The ServiceForge knowledge base holds 52 documented symptoms for the Welch Allyn PIC, with 141 ranked causes, 45 diagnostic checks with test points and expected values, and 48 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Failure to pace patient safety

Pacing function does not deliver output despite activation

Ranked causes

  1. 1. Faulty or disconnected ECG patient cable (common)
  2. 2. Poor quality electrodes (common)
  3. 3. Incorrect electrodes (common)
  4. 4. Faulty or disconnected pacer cable (common)
  5. 5. Defib-pacer circuit board failure (occasional)
  6. 6. Defib/pacer-Motherboard communications error (occasional)
  7. 7. Failure of power supply (occasional)
  8. 8. Internal cable failure (rare)
  9. 9. ECG preamp-Motherboard communications error (rare)
  10. 10. ECG preamp lead fault circuitry failure (rare)

10 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of ECG signal patient safety

ECG display is blank or no waveform appears

Ranked causes

  1. 1. Patient cable not connected (common)
  2. 2. Damaged patient cable (common)
  3. 3. Damaged hands-free cable (common)
  4. 4. Poor quality electrodes (common)
  5. 5. Faulty or damaged display (rare)
  6. 6. ECG preamp input cable damaged or disconnected (occasional)
  7. 7. Motherboard-ECG preamp communications error (occasional)
  8. 8. Motherboard-ECG preamp cable damaged or disconnected (occasional)
  9. 9. Preamp circuit failure (rare)
  10. 10. Failure of power supply (rare)

13 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

BP cuff does not inflate caution

Blood pressure cuff remains deflated when measurement is initiated

Ranked causes

  1. 1. Pump/valve assembly or cable failure (common)
  2. 2. BP-Motherboard communications error (occasional)
  3. 3. BP module circuit failure (occasional)
  4. 4. Damaged or leaky cuff or hose (common)
  5. 5. Cuff hose crimped or obstructed (common)
  6. 6. Internal hose crimped or obstructed (occasional)
  7. 7. Failure of power supply (rare)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure to charge high voltage

Unit does not charge defibrillator capacitor when charge button is pressed

Ranked causes

  1. 1. Failure of charge switch on front panel (common)
  2. 2. Front panel motherboard cable/connector faulty or damaged (common)
  3. 3. Failure of charge switch on paddle set (occasional)
  4. 4. Failure of paddle set cable/connector (occasional)
  5. 5. Defibrillator circuit error (occasional)
  6. 6. Motherboard circuit error (occasional)
  7. 7. Motherboard-defib communications error (rare)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure to obtain oximeter measurement patient safety

Pulse oximetry value does not display or reads as disconnected

Ranked causes

  1. 1. Probe failure (common)
  2. 2. Failure of internal cable (occasional)
  3. 3. Oximeter-Motherboard cable faulty or disconnected (occasional)
  4. 4. Failure of oximeter board (occasional)
  5. 5. Oximeter probe-oximeter board cable failure (rare)
  6. 6. Failure of power supply (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure to obtain respiration measurement patient safety

Respiration rate does not display or shows as disconnected

Ranked causes

  1. 1. Patient cable not connected (common)
  2. 2. Damaged patient cable (common)
  3. 3. Damaged hands-free cable (common)
  4. 4. ECG preamp circuit failure (5-lead) (occasional)
  5. 5. ECG preamp circuit failure (12-lead) (occasional)
  6. 6. ECG preamp-Motherboard cable damaged or disconnected (occasional)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Artifact on ECG caution

ECG display shows noise or electrical interference not from patient

Ranked causes

  1. 1. Damaged patient cable (common)
  2. 2. Incorrect or incompatible patient cable (common)
  3. 3. Poor quality electrodes (common)
  4. 4. ECG preamp circuit failure (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure of quick charger caution

Battery does not charge when connected to quick charger unit

Ranked causes

  1. 1. Damaged contact pins (common)
  2. 2. Incorrect line voltage (occasional)
  3. 3. Ingress of liquids (rare)
  4. 4. Use of charger to charge non-MRL battery (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure to deliver energy high voltage

Defibrillator discharges with audible/visual indication but no energy is delivered to patient

Ranked causes

  1. 1. Defibrillator circuit error (occasional)
  2. 2. Discharge (fire) switch failure (occasional)
  3. 3. Failed or damaged paddle cable (common)
  4. 4. Faulty or damaged paddle set (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure to fire in sync mode patient safety

Defibrillator does not deliver shock synchronized to QRS complex

Ranked causes

  1. 1. Failed or damaged paddle cable, patient cable, or hands-free cable (common)
  2. 2. QRS not sent to defibrillator (occasional)
  3. 3. Internal cable defective (rare)
  4. 4. Motherboard circuit failure (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure to obtain temperature measurement caution

Temperature reading does not display or shows as disconnected

Ranked causes

  1. 1. Faulty temperature probe (common)
  2. 2. BP/temp module circuit failure (occasional)
  3. 3. BP-Motherboard communications error (occasional)
  4. 4. Faulty temperature input cable/board (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure to turn on caution

Device does not power up when power button is pressed

Ranked causes

  1. 1. Failed power switch (common)
  2. 2. Power supply circuit failure (common)
  3. 3. Battery failure (common)
  4. 4. Motherboard circuit failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of display or lines on display caution

Display is blank, dark, or missing portions of image

Ranked causes

  1. 1. Display cable faulty or disconnected (common)
  2. 2. Faulty display adapter board (occasional)
  3. 3. Damaged or faulty display (occasional)
  4. 4. Motherboard circuit failure (rare)

9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Non-Invasive Blood Pressure Performance Test patient safety

Verify NIBP static pressure accuracy, leak rate, overpressure activation, and BP/HR reading agreement against NIBP analyzer reference

Ranked causes

  1. 1. Static pressure reading inaccuracy (common)
  2. 2. Excessive leak rate (occasional)
  3. 3. Overpressure activation out of range (rare)
First check

Verify stable static pressure reading at reference pressure

Test point
PIC display lower right corner, BP window
Expected
Pressure reading stabilizes after pressure adjustment

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure of Advisory module to prompt patient safety

Advisory (alarm/alert) function does not announce or display messages

Ranked causes

  1. 1. Speech board circuit failure (common)
  2. 2. Motherboard circuit error (occasional)
  3. 3. Speech failure - speaker cable or speaker (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure to respond to controls with device turned on caution

Device does not respond to button presses or control inputs

Ranked causes

  1. 1. Failure of power supply (occasional)
  2. 2. Front panel motherboard cable damaged or disconnected (common)
  3. 3. Motherboard circuit error (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inaccurate BP measurement caution

Blood pressure reading does not match manual measurement

Ranked causes

  1. 1. Improper cuff size or application (common)
  2. 2. Motion artifact (common)
  3. 3. BP module circuit failure or out of calibration (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inaccurate respiration measurement patient safety

Respiration rate reading does not match patient breathing rate

Ranked causes

  1. 1. Poor quality electrodes (common)
  2. 2. Incorrect electrodes (common)
  3. 3. Motion artifact (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to set defibrillator energy caution

Energy selector does not respond to user input or shows incorrect value

Ranked causes

  1. 1. Improper energy setting (unknown)
  2. 2. Damaged cable or connector in front panel (common)
  3. 3. Damaged energy switch on front panel (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not run on AC power caution

Device will not operate when connected to AC outlet, only on battery

Ranked causes

  1. 1. Failure of paddle tray/AC power supply (common)
  2. 2. Failure of battery board (occasional)
  3. 3. Failure of power supply board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit shuts off without low battery warning caution

Device powers down unexpectedly without displaying low battery message

Ranked causes

  1. 1. Battery failure (common)
  2. 2. Power supply circuit failure (common)
  3. 3. Motherboard circuit failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery reconditioning after failed charge time test caution

Battery charge time exceeds specification; battery age and usage frequency affect performance

Ranked causes

  1. 1. Degraded battery performance due to age or usage (common)
  2. 2. Battery failed after reconditioning (occasional)
First check

Verify battery has been fully charged before testing

Test point
Battery pack charging status
Expected
Battery fully charged per charger indication

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Delivers 10 or fewer shocks with charged battery patient safety

Defibrillator delivers insufficient number of shocks despite battery being charged

Ranked causes

  1. 1. Batteries not properly reconditioned (common)
  2. 2. Failure of power supply (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Delivers fewer than 11 shocks with charged battery high voltage

Unit delivers 10 or fewer defibrillation shocks despite battery is charged

Ranked causes

  1. 1. Batteries not properly reconditioned (common)
  2. 2. Failure of power supply (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG signal frequency response out of specification

Recorded ECG test signal amplitude not within acceptable range during diagnostic bandwidth verification

Ranked causes

  1. 1. ECG preamp board degraded or faulty (common)
  2. 2. ECG simulator output not constant across frequency range (occasional)
First check

Connect ECG simulator to PATIENT input; set for 0.05 Hz sine wave at 1 mV peak-to-peak

Test point
PATIENT input
Expected
0.05 Hz sinusoidal output at 1 mV peak-to-peak

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG sweep speed or chart paper speed inaccurate

Distance between R-wave peaks on monitor display or recorder chart does not meet specification

Ranked causes

  1. 1. Chart recorder speed out of specification (common)
  2. 2. ECG simulator rate inaccurate (occasional)
First check

Connect ECG simulator to PATIENT input; set rate to 60 bpm, amplitude to 1 mV

Test point
PATIENT input
Expected
60 bpm, 1 mV amplitude

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure to obtain capture while pacing patient safety

Pacing pulses delivered but patient heart does not respond

Ranked causes

  1. 1. Poor quality electrodes (common)
  2. 2. Incorrect current or pacing circuit failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Improper lead fault indication caution

System incorrectly indicates lead disconnection when leads are properly connected

Ranked causes

  1. 1. ECG preamp-Motherboard communication failure (common)
  2. 2. ECG preamp circuit failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inaccurate oximeter measurement patient safety

Pulse oximetry reading does not match expected clinical value

Ranked causes

  1. 1. Probe failure (common)
  2. 2. Oximeter circuit failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inaccurate temperature measurement caution

Temperature reading does not match clinical measurement

Ranked causes

  1. 1. BP/temp module circuit failure or out of calibration (common)
  2. 2. Improper probe or incompatible equipment (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inappropriate Advisory module prompt patient safety

Advisory module gives false alarms or incorrect messages

Ranked causes

  1. 1. Artifact or motion corrupts signal (common)
  2. 2. ECG preamp circuit failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Main Oximeter Cable disconnected or unseated

Cable not properly connected at J221 on Motherboard or J2 on Oximeter Board, causing oximeter function loss or intermittent operation.

Ranked causes

  1. 1. Main Oximeter Cable unseated at J221 connector on Motherboard (common)
  2. 2. Main Oximeter Cable unseated at J2 connector on Oximeter Board (common)
First check

Verify Main Oximeter Cable connection at J221 on Motherboard

Test point
J221
Expected
Cable fully seated with retaining clip engaged

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System does not recognize paddle set caution

Display shows 'Connect Paddle' warning message despite paddle set is connected

Ranked causes

  1. 1. Incompatible paddle set (common)
  2. 2. Paddle set connector not properly seated (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cable replacement required

Main Blood Pressure Cable damaged, corroded or non-functional and requires complete replacement

Ranked causes

  1. 1. Main Blood Pressure Cable faulty or damaged (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CO2 measurement

CO2 calibration procedure using 40 mmHg simulated trapezoidal signal to verify CO2 measurement accuracy

Ranked causes

  1. 1. CO2 calibration required (unknown)
First check

Simulate 40 mmHg trapezoidal signal

Test point
CO2 Cal Menu
Expected
CO2 measurement response to simulated signal

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Display

Calibrate the PIC System display via the CALIBRATION quick access button

Ranked causes

  1. 1. Display calibration required (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: ECG amplitude

Verify ECG amplitude display and chart recorder output matches 1 mV input signal

Ranked causes

  1. 1. ECG amplitude out of specification (unknown)
First check

Connect ECG simulator to PATIENT input

Test point
PATIENT input connector
Expected
ECG simulator connected

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Modulation Output

Mod Out calibration procedure for supervisor to verify modulation output settings

Ranked causes

  1. 1. Modulation output calibration required (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: NIBP blood pressure measurement patient safety

Authorized service personnel calibration procedure for PIC blood pressure measurement capabilities using supervisor menu

Ranked causes

  1. 1. NIBP calibration required (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Restore PIC variable settings

After Power Supply Board replacement, PIC variable settings must be restored from factory defaults before returning to use

Ranked causes

  1. 1. Power Supply Board replacement resets all PIC settings to factory defaults (common)
First check

Verify PIC powered on and operational

Test point
PIC display and status indicators
Expected
PIC boots successfully with factory default settings displayed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: SPO2 DC output gain

SPO2 calibration procedure to verify DC output gain where 1 mV ECG input equals 1 volt DC output, with bias calibration options

Ranked causes

  1. 1. SPO2 gain calibration required (unknown)
First check

Select bias calibration option

Test point
SPO2 Cal Menu bias selection
Expected
0, 1, or 3

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Delivers low energy high voltage

Defibrillator delivers less energy than selected or expected

Ranked causes

  1. 1. Defibrillator circuit error (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure to respond to pacer controls patient safety

Pacing rate, mode, or current controls do not affect pacing output

Ranked causes

  1. 1. Front panel motherboard cable failure (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inaccurate pacing current patient safety

Pacing output does not match selected current setting

Ranked causes

  1. 1. Circuit failure (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incorrect ECG frequency response caution

ECG waveform displays with improper filtering or frequency characteristics

Ranked causes

  1. 1. ECG preamp circuit failure (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD Display Adapter Board cable connections

Proper connection of display cables to LCD Display Adapter Board connectors with orientation requirements

Ranked causes

  1. 1. Flat Panel Display Cable not properly seated (occasional)
First check

Inspect Flat Panel Display Cable orientation at J1

Test point
J1 connector, cable color side
Expected
Blue side of cable facing Motherboard (Item 46)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD Display Adapter Board replacement

Removal and replacement of LCD CD Display Adapter Board (Item 49) with note on part number selection based on backlight inverter wire configuration

Ranked causes

  1. 1. LCD Display Adapter Board requires replacement (unknown)
First check

Verify correct part number for LCD Display Adapter Board based on backlight inverter wire configuration

Test point
Backlight Inverter Item 43 connections
Expected
Two white wires for Part Number 591382 OR two white and pink wires for Part Number 590382

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Main Blood Pressure Cable disconnected or loose

Blood pressure monitoring not functioning or readings unreliable due to cable disconnection at motherboard or blood pressure board connectors

Ranked causes

  1. 1. Main Blood Pressure Cable disconnected or improperly seated at J215 or J4 (common)
First check

Visually inspect Main Blood Pressure Cable connections at J215 and J4

Test point
J215 on Motherboard (Item 46) and J4 on Blood Pressure Board (Item 92)
Expected
Cable fully inserted with retaining clip engaged

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power Supply Board replacement required

Loss of PIC variable settings to factory defaults occurs when Power Supply Board is removed and replaced

Ranked causes

  1. 1. Power Supply Board removal during service (common)
First check

Verify all 16 Motherboard pins inserted into Power Supply Board

Test point
Power Supply Board (Item 44) connector to Motherboard (Item 46)
Expected
All 16 pins fully inserted

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Replace Defib Main Cable high voltage

Defib Main Cable (Item 19) requires removal and replacement as part of system maintenance or repair

Ranked causes

  1. 1. Cable replacement procedure (unknown)
First check

Verify blue side of Defib Main Cable faces outside of Card Cage Pre-Assembly

Test point
J218 on Motherboard (Item 46)
Expected
Cable seated straight with blue side facing outside of Card Cage Pre-Assembly (Item 48)

2 more checks with pass and fail actions, 13 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Replace Mono LCD Cable caution

Cable replacement procedure for Mono LCD Cable (Item 35) connecting LCD Display Adapter Board to Top Enclosure Assembly

Ranked causes

  1. 1. Mono LCD Cable requires replacement (unknown)
First check

Check that all cables are seated correctly in connectors and are clear of the sides of the unit

Test point
All connectors
Expected
Cables seated correctly and clear of unit sides

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Verify power supply voltages caution

Check output voltages on power supply board connector J1 (J4 on motherboard) to verify supply rails are within specification

Ranked causes

  1. 1. Power supply out of regulation (occasional)
First check

Measure voltage on PIN 1

Test point
J1 PIN 1
Expected
+11V +0.2V, +/-0.2V
Tools
DMM

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
ApexFireButnFailApex fire button failurePaddle apex electrode fire button not respondingerror
Bad SAED TransitionInvalid SAED state transitionMotherboard software state machine errorerror
BP Chksum ErrBlood pressure checksum errorBP data integrity failureerror
BP Comm FailBlood pressure communication failureBP module unable to communicate with motherboarderror
BP DC Path ErrBlood pressure DC signal path errorBP DC circuit malfunctionerror
BP RAM ErrBlood pressure RAM errorBP module memory failureerror
Card FailMemory card failureMemory card not detected or malfunctioningerror
DefibCommFailDefibrillator communication failureDefib unable to communicate with main systemerror
DefibErr:008Defibrillator error 008Specific defib circuit error conditionerror
DefibErr:009Defibrillator error 009Specific defib circuit error conditionerror
DefibErr:010Defibrillator error 010Specific defib circuit error conditionerror
DefibErr:011Defibrillator error 011Specific defib circuit error conditionerror
DefibErr:12V BulkDefibrillator 12V bulk supply errorPower supply or charging circuit failureerror
DefibErr:ChksumDefibrillator checksum errorDefib data integrity or circuit errorcritical
DefibErr:SPIDefibrillator SPI interface errorSPI communication failurecritical
Modem init failModem initialization failureModem hardware not respondingwarning
Modem timeout during modem responseModem communication timeoutModem response timeoutwarning
No answer at remote faxNo answer from remote faxReceiving fax not respondingwarning
No carrier detected or carrier lostModem carrier signal lostConnection lost during transmissionerror
No dialtone detectedNo dial tone on phone linePhone line not availableerror
Ox AC Path ErrOximeter AC signal path errorOximeter AC circuit or probe malfunctionerror
Ox Buffer ErrOximeter data buffer errorOximeter data buffering failureerror
Ox Chksum ErrOximeter checksum errorData integrity failure in oximeter moduleerror
Ox Comm FailOximeter communication failureOximeter unable to communicate with motherboarderror
Ox DC Path ErrOximeter DC signal path errorOximeter DC circuit malfunctionerror
PaceErr:013Pacemaker error 013Pacemaker circuit malfunctionerror
PaceErr:014Pacemaker error 014Pacemaker circuit malfunctionerror
PaceErr:015Pacemaker error 015Pacemaker circuit malfunctionerror
PaceErr:016Pacemaker error 016Test fixture impedance out of rangewarning
PaceErr:017Pacemaker error 017Pacemaker circuit malfunctionerror
PaceErr:018Pacemaker error 018Pacemaker circuit malfunctionerror
PaceErr:019Pacemaker error 019Pacemaker circuit malfunctionerror
PaceErr:020Pacemaker error 020Pacemaker circuit malfunctionerror
PaceErr:021Pacemaker error 021Pacemaker circuit malfunctionerror
Page send failedFax page transmission failedPage not successfully transmittedwarning
Page send failed (cellular)Cellular fax transmission failedCellular transmission errorwarning
PdlChargeButnFailPaddle charge button failureCharge button not respondingerror
PdlEnergyDnButnFailPaddle energy decrease button failureEnergy down button not respondingerror
PdlEnergyUpButnFailPaddle energy increase button failureEnergy up button not respondingerror
Preamp A/D ErrPreamp analog-digital conversion errorPreamp A/D circuit malfunctionerror
Preamp cableset failPreamp cable set failurePreamp cable disconnected or damagederror
Preamp Lead Test ErrPreamp lead test errorLead impedance test failureerror
Preamp QRS Det ErrPreamp QRS detection errorQRS detection circuit failureerror
Remote fax busyRemote fax line busyUnable to connect to receiving faxwarning
SpeechInitErrSpeech board initialization errorSpeech board startup failureerror
SternemFireButnFailSternum fire button failurePaddle sternum electrode fire button not respondingerror
Temp A/D ErrTemperature A/D conversion errorTemperature analog-digital conversion failureerror
xxx button failFront panel button malfunctionFaulty or damaged front panel button or control board issueerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI