Ziehm Imaging Vision RFD 3D: troubleshooting and error codes
The ServiceForge knowledge base holds 42 documented symptoms for the Ziehm Imaging Vision RFD 3D, with 58 ranked causes, 139 diagnostic checks with test points and expected values, and 372 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration influence on image quality assessment
Understanding and assessing the impact of calibration accuracy (0.14 superior vs 1.07 insufficient) on artifact presence, edge sharpness, and metal visualization
Ranked causes
- 1. Calibration value superior (0.14 mm) achieves optimal image quality (common)
- 2. Calibration value good (0.39 mm) produces acceptable image with minimal artifacts (common)
- 3. Calibration value insufficient (> 0.39 mm) produces unacceptable artifacts (occasional)
Compare calibration value 0.14 mm (superior) image to reference
- Test point
- Phantom layer reconstruction quality
- Expected
- Sharp phantom edges; metal pin without artifacts; clear crisp image
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: 3D image reconstruction accuracy radiation
Verify phantom positioning, ISO center alignment, C-arm collision limits, scan acquisition, accuracy calculation, and scaling factor validation for 3D imaging
Ranked causes
- 1. Initial phantom positioning and alignment not verified (common)
- 2. Mean accuracy value exceeds tolerance or scaling factor out of range (occasional)
- 3. Collision limits not verified (common)
Position phantom parallel to flat-panel detector in LAT position
- Test point
- Phantom position relative to detector
- Expected
- Parallel to detector; sides parallel to screen; central sphere inside crosshair
- Tools
- Laser alignment tool, fluoroscopy
10 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: 3D reconstruction image quality verification radiation
Verify phantom placement, C-arm positioning, image acquisition, and visual quality assessment of phantom grooves, metal pin, and diameter measurements
Ranked causes
- 1. Phantom not properly positioned for image quality check (common)
- 2. Phantom grooves not clearly visible or not properly formed (occasional)
- 3. Metal pin boundaries or diameter not within specification (occasional)
Mount attenuation device and position verification phantom
- Test point
- 25mm Al device mounted on generator; phantom Part No. 62109 in beam path
- Expected
- Attenuation device secured with tension strap; paper protective layer in place; phantom in beam path
- Tools
- Tension strap, 25mm Al attenuation device, paper
9 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Video adjustment radiation
Video signal levels must be calibrated using histogram function and grey step wedge. Black level, white level, and metal black level parameters control video signal output.
Ranked causes
- 1. Black level requires adjustment (common)
- 2. White level requires adjustment (common)
- 3. Metal black level requires adjustment (occasional)
Rotate image to mask circle position
- Test point
- Monitor display
- Expected
- Circular mask visible on image
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error E1246: Flat-panel detector communication fault
Ethernet cable disconnected or misconfigured between Image Memory and Command Processor during VIVA tool operation or system startup.
Ranked causes
- 1. Ethernet cable not connected between Image Memory and Command Processor during VIVA gain calibration setup (common)
- 2. System restarted with Ethernet cable disconnected from Image Memory (common)
- 3. VIVA tool Ethernet link not opened after physical connection established (occasional)
Verify physical Ethernet cable connection
- Test point
- Red connector ports on Image Memory and Command Processor 2
- Expected
- Straight Ethernet cable #24452 with inline coupler #217268 securely seated; cable not damaged or kinked
- Tools
- Visual inspection; straight Ethernet cable #24452; inline coupler #217268
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Center beam position radiation
Center beam test device image not concentric or centered on screen; may appear off-center to display
Ranked causes
- 1. Generator angle misalignment (left-right deviation) (common)
- 2. Generator attachment misalignment (top-down deviation) (common)
Visually inspect center beam test device image on screen for left-right centering
- Test point
- Screen display
- Expected
- Concentric shape centered horizontally on image
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Dose adjustment radiation
X-ray dose output must be calibrated using 25 mm Al + 1.5 mm copper filtration with dose measured in front of image intensifier grid. Dose Gain and CCD/FD Fill parameters control dose output.
Ranked causes
- 1. Dose Gain requires adjustment (common)
- 2. CCD/FD Fill requires adjustment (common)
Set CCD fill level to specified value
- Test point
- Dose 1 tab, CCD/FD-Fill input box
- Expected
- Value from System Settings CCD fill levels table
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: DSA (Digital Subtraction Angiography) dose adjustment radiation
Two separate dose adjustments available for DSA: HQ DSA and LPK DSA. Measurement uses dose rate with Frames/sec set to 25. Original setting must be restored after adjustment.
Ranked causes
- 1. HQ DSA dose requires adjustment (common)
- 2. LPK DSA dose requires adjustment (common)
Open Configuration Menu and set Frames/sec to 25
- Test point
- Configuration Menu parameter
- Expected
- Frames/sec = 25
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Module 10 Power Distribution Setup high voltage
Configure U539 board voltage selection and connector wiring for transformer RTO 2500-000585 or RTO 2500-001195 based on input AC voltage (105-240 VAC range).
Ranked causes
- 1. Voltage selector configuration required for primary AC input (common)
- 2. Voltage selector connectors X40-X50 require proper connection to transformer tap points (common)
Verify transformer primary taps are connected per selected input voltage
- Test point
- Transformer RTO 2500-000585 or 2500-001195 primary terminals A, B, C, D, E, F, H, I, K, L
- Expected
- Connections match table for selected input voltage (105, 110, 115, 120, 210, 220, 230, or 240 VAC)
- Tools
- Visual inspection, wiring diagram
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LED Indicator Verification
Verify all LEDs on U539g board illuminate correctly to confirm power distribution and controller status after configuration.
Ranked causes
- 1. LEDs indicate power rail and relay status during startup and normal operation (common)
- 2. HC08 processor monitoring status indicated by LEDs D45-D49, D15-D16 (common)
Verify D11, D8, D9 relay status LEDs illuminate in sequence during power-up
- Test point
- LEDs D11 (K1 mains 1), D8 (K2 inrush), D9 (K3 mains 2)
- Expected
- D8 illuminates briefly during inrush, then D11 and D9 steady on (green)
- Tools
- Visual inspection
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Standby LED flashing due to mains 2 contactor not switching caution
Standby LED flashes when CPU did not switch ON mains 2 contactor and input signal from PTH0 is LOW
Ranked causes
- 1. Short circuit from transistor T8 on PCB U539 (occasional)
- 2. CPU failed to control mains 2 contactor (occasional)
Measure voltage at PTH0 input
- Test point
- PTH0
- Expected
- HIGH (when mains 2 contactor should be ON)
- Tools
- DMM or oscilloscope
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: 26 inch monitor menu lock reset
Accessing locked menu on Radiance G2 HB monitor and restoring factory default settings
Ranked causes
- 1. Menu locked on 26 inch Radiance G2 HB monitor (common)
Verify menu access on 26 inch monitor
- Test point
- OSD display
- Expected
- Menu appears without MENU LOCKED message; Setup and Defaults sections accessible
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: 3D image quality verification
Visual inspection and measurement of 3D reconstruction quality including sagittal, axial and coronal views for contrast, sharpness and artifact assessment
Ranked causes
- 1. 3D calibration requires verification (common)
Measure length of drilling holes in sagittal view including artifacts
- Test point
- upper hole
- Expected
- between 11.8 mm and 12.4 mm, 11.8–12.4 mm
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Adipose Patient (LPK) mode dose adjustment radiation
LPK (Adipose Patient) is a modifier to each application program. Dose adjustment must be performed in all programs with same value. LPK has individual kV/mA curve.
Ranked causes
- 1. LPK mode dose requires adjustment in all programs (common)
Select LPK modifier and adjust dose in all programs
- Test point
- Adjustment dialog; all anatomical programs
- Expected
- Same dose value per System Settings in each program
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Asymmetric collimator DIP switch configuration
DIP switch settings on U659 module for asymmetric collimator operation and dose meter selection
Ranked causes
- 1. Incorrect DIP switch settings for operational mode (common)
Verify DIP switch 1 (Standalone mode) position
- Test point
- S2 Switch 1
- Expected
- OFF (standalone mode inactive)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Center Beam Adjustment radiation
Adjust generator beam center position relative to system screen after generator installation. Done before collimator installation using test device and pre-collimator for radiation protection.
Ranked causes
- 1. Generator mounted and requires center beam alignment before collimator installation (common)
Gather required tools and devices
- Test point
- Equipment inventory
- Expected
- Center beam test device (91116) and pre-collimator (65699) available
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Computed DAP / Air Kerma radiation
Adjustment of dose area product (DAP) and air kerma measurements using reference dose meter and Ziehm test phantom. Requires setting calibration points, pulse width correction, area factor, and kerma/DAP calculation factors.
Ranked causes
- 1. Calibration required for accurate dose measurement display (common)
Generator cover with filtration mounted and dose meter set to 3mm Al prefiltering
- Test point
- Generator cover, dose meter settings
- Expected
- Generator cover mounted, dose meter configured for 3mm Al prefilter
8 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Drilling hole measurement technique
Procedure to measure drilling hole lengths in sagittal view for calibration validation
Ranked causes
- 1. Measurement requires accuracy verification (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Flat-panel detector gain radiation
Adjust detector gain for image homogeneity and proper signal levels across panel modes via manual kV/mA tuning or VIVA tool software, with Gain Median validation.
Ranked causes
- 1. Gain calibration required for detector modes (common)
System warm-up time
- Test point
- System startup to calibration start
- Expected
- Minimum 1 hour elapsed
21 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: High Quality Mode dose adjustment radiation
HQ (High Quality) mode is a modifier to each application program. Dose adjustment must be performed in all programs with same value. HQ has individual kV/mA curve.
Ranked causes
- 1. HQ mode dose requires adjustment in all programs (common)
Select HQ modifier and adjust dose in all programs
- Test point
- Adjustment dialog; all anatomical programs
- Expected
- Same dose value per System Settings in each program
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Mains input voltage selection high voltage
Configure power supply board U539h for correct input voltage (105–240 VAC) by hardwiring connector pairs and selecting appropriate fuses (F1, F2) to match site mains supply.
Ranked causes
- 1. Voltage mismatch between system configuration and mains supply (common)
Verify mains supply voltage at site installation
- Test point
- Mains supply outlet
- Expected
- 105, 110, 115, 120, 210, 220, 230, or 240 VAC ±5%
2 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Measured DAP with VacuDAP-C radiation
Adjustment of measured dose area product (DAP) using VacuDAP-C external dose meter. Includes self-test procedure and state error diagnosis using 8-bit error code system.
Ranked causes
- 1. VacuDAP-C meter requires calibration factor adjustment (common)
VacuDAP-C firmware version and tube voltage reference
- Test point
- DAP Meter input box
- Expected
- Firmware version displayed, currently activated tube voltage shown
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Measuring kV high voltage
Measure kV output from generator to verify proper tube voltage. Measured values are divided by 10,000 (e.g., 6.7 V reading = 67 kV actual).
Ranked causes
- 1. kV measurement required during service or commissioning (common)
Set generator to continuous mode for accurate measurement
- Test point
- Adjustment dialog Dose 2
- Expected
- 100% pulse width or Pulse Length = 0
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Measuring mA high voltage
Measure mA (tube current) output from generator to verify proper tube current. Requires connector X3 on U564 board with jumper removed for measurement.
