Waters Medical Systems Oxicom 2100: troubleshooting and error codes
The ServiceForge knowledge base holds 32 documented symptoms for the Waters Medical Systems Oxicom 2100, with 44 ranked causes, 46 diagnostic checks with test points and expected values, and 20 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Display shows error code ER1-ER5 patient safety
Error codes ER1 through ER5 indicate measuring separated blood sample with streaks or improper sample handling
Ranked causes
- 1. Blood sample separated or diluted by heparin/saline (common)
- 2. Blood sample more than 10 minutes old (common)
- 3. Cuvette laid on edge during handling (occasional)
Inspect blood sample for separation or streaks
- Test point
- Cuvette contents
- Expected
- Uniform color, no streaks or separated areas
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Erroneous readings due to cuvette contamination patient safety
Inaccurate measurements caused by contamination of cuvette sampling area or improper handling
Ranked causes
- 1. Cuvette sampling area contaminated or touched (common)
- 2. Foreign material on cuvette external surface (occasional)
- 3. Air bubbles or micro bubbles in blood sample (occasional)
Inspect cuvette before use
- Test point
- Cuvette sampling area and exterior
- Expected
- Clean, no contamination, no powder, no visible damage
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows error code ER7 or ER8 patient safety
Error codes ER7 or ER8 indicate blood in sample chamber or poor calibration from bad blood sample
Ranked causes
- 1. Blood in sample chamber (occasional)
- 2. Poor calibration from bad blood sample (occasional)
Verify blood samples meet calibration parameters
- Test point
- Sample composition and saturation
- Expected
- Arterial 90-99.9%, venous 60-72%, THb 10-14 gm/dl, COHb <3%
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Erroneous readings due to blood sample separation or contamination patient safety
Inaccurate measurements from separated blood, heparin/saline dilution, or dye contamination
Ranked causes
- 1. Blood diluted by heparin or saline (common)
- 2. Presence of dyes affecting measurement (rare)
Verify blood sample is not separated
- Test point
- Cuvette contents visual inspection
- Expected
- Uniform color, no separated or stratified areas
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Erroneous readings due to cuvette reuse or filter reuse patient safety
Inaccurate measurements when cuvettes or white filters are reused
Ranked causes
- 1. Cuvette reused causing optical path obstruction (occasional)
- 2. White filter reused allowing oxygen contamination (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Erroneous readings due to dirty QC filters caution
Inaccurate QC readings when Quality Control filters are contaminated or not cleaned properly
Ranked causes
- 1. Dirty QC filters (common)
- 2. QC filters inserted in wrong sequence (occasional)
Inspect QC filters for cleanliness
- Test point
- QC filter surfaces
- Expected
- Clean, dry, no dust or contamination
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Erroneous readings due to improper sample insertion timing
Incorrect measurements when cuvette inserted before INS prompt or removed before reading complete
Ranked causes
- 1. Sample inserted before INS display prompt (common)
- 2. Sample removed before reading is completed (occasional)
Monitor display for INS prompt
- Test point
- Display screen
- Expected
- "INS" displayed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Erroneous readings due to liquid control sample issues caution
Inaccurate control readings when liquid QC samples are improperly mixed or contaminated
Ranked causes
- 1. Liquid control sample improperly mixed (occasional)
- 2. Blood or foreign substances in sample chamber affecting controls (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Quality control reading outside tolerance
QC filter reading differs from expected value by more than ±3% during internal QC or ±7% during external QC; 'CLN' or 'RMV' displayed when chamber contaminated
Ranked causes
- 1. Sample chamber contaminated (common)
- 2. Oxicom 2100 requires recalibration (occasional)
Verify sample chamber is clean before QC run
- Test point
- Sample chamber
- Expected
- No visible contamination
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sample not accepted or display shows RMV
Sample was inserted before 'INS' appeared on display, or sample inserted incorrectly; 'RMV' message displayed
Ranked causes
- 1. Sample inserted before INS prompt (common)
- 2. Sample inserted with cuvette orientation incorrect (occasional)
Verify INS prompt is displayed before sample insertion
- Test point
- Display screen
- Expected
- INS message visible
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxicom 2100 accuracy verification patient safety
Quarterly check to verify measurement accuracy within 3% of co-oximeter reference using venous blood sample in 60-72% saturation range
Ranked causes
- 1. Accuracy drift beyond 3% tolerance (common)
Compare Oxicom 2100 reading to co-oximeter functional saturation
- Test point
- Sample chamber
- Expected
- Within 3% of co-oximeter value
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxicom 2100 oxygen saturation patient safety
Procedure to calibrate the Oxicom 2100 medical gas analyzer using arterial and venous blood samples with Co-Oximeter reference values
Ranked causes
- 1. Calibration required for rated accuracy (common)
Warm-up time before calibration
- Test point
- Oxicom 2100 power on
- Expected
- Unit on for minimum 15 minutes; if not, wait for Display to read CAL
9 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxicom 2100 weekly quality control patient safety
Record weekly QC readings for QCHigh, QCMid, QCLow samples and verify they fall within ±7% of baseline range; inspect and clean sample chamber
Ranked causes
- 1. Out-of-range QC readings indicate analyzer drift or malfunction (occasional)
Establish baseline QC ranges for High, Mid, Low samples
- Test point
- QCHigh, QCMid, QCLow
- Expected
- Range ±7% Sat
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows asterisk (*) in left segment patient safety
Asterisk displayed indicating measured oxygen saturation exceeds 99.9%
Ranked causes
- 1. Measured saturation exceeds 99.9% saturation range (occasional)
Verify blood sample parameters
- Test point
- Sample composition
- Expected
- Arterial 90-99.9%, THb 10-14 gm/dl, COHb <3%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows error code CLN
Error code CLN displayed indicating blood or foreign material present in sample chamber
Ranked causes
- 1. Blood or foreign material in the sample chamber (common)
Inspect sample chamber for blood or contamination
- Test point
- Sample chamber interior surfaces
- Expected
- Clean, no visible blood or foreign material
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows error code ER6 patient safety
Error code ER6 indicates streaks visible in blood sample
Ranked causes
- 1. Streaks or abnormalities in blood sample (occasional)
Visually inspect blood sample in cuvette
- Test point
- Cuvette sampling area
- Expected
- Uniform color with no streaks or abnormalities
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows error code ER9 caution
Error code ER9 indicates optical density ratio (ODR/ODIR) outside range of 0.5 to 2.0
Ranked causes
- 1. Optical density ratio out of tolerance (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows error code ERa caution
Error code ERa indicates Red LED out of tolerance or not operating
Ranked causes
- 1. Red LED out of tolerance or not operating (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows error code ERb caution
Error code ERb indicates IR LED out of tolerance or not operating
Ranked causes
- 1. IR LED out of tolerance or not operating (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows error code ERc patient safety
Error code ERc indicates high and low sample values were entered in reversed order during calibration
Ranked causes
- 1. High and low sample target values reversed during calibration (occasional)
Verify correct sample target values during calibration entry
- Test point
- Calibration display prompts
- Expected
- High sample 90-99.9%, low sample 60-72%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows error code ERd patient safety
Error code ERd indicates calibration factors derived are out of range
Ranked causes
- 1. Bad blood samples or incorrect target values entered (occasional)
Verify blood sample parameters before calibration
- Test point
- Sample composition and saturation
- Expected
- Arterial 90-99.9%, venous 60-72%, THb 10-14 gm/dl, COHb <3%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows error code ERe caution
Error code ERe indicates optical density measured by Oxicom 2100 is out of tolerance
Ranked causes
- 1. Dirty QC filters, contaminated sample chamber, or calibration drift (common)
Verify QC filters are clean and dry
- Test point
- QC filter surfaces
- Expected
- No dust, blood, or contamination; filters dry
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows error code RMV
Error code RMV displayed when sample is inserted before INS prompt appears
Ranked causes
- 1. Sample inserted before INS display prompt (common)
Monitor display for INS prompt
- Test point
- Display screen
- Expected
- "INS" displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Erroneous readings due to display malfunction caution
Incorrect readings when display is malfunctioning or system check fails
Ranked causes
- 1. Display malfunction or failed system check (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Erroneous readings due to improper cuvette orientation
Inconsistent or inaccurate readings when cuvette not positioned correctly in chamber
Ranked causes
- 1. Cuvette not inserted with bottom panel down (occasional)
Verify cuvette orientation before insertion
- Test point
- Cuvette position in hand
- Expected
- Bottom panel facing downward toward chamber
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Erroneous readings due to incomplete warm-up
Inaccurate measurements taken before rated accuracy is achieved during warm-up period
Ranked causes
- 1. Unit not fully warmed up before measurement (common)
Verify warm-up completion
- Test point
- Display status
- Expected
- "RDY" or "CAL" displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error displayed during blood sample insertion
Error code appears when arterial or venous blood sample is inserted into sample chamber during calibration
Ranked causes
- 1. Blood or foreign material contamination in sample chamber (common)
Inspect sample chamber for visible contamination
- Test point
- Sample chamber interior
- Expected
- No blood residue, particles, or foreign material visible
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Factory reset calibration procedure patient safety
Procedure to reset Oxicom 2100 to factory default settings as starting point for recalibration
Ranked causes
- 1. Need to reset to factory defaults before recalibration (common)
Verify FCT mode is entered
- Test point
- Display screen
- Expected
- "FCT" displayed after holding arrow buttons and powering on
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
QC1 data not stored in memory
QC1 filter decimal points do not alternate or data storage does not complete during QC filter measurement phase
Ranked causes
- 1. Insufficient time for data storage completion (occasional)
Monitor decimal point alternation during QC1 measurement
- Test point
- Display decimal points
- Expected
- Decimal points alternate while measuring; stop alternating when measurement complete min, 1 to 2 minutes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Quality Control readings out of range
QC filter readings differ from expected calibration values or display shows 'ERe' indicating optical density variation or sample chamber contamination
Ranked causes
- 1. Sample chamber contaminated (common)
Insert QC1 filter and record reading
- Test point
- Sample chamber
- Expected
- Reading within ±3% of calibration value (e.g., 82-88% if QC1 nominal is 85%)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sample chamber contamination with blood caution
Blood or foreign material accidentally infused into sample chamber causes incorrect readings and error code CLN
Ranked causes
- 1. Blood or foreign material in sample chamber (occasional)
Verify sample chamber is clean before reassembly
- Test point
- Detector and fiber optic cable tip surfaces
- Expected
- No visible blood or foreign material
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sample measurement within time window
Ensuring blood samples are measured promptly to prevent cell separation and maintain sample integrity
Ranked causes
- 1. Delayed sample measurement causing cell separation (common)
Measure blood sample
- Test point
- Sample timing
- Expected
- Sample measured within 10 minutes of withdrawal
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
* | Saturation exceeds 99.9% | Measured oxygen saturation value exceeds 99.9% range | info |
CLN | Blood or foreign material in sample chamber | Detector senses contamination in sample chamber | warning |
ER1 | Blood sample separated or degraded | Measuring blood sample that has separated | error |
ER2 | Blood sample separated or degraded | Measuring degraded blood sample | error |
ER3 | Blood sample separated or degraded | Measuring degraded blood sample | error |
ER4 | Blood sample separated or degraded | Measuring degraded blood sample | error |
ER5 | Blood sample separated or degraded | Measuring degraded blood sample | error |
ER6 | Streaks visible in blood sample | Measuring blood sample with visible streaks | error |
ER7 | Blood in chamber or bad calibration sample | Blood detected in sample chamber or calibration on degraded sample | error |
ER8 | Blood in chamber or bad calibration sample | Blood detected in sample chamber or calibration on degraded sample | error |
ER9 | Optical density ratio out of tolerance | Ratio of optical densities (ODR/ODIR) outside 0.5 to 2.0 range | error |
ERa | Red LED out of tolerance or not operating | Red LED fails function or output test | error |
ERb | IR LED out of tolerance or not operating | IR LED fails function or output test | error |
ERc | High and low sample values entered in reversed order | During calibration, high sample entered as low value and vice versa | error |
ERd | Calibration factors out of range | Derived factors during calibration exceed acceptable limits | error |
ERe | Optical density out of tolerance | Measured optical density exceeds acceptable range | error |
ERe | Error in QC reading; optical densities differ or sample chamber contaminated | QC filter reading differs from expected value or optical density variation detected | error |
INS | Insert QC filter; device ready for filter insertion | Displayed after power-on warm-up completion or after START button pressed | info |
RMV | Sample inserted at wrong time | Cuvette inserted before INS display prompt appears | warning |
RMV | Remove QC filter; filter inserted before INS prompt appeared | QC filter inserted before "INS" displayed on screen | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI