Volcano s5: troubleshooting and error codes

The ServiceForge knowledge base holds 47 documented symptoms for the Volcano s5, with 113 ranked causes, 89 diagnostic checks with test points and expected values, and 108 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Archive to Network fails - DICOM/Network Configuration Check server fails

Error message 'DICOM Verification Failed - Failed to connect to remote host.' or C-STORE errors. Cannot connect to or transfer data to remote DICOM server.

Ranked causes

  1. 1. Invalid port number assignment for remote DICOM Server (common)
  2. 2. Remote DICOM Application not running or not on-line (common)
  3. 3. Remote DICOM Application not configured to accept socket connection from Volcano s5 System IP Address (occasional)
  4. 4. DICOM SCP server experiencing resource problem or hard drive full (occasional)
  5. 5. Interoperability problem with vendor device (rare)
  6. 6. Defective Volcano s5 System NIC (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

DICOM/Network Configuration - Find host fails

Error message 'PING Failed. The host name could not be found.' Unable to locate remote DICOM server by hostname or IP address.

Ranked causes

  1. 1. Network cable not connected to Volcano s5 System (common)
  2. 2. Incorrect network port enabled (common)
  3. 3. Defective network cable (occasional)
  4. 4. Invalid Volcano s5 System network configuration (occasional)
  5. 5. Invalid remote DICOM server IP address configuration (occasional)
  6. 6. Invalid remote DICOM server Gateway IP address or Subnet Mask configuration (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No display on Internal LCD

Internal LCD monitor shows no image

Ranked causes

  1. 1. AC power interruption to the system (common)
  2. 2. DVI video signal interruption (occasional)
  3. 3. Video board defective (occasional)
  4. 4. Display is damaged (rare)
  5. 5. Single Board Computer (SBC) failure (occasional)
First check

Check power LED status on monitor

Test point
Monitor power indicator
Expected
LED illuminated

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No Heart-beating icon with ECG input / FFR option - No ECG waveform / No VH data patient safety

ECG input not received; FFR and VH functions non-functional; no heartbeat icon displayed

Ranked causes

  1. 1. Improper ECG signal being used (common)
  2. 2. Hemodynamic system not configured for ECG output (occasional)
  3. 3. ECG cable defective (occasional)
  4. 4. Cabling not fully connected (occasional)
  5. 5. VH imaging card defective (occasional)
First check

Test ECG signal with simulator

Test point
ECG input connector
Expected
ECG waveform appears; heartbeat icon displayed

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PIM not Found Error Message

System displays 'PIM not Found' error; PIM module not detected

Ranked causes

  1. 1. Cabling not fully connected (common)
  2. 2. PIM connector not fully connected (common)
  3. 3. Bent connector pins (occasional)
  4. 4. PIM cord damaged (occasional)
  5. 5. PIM damaged (occasional)
First check

Verify PIM cable connection

Test point
PIM cabling connectors
Expected
All connectors fully seated and tight

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer does not work

Printer fails to produce output

Ranked causes

  1. 1. Cabling or extender not fully connected (common)
  2. 2. Printer is off or out of supplies (common)
  3. 3. Printer not set as default (occasional)
  4. 4. Print Manager not running (occasional)
  5. 5. Printer is defective (occasional)
First check

Verify USB and power cable connections

Test point
Printer USB and power connectors
Expected
All cables fully seated and secure

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Catheter Type not recognized

System does not identify catheter type; type not displayed

Ranked causes

  1. 1. Install error (common)
  2. 2. Contamination or corrosion on pins or sockets in PIM catheter connections (occasional)
  3. 3. Mechanical damage to PIM (occasional)
  4. 4. Damaged catheter (occasional)
First check

Disconnect and reinstall catheter

Test point
PIM catheter connector
Expected
Catheter type recognized after reinsertion

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No External VGA Video is displayed

External VGA monitor connected to I/O panel shows no image

Ranked causes

  1. 1. AC power interruption to the system (common)
  2. 2. Video signal interruption (occasional)
  3. 3. Video board defective (occasional)
  4. 4. Single Board Computer (SBC) failure (occasional)
First check

Check LED status on external monitor

Test point
Monitor power indicator
Expected
LED illuminated

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No IVUS display

IVUS imaging does not appear on screen

Ranked causes

  1. 1. Defective Catheter (common)
  2. 2. Defective PIM (occasional)
  3. 3. Defective hardware (occasional)
  4. 4. Interconnection failures (occasional)
First check

Ensure PIM is connected

Test point
PIM connector
Expected
Connector fully seated

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Operating System fails to boot

System powers on but OS does not load; boot fails before OS presentation

Ranked causes

  1. 1. Possible DC power interruption (occasional)
  2. 2. Mechanical error, parts not seated/mounted correctly (occasional)
  3. 3. Hard drive data corruption (occasional)
  4. 4. Defective Single Board Computer (SBC) (occasional)
First check

Measure system DC power

Test point
CPU and hard drive power connectors
Expected
Correct voltage present (per PSU spec)

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to archive to DICOM

System cannot send patient cases to DICOM network destination

Ranked causes

  1. 1. Cabling not fully connected (common)
  2. 2. Network set-up error (occasional)
  3. 3. Incorrect IP address (occasional)
  4. 4. Incorrect port - see FSB-006 (occasional)
First check

Verify network cable connection

Test point
Network cable to wall and system
Expected
Cables fully seated; LED indicators lit

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to archive to DVD

System cannot write patient cases to DVD media

Ranked causes

  1. 1. Improper media being used (common)
  2. 2. Cases not selected (common)
  3. 3. Software error in finding DVD drive (occasional)
  4. 4. Defective DVD drive (occasional)
First check

Verify media type

Test point
DVD media package or disc
Expected
DVD-R media present and readable

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Control Panel keys do not work

Keyboard buttons on control panel are unresponsive

Ranked causes

  1. 1. Cabling not fully connected (common)
  2. 2. Damage to or binding of keys of control panel (occasional)
  3. 3. Damaged or defective Control Panel (occasional)
First check

Verify control panel cable connection

Test point
Control panel cable connectors
Expected
Cable fully seated and secure

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

DICOM Verification Success with Warning - Abstract Syntax Rejected

Warning message 'DICOM Verification Success with Warning.' Abstract Syntax rejected for Worklist or Storage server configuration.

Ranked causes

  1. 1. Configured Worklist SCP is not a Worklist server (common)
  2. 2. Configured Storage SCP is not Storage server or does not accept Ultrasound Multi-frame Image SOP class (common)
  3. 3. Configured Storage SCP does not accept JPEG Lossy Transfer syntax (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Improper FFR measurements patient safety

FFR pressure values are inaccurate or out of expected range

Ranked causes

  1. 1. Faulty pressure wire (common)
  2. 2. Faulty PIM-ffr (occasional)
  3. 3. Faulty VH board (occasional)
First check

Test with alternate pressure wire and known reference

Test point
PIM-ffr pressure input
Expected
FFR values match reference standard within tolerance

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No FFR input patient safety

FFR (Fractional Flow Reserve) pressure measurement not received

Ranked causes

  1. 1. Faulty pressure wire (common)
  2. 2. Faulty PIM-ffr (occasional)
  3. 3. Faulty VH board (occasional)
First check

Test with another pressure wire

Test point
PIM-ffr connector
Expected
FFR input signal appears with alternate wire

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No Scheduled Procedures returned from Worklist Search

Worklist search returns no procedures. No items listed when querying HIS/CIS or local system Worklist server.

Ranked causes

  1. 1. Volcano s5 System clock not set to today's date (common)
  2. 2. 'This System' selected in Worklist search and HIS/CIS not configured with Volcano s5 Scheduled Station AE Title (common)
  3. 3. HIS/CIS does not schedule procedures for IVUS Modality (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

PIMr Pullback improper

PIMr motor pullback function malfunctions; movement improper or obstructed

Ranked causes

  1. 1. Obstruction to movement or faulty PIMr (common)
  2. 2. Defective imaging catheter (occasional)
  3. 3. Faulty PIMr (occasional)
First check

Inspect PIMr cable for obstructions

Test point
PIMr cable routing
Expected
Cable free of kinks or obstructions

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System does not power up high voltage

System fails to turn on; no indication of power

Ranked causes

  1. 1. AC power interruption to the system (common)
  2. 2. Internal power cord disconnected from AC inlet ATX Power Supply (occasional)
  3. 3. CPU system DC power failure (occasional)
First check

Verify AC power at mains

Test point
Power outlet
Expected
AC power present

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System only boots to Volcano logo screen; Mouse cursor is active, but no user interface appears

System displays Volcano logo but fails to load user interface

Ranked causes

  1. 1. User interface in the background (common)
  2. 2. System Software is corrupt (occasional)
  3. 3. System hard drive is corrupt or defective (occasional)
First check

Press Escape key

Test point
Keyboard
Expected
User interface appears on screen

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Trackball does not work properly

Trackball input is unresponsive or moves erratically

Ranked causes

  1. 1. Trackball is dirty or contaminated (common)
  2. 2. Cabling not fully connected (occasional)
  3. 3. Damaged or defective Control Panel or Trackball (occasional)
First check

Remove and clean trackball

Test point
Trackball housing and ball
Expected
Trackball clean and free of debris

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Video display appears to be lacking a fundamental color - or is only Red, Green, or Blue

Display shows only one color channel or color is missing

Ranked causes

  1. 1. DVI connector not fully seated and screwed in (common)
  2. 2. Video cable may have broken wire (occasional)
  3. 3. Defective Monitor (occasional)
First check

Inspect DVI connector

Test point
DVI connector on monitor and system
Expected
Connector fully seated with screws tight

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

DICOM Verification Failed - Association Negotiation Failed

Error message 'DICOM Verification Failed - Association Negotiation Failed. Unacceptable Called/Calling AP Title.' Application Entity Title mismatch between Volcano s5 System and remote DICOM server.

Ranked causes

  1. 1. Incorrect Application Entity Title for DICOM Storage or Worklist SCP (common)
  2. 2. Incorrect Volcano s5 System DICOM Storage or Worklist SCU AE Title (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

PIM-ffr auto-zero signal fluctuation error

System reports error during auto-zero process; input signal fluctuates excessively, preventing proper offset DAC adjustment

Ranked causes

  1. 1. Guide wire positioned in body sensing pulsing pressure (common)
  2. 2. Smart Wire faulty or defective (occasional)
First check

Verify guide wire position

Test point
Guide wire tip location
Expected
Wire outside patient body

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Aortic High Level input for FFR feature patient safety

Calibrate aortic signal input to rear panel so Pa (pressure–aortic) display matches Cath Lab physiological equipment. Required once per installation or when equipment changes. No field calibration for PIM-ffr.

Ranked causes

  1. 1. Initial installation or equipment change (common)
First check

Enter FFR Service Mode using service password

Test point
Settings Tab > Service
Expected
Service Mode Indicator displays

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Aortic High Level input signal patient safety

Configure the Aortic Signal input on rear panel so Pa (pressure – aortic) value on live display matches physiological monitor in Cath lab

Ranked causes

  1. 1. Initial setup or equipment change requiring aortic signal calibration (common)
First check

Input signal at zero level

Test point
Aortic Input rear panel
Expected
Zero level

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Input calibration restore to factory default

Restore Aortic High Level input calibration to factory-determined setting when problems are encountered or calibration has failed

Ranked causes

  1. 1. Failed calibration or calibration problems (unknown)
First check

Verify calibration failure or problems exist before restoring

Test point
Input Calibration display
Expected
Failed Calibration Message indication visible

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: PIM-ffr Input Pressure Accuracy

Digital Pd (distal pressure) value does not match expected reading from Pressure Guide Wire Simulator fixture during calibration check

Ranked causes

  1. 1. PIM-ffr miscalibrated or faulty (unknown)
First check

Verify zero pressure reading

Test point
PIM-ffr input with Pressure Guide Wire Simulator fixture
Expected
Digital Pd value indicates zero

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display Patient Cable Status

Tests PIM-ffr module and its cable connection; results also indicated by PIM icon in non-Service Mode screens.

Ranked causes

  1. 1. PIM-ffr module or cable failure (rare)
First check

Run Display Patient Cable Status diagnostic

Test point
Diagnostic Test Results field
Expected
PIM-ffr and cable status result; also monitor PIM icon in non-Service Mode screens

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Graphical user interface image is not aligned on screen

GUI appears slightly off-screen to one edge of display

Ranked causes

  1. 1. LCD Display did not switch modes correctly (common)
First check

Reboot display or execute re-scan

Test point
Display menu or system reboot
Expected
GUI properly aligned on screen edges

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image colors not displayed correctly on DICOM Viewer

DICOM images display with incorrect colors on remote workstation viewer. Color palette images not rendered correctly.

Ranked causes

  1. 1. Workstation does not handle Color Palette Images correctly (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incorrect measurement results from PIM cable caution

System produces wrong measurements when PIM cable length setting does not match installed cable

Ranked causes

  1. 1. PIM cable length setting does not match installed cable (common)
First check

Verify PIM cable length setting matches installed cable

Test point
Field Service Dialog - PIM Cable Length tab
Expected
Selected length equals installed PIM cable length

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incorrect PIM cable length configured caution

PIM cable length setting does not match the physical cable installed, causing incorrect measurement results

Ranked causes

  1. 1. Wrong PIM cable length selected in Field Service Dialog (common)
First check

Verify PIM cable length matches Field Service Dialog setting

Test point
PIM Cable Length tab
Expected
Selected cable length matches physically installed PIM cable

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

IVUS display is noisy / RF artifacts in image caution

IVUS image contains noise or RF artifacts

Ranked causes

  1. 1. Pick-up from extraneous equipment radiating EMI at transducer frequencies (common)
First check

Identify offending EMI source

Test point
Equipment near imaging system
Expected
Source identified or absent

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

IVUS image is vertically or horizontally ellipsoid shaped

IVUS image appears stretched or compressed in one axis

Ranked causes

  1. 1. Display is not set to correct aspect ratio display mode (common)
First check

Verify display resolution setting

Test point
Display or system video settings
Expected
Resolution set to 1280x1024

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

IVUS Images cannot be displayed on DICOM Viewer

IVUS intravascular ultrasound images do not display on remote DICOM workstation. Viewer does not support IVUS modality.

Ranked causes

  1. 1. DICOM Workstation/Viewer does not support IVUS Image Modality (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Network Transfer Failed - C-STORE Error Out of Resources

Error message 'Network Transfer Failed. C-STORE Error: Out of Resources. Status code: a700.' DICOM server resource exhausted or storage full.

Ranked causes

  1. 1. DICOM SCP server resource problem or system hard drive full (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No power output from ATX supply high voltage

All power supply LEDs out on backplane; system does not power up

Ranked causes

  1. 1. ATX power supply internal failure (occasional)
First check

Verify AC mains present at ATX supply input

Test point
AC input socket on ATX supply
Expected
AC voltage present

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

One or more power supply outputs shut down high voltage

Some backplane power LEDs out; specific voltage rail unavailable (+3.3V, +5V, +12V, -5V, -12V, or PGD)

Ranked causes

  1. 1. Electrical overload on affected power rail (common)
First check

Check backplane LED status

Test point
Backplane power supply LED indicators (+3.3, -12, PGD, -5, +5 SB, +5, +12)
Expected
LEDs illuminated per Figure 13 for each powered rail

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Revolution image noisy

Revolution ultrasound imaging produces noisy output

Ranked causes

  1. 1. Air in imaging catheter (common)
First check

Re-flush imaging catheter

Test point
Catheter lumen
Expected
Air bubbles eliminated; image noise reduced

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

s5i FFR hardware test

Test of FFR-related hardware modules; displays hex values for engineering use and pass/fail status.

Ranked causes

  1. 1. Hardware module failure detected (rare)
First check

Run s5i FFR hardware test from diagnostics menu

Test point
Diagnostic Test Results field
Expected
Either 's5 hardware test passed' or failure message with hex values

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scheduled Procedures discarded from Worklist Query

Error dialog displayed indicating scheduled procedures were discarded from Worklist Query. Multiple validation error codes listed.

Ranked causes

  1. 1. DICOM Worklist Server returns invalid attributes (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Search button is disabled in Worklist dialog

Search button cannot be clicked in Worklist search dialog. Button appears inactive or greyed out.

Ranked causes

  1. 1. Volcano s5 System does not allow unconstrained Worklist searches (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

System does not boot up

Unit fails to start; no audible computer beep heard during boot sequence.

Ranked causes

  1. 1. Power cycle required or boot process delayed (common)
First check

Listen for audible beep during boot sequence

Test point
System speaker
Expected
Audible computer beep

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Validation errors in Scheduled Procedures from Worklist Query

Warning dialog displayed indicating Scheduled Procedures from Worklist Query contain validation errors. Codes DISTA_MWL_000 through DISTA_MWL_006 listed.

Ranked causes

  1. 1. DICOM Worklist Server returns invalid attributes in Worklist Query results (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

VHFM Pipeline test

Checks FFR operations from VH board data input through FFR software process; indicates software, memory, or PC issues if failed.

Ranked causes

  1. 1. Software, memory, or PC problem (occasional)
First check

Start VHFM Pipeline test and select Live mode

Test point
Live screen display
Expected
Waveforms scrolling across screen, similar to demo data but more extensive

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Worklist button is disabled in Patient dialog

Worklist button cannot be selected or used in the Patient dialog. Button appears greyed out or inactive.

Ranked causes

  1. 1. No Worklist Servers have been configured (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
104Pressure cable bad codeError detected in pressure cable identification or dataerror
105Pressure cable error 2 - EEPROM errorPressure cable EEPROM has error or corruptionerror
107Pressure wire short conditionShort circuit detected in pressure wireerror
108Cable signal unstableSmartWire in patient body, intermittent connection, or fluctuating signal during auto zerowarning
109Cable error - calibration EEPROM errorCalibration EEPROM error with patient cableerror
110Cable error - +5 volt error+5 volt power supply error in patient cableerror
111Cable error - Vcc errorVcc power supply error in patient cableerror
112Cable error - no calibration performedCable calibration has not been performederror
113Cable error - patient cable ROM errorRead-only memory error in patient cableerror
114Cable error - patient cable RAM errorRandom access memory error in patient cableerror
115Cable error - offset errorCable status indicates offset calibration errorerror
116Cable error - gain errorCable status indicates gain calibration errorerror
202No factory calibrationCalibration EEPROM is corrupterror
204Low level aortic input errorInactive low level aortic input indicates error conditionwarning
205Calibration errorCalibration EEPROM error detectederror
206Isolated circuit power errorPower supply error for isolated circuits on rear panelerror
207High voltage limitHigh voltage limit condition detectedcritical
212EEPROM write errorError occurs while writing to EEPROM on S5 FFR System circuit boarderror
213DSP memory errorDSP memory error occurred during operationerror
215HVMON errorDSP detected an HVMON (high voltage monitor) errorerror
216Safety shutdown errorTransmit voltage safety shutdown activated; pulse repetition frequency, burst width, or power level exceeded limitscritical
Archivist failed to start; cannot save or recall dataArchivist failed to start; cannot save or recall dataFile storage subsystem failed to initialize properlycritical
Archivist internal error, data may be suspectArchivist internal error, data may be suspectGeneral error from file storage subsystemerror
Bad CommandCalibration cannot be performed because s5i Demo Mode is enabledDemo Mode ON during calibration attempterror
CardCoordinator: E10nnCard coordinator application errorCard coordinator module failureerror
CardMgr: E100nCard manager system errorCard manager module failure during operationcritical
CaseMgr: E1nnnCase manager error (n=any number)Case manager system error during operationcritical
Catheter FailureCatheter drawing excessive current from patient interface moduleCatheter plugged in draws abnormally high currenterror
Caution: hard disk filling, please archive files soonCaution: hard disk filling, please archive files soonLess than 800 MB free space (about 90 minutes of data) remainswarning
Caution: recording disallowed until disk space availableCaution: recording disallowed until disk space availableLess than 100 MB free space remains on hard diskcritical
Current case still open. End before switching to FFR Mode.System cannot switch to FFR mode while IVUS case is openUser attempts to switch to FFR mode with active casewarning
Current patient data may be corruptCurrent patient data may be corruptPatient identification data inconsistency detectedwarning
Data missing from recordingData missing from recordingGaps detected in sequence numbers of information packetserror
Directory file load read errorDirectory file load read errorError detected while loading directory fileerror
Directory file not valid, rebuildingDirectory file not valid, rebuildingDirectory file found inconsistent with studies on hard diskwarning
Directory file update write errorDirectory file update write errorError detected while writing directory fileerror
DISTA_MWL_000Unsupported Character SetDICOM Worklist Server returns query results with unsupported character set encoding.warning
DISTA_MWL_001Missing or empty Patient IDDICOM Worklist Server returns query results without Patient ID attribute.warning
DISTA_MWL_002Missing or empty Study Instance UIDDICOM Worklist Server returns query results without Study Instance UID.warning
DISTA_MWL_003Invalid Study Instance UIDDICOM Worklist Server returns query results with malformed Study Instance UID.warning
DISTA_MWL_004Missing or empty type 1 attributeDICOM Worklist Server returns query results missing required type 1 attribute.warning
DISTA_MWL_005Type 2 attribute missingDICOM Worklist Server returns query results missing optional type 2 attribute.info
DISTA_MWL_006MergeCOM-3 Validation failedDICOM Worklist Server returns query results that fail MergeCOM-3 validation.warning
Drive not bootable. Insert bootable disk.Hard drive cannot boot the operating systemFatal error on hard drive storage or boot sector corruptioncritical
DVD drive failureDVD drive failureDVD drive has malfunctionederror
DVD fullDVD fullCurrently inserted DVD has no remaining capacitywarning
DVD verify failedDVD verify failedPatient study file copied to DVD was not correctly writtenerror
E1Cannot communicate with RFID on catheterRFID communication failure detectederror
E2TGC memory failureTime Gain Compensation memory fault detectederror
E3Analog high voltage failureAnalog high voltage subsystem fault detectedcritical
E4One of the Pulse Width control pins is stuck high/lowPulse Width control pin detected stuck in high or low stateerror
E5Catheter motor driver failureCatheter motor driver malfunction detectederror
E6Pullback motor driver failedPullback motor driver malfunction detectederror
End of file encounteredEnd of file encounteredFinal data in recorded file has been played backinfo
Error during initialization.Unrecoverable error during system initialization; Health Checker encountered error on startupSystem startup; health checker operation failedcritical
ExecutionMgr: E1000-E1032Execution manager errorsExecution manager system failurecritical
ExecutionMgr: E1033Network configuration errorSystem startup with incorrect network setuperror
ExecutionMgr: E1034System initialization failureSystem startup or boot process failureerror
ExecutionMgr: E1035-1038Execution manager subsystem errorsSpecific execution module failure during operationcritical
ExecutionMgr: E1039-E1055Execution manager informational or non-critical errorsExecution manager non-critical conditioninfo
ExecutionMgr: E1056-1084Execution manager critical errorsCritical execution module failurecritical
FrameMgr: E1001-1057Frame manager errorsFrame processing or management failurecritical
Hard disk fullHard disk fullComputer hard disk has no remaining free spacewarning
Hardware did not respondSystem did not receive response to calibration command within allotted timeHardware communication timeout during calibrationerror
Hardware did not respondCommand to hardware not executed in allotted timeDuring calibration procedureerror
Hardware did not respond to commandHardware did not respond to commandHardware did not respond when command was sentwarning
Health Checker module error at startupError occurred trying to execute Health Checker module during initializationSystem startup; health checker determines hardware statecritical
ImageMgr: E1001-1211Image manager errorsImage processing or rendering failurecritical
Incorrect system time or dateTime or date set incorrectly on systemUser sets incorrect time/date during initialization or settingsinfo
Isolation transformer turned offExternal isolation transformer (s5i model) is not poweredUser error or power issue with isolation transformerwarning
Lights on but no displaySystem power available (console lights/tower lights, fans running) but monitor shows nothingMonitor inactive or disconnected; backup battery failureerror
Maximum session record time exceededMaximum session record time exceeded90 minutes of data recorded for patient; session limit reachedwarning
MeasurementMgr: E1001-1003Measurement manager errorsMeasurement processing failurecritical
Meridian could not restart VH+ and IVUS boards to Meridian MFailure to restart VH+ or IVUS board when switching back from FFR to Meridian modeSystem attempted to resume Meridian mode from FFR modeerror
Meridian could not stop VH+ and IVUS boards for FFR Mode swiFailure to stop VH+ or IVUS board when attempting FFR mode switchFFR mode switch requested while system actively running VHerror
Minit: E1001-1002Minit module errorsMinit subsystem failurecritical
MMAPI: E100-104MMAPI critical errors preventing IVUS imagingMMAPI subsystem critical failure during imaging initializationcritical
MMAPI: E108-121MMAPI errors affecting IVUS imaging capabilityMMAPI subsystem failure during imaging operationerror
MMAPI: E122-123MMAPI critical module errorsMMAPI critical subsystem failurecritical
MMAPI: E124-125MMAPI recoverable errorsMMAPI subsystem error with recovery optionerror
MMAPI: E126-131MMAPI automatic shutdown errorsMMAPI detects critical condition requiring system shutdowncritical
MMAPI: E132-133MMAPI shutdown errorsMMAPI critical error causing system shutdownerror
MMAPI: E134-136MMAPI errors preventing IVUS imagingMMAPI failure during imaging operationserror
MMAPI: E137MMAPI critical shutdown errorMMAPI critical condition triggering system shutdowncritical
MMAPI: E138-140MMAPI errors preventing IVUS imaging operationsMMAPI failure during imaging modeerror
MMAPI: E141-904MMAPI high-range error codesMMAPI error in extended error rangeerror
No CatheterPatient interface module does not have a catheter connected or recognition delayedCatheter not plugged in; PIM needs time to recognize new catheter; dialog boxes delay detectionwarning
No data in patient fileNo data in patient filePatient file contains only demographics without physiological datainfo
No Disk in Floppy DriveSystem cannot find or read floppy disk driveBackup battery failing; floppy drive not functionalwarning
No Paper or InkPrinter is out of paper or ink; or paper issues (incorrect brand, inserted incorrectly)Printer detects missing paper/ink; incorrect paper brand; wrong paper orientationwarning
No power, no lights, no fansSystem completely without electrical power or displayMains power failure, disconnected cable, or isolation transformer offcritical
Out of RangeInput signal level outside acceptable toleranceSignal too low or too high during calibrationerror
Out of RangeInput signal level deviates from selected calibration level by more than ±10%Signal too low or too high during calibrationerror
PIM Configuration in progress. Please wait.Hardware configuration in progress after PIM changeNew PIM installed or detectedinfo
PIM not DetectedSystem unable to communicate with patient interface module unitPIM connection lost or communication failureerror
PIM Power FailurePatient interface module detected undesirable high power condition for catheter in useHigh power condition detected on PIM output to cathetererror
Please insert a DVDPlease insert a DVDAttempt made to write to DVD when none insertedwarning
Power plug disconnectedSystem power cable not properly connected to wall or towerLoose or unplugged mains power connectorerror
Read error on playbackRead error on playbackA portion of file could not be read correctly during playback of recorded dataerror
Read error on recalling patient dataRead error on recalling patient dataError encountered while recalling patient identification dataerror
Shutting down will abort case archiving.Pressing power button during case archive will cancel archive processPower button pressed while system is archiving casecritical
StateMachine: E100nState machine errors affecting case operationsState machine failure during case operationerror
System error detected. Auto-repair in progress, please wait.Hardware failure detected; system attempting automatic self-repairVH or IVUS card failure detected during operationerror
System shutdown before ending case.System was powered down prematurely before case was ended; recover mode initiated on restartPower removed while case open; system restart after unexpected shutdownwarning
There are no patient files on the DVDThere are no patient files on the DVDAttempt made to recall data from DVD without S5 study fileswarning
Time does not maintain correct settingSystem does not retain correct time/date after rebootSystem powers off or reboots after time/date correctionerror
Unstable SignalInput signal fluctuates more than ±2.5 mmHg from the selected calibration levelSignal instability exceeds tolerance during calibrationerror
Write error on saving patient dataWrite error on saving patient dataError encountered while saving patient identification dataerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.