Volcano s5i: troubleshooting and error codes

The ServiceForge knowledge base holds 58 documented symptoms for the Volcano s5i, with 115 ranked causes, 119 diagnostic checks with test points and expected values, and 107 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

DICOM network configuration - host not found

System cannot locate DICOM server host; PING fails with host name not resolved

Ranked causes

  1. 1. Network cable not connected to Volcano s5i (common)
  2. 2. Remote server offline or not on network (common)
  3. 3. Defective network cable (occasional)
  4. 4. Invalid local Volcano s5i network configuration (occasional)
  5. 5. Invalid gateway IP address or subnet mask (occasional)
  6. 6. Invalid remote DICOM server IP address configuration (occasional)
  7. 7. Network speed/duplex auto negotiation failure (occasional)
First check

Verify network cable connection

Test point
Volcano s5i network port
Expected
Cable fully seated and connected

2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Control Panel keys do not work or respond

Control panel keyboard buttons not functioning; input not registering when buttons pressed

Ranked causes

  1. 1. USB cables or extender not fully connected (common)
  2. 2. Control panel keys damaged or binding (occasional)
  3. 3. Damaged or defective USB extender (occasional)
  4. 4. Control Panel hardware defective (rare)
First check

Verify USB cables and extender fully connected

Test point
USB connector ports on system and control panel; extender connector
Expected
All connectors fully seated and secured; no visible damage or bent pins

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No IVUS display or image shown

System boots and runs but IVUS imaging window does not appear or shows no image data

Ranked causes

  1. 1. PIM not connected or not powered (common)
  2. 2. Catheter or catheter simulator disconnected or defective (common)
  3. 3. Meridian analog/digital boards not properly seated or filter cards loose (occasional)
  4. 4. Meridian imaging boards defective (rare)
First check

Verify PIM connected to system interface

Test point
PIM connector to main system board
Expected
PIM cable fully seated and secured

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer does not work or print

Printer is connected but does not respond to print commands; nothing prints

Ranked causes

  1. 1. Printer not turned on or lacks paper/ink ribbon (common)
  2. 2. USB or power cabling not fully connected (common)
  3. 3. Printer not set as default or Print Manager not running (occasional)
  4. 4. Printer hardware defective (rare)
First check

Verify printer powered on and media loaded

Test point
Printer power button, paper tray, print ribbon
Expected
Printer powered on, LED lit, paper loaded, ribbon present and functional

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System does not power up caution

System will not turn on; workstation PC does not respond to power button

Ranked causes

  1. 1. AC power cord disconnected or improperly plugged in (common)
  2. 2. Isolation Transformer not enabled or set to wrong voltage (common)
  3. 3. Workstation PC power button location not found or power issue (occasional)
  4. 4. Workstation PC DC power supply failure (rare)
First check

Verify all AC power cords properly connected

Test point
Power inlet connectors at Isolation Transformer and system chassis
Expected
Connectors firmly seated, no visible damage

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System has no power

No display, indicators, or control feedback when power switch activated

Ranked causes

  1. 1. System powered off or power switch not activated (common)
  2. 2. Isolation transformer off (s5i only) (common)
  3. 3. Power plug disconnected (common)
  4. 4. Wall outlet power not available (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to archive to DVD

System cannot write to DVD media; DVD archive function fails with error or drive not accessible

Ranked causes

  1. 1. Improper DVD media type being used (common)
  2. 2. Cases not selected for archive operation (common)
  3. 3. DVD drive not detected by system BIOS or software (occasional)
  4. 4. DVD drive hardware defective (rare)
First check

Verify DVD-R media type being used

Test point
DVD media in drive and packaging
Expected
Media labeled DVD-R (not DVD-RW, DVD+R, or other type)

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Aortic High Level input

Procedure to calibrate the Aortic Signal input on rear panel so Pa (pressure-aortic) display value matches physiological monitor in Cath lab. Required at installation and when equipment changes.

Ranked causes

  1. 1. Aortic input requires calibration for display accuracy (common)
  2. 2. Demo Mode enabled prevents calibration (occasional)
  3. 3. Hardware communication timeout (rare)
First check

Aortic input connected to rear panel and at zero level

Test point
Aortic Input connector, rear panel
Expected
Zero level signal

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Catheter Type not recognized by system

System does not identify catheter type when connected; 'Catheter Type not recognized' error or message

Ranked causes

  1. 1. Contamination or corrosion on catheter connector pins or PIM sockets (common)
  2. 2. Install error; catheter not fully connected on first insertion (common)
  3. 3. Mechanical damage to PIM connector or defective catheter (occasional)
First check

Disconnect and reconnect catheter after 3 second delay

Test point
PIM catheter connector
Expected
Catheter fully seated and locked; type recognized by system

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

DICOM verification failed - server unreachable

System cannot connect to remote DICOM host; socket connection fails

Ranked causes

  1. 1. Invalid port number for remote DICOM server (common)
  2. 2. Remote DICOM application not running or offline (common)
  3. 3. DICOM server not configured to accept Volcano s5i IP (occasional)
First check

Verify DICOM SCP port number

Test point
DICOM server configuration
Expected
Port number matches Volcano s5i configuration

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

FFR measurements are incorrect or erratic patient safety

FFR pressure values displayed are inaccurate, unstable, or not matching expected values; measurements inconsistent

Ranked causes

  1. 1. Faulty or uncalibrated pressure wire (common)
  2. 2. Faulty PIM-ffr interface or signal conditioning (occasional)
  3. 3. Faulty VH board affecting FFR processing (rare)
First check

Test FFR measurements with known good pressure wire

Test point
Pressure wire connected to FFR input; display readings
Expected
Stable pressure readings; values consistent and within expected range

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No FFR input signal or measurement patient safety

Fractional Flow Reserve (FFR) system not acquiring or displaying pressure wire data; no FFR readings

Ranked causes

  1. 1. Faulty pressure wire (common)
  2. 2. Faulty PIM-ffr interface board (occasional)
  3. 3. Faulty VH imaging board (rare)
First check

Test FFR input with another known good pressure wire

Test point
FFR pressure wire connector on PIM-ffr or PIM
Expected
FFR input detected; pressure readings display on monitor

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No Heart-beating icon with ECG input or No VH data patient safety

ECG waveform not triggering heart-beat icon animation; VH (Virtual Histology) imaging card not functioning or ECG signal not detected

Ranked causes

  1. 1. Improper ECG signal or cabling not fully connected (common)
  2. 2. VH jumper cable not installed between Meridian digital board and VH board (occasional)
  3. 3. VH imaging board defective (rare)
First check

Test ECG input with ECG simulator

Test point
ECG input connector (rear panel)
Expected
Heart-beating icon animates when ECG simulator connected and running

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No scheduled procedures returned from worklist search patient safety

Worklist query returns empty result; no procedures displayed for patient or search criteria

Ranked causes

  1. 1. Volcano s5i system clock date not set to today (occasional)
  2. 2. HIS/CIS does not have Volcano s5i scheduled station AE title configured (occasional)
  3. 3. HIS/CIS does not schedule procedures for IVUS modality (occasional)
First check

Verify Volcano s5i system clock date

Test point
System settings
Expected
Current date set to today

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No video displayed on monitor

Monitor is powered on but shows no image; blank or black screen

Ranked causes

  1. 1. AC power interruption to system (common)
  2. 2. Video signal path interrupted or misconfigured (common)
  3. 3. Video board defective or display damaged (occasional)
First check

Check monitor power LED status

Test point
Monitor front panel LED
Expected
Power LED illuminated or monitor powered on

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Operating System fails to boot caution

System hangs or stops during OS boot sequence; does not reach Windows login or application startup

Ranked causes

  1. 1. DC power supply insufficient or failing (occasional)
  2. 2. Hard drive not detected or SATA not configured correctly (occasional)
  3. 3. Hard drive data corruption or operating system damage (occasional)
First check

Verify Workstation PC system DC power output

Test point
Main system board power connector
Expected
+12V, +5V, +3.3V supplies within 5% of nominal

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PIM not Found Error Message displayed

System displays 'PIM not Found' error; imaging interface not detected at startup

Ranked causes

  1. 1. PIM cabling not fully connected (common)
  2. 2. Bent connector pins on PIM or system connector (occasional)
  3. 3. PIM physically damaged or defective (rare)
First check

Verify PIM connector fully connected to system

Test point
PIM connector and system connector
Expected
Connectors fully seated and secured; no gaps visible

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System boots to Volcano logo screen but User Interface not visible

System boot process reaches logo screen but User Interface window does not appear; mouse cursor active

Ranked causes

  1. 1. User Interface window in background, not in foreground (common)
  2. 2. System hard drive corrupt or defective (occasional)
  3. 3. Hard drive defective and cannot be recovered (rare)
First check

Press Escape key to bring UI window to foreground

Test point
System keyboard
Expected
Escape key press triggers User Interface to appear

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Trackball does not work or responds improperly

Control panel trackball unresponsive; cursor does not move or moves erratically when trackball rotated

Ranked causes

  1. 1. USB cables or extender not fully connected (common)
  2. 2. Damaged or defective USB extender (occasional)
  3. 3. Control Panel or trackball hardware defective (rare)
First check

Verify USB cables and extender fully connected

Test point
USB connector ports at system and control panel; extender
Expected
All connectors fully seated; no damaged or bent pins

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

VHFM pipeline test failure or degraded waveforms

VHFM pipeline test waveforms differ greatly from demo display; data processing not functioning properly.

Ranked causes

  1. 1. Software, memory, or PC problem (no Volcano-unique hardware involved) (unknown)
  2. 2. System software or PC issue requiring full restart (unknown)
  3. 3. Software corruption or obsolete version (unknown)
First check

Initiate VHFM pipeline test from diagnostics menu

Test point
Diagnostic test Live screen
Expected
Waveforms scroll across screen similar to demo display with same settings

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display not functioning

System lights and fans operating but no video output to monitor

Ranked causes

  1. 1. Monitor powered off or disconnected (common)
  2. 2. Backup battery failed (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Joystick controller only has one axis of movement

Joystick responds to X or Y movement but not both axes; movement restricted to one direction

Ranked causes

  1. 1. Damaged or defective joystick (common)
  2. 2. Damaged USB extender affecting joystick signal (occasional)
First check

Test joystick movement in all directions

Test point
Joystick handle during system use
Expected
Smooth response to forward/back (Y) and left/right (X) movements

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Only one monitor displays User Interface

Dual display setup not working; only one of two monitors shows the system UI

Ranked causes

  1. 1. One display disconnected or not powered on during boot (common)
  2. 2. Display disabled in Windows or ATI control manager (occasional)
First check

Verify both displays connected and powered on

Test point
Display power buttons, DVI cable connectors at graphics card
Expected
Both display power indicators ON; both DVI cables firmly seated

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PIM-ffr not responding or cable fault patient safety

Display Patient Cable Status diagnostic indicates PIM-ffr module or cable malfunction; PIM icon at lower right of screen shows fault indication.

Ranked causes

  1. 1. PIM-ffr module cable loose or disconnected (common)
  2. 2. PIM-ffr module malfunction (occasional)
First check

Visually inspect PIM-ffr cable connection status

Test point
PIM-ffr connector
Expected
Cable fully seated; no visible damage; connector locked

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PIMr pullback function improper or not tracking correctly

PIMr (motorized pullback device) not retracting catheter smoothly; movement is obstructed, intermittent, or incorrect

Ranked causes

  1. 1. Obstruction to PIMr cable or catheter movement (common)
  2. 2. Faulty PIMr motor or drive mechanism (occasional)
First check

Verify PIMr cable and catheter free of obstruction

Test point
PIMr cable routing and imaging catheter path
Expected
Cable and catheter not kinked, bent, or pinched; free movement possible

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to archive to DICOM network

System cannot send DICOM data to network server; archive to DICOM fails with error or connection lost

Ranked causes

  1. 1. Network cabling not fully connected (common)
  2. 2. Network set-up error or incorrect IP address configuration (occasional)
First check

Verify network Ethernet cable connected to system and switch

Test point
Network port on system and switch/router
Expected
Cable fully seated; LED indicators on port showing connection

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Video display lacks fundamental color or shows only R, G, or B

Color display showing only one primary color (red, green, or blue) or missing colors; DVI signal incomplete

Ranked causes

  1. 1. DVI connector not fully seated or screwed in (common)
  2. 2. Video cable has broken or damaged wire in DVI connector (occasional)
First check

Inspect DVI connector seating and retaining screws

Test point
DVI connectors at display and graphics card
Expected
Connector fully inserted; retaining screws tight; no visible damage or missing pins

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Aortic High Level calibration fails with Hardware did not respond patient safety

System did not receive response from hardware within the allotted timeout period during Aortic High Level calibration procedure.

Ranked causes

  1. 1. Hardware communication timeout or connection failure (common)
First check

Verify hardware connection and responsiveness

Test point
Aortic Signal input connector on rear panel
Expected
Secure connection; hardware responds to commands

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Aortic High Level calibration fails with Out of Range error patient safety

Input signal level is too low or too high, exceeding the ±10% tolerance band of the selected calibration level (100, 200, or 300).

Ranked causes

  1. 1. Input signal outside ±10% tolerance of selected calibration level (common)
First check

Measure or verify input signal level

Test point
Aortic Signal input on rear panel
Expected
Within ±10% of selected calibration level (100, 200, or 300), ±10% of selected level
Tools
Physiological monitor or calibrator device

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Aortic High Level calibration fails with Unstable Signal error patient safety

Input signal fluctuates more than ±2.5 mmHg around the selected calibration level, preventing successful calibration.

Ranked causes

  1. 1. Input signal source or connection is unstable (common)
First check

Observe input signal stability during calibration attempt

Test point
Aortic Signal input on rear panel
Expected
Signal stable within ±2.5 mmHg of selected level, ±2.5 mmHg
Tools
Oscilloscope or physiological monitor display

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Aortic High Level for FFR patient safety

Setup and calibration of the Aortic Signal input on the rear panel to ensure Pa (pressure-aortic) values match the Cath lab's physiological equipment. Performed once at installation or when equipment changes.

Ranked causes

  1. 1. Initial installation or equipment changes requiring aortic signal calibration (common)
First check

Zero calibration of Aortic Input signal

Test point
Aortic Signal input on rear panel
Expected
System interprets input as zero; brief wait message displays

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Input aortic high level restore to factory default

Reset aortic high level calibration to factory-determined setting when calibration has failed or needs to be discarded

Ranked causes

  1. 1. Calibration failed or requires reset (unknown)
First check

Verify calibration failure or need for reset

Test point
Input Calibration section display
Expected
Failed Calibration Message indication visible

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: PIM-ffr input accuracy check

Verify PIM-ffr pressure reading accuracy using Pressure Guide Wire Simulator fixture; checks zero offset and 300 mmHg reading within tolerance

Ranked causes

  1. 1. PIM-ffr calibration out of specification (unknown)
First check

Verify PIM-ffr auto-zero function

Test point
Live tab digital Pd display
Expected
zero

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

DICOM association negotiation failed - invalid called AE title

DICOM verification fails; remote DICOM Storage or Worklist SCP AE title incorrect

Ranked causes

  1. 1. Incorrect Application Entity Title for DICOM Storage or Worklist SCP (common)
First check

Verify remote DICOM SCP AE title

Test point
Remote DICOM server configuration
Expected
AE title matches Volcano s5i configuration

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

DICOM association negotiation failed - invalid calling AE title

DICOM verification fails; Volcano s5i SCU AE title incorrect or not configured on remote

Ranked causes

  1. 1. Incorrect Volcano s5i DICOM Storage or Worklist SCU AE title (common)
First check

Verify Volcano s5i SCU AE title

Test point
Volcano s5i DICOM configuration
Expected
SCU AE title defined

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

DICOM verification warning - Storage abstract syntax rejected

DICOM verification succeeds with warning; Storage SCP does not accept Ultrasound Multiframe

Ranked causes

  1. 1. Storage SCP not a Storage server or does not accept Ultrasound Multiframe SOP class (occasional)
First check

Verify Storage SCP supports Ultrasound Multiframe

Test point
Remote DICOM server configuration
Expected
STANDARD_US_MF (Ultrasound Multiframe) supported

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

DICOM verification warning - Storage transfer syntax rejected

DICOM verification succeeds with warning; Storage SCP does not accept JPEG lossy compression

Ranked causes

  1. 1. Storage SCP does not accept JPEG lossy transfer syntax (occasional)
First check

Verify Storage SCP compression setting

Test point
DICOM Storage SCP configuration
Expected
Compression set to Uncompressed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

DICOM verification warning - Worklist abstract syntax rejected

DICOM verification succeeds with warning; Worklist SCP does not support Worklist server

Ranked causes

  1. 1. Configured Worklist SCP is not a Worklist server (occasional)
First check

Verify Worklist SCP type

Test point
DICOM Worklist SCP configuration
Expected
SCP configured as Worklist server

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive not bootable error

System displays 'Drive not bootable. Insert bootable disk.' message

Ranked causes

  1. 1. Hard drive failure or corruption (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

GUI image not aligned on screen or slightly off screen

User interface window positioned incorrectly on display; partially off edge or misaligned

Ranked causes

  1. 1. LCD display did not switch video modes correctly during boot (common)
First check

Reboot display or use display menu to rescan synchronization

Test point
Display power button or on-screen menu
Expected
Display re-detects video signal; GUI becomes properly centered

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image colors not displayed correctly on DICOM viewer

DICOM Viewer not rendering color palette images correctly

Ranked causes

  1. 1. Workstation does not handle color palette images correctly (occasional)
First check

Verify Storage SCP compression format

Test point
DICOM Storage SCP configuration
Expected
Compression set to JPEG for color palette support

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incorrect PIM cable length configuration caution

PIM cable length setting does not match physically installed cable, leading to inaccurate measurement results

Ranked causes

  1. 1. PIM cable length setting mismatch with installed hardware (common)
First check

Verify PIM cable length setting matches installed cable

Test point
Field Service Dialog, PIM Cable Length tab
Expected
Selected length matches physical cable

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incorrect PIM cable length setting caution

PIM cable length setting does not match the physical cable installed, resulting in incorrect measurement results

Ranked causes

  1. 1. PIM cable length misconfigured in Field Service Dialog (common)
First check

Verify PIM cable length setting matches installed cable

Test point
Field Service Dialog - PIM Cable Length tab
Expected
Selected length matches physical PIM cable installed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

IVUS display appears noisy or has excessive noise

IVUS image shows visual noise, artifacts, or degraded signal quality; not due to catheter issue

Ranked causes

  1. 1. Extraneous equipment radiating EMI at transducer frequencies (common)
First check

Identify EMI sources near system transducer cables

Test point
Environment around transducer and PIM cabling
Expected
No active RF/microwave/strong EM radiating equipment in close proximity

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

IVUS image vertically or horizontally ellipsoid shaped

IVUS image appears stretched or compressed in one axis; aspect ratio incorrect

Ranked causes

  1. 1. Display resolution not set to native 1280x1024 aspect ratio mode (common)
First check

Verify display resolution set to 1280x1024

Test point
Windows Display Settings resolution dropdown
Expected
Resolution shows 1280x1024; refresh rate standard

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

IVUS images cannot be displayed on DICOM viewer

DICOM Workstation/Viewer does not support IVUS image modality

Ranked causes

  1. 1. DICOM Workstation/Viewer does not support IVUS modality (occasional)
First check

Verify DICOM Storage Server modality configuration

Test point
DICOM Storage Server settings
Expected
Modality set to US or supported IVUS modality

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Multiple logical devices detected instead of single peripheral

'New Hardware Found' dialog displays repeatedly; system detects and configures multiple sub-devices for what is physically one peripheral

Ranked causes

  1. 1. Multi-function USB device with multiple logical sub-devices (occasional)
First check

Monitor for repeated 'New Hardware Found' dialogs

Test point
Screen display
Expected
Multiple dialogs appear sequentially, one per sub-device

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Network transfer failed - C-STORE interoperability error

DICOM archive to network fails with C-STORE error; status code and error description variable

Ranked causes

  1. 1. Interoperability problem with vendor device (rare)
First check

Debug via DICOM/Network Configuration dialog

Test point
DICOM Configuration menu
Expected
Detailed error diagnostics available

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Network transfer failed - C-STORE out of resources

DICOM archive to network fails; C-STORE error with status code a700

Ranked causes

  1. 1. DICOM SCP experiencing resource problem or disk full (common)
First check

Check remote DICOM server resources

Test point
Remote server
Expected
Adequate disk space and memory available

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No disk in floppy drive error

System displays 'No Disk in Floppy Drive' message

Ranked causes

  1. 1. Backup battery failing (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Restore input calibration to factory default

Reset Aortic High Level calibration to factory-determined setting when calibration fails or needs to be reset

Ranked causes

  1. 1. Calibration failure or need to reset to known state (unknown)
First check

Verify Restore button is available in Input Calibration section

Test point
Input Calibration section
Expected
Restore button visible and clickable

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Revolution image appears noisy or has noise artifacts

Revolution imaging mode showing visual noise or poor quality in image; excessive grain or interference

Ranked causes

  1. 1. Air in imaging catheter preventing proper signal transmission (common)
First check

Observe Revolution image quality for air artifacts

Test point
System display during Revolution imaging
Expected
Clear image without noise or shadowing characteristic of air in catheter

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

s5i FFR hardware test failure

FFR hardware test reports detected failure in FFR-related hardware modules; hex display shows error codes.

Ranked causes

  1. 1. FFR hardware module failure (unknown)
First check

Run s5i FFR hardware test from diagnostics menu

Test point
Diagnostic test selection
Expected
Test completes with 's5 hardware test passed' message or displays specific hardware failure code

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Search button disabled in worklist dialog

Search button unavailable in Worklist dialog; cannot query with selected criteria

Ranked causes

  1. 1. Unconstrained worklist searches not allowed (common)
First check

Verify minimum search criteria entered

Test point
Worklist dialog search fields
Expected
At least one search field populated or This System Only/date range/Modality selected

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

USB peripheral not automatically detected

New USB peripheral connected to Workstation PC or USB extender is not transparently detected and enabled; device does not work or 'New Hardware Found' dialog appears

Ranked causes

  1. 1. Automatic detection and enabling process failure (common)
First check

Power down system

Test point
Isolation transformer
Expected
System powered off

9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Windows hardware discovery windows display without Hardware Found dialog

Series of small windows display on-screen during peripheral connection; no 'New Hardware Found' dialog appears

Ranked causes

  1. 1. Normal Windows hardware discovery process in progress (occasional)
First check

Observe discovery window sequence

Test point
Screen display
Expected
Series of small windows appear and close as discovery progresses

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Worklist button disabled in patient dialog

Worklist button unavailable; cannot access worklist server from patient entry screen

Ranked causes

  1. 1. No Worklist servers configured or GE Innova interface enabled (common)
First check

Verify Worklist server configuration

Test point
DICOM configuration menu
Expected
At least one Worklist Server configured and GE Innova disabled

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Worklist query failed - unsupported character set patient safety

Worklist SCP sending patient information in language not supported by Volcano s5i

Ranked causes

  1. 1. Worklist SCP sending unsupported language character set (rare)
First check

Verify character set support

Test point
DICOM Worklist configuration
Expected
Supported character set in use

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
104Pressure cable bad codePressure cable malfunction detectederror
105Pressure cable error 2Pressure cable EEPROM error detectederror
107Pressure wire has a short conditionShort circuit detected in pressure wirecritical
108Cable signal unstableUnstable wire signal, intermittent connection, or wire in patient bodywarning
109Cable error - calibration EEPROM errorCalibration EEPROM error with patient cableerror
110Cable error - +5 volt error+5V supply error detected in patient cableerror
111Cable error - Vcc errorVcc power supply error detectederror
112Cable error - no calibration performedCable has not been calibratederror
113Cable error - patient cable ROM errorROM memory error in patient cableerror
114Cable error - patient cable RAM errorRAM memory error in patient cableerror
115Cable error - offset errorOffset calibration error detectederror
116Cable error - gain errorGain calibration error detectederror
1 beepRefresh circuits resetSystem powered on; normal startup sequenceinfo
202No factory calibrationCalibration EEPROM is corruptcritical
204Low level aortic input errorInactive or low aortic pressure input detectederror
205Calibration errorCalibration EEPROM error detectederror
206Isolated circuit power errorPower supply error for isolated circuits detectedcritical
207High voltage limit condition detectedHigh voltage limit exceedederror
212EEPROM write errorError writing to EEPROM on S5 FFR System circuit boarderror
213DSP memory errorDSP memory error occurrederror
215HVMON errorDSP detected HVMON errorerror
216Safety shutdown errorTransmit voltage safety shutdown triggered or failedwarning
5 short, 1 longMemory error - no memory detectedPOST routines detect no system memorycritical
8 beepsDisplay memory read/write errorPOST routines detect video adapter faultcritical
Archivist failed to startFile storage subsystem initialization failedSystem startupcritical
Archivist internal errorFile storage subsystem internal errorFile storage operationerror
Bad CommandCalibration cannot execute because Demo Mode is enabledAortic High Level calibration initiated with Demo Mode ONerror
Bad Commands5 Demo Mode is turned ON; real data not being transmittedAttempting Aortic High Level calibration in Demo Modeerror
Catheter FailureCatheter drawing excessive current from patient interface moduleCatheter plugged in draws abnormally high currenterror
Caution: hard disk fillingLow disk space warningLess than 800 MB remainingwarning
Caution: recording disallowedRecording blocked due to insufficient spaceLess than 100 MB remaining on diskwarning
Current case still open. End before switching to FFR Mode.System cannot switch to FFR mode while IVUS case is openUser attempts to switch to FFR mode with active casewarning
Current patient data may be corruptPatient identification data corruption detectedData validation failed on patient infoerror
Data missing from recordingGaps detected in recorded data sequenceRecording or playback verificationerror
DICOM/Network Configuration - Check server failsDICOM server verification failed after connection establishedDICOM verification process attempts to query configured remote DICOM servererror
DICOM/Network Configuration - Find host failsDICOM host resolution failed during network configurationDICOM configuration dialog attempts to locate and resolve DICOM server hostnameerror
DICOM Verification Failed - Association negotiation failed UDICOM association negotiation failed due to Volcano s5i SCU AE Title not acceptedDICOM verification initiates association with remote DICOM SCPerror
DICOM Verification Success with Warning: MODALITY_WORKLIST_FDICOM verification succeeds but remote SCP does not support Worklist serviceDICOM verification completes but Abstract Syntax negotiation for Worklist failswarning
DICOM Verification Success with Warning: STANDARD_US_MF AbstDICOM verification succeeds but remote Storage SCP does not support Ultrasound MultiframeDICOM verification completes but Abstract Syntax negotiation for STANDARD_US_MF failswarning
DICOM Verification Success with Warning: STANDARD_US_MF TranDICOM verification succeeds but Storage SCP does not accept configured transfer syntaxDICOM verification completes but Transfer Syntax negotiation failswarning
Directory file load read errorError loading directory fileLoading directory file from storageerror
Directory file not valid, rebuildingDirectory file inconsistency detectedDirectory file and disk studies mismatchwarning
Directory file update write errorError writing updated directory fileWriting directory file to storageerror
DVD drive failureDVD drive hardware malfunctionDVD drive operationcritical
DVD fullDVD storage capacity exceededAttempting to write to full DVDwarning
DVD verify failedVerify error after copying file to DVDWriting patient study file to DVDerror
E1Cannot communicate with RFID on catheterRFID communication failure detectederror
E1000-E1032ExecutionManager errorExecutionManager module experiences errorcritical
E1001-E1002Minit module error rangeMinit module experiences errorcritical
E1001-E1003MeasurementManager error rangeMeasurementManager module experiences errorcritical
E1001-E1057FrameManager error rangeFrameManager module experiences errorcritical
E1001-E1211ImageManager error rangeImageManager module experiences errorcritical
E100-E104MMAPI imaging module error rangeMMAPI imaging module failurecritical
E100nStateMachine error or CardManager errorStateMachine or CardManager module experiences errorerror
E1033ExecutionManager network configuration errorNetwork configuration failure on startupcritical
E1034ExecutionManager startup errorSystem startup initialization failurecritical
E1035-E1038ExecutionManager error rangeExecutionManager experiences error conditioncritical
E1039-E1055ExecutionManager informational error rangeExecutionManager logs informational eventinfo
E1056-E1084ExecutionManager error rangeExecutionManager experiences error conditioncritical
E108-E121MMAPI imaging module error rangeMMAPI imaging subsystem failureerror
E10nnCardCoordinator error rangeCardCoordinator module errorerror
E122-E123MMAPI critical error rangeMMAPI experiences critical failurecritical
E124-E125MMAPI error rangeMMAPI module experiences errorerror
E126-E131MMAPI automatic shutdown error rangeMMAPI detects critical condition, initiates auto-shutdowncritical
E132-E133MMAPI error rangeMMAPI module experiences errorerror
E134-E136MMAPI imaging error rangeMMAPI imaging subsystem failurecritical
E137MMAPI errorMMAPI module experiences errorerror
E138-E140MMAPI imaging error rangeMMAPI imaging subsystem failurecritical
E141-E904MMAPI error rangeMMAPI experiences error conditionerror
E1nnnCaseManager error (generic range)CaseManager module experiences error conditioncritical
E2TGC memory failureTime Gain Compensation memory error detectederror
E3Analog high voltage failureHigh voltage circuit malfunction detectedcritical
E4Pulse Width control pin stuck high or lowPulse Width control pin hardware faulterror
E5Catheter motor driver failureMotor driver circuit malfunctionerror
E6Pullback motor driver failedPullback motor driver circuit malfunctionerror
End of file encounteredEnd of recorded file reachedPlayback of recorded fileinfo
Error during initialization.Unrecoverable error during system initialization; Health Checker encountered error on startupSystem startup; health checker operation failedcritical
Hard disk fullHard disk storage capacity exceededAttempting to record new datawarning
Hardware did not respondSystem did not receive response to command within allotted timeCalibration sequence initiated without hardware responseerror
Hardware did not respondCommand to hardware not executed within allotted timeDuring Aortic High Level calibrationerror
Health Checker module error at startupError occurred trying to execute Health Checker module during initializationSystem startup; health checker determines hardware statecritical
Incorrect system time or dateTime or date set incorrectly on systemUser sets incorrect time/date during initialization or settingsinfo
Maximum session record time exceeded90-minute session recording limit reachedRecording 90 minutes for single patientwarning
Meridian could not restart VH+ and IVUS boards to Meridian MFailure to restart VH+ or IVUS board when switching back from FFR to Meridian modeSystem attempted to resume Meridian mode from FFR modeerror
Meridian could not stop VH+ and IVUS boards for FFR Mode swiFailure to stop VH+ or IVUS board when attempting FFR mode switchFFR mode switch requested while system actively running VHerror
Network Transfer Failed C-STORE Error: Out of Resources (a70DICOM C-STORE operation failed due to remote server resource exhaustionArchive operation attempts to store image on remote DICOM servererror
No CatheterPatient interface module does not have a catheter connected or recognition delayedCatheter not plugged in; PIM needs time to recognize new catheter; dialog boxes delay detectionwarning
No data in patient filePatient file contains only demographics, no study dataRecalling patient study fileinfo
No Paper or InkPrinter is out of paper or ink; or paper issues (incorrect brand, inserted incorrectly)Printer detects missing paper/ink; incorrect paper brand; wrong paper orientationwarning
Out of RangeInput signal level outside acceptable calibration rangeInput signal during calibration deviates more than ±10% from selected levelerror
PIM Configuration in progress. Please wait.Hardware configuration in progress after PIM changeNew PIM installed or detectedinfo
PIM-ffr failure or cable fault (from Display Patient Cable SPIM-ffr module not responding or cable connection malfunction detected.Display Patient Cable Status diagnostic run; triggered when PIM module does not respond to test.warning
PIM not DetectedSystem unable to communicate with patient interface module unitPIM connection lost or communication failureerror
PIM Power FailurePatient interface module detected undesirable high power condition for catheter in useHigh power condition detected on PIM output to cathetererror
Please insert a DVDNo DVD detected in driveAttempting DVD write operationwarning
Read error on playbackFile read error during playbackReading previously recorded data fileerror
Read error on recalling patient dataError reading patient identification dataRecalling patient name, address or identificationerror
s5i FFR hardware test failure (specific codes in hex displayOne or more FFR-related hardware modules detected as faulty during diagnostic test.Automatic on diagnostic run; triggered when FFR board or module fails self-test.error
Shutting down will abort case archiving.Pressing power button during case archive will cancel archive processPower button pressed while system is archiving casecritical
System does not boot upSystem boot failurePower-on boot sequence failedcritical
System error detected. Auto-repair in progress, please wait.Hardware failure detected; system attempting automatic self-repairVH or IVUS card failure detected during operationerror
System shutdown before ending case.System was powered down prematurely before case was ended; recover mode initiated on restartPower removed while case open; system restart after unexpected shutdownwarning
There are no patient files on the DVDDVD does not contain S5 FFR System study filesAttempting to recall patient data from DVDwarning
Time does not maintain correct settingSystem does not retain correct time/date after rebootSystem powers off or reboots after time/date correctionerror
Unstable SignalInput signal fluctuates excessively during calibrationSignal variation exceeds ±2.5 mmHg during high level calibrationwarning
Worklist Exception Unsupported Character Set (Tag: 00080005)Worklist SCP returning patient data in unsupported character setWorklist query receives response containing unsupported character encodingerror
Write error on saving patient dataError writing patient identification dataSaving patient name, address or other identificationerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI