Welch Allyn SureTemp Plus: troubleshooting and error codes
The ServiceForge knowledge base holds 12 documented symptoms for the Welch Allyn SureTemp Plus, with 14 ranked causes, 18 diagnostic checks with test points and expected values, and 26 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Patient temperature out of range patient safety
Display shows body outline with blinking up/down arrow and thermometer; measured patient temp outside limits
Ranked causes
- 1. Probe not properly positioned on patient (common)
- 2. Actual patient temperature outside normal limits (occasional)
- 3. Probe defect (occasional)
Visually verify probe placement on patient
- Test point
- Probe tip contact point
- Expected
- Firm direct contact with patient skin/tissue
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ambient temperature out of range
Display shows 'A' with blinking up/down arrow and thermometer; device measures out-of-range ambient
Ranked causes
- 2. Probe malfunction at high temperature (occasional)
Verify ambient temperature
- Test point
- Device location
- Expected
- 50°F to 110°F (10°C to 43.3°C), 50–110°F / 10–43.3°C
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ambient temperature too low (Av!)
Display shows 'A' with downward arrow and thermometer; ambient below minimum
Ranked causes
- 1. Ambient temperature below minimum (occasional)
Verify ambient temperature reading
- Test point
- Device environment
- Expected
- ≥50°F (≥10°C), 50–110°F / 10–43.3°C
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery above operational range
Display shows blinking battery frame, 'b' symbol, and upward arrow when battery voltage exceeds 5.375V.
Ranked causes
- 1. Battery voltage high (occasional)
Check battery voltage
- Test point
- Battery
- Expected
- ≤ 5.375 V
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery below operational range
Display shows blinking battery frame, 'b' symbol, and downward arrow when battery voltage drops below 3.00V.
Ranked causes
- 1. Battery voltage low (common)
Check battery voltage
- Test point
- Battery
- Expected
- ≥ 3.00 V
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery low warning
Single blinking battery segment in indicator; battery approaching minimum operational voltage
Ranked causes
- 1. Battery approaching depletion (common)
Check battery indicator segments
- Test point
- Battery display frame
- Expected
- Full segments or single blinking segment (low warning)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage above operational range
Display shows blinking battery frame with 'b' and upward arrow; device may malfunction
Ranked causes
- 1. Battery voltage too high (b^) (occasional)
Verify battery voltage output
- Test point
- Battery terminals
- Expected
- < 5.50 volts V, < 5.50 V
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage below operational range
Display shows blinking battery frame with 'b' and downward arrow; device may not function
Ranked causes
- 1. Battery voltage too low (bv) (common)
Verify battery voltage reading on display
- Test point
- Battery indicator frame
- Expected
- ≥3.1 volts (and <5.50 volts) V, >= 3.1 V and < 5.5 V
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Stored thermometer reactivation
Twelve-month maintenance for thermometers in extended storage before returning to service
Ranked causes
- 1. Routine twelve-month stored unit maintenance interval (common)
Remove batteries
- Test point
- battery compartment
- Expected
- batteries removed and stored separately
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SureTemp Plus thermometer
Six-month preventive maintenance calibration testing per Power-On Self Test and Model 9600/9600 Plus Calibration Testing procedure
Ranked causes
- 1. Routine six-month calibration interval (common)
Visual inspection for physical damage
- Test point
- entire thermometer exterior
- Expected
- no cracks, dents, or damage
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Instrument error with error code displayed
Display shows instrument symbol with X and error code in temperature field; internal circuitry failure
Ranked causes
- 1. Internal instrument circuit failure (rare)
Note instrument error code from display
- Test point
- Temperature display field
- Expected
- Error code (E4.x or E5.x series typically)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe error with error code displayed
Display shows probe symbol with X and error code in temperature field; probe or connector failure detected
Ranked causes
- 1. Broken or malfunctioning probe (common)
Note probe error code displayed
- Test point
- Temperature display field
- Expected
- Code specific to probe error type (E0.x series)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
23 | Host interface error | Communication error between thermometer and external host/computer during data transfer | warning |
26 | Invalid SureTemp mode | Device boots in unsupported or corrupted mode configuration | error |
29 | Battery voltage is not set | Device cannot read or store current battery voltage during initialization | error |
30 | Prediction mode is not set | Device fails to initialize predict mode configuration during startup | error |
71 | Device EEPROM acquire failure | Device cannot acquire (lock) EEPROM access during boot or operation | error |
72 | Device EEPROM release failure | Device cannot release (unlock) EEPROM after access during boot or operation | error |
73 | Device EEPROM invalid pointer failure | Memory pointer used for EEPROM access is corrupted or out of bounds | error |
E0.1 | Excessive Heater Energy during predict cycle | Heater draws excessive current or applies too much energy to probe | error |
E0.2 | A/D measurement out of range | Analog-to-digital converter receives voltage outside valid measurement range | error |
E0.4 | Probe over temperature (≥110°F / 43.3°C) | Probe tip temperature exceeds maximum safe threshold | warning |
E0.5 | No probe temperature rise (heater has no effect) | Heater powered but probe temperature does not increase during predict cycle | error |
E0.6 | Probe EEPROM read failure (bad handshake) | Device cannot establish communication handshake with probe EEPROM during startup or operation | error |
E0.7 | Probe EEPROM uninitialized | Probe EEPROM found but not initialized (missing calibration data) | error |
E0.8 | EEPROM CRC failure (data corrupted) | Cyclic redundancy check fails on probe or device EEPROM data | error |
E0.9 | Bad probe calibration | Probe EEPROM contains invalid or out-of-tolerance calibration data | error |
E4.0 | PTB measurement out of range | Patient Temperature Baseline measurement exceeds expected range during startup self-test | error |
E4.1 | RCAL measurement out of range | Reference calibration measurement during self-test outside expected bounds | error |
E4.2 | Reference power supply malfunction | Internal reference voltage supply voltage out of tolerance | error |
E4.3 | Device EEPROM save failure | Device cannot write calibration or configuration data to internal EEPROM | error |
E4.4 | Device EEPROM starting handshake failure | Device cannot establish initial communication with internal EEPROM during boot | error |
E4.5 | Device EEPROM response failure | EEPROM does not respond to read/write command within timeout | error |
E4.6 | Transmit failure to device EEPROM | Device data transmission to EEPROM fails or is corrupted | error |
E4.7 | Cannot initialize device EEPROM | Device boot sequence fails during EEPROM initialization | critical |
E4.8 | Device was not calibrated | Boot detects missing or invalid factory calibration data in EEPROM | error |
E5.0 | Heater not working properly | Internal heater fails to heat or maintain temperature during predict cycle | error |
E5.2 | Heater fail-safe failure | Heater safety shutdown mechanism malfunction; over-temperature protection triggered | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.