Welch Allyn ELI 380: troubleshooting and error codes
The ServiceForge knowledge base holds 25 documented symptoms for the Welch Allyn ELI 380, with 46 ranked causes, 40 diagnostic checks with test points and expected values, and 11 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Excess artifact on ECG tracing patient safety
ECG waveform contains excessive electrical noise or signal corruption
Ranked causes
- 1. Incorrect electrode type (common)
- 2. Corroded or contaminated banana plug or clip contacts (common)
- 3. Cracked or damaged lead wire insulation (occasional)
- 4. Loose lead wire connection at AM12/AM15E/WAM module (occasional)
- 5. Patient cable defect (rare)
Verify electrode type and application
- Test point
- electrodes on patient
- Expected
- Baxter-brand electrodes applied correctly
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No tracings, screen displays Searching for WAM continuously
WAM wireless module not communicating with instrument; Searching message persists
Ranked causes
- 1. WAM out of communication range (common)
- 2. WAM battery depleted or damaged (common)
- 3. Damaged WAM battery cap (occasional)
- 4. Wrong WAM model for instrument (occasional)
- 5. WAM not paired to instrument (occasional)
Verify communication range
- Test point
- WAM and instrument distance
- Expected
- within specified range
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper does not advance or advances incorrectly caution
Paper fails to advance, advances less than one inch, advances more than one page, or stops unexpectedly during print operation
Ranked causes
- 1. Incorrect paper type or old/stale paper (common)
- 2. Pages sticking together from compression or humidity damage (common)
- 3. Paper loaded incorrectly or cueing mark misaligned (occasional)
- 4. Cardboard separator sheets left under paper (occasional)
- 5. Paper advance mechanism jam or motor failure (rare)
Verify correct paper type and condition
- Test point
- Paper specification and condition
- Expected
- Paper part number 9100-026-52 or 9100-026-55; no old or damaged sheets; paper stored in humidity-controlled area
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High frequency noise on ECG signal
Electrical noise at high frequencies appearing on all leads of the ECG trace; interference from external sources or patient muscle tension.
Ranked causes
- 1. Improper low pass filter setting or AC filter setting (common)
- 2. Portable electronic devices in patient vicinity (occasional)
- 3. Patient has implanted muscle stimulator (rare)
- 4. Patient muscles tense or not relaxed (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper does not advance, grinding noise caution
Paper tray motor grinds without advancing paper through printer
Ranked causes
- 1. Incorrect paper type loaded (common)
- 2. Cardboard at bottom of paper stack (occasional)
- 3. Broken plastic in paper tray (occasional)
- 4. Gearbox misalignment or drive gear damage (rare)
Verify paper type
- Test point
- paper loaded in tray
- Expected
- Baxter brand paper
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Cue sensor
Cue sensor calibration required when cue sensor, writer assembly, or motherboard changed; or when 'Paper Queue Fault' message appears
Ranked causes
- 1. Cue sensor misalignment after component replacement (common)
- 2. Paper Queue Fault indication (occasional)
Verify cue mark on paper is offset from cue sensor on printer
- Test point
- Paper tray, perforation line
- Expected
- Paper not stopped at perforation line
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper comes out crooked or has lines missing
Printed ECG output misaligned or with missing print lines
Ranked causes
- 1. Incorrect paper type (common)
- 2. Dirty print head (common)
Verify paper type
- Test point
- paper loaded in tray
- Expected
- Baxter brand paper
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen is dark
Display shows no image; WLAN module not responding
Ranked causes
- 1. AC power cord not connected or damaged (common)
- 2. AC power supply failure (occasional)
Visually inspect AC power cord for damage
- Test point
- AC power cord
- Expected
- No visible damage
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device is not responding
Unit unresponsive to input; does not process commands or button presses
Ranked causes
- 1. System frozen or stuck in error state (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrocardiograph is in Standby Mode
Device powered but not active
Ranked causes
- 1. Standby mode activated (common)
Press On/Standby button
- Test point
- On/Standby button
- Expected
- Device returns to active use
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrocardiograph will not power on
Device does not respond to power button
Ranked causes
- 1. Power supply fault or reset needed (common)
Disconnect AC power and hold power button
- Test point
- Power button
- Expected
- Button held for >30 seconds
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lead fault; no ECG capture patient safety
Electrical lead connection failure; ECG waveform cannot be acquired
Ranked causes
- 1. Lead failure or improper configuration (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leads OFF message for one or more leads; square waves on display patient safety
Individual lead electrodes not sensing; displayed waveforms show square waves instead of ECG
Ranked causes
- 1. Lead electrode connection not secure or faulty electrode (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LEADS OFF or SEARCHING FOR WAM message patient safety
No signal from WAM, AM12 or AM15E module; unable to acquire ECG data
Ranked causes
- 1. WAM, AM12 or AM15E module not configured or not powered (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No tracings on screen or gaps in tracing, Leads Off shown patient safety
ECG waveform absent or intermittently missing with Leads Off message displayed
Ranked causes
- 1. Cable between AM12/AM15E and ELI 380 has broken wire (occasional)
Test cable continuity by flexing and observing display
- Test point
- interconnect cable from module to instrument
- Expected
- tracing present continuously
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper advance mechanism calibration out of spec caution
Paper advances incorrectly despite correct paper type and loading; advance distance not within expected range
Ranked causes
- 1. Paper advance cue sensor requires calibration adjustment (occasional)
Perform cue calibration procedure
- Test point
- Paper advance sensor and calibration routine
- Expected
- Paper advances smoothly to end of page and stops correctly after calibration
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper Queue Fault message displayed
Error message appears on screen indicating cue sensor calibration needed
Ranked causes
- 1. Cue sensor out of calibration (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor LA electrode or left arm tremor affecting Lead I and Lead III patient safety
Weak electrode contact on left arm or patient muscle tremor causing noise in leads I and III
Ranked causes
- 1. Poor LA electrode contact or patient arm tremor (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor LL electrode or left leg tremor affecting Lead II and Lead III patient safety
Weak electrode contact on left leg or patient muscle tremor causing noise in leads II and III
Ranked causes
- 1. Poor LL electrode contact or patient leg tremor (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor RA electrode or right arm tremor affecting Lead I and Lead II patient safety
Weak electrode contact on right arm or patient muscle tremor causing noise in leads I and II
Ranked causes
- 1. Poor RA electrode contact or patient arm tremor (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen does not turn on when lid is lifted
Display remains off after opening lid
Ranked causes
- 1. Display power or lid sensor issue (common)
Press power button once
- Test point
- Power button
- Expected
- Baxter logo appears within 10 seconds
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen shows Searching for WAM
Display indicates wireless acquisition module (WAM) not detected or out of range
Ranked causes
- 1. AM12/AM15E module selected instead of WAM in settings (common)
Verify WAM/AM-XX setting
- Test point
- settings menu gears icon
- Expected
- WAM mode selected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sweep square not advancing on screen patient safety
ECG time-base square does not move horizontally across display but time counter advances
Ranked causes
- 1. AM12/AM15E module plugged into wrong socket (common)
Verify AM12/AM15E module socket location
- Test point
- module socket label
- Expected
- socket marked ECG
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to acquire ECG or unable to print
Device cannot capture ECG waveform or produce printed output; battery indicator shows LOW
Ranked causes
- 1. Battery low (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
WAM battery low
Display shows low battery warning for wireless acquisition module
Ranked causes
- 1. WAM battery depleted (common)
Replace WAM battery
- Test point
- WAM battery
- Expected
- fresh AA battery installed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
BATTERY LOW – CHARGE UNIT | Battery charge level critically low; device cannot operate on battery power | Battery voltage drops below minimum operating threshold | warning |
CAN'T CONNECT TO ACCESS POINT | Unit unable to establish link to wireless access point | WLAN connection attempt failed after multiple retries | warning |
DHCP FAILURE | WLAN module failed to obtain IP address from DHCP server | WLAN initialization attempted but DHCP address assignment failed | error |
DPAC FAILURE | WLAN module failed to initialize | WLAN initialization failure during system startup or network reconnection | critical |
ERROR-DICOM Not Enabled | DICOM communication attempted but unit not configured for DICOM protocol | DICOM transmission request received while DICOM configuration disabled | error |
LEAD FAULT, NO ECG CAPTURE | Electrical connection failure on one or more ECG leads; ECG acquisition impossible | Lead impedance exceeds threshold; electrode not in contact; lead wire disconnected or damaged | error |
Paper Queue Fault | Paper advance system error or paper jam detected | Paper advance mechanism detects inability to advance paper or mechanical jam condition | warning |
Paper Queue Fault | Cue sensor unable to detect paper queue correctly | Cue sensor misalignment or cue sensor, writer assembly, or motherboard component change | warning |
Paper Queue Fault | Cue sensor is out of calibration or not functioning properly | Cue sensor offset from proper position or sensor malfunction | warning |
TRANSMIT FAILED | Unable to transmit ECG data to receiving system | ECG transmission attempt failed due to network or configuration error | error |
UNABLE TO SAVE ECG | ECG data cannot be stored due to excessive noise or full storage directory | Noise level too high for reliable ECG storage or directory storage capacity exceeded | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI