast fault - LCD not visible | LCD visibility check fails during startup self-check | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI