Welch Allyn PIC: troubleshooting and error codes
The ServiceForge knowledge base holds 52 documented symptoms for the Welch Allyn PIC, with 141 ranked causes, 45 diagnostic checks with test points and expected values, and 48 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Failure to pace patient safety
Pacing function does not deliver output despite activation
Ranked causes
- 1. Faulty or disconnected ECG patient cable (common)
- 2. Poor quality electrodes (common)
- 3. Incorrect electrodes (common)
- 4. Faulty or disconnected pacer cable (common)
- 5. Defib-pacer circuit board failure (occasional)
- 6. Defib/pacer-Motherboard communications error (occasional)
- 7. Failure of power supply (occasional)
- 8. Internal cable failure (rare)
- 9. ECG preamp-Motherboard communications error (rare)
- 10. ECG preamp lead fault circuitry failure (rare)
10 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of ECG signal patient safety
ECG display is blank or no waveform appears
Ranked causes
- 1. Patient cable not connected (common)
- 2. Damaged patient cable (common)
- 3. Damaged hands-free cable (common)
- 4. Poor quality electrodes (common)
- 5. Faulty or damaged display (rare)
- 6. ECG preamp input cable damaged or disconnected (occasional)
- 7. Motherboard-ECG preamp communications error (occasional)
- 8. Motherboard-ECG preamp cable damaged or disconnected (occasional)
- 9. Preamp circuit failure (rare)
- 10. Failure of power supply (rare)
13 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
BP cuff does not inflate caution
Blood pressure cuff remains deflated when measurement is initiated
Ranked causes
- 1. Pump/valve assembly or cable failure (common)
- 2. BP-Motherboard communications error (occasional)
- 3. BP module circuit failure (occasional)
- 4. Damaged or leaky cuff or hose (common)
- 5. Cuff hose crimped or obstructed (common)
- 6. Internal hose crimped or obstructed (occasional)
- 7. Failure of power supply (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure to charge high voltage
Unit does not charge defibrillator capacitor when charge button is pressed
Ranked causes
- 1. Failure of charge switch on front panel (common)
- 2. Front panel motherboard cable/connector faulty or damaged (common)
- 3. Failure of charge switch on paddle set (occasional)
- 4. Failure of paddle set cable/connector (occasional)
- 5. Defibrillator circuit error (occasional)
- 6. Motherboard circuit error (occasional)
- 7. Motherboard-defib communications error (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure to obtain oximeter measurement patient safety
Pulse oximetry value does not display or reads as disconnected
Ranked causes
- 1. Probe failure (common)
- 2. Failure of internal cable (occasional)
- 3. Oximeter-Motherboard cable faulty or disconnected (occasional)
- 4. Failure of oximeter board (occasional)
- 5. Oximeter probe-oximeter board cable failure (rare)
- 6. Failure of power supply (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure to obtain respiration measurement patient safety
Respiration rate does not display or shows as disconnected
Ranked causes
- 1. Patient cable not connected (common)
- 2. Damaged patient cable (common)
- 3. Damaged hands-free cable (common)
- 4. ECG preamp circuit failure (5-lead) (occasional)
- 5. ECG preamp circuit failure (12-lead) (occasional)
- 6. ECG preamp-Motherboard cable damaged or disconnected (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Artifact on ECG caution
ECG display shows noise or electrical interference not from patient
Ranked causes
- 1. Damaged patient cable (common)
- 2. Incorrect or incompatible patient cable (common)
- 3. Poor quality electrodes (common)
- 4. ECG preamp circuit failure (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure of quick charger caution
Battery does not charge when connected to quick charger unit
Ranked causes
- 1. Damaged contact pins (common)
- 2. Incorrect line voltage (occasional)
- 3. Ingress of liquids (rare)
- 4. Use of charger to charge non-MRL battery (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure to deliver energy high voltage
Defibrillator discharges with audible/visual indication but no energy is delivered to patient
Ranked causes
- 1. Defibrillator circuit error (occasional)
- 2. Discharge (fire) switch failure (occasional)
- 3. Failed or damaged paddle cable (common)
- 4. Faulty or damaged paddle set (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure to fire in sync mode patient safety
Defibrillator does not deliver shock synchronized to QRS complex
Ranked causes
- 1. Failed or damaged paddle cable, patient cable, or hands-free cable (common)
- 2. QRS not sent to defibrillator (occasional)
- 3. Internal cable defective (rare)
- 4. Motherboard circuit failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure to obtain temperature measurement caution
Temperature reading does not display or shows as disconnected
Ranked causes
- 1. Faulty temperature probe (common)
- 2. BP/temp module circuit failure (occasional)
- 3. BP-Motherboard communications error (occasional)
- 4. Faulty temperature input cable/board (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure to turn on caution
Device does not power up when power button is pressed
Ranked causes
- 1. Failed power switch (common)
- 2. Power supply circuit failure (common)
- 3. Battery failure (common)
- 4. Motherboard circuit failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of display or lines on display caution
Display is blank, dark, or missing portions of image
Ranked causes
- 1. Display cable faulty or disconnected (common)
- 2. Faulty display adapter board (occasional)
- 3. Damaged or faulty display (occasional)
- 4. Motherboard circuit failure (rare)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Non-Invasive Blood Pressure Performance Test patient safety
Verify NIBP static pressure accuracy, leak rate, overpressure activation, and BP/HR reading agreement against NIBP analyzer reference
Ranked causes
- 1. Static pressure reading inaccuracy (common)
- 2. Excessive leak rate (occasional)
- 3. Overpressure activation out of range (rare)
Verify stable static pressure reading at reference pressure
- Test point
- PIC display lower right corner, BP window
- Expected
- Pressure reading stabilizes after pressure adjustment
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure of Advisory module to prompt patient safety
Advisory (alarm/alert) function does not announce or display messages
Ranked causes
- 1. Speech board circuit failure (common)
- 2. Motherboard circuit error (occasional)
- 3. Speech failure - speaker cable or speaker (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure to respond to controls with device turned on caution
Device does not respond to button presses or control inputs
Ranked causes
- 1. Failure of power supply (occasional)
- 2. Front panel motherboard cable damaged or disconnected (common)
- 3. Motherboard circuit error (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate BP measurement caution
Blood pressure reading does not match manual measurement
Ranked causes
- 1. Improper cuff size or application (common)
- 2. Motion artifact (common)
- 3. BP module circuit failure or out of calibration (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate respiration measurement patient safety
Respiration rate reading does not match patient breathing rate
Ranked causes
- 1. Poor quality electrodes (common)
- 2. Incorrect electrodes (common)
- 3. Motion artifact (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to set defibrillator energy caution
Energy selector does not respond to user input or shows incorrect value
Ranked causes
- 1. Improper energy setting (unknown)
- 2. Damaged cable or connector in front panel (common)
- 3. Damaged energy switch on front panel (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not run on AC power caution
Device will not operate when connected to AC outlet, only on battery
Ranked causes
- 1. Failure of paddle tray/AC power supply (common)
- 2. Failure of battery board (occasional)
- 3. Failure of power supply board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit shuts off without low battery warning caution
Device powers down unexpectedly without displaying low battery message
Ranked causes
- 1. Battery failure (common)
- 2. Power supply circuit failure (common)
- 3. Motherboard circuit failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery reconditioning after failed charge time test caution
Battery charge time exceeds specification; battery age and usage frequency affect performance
Ranked causes
- 1. Degraded battery performance due to age or usage (common)
- 2. Battery failed after reconditioning (occasional)
Verify battery has been fully charged before testing
- Test point
- Battery pack charging status
- Expected
- Battery fully charged per charger indication
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Delivers 10 or fewer shocks with charged battery patient safety
Defibrillator delivers insufficient number of shocks despite battery being charged
Ranked causes
- 1. Batteries not properly reconditioned (common)
- 2. Failure of power supply (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Delivers fewer than 11 shocks with charged battery high voltage
Unit delivers 10 or fewer defibrillation shocks despite battery is charged
Ranked causes
- 1. Batteries not properly reconditioned (common)
- 2. Failure of power supply (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG signal frequency response out of specification
Recorded ECG test signal amplitude not within acceptable range during diagnostic bandwidth verification
Ranked causes
- 1. ECG preamp board degraded or faulty (common)
- 2. ECG simulator output not constant across frequency range (occasional)
Connect ECG simulator to PATIENT input; set for 0.05 Hz sine wave at 1 mV peak-to-peak
- Test point
- PATIENT input
- Expected
- 0.05 Hz sinusoidal output at 1 mV peak-to-peak
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG sweep speed or chart paper speed inaccurate
Distance between R-wave peaks on monitor display or recorder chart does not meet specification
Ranked causes
- 1. Chart recorder speed out of specification (common)
- 2. ECG simulator rate inaccurate (occasional)
Connect ECG simulator to PATIENT input; set rate to 60 bpm, amplitude to 1 mV
- Test point
- PATIENT input
- Expected
- 60 bpm, 1 mV amplitude
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure to obtain capture while pacing patient safety
Pacing pulses delivered but patient heart does not respond
Ranked causes
- 1. Poor quality electrodes (common)
- 2. Incorrect current or pacing circuit failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Improper lead fault indication caution
System incorrectly indicates lead disconnection when leads are properly connected
Ranked causes
- 1. ECG preamp-Motherboard communication failure (common)
- 2. ECG preamp circuit failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate oximeter measurement patient safety
Pulse oximetry reading does not match expected clinical value
Ranked causes
- 1. Probe failure (common)
- 2. Oximeter circuit failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate temperature measurement caution
Temperature reading does not match clinical measurement
Ranked causes
- 1. BP/temp module circuit failure or out of calibration (common)
- 2. Improper probe or incompatible equipment (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inappropriate Advisory module prompt patient safety
Advisory module gives false alarms or incorrect messages
Ranked causes
- 1. Artifact or motion corrupts signal (common)
- 2. ECG preamp circuit failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Main Oximeter Cable disconnected or unseated
Cable not properly connected at J221 on Motherboard or J2 on Oximeter Board, causing oximeter function loss or intermittent operation.
Ranked causes
- 1. Main Oximeter Cable unseated at J221 connector on Motherboard (common)
- 2. Main Oximeter Cable unseated at J2 connector on Oximeter Board (common)
Verify Main Oximeter Cable connection at J221 on Motherboard
- Test point
- J221
- Expected
- Cable fully seated with retaining clip engaged
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not recognize paddle set caution
Display shows 'Connect Paddle' warning message despite paddle set is connected
Ranked causes
- 1. Incompatible paddle set (common)
- 2. Paddle set connector not properly seated (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable replacement required
Main Blood Pressure Cable damaged, corroded or non-functional and requires complete replacement
Ranked causes
- 1. Main Blood Pressure Cable faulty or damaged (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 measurement
CO2 calibration procedure using 40 mmHg simulated trapezoidal signal to verify CO2 measurement accuracy
Ranked causes
- 1. CO2 calibration required (unknown)
Simulate 40 mmHg trapezoidal signal
- Test point
- CO2 Cal Menu
- Expected
- CO2 measurement response to simulated signal
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Display
Calibrate the PIC System display via the CALIBRATION quick access button
Ranked causes
- 1. Display calibration required (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ECG amplitude
Verify ECG amplitude display and chart recorder output matches 1 mV input signal
Ranked causes
- 1. ECG amplitude out of specification (unknown)
Connect ECG simulator to PATIENT input
- Test point
- PATIENT input connector
- Expected
- ECG simulator connected
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Modulation Output
Mod Out calibration procedure for supervisor to verify modulation output settings
Ranked causes
- 1. Modulation output calibration required (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP blood pressure measurement patient safety
Authorized service personnel calibration procedure for PIC blood pressure measurement capabilities using supervisor menu
Ranked causes
- 1. NIBP calibration required (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Restore PIC variable settings
After Power Supply Board replacement, PIC variable settings must be restored from factory defaults before returning to use
Ranked causes
- 1. Power Supply Board replacement resets all PIC settings to factory defaults (common)
Verify PIC powered on and operational
- Test point
- PIC display and status indicators
- Expected
- PIC boots successfully with factory default settings displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SPO2 DC output gain
SPO2 calibration procedure to verify DC output gain where 1 mV ECG input equals 1 volt DC output, with bias calibration options
Ranked causes
- 1. SPO2 gain calibration required (unknown)
Select bias calibration option
- Test point
- SPO2 Cal Menu bias selection
- Expected
- 0, 1, or 3
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Delivers low energy high voltage
Defibrillator delivers less energy than selected or expected
Ranked causes
- 1. Defibrillator circuit error (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure to respond to pacer controls patient safety
Pacing rate, mode, or current controls do not affect pacing output
Ranked causes
- 1. Front panel motherboard cable failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate pacing current patient safety
Pacing output does not match selected current setting
Ranked causes
- 1. Circuit failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect ECG frequency response caution
ECG waveform displays with improper filtering or frequency characteristics
Ranked causes
- 1. ECG preamp circuit failure (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD Display Adapter Board cable connections
Proper connection of display cables to LCD Display Adapter Board connectors with orientation requirements
Ranked causes
- 1. Flat Panel Display Cable not properly seated (occasional)
Inspect Flat Panel Display Cable orientation at J1
- Test point
- J1 connector, cable color side
- Expected
- Blue side of cable facing Motherboard (Item 46)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD Display Adapter Board replacement
Removal and replacement of LCD CD Display Adapter Board (Item 49) with note on part number selection based on backlight inverter wire configuration
Ranked causes
- 1. LCD Display Adapter Board requires replacement (unknown)
Verify correct part number for LCD Display Adapter Board based on backlight inverter wire configuration
- Test point
- Backlight Inverter Item 43 connections
- Expected
- Two white wires for Part Number 591382 OR two white and pink wires for Part Number 590382
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Main Blood Pressure Cable disconnected or loose
Blood pressure monitoring not functioning or readings unreliable due to cable disconnection at motherboard or blood pressure board connectors
Ranked causes
- 1. Main Blood Pressure Cable disconnected or improperly seated at J215 or J4 (common)
Visually inspect Main Blood Pressure Cable connections at J215 and J4
- Test point
- J215 on Motherboard (Item 46) and J4 on Blood Pressure Board (Item 92)
- Expected
- Cable fully inserted with retaining clip engaged
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power Supply Board replacement required
Loss of PIC variable settings to factory defaults occurs when Power Supply Board is removed and replaced
Ranked causes
- 1. Power Supply Board removal during service (common)
Verify all 16 Motherboard pins inserted into Power Supply Board
- Test point
- Power Supply Board (Item 44) connector to Motherboard (Item 46)
- Expected
- All 16 pins fully inserted
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace Defib Main Cable high voltage
Defib Main Cable (Item 19) requires removal and replacement as part of system maintenance or repair
Ranked causes
- 1. Cable replacement procedure (unknown)
Verify blue side of Defib Main Cable faces outside of Card Cage Pre-Assembly
- Test point
- J218 on Motherboard (Item 46)
- Expected
- Cable seated straight with blue side facing outside of Card Cage Pre-Assembly (Item 48)
2 more checks with pass and fail actions, 13 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace Mono LCD Cable caution
Cable replacement procedure for Mono LCD Cable (Item 35) connecting LCD Display Adapter Board to Top Enclosure Assembly
Ranked causes
- 1. Mono LCD Cable requires replacement (unknown)
Check that all cables are seated correctly in connectors and are clear of the sides of the unit
- Test point
- All connectors
- Expected
- Cables seated correctly and clear of unit sides
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verify power supply voltages caution
Check output voltages on power supply board connector J1 (J4 on motherboard) to verify supply rails are within specification
Ranked causes
- 1. Power supply out of regulation (occasional)
Measure voltage on PIN 1
- Test point
- J1 PIN 1
- Expected
- +11V +0.2V, +/-0.2V
- Tools
- DMM
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
ApexFireButnFail | Apex fire button failure | Paddle apex electrode fire button not responding | error |
Bad SAED Transition | Invalid SAED state transition | Motherboard software state machine error | error |
BP Chksum Err | Blood pressure checksum error | BP data integrity failure | error |
BP Comm Fail | Blood pressure communication failure | BP module unable to communicate with motherboard | error |
BP DC Path Err | Blood pressure DC signal path error | BP DC circuit malfunction | error |
BP RAM Err | Blood pressure RAM error | BP module memory failure | error |
Card Fail | Memory card failure | Memory card not detected or malfunctioning | error |
DefibCommFail | Defibrillator communication failure | Defib unable to communicate with main system | error |
DefibErr:008 | Defibrillator error 008 | Specific defib circuit error condition | error |
DefibErr:009 | Defibrillator error 009 | Specific defib circuit error condition | error |
DefibErr:010 | Defibrillator error 010 | Specific defib circuit error condition | error |
DefibErr:011 | Defibrillator error 011 | Specific defib circuit error condition | error |
DefibErr:12V Bulk | Defibrillator 12V bulk supply error | Power supply or charging circuit failure | error |
DefibErr:Chksum | Defibrillator checksum error | Defib data integrity or circuit error | critical |
DefibErr:SPI | Defibrillator SPI interface error | SPI communication failure | critical |
Modem init fail | Modem initialization failure | Modem hardware not responding | warning |
Modem timeout during modem response | Modem communication timeout | Modem response timeout | warning |
No answer at remote fax | No answer from remote fax | Receiving fax not responding | warning |
No carrier detected or carrier lost | Modem carrier signal lost | Connection lost during transmission | error |
No dialtone detected | No dial tone on phone line | Phone line not available | error |
Ox AC Path Err | Oximeter AC signal path error | Oximeter AC circuit or probe malfunction | error |
Ox Buffer Err | Oximeter data buffer error | Oximeter data buffering failure | error |
Ox Chksum Err | Oximeter checksum error | Data integrity failure in oximeter module | error |
Ox Comm Fail | Oximeter communication failure | Oximeter unable to communicate with motherboard | error |
Ox DC Path Err | Oximeter DC signal path error | Oximeter DC circuit malfunction | error |
PaceErr:013 | Pacemaker error 013 | Pacemaker circuit malfunction | error |
PaceErr:014 | Pacemaker error 014 | Pacemaker circuit malfunction | error |
PaceErr:015 | Pacemaker error 015 | Pacemaker circuit malfunction | error |
PaceErr:016 | Pacemaker error 016 | Test fixture impedance out of range | warning |
PaceErr:017 | Pacemaker error 017 | Pacemaker circuit malfunction | error |
PaceErr:018 | Pacemaker error 018 | Pacemaker circuit malfunction | error |
PaceErr:019 | Pacemaker error 019 | Pacemaker circuit malfunction | error |
PaceErr:020 | Pacemaker error 020 | Pacemaker circuit malfunction | error |
PaceErr:021 | Pacemaker error 021 | Pacemaker circuit malfunction | error |
Page send failed | Fax page transmission failed | Page not successfully transmitted | warning |
Page send failed (cellular) | Cellular fax transmission failed | Cellular transmission error | warning |
PdlChargeButnFail | Paddle charge button failure | Charge button not responding | error |
PdlEnergyDnButnFail | Paddle energy decrease button failure | Energy down button not responding | error |
PdlEnergyUpButnFail | Paddle energy increase button failure | Energy up button not responding | error |
Preamp A/D Err | Preamp analog-digital conversion error | Preamp A/D circuit malfunction | error |
Preamp cableset fail | Preamp cable set failure | Preamp cable disconnected or damaged | error |
Preamp Lead Test Err | Preamp lead test error | Lead impedance test failure | error |
Preamp QRS Det Err | Preamp QRS detection error | QRS detection circuit failure | error |
Remote fax busy | Remote fax line busy | Unable to connect to receiving fax | warning |
SpeechInitErr | Speech board initialization error | Speech board startup failure | error |
SternemFireButnFail | Sternum fire button failure | Paddle sternum electrode fire button not responding | error |
Temp A/D Err | Temperature A/D conversion error | Temperature analog-digital conversion failure | error |
xxx button fail | Front panel button malfunction | Faulty or damaged front panel button or control board issue | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.