Welch Allyn SureTemp 767: troubleshooting and error codes
The ServiceForge knowledge base holds 25 documented symptoms for the Welch Allyn SureTemp 767, with 54 ranked causes, 65 diagnostic checks with test points and expected values, and 13 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Temperature reading too high in normal mode patient safety
Thermometer displays elevated temperature readings
Ranked causes
- 1. Instrument calibration error (common)
- 2. Improper probe placement (common)
- 3. Probe malfunction or characterization error (occasional)
- 4. System calibration failure (rare)
- 5. Improper technique during measurement (common)
- 6. Probe inserted before display ready (common)
Verify display shows OrL before measurement
- Test point
- Display
- Expected
- Display shows OrL
1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Normal/Monitor mode switching problems patient safety
Unit switches between Normal and Monitor mode unexpectedly or incorrectly
Ranked causes
- 1. Ambient temperature above 33.9°C (93.0°F) (common)
- 2. Probe still cooling from previous use (occasional)
- 3. Auto-switch to Monitor mode after 60 seconds in Normal mode (common)
- 4. Defective mode button (occasional)
- 5. Defective probe (occasional)
Verify ambient temperature
- Test point
- Surrounding environment
- Expected
- Below 33.9°C (93.0°F)
- Tools
- Thermometer
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor mode temperature reading too low patient safety
Temperature reading in Monitor mode is lower than expected
Ranked causes
- 1. Probe malfunction or drift (common)
- 2. Instrument malfunction or calibration drift (occasional)
- 3. Improper probe placement (common)
- 4. Temperature not stabilized (common)
Test probe with calibration tester
- Test point
- Probe accuracy
- Expected
- Reading within tolerance per M9600 Calibration Tester
- Tools
- M9600 Calibration Tester
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No operation
Unit powered but not functioning or responding to inputs
Ranked causes
- 1. Short circuit in power circuit (common)
- 2. Power circuit component malfunction (occasional)
- 3. Oscillator failure (occasional)
- 4. LCD PCB loose (occasional)
Measure resistance at power input
- Test point
- Contacts J7 and J8
- Expected
- At least 3 M ohm
- Tools
- Ohm meter
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Probe type detection
Software determines probe type (Rectal, Calibration Key, or none) based on connector pin states at J1-3 (CAL), J1-5 (GND), and J1-6 (RCTL)
Ranked causes
- 1. Rectal probe connection (common)
- 2. Calibration Key connection (occasional)
- 3. No probe connected (occasional)
Measure voltage at J1-3 (CAL) relative to J1-5 (GND)
- Test point
- J1-3 (CAL) and J1-5 (GND)
- Expected
- 0 V (connected for Rectal probe, floating for other modes)
- Tools
- DMM
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display problems
LCD display unclear, crooked, dim, or not showing expected output
Ranked causes
- 1. Dirty or crooked elastomeric LCD conductor (zebra) strips (common)
- 2. Broken or shorted flex cable (occasional)
- 3. Cracked LCD (rare)
Inspect elastomeric strips and contacts for contamination
- Test point
- LCD elastomeric strips and glass contacts
- Expected
- Clean, properly aligned, no debris
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Normal mode temperature reading too low patient safety
Temperature reading in Normal mode is lower than expected
Ranked causes
- 1. Probe drift or calibration offset (common)
- 2. Defective probe (occasional)
- 3. Instrument malfunction or calibration drift (occasional)
Re-characterize probe by removal and reinstallation
- Test point
- Probe connection and recognition
- Expected
- Probe recharacterized, reading normalized
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe wrong type displayed
Incorrect probe type shown on display or probe not recognized
Ranked causes
- 1. Missing A/D power to probe connector (occasional)
- 2. Incorrect probe wiring (common)
- 3. Defective probe (occasional)
Check probe connector power pins without probe installed
- Test point
- Probe connector J1-3 and J1-6
- Expected
- Both pins pulled high during function active
- Tools
- DMM or logic probe
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on
No power supplied to thermometer unit
Ranked causes
- 1. Power supply not plugged in or turned off (common)
- 2. Power supply voltage out of range (occasional)
- 3. Power contacts not making contact (occasional)
Verify power supply is plugged in and powered on
- Test point
- Power supply
- Expected
- Power supply plugged in and on
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cal key does not activate thermometer
Calibration key not recognized or not activating thermometer
Ranked causes
- 1. Probe switch not activated (common)
- 2. Defective calibration key (occasional)
Connect cal key and activate probe switch
- Test point
- Display and probe switch activation
- Expected
- Display reads CAL for two seconds then blanks; probe switch activated
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Model 9600 Calibration Tester procedure
Testing entire thermometer system (instrument and probe) using Model 9600 Calibration Tester
Ranked causes
- 1. Complete thermometer system requires calibration verification (common)
- 2. Probe requires individual testing (common)
Verify 9600 Calibration Tester warm-up status
- Test point
- 9600 temperature display
- Expected
- Stable at one of two available temperature settings
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot enter Biotech mode
Unit cannot switch to Biotech mode operation
Ranked causes
- 1. Probe not connected or probe shaft sequence not completed (common)
- 2. Failed component or broken trace in mode button or probe switch (occasional)
Verify probe is connected to instrument
- Test point
- Probe connector
- Expected
- Probe fully inserted and connected
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor mode temperature reading too high patient safety
Temperature reading in Monitor mode is higher than expected
Ranked causes
- 1. Probe malfunction or drift (common)
- 2. Instrument malfunction or calibration drift (occasional)
Test probe and instrument with calibration tester
- Test point
- System calibration and probe accuracy
- Expected
- System within tolerance per M9600 Calibration Tester
- Tools
- M9600 Calibration Tester
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cal key displays OrL, rEC, or ALy
Calibration key shows error or invalid display codes
Ranked causes
- 1. Defective calibration key (common)
Observe cal key display codes
- Test point
- Display with cal key connected
- Expected
- CAL display; not OrL, rEC, or ALy
- Tools
- Observation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration Key test limitations
Calibration Key procedure does not test the probe; probe testing requires separate Model 9600 Calibration Tester equipment
Ranked causes
- 1. Probe accuracy verification requires Model 9600 Calibration Tester (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Calibration Key procedure
Testing thermometer accuracy using Calibration Key inserted into probe connector receptacle
Ranked causes
- 1. Thermometer requires calibration verification (common)
Display shows CAL indication
- Test point
- Display
- Expected
- CAL for two seconds, then blank
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Instrument and probe verification
Procedure to verify and adjust calibration of thermometer system
Ranked causes
- 1. Periodic calibration required (common)
Quick calibration check with Calibration Key
- Test point
- Calibration Key interface
- Expected
- Calibration reading within specification
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Power-up reset and self-tests
Microprocessor executes power-up reset sequence and self-test suite upon power application
Ranked causes
- 1. Power supply startup and reset initialization (common)
Monitor reset signal timing
- Test point
- U1-12 (reset line)
- Expected
- Active low reset after approximately 200 milliseconds from power application
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Self-tests on power-up
Microprocessor performs RAM, ROM, instruction set, self-calibration, probe warmer, probe, and ambient temperature tests; failures display specific error codes
Ranked causes
- 1. Self-test sequence initiated (common)
Verify RAM test passes
- Test point
- Display
- Expected
- No error code displayed after RAM test
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
F/C switching problems
Mode button or function/celsius switching not responding correctly
Ranked causes
- 1. Defective mode switch or broken trace (common)
Test mode button switch signal
- Test point
- U1-26
- Expected
- Low only when Mode button is pressed
- Tools
- DMM or oscilloscope
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Microcontroller clock generation
4.9152 MHz crystal (X1) with capacitors C26 and C27 generates oscillator; microcontroller divides internally by 2 to run at ~2.5 MHz
Ranked causes
- 1. Clock oscillator fault (rare)
Measure crystal oscillator frequency at X1
- Test point
- X1 crystal
- Expected
- 4.9152 MHz, +/- 50 ppm
- Tools
- Oscilloscope, frequency counter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Microprocessor power-up reset
Upon power-up, microprocessor receives reset signal from U1-12 (Econo reset chip U4) after ~200 ms delay, then launches self-tests
Ranked causes
- 1. Normal power-up reset sequence (common)
Measure voltage at U1-12 reset line
- Test point
- U1-12
- Expected
- Low (active) for ~200 ms after power applied, then returns high
- Tools
- Oscilloscope
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No backlight
Backlight function not working (optional feature)
Ranked causes
- 1. Broken trace or defective backlight component (occasional)
Measure backlight control signal at U1-42
- Test point
- U1-42
- Expected
- Low (Gnd) when LCD is on
- Tools
- DMM or oscilloscope
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply output voltage regulation high voltage
8VAC input converted to regulated 5V DC by full-wave bridge rectifier and low-dropout regulator; output range 4.75–5.25 V
Ranked causes
- 1. Power supply regulation fault (rare)
Measure 8VAC input voltage
- Test point
- Input to D7–D10 bridge rectifier
- Expected
- 8 VAC nominal
- Tools
- DMM (AC setting)
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe input noise filtering
Bypass capacitors C19 and C20 (0.01 µF each) filter spurious noise on J1-6 (PROBE_1) and J1-3 (PROBE_0) microprocessor input lines
Ranked causes
- 1. Noise on probe input lines (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
E0.1 | Probe heater accumulation test failure | Self-test detects probe heater malfunction | warning |
E0.2 | Probe A/D pulse width test failure | Self-test detects incorrect A/D conversion pulse width | warning |
E0.3 | Adaptive probe gain out of range | Self-test detects probe gain too high or too low | warning |
E1.1 | Ambient temperature high | Self-test detects ambient temperature exceeds safe limit | warning |
E1.2 | Ambient temperature low | Self-test detects ambient temperature below safe limit | warning |
E2.1 | Dead battery | Self-test detects battery voltage insufficient or battery failed | error |
E3.1 | RAM read/write test failure | Self-test detects RAM memory corruption or malfunction | critical |
E3.2 | ROM checksum test failure | Self-test detects ROM data integrity compromised | critical |
E3.3 | CPU instruction test failure | Self-test detects CPU cannot execute instructions properly | critical |
E4.0 | PTB resistor A/D pulse width test failure | Self-test detects incorrect PTB resistor analog-to-digital conversion | error |
E4.1 | RatioCal resistor A/D pulse width test failure | Self-test detects incorrect RatioCal resistor analog-to-digital conversion | error |
E5.0 | Heater circuit test failure | Self-test detects heater circuit malfunction | error |
E5.1 | Heater overheated | Self-test detects heater temperature exceeds safe limit | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.