Vantive TherMax: troubleshooting and error codes
The ServiceForge knowledge base holds 25 documented symptoms for the Vantive TherMax, with 61 ranked causes, 23 diagnostic checks with test points and expected values, and 39 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Built-In On-going Test (BIOT) fails during operation patient safety
Service LED illuminated during normal operation; heater power disconnected; device enters Malfunction mode
Ranked causes
- 1. Temperature sensor reading out of range (common)
- 2. Plate temperature variation exceeds 10°C (common)
- 3. PWM power variance exceeds 20% over 2-second window (occasional)
- 4. System voltages or currents out of specified range (occasional)
- 5. Blood leak detector absent or malfunctioning (occasional)
- 6. Outlet temperature change exceeds 20% deviation over 10 seconds (occasional)
Measure all temperature sensor readings
- Test point
- Plate, inlet, outlet, compensation sensors
- Expected
- >10°C and <50°C
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Power-On Self Test (POST) fails patient safety
Service LED illuminated, device enters Malfunction mode at startup after power applied or hardware reset
Ranked causes
- 1. Software version incompatible with installed circuit boards (unknown)
- 2. Operational voltages out of acceptable limits (unknown)
- 3. Temperature sensor reading out of range (>10°C and <50°C) (unknown)
- 4. Plate temperature sensor difference exceeds 10°C (unknown)
- 5. Bus or analog-to-digital converter malfunction (unknown)
- 6. Leakage current self-test failure (unknown)
Verify installed software version matches installed circuit boards
- Test point
- System firmware and hardware configuration
- Expected
- Software version compatible with circuit boards
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Leak Detector calibration fails
Self-Service Test Leak Detector calibration does not complete; leak sensing not responding.
Ranked causes
- 1. Dirty Leak Detection Sensors (common)
- 2. Defective Blood Leak Detection Sensor (occasional)
- 3. Defective Rear Leak Detection Sensor (occasional)
- 4. Defective Heater Control Board (occasional)
- 5. Cable continuity issue (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front Leak Detection test fails
Self-Service Test indicates Front Leak Detector is not functioning properly.
Ranked causes
- 1. Dirty Blood Leak Detection Sensors (common)
- 2. Defective Blood Leak Detection Sensor (occasional)
- 3. Defective Heater Control Board (occasional)
- 4. Cable continuity issue (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bag Detect calibration fails
Self-Service Test Bag Detect sensor calibration does not complete; bag detection not working properly.
Ranked causes
- 1. Dirty bag detector sensors (common)
- 2. Defective Bag Detection and Leak Detection Receive Board (occasional)
- 3. Cable continuity issue (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bag Type Verification test fails
Self-Service Test indicates Bag Detection Sensor is not properly identifying bag type.
Ranked causes
- 1. Dirty Bag Type Verification Sensors (common)
- 2. Defective Bag Detection and Leak Detection Receive Board (occasional)
- 3. Cable continuity issue (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: TherMax pairing with PrisMax patient safety
TherMax must be paired with PrisMax control unit after calibration and SSTs are completed. Pairing procedure is executed in Service Environment on Configuration page.
Ranked causes
- 1. TherMax not paired with PrisMax (common)
- 2. TherMax previously installed but not paired (occasional)
- 3. TherMax already in Paired List (common)
Verify Installation Procedure completed if TherMax is new
- Test point
- Heater Control Board memory
- Expected
- Calibration and SST data saved in internal memory
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power Delivery Test fails high voltage
Self-Service Test indicates Heater Control Board is not delivering adequate power during heating cycle.
Ranked causes
- 1. Defective Heater Control Board (common)
- 2. Defective Heater Plates (occasional)
- 3. Cable continuity issue (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power Switch Control test fails
Self-Service Test indicates power switch control circuits not responding properly.
Ranked causes
- 1. Defective Display Board (common)
- 2. Defective Display Control Panel (occasional)
- 3. Cable continuity issue (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rear Leak Detection test fails
Self-Service Test indicates Rear Leak Detector is not functioning properly.
Ranked causes
- 1. Dirty Rear Leak Detection Sensors (common)
- 2. Defective Rear Leak Detection Sensor (occasional)
- 3. Cable continuity issue (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Safety System test fails patient safety
Self-Service Test indicates overheat protection safety system is not functioning; Heater Control Board not cutting off power to heater plates when overheat detected.
Ranked causes
- 1. Defective Heater Control Board (common)
- 2. Defective Heater Plates (occasional)
- 3. Cable continuity issue (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature Accuracy Test fails
Self-Service Test (SST) Temperature Accuracy Test result indicates temperature sensor out of range relative to others.
Ranked causes
- 1. Faulty Bag Inlet or Outlet Temperature Sensor (common)
- 2. Defective Heater Plates (occasional)
- 3. Defective Ambient Temperature Sensor (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Therapy Alarm Test (TAT) detects abnormal outlet temperature patient safety
Alarm triggered during operational stage; heating mode disabled if outlet temperature conditions extreme
Ranked causes
- 1. Commanded plate temperature exceeds 45.4°C (occasional)
- 2. Outlet temperature exceeds set-point by >1°C for >10 minutes (occasional)
- 3. Outlet temperature below set-point by >1°C for >5 minutes (occasional)
Monitor outlet temperature against set-point
- Test point
- Outlet temperature sensor
- Expected
- Within ±1°C of set-point for >5 minutes
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Therapy Alarm Test (TAT) detects disposable not loaded or not locked patient safety
Alarm declared when disposable detection or lock sensor indicates device not properly loaded or secured
Ranked causes
- 1. Disposable not inserted in device (common)
- 2. Disposable not fully locked or seated (common)
Verify disposable is loaded in device
- Test point
- Disposable detection sensor
- Expected
- Disposable present and detected
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Blood warmer failure due to ESD damage patient safety
Device malfunction caused by improper handling of electronic circuits during disassembly without electrostatic protection
Ranked causes
- 1. Electrostatic discharge (ESD) damage to electronic circuits (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cycle Heating LED test fails
Self-Service Test indicates green Heating LED is not operating or responding.
Ranked causes
- 1. Defective Display Board (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cycle Power LED test fails
Self-Service Test indicates idle (amber) and on (green) power LEDs not cycling or functioning.
Ranked causes
- 1. Defective Display Board (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cycle Service LED test fails
Self-Service Test indicates red Service LED is not cycling or responding to test commands.
Ranked causes
- 1. Defective Display Board (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Live Outputs/Mechanical Switch test fails
Self-Service Test indicates bag lock mechanism output or mechanical switch not operating.
Ranked causes
- 1. Defective Bag Lock Assembly (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Therapy Alarm Test (TAT) detects cover not latched patient safety
Alarm triggered when cover lock sensor indicates cover is not properly latched
Ranked causes
- 1. Cover not fully closed or latched (common)
Verify cover is fully latched
- Test point
- Cover lock sensor
- Expected
- Cover latched and sensor activated
Therapy Alarm Test (TAT) detects high plate temperature patient safety
Alarm declared when any plate temperature sensor reads >46°C
Ranked causes
- 1. Plate temperature exceeds safe limit (occasional)
Measure all plate temperature sensors
- Test point
- Plate temperature sensors
- Expected
- ≤46°C
Therapy Alarm Test (TAT) detects leak patient safety
Alarm set when fluid leak detected by leak detection system
Ranked causes
- 1. Fluid leak in system (occasional)
Inspect all connections and tubing for visible leaks
- Test point
- Leak detection sensor output
- Expected
- No leaks detected; dry status
Therapy Alarm Test (TAT) detects low blood flow patient safety
Alarm declared when measured blood flow <50 ml/min and >0.5 ml/min; heating disabled if flow <50 ml/min
Ranked causes
- 1. Blood flow rate below 50 ml/min (occasional)
Measure blood flow rate
- Test point
- Flow sensor output
- Expected
- ≥50 ml/min for normal heating operation
Therapy Alarm Test (TAT) detects out-of-level condition caution
Alarm declared when blood warmer tilted >6° from level; second alarm at >21° tilt
Ranked causes
- 1. Device tilted beyond acceptable angle (occasional)
Check device leveling using orientation sensor
- Test point
- Orientation sensor reading
- Expected
- Within ±6° of level
TherMax pairing status unclear patient safety
Determine current pairing state of TherMax with PrisMax to identify required actions.
Ranked causes
- 1. Unknown pairing history of TherMax (unknown)
Determine if TherMax is new or previously installed
- Test point
- Device history or documentation
- Expected
- Installation status clearly identified
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0x0 | No malfunction detected | Normal operation | info |
0x03xxxx | I2C error reading board revision | I2C communication failure reading board revision | error |
0x1 | Software not compatible with board revision | Software version mismatch with hardware | error |
0x1000x | SPI bus post failure | SPI bus power-on self-test failure | critical |
0x1102x | Triac update BIOT failure | Triac control built-in self-test failure | error |
0x11030 | Mains voltage zone difference BIOT failure | Mains voltage differential self-test failure | error |
0x1104x | Mains voltage range BIOT failure | Mains voltage out-of-range self-test failure | error |
0x1105x | Mains RMS voltage BIOT failure | Mains RMS voltage measurement self-test failure | error |
0x1106x | Mains instant voltage BIOT failure | Mains instantaneous voltage measurement self-test failure | error |
0x1107x | Mains power BIOT failure | Mains power monitoring self-test failure | error |
0x15 | Front blood leak sensor BIOT failure | Front leak sensor built-in self-test failure | error |
0x1500x | Current leakage reduction self-test failure | Leakage current monitoring self-test failure | critical |
0x16 | Rear blood leak sensor BIOT failure | Rear leak sensor built-in self-test failure | error |
0x17 | Bag detect sensor 1 BIOT failure | Bag detector 1 built-in self-test failure | error |
0x18 | Bag detect sensor 2 BIOT failure | Bag detector 2 built-in self-test failure | error |
0x1Dxxxx | I2C error for accelerometer sensor | I2C communication failure with accelerometer | error |
0x1E | All on-board HCB I2C devices failed | All I2C devices on heater control board unresponsive | critical |
0x1F | Inlet, outlet, and compensation I2C device failures | All three temperature sensor I2C devices failed | critical |
0x2 | 3.3V power supply out of range | 3.3V rail voltage outside specifications | error |
0x2000x | Plate temperature sensor POST failure | Plate temperature sensor power-on self-test failure | error |
0x3 | 5.0V power supply out of range | 5.0V rail voltage outside specifications | error |
0x32 | Temperature sensors not in range or not consistent | Temperature sensor reading out of range or inconsistent | error |
0x33 | Power applied to heating elements when plate temps too high | Over-temperature condition during heating | critical |
0x34 | Power applied when outlet temp above 43C | Outlet temperature exceeds 43 degrees Celsius | critical |
0x35 | Power applied when outlet temp above setpoint by 1C | Outlet temperature exceeds setpoint plus 1 degree | critical |
0x38 | Controller failed to reach setpoint temperature | Temperature not reaching set value during normal operation | error |
0x3C | Inlet zone power mismatch between commanded and measured | Inlet heating element power does not match expected value | error |
0x3D | Outlet zone power mismatch between commanded and measured | Outlet heating element power does not match expected value | error |
0x3E | Both inlet and outlet zone power mismatch | Both heating zones power mismatch detected | critical |
0x41xxxx | I2C error for I/O bus extender | I2C communication failure with I/O expander | error |
0x48xxxx | I2C error for inlet temperature sensor | I2C communication failure with inlet temp sensor | error |
0x49xxxx | I2C error for outlet temperature sensor | I2C communication failure with outlet temp sensor | error |
0x4Axxxx | I2C error for compensation temperature sensor | I2C communication failure with compensation sensor | error |
0x5 | Preventive maintenance past due | Maintenance interval exceeded | warning |
0x50xxxx | I2C error for digital potentiometer | I2C communication failure with digital potentiometer | error |
0x6 | LRC shutdown protection circuit failure | Leakage reduction circuit failure | critical |
0x8 | FPGA CRC error | FPGA firmware CRC check failure | critical |
0x9 | FPGA diagnosed malfunction | FPGA self-test failure | critical |
0x9000 | Command to switch on membrane power switch failure | Membrane power switch activation failed | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI