Welch Allyn Vital Signs Monitor 300: troubleshooting and error codes
The ServiceForge knowledge base holds 67 documented symptoms for the Welch Allyn Vital Signs Monitor 300, with 84 ranked causes, 91 diagnostic checks with test points and expected values, and 26 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: SpO2 monitor function (Nellcor)
Verify SpO2 pulse rate and saturation percentage display accuracy using Nellcor SRC-MAX functional tester
Ranked causes
- 1. Functional verification of SpO2 monitoring circuit (unknown)
- 2. Functional verification of SpO2 monitoring circuit (Masimo) (unknown)
- 3. Functional verification of SpO2 sensor with monitor (unknown)
- 4. Functional verification of temperature monitor function (unknown)
- 5. Functional verification of temperature probe with monitor (unknown)
- 6. Functional verification of Nurse Call relay (unknown)
- 7. Functional verification of battery charging and voltage (unknown)
- 8. Functional verification of patient isolation (unknown)
Verify all monitor LEDs flash
- Test point
- Left panel, center panel, right panel LEDs
- Expected
- All LEDs flash continuously
23 more checks with pass and fail actions, 23 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP air leak exceeds limits patient safety
NIBP (non-invasive blood pressure) module leaks air
Ranked causes
- 1. Leaky check valve (common)
- 2. Leaky tubing or connector (common)
- 3. Leaky pump (occasional)
- 4. Leaky bleed or dump valve (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP leak test patient safety
Verify NIBP system for abnormal leakage using neonatal cuff at specified inflation pressures
Ranked causes
- 1. Leak in NIBP cuff (occasional)
- 2. Leak in NIBP hose or connections (occasional)
- 3. Leak in internal NIBP module (rare)
Put monitor in Service mode
- Test point
- Power button and control buttons
- Expected
- Monitor in SERVICE MODE
7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Repeated C07 error code patient safety
SpO2 module repeatedly reports error C07
Ranked causes
- 1. Incompatible SpO2 sensor (common)
- 2. Faulty SpO2 sensor (common)
- 4. Faulty SpO2 board (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Repeated C20 temperature error code patient safety
Temperature module repeatedly reports error C20
Ranked causes
- 1. Probe well not installed (common)
- 2. Faulty probe (common)
- 3. Faulty temperature board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Repeated C04 error code patient safety
NIBP module repeatedly reports error C04
Ranked causes
- 1. NIBP internal tubing disconnected (common)
- 2. Faulty NIBP module (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery run time is low
Device battery does not provide adequate operating duration
Ranked causes
- 1. Defective battery (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Button requires excessive pressure
Buttons on right side require noticeably more force to actuate
Ranked causes
- 1. Display board mounting screws loose (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery charging and voltage
Verify battery charging, runtime capability, and voltage
Ranked causes
- 1. Functional verification of battery charging and voltage (unknown)
Verify AC mains indicator illuminated
- Test point
- Monitor front panel AC mains indicator
- Expected
- Flashing (charging) or steady (90% charged)
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Blood pressure cuff pressure accuracy patient safety
Verify that SYS display matches digital pressure meter readings at 80, 150, and 300 mmHg target pressures during functional verification
Ranked causes
- 1. Pressure transducer or software calibration drift (unknown)
Verify SYS display at 80 mmHg target pressure
- Test point
- SYS window, digital pressure meter
- Expected
- Within 3 mmHg (0.4 kPa) of meter reading, 3 mmHg (0.4 kPa)
- Tools
- Digital pressure meter
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Blood pressure sensor accuracy at 150 mmHg
Verify systolic pressure reading accuracy at 150 mmHg (20 kPa) target against digital pressure meter
Ranked causes
- 1. Sensor calibration drift (unknown)
Compare SYS display reading to digital pressure meter at 150 mmHg
- Test point
- SYS window and digital pressure meter
- Expected
- Within 3 mmHg (0.4 kPa) of meter reading, ±3 mmHg (±0.4 kPa)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Blood pressure sensor accuracy at 300 mmHg
Verify systolic pressure reading accuracy at 300 mmHg (40 kPa) target against digital pressure meter
Ranked causes
- 1. Sensor calibration drift (unknown)
Compare SYS display reading to digital pressure meter at 300 mmHg
- Test point
- SYS window and digital pressure meter
- Expected
- Within 6 mmHg (0.8 kPa) of meter reading, ±6 mmHg (±0.8 kPa)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Blood pressure sensor accuracy at 80 mmHg
Verify systolic pressure reading accuracy at 80 mmHg (10.7 kPa) target against digital pressure meter
Ranked causes
- 1. Sensor calibration drift (unknown)
Compare SYS display reading to digital pressure meter at 80 mmHg
- Test point
- SYS window and digital pressure meter
- Expected
- Within 3 mmHg (0.4 kPa) of meter reading, ±3 mmHg (±0.4 kPa)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Cuff deflation rate
Verify deflation valve opens and pressure drops at specified rate from ~300 mmHg to <15 mmHg in 10 seconds
Ranked causes
- 1. Deflation valve stuck or blocked (unknown)
Measure pressure drop over 10 seconds after selecting 0 mmHg
- Test point
- Manometer reading at 10-second mark
- Expected
- Less than 15 mmHg remaining, <15 mmHg
- Tools
- Manometer, timer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Cuff inflation time from 5 to 250 mmHg
Verify pump inflation rate meets timing requirement when pressurizing from 5 mmHg to 250 mmHg
Ranked causes
- 1. Pump speed degradation or mechanical restriction (unknown)
Time cuff inflation from 5 mmHg to 250 mmHg using manometer
- Test point
- Manometer and timer
- Expected
- Less than 8 seconds, <8 seconds
- Tools
- Timer, manometer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Maximum allowable adult cuff pressure patient safety
Verify cuff cannot exceed 280 mmHg in adult mode even when manually pressurized beyond target
Ranked causes
- 1. Overpressure safety circuit malfunction (unknown)
Attempt to pressurize cuff above 280 mmHg in ADULT mode
- Test point
- Cuff pressure (manometer or SYS display)
- Expected
- Pressure does not exceed 280 mmHg; dump valve opens; C10 displayed, ≤280 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Maximum allowable cuff pressure (Adult mode) patient safety
Verify overpressure protection limits adult cuff pressure to 280 mmHg maximum before dump valve opens
Ranked causes
- 1. Software overpressure limit not enforced, or dump valve failure (unknown)
Pressure at which dump valve opens in Adult mode
- Test point
- Manometer, SYS window, visual observation of dump valve action
- Expected
- Dump valve opens at 280 mmHg; pressure drops to approximately 0; C10 error displays, 280 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Maximum allowable cuff pressure (Neonate mode) patient safety
Verify overpressure protection limits neonate cuff pressure to 141 mmHg maximum before dump valve opens
Ranked causes
- 1. Software overpressure limit not enforced for neonate mode, or dump valve failure (unknown)
Pressure at which dump valve opens in Neonate mode
- Test point
- Manometer, SYS window, visual observation of dump valve action
- Expected
- Dump valve opens at 141 mmHg; pressure drops to approximately 0; C10 error displays, 141 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Maximum allowable cuff pressure (Pediatric mode) patient safety
Verify overpressure protection limits pediatric cuff pressure to 200 mmHg maximum before dump valve opens
Ranked causes
- 1. Software overpressure limit not enforced for pediatric mode, or dump valve failure (unknown)
Pressure at which dump valve opens in Pediatric mode
- Test point
- Manometer, SYS window, visual observation of dump valve action
- Expected
- Dump valve opens at 200 mmHg; pressure drops to approximately 0; C10 error displays, 200 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Maximum allowable neonate cuff pressure patient safety
Verify cuff cannot exceed 141 mmHg in neonate mode even when manually pressurized beyond target
Ranked causes
- 1. Overpressure safety circuit malfunction (unknown)
Attempt to pressurize neonate cuff above 141 mmHg in NEONATE mode
- Test point
- Cuff pressure (manometer or SYS display)
- Expected
- Pressure does not exceed 141 mmHg; dump valve opens; C10 displayed, ≤141 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Maximum allowable pediatric cuff pressure patient safety
Verify cuff cannot exceed 200 mmHg in pediatric mode even when manually pressurized beyond target
Ranked causes
- 1. Overpressure safety circuit malfunction (unknown)
Attempt to pressurize cuff above 200 mmHg in PEDIATRIC mode
- Test point
- Cuff pressure (manometer or SYS display)
- Expected
- Pressure does not exceed 200 mmHg; dump valve opens; C10 displayed, ≤200 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP characterization test patient safety
Verify NIBP module error code C03 appears during characterization using 35 mL syringe test volume
Ranked causes
- 1. NIBP module not properly characterized or calibrated (common)
Attach neonate hose (part 008-0265-01) to NIBP fitting
- Test point
- NIBP fitting on monitor
- Expected
- Neonatal hose securely attached
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP pump current
Measure pump current draw during inflation to verify pump operation and detect motor or drive circuit faults
Ranked causes
- 1. Pump motor fault or excessive current draw (unknown)
Measure pump current rise during 80 mmHg inflation
- Test point
- Current meter, monitoring pump startup
- Expected
- Not more than 750 mA above baseline
- Tools
- Current meter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP pump deflation rate patient safety
Verify cuff deflates from approximately 300 mmHg to less than 15 mmHg within 10 seconds to ensure controlled pressure release
Ranked causes
- 1. Bleed valve partially stuck, or blockage in deflation path (unknown)
Measure deflation pressure after 10 seconds from ~300 mmHg
- Test point
- Manometer, timer, SYS window
- Expected
- Less than 15 mmHg, < 15 mmHg
- Tools
- Timer, manometer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP pump inflation rate patient safety
Verify pump inflates cuff from 5 mmHg to 250 mmHg within 8 seconds to ensure adequate inflation speed
Ranked causes
- 1. Pump worn, leak in tubing/cuff connection, or pressure relief valve stuck open (unknown)
Measure inflation time from 5 to 250 mmHg
- Test point
- Manometer, timer
- Expected
- Less than 8 seconds, < 8 sec
- Tools
- Timer, manometer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Nurse Call relay
Verify Nurse Call relay contact resistance in alarm and non-alarm states
Ranked causes
- 1. Functional verification of Nurse Call relay (unknown)
Measure Nurse Call relay resistance pins 1-2 (alarm OFF)
- Test point
- Nurse Call connector pins 1-2
- Expected
- > 1 MΩ
- Tools
- Ohmmeter
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Overpressure cutoff emergency safety (NIBP 4.20.00+) patient safety
Verify emergency overpressure cutoff stops pump and opens dump/bleed valves before pressure exceeds 280 mmHg in Adult mode
Ranked causes
- 1. Overpressure cutoff circuit or software failure (NIBP 4.20.00+) (unknown)
Verify overpressure cutoff maximum pressure during rapid inflation
- Test point
- Digital pressure meter, SYS window, E40 error display
- Expected
- Pressure does not exceed 280 mmHg; E40 error displays; any button press shuts down monitor, ≤ 280 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Overpressure cutoff emergency shutdown patient safety
Verify overpressure cutoff circuit prevents pressure exceeding 280 mmHg and triggers E40 error when manual bulb squeeze forces rapid pressure rise
Ranked causes
- 1. Overpressure cutoff circuit malfunction (unknown)
Rapid pressurization triggers overpressure cutoff
- Test point
- Digital pressure meter and SYS display
- Expected
- Pressure ≤280 mmHg; E40 error; dump/bleed valves open, ≤280 mmHg
- Tools
- Digital pressure meter, monitor display
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Patient isolation high voltage
Verify patient isolation using AC Withstand Voltage hi-pot tester on all external connectors
Ranked causes
- 1. Functional verification of patient isolation (unknown)
Test Nurse Call isolation current
- Test point
- Hi-pot tester current measurement
- Expected
- ≤ 2.50 mA (shielding sufficient)
- Tools
- AR 3605 or equivalent hi-pot tester
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Printer output and current draw
Verify printer produces legible settings report with no anomalies, missing, or faded sections; measure current draw during printing
Ranked causes
- 1. Printer paper depleted, printer motor or drive circuit fault, or excessive current draw (unknown)
Print settings report and inspect output quality
- Test point
- Printed report, visual inspection
- Expected
- Report completes with no printed anomalies, missing sections, or faded text
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pump current draw during 80 mmHg inflation
Verify pump motor current remains within specified limit during cuff inflation to 80 mmHg from 0 mmHg baseline
Ranked causes
- 1. Pump motor malfunction or excessive load (unknown)
Measure highest current during inflation from 0 to 80 mmHg
- Test point
- Current meter during pump operation
- Expected
- Not more than 750 mA above baseline current, Maximum 750 mA above baseline
- Tools
- Current meter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SpO2 monitor function (Masimo)
Verify SpO2 pulse rate and saturation percentage display accuracy using Masimo Tester REF 1593
Ranked causes
- 1. Functional verification of SpO2 monitoring circuit (Masimo) (unknown)
Measure pulse rate and SpO2 with Masimo tester
- Test point
- Monitor display
- Expected
- Pulse rate 61 ± 1 bpm, saturation 81 ± 3%, ± 1 bpm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SpO2 sensor with monitor
Verify SpO2 sensor and monitor function using BioTek Index2 simulator
Ranked causes
- 1. Functional verification of SpO2 sensor with monitor (unknown)
Measure SpO2 with BioTek simulator
- Test point
- Monitor display
- Expected
- SpO2 saturation 94 ± 4%, ± 4 %
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Temperature monitor function
Verify temperature display accuracy using Welch Allyn Sure Temp Plus test key
Ranked causes
- 1. Functional verification of temperature monitor function (unknown)
Measure temperature with test key
- Test point
- Monitor display
- Expected
- Temperature 97.3 ± 0.2 °F (36.3 ± 0.1 °C), ± 0.2 °F
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Temperature probe with monitor
Verify temperature probe accuracy using Welch Allyn Calibration Tester
Ranked causes
- 1. Functional verification of temperature probe with monitor (unknown)
Measure temperature at 96.4 °F setpoint (9600 model)
- Test point
- Monitor display
- Expected
- 96.4 ± 0.4 °F (35.8 ± 0.2 °C), ± 0.4 °F
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Erroneous LCD characters
LCD display showing corrupted or incorrect characters
Ranked causes
- 1. LCD partially connected to main board (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code E30
PROM memory corruption
Ranked causes
- 1. PROM memory corrupted (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code E31
RAM memory test failure
Ranked causes
- 1. RAM test failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code E32
Factory data failure
Ranked causes
- 1. Factory data corrupted (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code E33
User data error
Ranked causes
- 1. User settings corrupted (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code E34
A/D converter failure
Ranked causes
- 1. A/D converter malfunction (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code E35 patient safety
SpO2 module failure
Ranked causes
- 1. SpO2 module malfunction (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code E36
Temperature probe or module failure
Ranked causes
- 1. Faulty temperature probe (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code E37
Printer module failure
Ranked causes
- 1. Printer module malfunction (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code E38
Real-time clock failure
Ranked causes
- 1. Real-time clock malfunction (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code E40 patient safety
NIBP module failure
Ranked causes
- 1. NIBP module malfunction (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code E41 patient safety
NIBP module failure
Ranked causes
- 1. NIBP module malfunction (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code E44
POST queue overflow
Ranked causes
- 1. POST queue overflow (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front panel replacement required
Monitor requires front panel replacement
Ranked causes
- 1. Front panel failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Main board replacement required
Monitor requires main board replacement
Ranked causes
- 1. Main board failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing LED display
LED display on display board not illuminating
Ranked causes
- 1. LED failure on display board (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor configuration incompatible after NIBP module replacement patient safety
Monitor unable to recognize or function with new NIBP (POEM) module until service utility updates monitor configuration
Ranked causes
- 1. NIBP module firmware version not registered in monitor configuration (common)
Verify NIBP module installed has different firmware version than monitor expects
- Test point
- Service utility main screen, NIBP Module row under 'Installed Ver' and 'Update Ver'
- Expected
- Version numbers shown for old POEM and new POEM2 NIBP module differ
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor displays E35 error after temperature module replacement patient safety
E35 error appears on display when a new temperature module with different firmware version is installed without configuration update
Ranked causes
- 1. Temperature module firmware version not accepted by monitor (common)
Verify if new temperature module firmware version differs from previously installed version
- Test point
- Service utility Update Monitor screen, Installed Version column
- Expected
- New firmware version shown in orange if different from last configured version
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not power up
Monitor fails to turn on or show display
Ranked causes
- 2. Wrong power adapter used (common)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP air leak exceeds specified limits patient safety
Non-invasive blood pressure module has excessive air leakage
Ranked causes
- 1. Leaky check valve (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP cuff overinflation after component replacement patient safety
NIBP module not properly characterized after replacing pump, air filter, check valve, tubing, or NIBP board; cuff could overinflate posing hazard to neonatal patients
Ranked causes
- 1. NIBP component replaced without characterization (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP cuff overinflation hazard in neonatal patients patient safety
Risk of NIBP cuff overinflation causing injury to neonatal patients if NIBP module is not properly characterized after component replacement
Ranked causes
- 1. NIBP module not characterized after component replacement (common)
Verify NIBP characterization was performed
- Test point
- Service Utility logs or monitor status display
- Expected
- Characterization completion record present and current
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP module characterization required patient safety
NIBP module must be characterized after component replacement to ensure proper cuff inflation and patient safety
Ranked causes
- 1. NIBP component replaced (pump, air filter, check valve, tubing, or NIBP board) (common)
Verify NIBP characterization completed before returning monitor to service
- Test point
- NIBP module
- Expected
- Characterization procedure successfully completed using Service Utility
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP module replacement required patient safety
Monitor requires NIBP module replacement
Ranked causes
- 1. NIBP module failure (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP module requires characterization after component replacement patient safety
NIBP characterization procedure must be performed after replacing pump, air filter, check valve, tubing, or NIBP board to ensure proper cuff inflation and prevent hazard to neonatal patients
Ranked causes
- 1. NIBP component replaced (pump, air filter, check valve, tubing, or NIBP board) (common)
NIBP characterization completed using Monitor Service Utility
- Test point
- Monitor Service Utility
- Expected
- Characterization procedure successfully completed without errors
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Printer replacement required
- Repeated C03 error code
- Software update required
- SpO2 board replacement required
- SpO2 side panel replacement required
- Temperature board replacement required
- Warning: Do not connect to patient in Service mode
7 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
C03 | NIBP internal tubing is kinked or blocked | Air cannot flow through kinked tubing during measurement | warning |
C03 | NIBP module characterization successful; cuff overpressure condition detected or test sequence complete | Expected to appear during NIBP characterization test when cuff inflation reaches specified volume threshold | info |
C03 | NIBP characterization complete (success indicator) | NIBP module successfully characterizes against fixed volume test chamber | info |
C04 | Alarm condition indicator during Nurse Call relay verification test | Monitor enters alarm state after pump charges twice (approximately 10 seconds) during functional verification | info |
C04 | NIBP internal tubing is disconnected or faulty NIBP module | NIBP tubing disconnected from module or module failure | error |
C07 | SpO2 sensor incompatible, faulty, improperly connected, or faulty SpO2 board | SpO2 sensor not detected or communication failure | error |
C10 | Overpressure condition detected and cuff pressure dumped | Cuff pressure exceeds safe limits for selected mode (280 mmHg adult, 200 mmHg pediatric, 141 mmHg neonate) | warning |
C10 | Overpressure event occurred | Cuff pressure exceeded maximum allowable limit for patient mode (280 mmHg adult, 200 mmHg pediatric, 141 mmHg neonate) | warning |
C20 | Temperature probe well not installed, faulty probe, or faulty temperature board | Temperature measurement subsystem failure | error |
E30 | PROM is corrupt | PROM memory corruption detected | critical |
E31 | RAM test failure | RAM memory test failed | error |
E32 | Factory data failure | Factory calibration data corrupted | critical |
E33 | User data error | User settings data corrupted | warning |
E34 | A/D converter failure | Analog to digital converter malfunction | error |
E35 | Temperature module fault or incompatibility | New temperature module with different firmware version installed and monitor configuration not updated to accept it | error |
E35 | SpO2 module failure | SpO2 module communication or hardware failure | error |
E36 | Temperature module failure | Temperature module communication or hardware failure | error |
E37 | Printer failure | Printer module communication or hardware failure | error |
E38 | Real-time clock failure | Clock battery depleted or RTC malfunction | warning |
E38 | Power-on self-test error or initialization error | Displayed during or after power-on self-test sequence | error |
E38 | Temperature probe error or disconnection detected | When temperature probe is removed from probe well during initialization or when system detects probe fault | warning |
E40 | Emergency overpressure cutoff triggered; pump stopped and valves opened | Pressure rise rate exceeds cutoff threshold (rapid manual pressurization or pump malfunction) | critical |
E40 | Overpressure cutoff activated (NIBP 4.20.00+) | Pump pressure rise exceeds cutoff threshold (295-330 mmHg) during inflation; emergency safety feature activation | error |
E40 | NIBP module failure | NIBP module communication or hardware failure | error |
E41 | NIBP module failure | NIBP module communication or hardware failure | error |
E44 | POST queue full | Power-on self-test queue overflow | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.