Ranked causes
- 1. mA measurement required during service or commissioning (common)
Remove jumper from X3 connector
- Test point
- Connector X3 on U564 board
- Expected
- Jumper removed, connector accessible for meter connection
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monitor line voltage matching high voltage
Service technician must match facility input voltage to TR1 isolation transformer to ensure stable 230 VAC output for C-arm system
Ranked causes
- 1. Input voltage not matched to TR1 transformer tap settings (common)
Measure facility wall outlet voltage
- Test point
- Wall outlet connector
- Expected
- Stable within Voltage Selection table range
- Tools
- Digital volt meter
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monitor on-screen display menu lock/unlock
Setting or resetting keypad lock on 19 inch dual monitor via simultaneous key press, or locking/unlocking menu on 26 inch Radiance G2 HB monitor
Ranked causes
- 1. Keypad lock engaged on 19 inch monitor (common)
Verify keypad lock status on 19 inch monitor
- Test point
- OSD display
- Expected
- No lock message when keys pressed; Input and L/R keys responsive
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: RSA (Rotational Subtraction Angiography) dose adjustment radiation
RSA is a separate program requiring individual dose adjustment. Selected via RSA function in adjustment dialog dropdown menu.
Ranked causes
- 1. RSA mode dose requires adjustment (common)
Select RSA function in adjustment dialog dropdown
- Test point
- Adjustment dialog dropdown menu
- Expected
- RSA function selected
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Service Position for Generator Electronics Access caution
Position generator electronics module for access to test points and measurement connectors during service and calibration procedures.
Ranked causes
- 1. Generator electronics access required for measurement and calibration (common)
Remove specified screw from generator module
- Test point
- Generator electronics enclosure
- Expected
- Screw removed and set aside safely
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Snapshot mode dose adjustment radiation
Snapshot is a separate program requiring individual dose adjustment. Accumulated dose for one shot must be measured after regulation stabilizes, target <2 µGy.
Ranked causes
- 1. Snapshot mode dose requires calibration (common)
Select SSHT key in adjustment dialog
- Test point
- Adjustment dialog dropdown
- Expected
- SSHT (Snapshot) mode selected
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ethernet link error during image retrieval
Error message appears when attempting to retrieve images from detector; image data transfer fails.
Ranked causes
- 1. Ethernet link communication interrupted or lost (occasional)
Monitor for error message during image retrieval
- Test point
- Acquisition software display
- Expected
- No error message; successful image transfer
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inrush relay fault condition high voltage
Inrush relay switched ON but system does not detect proper inrush resistor voltage drop within timing window
Ranked causes
- 1. Inrush resistor failure or open circuit (occasional)
Measure analog voltage at PTB7 during inrush cycle
- Test point
- PTB7
- Expected
- Greater than 3.0 V, above 3.0 V
- Tools
- DMM or oscilloscope, scope set to capture inrush transient
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Jumper Configuration for Workstation Options
Set jumpers J1, J2, J3 on U539h/j to enable/disable workstation control options.
Ranked causes
- 1. Jumpers select workstation presence and control mode (common)
Verify jumper J1, J2, J3 positions match workstation configuration
- Test point
- Jumpers J1, J2, J3 on U539h/j board
- Expected
- J1, J2, J3 open if corresponding workstations not installed; closed if WS No.1 control enabled
- Tools
- Visual inspection
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Protective and Auxiliary Supply Fuse Configuration caution
Install or verify fuses F2, F3, F4, F5 per AC input voltage configuration (100/105-120/210-240 VAC).
Ranked causes
- 1. Supply voltage fuses must match system input voltage rating (common)
Verify F1A and F1B main fuses per selected voltage
- Test point
- Fuses F1A, F1B external on U539g board
- Expected
- 105-120 VAC: 20 A slow (17044); 210-240 VAC: 15 A slow (17020), exact rating required
- Tools
- Visual inspection, parts list
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Retrieved images appear incorrect or uncorrected
Detector images retrieved from system show appearance similar to uncorrected raw data with visible artifacts or non-uniform response.
Ranked causes
- 1. Image corrections disabled in VIVA tool configuration (common)
Verify image correction status in VIVA configuration
- Test point
- System Settings dialog
- Expected
- Image corrections enabled in configuration
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Standby LED and port PTF LEDs flashing caution
Standby LED and all LEDs from port PTF (D45, D46, D47, D48, D49, D15, D16) flash when PCB powered and analog voltage from PTB0 less than 1.0 V
Ranked causes
- 1. Remote button ON short circuit or pressed (common)
Measure analog voltage at port PTB0
- Test point
- PTB0
- Expected
- Greater than 1.0 V, above threshold
- Tools
- DMM or oscilloscope
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Standby LED flashing when OFF button pressed
Standby LED flashes when system in OFF state and OFF button (remote or local) is pressed
Ranked causes
- 1. Normal operation during OFF state transition (common)
Verify system is in OFF state
- Test point
- Power state indicator
- Expected
- System powered down, not in blocking state
VacuDAP-C state error: DAP overflow (Bit 2 = Decimal 4)
DAP meter accumulated dose has exceeded maximum measurable range.
Ranked causes
- 1. Accumulated DAP exceeds meter range during long screening session (occasional)
VacuDAP-C state error Bit 2
- Test point
- State field value
- Expected
- Bit 2 = 0 (no overflow) decimal
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
VacuDAP-C state error: Error test function (Bit 4 = Decimal 16)
Self-test result does not match reference value, indicating meter malfunction or calibration error.
Ranked causes
- 1. Self-test measured value differs from stored reference value (occasional)
Self-test result vs Normal Value Outer Field
- Test point
- State field, test result vs reference
- Expected
- Test result within tolerance of Normal Value Outer Field
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
VacuDAP-C state error: Firmware error (Bit 6 = Decimal 64)
Meter firmware is corrupted or malfunctioning.
Ranked causes
- 1. VacuDAP-C firmware corruption or hardware defect in microcontroller (rare)
VacuDAP-C state error Bit 6
- Test point
- State field value
- Expected
- Bit 6 = 0 (no firmware error) decimal
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
VacuDAP-C state error: TEST-Warning (Bit 1 = Decimal 2)
VacuDAP-C self-test warning indicating possible meter malfunction or calibration drift.
Ranked causes
- 1. Meter test value does not match reference specification (occasional)
VacuDAP-C self-test result vs reference value
- Test point
- State field, test result vs Normal Value Outer Field
- Expected
- Test result matches or is very close to Normal Value Outer Field reference
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
VacuDAP-C state error: Voltage for measuring chamber faulty (Bit 5 = Decimal 32)
Measuring chamber supply voltage is defective or out of specification.
Ranked causes
- 1. Power supply for ionization chamber damaged or disconnected (rare)
VacuDAP-C state error Bit 5
- Test point
- State field value
- Expected
- Bit 5 = 0 (no voltage fault) decimal
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
VacuDAP-C state error: Zero check error (Bit 3 = Decimal 8) radiation
Radiation was active during system reset, corrupting zero reference calibration.
Ranked causes
- 1. Radiation exposure during dose meter reset or power-up (rare)
Radiation status during system reset
- Test point
- X-ray enable signal, meter status
- Expected
- Radiation off during reset, zero calibration completes without exposure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
2900 | Kinematic Emergency Stop | Emergency stop button pressed or safety condition triggered | warning |
2901 | A Axis Fault | During kinematic operation | warning |
2902 | C Axis Fault | During kinematic operation | warning |
2903 | Y Axis Fault | During kinematic operation | warning |
2904 | Z Axis Fault | During kinematic operation | warning |
2905 | X-ray Fault | During operation | warning |
2906 | CAN Fault line active | During operation | warning |
2907 | Distance Control Fault | During kinematic operation | warning |
2914 | Internal Fault | During operation | warning |
2915 | Internal Fault | During power-up | warning |
2916 | Internal CAN Fault | During operation | warning |
2917 | Internal CAN Fault | During operation | warning |
2918 | Internal Fault | During operation | warning |
2919 | Internal Fault | During operation | warning |
2920 | Internal Fault | During operation | warning |
2921 | Internal Fault | During operation | warning |
2922 | Y Axis ISO speed Fault | During kinematic operation | warning |
2923 | Z Axis ISO speed Fault | During kinematic operation | warning |
2924 | ISO Fault | During kinematic operation | warning |
2925 | Internal Distance Control Fault | During operation | warning |
2926 | Internal Fault | During operation | warning |
2927 | Internal Distance Control Fault | During operation | warning |
2928 | Internal Distance Control Fault | During operation | warning |
2929 | Internal Distance Control Fault | During operation | warning |
2930 | Internal Distance Control Fault | During operation | warning |
2931 | Internal Fault | During operation | warning |
2932 | Emergency Stop Fault | When emergency stop button is pressed | warning |
2933 | Internal Distance Control Fault | During operation | warning |
2934 | Internal Distance Control Fault | During operation | warning |
2935 | Internal Fault | During operation | warning |
2936 | Navigation Fault | During operation | warning |
2937 | Parking brake engaged | During operation when brake is engaged | warning |
2938 | Swivel brake engaged | During operation when brake is engaged | warning |
2939 | Internal Fault | During operation | warning |
2940 | Internal Fault | During operation | warning |
2941 | Internal Fault | During power-up | warning |
2955 | Internal Fault Distance Control | During operation | warning |
2956 | Speed Fault | During operation | warning |
2957 | Collision Internal Fault | During operation | warning |
2958 | Internal Fault | During operation | warning |
2959 | Internal Fault | During operation | warning |
2960 | A Axis Fault | During kinematic operation | warning |
2961 | Y Axis Fault | During kinematic operation | warning |
2962 | C Axis Fault | During kinematic operation | warning |
2963 | Z Axis Fault | During kinematic operation | warning |
2964 | Kinematic module error | During operation | warning |
3000 | Axis not clutched | When axis is required but not engaged | warning |
3001 | Axis brake closed | When axis should move but brake is applied | warning |
3002 | Axis %d on limit A | Axis reaches limit position A | warning |
3003 | Axis %d on limit B | Axis reaches limit position B | warning |
3004 | Axis %d hardware over current | Hardware detects excessive current in axis | error |
3005 | Axis %d software over current | Software detects excessive current in axis | error |
3006 | Axis %d contouring error | Axis position deviates from commanded path | warning |
3007 | Axis %d extended error | Extended error condition detected on axis | error |
3008 | Axis %d not enabled | Axis control logic not activated | warning |
3010 | Axis %d solenoid current error | Solenoid current out of specification | error |
3011 | Axis %d engine temperature error | Motor temperature exceeds limit | error |
3012 | Axis %d hw voltage too low | Hardware power supply voltage below minimum | error |
3013 | Axis %d sw voltage too low | Software-detected power supply voltage below minimum | error |
3014 | Axis %d watchdog error | Watchdog timeout indicates processor hang | critical |
3015 | Axis %d tacho error | Tachometer signal missing or invalid | error |
3016 | Axis %d encoder error | Encoder signal invalid or missing | error |
3017 | Axis %d encoder error | Second encoder signal fault | error |
3018 | Axis %d encoder range error | Encoder position value out of valid range | error |
3019 | Axis %d gear error | Gear drive fault detected | error |
3020 | Axis %d port extender error | Port expander communication failure | error |
3021 | Axis %d emergency stop | Emergency stop button activated | warning |
3022 | Axis %d wigwag error | Oscillatory motion detected | warning |
3023 | Axis %d motor error | Motor control fault detected | error |
3024 | Axis %d motor error | Motor error with safety processor in reset | critical |
3025 | Axis %d motor error | Motor error with lift key active | error |
3026 | Axis %d in autarkic mode | Axis operating independently of main control | warning |
3027 | Axis %d unknown error | Unrecognized error condition | error |
E1000 | Filament not operating (small focus) | Small focus filament fails to initialize during bootup | error |
E1000 | Thin filament defective during power-up | System power-up detection of thin filament circuit failure | error |
E1001 | Thick filament defective during power-up | System power-up detection of thick filament circuit failure | error |
E1001 | Filament not operating (large focus) | Large focus filament fails to initialize during bootup | error |
E1002 | V-sync signal missing | V-sync signal not detected during bootup | error |
E1002 | No V-Sync | Video synchronization signal missing | error |
E1002 | No VSync at generator during power-up | Missing video sync signal from generator during startup | error |
E1003 | Generator temperature minimum not reached - please wait | Generator temperature below 13°C during startup | error |
E1004 | Oil temperature above maximum during bootup | Oil temperature maximum check fails | error |
E1004 | Generator must cool down below 72°C - please wait | Generator temperature exceeds 72°C during operation | warning |
E1005 | Generator temperature minimum not reached - please wait | Generator temperature below 13°C during startup (Sensor 2) | error |
E1006 | Generator must cool down below 72°C - please wait | Generator temperature exceeds 72°C during operation (Sensor 2) | warning |
E1006 | Oil temperature above maximum during runtime | Oil temperature maximum check fails during operation | error |
E1007 | Wrong EEPROM checksum during power-up | EEPROM memory checksum validation fails at startup | error |
E1007 | Machine data checksum error | Checksum validation fails during bootup | error |
E1008 | Generator fault - Interrupt 1 flip-flop defective | Interrupt control flip-flop 1 circuit failure detected | error |
E1008 | Hardware flip-flop for external interrupt error | Flip-flop not operating correctly during bootup | error |
E1009 | Hardware flip-flop for external interrupt error | Flip-flop not operating correctly during bootup | error |
E1009 | Generator fault - Interrupt 2 flip-flop defective | Interrupt control flip-flop 2 circuit failure detected | error |
E1010 | HW signal X-ray active during power-up | X-ray enable signal detected as active during system startup | error |
E1011 | CAN X-ray Enable active during power-up | CAN bus X-ray enable command detected during startup | error |
E1012 | Generator fault - D/A converter reference voltage defective | D/A converter reference voltage out of specification | error |
E1012 | Reference voltage D/A converter error | Reference voltage measurement fails | error |
E1013 | Generator fault - D/A converter kV defective | D/A converter kilovoltage output circuit failure | error |
E1013 | Actual voltage from kV D/A converter error | kV actual voltage measurement fails or out of range | error |
E1014 | Oil overtemperature switch active during power-up - call service | Oil temperature sensor alarm activated at startup | error |
E1015 | Machine data validity error | Invalid data bytes or version mismatch in machine data | error |
E1015 | Wrong machine data version detected during power-up | Machine configuration data version mismatch during initialization | error |
E1016 | Moisture sensor alarm during power-up - call service | Humidity sensor detects excess moisture during startup | error |
E1016 | Humidity sensor active/not active error | Humidity sensor status unexpected during bootup | warning |
E1017 | Test of FAULT signal failed | Internal fault signal test during diagnostics fails | error |
E1018 | Debug switch is on | Debug mode enabled at system startup | warning |
E1019 | Generator fault - Anode rotation monitoring does not work | Anode rotation feedback signal missing or abnormal during operation | error |
E1050 | High voltage fault | High voltage supply out of specification or unstable | warning |
E1050 | HVC fault hardware signal | High voltage controller detects fault condition | critical |
E1051 | High voltage fault (anode) | Anode high voltage supply malfunction | warning |
E1052 | High voltage fault (cathode) | Cathode high voltage supply malfunction | warning |
E1053 | mA peak error hardware signal | mA peak error detected in filament regulation | error |
E1053 | mA peak value fault | Tube current (mA) peak value exceeds safe limits | warning |
E1054 | kV nominal/actual value fault | Kilovoltage measured value deviates from nominal setpoint beyond tolerance | warning |
E1055 | mA nominal/actual value fault | Tube current measured value deviates from nominal setpoint beyond tolerance | warning |
E1056 | X-ray Enable fault | X-ray enable control signal abnormality | warning |
E1057 | Generator life monitoring missing | Generator usage lifetime tracking system failure | warning |
E1057 | CAN monitor monostable timeout during radiation | CAN monitor timing fails during active radiation | error |
E1058 | Generator fault - Hardware monitoring missing | Hardware monitoring circuit failure or data unavailable | warning |
E1058 | RUN monostable timeout during radiation | RUN signal timing fails during active radiation | error |
E1059 | X-ray break by external CAN module | External CAN module commanded X-ray shutdown | warning |
E1060 | Signal / X-ray missing | Expected X-ray signal or enable command not detected | warning |
E1060 | X-ray hardware signal missing during radiation | X-ray enable signal lost while radiation is active | critical |
E1061 | Generator fault - High-voltage arc | Arcing detected in high voltage circuit during operation | warning |
E1062 | Nominal kV value greater than maximum kV value | Operator set nominal kV exceeds system maximum specification | warning |
E1063 | Nominal mA value greater than maximum mA value | Operator set nominal mA exceeds system maximum specification | warning |
E1064 | Nominal power value greater than maximum power value | Calculated power output exceeds system maximum specification | warning |
E1065 | Nominal kV value less than minimum kV value | Operator set nominal kV below system minimum specification | warning |
E1066 | CAN fault line active | CAN bus fault condition detected | warning |
E1067 | Basic filament heating failure | Filament heating not detected during runtime | error |
E1068 | V-Sync faults | Synchronization timing mismatch between camera and generator | error |
E1068 | V-sync timing error during runtime | V-sync time check fails during operation | error |
E1069 | Oil temperature 1 error during runtime | Temperature sensor 1 reading out of specification | error |
E1069 | Temperature sensor 1 value beyond permissible range | Operation | warning |
E1070 | Temperature sensor 2 value beyond permissible range | Operation | warning |
E1070 | Oil temperature 2 error during runtime | Temperature sensor 2 reading out of specification | error |
E1071 | Signal / X-ray without radiation | Operation | warning |
E1072 | Oil over-temperature switch activated during runtime | Thermal overload switch trips during operation | critical |
E1073 | Humidity sensor error during runtime | Humidity sensor status unexpected during operation | warning |
E1074 | Positive or negative kV actual value exceeds 2 kV without radiation | Residual kV voltage detected when radiation inactive | error |
E1074 | kV value without X-ray command too high, radiation not allowed | Operation | warning |
E1075 | Test of rotating anode failed during operation | Operation | error |
E1076 | Serial numbers of generator and mobile stand do not match | Generator serial number does not match mobile stand | error |
E1082 | kV actual value after X-ray command too low, radiation not allowed | Generator output voltage insufficient after X-ray command | error |
E1083 | Generator fault - rotating anode could not be started | Operation | warning |
E1084 | Generator fault - rotating anode failed | Operation | warning |
E1085 | Average power in Burst mode > 600 Watt | Burst mode operation | warning |
E1086 | mA-value too high for thin filament | X-ray operation | warning |
E1087 | Max. anode capacity reached in Burst mode | Burst mode operation | warning |
E1088 | Energy reserve of the boost CAPs too small | Operation | warning |
E1095 | Generator overheated, recovery time %d seconds | Operation | warning |
E1098 | Generator power-up fault: %0 a | Power-up | error |
E1099 | Generator fault 26 | Generator software mismatch after replacement | error |
E1100 | Hand switch 1 pressed during power-up | Power-up | error |
E1101 | Hand switch 2 pressed during power-up | Power-up | error |
E1102 | Foot switch 1 pressed during power-up | Power-up | error |
E1103 | Foot switch 2 pressed during power-up | Power-up | error |
E1104 | Foot switch 3 pressed during power-up | Power-up | error |
E1105 | Foot switch 4 pressed during power-up | Power-up | error |
E1106 | Foot switch 5 pressed during power-up | Power-up | error |
E1107 | Foot switch 6 pressed during power-up | Power-up | error |
E1108 | Short circuit in hand switch 1 during power-up | Power-up | error |
E1109 | Short circuit in hand switch 2 during power-up | Power-up | error |
E1110 | Short circuit in foot switch 1 during power-up | Power-up | error |
E1111 | Short circuit in foot switch 2 during power-up | Power-up | error |
E1112 | Short circuit in foot switch 3 during power up | Power-up | error |
E1113 | Short circuit in foot switch 4 during power-up | Power-up | error |
E1114 | Short circuit in foot switch 5 during power-up | Power-up | error |
E1115 | Short circuit in foot switch 6 during power-up | Power-up | error |
E1116 | X-ray Enable active during power-up | Power-up | error |
E1117 | Internal communication error during power-up | Power-up | error |
E1118 | Wrong EEPROM checksum during power-up | Power-up | error |
E1119 | IRQ PTA5 of RS232 disabled during power-up | Power-up | error |
E1120 | Emergency stop | Manual activation or fault | warning |
E1121 | Hardware interface configuration for wireless module not OK | Power-up | warning |
E1122 | Wireless module not present | Power-up | warning |
E1132 | Short circuit in hand switch 1 during X-ray command | X-ray operation | warning |
E1133 | Short circuit in hand switch 2 during X-ray command | X-ray operation | warning |
E1134 | Short circuit in foot switch 1 during X-ray command | X-ray operation | warning |
E1135 | Short circuit in foot switch 2 during X-ray command | X-ray operation | warning |
E1136 | Short circuit in foot switch 3 during X-ray command | X-ray operation | warning |
E1137 | Short circuit in foot switch 4 during X-ray command | X-ray operation | warning |
E1138 | Short circuit in foot switch 12 without X-ray command | Operation | warning |
E1139 | Short circuit in foot switch 6 during X-ray command | X-ray operation | warning |
E1140 | X-ray Enable not active during X-ray command | X-ray operation | warning |
E1141 | CAN fault line active during X-ray command | X-ray operation | warning |
E1142 | Main interface fault during X-ray command | X-ray operation | warning |
E1143 | Image-memory life-monitoring fault during X-ray command | X-ray operation | error |
E1146 | No flow in cooling circuit | Operation | warning |
E1147 | Coolant pump inactive at X-ray start | X-ray start | warning |
E1148 | Timeout post-fluoro time (LIH) | Fluoroscopy operation | warning |
E1155 | X-ray break emergency stop | Operation | warning |
E1156 | Short circuit in hand switch 1 during X-ray command | X-ray operation | warning |
E1157 | Short circuit in hand switch 2 during X-ray command | X-ray operation | warning |
E1158 | Short circuit in foot switch 1 during X-ray command | X-ray operation | warning |
E1159 | Short circuit in foot switch 2 during X-ray command | X-ray operation | warning |
E1160 | Short circuit in foot switch 3 during X-ray command | X-ray operation | warning |
E1161 | Short circuit in foot switch 4 during X-ray command | X-ray operation | warning |
E1162 | Short circuit in foot switch 5 during X-ray command | X-ray operation | warning |
E1163 | Short circuit in foot switch 6 during X-ray command | X-ray operation | warning |
E1164 | Short circuit in hand switch 1 without X-ray command | Operation | warning |
E1165 | Short circuit in hand switch 2 without X-ray command | Operation | warning |
E1166 | Short circuit in foot switch 1 without X-ray command | Operation | warning |
E1167 | Short circuit in foot switch 2 without X-ray command | Operation | warning |
E1168 | Short circuit in foot switch 3 without X-ray command | Operation | warning |
E1169 | Short circuit in foot switch 4 without X-ray command | Operation | warning |
E1170 | Short circuit in foot switch 5 without X-ray command | Operation | warning |
E1171 | Short circuit in foot switch 6 without X-ray command | Operation | warning |
E1172 | X-ray Enable without pressing a hand switch or foot switch | Operation | warning |
E1173 | CAN fault line active | Operation | warning |
E1174 | Internal communication error during operation | Operation | warning |
E1175 | Image-memory life-monitoring fault | Operation | warning |
E1176 | Heat-exchanger fan blocked | Operation | warning |
E1177 | Flow in cooling circuit too low | Operation | warning |
E1178 | No flow in cooling circuit | Operation | error |
E1179 | Coolant pump inactive at X-ray start | X-ray start | error |
E1180 | Short circuit in hand switch 9 without X-ray command | Operation | warning |
E1181 | Short circuit in hand switch 10 without X-ray command | Operation | warning |
E1182 | Short circuit in hand switch 11 without X-ray command | Operation | warning |
E1184 | Short circuit in foot switch 13 without X-ray command | Operation | warning |
E1185 | Short circuit in foot switch 14 without X-ray command | Operation | warning |
E1186 | Short circuit in foot switch 15 without X-ray command | Operation | warning |
E1187 | Short circuit in foot switch 16 without X-ray command | Operation | warning |
E1188 | Emergency Stop | Manual activation | warning |
E1189 | Low battery on wireless footswitch | Operation | warning |
E1190 | Wireless module is not responding | Operation | warning |
E1199 | Main interface error: %0 | Operation | error |
E12000 | Function assigned to X-ray switch (%0), ignored | Configuration | warning |
E12001 | X-ray command assigned to function switch (%0), ignored | Configuration | warning |
E12002 | XRaySwitch / FunctionSwitch mismatch in mappings of machinedata.ini | Configuration | warning |
E12003 | CAN modules missing: CAN Module No. | CAN module cannot be found during startup | error |
E12004 | Hard disk full. Image cannot be stored | Image storage | warning |
E12005 | Maximum image storage capacity almost reached | Image storage | info |
E12006 | vision.md in main interface missing | Configuration | warning |
E12007 | vision.md written to main interface | Configuration update | info |
E12025 | Radiation allowed after mode values are displayed | Mode change | info |
E12026 | Unable to change APr: Value rejected, restored previous state | Parameter adjustment | info |
E12027 | Unable to change APr: Value rejected, no recovery possible | Parameter adjustment | info |
E12034 | License not available | Feature access | error |
E12041 | Flat-panel detector communication fault | Operation | warning |
E12042 | Signature check completed | Signature verification | info |
E12045 | Radiation blocked by key switch | Operation | warning |
E12046 | Radiation blocked by invalid APrConfig | Operation | warning |
E12050 | Option stick defective; Vision Center does not work properly; No radiation possible | USB Dongle not found or defective | error |
E12050 | Option stick defective | Operation | warning |
E12051 | Main Interface defective (serial number) | Operation | warning |
E12052 | Serial numbers of mobile stand and monitor cart do not match | Operation | warning |
E12053 | Cannot create dose report: Not current patient | Dose report creation | warning |
E12054 | Cannot create dose report: No dose data available | Dose report creation | warning |
E12055 | Failed to load file %0 | File access | warning |
E12057 | Machinedata.ini does not match to capture device frequency | Configuration check | error |
E12060 | CAN Manager transport error | CAN communication transport failure | error |
E12061 | CanManager AckBusOff | CAN communication | error |
E12062 | CanManager AckNoTxd | CAN communication | error |
E12063 | Thin filament defective! System switches to restricted mode with thick filament | Operation | warning |
E12064 | Internal fatal error | Operation | error |
E12070 | CanMaster failed | CAN communication | error |
E12071 | CAN module CameraIrisProxy failed | CAN communication | error |
E12072 | CAN module Collimator failed | CAN communication | warning |
E12073 | CAN module Generator failed | CAN communication | error |
E12074 | CAN module MainInterface failed | CAN communication | error |
E12075 | CAN module Regulation failed | CAN communication | error |
E12076 | CAN module MainPrePowerSupply failed | CAN communication | error |
E12077 | CAN module MainPrePowerSupply1 failed | CAN communication | error |
E12078 | CAN module MainPrePowerSupply2 failed | CAN communication | error |
E12079 | CAN module ELWMS failed | CAN communication | error |
E12080 | CAN module Kinematics failed | Kinematics module communication failure | error |
E12080 | Unknown error in CanMaster. Error state: %0 | CAN communication | error |
E12082 | Software update script wasn't found | Software update | error |
E12083 | Software update script couldn't be started | Software update | error |
E12084 | Software update script failed with error %0 | Software update | error |
E12096 | Flat-panel detector communication fault: %0 | Operation | warning |
E12098 | Cannot overwrite file '%s'. Access denied | File operation | warning |
E1246 | Flat-panel detector communication fault | Ethernet cable disconnected or not connected between Image Memory and Command Processor during startup or VIVA tool operation | critical |
E1300 | Data transfer to DAC failure (SPI interface) | Startup hardware check | error |
E1300 | PPS power-up fault | Power-up | error |
E1301 | PPS overtemperature fault during power-up | Power-up | warning |
E1301 | Temperature sensor check failure | Startup hardware check | error |
E1302 | PPS power-up fault | During power-up | error |
E1302 | Reference voltage DAC (MAX504) check failure | Startup hardware check | error |
E1303 | PPS power-up fault | During power-up | error |
E1303 | Output voltage check failure from DAC (MAX504) | Startup hardware check | error |
E1304 | Wrong EEPROM checksum during power-up | During power-up | error |
E1350 | SPI interface data transfer error (runtime) | Every nominal value change during runtime | error |
E1350 | PPS power-up fault | During power-up | error |
E1351 | PPS overtemperature fault during operation | During system operation | warning |
E1351 | Temperature error (runtime) | Permanent runtime temperature monitoring | warning |
E1355 | CAN fault detection | Permanent runtime CAN line monitoring | error |
E1358 | Output voltage beyond tolerance | During system operation | warning |
E1398 | PPS power-up fault | During power-up | error |
E1399 | PPS fault | During operation | error |
E1400 | Servo Rotation position fault | During operation | warning |
E1401 | Servo Rotation not connected | During operation | warning |
E1402 | Servo Rotation not ready | During operation | warning |
E1403 | Servo Iris position fault | During operation | warning |
E1404 | Servo Iris not connected | During operation | warning |
E1405 | Servo Iris not ready | During operation | warning |
E1406 | Servo Slot 1 position fault | During operation | warning |
E1407 | Servo Slot 1 not connected | During operation | warning |
E1408 | Servo Slot 1 not ready | During operation | warning |
E1409 | Servo Slot 2 position fault | During operation | warning |
E1410 | Servo Slot 2 not connected | During operation | warning |
E1411 | Servo Slot 2 not ready | During operation | warning |
E1412 | Servo Corner position fault | During operation | warning |
E1413 | Servo Corner not connected | During operation | warning |
E1414 | Servo Corner not ready | During operation | warning |
E1420 | DAP meter mismatch or not available | During operation | warning |
E1421 | DAP meter error | During operation | warning |
E1448 | Collimator startup error | During power-up or collimator initialization | warning |
E1449 | Vision Track unknown error | During Vision Track operation | warning |
E1450 | Servo position fault | During operation | warning |
E1451 | Servo not connected | During operation | warning |
E1452 | Servo not ready | During operation | warning |
E1499 | I.I. power supply unknown error | During operation | warning |
E1700 | kVmA table: kV not ascending | During table validation | warning |
E1701 | kVmA table: power not ascending | During table validation | warning |
E1702 | kVmA table not finished | During table validation | warning |
E1705 | No external video signal | During operation | warning |
E1708 | Generator pulse error | During operation | warning |
E1709 | No VSync | During operation | warning |
E1710 | No Interrupt | During operation | warning |
E1711 | No communication PPC-generator | During operation | warning |
E1712 | Wrong mA value from generator | During operation | warning |
E1713 | X-ray not allowed | During operation when X-ray should not emit | warning |
E1714 | Max. mA value cannot be read | During operation | warning |
E1716 | Image-memory power-up fault | During power-up | error |
E1717 | Image-memory power-up fault | During power-up | error |
E1718 | Image-memory power-up fault | During power-up | warning |
E1719 | Gain adjustment not possible | During power-up or calibration | error |
E1720 | Image-memory fault | During operation | error |
E1721 | Image-memory fault | During operation | error |
E1722 | V-Sync corrupt | At least one V-sync pulse missing at image memory input | error |
E1723 | Video standard unknown | During operation | error |
E1724 | Image-memory fault | During power-up | error |
E1725 | M-Data unknown | During operation | warning |
E1726 | Collimator does not support DAP values | During operation | warning |
E1727 | Flat-panel detector not supported by this version | During operation | warning |
E1728 | Internal fatal communication error | During operation | warning |
E1729 | Internal communication error | During operation | error |
E1732 | Image-memory power-up fault | During power-up | error |
E1733 | Image-memory power-up fault | During power-up | error |
E1734 | Image-memory power-up fault | During power-up | error |
E1735 | Image-memory power-up fault | During power-up | error |
E1736 | Image-memory power-up fault | During power-up | error |
E1737 | Image-memory power-up fault | During power-up | error |
E1738 | No external video signal | No video signal present at image memory input | error |
E1740 | Hardware doesn't support grid regulation | During operation | warning |
E1741 | Hardware doesn't support motion detection | During operation | warning |
E1742 | Hardware doesn't support DAP calculation | During operation | warning |
E1743 | Hardware doesn't support automatic dose reduction | During operation | warning |
E1747 | Black level >= White level | During calibration or operation | error |
E1756 | FFC board not available | During power-up or operation | error |
E1757 | FFC not working properly | During operation | error |
E1759 | Generator version and PPC version not compatible | During power-up | error |
E1760 | High resolution with pulse length 0 not allowed | During operation with high resolution mode | error |
E1789 | PPC unknown error | During operation | error |
E1790 | PPC CAN Exception error | During operation | warning |
E1800 | Radiation indication lamp defective during power-up | During power-up | warning |
E1801 | Wrong EEPROM checksum during power-up | During power-up | warning |
E1802 | Image-memory power-up fault | During power-up | warning |
E1803 | Image-memory power-up fault | During power-up | warning |
E1804 | No communication with left flatscreen | During power-up or operation | warning |
E1805 | No communication with right flatscreen | During power-up or operation | warning |
E1806 | Image-memory power-up fault | During power-up | warning |
E1807 | Image-memory power-up fault | During power-up | warning |
E1808 | Switch on printer | During operation | warning |
E1850 | Radiation indication lamp defective during operation | During system operation | warning |
E1865 | U578 execution failed | During U578 operation | warning |
E1898 | U578 Startup Error | During power-up | warning |
E1899 | U578 Error | During operation | warning |
Setting the network configuration failed | Network configuration cannot be applied to system | During network settings configuration | error |
Verification failed OM[5]: Association rejected: Wrong serve | DICOM AE title mismatch prevents server connection | Test button pressed during DICOM server configuration | error |
Verification failed OM[6]: Association request failed | DICOM server connection cannot be established | Test button pressed during DICOM server configuration | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.