Welch Allyn Patient Monitor 6200: troubleshooting and error codes
The ServiceForge knowledge base holds 49 documented symptoms for the Welch Allyn Patient Monitor 6200, with 130 ranked causes, 116 diagnostic checks with test points and expected values, and 1 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Printer motor does not run
Printer motor fails to start or spin when paper advance is initiated; device does not respond to print commands
Ranked causes
- 1. Unit not configured for printer operation (common)
- 2. Printer PCB to motor cable disconnected or damaged (occasional)
- 3. Printer PCB to main board cable disconnected or damaged (occasional)
- 4. Motor gear misaligned with first plastic gear (occasional)
- 5. Paper jam blocking motor (common)
- 6. Printer PCB failure (occasional)
- 7. CPU board failure preventing printer operation (rare)
Check system printer configuration setting
- Test point
- Configuration menu or display freeze duration
- Expected
- Unit configured for printer; no 10-second freeze on print
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure manometer accuracy out of specification patient safety
Measured pressure readings fail accuracy calibration test; readings consistently off from expected values; measurement drift or bias error
Ranked causes
- 1. Pump pressure capacity insufficient; needs verification and retest (common)
- 2. Check valve direction incorrect or malfunction (occasional)
- 3. Pneumatic leaks preventing stable pressure measurement (occasional)
- 4. Tubing kinked or blocked restricting accurate measurement (occasional)
- 5. Pressure transducer failure or measurement circuit fault (occasional)
- 6. Lifted pins on CPU connector JP8 (occasional)
Verify pump pressure capacity ≥300 mmHg
- Test point
- BP pump output
- Expected
- Pressure reaches 300+ mmHg
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure overpressure limit out of specification patient safety
Overpressure protection fails; system exceeds safe pressure limits; safety interlock may not function properly
Ranked causes
- 1. Pump capacity insufficient; needs verification (common)
- 2. Check valve reversed or failed (occasional)
- 3. System leak preventing overpressure detection (occasional)
- 4. Tubing pinched or blocked (occasional)
- 5. Pressure transducer or comparator circuit failure on main board (occasional)
- 6. CPU board failure in overpressure logic (rare)
Verify pump can exceed 300 mmHg
- Test point
- BP pump test circuit
- Expected
- Pressure reaches 300+ mmHg
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ETCO2 system failure, sensor unresponsive
ETCO2 system does not respond; pump does not activate; water trap switch failure or board malfunction
Ranked causes
- 1. Water trap receptacle switch broken or intermittent (common)
- 2. Service module board ID mismatch or configuration error (occasional)
- 3. Pump not responding to water trap switch activation (occasional)
- 4. Water trap receptacle switch intermittent; needs replacement (occasional)
- 5. Pryon board failure (occasional)
- 6. CPU board failure preventing ETCO2 operation (rare)
Test water trap receptacle switch activation
- Test point
- Water trap switch and pump response
- Expected
- Pump activates when switch is pressed or water trap inserted
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power up, no AC or power LED caution
Device completely unresponsive; AC indicator LED dark; power-on LED not illuminating
Ranked causes
- 1. AC power cord disconnected or damaged (common)
- 2. Power supply not connected to main board (occasional)
- 3. Power supply 12V output failure (occasional)
- 4. Power supply defective (occasional)
- 5. Main board to display board cable disconnected or damaged (occasional)
- 6. Display board failure (occasional)
Verify AC power present at outlet
- Test point
- Power outlet
- Expected
- AC power available
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure pump cannot exceed pressure limit, 'Can't pump beyond' error
System fails to reach target pressure; pump output limited; message displayed indicating maximum pressure not achieved
Ranked causes
- 1. Test incomplete or pump function normal; needs retest after verification (common)
- 2. Check valve installed in wrong direction or failed (occasional)
- 3. Pneumatic leak in system preventing pressure buildup (occasional)
- 4. Tubing pinched or blocked restricting flow (occasional)
- 5. Pump mechanical failure or worn pump (occasional)
Verify pump can reach 300 mmHg
- Test point
- BP pump pressure test circuit
- Expected
- Pressure reaches at least 300 mmHg without load
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CRT displays no image despite correct main to CRT connection high voltage
No visible output on CRT screen; main board to CRT connection verified; flex cable or high voltage connection may be compromised
Ranked causes
- 1. Flex cable from deflection board to tube disconnected or damaged (common)
- 2. High voltage anode connection disconnected or loose (common)
- 3. Deflection board failure (occasional)
- 4. CRT tube failure (occasional)
- 5. CPU board failure (rare)
Verify flex-cable from deflection board to tube
- Test point
- Flex-cable connectors at both ends
- Expected
- Cable fully seated and undamaged
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Earth resistance exceeds specification limit patient safety
Ground continuity measurement fails; resistance between ground reference and equipment chassis exceeds safe limits; electrical safety hazard
Ranked causes
- 1. Motor ground wire loose or disconnected from power supply can (common)
- 2. Ground tab fastening screws loose (common)
- 3. Power supply PCB mounting screws loose (occasional)
- 4. AC harness quick-connect terminals loose at power supply (occasional)
- 5. Power supply PCB failure with ground circuit defect (occasional)
Measure earth resistance (ground continuity)
- Test point
- Chassis ground reference
- Expected
- Less than equipment specification limit (typically <0.1 Ohm)
- Tools
- Digital multimeter with continuity or ohms measurement
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 system displays fault message, no signal detected patient safety
'SpO2 fault' message appears on CRT; system does not detect or respond to SpO2 probe; measurement unavailable
Ranked causes
- 1. SpO2 probe disconnected or not functioning in test (common)
- 2. SpO2 board to main board cable disconnected or damaged (occasional)
- 3. Main board failure affecting SpO2 circuit (occasional)
- 4. SpO2 board failure or defect (occasional)
- 5. CPU board failure preventing SpO2 processing (rare)
Test SpO2 probe functionality with finger placement
- Test point
- SpO2 finger cuff and signal detection
- Expected
- Signal detected when probe connected to finger; normal waveform visible
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power up, AC and power LEDs on caution
Device receives power (AC and power-on LEDs lit) but does not fully initialize or respond
Ranked causes
- 1. CPU board not fully seated (common)
- 2. CPU board failure (occasional)
- 3. Conformal coating contamination in J4 header (occasional)
- 4. Main board failure (occasional)
- 5. Display board failure (rare)
Verify CPU board fully seated in socket
- Test point
- CPU board connector
- Expected
- Board fully inserted with all pins making contact
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure leak test fails specification
BP system fails leak test; pneumatic leakage detected; system cannot maintain pressure during test cycle
Ranked causes
- 1. Intermittent leak in system that may pass individual test on retry (occasional)
- 2. Leaking seals on tubing connections in BP assembly (common)
- 3. Lifted pins on CPU connector JP8 causing air leak path (occasional)
- 4. Check valve failure or blockage (occasional)
Run leak test on individual unit
- Test point
- BP system leak test circuit
- Expected
- No pressure loss within specification during test cycle
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer motor runs but paper does not advance
Motor spins and makes noise but paper feed fails; thermal head chattering or grinding sounds; paper remains stationary during print cycle
Ranked causes
- 1. Missing dowel pin in paper roller (common)
- 2. Missing bearings or gears in paper feed mechanism (occasional)
- 3. Printer PCB failure affecting motor control (occasional)
- 4. CPU board failure affecting print control (rare)
Inspect roller for missing dowel pin
- Test point
- Paper roller shaft
- Expected
- Dowel pin present and secure
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer outputs garbled or unreadable characters
Printed text is jumbled, contains unrecognizable characters, wrong spacing, or is otherwise corrupted; thermal print quality issue or data transmission error
Ranked causes
- 1. CPU board not fully seated or loose connection (common)
- 2. CPU board failure (occasional)
- 3. Printer to main board cable damaged or loose (occasional)
- 4. Printer PCB failure (occasional)
Verify CPU board fully seated in socket
- Test point
- CPU board connector
- Expected
- All pins fully inserted and making contact
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer produces light or no print output
Printed output is faint, very light in color, or completely absent; thermal head may not be heating properly; paper may not be making proper contact
Ranked causes
- 1. Paper installed incorrectly with thermal side not visible (common)
- 2. Printer PCB to print head cable disconnected or damaged (occasional)
- 3. Printer thermistor circuit failure or strobe width setting too low (occasional)
- 4. Print head force or pressure too light on paper (occasional)
Verify paper installed correctly with thermal side visible
- Test point
- Paper orientation in printer
- Expected
- Thermal (shiny) side facing print head and visible
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit powers up but no image on CRT display caution
Device starts and LEDs power on but CRT screen remains blank or black; no image output to monitor
Ranked causes
- 1. Conformal coating in J4 header (occasional)
- 2. CPU board defective (occasional)
- 3. Main to deflection board connection incorrect or damaged (occasional)
- 4. CPU assembly failure (occasional)
Inspect J4 header for conformal coating
- Test point
- J4 connector on board
- Expected
- No coating contamination visible
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display segments stay on at power-up
Most or all of display segments remain illuminated when unit powers on; display does not clear to show normal startup state
Ranked causes
- 1. CPU board failure (common)
- 2. Display board to main board cable failure (occasional)
- 3. Power supply failure affecting display circuit (occasional)
Verify display board to main board cable connection
- Test point
- Cable connectors at both ends
- Expected
- Connectors fully seated
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP overpressure detected patient safety
Audible alarm and 'Check blood pressure cuff' message displayed; NIBP pump and valve drive signals opened
Ranked causes
- 1. Cuff pressure >=10 mmHg for 295 seconds (common)
- 2. Cuff pressure >=15 mmHg for 175 seconds (common)
- 3. Cuff pressure >295 mmHg for 0.5 seconds (occasional)
Monitor digitized primary transducer output for sustained overpressure condition
- Test point
- PrimaryPres A/D input
- Expected
- <10 mmHg (nominal)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power on self-test fails or displays incorrect response
Monitor does not display expected LEDs, beeps, dashes, or CRT screen content after power on; indicates hardware or firmware issue.
Ranked causes
- 1. Power button LED or audio alarm circuit failure (common)
- 2. CRT display failure to illuminate or show content (common)
- 3. Measurement parameters not displaying expected dash patterns or values (occasional)
Plug monitor into AC power outlet and verify AC~ LED indicator is lit
- Test point
- AC power input connector; AC~ LED on front panel
- Expected
- AC~ LED is illuminated (no flickering); stable power detected
- Tools
- AC power outlet with ground; multimeter for voltage check if needed
7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature measurement system failure or disconnection
Temperature probe disconnected, unresponsive, or measurement circuit failure; no temperature reading displayed
Ranked causes
- 1. Temperature probe to main board connection loose or disconnected (common)
- 2. Temperature probe connector pins not properly seated (occasional)
- 3. Main board temperature measurement circuit failure (occasional)
Verify temperature probe cable connected to main board jack
- Test point
- Main board temperature connector
- Expected
- Cable fully and securely connected
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure deflation profile out of specification
Cuff deflation rate or curve does not match expected specification; deflation too fast or irregular; system dump valve malfunction
Ranked causes
- 1. Dump valve failure or malfunction (common)
- 2. Main board valve drive circuitry fault (occasional)
Verify deflation profile rate and regularity
- Test point
- System deflation during measurement cycle
- Expected
- Smooth, continuous deflation at specified rate (exact rate per specification document)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure dwell pressure out of specification
Pressure level at dwell point does not match specified value; typically 15 mmHg detection point accuracy or consistency issue
Ranked causes
- 1. Main board 15 mmHg detection circuit fault (occasional)
- 2. CPU board firmware or logic error affecting dwell detection (occasional)
Measure dwell pressure value
- Test point
- System dwell detection at specified point
- Expected
- Dwell pressure equals specification value, typically 15 mmHg (see specification doc)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure dwell time out of specification
Time interval between start of measurement and 15 mmHg detection point incorrect; dwell timing does not meet specification
Ranked causes
- 1. Main board 15 mmHg detection circuit fault (occasional)
- 2. CPU board timer circuit fault (occasional)
Measure dwell time interval
- Test point
- System dwell measurement
- Expected
- Dwell time within specification (consult specification document for exact limits)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
BP SYSTEM FAULT displayed patient safety
'BP SYSTEM FAULT' message on CRT display; NIBP system disabled; redundant safety transistor opens to ensure pump off and valve open
Ranked causes
- 1. Primary transducer pressure >330 mmHg (nominal trip 314 mmHg) (rare)
- 2. Primary transducer pressure >15 mmHg (nominal 13.3 mmHg) for three minutes (rare)
Monitor primary transducer output voltage for hardware overpressure fault conditions
- Test point
- U203A/B comparator inputs, PrimaryPres signal
- Expected
- 0.75 V at 0 mmHg; 4.0 V at 300 mmHg (with 5 V supply)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CRT displays only horizontal center line
CRT shows only a horizontal line across center of screen; no vertical scan or full raster; vertical deflection absent
Ranked causes
- 1. Yoke to CRT board cable problem (common)
- 2. Deflection board failure (occasional)
Verify yoke cable connected to CRT board
- Test point
- Yoke cable connectors
- Expected
- Cable fully seated at CRT board end
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CRT shows glow only at bottom of screen
Dark glow visible at bottom edge of CRT display; minimal or no image in upper and middle portions of screen
Ranked causes
- 1. Conformal coating in J4 header on circuit board (occasional)
- 2. CPU board failure affecting display circuitry (occasional)
Inspect J4 header on CPU or main board
- Test point
- J4 connector
- Expected
- No conformal coating residue visible
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ETCO2 system fails specification for flow rate or percentage
Capnography measurements fail accuracy or flow rate specification limits; readings outside acceptable range
Ranked causes
- 1. System requires calibration or calibration needs repeating (common)
- 2. Pryon board failure affecting measurement accuracy (occasional)
Perform ETCO2 calibration procedure
- Test point
- Calibration sequence and results
- Expected
- Specification limits met after calibration
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not power on or immediately powers off caution
No response to power button; monitor shuts down shortly after turning on; AC LED not lit or unstable.
Ranked causes
- 1. AC power not present or unstable (common)
- 2. Battery depleted (models 622/623 only) (occasional)
Verify AC power outlet is functional using a known-good device (lamp, charger, etc.)
- Test point
- AC power outlet receptacle
- Expected
- Test device powers on normally; outlet is live with stable voltage
- Tools
- Lamp or another AC device to test outlet
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP Fault Message displayed patient safety
NIBP system disabled; transducer outputs disagree or A/D reference voltage measurement error detected
Ranked causes
- 1. Primary and safety transducer outputs disagree (occasional)
- 2. A/D reference voltage measurement error (occasional)
Compare digitized primary and safety transducer outputs
- Test point
- SafetyPres and PrimaryPres A/D inputs
- Expected
- Within calibrated range; typically <5% disagreement
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Print on Alarm function not working
Printer does not print automatically when alarm condition triggers, or prints when it should not
Ranked causes
- 1. Print on Alarm setting disabled (common)
- 2. Alarm condition not triggering (occasional)
Access Advanced Configuration menu and disable Print on Alarm
- Test point
- Print on Alarm setting in Configuration menu
- Expected
- Print on alarm set to No
7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 signal accuracy out of specification patient safety
Measured SpO2 values do not match expected accuracy; readings drift or are inconsistent with reference values
Ranked causes
- 1. Nellcor probe ribbon cable or harness damaged (occasional)
- 2. SpO2 board failure or sensor circuit defect (occasional)
Visually inspect probe ribbon cable for damage
- Test point
- Probe connector cable from SpO2 board
- Expected
- No visible kinks, tears, or pinching
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit powers on intermittently or power button sticks caution
Device intermittently powers on and off; power button physically sticks or power-on operation cannot be consistently reproduced; unit passes burn-in testing
Ranked causes
- 1. Flashing on bezel interfering with power button (occasional)
- 2. Power button LED tipped or misaligned on display board (occasional)
Inspect bezel flashing around power button
- Test point
- Bezel surface near power button
- Expected
- No flashing material present
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure alarm cannot be silenced patient safety
Audio alarm continues after pressing BP ALARMS OFF button; visual alarms remain active; red LED in button not illuminating
Ranked causes
- 1. BP ALARMS OFF button or alarm mute circuit failure (unknown)
Verify audio alarm silences
- Test point
- Atlas speaker
- Expected
- Audio alarm silent
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure alarm not triggering on overpressure patient safety
No audio alarm, no visual 'Check Blood Pressure Cuff' message, no flashing SYSTOLIC/DIASTOLIC LEDs when finger blocks BP port
Ranked causes
- 1. Alarm system malfunction (unknown)
Verify audio alarm sounds
- Test point
- Atlas speaker
- Expected
- Audio alarm audible
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure alarm triggers at timeout without cuff connection patient safety
Audio and visual alarms trigger at 60 seconds with no cuff attached; SYSTOLIC and DIASTOLIC LEDs flash
Ranked causes
- 1. Normal cuff disconnection detection behavior (common)
Verify alarm triggers at 60 seconds
- Test point
- System timer and alarm circuit
- Expected
- Audio and visual alarm at 60 second mark
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure diastolic high alarm does not trigger patient safety
No audio alarm or LED flash when diastolic pressure exceeds high alarm threshold of 70 mmHg
Ranked causes
- 1. Diastolic HI alarm threshold not set or alarm circuit failure (unknown)
Verify DIASTOLIC HI alarm threshold set
- Test point
- SELECT and SET buttons, DIASTOLIC LED display
- Expected
- 70 mmHg displayed after setting
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure diastolic low alarm does not trigger patient safety
No audio alarm or LED flash when diastolic pressure drops below low alarm threshold of 90 mmHg
Ranked causes
- 1. Diastolic LO alarm threshold not set or alarm circuit failure (unknown)
Verify DIASTOLIC LO alarm threshold set
- Test point
- SELECT and SET buttons, DIASTOLIC LED display
- Expected
- 90 mmHg displayed after setting
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure systolic high alarm does not trigger patient safety
No audio alarm or LED flash when systolic pressure exceeds high alarm threshold of 130 mmHg with simulator set to 140/80
Ranked causes
- 1. Systolic HI alarm threshold not set or alarm circuit malfunction (unknown)
Verify SYSTOLIC HI alarm threshold is set correctly
- Test point
- SELECT and SET buttons, SYSTOLIC LED display
- Expected
- 130 mmHg in LED display after setting
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure systolic low alarm does not trigger patient safety
No audio alarm or LED flash when systolic pressure drops below low alarm threshold of 160 mmHg with simulator at 140/80
Ranked causes
- 1. Systolic LO alarm threshold not set or alarm circuit malfunction (unknown)
Verify SYSTOLIC LO alarm is set correctly
- Test point
- SELECT and SET buttons, SYSTOLIC LED display
- Expected
- 160 mmHg displayed after setting
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration Failed message
Calibration fails during ET CO2 calibration procedure
Ranked causes
- 1. System date set to 2022 or later (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Blood pressure patient safety
BP calibration procedure using external pressure reference and manometer connected to BP port via T-connectors
Ranked causes
- 1. Blood pressure calibration required (unknown)
Verify HyperTerminal connection to COM1
- Test point
- PC COM1 port
- Expected
- Pangea> prompt displayed in HyperTerminal after pressing Enter
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: BP (Blood Pressure)
BP calibration procedure using external pressure source and manometer to verify monitor accuracy
Ranked causes
- 1. Routine calibration setup (common)
Verify date and time are correct on Atlas monitor
- Test point
- CLOCK button display
- Expected
- Current date and time match system time
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Safety transducer two-point
Factory or field recalibration of safety transducer gain and zero using two known pressure points and NVRAM storage
Ranked causes
- 1. Safety transducer loose initial accuracy (+/-50%) requires calibration (common)
Measure safety transducer output at zero pressure
- Test point
- SafetyPres signal at P202-Out+/Out- (differential), or TP226
- Expected
- 0 mmHg reference; nominal gain 0.25 mV/mmHg with +/-25 mV zero offset
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Capnography waveform not displayed
After capnography sensor warmup, pump motor sounds and initial waveform with respiration dashes shown, but waveform not updating with breathing
Ranked causes
- 1. Capnography sensor not detecting breath gas flow (common)
Observe waveform response to gentle breathing
- Test point
- Display lower trace
- Expected
- Waveform visible within seconds of breathing
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CRT displays dark vertical scan lines in image
Multiple dark vertical lines visible across CRT image, appearing as scan line artifacts or loss of horizontal resolution segments
Ranked causes
- 1. CPU board failure affecting display control (common)
Verify CPU board operation
- Test point
- System startup and display processing
- Expected
- No scan line artifacts; clear image without vertical lines
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CRT displays only center dot
CRT screen shows single dot in center of display with no horizontal or vertical scan; no image raster visible
Ranked causes
- 1. Yoke to CRT board cable disconnected or damaged (common)
Verify yoke to CRT board cable connected
- Test point
- Cable connectors at both ends
- Expected
- Cable fully seated at yoke and CRT board
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CRT image rotated or poorly aligned
Image displayed on CRT is tilted, skewed, or not properly centered on screen; alignment is off from expected position
Ranked causes
- 1. Yoke position incorrect or magnetic deflection misaligned (common)
Inspect yoke positioning
- Test point
- CRT yoke mounting
- Expected
- Yoke aligned to display centered and upright image
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CRT segment or LED out, buttons non-responsive
One or more display segments do not illuminate; individual LED indicators dark; front panel buttons unresponsive to input
Ranked causes
- 1. Display board failure (common)
Verify display board power and connections
- Test point
- Display board connector
- Expected
- All connections secure and power present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pangea Communication Protocol
Serial interface protocol for Atlas monitor interaction with computer; case-sensitive English commands with prompts
Ranked causes
- 1. Enable serial communication for calibration (common)
Monitor for Pangea prompt after Atlas power-up
- Test point
- HyperTerminal display
- Expected
- Pangea> prompt appears
- Tools
- HyperTerminal on PC
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer output darkness or legibility issues
Waveform printout too dark, too light, lines too thick/thin, text illegible, blurring or blooming of text
Ranked causes
- 1. Printer calibration not optimized (common)
Connect ECG simulator and generate sample waveform
- Test point
- ECG simulator output to Atlas ECG input
- Expected
- Waveform appears on printout
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
ECG Lead fault | ECG cable disconnected or not properly connected | ECG cable is disconnected from the patient monitor | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI