Welch Allyn 5002: troubleshooting and error codes

The ServiceForge knowledge base holds 18 documented symptoms for the Welch Allyn 5002, with 34 ranked causes, 27 diagnostic checks with test points and expected values, and 22 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Scale beep codes and messages

Scale emits various audible signals and displays messages; user identifies meaning

Ranked causes

  1. 1. Print requested without valid weight displayed (common)
  2. 2. Printer in use or disabled (occasional)
  3. 3. Printer hardware problem (occasional)
  4. 4. Battery too weak to operate (occasional)
  5. 5. Scale general problem (rare)
First check

Identify beep pattern and count

Test point
Audio output from scale
Expected
One/two/three beeps (normal); four short/long or continuous beeps (error)

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No print, four long beeps

Four long beeps sound when printer has internal problem

Ranked causes

  1. 1. Jammed print head (occasional)
  2. 2. Out of paper (common)
  3. 3. Disconnected printer cable (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No print, four short beeps

Four short beeps sound when print is requested but conditions are not met

Ranked causes

  1. 1. Weight display shows invalid value (e.g. '0.0') (common)
  2. 2. Printer in use (occasional)
  3. 3. Print function disabled in BMI calculation mode (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Reading does not change when weight applied

Scale shows no change in reading when object placed on platform

Ranked causes

  1. 1. Platform not plugged into readout (common)
  2. 2. Defective platform cable (occasional)
  3. 3. Defective load cell transducer (occasional)
First check

Verify platform cable plugged into readout

Test point
Platform cable connector at readout
Expected
Cable firmly connected

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scale displays CABLE or O-LOAd error

Console shows CABLE or O-LOAd message indicating signal issue

Ranked causes

  1. 1. Weight exceeds scale capacity (O-LOAd) (common)
  2. 2. Damaged transducer or defective instrument board (O-LOAd with valid reading) (occasional)
  3. 3. Damaged or disconnected transducer cable (CABLE) (common)
First check

Verify patient weight within capacity

Test point
Patient on scale
Expected
Weight reading within scale specifications (see inside front cover)

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Weight reading takes excessive time to display

Scale delays before showing stable weight value on display

Ranked causes

  1. 1. Platform in motion or excessive patient motion (common)
  2. 2. Platform rubbing against foreign object (occasional)
  3. 3. Loose cable connection (occasional)
First check

Verify platform and patient stability

Test point
Weighing platform
Expected
Platform stationary; patient not moving

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Four long beeps

Scale produces four long beeps indicating battery or mode issue

Ranked causes

  1. 1. Battery too weak to operate scale (common)
  2. 2. Scale in custom/factory/service/calibration mode (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Weight reading not accurate

Scale displays incorrect or unstable weight measurements on platform

Ranked causes

  1. 1. Mechanical obstruction preventing platform downward movement (common)
  2. 2. Loose or disconnected transducer cable (common)
First check

Inspect weighing platform for obstructions

Test point
Platform underside and surrounding area
Expected
No foreign objects visible; platform moves freely downward

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery replacement and storage

Annual battery replacement and proper storage procedures to prevent corrosion and maintain operation

Ranked causes

  1. 1. Routine battery maintenance (common)
First check

Replace batteries annually

Test point
Battery compartment
Expected
Fresh batteries installed, no corrosion

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Annual verification patient safety

Periodic service to confirm device functions within design specifications

Ranked causes

  1. 1. Routine maintenance requirement (common)
First check

Verify scale calibration

Test point
Scale output
Expected
Within design specifications

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Scale zero and span high voltage

Factory calibration drift or need to recalibrate zero point and measurement span using certified test weights

Ranked causes

  1. 1. Normal recalibration requirement or load cell/readout component damage (unknown)
First check

Verify scale is on flat hard surface with no physical interference

Test point
Scale platform
Expected
Flat, level, no obstructions

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Span only (default capacity) high voltage

Span calibration using default capacity and mV/V from scale type table without manual weight entry

Ranked causes

  1. 1. Need to recalibrate span using factory defaults (occasional)
First check

Verify scale is empty before default span calibration

Test point
Scale platform
Expected
Empty, no residual weight

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Zero only (default) high voltage

Quick zero calibration using default 0 kg = 0 mV/V without span adjustment

Ranked causes

  1. 1. Need to reset zero reference point only (occasional)
First check

Ensure scale is empty and on flat hard surface

Test point
Scale platform
Expected
No weight, flat level surface

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Continuous long beep

Scale produces continuous long beeping sound

Ranked causes

  1. 1. Scale malfunction (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cradle inspection required patient safety

Visual inspection for physical damage that may affect scale stability or patient safety

Ranked causes

  1. 1. Preventive maintenance requirement (common)
First check

Inspect cradle for damage

Test point
Cradle assembly
Expected
No cracks, all hardware present and tight

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Enclosure inspection required patient safety

Visual inspection of scale housing for damage, cracks, or loose components

Ranked causes

  1. 1. Preventive maintenance requirement (common)
First check

Inspect enclosure for damage

Test point
Scale housing
Expected
No damage, all hardware present and tight

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Long/short beeps

Irregular beep pattern indicates scale problem

Ranked causes

  1. 1. Scale malfunction (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power adapter cable inspection required high voltage

Visual check for signs of wear, abrasion, or damage to the power cable

Ranked causes

  1. 1. Preventive maintenance requirement (common)
First check

Inspect power adapter cable

Test point
Cable outer sheath and connectors
Expected
No abrasions, cracks, or visible wear

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
------Stationary display indicates a negative weightNegative weight value displayedwarning
bAttrYBattery power insufficient to operate scaleBattery voltage drops below minimum operating thresholdwarning
CABLEMeasurement board connection lost or signal out of rangeSignal from transducer outside a-d converter range; disconnected or damaged cableerror
ER OLSpan weight exceeds scale capacityerror
ERR ADMeasurement board cannot be readCommunication failure between scale processor and measurement boarderror
ERR AVMeasurement board version incompatibleMeasurement board firmware or hardware version not compatible with main scale firmwareerror
ERR CLMeasurement board error; weight display blockedMeasurement board malfunction preventing weight calculation or displayerror
ERR DBScale application database integrity compromisedCorruption or loss of application data in memorycritical
ERR FDScale factory database integrity compromisedCorruption or loss of factory configuration data in memorycritical
ERRoRScale flash memory integrity compromisedFlash memory corruption detected during self-testcritical
ER SP1Span weight outside acceptable margin for mV/V readingerror
ER SP2Calibration weight message failed to senderror
ER SP4Calibration weight message failed to senderror
ER SP5Error while waiting for scale to calibrate or measurement board internal errorerror
ER SP6Calibration did not saveerror
ER ZR0Reading is outside the scale type's acceptable mV/V rangeerror
ER ZR1Calibration message failed to senderror
ER ZR2Error while waiting for the scale to calibrateerror
ER ZR3Calibration did not saveerror
LoBATTBattery power insufficient to printBattery voltage too low for printer operationwarning
OFF SCALE PLEASERemove patient from weighing platform (wheelchair scales with ramps)Patient on scale during power-up before ramp calibration completesinfo
O-LOAdWeight exceeds scale maximum capacityWeight signal exceeds maximum design limit for scale modelwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.