Volcano s5: troubleshooting and error codes
The ServiceForge knowledge base holds 47 documented symptoms for the Volcano s5, with 113 ranked causes, 89 diagnostic checks with test points and expected values, and 108 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Archive to Network fails - DICOM/Network Configuration Check server fails
Error message 'DICOM Verification Failed - Failed to connect to remote host.' or C-STORE errors. Cannot connect to or transfer data to remote DICOM server.
Ranked causes
- 1. Invalid port number assignment for remote DICOM Server (common)
- 2. Remote DICOM Application not running or not on-line (common)
- 3. Remote DICOM Application not configured to accept socket connection from Volcano s5 System IP Address (occasional)
- 4. DICOM SCP server experiencing resource problem or hard drive full (occasional)
- 5. Interoperability problem with vendor device (rare)
- 6. Defective Volcano s5 System NIC (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM/Network Configuration - Find host fails
Error message 'PING Failed. The host name could not be found.' Unable to locate remote DICOM server by hostname or IP address.
Ranked causes
- 1. Network cable not connected to Volcano s5 System (common)
- 2. Incorrect network port enabled (common)
- 3. Defective network cable (occasional)
- 4. Invalid Volcano s5 System network configuration (occasional)
- 5. Invalid remote DICOM server IP address configuration (occasional)
- 6. Invalid remote DICOM server Gateway IP address or Subnet Mask configuration (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No display on Internal LCD
Internal LCD monitor shows no image
Ranked causes
- 1. AC power interruption to the system (common)
- 2. DVI video signal interruption (occasional)
- 3. Video board defective (occasional)
- 4. Display is damaged (rare)
- 5. Single Board Computer (SBC) failure (occasional)
Check power LED status on monitor
- Test point
- Monitor power indicator
- Expected
- LED illuminated
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Heart-beating icon with ECG input / FFR option - No ECG waveform / No VH data patient safety
ECG input not received; FFR and VH functions non-functional; no heartbeat icon displayed
Ranked causes
- 1. Improper ECG signal being used (common)
- 2. Hemodynamic system not configured for ECG output (occasional)
- 3. ECG cable defective (occasional)
- 4. Cabling not fully connected (occasional)
- 5. VH imaging card defective (occasional)
Test ECG signal with simulator
- Test point
- ECG input connector
- Expected
- ECG waveform appears; heartbeat icon displayed
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIM not Found Error Message
System displays 'PIM not Found' error; PIM module not detected
Ranked causes
- 1. Cabling not fully connected (common)
- 2. PIM connector not fully connected (common)
- 3. Bent connector pins (occasional)
- 4. PIM cord damaged (occasional)
- 5. PIM damaged (occasional)
Verify PIM cable connection
- Test point
- PIM cabling connectors
- Expected
- All connectors fully seated and tight
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer does not work
Printer fails to produce output
Ranked causes
- 1. Cabling or extender not fully connected (common)
- 2. Printer is off or out of supplies (common)
- 3. Printer not set as default (occasional)
- 4. Print Manager not running (occasional)
- 5. Printer is defective (occasional)
Verify USB and power cable connections
- Test point
- Printer USB and power connectors
- Expected
- All cables fully seated and secure
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Catheter Type not recognized
System does not identify catheter type; type not displayed
Ranked causes
- 1. Install error (common)
- 2. Contamination or corrosion on pins or sockets in PIM catheter connections (occasional)
- 3. Mechanical damage to PIM (occasional)
- 4. Damaged catheter (occasional)
Disconnect and reinstall catheter
- Test point
- PIM catheter connector
- Expected
- Catheter type recognized after reinsertion
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No External VGA Video is displayed
External VGA monitor connected to I/O panel shows no image
Ranked causes
- 1. AC power interruption to the system (common)
- 2. Video signal interruption (occasional)
- 3. Video board defective (occasional)
- 4. Single Board Computer (SBC) failure (occasional)
Check LED status on external monitor
- Test point
- Monitor power indicator
- Expected
- LED illuminated
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No IVUS display
IVUS imaging does not appear on screen
Ranked causes
- 1. Defective Catheter (common)
- 2. Defective PIM (occasional)
- 3. Defective hardware (occasional)
- 4. Interconnection failures (occasional)
Ensure PIM is connected
- Test point
- PIM connector
- Expected
- Connector fully seated
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Operating System fails to boot
System powers on but OS does not load; boot fails before OS presentation
Ranked causes
- 1. Possible DC power interruption (occasional)
- 2. Mechanical error, parts not seated/mounted correctly (occasional)
- 3. Hard drive data corruption (occasional)
- 4. Defective Single Board Computer (SBC) (occasional)
Measure system DC power
- Test point
- CPU and hard drive power connectors
- Expected
- Correct voltage present (per PSU spec)
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to archive to DICOM
System cannot send patient cases to DICOM network destination
Ranked causes
- 1. Cabling not fully connected (common)
- 2. Network set-up error (occasional)
- 3. Incorrect IP address (occasional)
- 4. Incorrect port - see FSB-006 (occasional)
Verify network cable connection
- Test point
- Network cable to wall and system
- Expected
- Cables fully seated; LED indicators lit
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to archive to DVD
System cannot write patient cases to DVD media
Ranked causes
- 1. Improper media being used (common)
- 2. Cases not selected (common)
- 3. Software error in finding DVD drive (occasional)
- 4. Defective DVD drive (occasional)
Verify media type
- Test point
- DVD media package or disc
- Expected
- DVD-R media present and readable
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control Panel keys do not work
Keyboard buttons on control panel are unresponsive
Ranked causes
- 1. Cabling not fully connected (common)
- 2. Damage to or binding of keys of control panel (occasional)
- 3. Damaged or defective Control Panel (occasional)
Verify control panel cable connection
- Test point
- Control panel cable connectors
- Expected
- Cable fully seated and secure
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM Verification Success with Warning - Abstract Syntax Rejected
Warning message 'DICOM Verification Success with Warning.' Abstract Syntax rejected for Worklist or Storage server configuration.
Ranked causes
- 1. Configured Worklist SCP is not a Worklist server (common)
- 2. Configured Storage SCP is not Storage server or does not accept Ultrasound Multi-frame Image SOP class (common)
- 3. Configured Storage SCP does not accept JPEG Lossy Transfer syntax (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Improper FFR measurements patient safety
FFR pressure values are inaccurate or out of expected range
Ranked causes
- 1. Faulty pressure wire (common)
- 2. Faulty PIM-ffr (occasional)
- 3. Faulty VH board (occasional)
Test with alternate pressure wire and known reference
- Test point
- PIM-ffr pressure input
- Expected
- FFR values match reference standard within tolerance
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No FFR input patient safety
FFR (Fractional Flow Reserve) pressure measurement not received
Ranked causes
- 1. Faulty pressure wire (common)
- 2. Faulty PIM-ffr (occasional)
- 3. Faulty VH board (occasional)
Test with another pressure wire
- Test point
- PIM-ffr connector
- Expected
- FFR input signal appears with alternate wire
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Scheduled Procedures returned from Worklist Search
Worklist search returns no procedures. No items listed when querying HIS/CIS or local system Worklist server.
Ranked causes
- 1. Volcano s5 System clock not set to today's date (common)
- 2. 'This System' selected in Worklist search and HIS/CIS not configured with Volcano s5 Scheduled Station AE Title (common)
- 3. HIS/CIS does not schedule procedures for IVUS Modality (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIMr Pullback improper
PIMr motor pullback function malfunctions; movement improper or obstructed
Ranked causes
- 1. Obstruction to movement or faulty PIMr (common)
- 2. Defective imaging catheter (occasional)
- 3. Faulty PIMr (occasional)
Inspect PIMr cable for obstructions
- Test point
- PIMr cable routing
- Expected
- Cable free of kinks or obstructions
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not power up high voltage
System fails to turn on; no indication of power
Ranked causes
- 1. AC power interruption to the system (common)
- 2. Internal power cord disconnected from AC inlet ATX Power Supply (occasional)
- 3. CPU system DC power failure (occasional)
Verify AC power at mains
- Test point
- Power outlet
- Expected
- AC power present
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System only boots to Volcano logo screen; Mouse cursor is active, but no user interface appears
System displays Volcano logo but fails to load user interface
Ranked causes
- 1. User interface in the background (common)
- 2. System Software is corrupt (occasional)
- 3. System hard drive is corrupt or defective (occasional)
Press Escape key
- Test point
- Keyboard
- Expected
- User interface appears on screen
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trackball does not work properly
Trackball input is unresponsive or moves erratically
Ranked causes
- 1. Trackball is dirty or contaminated (common)
- 2. Cabling not fully connected (occasional)
- 3. Damaged or defective Control Panel or Trackball (occasional)
Remove and clean trackball
- Test point
- Trackball housing and ball
- Expected
- Trackball clean and free of debris
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Video display appears to be lacking a fundamental color - or is only Red, Green, or Blue
Display shows only one color channel or color is missing
Ranked causes
- 1. DVI connector not fully seated and screwed in (common)
- 2. Video cable may have broken wire (occasional)
- 3. Defective Monitor (occasional)
Inspect DVI connector
- Test point
- DVI connector on monitor and system
- Expected
- Connector fully seated with screws tight
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM Verification Failed - Association Negotiation Failed
Error message 'DICOM Verification Failed - Association Negotiation Failed. Unacceptable Called/Calling AP Title.' Application Entity Title mismatch between Volcano s5 System and remote DICOM server.
Ranked causes
- 1. Incorrect Application Entity Title for DICOM Storage or Worklist SCP (common)
- 2. Incorrect Volcano s5 System DICOM Storage or Worklist SCU AE Title (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIM-ffr auto-zero signal fluctuation error
System reports error during auto-zero process; input signal fluctuates excessively, preventing proper offset DAC adjustment
Ranked causes
- 1. Guide wire positioned in body sensing pulsing pressure (common)
- 2. Smart Wire faulty or defective (occasional)
Verify guide wire position
- Test point
- Guide wire tip location
- Expected
- Wire outside patient body
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Aortic High Level input for FFR feature patient safety
Calibrate aortic signal input to rear panel so Pa (pressure–aortic) display matches Cath Lab physiological equipment. Required once per installation or when equipment changes. No field calibration for PIM-ffr.
Ranked causes
- 1. Initial installation or equipment change (common)
Enter FFR Service Mode using service password
- Test point
- Settings Tab > Service
- Expected
- Service Mode Indicator displays
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Aortic High Level input signal patient safety
Configure the Aortic Signal input on rear panel so Pa (pressure – aortic) value on live display matches physiological monitor in Cath lab
Ranked causes
- 1. Initial setup or equipment change requiring aortic signal calibration (common)
Input signal at zero level
- Test point
- Aortic Input rear panel
- Expected
- Zero level
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Input calibration restore to factory default
Restore Aortic High Level input calibration to factory-determined setting when problems are encountered or calibration has failed
Ranked causes
- 1. Failed calibration or calibration problems (unknown)
Verify calibration failure or problems exist before restoring
- Test point
- Input Calibration display
- Expected
- Failed Calibration Message indication visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PIM-ffr Input Pressure Accuracy
Digital Pd (distal pressure) value does not match expected reading from Pressure Guide Wire Simulator fixture during calibration check
Ranked causes
- 1. PIM-ffr miscalibrated or faulty (unknown)
Verify zero pressure reading
- Test point
- PIM-ffr input with Pressure Guide Wire Simulator fixture
- Expected
- Digital Pd value indicates zero
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display Patient Cable Status
Tests PIM-ffr module and its cable connection; results also indicated by PIM icon in non-Service Mode screens.
Ranked causes
- 1. PIM-ffr module or cable failure (rare)
Run Display Patient Cable Status diagnostic
- Test point
- Diagnostic Test Results field
- Expected
- PIM-ffr and cable status result; also monitor PIM icon in non-Service Mode screens
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Graphical user interface image is not aligned on screen
GUI appears slightly off-screen to one edge of display
Ranked causes
- 1. LCD Display did not switch modes correctly (common)
Reboot display or execute re-scan
- Test point
- Display menu or system reboot
- Expected
- GUI properly aligned on screen edges
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image colors not displayed correctly on DICOM Viewer
DICOM images display with incorrect colors on remote workstation viewer. Color palette images not rendered correctly.
Ranked causes
- 1. Workstation does not handle Color Palette Images correctly (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect measurement results from PIM cable caution
System produces wrong measurements when PIM cable length setting does not match installed cable
Ranked causes
- 1. PIM cable length setting does not match installed cable (common)
Verify PIM cable length setting matches installed cable
- Test point
- Field Service Dialog - PIM Cable Length tab
- Expected
- Selected length equals installed PIM cable length
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect PIM cable length configured caution
PIM cable length setting does not match the physical cable installed, causing incorrect measurement results
Ranked causes
- 1. Wrong PIM cable length selected in Field Service Dialog (common)
Verify PIM cable length matches Field Service Dialog setting
- Test point
- PIM Cable Length tab
- Expected
- Selected cable length matches physically installed PIM cable
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
IVUS display is noisy / RF artifacts in image caution
IVUS image contains noise or RF artifacts
Ranked causes
- 1. Pick-up from extraneous equipment radiating EMI at transducer frequencies (common)
Identify offending EMI source
- Test point
- Equipment near imaging system
- Expected
- Source identified or absent
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IVUS image is vertically or horizontally ellipsoid shaped
IVUS image appears stretched or compressed in one axis
Ranked causes
- 1. Display is not set to correct aspect ratio display mode (common)
Verify display resolution setting
- Test point
- Display or system video settings
- Expected
- Resolution set to 1280x1024
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IVUS Images cannot be displayed on DICOM Viewer
IVUS intravascular ultrasound images do not display on remote DICOM workstation. Viewer does not support IVUS modality.
Ranked causes
- 1. DICOM Workstation/Viewer does not support IVUS Image Modality (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network Transfer Failed - C-STORE Error Out of Resources
Error message 'Network Transfer Failed. C-STORE Error: Out of Resources. Status code: a700.' DICOM server resource exhausted or storage full.
Ranked causes
- 1. DICOM SCP server resource problem or system hard drive full (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No power output from ATX supply high voltage
All power supply LEDs out on backplane; system does not power up
Ranked causes
- 1. ATX power supply internal failure (occasional)
Verify AC mains present at ATX supply input
- Test point
- AC input socket on ATX supply
- Expected
- AC voltage present
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
One or more power supply outputs shut down high voltage
Some backplane power LEDs out; specific voltage rail unavailable (+3.3V, +5V, +12V, -5V, -12V, or PGD)
Ranked causes
- 1. Electrical overload on affected power rail (common)
Check backplane LED status
- Test point
- Backplane power supply LED indicators (+3.3, -12, PGD, -5, +5 SB, +5, +12)
- Expected
- LEDs illuminated per Figure 13 for each powered rail
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Revolution image noisy
Revolution ultrasound imaging produces noisy output
Ranked causes
- 1. Air in imaging catheter (common)
Re-flush imaging catheter
- Test point
- Catheter lumen
- Expected
- Air bubbles eliminated; image noise reduced
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
s5i FFR hardware test
Test of FFR-related hardware modules; displays hex values for engineering use and pass/fail status.
Ranked causes
- 1. Hardware module failure detected (rare)
Run s5i FFR hardware test from diagnostics menu
- Test point
- Diagnostic Test Results field
- Expected
- Either 's5 hardware test passed' or failure message with hex values
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scheduled Procedures discarded from Worklist Query
Error dialog displayed indicating scheduled procedures were discarded from Worklist Query. Multiple validation error codes listed.
Ranked causes
- 1. DICOM Worklist Server returns invalid attributes (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Search button is disabled in Worklist dialog
Search button cannot be clicked in Worklist search dialog. Button appears inactive or greyed out.
Ranked causes
- 1. Volcano s5 System does not allow unconstrained Worklist searches (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not boot up
Unit fails to start; no audible computer beep heard during boot sequence.
Ranked causes
- 1. Power cycle required or boot process delayed (common)
Listen for audible beep during boot sequence
- Test point
- System speaker
- Expected
- Audible computer beep
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Validation errors in Scheduled Procedures from Worklist Query
Warning dialog displayed indicating Scheduled Procedures from Worklist Query contain validation errors. Codes DISTA_MWL_000 through DISTA_MWL_006 listed.
Ranked causes
- 1. DICOM Worklist Server returns invalid attributes in Worklist Query results (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
VHFM Pipeline test
Checks FFR operations from VH board data input through FFR software process; indicates software, memory, or PC issues if failed.
Ranked causes
- 1. Software, memory, or PC problem (occasional)
Start VHFM Pipeline test and select Live mode
- Test point
- Live screen display
- Expected
- Waveforms scrolling across screen, similar to demo data but more extensive
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Worklist button is disabled in Patient dialog
Worklist button cannot be selected or used in the Patient dialog. Button appears greyed out or inactive.
Ranked causes
- 1. No Worklist Servers have been configured (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
104 | Pressure cable bad code | Error detected in pressure cable identification or data | error |
105 | Pressure cable error 2 - EEPROM error | Pressure cable EEPROM has error or corruption | error |
107 | Pressure wire short condition | Short circuit detected in pressure wire | error |
108 | Cable signal unstable | SmartWire in patient body, intermittent connection, or fluctuating signal during auto zero | warning |
109 | Cable error - calibration EEPROM error | Calibration EEPROM error with patient cable | error |
110 | Cable error - +5 volt error | +5 volt power supply error in patient cable | error |
111 | Cable error - Vcc error | Vcc power supply error in patient cable | error |
112 | Cable error - no calibration performed | Cable calibration has not been performed | error |
113 | Cable error - patient cable ROM error | Read-only memory error in patient cable | error |
114 | Cable error - patient cable RAM error | Random access memory error in patient cable | error |
115 | Cable error - offset error | Cable status indicates offset calibration error | error |
116 | Cable error - gain error | Cable status indicates gain calibration error | error |
202 | No factory calibration | Calibration EEPROM is corrupt | error |
204 | Low level aortic input error | Inactive low level aortic input indicates error condition | warning |
205 | Calibration error | Calibration EEPROM error detected | error |
206 | Isolated circuit power error | Power supply error for isolated circuits on rear panel | error |
207 | High voltage limit | High voltage limit condition detected | critical |
212 | EEPROM write error | Error occurs while writing to EEPROM on S5 FFR System circuit board | error |
213 | DSP memory error | DSP memory error occurred during operation | error |
215 | HVMON error | DSP detected an HVMON (high voltage monitor) error | error |
216 | Safety shutdown error | Transmit voltage safety shutdown activated; pulse repetition frequency, burst width, or power level exceeded limits | critical |
Archivist failed to start; cannot save or recall data | Archivist failed to start; cannot save or recall data | File storage subsystem failed to initialize properly | critical |
Archivist internal error, data may be suspect | Archivist internal error, data may be suspect | General error from file storage subsystem | error |
Bad Command | Calibration cannot be performed because s5i Demo Mode is enabled | Demo Mode ON during calibration attempt | error |
CardCoordinator: E10nn | Card coordinator application error | Card coordinator module failure | error |
CardMgr: E100n | Card manager system error | Card manager module failure during operation | critical |
CaseMgr: E1nnn | Case manager error (n=any number) | Case manager system error during operation | critical |
Catheter Failure | Catheter drawing excessive current from patient interface module | Catheter plugged in draws abnormally high current | error |
Caution: hard disk filling, please archive files soon | Caution: hard disk filling, please archive files soon | Less than 800 MB free space (about 90 minutes of data) remains | warning |
Caution: recording disallowed until disk space available | Caution: recording disallowed until disk space available | Less than 100 MB free space remains on hard disk | critical |
Current case still open. End before switching to FFR Mode. | System cannot switch to FFR mode while IVUS case is open | User attempts to switch to FFR mode with active case | warning |
Current patient data may be corrupt | Current patient data may be corrupt | Patient identification data inconsistency detected | warning |
Data missing from recording | Data missing from recording | Gaps detected in sequence numbers of information packets | error |
Directory file load read error | Directory file load read error | Error detected while loading directory file | error |
Directory file not valid, rebuilding | Directory file not valid, rebuilding | Directory file found inconsistent with studies on hard disk | warning |
Directory file update write error | Directory file update write error | Error detected while writing directory file | error |
DISTA_MWL_000 | Unsupported Character Set | DICOM Worklist Server returns query results with unsupported character set encoding. | warning |
DISTA_MWL_001 | Missing or empty Patient ID | DICOM Worklist Server returns query results without Patient ID attribute. | warning |
DISTA_MWL_002 | Missing or empty Study Instance UID | DICOM Worklist Server returns query results without Study Instance UID. | warning |
DISTA_MWL_003 | Invalid Study Instance UID | DICOM Worklist Server returns query results with malformed Study Instance UID. | warning |
DISTA_MWL_004 | Missing or empty type 1 attribute | DICOM Worklist Server returns query results missing required type 1 attribute. | warning |
DISTA_MWL_005 | Type 2 attribute missing | DICOM Worklist Server returns query results missing optional type 2 attribute. | info |
DISTA_MWL_006 | MergeCOM-3 Validation failed | DICOM Worklist Server returns query results that fail MergeCOM-3 validation. | warning |
Drive not bootable. Insert bootable disk. | Hard drive cannot boot the operating system | Fatal error on hard drive storage or boot sector corruption | critical |
DVD drive failure | DVD drive failure | DVD drive has malfunctioned | error |
DVD full | DVD full | Currently inserted DVD has no remaining capacity | warning |
DVD verify failed | DVD verify failed | Patient study file copied to DVD was not correctly written | error |
E1 | Cannot communicate with RFID on catheter | RFID communication failure detected | error |
E2 | TGC memory failure | Time Gain Compensation memory fault detected | error |
E3 | Analog high voltage failure | Analog high voltage subsystem fault detected | critical |
E4 | One of the Pulse Width control pins is stuck high/low | Pulse Width control pin detected stuck in high or low state | error |
E5 | Catheter motor driver failure | Catheter motor driver malfunction detected | error |
E6 | Pullback motor driver failed | Pullback motor driver malfunction detected | error |
End of file encountered | End of file encountered | Final data in recorded file has been played back | info |
Error during initialization. | Unrecoverable error during system initialization; Health Checker encountered error on startup | System startup; health checker operation failed | critical |
ExecutionMgr: E1000-E1032 | Execution manager errors | Execution manager system failure | critical |
ExecutionMgr: E1033 | Network configuration error | System startup with incorrect network setup | error |
ExecutionMgr: E1034 | System initialization failure | System startup or boot process failure | error |
ExecutionMgr: E1035-1038 | Execution manager subsystem errors | Specific execution module failure during operation | critical |
ExecutionMgr: E1039-E1055 | Execution manager informational or non-critical errors | Execution manager non-critical condition | info |
ExecutionMgr: E1056-1084 | Execution manager critical errors | Critical execution module failure | critical |
FrameMgr: E1001-1057 | Frame manager errors | Frame processing or management failure | critical |
Hard disk full | Hard disk full | Computer hard disk has no remaining free space | warning |
Hardware did not respond | System did not receive response to calibration command within allotted time | Hardware communication timeout during calibration | error |
Hardware did not respond | Command to hardware not executed in allotted time | During calibration procedure | error |
Hardware did not respond to command | Hardware did not respond to command | Hardware did not respond when command was sent | warning |
Health Checker module error at startup | Error occurred trying to execute Health Checker module during initialization | System startup; health checker determines hardware state | critical |
ImageMgr: E1001-1211 | Image manager errors | Image processing or rendering failure | critical |
Incorrect system time or date | Time or date set incorrectly on system | User sets incorrect time/date during initialization or settings | info |
Isolation transformer turned off | External isolation transformer (s5i model) is not powered | User error or power issue with isolation transformer | warning |
Lights on but no display | System power available (console lights/tower lights, fans running) but monitor shows nothing | Monitor inactive or disconnected; backup battery failure | error |
Maximum session record time exceeded | Maximum session record time exceeded | 90 minutes of data recorded for patient; session limit reached | warning |
MeasurementMgr: E1001-1003 | Measurement manager errors | Measurement processing failure | critical |
Meridian could not restart VH+ and IVUS boards to Meridian M | Failure to restart VH+ or IVUS board when switching back from FFR to Meridian mode | System attempted to resume Meridian mode from FFR mode | error |
Meridian could not stop VH+ and IVUS boards for FFR Mode swi | Failure to stop VH+ or IVUS board when attempting FFR mode switch | FFR mode switch requested while system actively running VH | error |
Minit: E1001-1002 | Minit module errors | Minit subsystem failure | critical |
MMAPI: E100-104 | MMAPI critical errors preventing IVUS imaging | MMAPI subsystem critical failure during imaging initialization | critical |
MMAPI: E108-121 | MMAPI errors affecting IVUS imaging capability | MMAPI subsystem failure during imaging operation | error |
MMAPI: E122-123 | MMAPI critical module errors | MMAPI critical subsystem failure | critical |
MMAPI: E124-125 | MMAPI recoverable errors | MMAPI subsystem error with recovery option | error |
MMAPI: E126-131 | MMAPI automatic shutdown errors | MMAPI detects critical condition requiring system shutdown | critical |
MMAPI: E132-133 | MMAPI shutdown errors | MMAPI critical error causing system shutdown | error |
MMAPI: E134-136 | MMAPI errors preventing IVUS imaging | MMAPI failure during imaging operations | error |
MMAPI: E137 | MMAPI critical shutdown error | MMAPI critical condition triggering system shutdown | critical |
MMAPI: E138-140 | MMAPI errors preventing IVUS imaging operations | MMAPI failure during imaging mode | error |
MMAPI: E141-904 | MMAPI high-range error codes | MMAPI error in extended error range | error |
No Catheter | Patient interface module does not have a catheter connected or recognition delayed | Catheter not plugged in; PIM needs time to recognize new catheter; dialog boxes delay detection | warning |
No data in patient file | No data in patient file | Patient file contains only demographics without physiological data | info |
No Disk in Floppy Drive | System cannot find or read floppy disk drive | Backup battery failing; floppy drive not functional | warning |
No Paper or Ink | Printer is out of paper or ink; or paper issues (incorrect brand, inserted incorrectly) | Printer detects missing paper/ink; incorrect paper brand; wrong paper orientation | warning |
No power, no lights, no fans | System completely without electrical power or display | Mains power failure, disconnected cable, or isolation transformer off | critical |
Out of Range | Input signal level outside acceptable tolerance | Signal too low or too high during calibration | error |
Out of Range | Input signal level deviates from selected calibration level by more than ±10% | Signal too low or too high during calibration | error |
PIM Configuration in progress. Please wait. | Hardware configuration in progress after PIM change | New PIM installed or detected | info |
PIM not Detected | System unable to communicate with patient interface module unit | PIM connection lost or communication failure | error |
PIM Power Failure | Patient interface module detected undesirable high power condition for catheter in use | High power condition detected on PIM output to catheter | error |
Please insert a DVD | Please insert a DVD | Attempt made to write to DVD when none inserted | warning |
Power plug disconnected | System power cable not properly connected to wall or tower | Loose or unplugged mains power connector | error |
Read error on playback | Read error on playback | A portion of file could not be read correctly during playback of recorded data | error |
Read error on recalling patient data | Read error on recalling patient data | Error encountered while recalling patient identification data | error |
Shutting down will abort case archiving. | Pressing power button during case archive will cancel archive process | Power button pressed while system is archiving case | critical |
StateMachine: E100n | State machine errors affecting case operations | State machine failure during case operation | error |
System error detected. Auto-repair in progress, please wait. | Hardware failure detected; system attempting automatic self-repair | VH or IVUS card failure detected during operation | error |
System shutdown before ending case. | System was powered down prematurely before case was ended; recover mode initiated on restart | Power removed while case open; system restart after unexpected shutdown | warning |
There are no patient files on the DVD | There are no patient files on the DVD | Attempt made to recall data from DVD without S5 study files | warning |
Time does not maintain correct setting | System does not retain correct time/date after reboot | System powers off or reboots after time/date correction | error |
Unstable Signal | Input signal fluctuates more than ±2.5 mmHg from the selected calibration level | Signal instability exceeds tolerance during calibration | error |
Write error on saving patient data | Write error on saving patient data | Error encountered while saving patient identification data | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI