Welch Allyn SureTemp 678: troubleshooting and error codes

The ServiceForge knowledge base holds 20 documented symptoms for the Welch Allyn SureTemp 678, with 48 ranked causes, 28 diagnostic checks with test points and expected values, and 15 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Normal/Monitor mode switching problems

Unit switches between Normal and Monitor modes unexpectedly or fails to switch when intended

Ranked causes

  1. 1. Ambient temperature above 33.9°C (93.0°F) (common)
  2. 2. Probe cooling from previous measurement (occasional)
  3. 3. Mode button inoperative or low signal (occasional)
  4. 4. Defective probe (occasional)
  5. 5. Failed component or broken trace (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery life problems

Battery drains quickly or does not maintain full charge during normal use

Ranked causes

  1. 1. Excessive alarms (common)
  2. 2. Excessive use in Monitor mode (common)
  3. 3. Dead cell in battery (occasional)
  4. 4. First set shelf life reduction due to long storage (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display problems

LCD inactive, distorted, or showing incorrect information

Ranked causes

  1. 1. LCD frame loose or misaligned (common)
  2. 2. Dirty LCD elastomeric conductor strips (common)
  3. 3. Cracked LCD (occasional)
  4. 4. Microprocessor failure (rare)
First check

Inspect LCD frame assembly

Test point
LCD frame plastic hooks
Expected
All 4 hooks tight, frame flush against PCB

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor mode temperature reading too low

Temperature displayed in Monitor mode is lower than expected during measurement

Ranked causes

  1. 1. Probe malfunction (common)
  2. 2. Instrument malfunction (occasional)
  3. 3. Improper placement of probe (common)
  4. 4. Temperature not stable (common)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Normal mode temperature reading too high

Temperature displayed in Normal mode is higher than expected after stabilization

Ranked causes

  1. 1. Probe malfunction (common)
  2. 2. Instrument malfunction (occasional)
  3. 3. Improper technique - movement in mouth (common)
  4. 4. Improper technique - probe insertion timing (common)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not power on

No operation, display inactive, no response to button presses

Ranked causes

  1. 1. Dead, missing, or incorrectly installed batteries (common)
  2. 2. Broken battery wire (occasional)
  3. 3. Short circuit preventing operation (occasional)
  4. 4. Failed component (oscillator or microprocessor) (rare)
First check

Verify battery installation

Test point
Battery compartment
Expected
All batteries installed in correct polarity

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot enter Biotech Mode

Mode button press does not activate Biotech Mode on unit in wall holder

Ranked causes

  1. 1. Mode button not pressed or instrument not in wall holder (678 only) (common)
  2. 2. Probe not connected or probe shaft not inserted and removed (common)
  3. 3. Failed component or broken trace (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No beeper sound

Horn does not sound during operation, errors, or timer cycles

Ranked causes

  1. 1. Defective horn speaker (common)
  2. 2. Broken connection between microprocessor and horn (occasional)
  3. 3. Defective microprocessor U1 (rare)
First check

Check horn control signal

Test point
U1-49
Expected
Square wave signal present during operation or error condition
Tools
Oscilloscope

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Normal mode temperature reading too low

Temperature displayed in Normal mode is lower than expected after stabilization

Ranked causes

  1. 1. Probe malfunction (common)
  2. 2. Instrument malfunction (occasional)
  3. 3. Improper placement of probe (common)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor mode temperature reading too high

Temperature displayed in Monitor mode is higher than expected during measurement

Ranked causes

  1. 1. Probe malfunction (common)
  2. 2. Instrument malfunction (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No Backlight

LCD backlight does not illuminate during operation

Ranked causes

  1. 1. Defective backlight LED D3 or resistor R23 (common)
  2. 2. No backlight control signal from microprocessor (occasional)
First check

Inspect backlight LED and resistor

Test point
LED D3 and resistor R23
Expected
LED intact with no visible cracks; resistor not burnt or open
Tools
Visual inspection, DMM for resistance measurement

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No Timer function

Timer button S1 does not activate or timer mode does not operate

Ranked causes

  1. 1. Defective timer button S1 (common)
  2. 2. Broken PCB trace to timer button (occasional)
First check

Measure signal at U1-26 when timer button pressed

Test point
U1-26 (microprocessor /PTIMER input)
Expected
Logic low (ground/0 V) when button S1 depressed; logic high (VCC) when released
Tools
DMM or oscilloscope

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cal Key displays OrL, rEC, or ALy

Cal Key shows error text instead of normal CAL display

Ranked causes

  1. 1. Defective Cal Key (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cal Key does not activate thermometer

Cal Key connected but does not activate calibration mode or display remains blank

Ranked causes

  1. 1. Probe switch not activated (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Self-tests

Verify microprocessor self-checks including RAM, ROM, instruction set, calibration, and sensor tests

Ranked causes

  1. 1. Power-up reset sequence initiated after battery installation (common)
First check

Verify RAM test

Test point
Microprocessor memory
Expected
RAM test passes without error code

9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Erroneous temperature readings patient safety

Temperature readings display incorrect values during operation

Ranked causes

  1. 1. Battery voltage degraded below functional threshold (occasional)
First check

Perform self-calibration test

Test point
Electronics accuracy, hi cal, low cal
Expected
All self calibration tests pass without error codes
Tools
Thermometer self-test function

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low battery error condition

E2.1 error message displayed, thermometer operation halted

Ranked causes

  1. 1. Battery voltage fallen to approximately 3.0 volts or lower (common)
First check

Measure battery voltage

Test point
Battery terminals
Expected
approximately 3.0 volts or lower indicates error condition V, below 3.0V indicates dead battery
Tools
Voltmeter

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low battery warning

Battery icon flashes in display indicating batteries are getting low

Ranked causes

  1. 1. Battery voltage fallen to approximately 3.2 volts (common)
First check

Measure battery voltage

Test point
Battery terminals
Expected
approximately 3.2 volts or above for normal operation V, above 3.2V for normal operation
Tools
Voltmeter

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Probe: wrong type displayed

Instrument displays incorrect probe type in normal or monitor mode operation

Ranked causes

  1. 1. Incorrect wiring of probe (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Recalled temperature is not correct

Unit displays temperature lower than expected when temperature is recalled from memory

Ranked causes

  1. 1. Unit switched to Monitor mode (common)
First check

Verify thermometer operating mode

Test point
Unit mode display or settings
Expected
Normal mode active (not Monitor mode)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
E0.1Probe heater accumulation test failureSelf-test detects error during power-on or operational self-testerror
E0.2Probe A/D pulse width test failureSelf-test detects error during analog-to-digital conversion measurementerror
E0.3Adaptive probe gain too high or too lowSelf-test detects out-of-range probe signal conditioning gain during startupwarning
E1.1Ambient temperature high test failureAmbient probe temperature measured above 104.0°F during self-test or operationwarning
E1.2Ambient temperature low test failureAmbient probe temperature measured below 60.8°F during self-test or operationwarning
E2.1Low battery error conditionBattery voltage falls to approximately 3.0 volts or lowererror
E3.1RAM read/write test failureInternal RAM memory test fails during self-test or power-onerror
E3.2ROM checksum test failureInternal ROM memory checksum validation fails during self-testcritical
E3.3CPU instruction test failureMicroprocessor instruction execution self-test fails during startupcritical
E4.0PTB resistor A/D pulse width test failurePrecision temperature bridge resistor analog-to-digital conversion failserror
E4.1RatioCal resistor A/D pulse width test failureRatio calibration resistor analog-to-digital conversion fails during self-testerror
E5.0Heater circuit test failureProbe heater power circuit test fails during self-testerror
E5.1Heater overheated test failureProbe heater temperature exceeds safe operating limit during self-test or operationerror
E5.2Heater watchdog timeout test failureHeater circuit control watchdog timer expires during probe heating cycleerror
E6.0PTB resistor temperature test failurePrecision temperature bridge resistor temperature measurement fails during self-testerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI