Welch Allyn 5002: troubleshooting and error codes
The ServiceForge knowledge base holds 18 documented symptoms for the Welch Allyn 5002, with 34 ranked causes, 27 diagnostic checks with test points and expected values, and 22 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Scale beep codes and messages
Scale emits various audible signals and displays messages; user identifies meaning
Ranked causes
- 1. Print requested without valid weight displayed (common)
- 2. Printer in use or disabled (occasional)
- 3. Printer hardware problem (occasional)
- 4. Battery too weak to operate (occasional)
- 5. Scale general problem (rare)
Identify beep pattern and count
- Test point
- Audio output from scale
- Expected
- One/two/three beeps (normal); four short/long or continuous beeps (error)
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No print, four long beeps
Four long beeps sound when printer has internal problem
Ranked causes
- 1. Jammed print head (occasional)
- 2. Out of paper (common)
- 3. Disconnected printer cable (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No print, four short beeps
Four short beeps sound when print is requested but conditions are not met
Ranked causes
- 1. Weight display shows invalid value (e.g. '0.0') (common)
- 2. Printer in use (occasional)
- 3. Print function disabled in BMI calculation mode (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reading does not change when weight applied
Scale shows no change in reading when object placed on platform
Ranked causes
- 1. Platform not plugged into readout (common)
- 2. Defective platform cable (occasional)
- 3. Defective load cell transducer (occasional)
Verify platform cable plugged into readout
- Test point
- Platform cable connector at readout
- Expected
- Cable firmly connected
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scale displays CABLE or O-LOAd error
Console shows CABLE or O-LOAd message indicating signal issue
Ranked causes
- 1. Weight exceeds scale capacity (O-LOAd) (common)
- 2. Damaged transducer or defective instrument board (O-LOAd with valid reading) (occasional)
- 3. Damaged or disconnected transducer cable (CABLE) (common)
Verify patient weight within capacity
- Test point
- Patient on scale
- Expected
- Weight reading within scale specifications (see inside front cover)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Weight reading takes excessive time to display
Scale delays before showing stable weight value on display
Ranked causes
- 1. Platform in motion or excessive patient motion (common)
- 2. Platform rubbing against foreign object (occasional)
- 3. Loose cable connection (occasional)
Verify platform and patient stability
- Test point
- Weighing platform
- Expected
- Platform stationary; patient not moving
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Four long beeps
Scale produces four long beeps indicating battery or mode issue
Ranked causes
- 1. Battery too weak to operate scale (common)
- 2. Scale in custom/factory/service/calibration mode (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Weight reading not accurate
Scale displays incorrect or unstable weight measurements on platform
Ranked causes
- 1. Mechanical obstruction preventing platform downward movement (common)
- 2. Loose or disconnected transducer cable (common)
Inspect weighing platform for obstructions
- Test point
- Platform underside and surrounding area
- Expected
- No foreign objects visible; platform moves freely downward
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery replacement and storage
Annual battery replacement and proper storage procedures to prevent corrosion and maintain operation
Ranked causes
- 1. Routine battery maintenance (common)
Replace batteries annually
- Test point
- Battery compartment
- Expected
- Fresh batteries installed, no corrosion
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Annual verification patient safety
Periodic service to confirm device functions within design specifications
Ranked causes
- 1. Routine maintenance requirement (common)
Verify scale calibration
- Test point
- Scale output
- Expected
- Within design specifications
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Scale zero and span high voltage
Factory calibration drift or need to recalibrate zero point and measurement span using certified test weights
Ranked causes
- 1. Normal recalibration requirement or load cell/readout component damage (unknown)
Verify scale is on flat hard surface with no physical interference
- Test point
- Scale platform
- Expected
- Flat, level, no obstructions
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Span only (default capacity) high voltage
Span calibration using default capacity and mV/V from scale type table without manual weight entry
Ranked causes
- 1. Need to recalibrate span using factory defaults (occasional)
Verify scale is empty before default span calibration
- Test point
- Scale platform
- Expected
- Empty, no residual weight
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Zero only (default) high voltage
Quick zero calibration using default 0 kg = 0 mV/V without span adjustment
Ranked causes
- 1. Need to reset zero reference point only (occasional)
Ensure scale is empty and on flat hard surface
- Test point
- Scale platform
- Expected
- No weight, flat level surface
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Continuous long beep
Scale produces continuous long beeping sound
Ranked causes
- 1. Scale malfunction (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cradle inspection required patient safety
Visual inspection for physical damage that may affect scale stability or patient safety
Ranked causes
- 1. Preventive maintenance requirement (common)
Inspect cradle for damage
- Test point
- Cradle assembly
- Expected
- No cracks, all hardware present and tight
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Enclosure inspection required patient safety
Visual inspection of scale housing for damage, cracks, or loose components
Ranked causes
- 1. Preventive maintenance requirement (common)
Inspect enclosure for damage
- Test point
- Scale housing
- Expected
- No damage, all hardware present and tight
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Long/short beeps
Irregular beep pattern indicates scale problem
Ranked causes
- 1. Scale malfunction (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power adapter cable inspection required high voltage
Visual check for signs of wear, abrasion, or damage to the power cable
Ranked causes
- 1. Preventive maintenance requirement (common)
Inspect power adapter cable
- Test point
- Cable outer sheath and connectors
- Expected
- No abrasions, cracks, or visible wear
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
------ | Stationary display indicates a negative weight | Negative weight value displayed | warning |
bAttrY | Battery power insufficient to operate scale | Battery voltage drops below minimum operating threshold | warning |
CABLE | Measurement board connection lost or signal out of range | Signal from transducer outside a-d converter range; disconnected or damaged cable | error |
ER OL | Span weight exceeds scale capacity | error | |
ERR AD | Measurement board cannot be read | Communication failure between scale processor and measurement board | error |
ERR AV | Measurement board version incompatible | Measurement board firmware or hardware version not compatible with main scale firmware | error |
ERR CL | Measurement board error; weight display blocked | Measurement board malfunction preventing weight calculation or display | error |
ERR DB | Scale application database integrity compromised | Corruption or loss of application data in memory | critical |
ERR FD | Scale factory database integrity compromised | Corruption or loss of factory configuration data in memory | critical |
ERRoR | Scale flash memory integrity compromised | Flash memory corruption detected during self-test | critical |
ER SP1 | Span weight outside acceptable margin for mV/V reading | error | |
ER SP2 | Calibration weight message failed to send | error | |
ER SP4 | Calibration weight message failed to send | error | |
ER SP5 | Error while waiting for scale to calibrate or measurement board internal error | error | |
ER SP6 | Calibration did not save | error | |
ER ZR0 | Reading is outside the scale type's acceptable mV/V range | error | |
ER ZR1 | Calibration message failed to send | error | |
ER ZR2 | Error while waiting for the scale to calibrate | error | |
ER ZR3 | Calibration did not save | error | |
LoBATT | Battery power insufficient to print | Battery voltage too low for printer operation | warning |
OFF SCALE PLEASE | Remove patient from weighing platform (wheelchair scales with ramps) | Patient on scale during power-up before ramp calibration completes | info |
O-LOAd | Weight exceeds scale maximum capacity | Weight signal exceeds maximum design limit for scale model | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI