VersaMed iVent201: troubleshooting and error codes
The ServiceForge knowledge base holds 73 documented symptoms for the VersaMed iVent201, with 101 ranked causes, 128 diagnostic checks with test points and expected values, and 12 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Erratic exhale tidal volumes patient safety
Delivered breaths vary unpredictably in size despite volume control mode setting, some breaths smaller than normal with pink person breath type icon
Ranked causes
- 1. Pressure support extra breaths triggered during volume control (common)
- 2. Flow offset due to calibration error (occasional)
- 3. Leak at patient interface or test lung (common)
- 4. Flow sensitivity set too low (occasional)
Observe breath types and sizes during operation
- Test point
- Ventilator display breath type icon
- Expected
- Pink person icon indicates extra breaths during pressure support
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery run time too short or Battery indicator never displays full patient safety
Battery depletes faster than expected; indicator remains empty even after charging
Ranked causes
- 1. Battery capacity is inadequate (common)
- 2. Unit goes into functional failure but no Empty Battery alarm (occasional)
- 3. Battery indicator displays empty after recharge; Defective Electronic Module (rare)
Measure actual battery run time under standard load
- Test point
- Unit operation timer
- Expected
- Run time meets or exceeds manufacturer specifications
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhale tidal volume measurement inaccuracy patient safety
Exhale VT measurement differs from set tidal volume by more than 10% under controlled test bench conditions with verified calibration and patient circuit substitution
Ranked causes
- 1. Leak in test lung or circuit (common)
- 2. High velocity exhale from undersized test lung (occasional)
- 3. MAQUET (Siemens 190) test lung characteristics interfere with measurement (occasional)
Verify exhale volume matches set tidal volume within ±10%
- Test point
- Flow graph display
- Expected
- Exhale volume within 10% of set tidal volume
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Black screen on display
Screen appears completely black; unit may be reported as not booting or display defective
Ranked causes
- 1. LCD backlight or inverter failure (common)
- 2. Defective LCD display (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blower pressure branch pneumatic leak caution
Air escapes from blower pressure sensing tube or exhalation valve control line; exhalation valve cannot close fully; negative flow measurement observed on display
Ranked causes
- 1. Barbed connector leak behind front panel luer of exhalation valve control line (occasional)
- 2. Small barb of blower pressure connection tee leak (occasional)
Set up ventilator and patient circuit
- Test point
- Per Section 3.4.1
- Expected
- Ventilator operational and patient circuit connected
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Buzzer is audibly too low in volume caution
Alarm volume insufficient; buzzer output is weak or inaudible during alarms
Ranked causes
- 1. Alarm Volume level is not adequate in options menu (common)
- 2. Defective Interface Board (occasional)
Check Alarm Volume setting
- Test point
- Options menu Alarm Volume control
- Expected
- Volume slider set to adequate level; alarm sounds audible
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Demand valve leakage or fails VVT caution
Device leaks oxygen flow or fails to shut off with back pressure; may fail VVT when reinstalled
Ranked causes
- 1. Demand valve positioned ahead of pneumatic unit O2 intake port due to flexible tubing loss of flexibility (occasional)
- 2. Two versions of demand valve with different actuation requirements (unknown)
Connect O2 to demand valve DISS fitting and apply pressure
- Test point
- DISS fitting on demand valve
- Expected
- O2 flows when manual button actuated; no flow when button released psi
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code 316 found in error log caution
Technical log contains error code 316; user likely held key or control knob for extended duration
Ranked causes
- 1. User pressed key or control knob for more than 5 seconds (common)
- 2. Keypad is defective or intermittent (occasional)
Review error log for code 316 frequency
- Test point
- Technical log file (error.log)
- Expected
- Code 316 absent or infrequent after user instruction
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error Code 359 in technical log; O2 system out of calibration patient safety
Error code 359 logged; O2 mixer started in extreme 21% position during last power-up or system out of calibration
Ranked causes
- 1. Mixer started in extreme 21% position (common)
- 2. O2 system out of calibration (occasional)
Inspect stepper motor and microswitch cam position
- Test point
- Mixer stepper motor and microswitch cam
- Expected
- Microswitch cam positioned in mid-range position
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fails VVT, O2 Pressure Switch on 100% patient safety
VVT O2 pressure switch test fails with 100% reading; O2 supply pressure insufficient or switch defective
Ranked causes
- 1. O2 supply pressure is less than 40 psi (common)
- 2. O2 Pressure Switch out of acceptable range (occasional)
Measure O2 supply pressure at input
- Test point
- O2 inlet connection to unit
- Expected
- Pressure reads 40 psi or greater throughout VVT, >= 40 psi
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fails VVT, Solenoid failure; Error code 330 in technical log patient safety
VVT test fails on solenoid check; error code 330 logged; solenoid valves stuck or defective
Ranked causes
- 1. Defective contact on Main Pneumatic connector (common)
- 2. Defective Pneumatic Unit (occasional)
Inspect Main Pneumatic connector contacts
- Test point
- Main Pneumatic connector pins and sockets
- Expected
- All contacts clean, shiny, and fully engaged with low resistance
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow offset on graph with different circuits; Fails OVT, Patient Circuit Failed patient safety
Flow offset persists across multiple patient circuits; patient circuit malfunction or internal tubing contamination
Ranked causes
- 1. Internal tubing is contaminated with fluid (common)
- 2. Missing or non-functioning one-way valve in patient circuit (occasional)
Disconnect patient circuit and manually inspect tubing
- Test point
- Internal sensor tubing
- Expected
- No fluid present in internal tubing; clear and dry
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front LED doesn't light or is intermittent caution
Front panel LED indicator fails to illuminate or flickers intermittently during operation
Ranked causes
- 1. Conductive coating shorting LED leads (common)
- 2. Defective keypad (occasional)
Visually inspect inside front enclosure for conductive coating
- Test point
- LED assembly area and surrounding enclosure
- Expected
- No conductive residue on LED leads or surrounding components
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leakage at Demand Valve outlet; O2% higher than 21% at 21% setting only with high pressure O2 patient safety
O2 concentration reads above 21% when set to 21%; demand valve entrains O2 when high pressure supplied
Ranked causes
- 1. Leakage at Demand Valve outlet (silicone) connection (common)
- 2. Demand Valve is entraining O2 (occasional)
Measure O2% output with unit set to 21% and high pressure O2 supplied
- Test point
- O2 output sensor
- Expected
- O2% reads exactly 21.0% ±2% when set to 21%, ±2%
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No blower pressure or turbine doesn't turn; Error Codes 345, 347 in technical log patient safety
Blower fails to generate pressure; turbine rotor does not rotate; error codes 345 and 347 logged
Ranked causes
- 1. Defective Electronic Module (common)
- 2. Defective Turbine (occasional)
Verify Electronic Module output control signal to blower
- Test point
- Blower control signal at Electronic Module
- Expected
- Control signal present and varying during operation
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 microswitch malfunction or misalignment caution
O2 microswitch does not change state properly when actuated by cam; fails closed or open conditions
Ranked causes
- 1. Microswitch actuator not properly positioned on cam or switch defective (unknown)
- 2. Stepper motor shaft binding or misalignment preventing proper cam rotation (unknown)
Connect ohmmeter to microswitch and rotate shaft to end position
- Test point
- Pins 7 & 8 in PU main connector
- Expected
- Switch in open state at end position on cam
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure switch malfunction or incorrect reading caution
O2 pressure switch does not open or close properly; fails to detect pressure conditions
Ranked causes
- 1. Pressure switch not properly connected or defective (unknown)
- 2. Two versions with different specifications and pin configurations (unknown)
Connect ohmmeter to pressure switch terminals per type and measure state
- Test point
- Pins 1 & 2 (silver type) or rear tabs (brass type)
- Expected
- Switch normally open (no O2); closes with high pressure O2 present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit displays Call Service/Battery Disconnect or Battery Damaged error patient safety
Error alarm indicates battery connection or function problem; charging LED lights with AC power
Ranked causes
- 1. Power Pack fuse is blown (common)
- 2. Battery capacity is inadequate; Fails VVT, Battery check (occasional)
Inspect Power Pack fuse for rupture
- Test point
- Power Pack fuse element
- Expected
- Fuse element intact with continuity
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit displays Patient Circuit Failed alarm during normal operation patient safety
Patient Circuit Failed alarm displays during normal ventilation; sensing line contaminated or one-way valve failed
Ranked causes
- 1. Sensing lines of patient circuit partially contaminated with fluid (common)
- 2. Pressure graph waveform is noisy from contamination (occasional)
Inspect patient circuit sensing tubes for fluid or debris
- Test point
- Patient circuit sensing line inlets at front panel
- Expected
- Sensing tubes clean, dry, and clear
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit displays temperature error; Error Code 304 in technical log caution
Temperature error alarm displays; error code 304 logged; cooling fan issue suspected
Ranked causes
- 1. Cooling fan disconnected or defective connector contact (common)
- 2. Cabling interfering with cooling fan (occasional)
Inspect cooling fan connector
- Test point
- Cooling fan connector J
- Expected
- Connector fully seated with no corrosion; contact resistance low
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power up on AC power high voltage
Unit fails to power on when AC power is supplied; no response to power switch
Ranked causes
- 1. No continuity through power switch terminals (common)
- 2. Defective Electronic Module (occasional)
Verify power switch continuity
- Test point
- Power switch terminals
- Expected
- Continuity present across switch terminals when in ON position
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit powered off without user intervention
Unit unexpectedly powers down during operation; AC power lost or battery depleted
Ranked causes
- 1. AC fuse blown or discharged battery (common)
- 2. Unit running on battery with AC power disconnected (occasional)
Review technical log for AC disconnect events
- Test point
- Event log file
- Expected
- No AC power disconnect entries during operation
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit powers off when power switch is released
Power switch does not latch into the ON position; unit immediately loses power
Ranked causes
- 1. Gasket still present in switch body (common)
- 2. Defective On/Off switch (occasional)
Inspect power switch body for gasket material
- Test point
- Power switch
- Expected
- No gasket present in switch body
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit powers up but does not boot up
Green LED lights but unit fails to complete boot sequence; stuck during startup
Ranked causes
- 1. Wrong software revision installed or installation interrupted (common)
- 2. Defective Electronic Module (occasional)
Verify software revision via technical log
- Test point
- Technical log files (error.log, events.log)
- Expected
- Current released software revision installed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
100% O2 suction mode activation patient safety
Verify 100% O2 button triggers green pop-up, FiO2 elevates to 100%, and 2-minute alarm silence timer starts
Ranked causes
- 1. Front panel 100% O2 button malfunction or O2 delivery system failure (unknown)
Green pop-up appears next to FiO2 field
- Test point
- FiO2 field display area
- Expected
- Green pop-up window visible
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Apnea alarm detection and recovery patient safety
Verify apnea alarm triggers after 20-25 seconds at low rate, mode changes to blinking APNEA, and unit recovers on patient triggers after first minute
Ranked causes
- 1. Apnea detection sensor failure or alarm timer logic malfunction (unknown)
Set ventilator rate to 2 bpm
- Test point
- Rate setting field
- Expected
- 2 bpm
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charge LED doesn't light with AC power cord plugged in high voltage
Battery Charge indicator LED fails to illuminate when AC power connected; charging circuit issue
Ranked causes
- 1. AC Fuse blown (common)
Inspect AC fuse elements
- Test point
- AC fuse cartridges
- Expected
- Fuse element intact with continuity across terminals
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charge LED lights with AC power cord plugged in but battery remains discharged
Charge indicator LED illuminates but battery does not charge; battery depleted despite powered charger
Ranked causes
- 1. Battery is discharged (common)
Monitor battery voltage during charging
- Test point
- Battery voltage at internal terminals
- Expected
- Battery voltage increases steadily during 8-hour charge cycle
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control knob feels loose or worn out caution
Control knob has excessive play or exhibits wear in mechanical response
Ranked causes
- 1. Encoder mounting shaft nut is loose (common)
Inspect encoder shaft nut tightness
- Test point
- Encoder mounting shaft
- Expected
- Shaft nut firmly secured with no play in knob
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cursor moves without touching control knob caution
On-screen cursor or selection indicator moves without user input on the control knob
Ranked causes
- 1. Defective encoder (common)
Monitor cursor movement without touching knob
- Test point
- Control knob encoder output
- Expected
- No cursor movement without physical knob input
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Defective O2 sensor; Error Code 307 in technical log patient safety
Error code 307 logged; O2 sensor reading out of range or non-responsive
Ranked causes
- 1. Defective O2 sensor (common)
Test O2 sensor response to known O2 concentrations
- Test point
- O2 sensor output signal
- Expected
- Sensor output varies linearly with O2 concentration; range 0-25%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display image is upside-down, reversed, monocolored, fuzzy, etc.
LCD display shows distorted, inverted, or incorrect image; Screen Capture image is correct
Ranked causes
- 1. LCD connector is partially dislodged (common)
Visually inspect LCD connector at display module
- Test point
- LCD connector J1
- Expected
- Connector fully seated with retaining bracket locked
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display is dim or flickers
LCD display shows reduced brightness or intermittent flickering during operation
Ranked causes
- 1. Defective LCD Display or Inverter Board (common)
Test display brightness and flicker frequency
- Test point
- LCD display panel
- Expected
- Steady bright image with no flicker or dimming
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Erratic O2 control patient safety
FiO2 setpoint control is unstable or fails to stabilize at targeted O2 mix value
Ranked causes
- 1. Unstable minute volume caused by auto-triggering (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error Code 360 found in technical log; Sensors Disconnect alarm caution
Error code 360 logged; Sensors Disconnect alarm generated; unit passes VVT and Functional Tests
Ranked causes
- 1. Ventilator settings not optimized for patient or connecting interface (common)
Review ventilator settings and patient interface configuration
- Test point
- Operating parameters and patient circuit type
- Expected
- Settings matched to patient category and circuit type; no alarm after rerun
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fails calibration: Calibrate Flow Sensor step
Unit cannot complete the Calibrate Flow Sensor calibration procedure
Ranked causes
- 1. O2 mixer set to 100% O2 position (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fails O2 Calibration; Error Code 375 in technical log patient safety
O2 system calibration test fails; error code 375 indicates O2 sensor range inadequate
Ranked causes
- 1. O2 sensor range is inadequate (common)
Test O2 sensor response across full operating range
- Test point
- O2 sensor output signal
- Expected
- Sensor output spans full range 0-25% O2 concentration
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fails VVT, Blower or Patient Pressure; PIP reaches <65 cmH2O during VVT patient safety
VVT shows insufficient peak inspiratory pressure; unit out of calibration or hardware fault
Ranked causes
- 1. Unit out of calibration (common)
Run VVT with baseline calibration
- Test point
- Pressure measurement during VVT
- Expected
- PIP reaches minimum 65 cmH2O during test
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fails VVT, Blower or Patient Pressure; PIP reaches <65 cmH2O when calibrated patient safety
Even after calibration, peak inspiratory pressure insufficient; hardware fault suspected
Ranked causes
- 1. Outlet Muffler broken and leaking (common)
Visually inspect outlet muffler for cracks or damage
- Test point
- Ventilator outlet muffler housing
- Expected
- Muffler intact with no visible cracks or leaks
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fails VVT, Flow Zeroing right after Flow Performance patient safety
VVT test fails on Flow Zeroing step immediately after Flow Performance check; zeroing/purge issue
Ranked causes
- 1. Zeroing/ Purge Board disconnected (common)
Inspect Zeroing/Purge Board connector seating
- Test point
- Zeroing/Purge Board connector at Main Board
- Expected
- Board fully seated and retaining mechanism engaged
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
FiO2 return to baseline after 100% O2 timeout patient safety
Verify FiO2 setting and delivery return to 21% one minute after 2-minute alarm silence timer expires
Ranked causes
- 1. 100% O2 mode timeout logic failure or O2 control system malfunction (unknown)
FiO2 return to 21% after timeout
- Test point
- FiO2 display and delivery measurement
- Expected
- 21%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow sensor leak VVT failure with no actual leak patient safety
Unit reports flow sensor leak failure during VVT testing but no pneumatic leak is present; operator suspects false alarm
Ranked causes
- 1. Electronic module fault detecting non-existent leak (occasional)
Set up ventilator and patient circuit per Section 6.2.4
- Test point
- VVT setup
- Expected
- Ventilator and patient circuit properly configured
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow sensor leakage in patient circuit caution
Perceptible leakage detected in patient circuit pneumatic path; oxygen escaping from tubing connections or components
Ranked causes
- 1. Leak at one of approximately 50 tubing connection points or component connections (common)
High pressure alarm threshold validation patient safety
Verify high pressure alarm triggers at set threshold and can be adjusted in Alarm Settings
Ranked causes
- 1. High pressure alarm sensor failure or threshold circuitry malfunction (unknown)
Access Alarm Settings and Pressure group
- Test point
- Main Menu > Alarm Settings > Pressure group
- Expected
- Pressure group highlighted
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inhale tidal volume accuracy verification
Verification that inhale VT measurement is accurate using calibration manometer and test lung
Ranked causes
- 1. Pressure calibration verified with test lung (common)
Measure peak inspiratory pressure (PIP) at 70+ kg default settings
- Test point
- Pressure display or external manometer
- Expected
- Approximately 34 cmH2O, ±1 cmH2O
- Tools
- Certified manometer, VersaMed 2-liter test lung P/N 910V0005-01
Internal O2 measurement verification needed
Validate O2 sensor reads correct values at different FiO2 settings (21%, 60%, 100%) across patient weights
Ranked causes
- 1. O2 sensor calibration or accuracy drift (unknown)
View current O2 measurement value
- Test point
- Alarm Settings, O2 value field (blue)
- Expected
- O2 value displayed for current FiO2 setting
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low pressure during calibration or VVT patient safety
Calibration or Volume Verification Test (VVT) shows pressures lower than normal, indicating power source issue
Ranked causes
- 1. Ventilator running on battery power instead of AC mains (common)
Verify AC power connection
- Test point
- AC power inlet and indicator light
- Expected
- AC power connected and power indicator illuminated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Need Cal alarm after power up; Error Code 306 in technical log patient safety
Calibration required alarm displays; error code 306 indicates O2 system is out of calibration
Ranked causes
- 1. O2 system out of calibration (common)
Run O2 system calibration procedure
- Test point
- O2 sensor and control system
- Expected
- Calibration completes successfully; alarm clears
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Negative (-) flow sensor port leak isolation
Diagnostic procedure to isolate leaks in the negative flow sensor branch
Ranked causes
- 1. Leak in negative flow sensor branch (unknown)
Occlude negative flow sensor port
- Test point
- Negative flow sensor port tube
- Expected
- Flow graph trace goes negative or stays in negative range
Negative flow port and patient pressure branch leak caution
Leak in shared negative flow and patient pressure tubing system with multiple connection junctions; requires partial disassembly for isolation
Ranked causes
- 1. Leak at tubing junctions or connections in negative/patient pressure branch (occasional)
Disassemble and reconfigure unit for testing
- Test point
- Electronic module positioned for tube access
- Expected
- Module configured per Figure 8-7 with tubing reconnected to external fittings
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Negative flow port pneumatic leak caution
Leak in left-side (negative) flow sensing tube; flow graph trace goes above baseline; leftmost metal male locking luer connector affected
Ranked causes
- 1. Leak in negative flow port tubing or connectors (common)
Set up ventilator and patient circuit
- Test point
- Per Section 6.2.4 page 90
- Expected
- Ventilator and circuit operational
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2% higher than 21% at 21% setting only when high pressure O2 not supplied patient safety
O2 concentration measures above 21% when set to 21%; issue only present without high pressure O2
Ranked causes
- 1. O2 system out of calibration at 21% (common)
Measure O2% with unit at 21% setting and normal room air O2 only
- Test point
- O2 output sensor
- Expected
- O2% reads 21.0% ±2%, ±2%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Circuit connections detect anomaly; Fails VVT, Flow sensor leak patient safety
VVT detects flow sensor malfunction; patient circuit connections or Electronic Module defective
Ranked causes
- 1. Defective Electronic Module (common)
Verify patient circuit connections are secure
- Test point
- Patient circuit connectors at front panel
- Expected
- All connectors fully seated and locked
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient disconnect alarm detection patient safety
Verify alarm and indicator response when test lung and Rp20 resistor disconnected from patient circuit
Ranked causes
- 1. Patient disconnect sensor failure or alarm output malfunction (unknown)
Disconnect test lung and Rp20 resistor
- Test point
- Patient circuit connection point
- Expected
- Components removed from patient circuit
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient disconnect detection during 100% O2 mode patient safety
Verify alarm and indicator behavior when test lung and Rp20 resistor disconnected during first 2 minutes and after timer expires
Ranked causes
- 1. Patient disconnect sensor failure or alarm circuitry malfunction (unknown)
Patient disconnect alarm during first 2 minutes
- Test point
- Disconnect test lung and Rp20 resistor
- Expected
- Alarm sounds once; red Pat disc indicator appears; Low press indicator may appear
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pneumatic leak at internal tubing; Fails VVT, Flow sensor leak patient safety
VVT detects pneumatic leak in internal tubing; flow sensor measurement compromised
Ranked causes
- 1. Internal tubing leak identified during testing (common)
Perform internal tubing leak location procedure
- Test point
- Internal pneumatic pathways (see Section 8.2.5)
- Expected
- No fluid leakage; all tubing connections intact
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pneumatic leak at sensing tubes connection patient safety
VVT detects pneumatic leak at patient circuit sensing tube connections; Flow sensor leak suspected
Ranked causes
- 1. Patient Circuit connections loose or defective (common)
Inspect all sensing tube connections at front panel
- Test point
- Patient circuit sensing tube connectors
- Expected
- All connections finger-tight and secure with no audible hiss
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Positive (+) flow sensor port leak isolation
Diagnostic procedure to isolate leaks in the positive flow sensor branch
Ranked causes
- 1. Leak in positive flow sensor branch (unknown)
Occlude positive flow sensor port sensing tube
- Test point
- Positive flow sensor port tube
- Expected
- Flow graph trace goes positive or hugs baseline; remains in positive region even during ventilation
Positive flow port pneumatic leak caution
Leak in right-side (positive) flow sensing tube; flow graph trace goes below baseline; female locking luer connector affected
Ranked causes
- 1. Leak in positive flow port tubing or connectors (common)
Set up ventilator and patient circuit
- Test point
- Per Section 6.2.4 page 90
- Expected
- Ventilator and circuit operational
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor disconnect alarm detection patient safety
Verify alarm and indicator response when patient circuit flow sensor connectors disconnected from ventilator front panel
Ranked causes
- 1. Flow sensor connector failure or alarm output malfunction (unknown)
Disconnect flow sensor connectors
- Test point
- Two clear lines at lower left of ventilator front panel
- Expected
- Connectors removed
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Service Notice! Alarm after power up; Error Codes 330, 333, 345, 347, 359
- Slow rise or decay of pressure on graph and PIP bar
- Tube disconnect alarm detection
- Turbine is noisy; Unit appears to pressurize normally
- Unit blows AC fuse immediately upon power up
- Unit continues to function but displays Low Battery or Empty Battery alarm
- Unit doesn't boot up on battery power
- Unit power up sequence is normal but the display is black
- Unit powers on only while holding switch button
- Unit powers up but restarts over and over again
- Unit won't pass OVT; Flow offset on graph
- Unstable O2 control and minute volume
- User interface timeout behavior
13 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1 | Error logged during operation | Power surge, power dropout, or defective Electronic Module | warning |
304 | Temperature error detected | Cooling fan disconnected, defective connector, or cable interference | error |
306 | O2 system requires calibration | O2 system detected out of acceptable operating range | warning |
307 | O2 sensor malfunction | O2 sensor reading outside acceptable range or unresponsive | error |
316 | Keypad input held too long | User pressed and held key or control knob for more than 5 seconds | info |
330 | Solenoid valve malfunction detected | Main Pneumatic connector loose, defective solenoid contact, or defective Pneumatic Unit | error |
333 | Pneumatic system malfunction | Main Pneumatic connector disconnected or defective pneumatic component | error |
345 | Blower or pressure system failure | No blower pressure output, turbine not rotating, or Electronic Module defect | critical |
347 | Turbine or blower control failure | Turbine not responding to control signals or Electronic Module defective | critical |
359 | O2 mixer stepper motor position error or O2 system out of calibration | Mixer started in extreme 21% position or O2 system calibration drift | error |
360 | Sensor disconnect or sensitivity issue detected | Sensor signal loss, contamination, or ventilator settings not optimized for interface | warning |
375 | O2 sensor calibration failure | O2 sensor range inadequate or sensor malfunction during calibration | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI