Welch Allyn SureTemp Plus 690: troubleshooting and error codes
The ServiceForge knowledge base holds 19 documented symptoms for the Welch Allyn SureTemp Plus 690, with 23 ranked causes, 39 diagnostic checks with test points and expected values, and 26 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Ambient temperature out of range
Display shows capital 'A' followed by blinking up or down arrow and blinking thermometer bulb when ambient temperature exceeds safe range.
Ranked causes
- 1. Ambient temperature too high (occasional)
- 2. Ambient temperature too low (occasional)
Verify ambient temperature
- Test point
- Ambient probe
- Expected
- 50.0 to 104.0°F, 50.0-104.0°F
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SureTemp Plus 690
Periodic calibration and self-test procedure for units in service or after extended storage
Ranked causes
- 1. Scheduled preventive maintenance for units in regular service (common)
- 2. Maintenance required before use after extended storage (common)
Perform Power-On Self Test
- Test point
- Unit startup
- Expected
- Self test completes without errors
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cleaning with disinfectants caution
Routine and deep cleaning procedures for thermometer and probe
Ranked causes
- 1. Routine cleaning requirement (common)
- 2. Deep cleaning with approved disinfectants (common)
Wipe with cloth and warm water with mild detergent
- Test point
- All exterior surfaces
- Expected
- Visibly clean, no residue
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient temperature out of range patient safety
Entire body (head, torso, legs) displays blinking with solid up or down arrow and thermometer bulb
Ranked causes
- 1. Patient temperature above 110.0°F or 43.3°C (occasional)
- 2. Patient temperature below 80.0°F or 26.7°C (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery above operational range
Display shows blinking battery frame, 'b' symbol, and upward arrow when battery voltage exceeds 5.375V.
Ranked causes
- 1. Battery voltage high (occasional)
Check battery voltage
- Test point
- Battery
- Expected
- ≤ 5.375 V
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery below operational range
Display shows blinking battery frame, 'b' symbol, and downward arrow when battery voltage drops below 3.00V.
Ranked causes
- 1. Battery voltage low (common)
Check battery voltage
- Test point
- Battery
- Expected
- ≥ 3.00 V
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery low warning
Single blinking battery segment in battery indicator when voltage is between 3.0V and 3.4V.
Ranked causes
- 1. Battery voltage declining (common)
Monitor battery voltage trend
- Test point
- Battery indicator
- Expected
- Single blinking segment V, > 3.0 and <= 3.4
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage above operational range
Blinking battery frame with 'b' and upward arrow on main display
Ranked causes
- 1. Battery voltage too high (> 5.375v) (common)
Measure battery voltage
- Test point
- Battery terminals
- Expected
- < 5.50 volts
- Tools
- Voltmeter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage below operational range
Blinking battery frame with 'b' and downward arrow on main display
Ranked causes
- 1. Battery voltage too low (< 3.00v) (common)
Measure battery voltage
- Test point
- Battery terminals
- Expected
- >= 3.1 volts
- Tools
- Voltmeter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SureTemp Plus 690 using Cal-Key
Verify thermometer calibration using the fixed-value Cal-Key (PN 06138-000) method to ensure clinical accuracy
Ranked causes
- 1. Calibration verification required (common)
Remove any probe currently connected to thermometer
- Test point
- Device probe port
- Expected
- Probe disconnected
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SureTemp Plus 690 using Model 9600 Plus Calibration Tester caution
Verify thermometer and probe calibration at three temperature points (36.0ºC, 38.5ºC, 41.0ºC) using Model 9600 Plus Calibration Tester
Ranked causes
- 1. Three-point calibration verification needed (common)
Store thermometers in same room as Model 9600 Plus for thermal accommodation
- Test point
- Room temperature environment
- Expected
- Approximately 30 minutes storage time
12 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Instrument error detected
Instrument with X through it displayed with error code in temperature field; temperature measuring prohibited
Ranked causes
- 1. Internal instrument circuitry failure (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery shutdown
Device automatically shuts off when battery voltage degrades below threshold to prevent erroneous temperature readings
Ranked causes
- 1. Battery voltage below 3 volts (common)
Measure battery voltage
- Test point
- Battery terminals
- Expected
- 3.60V nominal (fresh), minimum 3.0V operating, minimum threshold
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power-on self test failure
Device performs POST when battery power is applied to check internal hardware and stored calibration values
Ranked causes
- 1. Factory calibration not intact (unknown)
Verify battery voltage and POST completion
- Test point
- Power supply output
- Expected
- Peak 4.8V, average 3.60V
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe error detected
Probe with X through it displayed with error code in temperature field; temperature measuring prohibited
Ranked causes
- 1. Broken or malfunctioning probe (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe type detection
Thermometer reads probe EEPROM to determine probe type (Oral/Axillary or Rectal) and load appropriate calibration values
Ranked causes
- 1. Probe EEPROM not readable or corrupted (rare)
Verify probe detection on insertion
- Test point
- Optical probe detection switch
- Expected
- High-to-Low transition when probe pulled from well
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Six-month visual inspection and cleaning
Regular inspection for physical damage and cleaning of units in active service
Ranked causes
- 1. Regular preventive maintenance for active-service units (common)
Visually inspect for physical damage
- Test point
- Entire thermometer exterior and probe
- Expected
- No cracks, dents, or visible defects
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermistor measurement accuracy
Thermometer measures thermistor resistance to calculate patient temperature using nonlinear equation
Ranked causes
- 1. Thermistor degradation or out-of-spec resistance (occasional)
Measure thermistor resistance at room temperature
- Test point
- Thermistor element
- Expected
- Approximately 20K ohms
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit storage maintenance
Procedures for thermometers stored for extended periods to ensure functionality before reuse
Ranked causes
- 1. Stored unit requires reactivation maintenance (common)
Remove and reinstall batteries
- Test point
- Battery compartment
- Expected
- Batteries removed then reinstalled
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
23 | Host Interface Error | Communication failure with host interface | error |
26 | Invalid SureTemp mode | Device mode setting is invalid or corrupted | warning |
29 | Battery Voltage Is Not Set | Battery voltage parameter not configured in device memory | warning |
30 | Prediction mode Is Not Set | Prediction mode parameter not configured in device memory | warning |
71 | Device EEPROM acquire failure | Device cannot acquire lock or access to EEPROM | error |
72 | Device EEPROM release failure | Device cannot release EEPROM after access | error |
73 | DEVICES EEPROM invalid pointer failure | EEPROM memory pointer is invalid or corrupted | error |
E0.1 | Excessive Heater Energy | Heater produces excessive energy during measurement cycle | error |
E0.2 | A/D Measurement out of range error | Analog-to-digital converter measures value outside valid range | error |
E0.4 | Probe Over Temperature | Probe tip temperature exceeds 110°F/43.3°C | warning |
E0.5 | No Probe Temperature Rise | Probe heater has no effect on temperature measurement | error |
E0.6 | Probe EEPROM read failure | Bad EEPROM handshake between probe and device | error |
E0.7 | Probe EEPROM is un-initialized | Probe memory not properly initialized | error |
E0.8 | EEPROM CRC Failure | EEPROM data integrity check fails | error |
E0.9 | Bad probe calibration | Probe calibration data is invalid or corrupted | error |
E4.0 | PTB measurement out of range | Patient Temperature Buffer measurement exceeds valid range | error |
E4.1 | RCAL measurement out of range | Reference calibration measurement exceeds valid range | error |
E4.2 | Reference power supply malfunction | Internal reference power supply fails operational test | error |
E4.3 | Device EEPROM save failure | Device unable to write to internal EEPROM memory | error |
E4.4 | Device EEPROM starting handshake failure | Initial communication with EEPROM fails at startup | error |
E4.5 | Device EEPROM response failure | EEPROM does not respond to read/write commands | error |
E4.6 | Transmit failure to the Device EEPROM | Device unable to transmit data to internal EEPROM | error |
E4.7 | Can't initialize the device EEPROM | Device EEPROM initialization sequence fails | error |
E4.8 | Device was not calibrated | Device does not have valid calibration data loaded | error |
E5.0 | Heater is not working properly | Internal heater circuit malfunction detected | error |
E5.2 | Heater fail-safe failure | Heater fail-safe protective circuit malfunction | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.