Vyaire Mk II Infant Flow Driver: troubleshooting and error codes
The ServiceForge knowledge base holds 11 documented symptoms for the Vyaire Mk II Infant Flow Driver, with 14 ranked causes, 11 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unit does not pass performance verification test patient safety
One or more calibration checks fail to meet specification requirements
Ranked causes
- 1. Mixer output does not meet ±3% accuracy specification (unknown)
- 2. Oxygen analyzer readings outside tolerance range (unknown)
- 3. Pressure manometer does not track accurately (unknown)
- 4. Alarm functions do not operate correctly (unknown)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air inlet filter inspection
Inspect air inlet filter for dirt or moisture contamination. Replace if dirty or wet.
Ranked causes
- 1. Air inlet filter dirty or wet (occasional)
Observe air inlet filter condition
- Test point
- Air inlet filter
- Expected
- Clean and dry
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Air/oxygen mixer and electronic pressure manometer
Periodic 4-monthly calibration check of air/oxygen mixer blender output and electronic pressure manometer required.
Ranked causes
- 1. Mixer or pressure manometer calibration drift (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Electronic safety relief valve regulator setting patient safety
Verify electronic patient safety relief valve regulator pressure setting
Ranked causes
- 1. Electronic regulator requires periodic verification (common)
Measure regulator output pressure
- Test point
- Pressure regulator outlet (disconnected from solenoid)
- Expected
- Between 160 & 180 mm Hg, 160-180 mm Hg
- Tools
- Mercury column or accurate pressure measuring device
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen analyzer
Fuel cell oxygen analyzer requires regular calibration check to ensure accurate display readings at 21% and 100% oxygen concentrations.
Ranked causes
- 1. Oxygen analyzer drift or miscalibration (common)
Set mixer to 21% oxygen and verify display reading
- Test point
- Oxygen display
- Expected
- 21%
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Safety pressure relief valve setting patient safety
Verify mechanical safety relief valve is set to protect equipment and delivery circuit
Ranked causes
- 1. Safety pressure relief valve requires periodic verification (common)
Measure pressure at fully occluded delivery tube
- Test point
- Patient circuit
- Expected
- 205 cm H2O, ±20 cm H2O
- Tools
- Calibrated water column or pressure measuring system
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check alarm system integrity patient safety
Periodic checking of alarm systems integrity required as part of routine maintenance.
Ranked causes
- 1. Alarm system malfunction or failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check exterior air filter condition caution
Exterior air filter visible through polycarbonate bowl; replace if discolored or wet.
Ranked causes
- 1. Discolored or wet exterior air filter (occasional)
Inspect exterior air filter through polycarbonate bowl
- Test point
- Air filter
- Expected
- Clear, dry, no discoloration
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flowmeter calibration drift
Flowmeter indicates incorrect flow rate compared to test device
Ranked causes
- 1. Flowmeter requires recalibration (common)
Verify flowmeter reading against external test device
- Test point
- Outlet of Infant Flow Driver
- Expected
- Within ±10% of test device reading lpm, ±10%
- Tools
- Test flowmeter or flow test device
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen sensor end of life
Oxygen sensor no longer calibrates accurately or reaches one year of service
Ranked causes
- 1. Oxygen sensor aged beyond useful life (1 year typical) (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen sensor replacement caution
Oxygen sensor must be changed annually or when calibration is no longer possible. Device must be disconnected from power before replacement.
Ranked causes
- 1. Annual oxygen sensor replacement or failed calibration (common)
Allow oxygen sensor to stabilize after installation
- Test point
- Time elapsed
- Expected
- 2 hours minimum
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI