Vapotherm Precision Flow Plus: troubleshooting and error codes
The ServiceForge knowledge base holds 14 documented symptoms for the Vapotherm Precision Flow Plus, with 20 ranked causes, 2 diagnostic checks with test points and expected values, and 17 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
O2 sensor fault with display dashes patient safety
General fault icon flashing, % O2 displays dashes (--), medium priority audio signal, cannot be muted
Ranked causes
- 1. Depleted or defective O2 sensor (common)
- 2. Bad 3-way solenoid valve (occasional)
- 3. Bad O2 sensor wire harness (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water out alarm patient safety
Water out icon flashing, medium priority audio signal, no water in Disposable Water Path, gas flow continues without heating or circulation
Ranked causes
- 1. Sterile water empty (common)
- 2. Obstructed inlet tube (occasional)
- 3. Faulty DPC water float or water level sensor (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blocked tube alarm patient safety
Blocked tube icon flashing, medium priority audio signal, mutes only during brief reset period, gas flow continues
Ranked causes
- 1. Obstructed or kinked cannula or delivery tube (common)
- 2. DPC not fully seated (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disposable Water Path faulty or not detected patient safety
DWP icon flashing, medium priority audio signal, unit will not run
Ranked causes
- 1. DPC not assembled, defective, not properly seated, or not installed (common)
- 2. Faulty stator motor (water pump) (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery mode operation patient safety
Battery charging icon flashing, medium priority audio signal, unit running on internal backup battery, gas flow and blending continue without heat or circulation
Ranked causes
- 1. AC power disconnected (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cartridge fault detected patient safety
Cartridge alarm icon flashing, medium priority audio signal, cartridge not detected, unit will not run
Ranked causes
- 1. Faulty sensor or cartridge not detected (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas bubbles in water circulation patient safety
Cartridge fault icon continuous (not flashing), low priority audio signal, unit continues to operate
Ranked causes
- 1. Excessive gas diffusion or DPC not fully primed (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas supply pressure out of range caution
Gas supply icon flashing, medium priority audio signal, gas supply pressure outside 4–85 psi (28–586 kPa) range
Ranked causes
- 1. Gas supply disconnected or exhausted (common)
Verify gas supply pressure within operating range
- Test point
- Gas inlet
- Expected
- 40–85 psi (optimal 55±10 psi), ±10 psi
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
General fault alarm triggered patient safety
Flashing icon on LED panel, medium priority audio signal (rapid triple beeps), cannot be muted
Ranked causes
- 1. Sensor or control system malfunction; internal component failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
General fault with dashes and temperature out of range patient safety
General fault icon flashing, % O2 and L/min display dashes (--), temp displays 52–84°C, medium priority cannot be muted
Ranked causes
- 1. System control failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Selected flow cannot be provided caution
Gas supply icon continuous and flow rate parameter display flashes, lowers in value, medium priority audio signal
Ranked causes
- 1. Inlet gas pressure too low for selected flow rate (common)
Verify gas supply pressure sufficient for selected flow
- Test point
- Gas inlet pressure gauge
- Expected
- 40–85 psi minimum; optimal 55±10 psi, ±10 psi
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature flashing above setpoint caution
Temperature parameter display flashes, temperature at least 2°C above set point
Ranked causes
- 1. User entered setpoint much lower than previous temperature (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature flashing below setpoint caution
Temperature parameter display flashes, temperature at least 2°C below set point
Ranked causes
- 1. Low water temperature or new water bag with low initial temperature (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature out of range patient safety
Temperature display shows two dashes flashing, general fault icon, medium priority cannot be muted, unit will not operate
Ranked causes
- 1. Overheating or temperature sensor malfunction (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
52 | Pressure sensor 1 (AIR) out of range | Air pressure sensor reading exceeds operating range | error |
53 | Pressure sensor 2 (O2) out of range | O2 pressure sensor reading exceeds operating range | error |
54 | Mass flow sensor 1 (AIR) out of range | Air mass flow sensor malfunction | error |
55 | Mass flow sensor 2 (O2) out of range | O2 mass flow sensor malfunction | error |
57 | PS4 back pressure out of range | Back pressure sensor 4 reading exceeds operating range | error |
58 | O2 sensor disconnect – unit will not start | O2 sensor is not connected or not present | critical |
59 | Pump tachometer out of range | Water pump motor speed sensor reading out of range | error |
61 | Battery voltage below range | Internal backup battery voltage drops below minimum operating level | warning |
63 | POST test failure | Power-on self-test routine failed during startup | critical |
64 | Spurious reset; Battery voltage below range | Unexpected system reset with low battery voltage | warning |
65 | Watchdog timer failed | System watchdog timer did not respond correctly | critical |
67 | Post RAM test failed | RAM memory test failed during power-on self-test | critical |
68 | Flash memory corrupt | Flash memory integrity check failed | critical |
77 | Heater thermocouple failure | Temperature measurement circuit failure | error |
78 | IR sensor TS1 failure or disconnected | Infrared temperature sensor 1 not responding | error |
79 | IR sensor TS2 failure or disconnected | Infrared temperature sensor 2 not responding | error |
84 | SW or hardware inconsistency | Software and hardware configuration mismatch detected | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.