Verathon BladderScan Prime: troubleshooting and error codes
The ServiceForge knowledge base holds 25 documented symptoms for the Verathon BladderScan Prime, with 53 ranked causes, 40 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Lithium-ion battery hazard: leakage, explosion, fire, or injury caution
Battery leaks, explodes, catches fire, or causes injury during storage, handling, charging, or transportation
Ranked causes
- 1. Battery stored in console for extended period (occasional)
- 2. Battery short-circuited by contact between terminals and conductive object (rare)
- 3. Battery exposed to abnormal shock, vibration, or pressure (rare)
- 4. Battery disassembled, heated above 60°C, or incinerated (rare)
- 5. Battery leaking or case cracked (rare)
- 6. Battery accessible to children before use (rare)
- 7. Battery terminals not insulated during transportation (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device damage from improper cleaning or disinfection caution
Equipment damaged due to immersion, water exposure, sterilization by prohibited methods, bleach use, or abrasive brushes
Ranked causes
- 1. Instrument immersed in cleaning or disinfectant solution (occasional)
- 2. Console exposed to water during probe cleaning (occasional)
- 3. Instrument subjected to steam, ethylene oxide, radiation, or autoclaving (occasional)
- 4. Bleach-based products used on touchscreen (occasional)
- 5. Metal or abrasive brushes used on instrument surface (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical safety hazard from improper power supply or battery charger high voltage
Risk of electric shock, burns, or fires due to use of non-approved power supply, battery charger, or improper electrical connections
Ranked causes
- 1. Non-approved power supply or battery charger used (occasional)
- 2. Power cord or adapter not connected to properly grounded plug or disconnect not accessible (occasional)
- 3. Battery charger used with visible cracks or damage (occasional)
- 4. Operator or patient touching power supply during use or in contact with battery charger power cord (occasional)
- 5. Non-Verathon approved accessories or peripherals connected (occasional)
Inspect battery charger for physical damage
- Test point
- Charger housing and power cord
- Expected
- No visible cracks, breaks, or damage
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electromagnetic interference affecting system or adjacent equipment caution
System radiates RF energy causing interference with nearby devices or experiences degraded performance due to external RF sources
Ranked causes
- 1. System used adjacent to or stacked with other equipment (occasional)
- 2. Portable or mobile RF communications equipment (cellular phones) in vicinity (occasional)
- 3. Unshielded cables or non-specified accessories used (occasional)
Verify normal system operation in intended configuration
- Test point
- System display and functionality
- Expected
- Normal operation without degradation
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate bladder volume measurement patient safety
Scan result shows off-target aiming, EdgeScan condition, or pubic bone interference affecting measurement accuracy
Ranked causes
- 1. Probe not properly positioned or angled (common)
- 2. Excessive pressure applied to probe (occasional)
- 3. Pubic bone obscuring bladder in image (occasional)
Confirm bladder position in ultrasound image
- Test point
- Ultrasound display crosshairs center
- Expected
- Bladder appears in center of crosshairs on Results screen
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe aiming inaccurate or off-target patient safety
Scan unsuccessful during aiming; moving probe in expected direction moves bladder further off-target instead of toward it.
Ranked causes
- 1. Incorrect position selection on console (occasional)
- 2. Improper probe holding technique (common)
- 3. Probe display not aligned with patient anatomy (occasional)
Verify correct patient position selected on screen
- Test point
- Before you Scan screen position selector
- Expected
- Position matches operator location relative to patient
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to locate bladder on ultrasound image
Operator cannot visualize bladder during pre-scan or scanning procedure despite proper probe positioning technique
Ranked causes
- 1. Inadequate ultrasound gel application (common)
- 2. Air gaps between probe and patient skin (common)
- 3. Probe not properly angled toward bladder (common)
Verify adequate ultrasound gel presence on abdomen
- Test point
- Contact surface between probe and patient skin
- Expected
- Ample gel visible with minimal air bubbles at contact site
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on
Instrument fails to turn on; may indicate dead or discharged battery or system malfunction.
Ranked causes
- 1. Battery dead or discharged (common)
- 2. Battery charge too low to function (common)
- 3. System not responding; requires reset (occasional)
Verify battery status and charge level
- Test point
- Battery well
- Expected
- Battery present and charged
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Irregular printer printouts caution
Printer produces distorted, streaked, or poor-quality thermal paper output; may indicate dirty print head or misaligned drive roller.
Ranked causes
- 1. Dirty or contaminated print head (common)
- 2. Printer drive roller fallen out or misaligned (occasional)
Verify console in standby mode
- Test point
- Console Standby button
- Expected
- Console in standby; ready for printer access
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper jam in printer caution
Printer jammed; paper stuck, ripped, or folded inside mechanism; printing not possible until jam cleared.
Ranked causes
- 1. Paper jammed, ripped, or folded in printer mechanism (common)
- 2. Printer drive roller fallen out during jam (occasional)
Verify console in standby mode
- Test point
- Console Standby button
- Expected
- Console in standby mode; safe to access printer
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe not detected by console
Console displays error message indicating probe is not attached or not recognized.
Ranked causes
- 1. Probe connection loose or disconnected (common)
- 2. Battery issue causing probe detection failure (occasional)
Enter and exit standby mode
- Test point
- Standby button
- Expected
- Console successfully enters and exits standby; probe detected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ultrasound image quality compromised patient safety
Optimal, repeatable image not obtained; accuracy of bladder volume measurement affected
Ranked causes
- 1. Catheter in bladder introducing air or interfering with balloon (occasional)
- 2. Scar tissue, surgical incisions, sutures, or staples from abdominal surgery (occasional)
Verify optimal, repeatable image obtained
- Test point
- Bladder ultrasound display
- Expected
- Clear, consistent image of bladder
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable or accessory damage from excessive bending or twisting caution
Cables or connected accessories damaged due to improper handling during use or storage
Ranked causes
- 1. Cables excessively twisted or bent during use or storage (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cracks in console or probe allowing fluid ingress patient safety
Visible cracks in console, probe, probe cable allowing fluid to enter device may affect instrument performance
Ranked causes
- 1. Physical damage to console or probe housing (unknown)
Visual inspection for cracks or damage
- Test point
- console, probe, probe cable, power supply, power cords, batteries, plugs
- Expected
- no visible cracks or damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device exposure contraindication: fetal or pregnant patient patient safety
System used on fetal or pregnant patient populations for which it is not intended
Ranked causes
- 1. System not intended for fetal use or pregnant patients (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device exposure contraindication: open skin wounds or ascites patient safety
System used on patients with contraindicated skin condition or abdominal fluid condition
Ranked causes
- 1. Open skin or wounds in suprapubic region or ascites present (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical shock risk from opening system components high voltage
Operator attempts to disassemble console or internal components, risking electrocution or equipment damage
Ranked causes
- 1. System components opened or disassembled by operator (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Improper cleaning or disinfection affecting device contamination status patient safety
Device not adequately cleaned before disinfection, resulting in contaminated instrument after disinfection procedure
Ranked causes
- 1. Cleaning performed inadequately or skipped before disinfection (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer drive roller fallen out or misaligned caution
Printer drive roller has fallen from gray clips in door; may result in irregular or no printing and inability to close printer door.
Ranked causes
- 1. Printer dropped or impact to printer door area (occasional)
Verify console in standby mode
- Test point
- Console Standby button
- Expected
- Console in standby; safe to access printer
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer not responding when print commanded caution
Printer does not respond when Print is tapped on console; may indicate electrical contact issue.
Ranked causes
- 1. Dirty or corroded electrical contacts between console and printer (common)
Remove battery and prepare console for access
- Test point
- Battery well and mobile stand mounting (if applicable)
- Expected
- Battery removed; console secured and ready for disassembly
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scan measurement exceeds bladder volume due to ultrasound cone size
Measurement shows greater than symbol indicating bladder extends beyond ultrasound image cone boundaries
Ranked causes
- 1. Large bladder volume exceeds ultrasound scan field of view (occasional)
Observe measurement display for greater than symbol
- Test point
- Volume measurement on Results screen
- Expected
- Volume measurement without greater than symbol for bladders within scan cone; greater than symbol for oversized bladders
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System modified or unauthorized alterations made patient safety
Equipment modified or altered from manufacturer specifications, affecting safety and performance
Ranked causes
- 1. Equipment modified or altered after delivery (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System used for imaging biological structures other than bladder patient safety
Device used to display or measure anatomical features not intended for measurement (non-bladder imaging)
Ranked causes
- 1. Operator attempts to use system for off-label imaging (non-bladder anatomy) (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System used with flammable anesthetics patient safety
Device operated in environment with flammable anesthetic gases present, creating explosion risk
Ranked causes
- 1. System operated in presence of flammable anesthetics (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System used with high-frequency surgical equipment high voltage
Device operated in conjunction with high-frequency surgical unit, creating risk of electric shock or burns
Ranked causes
- 1. System connected or operated simultaneously with high-frequency surgical equipment (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI