Welch Allyn Connex Spot Monitor: troubleshooting and error codes
The ServiceForge knowledge base holds 104 documented symptoms for the Welch Allyn Connex Spot Monitor, with 180 ranked causes, 76 diagnostic checks with test points and expected values, and 310 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Blood pressure readings inaccurate patient safety
NIBP measurements do not match expected values or clinical norms; inconsistent results across multiple measurements
Ranked causes
- 1. Incorrect cuff size, improper placement, or incompatible cuff (common)
- 2. Patient arm not at heart level or movement during measurement (common)
- 3. Cuff placed over clothing (common)
- 4. Patient has cardiac arrhythmia (occasional)
- 5. Leak in pneumatic system (occasional)
- 6. NIBP module requires calibration (occasional)
- 7. NIBP module malfunction (occasional)
- 8. Single or dual lumen switch failure (rare)
18 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature measurements are not displayed patient safety
Temperature readings do not appear on monitor display
Ranked causes
- 1. The probe is disconnected from the monitor (common)
- 2. Probe is placed improperly (common)
- 3. The probe malfunctioned (common)
- 4. The temperature module malfunctioned (occasional)
- 5. A software issue occurred (occasional)
- 6. The probe or the window is dirty or contaminated (occasional)
- 7. The main board malfunctioned (rare)
13 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device does not power up
Device fails to turn on when power switch is activated
Ranked causes
- 1. New battery installed (occasional)
- 2. Battery discharged (common)
- 3. Power switch defective (occasional)
- 4. Power supply defective (occasional)
- 5. Battery dead or defective (occasional)
- 6. Main board defective (rare)
If new battery installed, apply AC power
- Test point
- AC power connector
- Expected
- Device powers on with AC power
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor cannot connect with Bluetooth device
Monitor fails to establish Bluetooth connection with external device
Ranked causes
- 1. Pairing issue occurred (common)
- 2. Bluetooth device driver not installed properly (occasional)
- 3. PC Service for device has not started (occasional)
- 4. Connectivity issue occurred (occasional)
- 5. Hardware component has failed (rare)
- 6. Software has failed (rare)
17 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature measurements are inaccurate patient safety
Temperature readings do not match expected values or are unreliable
Ranked causes
- 1. Probe is placed improperly (common)
- 2. The probe malfunctioned (common)
- 3. The probe cover is faulty or incorrectly attached (occasional)
- 4. The probe or the window is dirty or contaminated (occasional)
- 5. The temperature module malfunctioned (rare)
- 6. The main board malfunctioned (rare)
10 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Complete functional verification and calibration
Use Gold edition service tool to verify device meets specifications and perform calibration after damage, repair, case opening, or malfunction
Ranked causes
- 1. Device does not meet specifications (common)
- 2. Device dropped or damaged (occasional)
- 3. Device is malfunctioning (common)
- 4. Case has been opened (occasional)
- 5. Internal part has been replaced (occasional)
Verify device meets specifications using basic functional verification
- Test point
- device under test
- Expected
- Device meets all specifications
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Complete functional verification required patient safety
Full device calibration and verification needed when specific conditions trigger comprehensive service
Ranked causes
- 1. Basic functional verification shows device does not meet specifications (occasional)
- 2. Device has been dropped or damaged (occasional)
- 3. Device is malfunctioning (occasional)
- 4. Case has been opened (occasional)
- 5. Internal part has been replaced (battery excluded) (occasional)
Determine if basic functional verification met specifications
- Test point
- Device
- Expected
- Meets design specifications
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP cuff does not inflate patient safety
Blood pressure cuff fails to pressurize when measurement cycle initiated; NIBP function unavailable or degraded
Ranked causes
- 1. Leak in pneumatic system (common)
- 2. NIBP module malfunction (occasional)
- 3. Main board malfunction (occasional)
- 4. Single or dual lumen switch failure (rare)
- 5. NIBP module requires calibration (occasional)
17 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Attempt to upgrade the software failed
Software update process did not complete successfully
Ranked causes
- 1. The hardware or software between the Host and the Ethernet module is incompatible (occasional)
- 2. The monitor lost communication with the network or a timing issue occurred during the upgrade (occasional)
- 3. The temperature module malfunctioned (rare)
- 4. The main board malfunctioned (rare)
11 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bluetooth device did not connect with monitor
Bluetooth connection cannot be established after pairing attempt
Ranked causes
- 1. Bluetooth connection dropped due to distance or application crash (unknown)
- 2. RF connection not possible (unknown)
- 3. RF connection possible but cannot find device (unknown)
- 4. Bluetooth connection established but WACP connection not possible (unknown)
Power-on self test failure patient safety
Error message and error code appear in system status area during power-up; system enters safe mode and stops monitoring functions.
Ranked causes
- 1. CVC-PIC test failure (unknown)
- 2. LED power test failure (unknown)
- 3. eMMC memory test failure (unknown)
- 4. EEPROM test failure (unknown)
Verify buzzer generates sound on power-up
- Test point
- Buzzer output
- Expected
- Sound audible
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Software upgrade failed
Attempt to upgrade monitor software was unsuccessful
Ranked causes
- 1. Hardware or software incompatibility between Host and SpO2 module (occasional)
- 2. Network communication loss or timing issue during upgrade (occasional)
- 3. Temperature module malfunctioned (rare)
- 4. Main board malfunctioned (rare)
11 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 measurements inaccurate patient safety
SpO2 readings are displayed but values are incorrect or unreliable
Ranked causes
- 1. Sensor faulty or incorrectly attached (common)
- 2. Sensor contaminated or dirty (occasional)
- 3. SpO2 module malfunctioned (occasional)
- 4. Main board malfunctioned (rare)
12 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 measurements not displayed patient safety
SpO2 values do not appear on the monitor display
Ranked causes
- 1. Sensor faulty or incorrectly attached (common)
- 2. Sensor contaminated or dirty (occasional)
- 3. SpO2 module malfunctioned (occasional)
- 4. Main board malfunctioned (rare)
12 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature module not functional
Temperature measurement fails with error codes 3FFF01-3FFF04 indicating sensor communication or processing failure
Ranked causes
- 1. Unrecognized WACP parameter from sensor (unknown)
- 2. Timeout waiting for sensor response (unknown)
- 3. Error deserializing WACP message from sensor (unknown)
- 4. WACP stack send message failure (unknown)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery operating time
Battery performance depends on display brightness, sleep mode, and automatic power-down settings
Ranked causes
- 1. Suboptimal display brightness setting (common)
- 2. Sleep mode not configured (common)
- 3. Automatic power-down disabled or misconfigured (common)
Verify display brightness setting
- Test point
- Display brightness control
- Expected
- 5
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery performance optimization
Extend battery operating time through recommended display, sleep mode, and power-down settings
Ranked causes
- 1. Suboptimal display brightness setting (common)
- 2. Sleep mode not configured for energy efficiency (common)
- 3. Automatic power-down interval too long (common)
Verify display brightness setting
- Test point
- Display brightness menu
- Expected
- 5
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device does not signal alarm conditions for Nurse Call patient safety
Alarm notifications not transmitted to Nurse Call system despite alert conditions
Ranked causes
- 1. Alarm limit settings too high or too low (common)
- 2. Audio alarms silenced (common)
- 3. Hardware component failure (occasional)
Verify alarm limit settings
- Test point
- Device alarm settings menu
- Expected
- Alarm limits set appropriately for monitored parameters
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor cannot connect to network
System unable to establish network connectivity; Ethernet port unresponsive or unreachable
Ranked causes
- 1. Network connectivity issue (common)
- 2. Hardware failure in network interface (occasional)
- 3. Software failure in network stack (occasional)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor cannot connect with wireless network
Device fails to establish wireless network connection
Ranked causes
- 1. Connectivity issue (common)
- 2. Hardware component failure (occasional)
- 3. Software failure (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touchscreen unresponsive
Touchscreen does not respond to touch inputs or responds inconsistently
Ranked causes
- 1. Touchscreen needs calibration (common)
- 2. LCD/touchscreen hardware failure (occasional)
- 3. Main board defective (rare)
Run Host Calibration Sequence
- Test point
- Service Tool on connected PC
- Expected
- Calibration completes successfully and touchscreen responds to touch
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
All tests for device not run with Service Tool
Device fails to complete diagnostic test suite when connected to Welch Allyn Service Tool
Ranked causes
- 1. Device not connected to power supply (common)
- 2. Initialization not completed (common)
Verify power connection
- Test point
- electrical outlet and power supply connector
- Expected
- Device plugged into outlet and power supply connected to device
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device powered on but display is blank
Device powers on normally but screen shows no image or content
Ranked causes
- 1. Device in Sleep mode (common)
- 2. Hardware component failure (occasional)
Attempt to wake device
- Test point
- Touchscreen surface
- Expected
- Display activates and shows normal boot or operating screen
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermometer not functioning or charging
Thermometer does not operate, charge, or communicate with monitor
Ranked causes
- 1. The thermometer contacts are dirty (common)
- 2. The thermometer batteries are dead (common)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Advanced settings unavailable
Advanced settings cannot be accessed because monitor is not in idle state
Ranked causes
- 1. Monitor not in idle state (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ambient temperature out of range - high
Temperature measurement above maximum allowable ambient or patient temperature threshold
Ranked causes
- 1. Ambient temperature above 104 °F or 40 °C (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ambient temperature out of range - low
Temperature measurement below minimum allowable ambient or patient temperature threshold
Ranked causes
- 1. Ambient temperature below 50 °F or 10 °C (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
APM battery absent or faulty (error 1C100F)
APM battery is not installed or not communicating
Ranked causes
- 1. APM battery missing or not installed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
APM battery not communicating (error 1C100B)
APM battery is installed but does not communicate with the monitor
Ranked causes
- 1. APM battery not properly initialized or connected to mains (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
APM battery not recharging (error 1C100B)
APM battery is not recharging
Ranked causes
- 1. APM battery charging circuit failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
APM is disconnected (error 1C1002)
APM is unplugged from monitor while powered on
Ranked causes
- 1. APM unplugged from monitor during operation (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
APM not functional (error 1C1001)
APM is detected as connected but no communication through APM serial port
Ranked causes
- 1. No communication through APM serial port (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
APM not functional (error 1C1010)
APM USB hub plugged in while monitor communication cable is disconnected
Ranked causes
- 1. APM USB hub plugged in without communication cable (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
APM PIC cannot communicate with accelerometer (error 1C1004)
APM PIC fails to communicate with the accelerometer
Ranked causes
- 1. APM PIC communication failure with accelerometer (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
APM plugged in during operation
APM was plugged in while the monitor is powered on
Ranked causes
- 1. APM plugged in with monitor powered on (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
APM software update failed (error 1C1009)
APM PIC software update and retries have failed
Ranked causes
- 1. APM PIC software update failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
APM USB changes after startup
APM USB changes from unplugged to plugged after monitor startup
Ranked causes
- 1. APM USB connection unstable after startup (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio system not functional patient safety
Speaker or audio codec is faulty; no audio output
Ranked causes
- 1. Speaker or audio codec failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery critically low; device shutting down patient safety
Battery is critically depleted; monitor shutting down
Ranked causes
- 1. Battery power critically depleted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery replacement caution
Battery service after specified charge cycles to maintain device operation
Ranked causes
- 1. Battery reached 300 charge cycles (common)
Count battery charge cycles or assess battery performance degradation
- Test point
- Battery
- Expected
- Less than 300 charge cycles
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bluetooth device disconnected
Bluetooth device disconnects from monitor
Ranked causes
- 1. Bluetooth device disconnected (unknown)
Bluetooth device failed to pair with monitor
Device cannot establish pairing connection with the monitor via Bluetooth
Ranked causes
- 1. Bluetooth stack on PC does not support simple secure pairing (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery performance
Verify battery operating time under recommended settings and conditions
Ranked causes
- 1. Battery operating time affected by display and power settings (common)
Display brightness setting
- Test point
- Device display settings
- Expected
- 5
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery performance and operating time optimization
Extend battery operating time by configuring display brightness, sleep mode, and automatic power-down settings
Ranked causes
- 1. Suboptimal battery settings affecting operating time (common)
Verify display brightness setting
- Test point
- Display brightness setting
- Expected
- 5
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Braun ThermoScan PRO 4000 patient safety
Annual functional verification to confirm device meets design specifications
Ranked causes
- 1. Routine maintenance per service interval (common)
Verify Braun ThermoScan PRO 4000 functions within design specifications
- Test point
- Braun ThermoScan PRO 4000
- Expected
- Meets design specifications
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: First Boot Settings configuration
After repair and calibration, device can be placed into First Boot states using Service Tool Configuration screen
Ranked causes
- 1. Device repair and calibration complete (common)
Verify repair completion and calibration
- Test point
- Device status
- Expected
- All repairs completed and device calibrated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP board initialization
New PCA boards must be initialized for first use after installation
Ranked causes
- 1. New PCA board installed (common)
Attach pump assembly to main PCA board
- Test point
- Pump manifold on main PCA board transducer
- Expected
- Pump manifold fully seated on transducer
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP module patient safety
Annual functional verification to confirm device meets design specifications
Ranked causes
- 1. Routine maintenance per service interval (common)
Verify NIBP module functions within design specifications
- Test point
- NIBP module
- Expected
- Meets design specifications
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SpO2 module patient safety
Annual functional verification to confirm device meets design specifications
Ranked causes
- 1. Routine maintenance per service interval (common)
Verify SpO2 module functions within design specifications
- Test point
- SpO2 module
- Expected
- Meets design specifications
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SureTemp Plus patient safety
Annual functional verification to confirm device meets design specifications
Ranked causes
- 1. Routine maintenance per service interval (common)
Verify SureTemp Plus functions within design specifications
- Test point
- SureTemp Plus
- Expected
- Meets design specifications
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Client certificate expires within 30 days
Certificate approaching expiration date
Ranked causes
- 1. Certificate close to expiration (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Client certificate load failed
Certificate dates misaligned or outside valid date range
Ranked causes
- 1. Certificate dates misaligned (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Client certificate not loaded
Client authentication enabled but no certificate loaded in system
Ranked causes
- 1. Client authentication enabled without certificate (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CSM battery not installed patient safety
No battery installed in the monitor
Ranked causes
- 1. Missing battery (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device operating in battery mode
AC power cord has been disconnected
Ranked causes
- 1. AC power cord disconnected (unknown)
Device shutdown not available caution
Monitor cannot perform immediate shutdown; AC power and battery must be disconnected
Ranked causes
- 1. System shutdown failure (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display lock unavailable without patient context
Lockout is not allowed without active patient information
Ranked causes
- 1. Missing patient information (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
EEPROM access failed
Internal hardware failure; device boots but wired communications disabled
Ranked causes
- 1. EEPROM access failure (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Factory default settings now active
Monitor configured to factory defaults; user settings reset
Ranked causes
- 1. Factory reset applied (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Host battery not charging patient safety
Monitor battery not charging when connected to power
Ranked causes
- 1. Battery charging circuit failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Input voltage too low; device shutting down
- LCD touchscreen integration verification
- LCD touchscreen replacement
- Low battery 30 minutes or less remaining
- Low battery 5 minutes or less remaining
- Low RRp confidence
- Low SpO2 perfusion index
- Monitor PIC communications failure
- NIBP primary transducer failure risk
- NIBP safety transducer calibration error risk
- No valid files on USB flash drive
- Power cable disconnected (error 1C1011)
- Power supply issue; PMIC too hot
- Pressure Calibration Test setup and adjustment
- Probe well missing
- Replace temperature probe
- Root file system corrupted
- Searching for pulse signal
- Sensor not connected
- Service tool functional verification and calibration
- Sleep mode unavailable during intervals
- Sleep mode unavailable with active alarm
- Software update failed
- SPL memory test failure
- SpO2 defective sensor (044100)
- SpO2 interference detected
- SpO2 low PR confidence
- SpO2 low signal quality
- SpO2 module hardware error (043d00)
- SpO2 module hardware error (043e00)
- SpO2 module received bad message (044000)
- SpO2 module software error (043f00)
- SpO2 rebooting
- SpO2 rebooting with low PR confidence
- SpO2 sensor not connected
- System failure (codes 000001-000006)
- Temperature A/D measurement timeout
- Temperature probe not detected
- The thermometer is not in the cradle
- Thermometer anti-theft timer expired
- Tube type does not match device configuration
- Unable to read configuration from USB
- Unexpected restart occurred
- USB cable disconnected (error 1C1003)
44 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
000001 | System failure | Unspecified system fault detected | error |
000002 | System failure | Unspecified system fault detected | error |
000003 | System failure | Unspecified system fault detected | error |
000004 | System failure | Unspecified system fault detected | error |
000005 | System failure | Unspecified system fault detected | error |
000006 | System failure | Unspecified system fault detected | error |
030105 | Temperature not functional - CRC mismatch | CRC checksum mismatch on temperature module communication | error |
030201 | Temperature not functional - unimplemented message | Temperature module received unimplemented message type | error |
030202 | Temperature not functional - unsupported message | Temperature module received unsupported message type | error |
030203 | Temperature not functional - out of memory | Temperature module memory exhausted | error |
030205 | Temperature not functional - invalid parameter | Temperature module received invalid parameter | error |
030206 | Temperature not functional - parameter out of range | Parameter provided by temperature module outside allowable range | error |
030207 | Temperature not functional - missing object | Message requires object but object not contained in message | error |
030208 | Temperature not functional - deserialization failed | Temperature module object could not be deserialized | error |
030209 | Temperature not functional - serialization failed | Temperature module object could not be serialized | error |
03020A | Temperature not functional - invalid state request | Module attempted request/action prohibited by current module state | error |
03020B | Temperature not functional - item unavailable | Requested item not available due to current module state | error |
030503 | Temperature not functional - corrupted factory settings | Factory settings and calibration information corrupted | error |
030504 | Temperature not functional - corrupted user settings | User settings in module memory are corrupted | error |
030509 | Temperature not functional - calibration not set | Temperature module calibration has not been performed | error |
03050C | Temperature not functional - error log corrupted | Error log data structure in module memory corrupted | error |
030516 | Temperature not functional - hardware malfunction | Hardware malfunction detected on temperature module | error |
030518 | Temperature not functional - power rail too low | Temperature module power rail voltage below minimum threshold | error |
030519 | Temperature not functional - power rail too high | Temperature module power rail voltage above maximum threshold | error |
03051A | Temperature not functional - reference voltage unstable | Reference voltage circuit under voltage or unstable | error |
030801 | Ambient temperature out of range - low | Ambient temperature measurement below 50 °F or 10 °C | warning |
030802 | Ambient temperature out of range - high | Ambient temperature measurement above 104 °F or 40 °C | warning |
030803 | Temperature not functional - calibration resistor damaged (long pulse) | Internal calibration resistor RCAL pulse too long | error |
030804 | Temperature not functional - calibration resistor damaged (short pulse) | Internal calibration resistor RCAL pulse too short | error |
030805 | Temperature not functional - validation resistor over | Internal validation resistor PTB value over specification | error |
030806 | Temperature not functional - validation resistor under | Internal validation resistor PTB value under specification | error |
030807 | Ambient temperature out of range - A/D timeout | Analog-to-digital measurement timed out | warning |
030808 | Replace temperature probe - not calibrated | Temperature probe was not characterized or calibrated | warning |
030809 | Insert probe well - missing probe well | Temperature module probe well not inserted | warning |
03080A | Temperature not functional - EEPROM save failed | Problem saving to monitor EEPROM in biotech mode | error |
03080B | Temperature not functional - error detection mechanism triggered | Module error detection mechanism detected error | error |
03080C | Replace temperature probe - probe error detected | Probe error detection mechanism triggered | warning |
03080D | Temperature not functional - log error detected | Log error detection mechanism detected error | warning |
03080E | Temperature not functional - calibration error detected | Calibration error detection mechanism triggered | warning |
03080F | Connect temperature probe - no probe detected | Temperature module detected no probe connected | warning |
030810 | Replace temperature probe - EEPROM unreadable | Cannot read probe EEPROM or probe untested from factory | warning |
030811 | Temperature not functional - invalid events index | Temperature module has invalid events index | error |
030812 | Temperature not functional - EEPROM read/write error | Problem reading temperature module EEPROM or saving to monitor | error |
030813 | Replace temperature probe - probe EEPROM error | Problem reading the probe EEPROM | warning |
030814 | Temperature not functional - config acquire failure | Temperature module config acquire failure | error |
030815 | Temperature not functional - config release failure | Temperature module config release failure | error |
030816 | Temperature not functional - invalid pointer | Temperature module config invalid pointer failure | error |
030817 | Temperature not functional - EEPROM not initialized | Temperature module EEPROM not initialized | error |
030818 | Unable to detect new temperature. Retry measurement. | Temperature module heater indicates on when turned off | warning |
030819 | Unable to detect new temperature. Retry measurement. | Temperature module heater indicates off when turned on | warning |
03081A | Temperature not functional | Temperature module HTR_Q is on and HTRC is off but still has voltage | warning |
03081B | Temperature not functional | Temperature module HTR_Q is tri-stated with HTRC enabled and has heater power | warning |
03081C | Temperature not functional | Temperature module turned Q&C on and the heater voltage is not high enough | warning |
03081D | Temperature not functional | Temperature module heater hardware failsafe should have turned off but did not | warning |
03081E | Replace temperature probe | Temperature module probe is above 112°F or 43.3°C | warning |
03081F | Replace temperature probe | Temperature module has excessive heater energy | warning |
030820 | Temperature not functional | Temperature module host interface error | warning |
030821 | Ambient temperature out of range. Clear to retry. | Temperature module over ambient 45°C temp | warning |
030822 | Ambient temperature out of range. Clear to retry. | Temperature module under ambient temp | warning |
030823 | Temperature not functional | Temperature module has an invalid SureTemp algorithm | warning |
030824 | Temperature not functional | Temperature module is over max battery volts | warning |
030825 | Temperature not functional | Temperature module is under min battery volts | warning |
030826 | Temperature not functional | Temperature module battery volts not set | warning |
030827 | Temperature not functional | Temperature module predict algorithm is not set | warning |
030828 | Temperature not functional | Temperature module ambient temp is not set | warning |
030829 | Temperature not functional | Temperature module has a non-responsive probe. Thermistor pulled away or heater broken. | warning |
03082A | Temperature not functional | Temperature module is experiencing bad probe gain | warning |
03082B | Temperature not functional | Temperature module has a bad probe response value | warning |
03C800 | Temperature not functional | Temperature module is not functional | warning |
03C900 | Temperature not functional | Unable to deserialize messages from the temperature module | warning |
03CA00 | Temperature not functional | Unsupported message received from the Temperature module | warning |
03CB00 | Temperature not functional | Unable to send message to the Temperature module | warning |
03CC00 | Temperature not functional | Temperature module communication times out | warning |
03CD00 | Temperature not functional | Failed to upgrade the temperature module | warning |
03CE00 | Temperature not functional | Unable to read PIM file | warning |
03CE01 | Temperature not functional | Upgrade file directory could not be accessed | warning |
03D000 | Temperature module reset | Temperature sensor reset unexpectedly | info |
040100 | Sensor not connected | SpO2 sensor disconnected or connection lost | warning |
040200 | Searching for pulse signal | SpO2 module actively searching for patient pulse | info |
040400 | SpO2 interference detected | Electromagnetic or optical interference detected by sensor | warning |
040500 | Low SpO2 perfusion index | Marginal pulse quality or artifact detected | warning |
040600 | SpO2 cable not connected | SpO2 cable is not connected | warning |
040700 | SpO2 cable life expired | SpO2 cable life has expired | warning |
040800 | SpO2 cable incompatible | SpO2 cable is incompatible with monitor | warning |
040900 | SpO2 cable not recognized | SpO2 cable is not recognized by monitor | warning |
040a00 | SpO2 cable defective | SpO2 cable is defective | warning |
040b00 | SpO2 sensor not connected | SpO2 sensor not connected to monitor | warning |
040c00 | SpO2 sensor expired | SpO2 sensor has expired | warning |
040d00 | SpO2 sensor not recognized | SpO2 sensor not recognized by monitor | warning |
040e00 | SpO2 sensor not recognized | SpO2 sensor is not recognized | warning |
040f00 | SpO2 sensor defective | SpO2 sensor is defective | warning |
041000 | SpO2 sensor and cable fault | Fault with both SpO2 sensor and cable | warning |
041100 | Adhesive SpO2 sensor not connected | Adhesive SpO2 sensor not connected | warning |
041200 | Adhesive SpO2 sensor expired | Adhesive SpO2 sensor has expired | warning |
041300 | Adhesive SpO2 sensor incompatible | Adhesive SpO2 sensor is incompatible | warning |
041400 | Adhesive SpO2 sensor not recognized | Adhesive SpO2 sensor not recognized | warning |
041500 | Adhesive SpO2 sensor defective | Adhesive SpO2 sensor is defective | warning |
041800 | SpO2 pulse search active | SpO2 pulse search normal operation | info |
041900 | SpO2 interference detected | SpO2 module interference detected | info |
041a00 | Low SpO2 perfusion index | Marginal SpO2 pulse quality or artifact | warning |
041b00 | SpO2 demo mode active | SpO2 parameter is in demo mode | info |
041c00 | SpO2 sensor connection issue | Check SpO2 sensor connection | warning |
041e00 | SpO2 raw queue overrun | SpO2 raw queue overrun detected | error |
041f00 | SpO2 hardware failure | SpO2 hardware failure detected | error |
042000 | SpO2 MCU failure | SpO2 MCU failure detected | error |
042100 | SpO2 watchdog failure | SpO2 watchdog timer failure | error |
042200 | SpO2 invalid board type | Invalid SpO2 board type detected | error |
042300 | Invalid SpO2 master control state | Invalid master control state detected | error |
042400 | SpO2 SRAM transfer failure | SRAM transfer failure detected | error |
042500 | SpO2 SRAM task queue overrun | SRAM task queue overrun detected | error |
042600 | SpO2 database failure | Database failure detected | error |
042700 | Invalid SpO2 flash memory device | Invalid flash memory device detected | error |
042800 | SpO2 anode voltage configuration failure | Anode voltage configuration failure detected | error |
042900 | SpO2 analog ground problem | Analog ground voltage out of spec | error |
042a00 | SpO2 digital ground problem | Digital ground voltage out of spec | error |
042b00 | SpO2 LED ground problem | LED ground voltage out of spec | error |
042c00 | SpO2 reference voltage problem | Reference voltage out of spec | error |
042d00 | SpO2 DSP core voltage problem | DSP core voltage out of spec | error |
042e00 | SpO2 filtered input voltage problem | Filtered input voltage out of spec | error |
042f00 | SpO2 DSP I/O voltage problem | DSP I/O voltage out of spec | error |
043000 | SpO2 positive detector voltage problem | Positive detector voltage out of spec | error |
043100 | SpO2 negative detector voltage problem | Negative detector voltage out of spec | error |
043200 | SpO2 positive LED voltage problem | Positive LED voltage out of spec | error |
043300 | SpO2 LED drive voltage problem | LED drive voltage out of spec | error |
043400 | SpO2 positive preamp voltage problem | Positive preamp voltage out of spec | error |
043500 | SpO2 Sensor ID problem | Sensor ID detection failure | error |
043600 | SpO2 thermistor problem | Thermistor circuit failure detected | error |
043700 | SpO2 LED current problem | LED current out of spec | error |
043800 | SpO2 preamp problem | Preamp circuit malfunction detected | error |
043900 | Sensor not connected | SpO2 sensor is not detected as physically connected | warning |
043a00 | Searching for pulse signal | Module is acquiring sensor signal after connection or signal loss | info |
043c00 | SpO2 interference detected | Electromagnetic or optical interference detected by module | warning |
043d00 | SpO2 module hardware error | Module detects internal hardware failure | critical |
043e00 | SpO2 module hardware error | Module detects internal hardware failure | critical |
043f00 | SpO2 module software error | Module detects internal software error during operation | error |
044000 | SpO2 module received bad message | Module receives malformed or corrupt communication message | error |
044100 | SpO2 defective sensor | Module detects sensor as non-functional or defective | warning |
044200 | SpO2 module rebooting | SpO2 module received bad/corrupted message | warning |
044300 | SpO2 module received bad packet | Bad data packet received by SpO2 module | error |
044400 | SpO2 module received invalid command | Invalid command received by module | error |
044500 | SpO2 output exceeds baud rate capacity | Command output exceeds baud rate limit | error |
044600 | SpO2 command requires missing application | Command received for application not present | error |
044700 | SpO2 command received while locked | Command received while module is locked | error |
044900 | SpO2 module not responding | Internal hardware malfunction in SpO2 module | error |
044a00 | SpO2 module rebooting - bad response | SpO2 module is not responding | info |
044b00 | SpO2 module rebooting - no data | SpO2 module stopped sending data | info |
044c00 | SpO2 rebooting - bad CRC | Packet received with bad CRC from module | info |
044d00 | SpO2 module rebooting - POST failed | SpO2 power on self test failed | error |
044e00 | SpO2 module rebooting - POST timeout | SpO2 power on self test timed out | error |
044f00 | SpO2 low saturation signal quality | Sensor signal quality falls below acceptable threshold | warning |
045000 | SpO2 rebooting with low pulse rate confidence | Module reboots while detecting low pulse rate confidence | warning |
045100 | Low pulse rate confidence (PI confidence) | Pulse interval confidence falls below acceptable threshold | warning |
045200 | Low respiratory rate confidence | Respiratory rate parameter confidence falls below threshold | warning |
050002 | NIBP not functional | Internal malfunction in NIBP module | error |
050003 | Unable to determine NIBP; check connections; limit patient movement | NIBP measurement may be inaccurate, patient motion occurred, or settings for patient readings not accurate | error |
050004 | Unable to determine NIBP; check connections; limit patient movement. | Excessive artifact, no blood pressure parameters could be calculated | warning |
050005 | Unable to determine NIBP; check inflation settings | Low inflation on blood pressure measurement attempt | warning |
050006 | Unable to determine NIBP; check connections and tubing for kinks. | NIBP tubing has a kink or transducer calibration failure | error |
050007 | Unable to determine NIBP; check connections; limit patient movement. | Blood pressure measurement dumped too soon | warning |
050008 | Unable to determine NIBP; check connections; limit patient movement. | Not enough steps on the measurement attempt | warning |
050009 | Unable to determine NIBP; check inflation settings. | Invalid patient information for the selected mode | error |
05000A | Unable to determine NIBP; check connections; limit patient movement. | Re-inflation was too late on the measurement attempt | warning |
05000B | Unable to determine NIBP; check inflation settings. | Many re-inflate attempts on the measurement attempt | warning |
05000C | Unable to determine NIBP; check connections and tubing for kinks. | Couldn't bleed pressure below safe venous return pressure | error |
05000D | NIBP air leak; check cuff and tubing connections. | Leak was detected in BP cycle | warning |
05000F | Unable to determine NIBP; check connections; limit patient movement. | Safety check failed on the measurement attempt or Auto Zero check failure | error |
050105 | NIBP not functional. | WACP message CRC mismatch on NIBP module | warning |
050201 | NIBP not functional. | This message is not implemented by the NIBP module | warning |
050202 | NIBP not functional. | This message is not supported by the NIBP module | warning |
050203 | NIBP not functional. | The NIBP module has run out of memory | warning |
050205 | NIBP not functional. | The NIBP module has received an invalid parameter | warning |
050206 | NIBP not functional. | The parameter provided by the NIBP module is outside of the allowable range | warning |
050207 | NIBP not functional. | The NIBP module message requires an object, but did not contain one | warning |
050208 | NIBP not functional. | The NIBP module object could not be deserialized | warning |
050209 | NIBP not functional. | The NIBP module object could not be serialized | warning |
05020A | NIBP not functional. | The NIBP module message is performing a request when module state prohibits it | warning |
050503 | NIBP not calibrated. | Factory EEPROM checksum error on NIBP, units internal configuration corrupted | warning |
050504 | NIBP not functional. | User EEPROM checksum error, configuration data damaged or lost | warning |
050505 | NIBP not functional. | Post failure of A/D convertor | warning |
050509 | NIBP not calibrated. | NIBP module calibration failure, calibration signature is zero | warning |
05050A | Invalid algorithm. Select correct algorithm and retry. | Invalid NIBP Algorithm, sensor configured illegally | warning |
050513 | NIBP not functional. | Invalid NIBP initiation code | warning |
050514 | Invalid patient mode. Select correct patient mode and retry. | Invalid patient mode on NIBP, sensor configured illegally | warning |
050515 | NIBP not functional. | Invalid module configuration for NIBP | warning |
050516 | NIBP not functional. | NIBP module malfunction | warning |
050517 | Ambient temperature out of range. Clear error and retry. | Ambient temp out of range on NIBP | warning |
050518 | Low battery. Plug into outlet. | The NIBP module power rail is too low | warning |
0505191 | Battery overcharged. Disconnect from outlet. | The NIBP module power rail is too high | warning |
050601 | NIBP not calibrated. | NIBP failed to load safety processors calibration record from EEPROM | warning |
050602 | NIBP not functional. | NIBP safety processor failed ROM checksum | warning |
050603 | NIBP not calibrated. | NIBP safety processor not calibrated, missing calibration signature | warning |
050604 | Cuff pressure limits exceeded. | NIBP system failure, overpressure | error |
050605 | Premature auto cycle skipped. | NIBP auto cycle skipped, SVRP requirement not satisfied | warning |
050606 | Cuff pressure too high. Clear error to retry. | NIBP cuff pressure above SVRP for too long | error |
050607 | NIBP not functional. | NIBP cannot clear the failsafe alerts | warning |
050608 | NIBP not functional. | NIBP safety processor has stopped responding | warning |
050609 | Stat mode requested too soon. Clear to retry. | NIBP excessive stat mode time, readings exceed 15 minutes averaging cycle | warning |
05060A | Unable to determine NIBP; check connections and tubing for kinks. | NIBP transducers are not matched, pressure difference exceeds 40 mmHg | error |
05060B | NIBP not calibrated. | Factory EEPROM checksum error on NIBP, internal configuration corrupted | warning |
05060C | NIBP not functional. | NIBP command not implemented | warning |
05060D | NIBP not functional. | NIBP wrong data count | warning |
05060E | NIBP not functional. | NIBP data range error | warning |
05060F | NIBP not functional. | NIBP no POST error to clear | warning |
050610 | NIBP not functional. | NIBP cannot clear this POST error | warning |
050611 | NIBP not functional. | NIBP command not command type | warning |
050612 | NIBP not functional. | NIBP communication timeout | warning |
050613 | NIBP not functional. | NIBP response header wrong | warning |
050614 | NIBP not functional. | NIBP response checksum wrong | warning |
050615 | NIBP not functional. | Too much NIBP data was received | warning |
050616 | NIBP not functional. | NIBP FPROM erase error | warning |
050617 | NIBP not functional. | NIBP FPROM programming error | warning |
050618 | NIBP not functional. | Invalid NIBP target pressure | warning |
05FF01 | NIBP not functional. | Unrecognized WACP parameter received from sensor | warning |
05FF02 | NIBP not functional. | Timeout waiting for sensor response | warning |
05FF03 | NIBP not functional. | Error deserializing the WACP message received from the sensor | warning |
05FF04 | NIBP not functional. | WACP stack send message failure | warning |
05FF05 | NIBP not functional. | Timeout waiting for asynchronous sensor message | warning |
05FF06 | NIBP not functional. | One or more undetermined numerics when reading status indicates OK | warning |
05FF07 | NIBP not functional. | Unrecognized sensor reading status code | warning |
05FF08 | NIBP not functional. | Sensor power up failure | warning |
05FF09 | NIBP not functional. | WACP rendezvous failure | warning |
05FF0A | NIBP not functional. | Application firmware retrieval error during POST | warning |
05FF0B | NIBP not functional. | Upgrade .pim file is corrupt | warning |
05FF0C | NIBP not functional. | Configured upgrade firmware directory cannot be accessed | warning |
05FF0D | Device configuration issue. | Configured parameter (NIBP or SpO2) used in Intervals is missing | warning |
05FF0E | NIBP not functional. | The NIBP sensor unexpectedly reset | warning |
05FF0F | NIBP not functional. | The NIBP sensor firmware failed to upgrade | warning |
140100 | EEPROM access failed | EEPROM cannot be accessed during boot | error |
1C1000 | Monitor PIC communications failure | PIC interface communication never starts or quits during operation | error |
1C1001 | APM not functional | APM is detected as connected but there is no communication through the APM serial port | error |
1C1002 | The APM is disconnected | The APM is unplugged from the monitor while the monitor is powered on | error |
1C1003 | USB cable is disconnected | The APM USB hub is unplugged from the monitor while the monitor is powered on | error |
1C1004 | APM not functional | The APM PIC cannot communicate with the accelerometer | error |
1C1005 | Low battery 30 minutes or less remaining | Battery charge drops to critical level | warning |
1C1006 | Low battery 5 minutes or less remaining | Battery charge critically low | critical |
1C1007 | Battery critically low; device shutting down | Battery power at critical minimum; shutdown initiated | critical |
1C1008 | Update unsuccessful; reboot and retry | Software update process fails to complete | error |
1C1009 | APM not functional | The APM PIC software update and any retries have failed | error |
1C100A | Host battery not charging | Battery charge circuit inactive during AC power connection | warning |
1C100B | APM not functional - battery not communicating | APM battery is installed but does not communicate with the monitor | error |
1C100C | Input voltage too low; device shutting down | PMIC input voltage falls below minimum threshold | critical |
1C100D | Internal hardware failure; device shutting down | PMIC overheats beyond safe operating limit | critical |
1C100E | CSM battery is not installed | No battery detected during startup | warning |
1C100F | APM battery absent or faulty | The APM battery is not installed | error |
1C1010 | APM not functional | The APM USB hub is plugged in while the monitor communication cable is disconnected | error |
1C1011 | Power cable is disconnected | The APM communication cable is plugged in while the APM USB cable is disconnected | error |
1C1012 | Unexpected restart occurred | Monitor spontaneously restarts without operator action | warning |
1D0100 | Audio system not functional | Speaker or audio codec fails to output sound | warning |
1F0001 | Database unreadable during startup; recreating | Database file corrupted at system startup | warning |
1F0002 | Error accessing PDM database; restarting PDM | Database corruption detected during device operation | warning |
3A0001 | Factory default settings now active | Monitor configuration reset to factory defaults | info |
3A0002 | Unable to read configuration from USB | File load from external USB memory device fails | warning |
3F0105 | Temperature not functional | WACP message CRC mismatch | warning |
3F0201 | Temperature not functional | This message is not implemented by the module | warning |
3F0202 | Temperature not functional | This message is not supported by the module | warning |
3F0203 | Temperature not functional | The module has run out of memory | warning |
3F0204 | Temperature not functional | No parameter provided for the specified message | warning |
3F0205 | Temperature not functional | The parameter provided is invalid for the specified message | warning |
3F0206 | Temperature not functional | The parameter provided is outside of the allowable range | warning |
3F0207 | Temperature not functional | The message requires an object, but did not contain one | warning |
3F0208 | Temperature not functional | The object provided with the message could not be deserialized | warning |
3F0209 | Temperature not functional | The object could not be serialized | warning |
3F020A | Temperature not functional | The message is performing a request/action when module state prohibits it | warning |
3F020B | Temperature not functional | The requested item is not currently available due to module state | warning |
3F0503 | Temperature not functional | The factory settings and calibration information is corrupt | warning |
3F0504 | Temperature not functional | The user settings are corrupt | warning |
3F0509 | Temperature not functional | The calibration is not set | warning |
3F050C | Temperature not functional | The error log is corrupt | warning |
3F0516 | Temperature not functional | A hardware malfunction has been detected | warning |
3F0518 | Temperature not functional | The module power rail is too low | warning |
3F0519 | Temperature not functional | The module power rail is too high | warning |
3F051A | Temperature not functional | The reference voltage circuit was detected to be under voltage or unstable | warning |
3F0821 | Temperature not functional | The ambient temperature is too high | warning |
3F0822 | Temperature not functional | The ambient temperature is too low | warning |
3F0824 | Temperature not functional | The battery is over the maximum voltage | warning |
3F0833 | Temperature not functional | The sensor not functional | warning |
3F0E04 | Temperature not functional | Low battery | warning |
3FFF01 | Unrecognized WACP parameter from temperature sensor | Temperature module receives invalid sensor parameter | warning |
3FFF02 | Timeout waiting for temperature sensor response | Temperature sensor does not respond within timeout period | warning |
3FFF03 | Error deserializing WACP message from temperature sensor | Temperature module cannot parse sensor message | warning |
3FFF04 | WACP stack send message failure to temperature sensor | Temperature module fails to send message to sensor | warning |
3FFF05 | Re-dock Braun - Anti-theft timer expired | Thermometer not re-docked after measurement within timeout | info |
Bluetooth device connection successful | The Bluetooth device connected successfully | Bluetooth device successfully established connection | info |
Bluetooth not functional. 370001 | Monitor detected a Bluetooth device that is not functional | Bluetooth device detected but not functioning properly | warning |
Bluetooth not functional. 370002 | Monitor cannot detect a Bluetooth module | Bluetooth module not detected during initialization | warning |
Check cuff inflation settings. | Cuff inflation target overridden due to max pressure being too low | Cuff inflation target configuration issue | info |
Configuration upload successful. | Device configuration has successfully uploaded | Completion of configuration upload operation | info |
Database schema out of date; recreating | Database schema requires update and recreation | Database version mismatch on startup | info |
Factory reset failed. | Monitor factory reset has failed; custom configuration file not deleted | Attempted factory reset operation | info |
Factory reset successful. | Monitor was reset to factory settings | Completion of factory reset operation | info |
Radio software upgrade failed. 350600 | The radio software was not upgraded successfully | Radio software upgrade process fails | warning |
Save not successful. | Device configuration or logs were not saved to the USB device | Attempted to save device configuration or logs to USB device | info |
Save successful. | Device configuration or logs were saved to the USB device | Successful completion of save operation to USB device | info |
Software Update: Installation failed. Reboot and retry | Failure to install subsystem during software update | Subsystem installation fails during update process | info |
Software Update: Invalid token file. | Invalid token file encountered during software update | Software update process attempting to verify token file | info |
Software Update: Manager internal error | Software update manager internal error | Internal error in software update manager | info |
Software Update: Manifest transfer timed out. Verify connect | Manifest file transfer timed out or connection lost during download | Software update manifest file download | info |
Software Update: Package file corrupted | The package file is corrupt and does not have expected SHA256 hash | Software update process detects corrupt package | info |
Software Update: Package file transfer timed out. Verify con | Package file transfer timed out or connection lost during download | Software update package file download | info |
Software Update: SWUP internal error | SWUP NIBP is not functional | Internal error in software update SWUP module | info |
Software Update: Unable to find manifest file on server. | Manifest file was not found on the server | Software update process searching for manifest file on server | info |
Software Update: Unable to find package file | The package file cannot be found on the server | Software update process cannot locate package file | info |
Software Update: Unable to verify manifest file signature. | Manifest file signature verification failed | Software update process attempting to verify manifest file signature | info |
Software Update: Update unsuccessful. Incompatible firmware | Current firmware version is lower than required by token file | Firmware version check fails during update | info |
Software Update: Upgrade unsuccessful. Insufficient disk spa | The partition is running out of space during upgrade | Insufficient disk space detected during software upgrade | info |
Software upgrade is downloading. Do not shutdown. | Software installation is in progress; device cannot be powered down | Device attempting to power down while software install is in progress | info |
Tube type does not match device configuration. | Tube type settings and actual type do not match | Tube type mismatch between settings and actual configuration | info |
Unable to load configuration. | Device configuration was not successfully uploaded | Attempted to load device configuration | info |
User canceled NIBP reading. | The NIBP measurement was canceled by user | User action | info |
User canceled NIBP reading | User canceled NIBP reading | NIBP measurement was canceled by user | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.