Welch Allyn SureTemp Plus: troubleshooting and error codes

The ServiceForge knowledge base holds 12 documented symptoms for the Welch Allyn SureTemp Plus, with 14 ranked causes, 18 diagnostic checks with test points and expected values, and 26 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Patient temperature out of range patient safety

Display shows body outline with blinking up/down arrow and thermometer; measured patient temp outside limits

Ranked causes

  1. 1. Probe not properly positioned on patient (common)
  2. 2. Actual patient temperature outside normal limits (occasional)
  3. 3. Probe defect (occasional)
First check

Visually verify probe placement on patient

Test point
Probe tip contact point
Expected
Firm direct contact with patient skin/tissue

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ambient temperature out of range

Display shows 'A' with blinking up/down arrow and thermometer; device measures out-of-range ambient

Ranked causes

  1. 2. Probe malfunction at high temperature (occasional)
First check

Verify ambient temperature

Test point
Device location
Expected
50°F to 110°F (10°C to 43.3°C), 50–110°F / 10–43.3°C

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ambient temperature too low (Av!)

Display shows 'A' with downward arrow and thermometer; ambient below minimum

Ranked causes

  1. 1. Ambient temperature below minimum (occasional)
First check

Verify ambient temperature reading

Test point
Device environment
Expected
≥50°F (≥10°C), 50–110°F / 10–43.3°C

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery above operational range

Display shows blinking battery frame, 'b' symbol, and upward arrow when battery voltage exceeds 5.375V.

Ranked causes

  1. 1. Battery voltage high (occasional)
First check

Check battery voltage

Test point
Battery
Expected
≤ 5.375 V

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery below operational range

Display shows blinking battery frame, 'b' symbol, and downward arrow when battery voltage drops below 3.00V.

Ranked causes

  1. 1. Battery voltage low (common)
First check

Check battery voltage

Test point
Battery
Expected
≥ 3.00 V

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery low warning

Single blinking battery segment in indicator; battery approaching minimum operational voltage

Ranked causes

  1. 1. Battery approaching depletion (common)
First check

Check battery indicator segments

Test point
Battery display frame
Expected
Full segments or single blinking segment (low warning)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery voltage above operational range

Display shows blinking battery frame with 'b' and upward arrow; device may malfunction

Ranked causes

  1. 1. Battery voltage too high (b^) (occasional)
First check

Verify battery voltage output

Test point
Battery terminals
Expected
< 5.50 volts V, < 5.50 V

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery voltage below operational range

Display shows blinking battery frame with 'b' and downward arrow; device may not function

Ranked causes

  1. 1. Battery voltage too low (bv) (common)
First check

Verify battery voltage reading on display

Test point
Battery indicator frame
Expected
≥3.1 volts (and <5.50 volts) V, >= 3.1 V and < 5.5 V

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Stored thermometer reactivation

Twelve-month maintenance for thermometers in extended storage before returning to service

Ranked causes

  1. 1. Routine twelve-month stored unit maintenance interval (common)
First check

Remove batteries

Test point
battery compartment
Expected
batteries removed and stored separately

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: SureTemp Plus thermometer

Six-month preventive maintenance calibration testing per Power-On Self Test and Model 9600/9600 Plus Calibration Testing procedure

Ranked causes

  1. 1. Routine six-month calibration interval (common)
First check

Visual inspection for physical damage

Test point
entire thermometer exterior
Expected
no cracks, dents, or damage

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Instrument error with error code displayed

Display shows instrument symbol with X and error code in temperature field; internal circuitry failure

Ranked causes

  1. 1. Internal instrument circuit failure (rare)
First check

Note instrument error code from display

Test point
Temperature display field
Expected
Error code (E4.x or E5.x series typically)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Probe error with error code displayed

Display shows probe symbol with X and error code in temperature field; probe or connector failure detected

Ranked causes

  1. 1. Broken or malfunctioning probe (common)
First check

Note probe error code displayed

Test point
Temperature display field
Expected
Code specific to probe error type (E0.x series)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
23Host interface errorCommunication error between thermometer and external host/computer during data transferwarning
26Invalid SureTemp modeDevice boots in unsupported or corrupted mode configurationerror
29Battery voltage is not setDevice cannot read or store current battery voltage during initializationerror
30Prediction mode is not setDevice fails to initialize predict mode configuration during startuperror
71Device EEPROM acquire failureDevice cannot acquire (lock) EEPROM access during boot or operationerror
72Device EEPROM release failureDevice cannot release (unlock) EEPROM after access during boot or operationerror
73Device EEPROM invalid pointer failureMemory pointer used for EEPROM access is corrupted or out of boundserror
E0.1Excessive Heater Energy during predict cycleHeater draws excessive current or applies too much energy to probeerror
E0.2A/D measurement out of rangeAnalog-to-digital converter receives voltage outside valid measurement rangeerror
E0.4Probe over temperature (≥110°F / 43.3°C)Probe tip temperature exceeds maximum safe thresholdwarning
E0.5No probe temperature rise (heater has no effect)Heater powered but probe temperature does not increase during predict cycleerror
E0.6Probe EEPROM read failure (bad handshake)Device cannot establish communication handshake with probe EEPROM during startup or operationerror
E0.7Probe EEPROM uninitializedProbe EEPROM found but not initialized (missing calibration data)error
E0.8EEPROM CRC failure (data corrupted)Cyclic redundancy check fails on probe or device EEPROM dataerror
E0.9Bad probe calibrationProbe EEPROM contains invalid or out-of-tolerance calibration dataerror
E4.0PTB measurement out of rangePatient Temperature Baseline measurement exceeds expected range during startup self-testerror
E4.1RCAL measurement out of rangeReference calibration measurement during self-test outside expected boundserror
E4.2Reference power supply malfunctionInternal reference voltage supply voltage out of toleranceerror
E4.3Device EEPROM save failureDevice cannot write calibration or configuration data to internal EEPROMerror
E4.4Device EEPROM starting handshake failureDevice cannot establish initial communication with internal EEPROM during booterror
E4.5Device EEPROM response failureEEPROM does not respond to read/write command within timeouterror
E4.6Transmit failure to device EEPROMDevice data transmission to EEPROM fails or is corruptederror
E4.7Cannot initialize device EEPROMDevice boot sequence fails during EEPROM initializationcritical
E4.8Device was not calibratedBoot detects missing or invalid factory calibration data in EEPROMerror
E5.0Heater not working properlyInternal heater fails to heat or maintain temperature during predict cycleerror
E5.2Heater fail-safe failureHeater safety shutdown mechanism malfunction; over-temperature protection triggeredcritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.