Ziehm Imaging Vision RFD 3D: troubleshooting and error codes

The ServiceForge knowledge base holds 42 documented symptoms for the Ziehm Imaging Vision RFD 3D, with 58 ranked causes, 139 diagnostic checks with test points and expected values, and 372 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Calibration influence on image quality assessment

Understanding and assessing the impact of calibration accuracy (0.14 superior vs 1.07 insufficient) on artifact presence, edge sharpness, and metal visualization

Ranked causes

  1. 1. Calibration value superior (0.14 mm) achieves optimal image quality (common)
  2. 2. Calibration value good (0.39 mm) produces acceptable image with minimal artifacts (common)
  3. 3. Calibration value insufficient (> 0.39 mm) produces unacceptable artifacts (occasional)
First check

Compare calibration value 0.14 mm (superior) image to reference

Test point
Phantom layer reconstruction quality
Expected
Sharp phantom edges; metal pin without artifacts; clear crisp image

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: 3D image reconstruction accuracy radiation

Verify phantom positioning, ISO center alignment, C-arm collision limits, scan acquisition, accuracy calculation, and scaling factor validation for 3D imaging

Ranked causes

  1. 1. Initial phantom positioning and alignment not verified (common)
  2. 2. Mean accuracy value exceeds tolerance or scaling factor out of range (occasional)
  3. 3. Collision limits not verified (common)
First check

Position phantom parallel to flat-panel detector in LAT position

Test point
Phantom position relative to detector
Expected
Parallel to detector; sides parallel to screen; central sphere inside crosshair
Tools
Laser alignment tool, fluoroscopy

10 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: 3D reconstruction image quality verification radiation

Verify phantom placement, C-arm positioning, image acquisition, and visual quality assessment of phantom grooves, metal pin, and diameter measurements

Ranked causes

  1. 1. Phantom not properly positioned for image quality check (common)
  2. 2. Phantom grooves not clearly visible or not properly formed (occasional)
  3. 3. Metal pin boundaries or diameter not within specification (occasional)
First check

Mount attenuation device and position verification phantom

Test point
25mm Al device mounted on generator; phantom Part No. 62109 in beam path
Expected
Attenuation device secured with tension strap; paper protective layer in place; phantom in beam path
Tools
Tension strap, 25mm Al attenuation device, paper

9 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Video adjustment radiation

Video signal levels must be calibrated using histogram function and grey step wedge. Black level, white level, and metal black level parameters control video signal output.

Ranked causes

  1. 1. Black level requires adjustment (common)
  2. 2. White level requires adjustment (common)
  3. 3. Metal black level requires adjustment (occasional)
First check

Rotate image to mask circle position

Test point
Monitor display
Expected
Circular mask visible on image

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error E1246: Flat-panel detector communication fault

Ethernet cable disconnected or misconfigured between Image Memory and Command Processor during VIVA tool operation or system startup.

Ranked causes

  1. 1. Ethernet cable not connected between Image Memory and Command Processor during VIVA gain calibration setup (common)
  2. 2. System restarted with Ethernet cable disconnected from Image Memory (common)
  3. 3. VIVA tool Ethernet link not opened after physical connection established (occasional)
First check

Verify physical Ethernet cable connection

Test point
Red connector ports on Image Memory and Command Processor 2
Expected
Straight Ethernet cable #24452 with inline coupler #217268 securely seated; cable not damaged or kinked
Tools
Visual inspection; straight Ethernet cable #24452; inline coupler #217268

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Center beam position radiation

Center beam test device image not concentric or centered on screen; may appear off-center to display

Ranked causes

  1. 1. Generator angle misalignment (left-right deviation) (common)
  2. 2. Generator attachment misalignment (top-down deviation) (common)
First check

Visually inspect center beam test device image on screen for left-right centering

Test point
Screen display
Expected
Concentric shape centered horizontally on image

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Dose adjustment radiation

X-ray dose output must be calibrated using 25 mm Al + 1.5 mm copper filtration with dose measured in front of image intensifier grid. Dose Gain and CCD/FD Fill parameters control dose output.

Ranked causes

  1. 1. Dose Gain requires adjustment (common)
  2. 2. CCD/FD Fill requires adjustment (common)
First check

Set CCD fill level to specified value

Test point
Dose 1 tab, CCD/FD-Fill input box
Expected
Value from System Settings CCD fill levels table

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: DSA (Digital Subtraction Angiography) dose adjustment radiation

Two separate dose adjustments available for DSA: HQ DSA and LPK DSA. Measurement uses dose rate with Frames/sec set to 25. Original setting must be restored after adjustment.

Ranked causes

  1. 1. HQ DSA dose requires adjustment (common)
  2. 2. LPK DSA dose requires adjustment (common)
First check

Open Configuration Menu and set Frames/sec to 25

Test point
Configuration Menu parameter
Expected
Frames/sec = 25

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Module 10 Power Distribution Setup high voltage

Configure U539 board voltage selection and connector wiring for transformer RTO 2500-000585 or RTO 2500-001195 based on input AC voltage (105-240 VAC range).

Ranked causes

  1. 1. Voltage selector configuration required for primary AC input (common)
  2. 2. Voltage selector connectors X40-X50 require proper connection to transformer tap points (common)
First check

Verify transformer primary taps are connected per selected input voltage

Test point
Transformer RTO 2500-000585 or 2500-001195 primary terminals A, B, C, D, E, F, H, I, K, L
Expected
Connections match table for selected input voltage (105, 110, 115, 120, 210, 220, 230, or 240 VAC)
Tools
Visual inspection, wiring diagram

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

LED Indicator Verification

Verify all LEDs on U539g board illuminate correctly to confirm power distribution and controller status after configuration.

Ranked causes

  1. 1. LEDs indicate power rail and relay status during startup and normal operation (common)
  2. 2. HC08 processor monitoring status indicated by LEDs D45-D49, D15-D16 (common)
First check

Verify D11, D8, D9 relay status LEDs illuminate in sequence during power-up

Test point
LEDs D11 (K1 mains 1), D8 (K2 inrush), D9 (K3 mains 2)
Expected
D8 illuminates briefly during inrush, then D11 and D9 steady on (green)
Tools
Visual inspection

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Standby LED flashing due to mains 2 contactor not switching caution

Standby LED flashes when CPU did not switch ON mains 2 contactor and input signal from PTH0 is LOW

Ranked causes

  1. 1. Short circuit from transistor T8 on PCB U539 (occasional)
  2. 2. CPU failed to control mains 2 contactor (occasional)
First check

Measure voltage at PTH0 input

Test point
PTH0
Expected
HIGH (when mains 2 contactor should be ON)
Tools
DMM or oscilloscope

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: 26 inch monitor menu lock reset

Accessing locked menu on Radiance G2 HB monitor and restoring factory default settings

Ranked causes

  1. 1. Menu locked on 26 inch Radiance G2 HB monitor (common)
First check

Verify menu access on 26 inch monitor

Test point
OSD display
Expected
Menu appears without MENU LOCKED message; Setup and Defaults sections accessible

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: 3D image quality verification

Visual inspection and measurement of 3D reconstruction quality including sagittal, axial and coronal views for contrast, sharpness and artifact assessment

Ranked causes

  1. 1. 3D calibration requires verification (common)
First check

Measure length of drilling holes in sagittal view including artifacts

Test point
upper hole
Expected
between 11.8 mm and 12.4 mm, 11.8–12.4 mm

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Adipose Patient (LPK) mode dose adjustment radiation

LPK (Adipose Patient) is a modifier to each application program. Dose adjustment must be performed in all programs with same value. LPK has individual kV/mA curve.

Ranked causes

  1. 1. LPK mode dose requires adjustment in all programs (common)
First check

Select LPK modifier and adjust dose in all programs

Test point
Adjustment dialog; all anatomical programs
Expected
Same dose value per System Settings in each program

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Asymmetric collimator DIP switch configuration

DIP switch settings on U659 module for asymmetric collimator operation and dose meter selection

Ranked causes

  1. 1. Incorrect DIP switch settings for operational mode (common)
First check

Verify DIP switch 1 (Standalone mode) position

Test point
S2 Switch 1
Expected
OFF (standalone mode inactive)

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Center Beam Adjustment radiation

Adjust generator beam center position relative to system screen after generator installation. Done before collimator installation using test device and pre-collimator for radiation protection.

Ranked causes

  1. 1. Generator mounted and requires center beam alignment before collimator installation (common)
First check

Gather required tools and devices

Test point
Equipment inventory
Expected
Center beam test device (91116) and pre-collimator (65699) available

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Computed DAP / Air Kerma radiation

Adjustment of dose area product (DAP) and air kerma measurements using reference dose meter and Ziehm test phantom. Requires setting calibration points, pulse width correction, area factor, and kerma/DAP calculation factors.

Ranked causes

  1. 1. Calibration required for accurate dose measurement display (common)
First check

Generator cover with filtration mounted and dose meter set to 3mm Al prefiltering

Test point
Generator cover, dose meter settings
Expected
Generator cover mounted, dose meter configured for 3mm Al prefilter

8 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Drilling hole measurement technique

Procedure to measure drilling hole lengths in sagittal view for calibration validation

Ranked causes

  1. 1. Measurement requires accuracy verification (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Flat-panel detector gain radiation

Adjust detector gain for image homogeneity and proper signal levels across panel modes via manual kV/mA tuning or VIVA tool software, with Gain Median validation.

Ranked causes

  1. 1. Gain calibration required for detector modes (common)
First check

System warm-up time

Test point
System startup to calibration start
Expected
Minimum 1 hour elapsed

21 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: High Quality Mode dose adjustment radiation

HQ (High Quality) mode is a modifier to each application program. Dose adjustment must be performed in all programs with same value. HQ has individual kV/mA curve.

Ranked causes

  1. 1. HQ mode dose requires adjustment in all programs (common)
First check

Select HQ modifier and adjust dose in all programs

Test point
Adjustment dialog; all anatomical programs
Expected
Same dose value per System Settings in each program

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Mains input voltage selection high voltage

Configure power supply board U539h for correct input voltage (105–240 VAC) by hardwiring connector pairs and selecting appropriate fuses (F1, F2) to match site mains supply.

Ranked causes

  1. 1. Voltage mismatch between system configuration and mains supply (common)
First check

Verify mains supply voltage at site installation

Test point
Mains supply outlet
Expected
105, 110, 115, 120, 210, 220, 230, or 240 VAC ±5%

2 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Measured DAP with VacuDAP-C radiation

Adjustment of measured dose area product (DAP) using VacuDAP-C external dose meter. Includes self-test procedure and state error diagnosis using 8-bit error code system.

Ranked causes

  1. 1. VacuDAP-C meter requires calibration factor adjustment (common)
First check

VacuDAP-C firmware version and tube voltage reference

Test point
DAP Meter input box
Expected
Firmware version displayed, currently activated tube voltage shown

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Measuring kV high voltage

Measure kV output from generator to verify proper tube voltage. Measured values are divided by 10,000 (e.g., 6.7 V reading = 67 kV actual).

Ranked causes

  1. 1. kV measurement required during service or commissioning (common)
First check

Set generator to continuous mode for accurate measurement

Test point
Adjustment dialog Dose 2
Expected
100% pulse width or Pulse Length = 0

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Measuring mA high voltage

Measure mA (tube current) output from generator to verify proper tube current. Requires connector X3 on U564 board with jumper removed for measurement.

Ranked causes

  1. 1. mA measurement required during service or commissioning (common)
First check

Remove jumper from X3 connector

Test point
Connector X3 on U564 board
Expected
Jumper removed, connector accessible for meter connection

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Monitor line voltage matching high voltage

Service technician must match facility input voltage to TR1 isolation transformer to ensure stable 230 VAC output for C-arm system

Ranked causes

  1. 1. Input voltage not matched to TR1 transformer tap settings (common)
First check

Measure facility wall outlet voltage

Test point
Wall outlet connector
Expected
Stable within Voltage Selection table range
Tools
Digital volt meter

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Monitor on-screen display menu lock/unlock

Setting or resetting keypad lock on 19 inch dual monitor via simultaneous key press, or locking/unlocking menu on 26 inch Radiance G2 HB monitor

Ranked causes

  1. 1. Keypad lock engaged on 19 inch monitor (common)
First check

Verify keypad lock status on 19 inch monitor

Test point
OSD display
Expected
No lock message when keys pressed; Input and L/R keys responsive

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: RSA (Rotational Subtraction Angiography) dose adjustment radiation

RSA is a separate program requiring individual dose adjustment. Selected via RSA function in adjustment dialog dropdown menu.

Ranked causes

  1. 1. RSA mode dose requires adjustment (common)
First check

Select RSA function in adjustment dialog dropdown

Test point
Adjustment dialog dropdown menu
Expected
RSA function selected

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Service Position for Generator Electronics Access caution

Position generator electronics module for access to test points and measurement connectors during service and calibration procedures.

Ranked causes

  1. 1. Generator electronics access required for measurement and calibration (common)
First check

Remove specified screw from generator module

Test point
Generator electronics enclosure
Expected
Screw removed and set aside safely

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Snapshot mode dose adjustment radiation

Snapshot is a separate program requiring individual dose adjustment. Accumulated dose for one shot must be measured after regulation stabilizes, target <2 µGy.

Ranked causes

  1. 1. Snapshot mode dose requires calibration (common)
First check

Select SSHT key in adjustment dialog

Test point
Adjustment dialog dropdown
Expected
SSHT (Snapshot) mode selected

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ethernet link error during image retrieval

Error message appears when attempting to retrieve images from detector; image data transfer fails.

Ranked causes

  1. 1. Ethernet link communication interrupted or lost (occasional)
First check

Monitor for error message during image retrieval

Test point
Acquisition software display
Expected
No error message; successful image transfer

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inrush relay fault condition high voltage

Inrush relay switched ON but system does not detect proper inrush resistor voltage drop within timing window

Ranked causes

  1. 1. Inrush resistor failure or open circuit (occasional)
First check

Measure analog voltage at PTB7 during inrush cycle

Test point
PTB7
Expected
Greater than 3.0 V, above 3.0 V
Tools
DMM or oscilloscope, scope set to capture inrush transient

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Jumper Configuration for Workstation Options

Set jumpers J1, J2, J3 on U539h/j to enable/disable workstation control options.

Ranked causes

  1. 1. Jumpers select workstation presence and control mode (common)
First check

Verify jumper J1, J2, J3 positions match workstation configuration

Test point
Jumpers J1, J2, J3 on U539h/j board
Expected
J1, J2, J3 open if corresponding workstations not installed; closed if WS No.1 control enabled
Tools
Visual inspection

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Protective and Auxiliary Supply Fuse Configuration caution

Install or verify fuses F2, F3, F4, F5 per AC input voltage configuration (100/105-120/210-240 VAC).

Ranked causes

  1. 1. Supply voltage fuses must match system input voltage rating (common)
First check

Verify F1A and F1B main fuses per selected voltage

Test point
Fuses F1A, F1B external on U539g board
Expected
105-120 VAC: 20 A slow (17044); 210-240 VAC: 15 A slow (17020), exact rating required
Tools
Visual inspection, parts list

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Retrieved images appear incorrect or uncorrected

Detector images retrieved from system show appearance similar to uncorrected raw data with visible artifacts or non-uniform response.

Ranked causes

  1. 1. Image corrections disabled in VIVA tool configuration (common)
First check

Verify image correction status in VIVA configuration

Test point
System Settings dialog
Expected
Image corrections enabled in configuration

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Standby LED and port PTF LEDs flashing caution

Standby LED and all LEDs from port PTF (D45, D46, D47, D48, D49, D15, D16) flash when PCB powered and analog voltage from PTB0 less than 1.0 V

Ranked causes

  1. 1. Remote button ON short circuit or pressed (common)
First check

Measure analog voltage at port PTB0

Test point
PTB0
Expected
Greater than 1.0 V, above threshold
Tools
DMM or oscilloscope

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Standby LED flashing when OFF button pressed

Standby LED flashes when system in OFF state and OFF button (remote or local) is pressed

Ranked causes

  1. 1. Normal operation during OFF state transition (common)
First check

Verify system is in OFF state

Test point
Power state indicator
Expected
System powered down, not in blocking state

VacuDAP-C state error: DAP overflow (Bit 2 = Decimal 4)

DAP meter accumulated dose has exceeded maximum measurable range.

Ranked causes

  1. 1. Accumulated DAP exceeds meter range during long screening session (occasional)
First check

VacuDAP-C state error Bit 2

Test point
State field value
Expected
Bit 2 = 0 (no overflow) decimal

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

VacuDAP-C state error: Error test function (Bit 4 = Decimal 16)

Self-test result does not match reference value, indicating meter malfunction or calibration error.

Ranked causes

  1. 1. Self-test measured value differs from stored reference value (occasional)
First check

Self-test result vs Normal Value Outer Field

Test point
State field, test result vs reference
Expected
Test result within tolerance of Normal Value Outer Field

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

VacuDAP-C state error: Firmware error (Bit 6 = Decimal 64)

Meter firmware is corrupted or malfunctioning.

Ranked causes

  1. 1. VacuDAP-C firmware corruption or hardware defect in microcontroller (rare)
First check

VacuDAP-C state error Bit 6

Test point
State field value
Expected
Bit 6 = 0 (no firmware error) decimal

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

VacuDAP-C state error: TEST-Warning (Bit 1 = Decimal 2)

VacuDAP-C self-test warning indicating possible meter malfunction or calibration drift.

Ranked causes

  1. 1. Meter test value does not match reference specification (occasional)
First check

VacuDAP-C self-test result vs reference value

Test point
State field, test result vs Normal Value Outer Field
Expected
Test result matches or is very close to Normal Value Outer Field reference

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

VacuDAP-C state error: Voltage for measuring chamber faulty (Bit 5 = Decimal 32)

Measuring chamber supply voltage is defective or out of specification.

Ranked causes

  1. 1. Power supply for ionization chamber damaged or disconnected (rare)
First check

VacuDAP-C state error Bit 5

Test point
State field value
Expected
Bit 5 = 0 (no voltage fault) decimal

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

VacuDAP-C state error: Zero check error (Bit 3 = Decimal 8) radiation

Radiation was active during system reset, corrupting zero reference calibration.

Ranked causes

  1. 1. Radiation exposure during dose meter reset or power-up (rare)
First check

Radiation status during system reset

Test point
X-ray enable signal, meter status
Expected
Radiation off during reset, zero calibration completes without exposure

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
2900Kinematic Emergency StopEmergency stop button pressed or safety condition triggeredwarning
2901A Axis FaultDuring kinematic operationwarning
2902C Axis FaultDuring kinematic operationwarning
2903Y Axis FaultDuring kinematic operationwarning
2904Z Axis FaultDuring kinematic operationwarning
2905X-ray FaultDuring operationwarning
2906CAN Fault line activeDuring operationwarning
2907Distance Control FaultDuring kinematic operationwarning
2914Internal FaultDuring operationwarning
2915Internal FaultDuring power-upwarning
2916Internal CAN FaultDuring operationwarning
2917Internal CAN FaultDuring operationwarning
2918Internal FaultDuring operationwarning
2919Internal FaultDuring operationwarning
2920Internal FaultDuring operationwarning
2921Internal FaultDuring operationwarning
2922Y Axis ISO speed FaultDuring kinematic operationwarning
2923Z Axis ISO speed FaultDuring kinematic operationwarning
2924ISO FaultDuring kinematic operationwarning
2925Internal Distance Control FaultDuring operationwarning
2926Internal FaultDuring operationwarning
2927Internal Distance Control FaultDuring operationwarning
2928Internal Distance Control FaultDuring operationwarning
2929Internal Distance Control FaultDuring operationwarning
2930Internal Distance Control FaultDuring operationwarning
2931Internal FaultDuring operationwarning
2932Emergency Stop FaultWhen emergency stop button is pressedwarning
2933Internal Distance Control FaultDuring operationwarning
2934Internal Distance Control FaultDuring operationwarning
2935Internal FaultDuring operationwarning
2936Navigation FaultDuring operationwarning
2937Parking brake engagedDuring operation when brake is engagedwarning
2938Swivel brake engagedDuring operation when brake is engagedwarning
2939Internal FaultDuring operationwarning
2940Internal FaultDuring operationwarning
2941Internal FaultDuring power-upwarning
2955Internal Fault Distance ControlDuring operationwarning
2956Speed FaultDuring operationwarning
2957Collision Internal FaultDuring operationwarning
2958Internal FaultDuring operationwarning
2959Internal FaultDuring operationwarning
2960A Axis FaultDuring kinematic operationwarning
2961Y Axis FaultDuring kinematic operationwarning
2962C Axis FaultDuring kinematic operationwarning
2963Z Axis FaultDuring kinematic operationwarning
2964Kinematic module errorDuring operationwarning
3000Axis not clutchedWhen axis is required but not engagedwarning
3001Axis brake closedWhen axis should move but brake is appliedwarning
3002Axis %d on limit AAxis reaches limit position Awarning
3003Axis %d on limit BAxis reaches limit position Bwarning
3004Axis %d hardware over currentHardware detects excessive current in axiserror
3005Axis %d software over currentSoftware detects excessive current in axiserror
3006Axis %d contouring errorAxis position deviates from commanded pathwarning
3007Axis %d extended errorExtended error condition detected on axiserror
3008Axis %d not enabledAxis control logic not activatedwarning
3010Axis %d solenoid current errorSolenoid current out of specificationerror
3011Axis %d engine temperature errorMotor temperature exceeds limiterror
3012Axis %d hw voltage too lowHardware power supply voltage below minimumerror
3013Axis %d sw voltage too lowSoftware-detected power supply voltage below minimumerror
3014Axis %d watchdog errorWatchdog timeout indicates processor hangcritical
3015Axis %d tacho errorTachometer signal missing or invaliderror
3016Axis %d encoder errorEncoder signal invalid or missingerror
3017Axis %d encoder errorSecond encoder signal faulterror
3018Axis %d encoder range errorEncoder position value out of valid rangeerror
3019Axis %d gear errorGear drive fault detectederror
3020Axis %d port extender errorPort expander communication failureerror
3021Axis %d emergency stopEmergency stop button activatedwarning
3022Axis %d wigwag errorOscillatory motion detectedwarning
3023Axis %d motor errorMotor control fault detectederror
3024Axis %d motor errorMotor error with safety processor in resetcritical
3025Axis %d motor errorMotor error with lift key activeerror
3026Axis %d in autarkic modeAxis operating independently of main controlwarning
3027Axis %d unknown errorUnrecognized error conditionerror
E1000Filament not operating (small focus)Small focus filament fails to initialize during bootuperror
E1000Thin filament defective during power-upSystem power-up detection of thin filament circuit failureerror
E1001Thick filament defective during power-upSystem power-up detection of thick filament circuit failureerror
E1001Filament not operating (large focus)Large focus filament fails to initialize during bootuperror
E1002V-sync signal missingV-sync signal not detected during bootuperror
E1002No V-SyncVideo synchronization signal missingerror
E1002No VSync at generator during power-upMissing video sync signal from generator during startuperror
E1003Generator temperature minimum not reached - please waitGenerator temperature below 13°C during startuperror
E1004Oil temperature above maximum during bootupOil temperature maximum check failserror
E1004Generator must cool down below 72°C - please waitGenerator temperature exceeds 72°C during operationwarning
E1005Generator temperature minimum not reached - please waitGenerator temperature below 13°C during startup (Sensor 2)error
E1006Generator must cool down below 72°C - please waitGenerator temperature exceeds 72°C during operation (Sensor 2)warning
E1006Oil temperature above maximum during runtimeOil temperature maximum check fails during operationerror
E1007Wrong EEPROM checksum during power-upEEPROM memory checksum validation fails at startuperror
E1007Machine data checksum errorChecksum validation fails during bootuperror
E1008Generator fault - Interrupt 1 flip-flop defectiveInterrupt control flip-flop 1 circuit failure detectederror
E1008Hardware flip-flop for external interrupt errorFlip-flop not operating correctly during bootuperror
E1009Hardware flip-flop for external interrupt errorFlip-flop not operating correctly during bootuperror
E1009Generator fault - Interrupt 2 flip-flop defectiveInterrupt control flip-flop 2 circuit failure detectederror
E1010HW signal X-ray active during power-upX-ray enable signal detected as active during system startuperror
E1011CAN X-ray Enable active during power-upCAN bus X-ray enable command detected during startuperror
E1012Generator fault - D/A converter reference voltage defectiveD/A converter reference voltage out of specificationerror
E1012Reference voltage D/A converter errorReference voltage measurement failserror
E1013Generator fault - D/A converter kV defectiveD/A converter kilovoltage output circuit failureerror
E1013Actual voltage from kV D/A converter errorkV actual voltage measurement fails or out of rangeerror
E1014Oil overtemperature switch active during power-up - call serviceOil temperature sensor alarm activated at startuperror
E1015Machine data validity errorInvalid data bytes or version mismatch in machine dataerror
E1015Wrong machine data version detected during power-upMachine configuration data version mismatch during initializationerror
E1016Moisture sensor alarm during power-up - call serviceHumidity sensor detects excess moisture during startuperror
E1016Humidity sensor active/not active errorHumidity sensor status unexpected during bootupwarning
E1017Test of FAULT signal failedInternal fault signal test during diagnostics failserror
E1018Debug switch is onDebug mode enabled at system startupwarning
E1019Generator fault - Anode rotation monitoring does not workAnode rotation feedback signal missing or abnormal during operationerror
E1050High voltage faultHigh voltage supply out of specification or unstablewarning
E1050HVC fault hardware signalHigh voltage controller detects fault conditioncritical
E1051High voltage fault (anode)Anode high voltage supply malfunctionwarning
E1052High voltage fault (cathode)Cathode high voltage supply malfunctionwarning
E1053mA peak error hardware signalmA peak error detected in filament regulationerror
E1053mA peak value faultTube current (mA) peak value exceeds safe limitswarning
E1054kV nominal/actual value faultKilovoltage measured value deviates from nominal setpoint beyond tolerancewarning
E1055mA nominal/actual value faultTube current measured value deviates from nominal setpoint beyond tolerancewarning
E1056X-ray Enable faultX-ray enable control signal abnormalitywarning
E1057Generator life monitoring missingGenerator usage lifetime tracking system failurewarning
E1057CAN monitor monostable timeout during radiationCAN monitor timing fails during active radiationerror
E1058Generator fault - Hardware monitoring missingHardware monitoring circuit failure or data unavailablewarning
E1058RUN monostable timeout during radiationRUN signal timing fails during active radiationerror
E1059X-ray break by external CAN moduleExternal CAN module commanded X-ray shutdownwarning
E1060Signal / X-ray missingExpected X-ray signal or enable command not detectedwarning
E1060X-ray hardware signal missing during radiationX-ray enable signal lost while radiation is activecritical
E1061Generator fault - High-voltage arcArcing detected in high voltage circuit during operationwarning
E1062Nominal kV value greater than maximum kV valueOperator set nominal kV exceeds system maximum specificationwarning
E1063Nominal mA value greater than maximum mA valueOperator set nominal mA exceeds system maximum specificationwarning
E1064Nominal power value greater than maximum power valueCalculated power output exceeds system maximum specificationwarning
E1065Nominal kV value less than minimum kV valueOperator set nominal kV below system minimum specificationwarning
E1066CAN fault line activeCAN bus fault condition detectedwarning
E1067Basic filament heating failureFilament heating not detected during runtimeerror
E1068V-Sync faultsSynchronization timing mismatch between camera and generatorerror
E1068V-sync timing error during runtimeV-sync time check fails during operationerror
E1069Oil temperature 1 error during runtimeTemperature sensor 1 reading out of specificationerror
E1069Temperature sensor 1 value beyond permissible rangeOperationwarning
E1070Temperature sensor 2 value beyond permissible rangeOperationwarning
E1070Oil temperature 2 error during runtimeTemperature sensor 2 reading out of specificationerror
E1071Signal / X-ray without radiationOperationwarning
E1072Oil over-temperature switch activated during runtimeThermal overload switch trips during operationcritical
E1073Humidity sensor error during runtimeHumidity sensor status unexpected during operationwarning
E1074Positive or negative kV actual value exceeds 2 kV without radiationResidual kV voltage detected when radiation inactiveerror
E1074kV value without X-ray command too high, radiation not allowedOperationwarning
E1075Test of rotating anode failed during operationOperationerror
E1076Serial numbers of generator and mobile stand do not matchGenerator serial number does not match mobile standerror
E1082kV actual value after X-ray command too low, radiation not allowedGenerator output voltage insufficient after X-ray commanderror
E1083Generator fault - rotating anode could not be startedOperationwarning
E1084Generator fault - rotating anode failedOperationwarning
E1085Average power in Burst mode > 600 WattBurst mode operationwarning
E1086mA-value too high for thin filamentX-ray operationwarning
E1087Max. anode capacity reached in Burst modeBurst mode operationwarning
E1088Energy reserve of the boost CAPs too smallOperationwarning
E1095Generator overheated, recovery time %d secondsOperationwarning
E1098Generator power-up fault: %0 aPower-uperror
E1099Generator fault 26Generator software mismatch after replacementerror
E1100Hand switch 1 pressed during power-upPower-uperror
E1101Hand switch 2 pressed during power-upPower-uperror
E1102Foot switch 1 pressed during power-upPower-uperror
E1103Foot switch 2 pressed during power-upPower-uperror
E1104Foot switch 3 pressed during power-upPower-uperror
E1105Foot switch 4 pressed during power-upPower-uperror
E1106Foot switch 5 pressed during power-upPower-uperror
E1107Foot switch 6 pressed during power-upPower-uperror
E1108Short circuit in hand switch 1 during power-upPower-uperror
E1109Short circuit in hand switch 2 during power-upPower-uperror
E1110Short circuit in foot switch 1 during power-upPower-uperror
E1111Short circuit in foot switch 2 during power-upPower-uperror
E1112Short circuit in foot switch 3 during power upPower-uperror
E1113Short circuit in foot switch 4 during power-upPower-uperror
E1114Short circuit in foot switch 5 during power-upPower-uperror
E1115Short circuit in foot switch 6 during power-upPower-uperror
E1116X-ray Enable active during power-upPower-uperror
E1117Internal communication error during power-upPower-uperror
E1118Wrong EEPROM checksum during power-upPower-uperror
E1119IRQ PTA5 of RS232 disabled during power-upPower-uperror
E1120Emergency stopManual activation or faultwarning
E1121Hardware interface configuration for wireless module not OKPower-upwarning
E1122Wireless module not presentPower-upwarning
E1132Short circuit in hand switch 1 during X-ray commandX-ray operationwarning
E1133Short circuit in hand switch 2 during X-ray commandX-ray operationwarning
E1134Short circuit in foot switch 1 during X-ray commandX-ray operationwarning
E1135Short circuit in foot switch 2 during X-ray commandX-ray operationwarning
E1136Short circuit in foot switch 3 during X-ray commandX-ray operationwarning
E1137Short circuit in foot switch 4 during X-ray commandX-ray operationwarning
E1138Short circuit in foot switch 12 without X-ray commandOperationwarning
E1139Short circuit in foot switch 6 during X-ray commandX-ray operationwarning
E1140X-ray Enable not active during X-ray commandX-ray operationwarning
E1141CAN fault line active during X-ray commandX-ray operationwarning
E1142Main interface fault during X-ray commandX-ray operationwarning
E1143Image-memory life-monitoring fault during X-ray commandX-ray operationerror
E1146No flow in cooling circuitOperationwarning
E1147Coolant pump inactive at X-ray startX-ray startwarning
E1148Timeout post-fluoro time (LIH)Fluoroscopy operationwarning
E1155X-ray break emergency stopOperationwarning
E1156Short circuit in hand switch 1 during X-ray commandX-ray operationwarning
E1157Short circuit in hand switch 2 during X-ray commandX-ray operationwarning
E1158Short circuit in foot switch 1 during X-ray commandX-ray operationwarning
E1159Short circuit in foot switch 2 during X-ray commandX-ray operationwarning
E1160Short circuit in foot switch 3 during X-ray commandX-ray operationwarning
E1161Short circuit in foot switch 4 during X-ray commandX-ray operationwarning
E1162Short circuit in foot switch 5 during X-ray commandX-ray operationwarning
E1163Short circuit in foot switch 6 during X-ray commandX-ray operationwarning
E1164Short circuit in hand switch 1 without X-ray commandOperationwarning
E1165Short circuit in hand switch 2 without X-ray commandOperationwarning
E1166Short circuit in foot switch 1 without X-ray commandOperationwarning
E1167Short circuit in foot switch 2 without X-ray commandOperationwarning
E1168Short circuit in foot switch 3 without X-ray commandOperationwarning
E1169Short circuit in foot switch 4 without X-ray commandOperationwarning
E1170Short circuit in foot switch 5 without X-ray commandOperationwarning
E1171Short circuit in foot switch 6 without X-ray commandOperationwarning
E1172X-ray Enable without pressing a hand switch or foot switchOperationwarning
E1173CAN fault line activeOperationwarning
E1174Internal communication error during operationOperationwarning
E1175Image-memory life-monitoring faultOperationwarning
E1176Heat-exchanger fan blockedOperationwarning
E1177Flow in cooling circuit too lowOperationwarning
E1178No flow in cooling circuitOperationerror
E1179Coolant pump inactive at X-ray startX-ray starterror
E1180Short circuit in hand switch 9 without X-ray commandOperationwarning
E1181Short circuit in hand switch 10 without X-ray commandOperationwarning
E1182Short circuit in hand switch 11 without X-ray commandOperationwarning
E1184Short circuit in foot switch 13 without X-ray commandOperationwarning
E1185Short circuit in foot switch 14 without X-ray commandOperationwarning
E1186Short circuit in foot switch 15 without X-ray commandOperationwarning
E1187Short circuit in foot switch 16 without X-ray commandOperationwarning
E1188Emergency StopManual activationwarning
E1189Low battery on wireless footswitchOperationwarning
E1190Wireless module is not respondingOperationwarning
E1199Main interface error: %0Operationerror
E12000Function assigned to X-ray switch (%0), ignoredConfigurationwarning
E12001X-ray command assigned to function switch (%0), ignoredConfigurationwarning
E12002XRaySwitch / FunctionSwitch mismatch in mappings of machinedata.iniConfigurationwarning
E12003CAN modules missing: CAN Module No.CAN module cannot be found during startuperror
E12004Hard disk full. Image cannot be storedImage storagewarning
E12005Maximum image storage capacity almost reachedImage storageinfo
E12006vision.md in main interface missingConfigurationwarning
E12007vision.md written to main interfaceConfiguration updateinfo
E12025Radiation allowed after mode values are displayedMode changeinfo
E12026Unable to change APr: Value rejected, restored previous stateParameter adjustmentinfo
E12027Unable to change APr: Value rejected, no recovery possibleParameter adjustmentinfo
E12034License not availableFeature accesserror
E12041Flat-panel detector communication faultOperationwarning
E12042Signature check completedSignature verificationinfo
E12045Radiation blocked by key switchOperationwarning
E12046Radiation blocked by invalid APrConfigOperationwarning
E12050Option stick defective; Vision Center does not work properly; No radiation possibleUSB Dongle not found or defectiveerror
E12050Option stick defectiveOperationwarning
E12051Main Interface defective (serial number)Operationwarning
E12052Serial numbers of mobile stand and monitor cart do not matchOperationwarning
E12053Cannot create dose report: Not current patientDose report creationwarning
E12054Cannot create dose report: No dose data availableDose report creationwarning
E12055Failed to load file %0File accesswarning
E12057Machinedata.ini does not match to capture device frequencyConfiguration checkerror
E12060CAN Manager transport errorCAN communication transport failureerror
E12061CanManager AckBusOffCAN communicationerror
E12062CanManager AckNoTxdCAN communicationerror
E12063Thin filament defective! System switches to restricted mode with thick filamentOperationwarning
E12064Internal fatal errorOperationerror
E12070CanMaster failedCAN communicationerror
E12071CAN module CameraIrisProxy failedCAN communicationerror
E12072CAN module Collimator failedCAN communicationwarning
E12073CAN module Generator failedCAN communicationerror
E12074CAN module MainInterface failedCAN communicationerror
E12075CAN module Regulation failedCAN communicationerror
E12076CAN module MainPrePowerSupply failedCAN communicationerror
E12077CAN module MainPrePowerSupply1 failedCAN communicationerror
E12078CAN module MainPrePowerSupply2 failedCAN communicationerror
E12079CAN module ELWMS failedCAN communicationerror
E12080CAN module Kinematics failedKinematics module communication failureerror
E12080Unknown error in CanMaster. Error state: %0CAN communicationerror
E12082Software update script wasn't foundSoftware updateerror
E12083Software update script couldn't be startedSoftware updateerror
E12084Software update script failed with error %0Software updateerror
E12096Flat-panel detector communication fault: %0Operationwarning
E12098Cannot overwrite file '%s'. Access deniedFile operationwarning
E1246Flat-panel detector communication faultEthernet cable disconnected or not connected between Image Memory and Command Processor during startup or VIVA tool operationcritical
E1300Data transfer to DAC failure (SPI interface)Startup hardware checkerror
E1300PPS power-up faultPower-uperror
E1301PPS overtemperature fault during power-upPower-upwarning
E1301Temperature sensor check failureStartup hardware checkerror
E1302PPS power-up faultDuring power-uperror
E1302Reference voltage DAC (MAX504) check failureStartup hardware checkerror
E1303PPS power-up faultDuring power-uperror
E1303Output voltage check failure from DAC (MAX504)Startup hardware checkerror
E1304Wrong EEPROM checksum during power-upDuring power-uperror
E1350SPI interface data transfer error (runtime)Every nominal value change during runtimeerror
E1350PPS power-up faultDuring power-uperror
E1351PPS overtemperature fault during operationDuring system operationwarning
E1351Temperature error (runtime)Permanent runtime temperature monitoringwarning
E1355CAN fault detectionPermanent runtime CAN line monitoringerror
E1358Output voltage beyond toleranceDuring system operationwarning
E1398PPS power-up faultDuring power-uperror
E1399PPS faultDuring operationerror
E1400Servo Rotation position faultDuring operationwarning
E1401Servo Rotation not connectedDuring operationwarning
E1402Servo Rotation not readyDuring operationwarning
E1403Servo Iris position faultDuring operationwarning
E1404Servo Iris not connectedDuring operationwarning
E1405Servo Iris not readyDuring operationwarning
E1406Servo Slot 1 position faultDuring operationwarning
E1407Servo Slot 1 not connectedDuring operationwarning
E1408Servo Slot 1 not readyDuring operationwarning
E1409Servo Slot 2 position faultDuring operationwarning
E1410Servo Slot 2 not connectedDuring operationwarning
E1411Servo Slot 2 not readyDuring operationwarning
E1412Servo Corner position faultDuring operationwarning
E1413Servo Corner not connectedDuring operationwarning
E1414Servo Corner not readyDuring operationwarning
E1420DAP meter mismatch or not availableDuring operationwarning
E1421DAP meter errorDuring operationwarning
E1448Collimator startup errorDuring power-up or collimator initializationwarning
E1449Vision Track unknown errorDuring Vision Track operationwarning
E1450Servo position faultDuring operationwarning
E1451Servo not connectedDuring operationwarning
E1452Servo not readyDuring operationwarning
E1499I.I. power supply unknown errorDuring operationwarning
E1700kVmA table: kV not ascendingDuring table validationwarning
E1701kVmA table: power not ascendingDuring table validationwarning
E1702kVmA table not finishedDuring table validationwarning
E1705No external video signalDuring operationwarning
E1708Generator pulse errorDuring operationwarning
E1709No VSyncDuring operationwarning
E1710No InterruptDuring operationwarning
E1711No communication PPC-generatorDuring operationwarning
E1712Wrong mA value from generatorDuring operationwarning
E1713X-ray not allowedDuring operation when X-ray should not emitwarning
E1714Max. mA value cannot be readDuring operationwarning
E1716Image-memory power-up faultDuring power-uperror
E1717Image-memory power-up faultDuring power-uperror
E1718Image-memory power-up faultDuring power-upwarning
E1719Gain adjustment not possibleDuring power-up or calibrationerror
E1720Image-memory faultDuring operationerror
E1721Image-memory faultDuring operationerror
E1722V-Sync corruptAt least one V-sync pulse missing at image memory inputerror
E1723Video standard unknownDuring operationerror
E1724Image-memory faultDuring power-uperror
E1725M-Data unknownDuring operationwarning
E1726Collimator does not support DAP valuesDuring operationwarning
E1727Flat-panel detector not supported by this versionDuring operationwarning
E1728Internal fatal communication errorDuring operationwarning
E1729Internal communication errorDuring operationerror
E1732Image-memory power-up faultDuring power-uperror
E1733Image-memory power-up faultDuring power-uperror
E1734Image-memory power-up faultDuring power-uperror
E1735Image-memory power-up faultDuring power-uperror
E1736Image-memory power-up faultDuring power-uperror
E1737Image-memory power-up faultDuring power-uperror
E1738No external video signalNo video signal present at image memory inputerror
E1740Hardware doesn't support grid regulationDuring operationwarning
E1741Hardware doesn't support motion detectionDuring operationwarning
E1742Hardware doesn't support DAP calculationDuring operationwarning
E1743Hardware doesn't support automatic dose reductionDuring operationwarning
E1747Black level >= White levelDuring calibration or operationerror
E1756FFC board not availableDuring power-up or operationerror
E1757FFC not working properlyDuring operationerror
E1759Generator version and PPC version not compatibleDuring power-uperror
E1760High resolution with pulse length 0 not allowedDuring operation with high resolution modeerror
E1789PPC unknown errorDuring operationerror
E1790PPC CAN Exception errorDuring operationwarning
E1800Radiation indication lamp defective during power-upDuring power-upwarning
E1801Wrong EEPROM checksum during power-upDuring power-upwarning
E1802Image-memory power-up faultDuring power-upwarning
E1803Image-memory power-up faultDuring power-upwarning
E1804No communication with left flatscreenDuring power-up or operationwarning
E1805No communication with right flatscreenDuring power-up or operationwarning
E1806Image-memory power-up faultDuring power-upwarning
E1807Image-memory power-up faultDuring power-upwarning
E1808Switch on printerDuring operationwarning
E1850Radiation indication lamp defective during operationDuring system operationwarning
E1865U578 execution failedDuring U578 operationwarning
E1898U578 Startup ErrorDuring power-upwarning
E1899U578 ErrorDuring operationwarning
Setting the network configuration failedNetwork configuration cannot be applied to systemDuring network settings configurationerror
Verification failed OM[5]: Association rejected: Wrong serveDICOM AE title mismatch prevents server connectionTest button pressed during DICOM server configurationerror
Verification failed OM[6]: Association request failedDICOM server connection cannot be establishedTest button pressed during DICOM server configurationerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI