Volcano s5i: troubleshooting and error codes
The ServiceForge knowledge base holds 58 documented symptoms for the Volcano s5i, with 115 ranked causes, 119 diagnostic checks with test points and expected values, and 107 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
DICOM network configuration - host not found
System cannot locate DICOM server host; PING fails with host name not resolved
Ranked causes
- 1. Network cable not connected to Volcano s5i (common)
- 2. Remote server offline or not on network (common)
- 3. Defective network cable (occasional)
- 4. Invalid local Volcano s5i network configuration (occasional)
- 5. Invalid gateway IP address or subnet mask (occasional)
- 6. Invalid remote DICOM server IP address configuration (occasional)
- 7. Network speed/duplex auto negotiation failure (occasional)
Verify network cable connection
- Test point
- Volcano s5i network port
- Expected
- Cable fully seated and connected
2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control Panel keys do not work or respond
Control panel keyboard buttons not functioning; input not registering when buttons pressed
Ranked causes
- 1. USB cables or extender not fully connected (common)
- 2. Control panel keys damaged or binding (occasional)
- 3. Damaged or defective USB extender (occasional)
- 4. Control Panel hardware defective (rare)
Verify USB cables and extender fully connected
- Test point
- USB connector ports on system and control panel; extender connector
- Expected
- All connectors fully seated and secured; no visible damage or bent pins
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No IVUS display or image shown
System boots and runs but IVUS imaging window does not appear or shows no image data
Ranked causes
- 1. PIM not connected or not powered (common)
- 2. Catheter or catheter simulator disconnected or defective (common)
- 3. Meridian analog/digital boards not properly seated or filter cards loose (occasional)
- 4. Meridian imaging boards defective (rare)
Verify PIM connected to system interface
- Test point
- PIM connector to main system board
- Expected
- PIM cable fully seated and secured
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer does not work or print
Printer is connected but does not respond to print commands; nothing prints
Ranked causes
- 1. Printer not turned on or lacks paper/ink ribbon (common)
- 2. USB or power cabling not fully connected (common)
- 3. Printer not set as default or Print Manager not running (occasional)
- 4. Printer hardware defective (rare)
Verify printer powered on and media loaded
- Test point
- Printer power button, paper tray, print ribbon
- Expected
- Printer powered on, LED lit, paper loaded, ribbon present and functional
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not power up caution
System will not turn on; workstation PC does not respond to power button
Ranked causes
- 1. AC power cord disconnected or improperly plugged in (common)
- 2. Isolation Transformer not enabled or set to wrong voltage (common)
- 3. Workstation PC power button location not found or power issue (occasional)
- 4. Workstation PC DC power supply failure (rare)
Verify all AC power cords properly connected
- Test point
- Power inlet connectors at Isolation Transformer and system chassis
- Expected
- Connectors firmly seated, no visible damage
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System has no power
No display, indicators, or control feedback when power switch activated
Ranked causes
- 1. System powered off or power switch not activated (common)
- 2. Isolation transformer off (s5i only) (common)
- 3. Power plug disconnected (common)
- 4. Wall outlet power not available (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to archive to DVD
System cannot write to DVD media; DVD archive function fails with error or drive not accessible
Ranked causes
- 1. Improper DVD media type being used (common)
- 2. Cases not selected for archive operation (common)
- 3. DVD drive not detected by system BIOS or software (occasional)
- 4. DVD drive hardware defective (rare)
Verify DVD-R media type being used
- Test point
- DVD media in drive and packaging
- Expected
- Media labeled DVD-R (not DVD-RW, DVD+R, or other type)
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Aortic High Level input
Procedure to calibrate the Aortic Signal input on rear panel so Pa (pressure-aortic) display value matches physiological monitor in Cath lab. Required at installation and when equipment changes.
Ranked causes
- 1. Aortic input requires calibration for display accuracy (common)
- 2. Demo Mode enabled prevents calibration (occasional)
- 3. Hardware communication timeout (rare)
Aortic input connected to rear panel and at zero level
- Test point
- Aortic Input connector, rear panel
- Expected
- Zero level signal
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Catheter Type not recognized by system
System does not identify catheter type when connected; 'Catheter Type not recognized' error or message
Ranked causes
- 1. Contamination or corrosion on catheter connector pins or PIM sockets (common)
- 2. Install error; catheter not fully connected on first insertion (common)
- 3. Mechanical damage to PIM connector or defective catheter (occasional)
Disconnect and reconnect catheter after 3 second delay
- Test point
- PIM catheter connector
- Expected
- Catheter fully seated and locked; type recognized by system
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM verification failed - server unreachable
System cannot connect to remote DICOM host; socket connection fails
Ranked causes
- 1. Invalid port number for remote DICOM server (common)
- 2. Remote DICOM application not running or offline (common)
- 3. DICOM server not configured to accept Volcano s5i IP (occasional)
Verify DICOM SCP port number
- Test point
- DICOM server configuration
- Expected
- Port number matches Volcano s5i configuration
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
FFR measurements are incorrect or erratic patient safety
FFR pressure values displayed are inaccurate, unstable, or not matching expected values; measurements inconsistent
Ranked causes
- 1. Faulty or uncalibrated pressure wire (common)
- 2. Faulty PIM-ffr interface or signal conditioning (occasional)
- 3. Faulty VH board affecting FFR processing (rare)
Test FFR measurements with known good pressure wire
- Test point
- Pressure wire connected to FFR input; display readings
- Expected
- Stable pressure readings; values consistent and within expected range
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No FFR input signal or measurement patient safety
Fractional Flow Reserve (FFR) system not acquiring or displaying pressure wire data; no FFR readings
Ranked causes
- 1. Faulty pressure wire (common)
- 2. Faulty PIM-ffr interface board (occasional)
- 3. Faulty VH imaging board (rare)
Test FFR input with another known good pressure wire
- Test point
- FFR pressure wire connector on PIM-ffr or PIM
- Expected
- FFR input detected; pressure readings display on monitor
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Heart-beating icon with ECG input or No VH data patient safety
ECG waveform not triggering heart-beat icon animation; VH (Virtual Histology) imaging card not functioning or ECG signal not detected
Ranked causes
- 1. Improper ECG signal or cabling not fully connected (common)
- 2. VH jumper cable not installed between Meridian digital board and VH board (occasional)
- 3. VH imaging board defective (rare)
Test ECG input with ECG simulator
- Test point
- ECG input connector (rear panel)
- Expected
- Heart-beating icon animates when ECG simulator connected and running
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No scheduled procedures returned from worklist search patient safety
Worklist query returns empty result; no procedures displayed for patient or search criteria
Ranked causes
- 1. Volcano s5i system clock date not set to today (occasional)
- 2. HIS/CIS does not have Volcano s5i scheduled station AE title configured (occasional)
- 3. HIS/CIS does not schedule procedures for IVUS modality (occasional)
Verify Volcano s5i system clock date
- Test point
- System settings
- Expected
- Current date set to today
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No video displayed on monitor
Monitor is powered on but shows no image; blank or black screen
Ranked causes
- 1. AC power interruption to system (common)
- 2. Video signal path interrupted or misconfigured (common)
- 3. Video board defective or display damaged (occasional)
Check monitor power LED status
- Test point
- Monitor front panel LED
- Expected
- Power LED illuminated or monitor powered on
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Operating System fails to boot caution
System hangs or stops during OS boot sequence; does not reach Windows login or application startup
Ranked causes
- 1. DC power supply insufficient or failing (occasional)
- 2. Hard drive not detected or SATA not configured correctly (occasional)
- 3. Hard drive data corruption or operating system damage (occasional)
Verify Workstation PC system DC power output
- Test point
- Main system board power connector
- Expected
- +12V, +5V, +3.3V supplies within 5% of nominal
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIM not Found Error Message displayed
System displays 'PIM not Found' error; imaging interface not detected at startup
Ranked causes
- 1. PIM cabling not fully connected (common)
- 2. Bent connector pins on PIM or system connector (occasional)
- 3. PIM physically damaged or defective (rare)
Verify PIM connector fully connected to system
- Test point
- PIM connector and system connector
- Expected
- Connectors fully seated and secured; no gaps visible
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System boots to Volcano logo screen but User Interface not visible
System boot process reaches logo screen but User Interface window does not appear; mouse cursor active
Ranked causes
- 1. User Interface window in background, not in foreground (common)
- 2. System hard drive corrupt or defective (occasional)
- 3. Hard drive defective and cannot be recovered (rare)
Press Escape key to bring UI window to foreground
- Test point
- System keyboard
- Expected
- Escape key press triggers User Interface to appear
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trackball does not work or responds improperly
Control panel trackball unresponsive; cursor does not move or moves erratically when trackball rotated
Ranked causes
- 1. USB cables or extender not fully connected (common)
- 2. Damaged or defective USB extender (occasional)
- 3. Control Panel or trackball hardware defective (rare)
Verify USB cables and extender fully connected
- Test point
- USB connector ports at system and control panel; extender
- Expected
- All connectors fully seated; no damaged or bent pins
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
VHFM pipeline test failure or degraded waveforms
VHFM pipeline test waveforms differ greatly from demo display; data processing not functioning properly.
Ranked causes
- 1. Software, memory, or PC problem (no Volcano-unique hardware involved) (unknown)
- 2. System software or PC issue requiring full restart (unknown)
- 3. Software corruption or obsolete version (unknown)
Initiate VHFM pipeline test from diagnostics menu
- Test point
- Diagnostic test Live screen
- Expected
- Waveforms scroll across screen similar to demo display with same settings
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display not functioning
System lights and fans operating but no video output to monitor
Ranked causes
- 1. Monitor powered off or disconnected (common)
- 2. Backup battery failed (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Joystick controller only has one axis of movement
Joystick responds to X or Y movement but not both axes; movement restricted to one direction
Ranked causes
- 1. Damaged or defective joystick (common)
- 2. Damaged USB extender affecting joystick signal (occasional)
Test joystick movement in all directions
- Test point
- Joystick handle during system use
- Expected
- Smooth response to forward/back (Y) and left/right (X) movements
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Only one monitor displays User Interface
Dual display setup not working; only one of two monitors shows the system UI
Ranked causes
- 1. One display disconnected or not powered on during boot (common)
- 2. Display disabled in Windows or ATI control manager (occasional)
Verify both displays connected and powered on
- Test point
- Display power buttons, DVI cable connectors at graphics card
- Expected
- Both display power indicators ON; both DVI cables firmly seated
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIM-ffr not responding or cable fault patient safety
Display Patient Cable Status diagnostic indicates PIM-ffr module or cable malfunction; PIM icon at lower right of screen shows fault indication.
Ranked causes
- 1. PIM-ffr module cable loose or disconnected (common)
- 2. PIM-ffr module malfunction (occasional)
Visually inspect PIM-ffr cable connection status
- Test point
- PIM-ffr connector
- Expected
- Cable fully seated; no visible damage; connector locked
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIMr pullback function improper or not tracking correctly
PIMr (motorized pullback device) not retracting catheter smoothly; movement is obstructed, intermittent, or incorrect
Ranked causes
- 1. Obstruction to PIMr cable or catheter movement (common)
- 2. Faulty PIMr motor or drive mechanism (occasional)
Verify PIMr cable and catheter free of obstruction
- Test point
- PIMr cable routing and imaging catheter path
- Expected
- Cable and catheter not kinked, bent, or pinched; free movement possible
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to archive to DICOM network
System cannot send DICOM data to network server; archive to DICOM fails with error or connection lost
Ranked causes
- 1. Network cabling not fully connected (common)
- 2. Network set-up error or incorrect IP address configuration (occasional)
Verify network Ethernet cable connected to system and switch
- Test point
- Network port on system and switch/router
- Expected
- Cable fully seated; LED indicators on port showing connection
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Video display lacks fundamental color or shows only R, G, or B
Color display showing only one primary color (red, green, or blue) or missing colors; DVI signal incomplete
Ranked causes
- 1. DVI connector not fully seated or screwed in (common)
- 2. Video cable has broken or damaged wire in DVI connector (occasional)
Inspect DVI connector seating and retaining screws
- Test point
- DVI connectors at display and graphics card
- Expected
- Connector fully inserted; retaining screws tight; no visible damage or missing pins
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Aortic High Level calibration fails with Hardware did not respond patient safety
System did not receive response from hardware within the allotted timeout period during Aortic High Level calibration procedure.
Ranked causes
- 1. Hardware communication timeout or connection failure (common)
Verify hardware connection and responsiveness
- Test point
- Aortic Signal input connector on rear panel
- Expected
- Secure connection; hardware responds to commands
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Aortic High Level calibration fails with Out of Range error patient safety
Input signal level is too low or too high, exceeding the ±10% tolerance band of the selected calibration level (100, 200, or 300).
Ranked causes
- 1. Input signal outside ±10% tolerance of selected calibration level (common)
Measure or verify input signal level
- Test point
- Aortic Signal input on rear panel
- Expected
- Within ±10% of selected calibration level (100, 200, or 300), ±10% of selected level
- Tools
- Physiological monitor or calibrator device
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Aortic High Level calibration fails with Unstable Signal error patient safety
Input signal fluctuates more than ±2.5 mmHg around the selected calibration level, preventing successful calibration.
Ranked causes
- 1. Input signal source or connection is unstable (common)
Observe input signal stability during calibration attempt
- Test point
- Aortic Signal input on rear panel
- Expected
- Signal stable within ±2.5 mmHg of selected level, ±2.5 mmHg
- Tools
- Oscilloscope or physiological monitor display
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Aortic High Level for FFR patient safety
Setup and calibration of the Aortic Signal input on the rear panel to ensure Pa (pressure-aortic) values match the Cath lab's physiological equipment. Performed once at installation or when equipment changes.
Ranked causes
- 1. Initial installation or equipment changes requiring aortic signal calibration (common)
Zero calibration of Aortic Input signal
- Test point
- Aortic Signal input on rear panel
- Expected
- System interprets input as zero; brief wait message displays
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Input aortic high level restore to factory default
Reset aortic high level calibration to factory-determined setting when calibration has failed or needs to be discarded
Ranked causes
- 1. Calibration failed or requires reset (unknown)
Verify calibration failure or need for reset
- Test point
- Input Calibration section display
- Expected
- Failed Calibration Message indication visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PIM-ffr input accuracy check
Verify PIM-ffr pressure reading accuracy using Pressure Guide Wire Simulator fixture; checks zero offset and 300 mmHg reading within tolerance
Ranked causes
- 1. PIM-ffr calibration out of specification (unknown)
Verify PIM-ffr auto-zero function
- Test point
- Live tab digital Pd display
- Expected
- zero
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM association negotiation failed - invalid called AE title
DICOM verification fails; remote DICOM Storage or Worklist SCP AE title incorrect
Ranked causes
- 1. Incorrect Application Entity Title for DICOM Storage or Worklist SCP (common)
Verify remote DICOM SCP AE title
- Test point
- Remote DICOM server configuration
- Expected
- AE title matches Volcano s5i configuration
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM association negotiation failed - invalid calling AE title
DICOM verification fails; Volcano s5i SCU AE title incorrect or not configured on remote
Ranked causes
- 1. Incorrect Volcano s5i DICOM Storage or Worklist SCU AE title (common)
Verify Volcano s5i SCU AE title
- Test point
- Volcano s5i DICOM configuration
- Expected
- SCU AE title defined
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM verification warning - Storage abstract syntax rejected
DICOM verification succeeds with warning; Storage SCP does not accept Ultrasound Multiframe
Ranked causes
- 1. Storage SCP not a Storage server or does not accept Ultrasound Multiframe SOP class (occasional)
Verify Storage SCP supports Ultrasound Multiframe
- Test point
- Remote DICOM server configuration
- Expected
- STANDARD_US_MF (Ultrasound Multiframe) supported
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM verification warning - Storage transfer syntax rejected
DICOM verification succeeds with warning; Storage SCP does not accept JPEG lossy compression
Ranked causes
- 1. Storage SCP does not accept JPEG lossy transfer syntax (occasional)
Verify Storage SCP compression setting
- Test point
- DICOM Storage SCP configuration
- Expected
- Compression set to Uncompressed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM verification warning - Worklist abstract syntax rejected
DICOM verification succeeds with warning; Worklist SCP does not support Worklist server
Ranked causes
- 1. Configured Worklist SCP is not a Worklist server (occasional)
Verify Worklist SCP type
- Test point
- DICOM Worklist SCP configuration
- Expected
- SCP configured as Worklist server
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive not bootable error
System displays 'Drive not bootable. Insert bootable disk.' message
Ranked causes
- 1. Hard drive failure or corruption (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
GUI image not aligned on screen or slightly off screen
User interface window positioned incorrectly on display; partially off edge or misaligned
Ranked causes
- 1. LCD display did not switch video modes correctly during boot (common)
Reboot display or use display menu to rescan synchronization
- Test point
- Display power button or on-screen menu
- Expected
- Display re-detects video signal; GUI becomes properly centered
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image colors not displayed correctly on DICOM viewer
DICOM Viewer not rendering color palette images correctly
Ranked causes
- 1. Workstation does not handle color palette images correctly (occasional)
Verify Storage SCP compression format
- Test point
- DICOM Storage SCP configuration
- Expected
- Compression set to JPEG for color palette support
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect PIM cable length configuration caution
PIM cable length setting does not match physically installed cable, leading to inaccurate measurement results
Ranked causes
- 1. PIM cable length setting mismatch with installed hardware (common)
Verify PIM cable length setting matches installed cable
- Test point
- Field Service Dialog, PIM Cable Length tab
- Expected
- Selected length matches physical cable
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect PIM cable length setting caution
PIM cable length setting does not match the physical cable installed, resulting in incorrect measurement results
Ranked causes
- 1. PIM cable length misconfigured in Field Service Dialog (common)
Verify PIM cable length setting matches installed cable
- Test point
- Field Service Dialog - PIM Cable Length tab
- Expected
- Selected length matches physical PIM cable installed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IVUS display appears noisy or has excessive noise
IVUS image shows visual noise, artifacts, or degraded signal quality; not due to catheter issue
Ranked causes
- 1. Extraneous equipment radiating EMI at transducer frequencies (common)
Identify EMI sources near system transducer cables
- Test point
- Environment around transducer and PIM cabling
- Expected
- No active RF/microwave/strong EM radiating equipment in close proximity
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IVUS image vertically or horizontally ellipsoid shaped
IVUS image appears stretched or compressed in one axis; aspect ratio incorrect
Ranked causes
- 1. Display resolution not set to native 1280x1024 aspect ratio mode (common)
Verify display resolution set to 1280x1024
- Test point
- Windows Display Settings resolution dropdown
- Expected
- Resolution shows 1280x1024; refresh rate standard
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IVUS images cannot be displayed on DICOM viewer
DICOM Workstation/Viewer does not support IVUS image modality
Ranked causes
- 1. DICOM Workstation/Viewer does not support IVUS modality (occasional)
Verify DICOM Storage Server modality configuration
- Test point
- DICOM Storage Server settings
- Expected
- Modality set to US or supported IVUS modality
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Multiple logical devices detected instead of single peripheral
'New Hardware Found' dialog displays repeatedly; system detects and configures multiple sub-devices for what is physically one peripheral
Ranked causes
- 1. Multi-function USB device with multiple logical sub-devices (occasional)
Monitor for repeated 'New Hardware Found' dialogs
- Test point
- Screen display
- Expected
- Multiple dialogs appear sequentially, one per sub-device
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network transfer failed - C-STORE interoperability error
DICOM archive to network fails with C-STORE error; status code and error description variable
Ranked causes
- 1. Interoperability problem with vendor device (rare)
Debug via DICOM/Network Configuration dialog
- Test point
- DICOM Configuration menu
- Expected
- Detailed error diagnostics available
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network transfer failed - C-STORE out of resources
DICOM archive to network fails; C-STORE error with status code a700
Ranked causes
- 1. DICOM SCP experiencing resource problem or disk full (common)
Check remote DICOM server resources
- Test point
- Remote server
- Expected
- Adequate disk space and memory available
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No disk in floppy drive error
System displays 'No Disk in Floppy Drive' message
Ranked causes
- 1. Backup battery failing (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Restore input calibration to factory default
Reset Aortic High Level calibration to factory-determined setting when calibration fails or needs to be reset
Ranked causes
- 1. Calibration failure or need to reset to known state (unknown)
Verify Restore button is available in Input Calibration section
- Test point
- Input Calibration section
- Expected
- Restore button visible and clickable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Revolution image appears noisy or has noise artifacts
Revolution imaging mode showing visual noise or poor quality in image; excessive grain or interference
Ranked causes
- 1. Air in imaging catheter preventing proper signal transmission (common)
Observe Revolution image quality for air artifacts
- Test point
- System display during Revolution imaging
- Expected
- Clear image without noise or shadowing characteristic of air in catheter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
s5i FFR hardware test failure
FFR hardware test reports detected failure in FFR-related hardware modules; hex display shows error codes.
Ranked causes
- 1. FFR hardware module failure (unknown)
Run s5i FFR hardware test from diagnostics menu
- Test point
- Diagnostic test selection
- Expected
- Test completes with 's5 hardware test passed' message or displays specific hardware failure code
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Search button disabled in worklist dialog
Search button unavailable in Worklist dialog; cannot query with selected criteria
Ranked causes
- 1. Unconstrained worklist searches not allowed (common)
Verify minimum search criteria entered
- Test point
- Worklist dialog search fields
- Expected
- At least one search field populated or This System Only/date range/Modality selected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB peripheral not automatically detected
New USB peripheral connected to Workstation PC or USB extender is not transparently detected and enabled; device does not work or 'New Hardware Found' dialog appears
Ranked causes
- 1. Automatic detection and enabling process failure (common)
Power down system
- Test point
- Isolation transformer
- Expected
- System powered off
9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Windows hardware discovery windows display without Hardware Found dialog
Series of small windows display on-screen during peripheral connection; no 'New Hardware Found' dialog appears
Ranked causes
- 1. Normal Windows hardware discovery process in progress (occasional)
Observe discovery window sequence
- Test point
- Screen display
- Expected
- Series of small windows appear and close as discovery progresses
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Worklist button disabled in patient dialog
Worklist button unavailable; cannot access worklist server from patient entry screen
Ranked causes
- 1. No Worklist servers configured or GE Innova interface enabled (common)
Verify Worklist server configuration
- Test point
- DICOM configuration menu
- Expected
- At least one Worklist Server configured and GE Innova disabled
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Worklist query failed - unsupported character set patient safety
Worklist SCP sending patient information in language not supported by Volcano s5i
Ranked causes
- 1. Worklist SCP sending unsupported language character set (rare)
Verify character set support
- Test point
- DICOM Worklist configuration
- Expected
- Supported character set in use
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
104 | Pressure cable bad code | Pressure cable malfunction detected | error |
105 | Pressure cable error 2 | Pressure cable EEPROM error detected | error |
107 | Pressure wire has a short condition | Short circuit detected in pressure wire | critical |
108 | Cable signal unstable | Unstable wire signal, intermittent connection, or wire in patient body | warning |
109 | Cable error - calibration EEPROM error | Calibration EEPROM error with patient cable | error |
110 | Cable error - +5 volt error | +5V supply error detected in patient cable | error |
111 | Cable error - Vcc error | Vcc power supply error detected | error |
112 | Cable error - no calibration performed | Cable has not been calibrated | error |
113 | Cable error - patient cable ROM error | ROM memory error in patient cable | error |
114 | Cable error - patient cable RAM error | RAM memory error in patient cable | error |
115 | Cable error - offset error | Offset calibration error detected | error |
116 | Cable error - gain error | Gain calibration error detected | error |
1 beep | Refresh circuits reset | System powered on; normal startup sequence | info |
202 | No factory calibration | Calibration EEPROM is corrupt | critical |
204 | Low level aortic input error | Inactive or low aortic pressure input detected | error |
205 | Calibration error | Calibration EEPROM error detected | error |
206 | Isolated circuit power error | Power supply error for isolated circuits detected | critical |
207 | High voltage limit condition detected | High voltage limit exceeded | error |
212 | EEPROM write error | Error writing to EEPROM on S5 FFR System circuit board | error |
213 | DSP memory error | DSP memory error occurred | error |
215 | HVMON error | DSP detected HVMON error | error |
216 | Safety shutdown error | Transmit voltage safety shutdown triggered or failed | warning |
5 short, 1 long | Memory error - no memory detected | POST routines detect no system memory | critical |
8 beeps | Display memory read/write error | POST routines detect video adapter fault | critical |
Archivist failed to start | File storage subsystem initialization failed | System startup | critical |
Archivist internal error | File storage subsystem internal error | File storage operation | error |
Bad Command | Calibration cannot execute because Demo Mode is enabled | Aortic High Level calibration initiated with Demo Mode ON | error |
Bad Command | s5 Demo Mode is turned ON; real data not being transmitted | Attempting Aortic High Level calibration in Demo Mode | error |
Catheter Failure | Catheter drawing excessive current from patient interface module | Catheter plugged in draws abnormally high current | error |
Caution: hard disk filling | Low disk space warning | Less than 800 MB remaining | warning |
Caution: recording disallowed | Recording blocked due to insufficient space | Less than 100 MB remaining on disk | warning |
Current case still open. End before switching to FFR Mode. | System cannot switch to FFR mode while IVUS case is open | User attempts to switch to FFR mode with active case | warning |
Current patient data may be corrupt | Patient identification data corruption detected | Data validation failed on patient info | error |
Data missing from recording | Gaps detected in recorded data sequence | Recording or playback verification | error |
DICOM/Network Configuration - Check server fails | DICOM server verification failed after connection established | DICOM verification process attempts to query configured remote DICOM server | error |
DICOM/Network Configuration - Find host fails | DICOM host resolution failed during network configuration | DICOM configuration dialog attempts to locate and resolve DICOM server hostname | error |
DICOM Verification Failed - Association negotiation failed U | DICOM association negotiation failed due to Volcano s5i SCU AE Title not accepted | DICOM verification initiates association with remote DICOM SCP | error |
DICOM Verification Success with Warning: MODALITY_WORKLIST_F | DICOM verification succeeds but remote SCP does not support Worklist service | DICOM verification completes but Abstract Syntax negotiation for Worklist fails | warning |
DICOM Verification Success with Warning: STANDARD_US_MF Abst | DICOM verification succeeds but remote Storage SCP does not support Ultrasound Multiframe | DICOM verification completes but Abstract Syntax negotiation for STANDARD_US_MF fails | warning |
DICOM Verification Success with Warning: STANDARD_US_MF Tran | DICOM verification succeeds but Storage SCP does not accept configured transfer syntax | DICOM verification completes but Transfer Syntax negotiation fails | warning |
Directory file load read error | Error loading directory file | Loading directory file from storage | error |
Directory file not valid, rebuilding | Directory file inconsistency detected | Directory file and disk studies mismatch | warning |
Directory file update write error | Error writing updated directory file | Writing directory file to storage | error |
DVD drive failure | DVD drive hardware malfunction | DVD drive operation | critical |
DVD full | DVD storage capacity exceeded | Attempting to write to full DVD | warning |
DVD verify failed | Verify error after copying file to DVD | Writing patient study file to DVD | error |
E1 | Cannot communicate with RFID on catheter | RFID communication failure detected | error |
E1000-E1032 | ExecutionManager error | ExecutionManager module experiences error | critical |
E1001-E1002 | Minit module error range | Minit module experiences error | critical |
E1001-E1003 | MeasurementManager error range | MeasurementManager module experiences error | critical |
E1001-E1057 | FrameManager error range | FrameManager module experiences error | critical |
E1001-E1211 | ImageManager error range | ImageManager module experiences error | critical |
E100-E104 | MMAPI imaging module error range | MMAPI imaging module failure | critical |
E100n | StateMachine error or CardManager error | StateMachine or CardManager module experiences error | error |
E1033 | ExecutionManager network configuration error | Network configuration failure on startup | critical |
E1034 | ExecutionManager startup error | System startup initialization failure | critical |
E1035-E1038 | ExecutionManager error range | ExecutionManager experiences error condition | critical |
E1039-E1055 | ExecutionManager informational error range | ExecutionManager logs informational event | info |
E1056-E1084 | ExecutionManager error range | ExecutionManager experiences error condition | critical |
E108-E121 | MMAPI imaging module error range | MMAPI imaging subsystem failure | error |
E10nn | CardCoordinator error range | CardCoordinator module error | error |
E122-E123 | MMAPI critical error range | MMAPI experiences critical failure | critical |
E124-E125 | MMAPI error range | MMAPI module experiences error | error |
E126-E131 | MMAPI automatic shutdown error range | MMAPI detects critical condition, initiates auto-shutdown | critical |
E132-E133 | MMAPI error range | MMAPI module experiences error | error |
E134-E136 | MMAPI imaging error range | MMAPI imaging subsystem failure | critical |
E137 | MMAPI error | MMAPI module experiences error | error |
E138-E140 | MMAPI imaging error range | MMAPI imaging subsystem failure | critical |
E141-E904 | MMAPI error range | MMAPI experiences error condition | error |
E1nnn | CaseManager error (generic range) | CaseManager module experiences error condition | critical |
E2 | TGC memory failure | Time Gain Compensation memory error detected | error |
E3 | Analog high voltage failure | High voltage circuit malfunction detected | critical |
E4 | Pulse Width control pin stuck high or low | Pulse Width control pin hardware fault | error |
E5 | Catheter motor driver failure | Motor driver circuit malfunction | error |
E6 | Pullback motor driver failed | Pullback motor driver circuit malfunction | error |
End of file encountered | End of recorded file reached | Playback of recorded file | info |
Error during initialization. | Unrecoverable error during system initialization; Health Checker encountered error on startup | System startup; health checker operation failed | critical |
Hard disk full | Hard disk storage capacity exceeded | Attempting to record new data | warning |
Hardware did not respond | System did not receive response to command within allotted time | Calibration sequence initiated without hardware response | error |
Hardware did not respond | Command to hardware not executed within allotted time | During Aortic High Level calibration | error |
Health Checker module error at startup | Error occurred trying to execute Health Checker module during initialization | System startup; health checker determines hardware state | critical |
Incorrect system time or date | Time or date set incorrectly on system | User sets incorrect time/date during initialization or settings | info |
Maximum session record time exceeded | 90-minute session recording limit reached | Recording 90 minutes for single patient | warning |
Meridian could not restart VH+ and IVUS boards to Meridian M | Failure to restart VH+ or IVUS board when switching back from FFR to Meridian mode | System attempted to resume Meridian mode from FFR mode | error |
Meridian could not stop VH+ and IVUS boards for FFR Mode swi | Failure to stop VH+ or IVUS board when attempting FFR mode switch | FFR mode switch requested while system actively running VH | error |
Network Transfer Failed C-STORE Error: Out of Resources (a70 | DICOM C-STORE operation failed due to remote server resource exhaustion | Archive operation attempts to store image on remote DICOM server | error |
No Catheter | Patient interface module does not have a catheter connected or recognition delayed | Catheter not plugged in; PIM needs time to recognize new catheter; dialog boxes delay detection | warning |
No data in patient file | Patient file contains only demographics, no study data | Recalling patient study file | info |
No Paper or Ink | Printer is out of paper or ink; or paper issues (incorrect brand, inserted incorrectly) | Printer detects missing paper/ink; incorrect paper brand; wrong paper orientation | warning |
Out of Range | Input signal level outside acceptable calibration range | Input signal during calibration deviates more than ±10% from selected level | error |
PIM Configuration in progress. Please wait. | Hardware configuration in progress after PIM change | New PIM installed or detected | info |
PIM-ffr failure or cable fault (from Display Patient Cable S | PIM-ffr module not responding or cable connection malfunction detected. | Display Patient Cable Status diagnostic run; triggered when PIM module does not respond to test. | warning |
PIM not Detected | System unable to communicate with patient interface module unit | PIM connection lost or communication failure | error |
PIM Power Failure | Patient interface module detected undesirable high power condition for catheter in use | High power condition detected on PIM output to catheter | error |
Please insert a DVD | No DVD detected in drive | Attempting DVD write operation | warning |
Read error on playback | File read error during playback | Reading previously recorded data file | error |
Read error on recalling patient data | Error reading patient identification data | Recalling patient name, address or identification | error |
s5i FFR hardware test failure (specific codes in hex display | One or more FFR-related hardware modules detected as faulty during diagnostic test. | Automatic on diagnostic run; triggered when FFR board or module fails self-test. | error |
Shutting down will abort case archiving. | Pressing power button during case archive will cancel archive process | Power button pressed while system is archiving case | critical |
System does not boot up | System boot failure | Power-on boot sequence failed | critical |
System error detected. Auto-repair in progress, please wait. | Hardware failure detected; system attempting automatic self-repair | VH or IVUS card failure detected during operation | error |
System shutdown before ending case. | System was powered down prematurely before case was ended; recover mode initiated on restart | Power removed while case open; system restart after unexpected shutdown | warning |
There are no patient files on the DVD | DVD does not contain S5 FFR System study files | Attempting to recall patient data from DVD | warning |
Time does not maintain correct setting | System does not retain correct time/date after reboot | System powers off or reboots after time/date correction | error |
Unstable Signal | Input signal fluctuates excessively during calibration | Signal variation exceeds ±2.5 mmHg during high level calibration | warning |
Worklist Exception Unsupported Character Set (Tag: 00080005) | Worklist SCP returning patient data in unsupported character set | Worklist query receives response containing unsupported character encoding | error |
Write error on saving patient data | Error writing patient identification data | Saving patient name, address or other identification | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.