Welch Allyn AT-10 Plus: troubleshooting and error codes
The ServiceForge knowledge base holds 42 documented symptoms for the Welch Allyn AT-10 Plus, with 61 ranked causes, 98 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Print quality degraded or misaligned
Printout shows fading, misalignment, faulty pixels, or irregular blackness across print width
Ranked causes
- 1. Thermal print head contamination or wear (common)
- 2. Low quality or expired paper installed (common)
- 3. Electrical or mechanical print head misalignment (occasional)
- 4. Printer mechanism defective (occasional)
Enter test screen and initiate continuous test print
- Test point
- Test screen
- Expected
- Continuous test print initiated
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mechanical condition degraded or safety labels missing caution
Cracks or chips in casing; damaged cables; loose connectors; unreadable or missing safety labels
Ranked causes
- 1. Physical damage from drops, impacts, or environmental exposure (occasional)
- 2. Cable insulation damaged by crushing, chafing, or cuts (occasional)
- 3. Safety label worn, faded, or detached (occasional)
Inspect external casing for cracks, chips, or impact damage
- Test point
- Entire device exterior
- Expected
- No visible cracks or chips in plastic or metal casing
7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer paper feed misaligned caution
Paper does not stop at perforation point after manual start/stop cycle
Ranked causes
- 1. Non-SCHILLER paper or wrong paper type used (common)
- 2. Paper detection opto window contaminated (occasional)
- 3. Paper mark detection circuit defective (rare)
Load SCHILLER paper and press Man Start
- Test point
- Paper transport mechanism
- Expected
- Paper feeds through printer
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software update fails or is interrupted during PC-based update caution
Unit software update via SWUP program does not complete; device may not restart automatically after download
Ranked causes
- 1. Unit powered off during update process (common)
- 2. PC COM port misconfiguration (occasional)
- 3. Incompatible operating system (occasional)
Verify unit power status during update
- Test point
- Unit power indicator
- Expected
- Unit powered on and AC adapter connected
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not operate caution
Device does not power on when switched with On/Off key
Ranked causes
- 1. Battery fully discharged (common)
- 2. Mains power not connected (common)
- 3. Damaged mains lead (occasional)
Verify mains power connection
- Test point
- Power cable connection at rear of unit
- Expected
- Cable connected and Mains indicator lamp lit
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery indicator lamp not lit
LED indicator (2) does not illuminate or battery symbol missing from LCD (4) when on battery power
Ranked causes
- 1. Battery not charged or discharged (common)
- 2. Battery capacity depleted (occasional)
Verify device powered by mains before disconnecting
- Test point
- Mains indicator LED (1)
- Expected
- Mains indicator lit
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG amplifier or patient cable defective patient safety
ECG leads show incorrect mV readings or excessive noise; missing or intermittent electrode signals
Ranked causes
- 1. Patient cable disconnected or loose connection (common)
- 2. ECG amplifier circuit defective or gain mismatch (rare)
Connect patient simulator and press ECG key
- Test point
- ECG input connector
- Expected
- Electrode screen displayed; mV readings visible for all leads
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG traces not displayed on screen
Patient simulator connected but ECG waveform does not appear on display
Ranked causes
- 1. Patient cable not properly connected (unknown)
- 2. Patient simulator not powered on (unknown)
Verify patient cable connection between simulator and unit
- Test point
- SCHILLER patient cable connectors
- Expected
- Cable properly seated and connected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical safety hazard during operation high voltage
Risk of electrical shock or leakage current when unit is not properly grounded or mains lead is damaged
Ranked causes
- 1. Missing or suspect earth connection (occasional)
- 2. Leakage current from connected external equipment (occasional)
Verify mains lead earth connection
- Test point
- Mains plug and wall socket
- Expected
- Earth pin properly inserted and secured in grounded socket
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External monitor not displaying or incorrect display settings
External VGA monitor shows no image or different data than LCD display after connection
Ranked causes
- 1. External Display config not enabled in settings (common)
- 2. Monitor cable not connected or VGA connector loose (occasional)
Connect external VGA monitor to back panel external monitor connector
- Test point
- External monitor VGA connector on back panel
- Expected
- Monitor cable seated firmly; monitor powered on
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard non-responsive or keys stuck caution
Function keys do not respond, no acoustic confirmation when pressed, or visible mechanical damage
Ranked causes
- 1. Mechanical damage or excessive wear on keyboard (occasional)
- 2. Key contacts dirty or non-conductive (occasional)
Inspect keyboard for cracks, chips, or visible wear
- Test point
- All keys on keyboard
- Expected
- No mechanical damage visible; keys operate freely
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No printout after automatic ECG recording
Automatic ECG recording does not produce printout after approximately 10 seconds
Ranked causes
- 1. Printer not operational or out of paper (unknown)
- 2. Auto Mode Format configuration not set (unknown)
Verify ECG was recorded and result displays on screen
- Test point
- LCD display
- Expected
- Result displayed on screen
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer output speed incorrect caution
Printed waveform spacing does not match expected calibration distances at specified print speeds
Ranked causes
- 1. Incorrect print speed setting (common)
- 2. Paper drive motor or speed control circuit defective (occasional)
Connect calibrated simulator with HR 60/min, no arrhythmias
- Test point
- ECG input from patient simulator
- Expected
- HR displays 60 on LCD
4 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Analogue ports not functioning
DC Out, DC In, QRS Out, Load Out, Load In, RPM In, Fuss In outputs not responding correctly during Ergo test
Ranked causes
- 1. Analogue port communication or signal integrity issue (unknown)
Connect DC In/Out test cable to analogue ports
- Test point
- DC OUT 1, DC OUT 2, DC IN 1, DC IN 2, QRS OUT, RPM IN, LOAD OUT, LOAD IN, FUSS IN, TM UP, TM DOWN
- Expected
- Signals transmit and receive without errors on all tested analogue channels
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery capacity insufficient
Device does not provide minimum one hour runtime on battery power after full charge
Ranked causes
- 1. Battery capacity degraded or defective (unknown)
Fully charge battery by leaving unit connected to mains
- Test point
- Mains power supply
- Expected
- Eight hours charging time
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery capacity limited
Battery LED blinking indicates low battery capacity; unit should be reconnected to mains supply for charging
Ranked causes
- 1. Internal battery not fully charged (common)
Observe Battery LED indicator status
- Test point
- Battery LED on unit casing
- Expected
- Blinking indicates limited capacity
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery replacement needed high voltage
Battery running time falls substantially under 2 hours; replacement required approximately every 4 years depending on application
Ranked causes
- 1. Battery aged or degraded capacity (common)
Verify actual running time is substantially under 2 hours
- Test point
- Device operation
- Expected
- Less than 2 hours running time indicates replacement needed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery stored for extended period
Battery recharge requirement after storage exceeding three months
Ranked causes
- 1. Battery discharge during storage (common)
Verify storage period
- Test point
- Device history
- Expected
- Storage duration exceeds three months
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibrate battery capacity
Battery capacity appears limited or inaccurate; recalibration needed after battery replacement or during service.
Ranked causes
- 1. Battery installed or capacity needs recalibration (common)
Unit powered only by battery until depleted
- Test point
- Battery voltage via hardware counter
- Expected
- Unit switches off when battery empty
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: battery replacement high voltage
Battery replacement procedure with calibration and capacity verification after installation
Ranked causes
- 1. Battery end of life or insufficient runtime (common)
Disconnect device from mains
- Test point
- Power inlet
- Expected
- Device unplugged
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Connected equipment not compliant with IEC standards patient safety
Ancillary equipment connected to analogue or digital interfaces does not meet IEC certification requirements
Ranked causes
- 1. Non-IEC certified equipment connected to interfaces (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Damaged mains lead or earth connection suspect high voltage
Mains lead is damaged, suspected of being damaged, or earth connection is suspect
Ranked causes
- 1. Defective or damaged mains cable (common)
Inspect mains lead for damage or degradation
- Test point
- Mains lead
- Expected
- No visible damage, cracks, or degradation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device opened without disconnecting from mains high voltage
Risk of electric shock when device is opened while connected to power
Ranked causes
- 1. Device opened while powered (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG printout waveform does not match reference patient safety
Printed ECG shape or polarity differs from standard reference; measurement values outside tolerance
Ranked causes
- 1. Auto mode settings incorrect or corrupted (occasional)
Connect simulator and select calibrated reference EN 60601-2-51 > CAL 20160
- Test point
- Simulator output configuration
- Expected
- Reference waveform selected
9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Faulty pixels on printout
ECG printout exhibits defective or missing pixels in the output
Ranked causes
- 1. Printer head defect or paper quality issue (unknown)
Inspect ECG printout for faulty or missing pixels
- Test point
- Printed ECG paper
- Expected
- No faulty pixels in printed ECG
Fuse needs replacement high voltage
Device requires fuse replacement due to blown fuse or routine maintenance
Ranked causes
- 1. Blown fuse (unknown)
Verify correct fuse type for mains voltage
- Test point
- Fuse assembly voltage indication window
- Expected
- Voltage range matches device setting: 220-240 VAC or 100-115 VAC
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fuse replacement with incorrect type or rating high voltage
Defective or ruptured fuse replaced with wrong type or rating
Ranked causes
- 1. Incorrect fuse type or rating installed (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Language, date, or time not set correctly
Device displays wrong language, date, or time; menu settings not saved
Ranked causes
- 1. User did not confirm language, date, or time selection (common)
Press Menu key to open system menu
- Test point
- Menu button on device
- Expected
- System menu displayed with tabs
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD screen defective or uneven contrast caution
Visible spots, black fields, or uneven contrast/brightness across display; backlight appears faulty
Ranked causes
- 1. LCD backlight failure or defective LCD panel (occasional)
Visually inspect LCD for dark spots or black fields
- Test point
- Entire LCD display area
- Expected
- No spots or minimal defects (few faulty pixels normal); even contrast and brightness
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD screen does not light
Display fails to illuminate or ECG acquisition screen not displayed after powering on
Ranked causes
- 1. Unit not powered on or startup delay (unknown)
Press On key and observe LCD screen illumination
- Test point
- LCD display
- Expected
- LCD screen lights and ECG acquisition screen is displayed after a few moments
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Liquid spilled on device high voltage
Liquid has been spilled over the AT-10 Plus ECG device
Ranked causes
- 1. Liquid contamination (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mains indicator lamp not lit
LED indicator (1) does not illuminate when power cable connected to mains supply
Ranked causes
- 1. Power cable not connected or loose connection (common)
Verify power cable connected to mains supply
- Test point
- Rear panel power connector
- Expected
- Mains indicator lamp (1) lit
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mains LED does not light
Green mains LED fails to illuminate when power is connected
Ranked causes
- 1. Mains power not connected or defective (unknown)
Verify mains power connection and check green mains LED illuminates
- Test point
- Mains LED indicator
- Expected
- Green mains LED lights
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mains voltage needs adjustment high voltage
Device mains voltage selector requires adjustment for different supply voltage
Ranked causes
- 1. Mains voltage setting incorrect for local supply (unknown)
Verify voltage indication after rotating inset
- Test point
- Voltage indication window on fuse assembly
- Expected
- Correct voltage range displayed: 220-240 VAC or 100-115 VAC
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing pixels on LCD screen
LCD display exhibits defective or missing pixels
Ranked causes
- 1. Display panel defect (unknown)
Inspect LCD screen for missing or defective pixels
- Test point
- LCD display surface
- Expected
- No missing pixels visible
Non-recommended accessories or parts used patient safety
Device operated with accessories or parts not recommended or supplied by Schiller AG
Ranked causes
- 1. Incompatible or non-approved accessories installed (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient proximity distance insufficient patient safety
Patient positioned less than 1.5 meters from device when connected to ancillary equipment without isolation
Ranked causes
- 1. Insufficient patient clearance from device (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Print quality degraded or faint output caution
Printed ECG waveforms are faint, unclear, or have poor contrast on paper
Ranked causes
- 1. Print head worn or thermal print mechanism degraded (occasional)
Access service screen and run print quality diagnostic
- Test point
- Service screen menu
- Expected
- Print quality test executed and results displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer head misaligned or calibration waveforms offset caution
Calibration waveforms not vertically aligned on paper; lateral deviation exceeds tolerance
Ranked causes
- 1. Mechanical adjustment of print head out of alignment (occasional)
Remove simulator and press Man Start
- Test point
- Printer manual feed
- Expected
- Printer operates without patient cable connected
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RS-232 port not functioning
Red indicator square does not turn green when RS-232 test plug is inserted into Treadmill, Ergo, or RS-232 port
Ranked causes
- 1. RS-232 port communication failure (unknown)
Insert RS-232 test plug into port
- Test point
- Treadmill (back panel), Ergo (back panel), or RS-232 (side panel)
- Expected
- Red square indicator changes to green
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software option installation incomplete
Software option code accepted but option may not be properly saved in the unit
Ranked causes
- 1. Option not saved after code entry (occasional)
Verify option installation in Software screen
- Test point
- Software screen menu
- Expected
- Option displayed and saved status confirmed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software option upgrade fails or freezes after multiple attempts
Unit freezes during program option installation after approximately 10 or more unsuccessful upgrade attempts; unit cannot be upgraded and must be returned to manufacturer
Ranked causes
- 1. Excessive failed upgrade attempts (approximately 10 or more) (common)
Count unsuccessful upgrade attempts
- Test point
- Number displayed to right of code entry field (4)
- Expected
- Less than approximately 10 attempts
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.