Vantive TherMax: troubleshooting and error codes

The ServiceForge knowledge base holds 25 documented symptoms for the Vantive TherMax, with 61 ranked causes, 23 diagnostic checks with test points and expected values, and 39 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Built-In On-going Test (BIOT) fails during operation patient safety

Service LED illuminated during normal operation; heater power disconnected; device enters Malfunction mode

Ranked causes

  1. 1. Temperature sensor reading out of range (common)
  2. 2. Plate temperature variation exceeds 10°C (common)
  3. 3. PWM power variance exceeds 20% over 2-second window (occasional)
  4. 4. System voltages or currents out of specified range (occasional)
  5. 5. Blood leak detector absent or malfunctioning (occasional)
  6. 6. Outlet temperature change exceeds 20% deviation over 10 seconds (occasional)
First check

Measure all temperature sensor readings

Test point
Plate, inlet, outlet, compensation sensors
Expected
>10°C and <50°C

3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Power-On Self Test (POST) fails patient safety

Service LED illuminated, device enters Malfunction mode at startup after power applied or hardware reset

Ranked causes

  1. 1. Software version incompatible with installed circuit boards (unknown)
  2. 2. Operational voltages out of acceptable limits (unknown)
  3. 3. Temperature sensor reading out of range (>10°C and <50°C) (unknown)
  4. 4. Plate temperature sensor difference exceeds 10°C (unknown)
  5. 5. Bus or analog-to-digital converter malfunction (unknown)
  6. 6. Leakage current self-test failure (unknown)
First check

Verify installed software version matches installed circuit boards

Test point
System firmware and hardware configuration
Expected
Software version compatible with circuit boards

3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Leak Detector calibration fails

Self-Service Test Leak Detector calibration does not complete; leak sensing not responding.

Ranked causes

  1. 1. Dirty Leak Detection Sensors (common)
  2. 2. Defective Blood Leak Detection Sensor (occasional)
  3. 3. Defective Rear Leak Detection Sensor (occasional)
  4. 4. Defective Heater Control Board (occasional)
  5. 5. Cable continuity issue (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Front Leak Detection test fails

Self-Service Test indicates Front Leak Detector is not functioning properly.

Ranked causes

  1. 1. Dirty Blood Leak Detection Sensors (common)
  2. 2. Defective Blood Leak Detection Sensor (occasional)
  3. 3. Defective Heater Control Board (occasional)
  4. 4. Cable continuity issue (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bag Detect calibration fails

Self-Service Test Bag Detect sensor calibration does not complete; bag detection not working properly.

Ranked causes

  1. 1. Dirty bag detector sensors (common)
  2. 2. Defective Bag Detection and Leak Detection Receive Board (occasional)
  3. 3. Cable continuity issue (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bag Type Verification test fails

Self-Service Test indicates Bag Detection Sensor is not properly identifying bag type.

Ranked causes

  1. 1. Dirty Bag Type Verification Sensors (common)
  2. 2. Defective Bag Detection and Leak Detection Receive Board (occasional)
  3. 3. Cable continuity issue (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: TherMax pairing with PrisMax patient safety

TherMax must be paired with PrisMax control unit after calibration and SSTs are completed. Pairing procedure is executed in Service Environment on Configuration page.

Ranked causes

  1. 1. TherMax not paired with PrisMax (common)
  2. 2. TherMax previously installed but not paired (occasional)
  3. 3. TherMax already in Paired List (common)
First check

Verify Installation Procedure completed if TherMax is new

Test point
Heater Control Board memory
Expected
Calibration and SST data saved in internal memory

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power Delivery Test fails high voltage

Self-Service Test indicates Heater Control Board is not delivering adequate power during heating cycle.

Ranked causes

  1. 1. Defective Heater Control Board (common)
  2. 2. Defective Heater Plates (occasional)
  3. 3. Cable continuity issue (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power Switch Control test fails

Self-Service Test indicates power switch control circuits not responding properly.

Ranked causes

  1. 1. Defective Display Board (common)
  2. 2. Defective Display Control Panel (occasional)
  3. 3. Cable continuity issue (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rear Leak Detection test fails

Self-Service Test indicates Rear Leak Detector is not functioning properly.

Ranked causes

  1. 1. Dirty Rear Leak Detection Sensors (common)
  2. 2. Defective Rear Leak Detection Sensor (occasional)
  3. 3. Cable continuity issue (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Safety System test fails patient safety

Self-Service Test indicates overheat protection safety system is not functioning; Heater Control Board not cutting off power to heater plates when overheat detected.

Ranked causes

  1. 1. Defective Heater Control Board (common)
  2. 2. Defective Heater Plates (occasional)
  3. 3. Cable continuity issue (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature Accuracy Test fails

Self-Service Test (SST) Temperature Accuracy Test result indicates temperature sensor out of range relative to others.

Ranked causes

  1. 1. Faulty Bag Inlet or Outlet Temperature Sensor (common)
  2. 2. Defective Heater Plates (occasional)
  3. 3. Defective Ambient Temperature Sensor (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Therapy Alarm Test (TAT) detects abnormal outlet temperature patient safety

Alarm triggered during operational stage; heating mode disabled if outlet temperature conditions extreme

Ranked causes

  1. 1. Commanded plate temperature exceeds 45.4°C (occasional)
  2. 2. Outlet temperature exceeds set-point by >1°C for >10 minutes (occasional)
  3. 3. Outlet temperature below set-point by >1°C for >5 minutes (occasional)
First check

Monitor outlet temperature against set-point

Test point
Outlet temperature sensor
Expected
Within ±1°C of set-point for >5 minutes

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Therapy Alarm Test (TAT) detects disposable not loaded or not locked patient safety

Alarm declared when disposable detection or lock sensor indicates device not properly loaded or secured

Ranked causes

  1. 1. Disposable not inserted in device (common)
  2. 2. Disposable not fully locked or seated (common)
First check

Verify disposable is loaded in device

Test point
Disposable detection sensor
Expected
Disposable present and detected

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Blood warmer failure due to ESD damage patient safety

Device malfunction caused by improper handling of electronic circuits during disassembly without electrostatic protection

Ranked causes

  1. 1. Electrostatic discharge (ESD) damage to electronic circuits (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cycle Heating LED test fails

Self-Service Test indicates green Heating LED is not operating or responding.

Ranked causes

  1. 1. Defective Display Board (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cycle Power LED test fails

Self-Service Test indicates idle (amber) and on (green) power LEDs not cycling or functioning.

Ranked causes

  1. 1. Defective Display Board (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cycle Service LED test fails

Self-Service Test indicates red Service LED is not cycling or responding to test commands.

Ranked causes

  1. 1. Defective Display Board (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Live Outputs/Mechanical Switch test fails

Self-Service Test indicates bag lock mechanism output or mechanical switch not operating.

Ranked causes

  1. 1. Defective Bag Lock Assembly (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Therapy Alarm Test (TAT) detects cover not latched patient safety

Alarm triggered when cover lock sensor indicates cover is not properly latched

Ranked causes

  1. 1. Cover not fully closed or latched (common)
First check

Verify cover is fully latched

Test point
Cover lock sensor
Expected
Cover latched and sensor activated

Therapy Alarm Test (TAT) detects high plate temperature patient safety

Alarm declared when any plate temperature sensor reads >46°C

Ranked causes

  1. 1. Plate temperature exceeds safe limit (occasional)
First check

Measure all plate temperature sensors

Test point
Plate temperature sensors
Expected
≤46°C

Therapy Alarm Test (TAT) detects leak patient safety

Alarm set when fluid leak detected by leak detection system

Ranked causes

  1. 1. Fluid leak in system (occasional)
First check

Inspect all connections and tubing for visible leaks

Test point
Leak detection sensor output
Expected
No leaks detected; dry status

Therapy Alarm Test (TAT) detects low blood flow patient safety

Alarm declared when measured blood flow <50 ml/min and >0.5 ml/min; heating disabled if flow <50 ml/min

Ranked causes

  1. 1. Blood flow rate below 50 ml/min (occasional)
First check

Measure blood flow rate

Test point
Flow sensor output
Expected
≥50 ml/min for normal heating operation

Therapy Alarm Test (TAT) detects out-of-level condition caution

Alarm declared when blood warmer tilted >6° from level; second alarm at >21° tilt

Ranked causes

  1. 1. Device tilted beyond acceptable angle (occasional)
First check

Check device leveling using orientation sensor

Test point
Orientation sensor reading
Expected
Within ±6° of level

TherMax pairing status unclear patient safety

Determine current pairing state of TherMax with PrisMax to identify required actions.

Ranked causes

  1. 1. Unknown pairing history of TherMax (unknown)
First check

Determine if TherMax is new or previously installed

Test point
Device history or documentation
Expected
Installation status clearly identified

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
0x0No malfunction detectedNormal operationinfo
0x03xxxxI2C error reading board revisionI2C communication failure reading board revisionerror
0x1Software not compatible with board revisionSoftware version mismatch with hardwareerror
0x1000xSPI bus post failureSPI bus power-on self-test failurecritical
0x1102xTriac update BIOT failureTriac control built-in self-test failureerror
0x11030Mains voltage zone difference BIOT failureMains voltage differential self-test failureerror
0x1104xMains voltage range BIOT failureMains voltage out-of-range self-test failureerror
0x1105xMains RMS voltage BIOT failureMains RMS voltage measurement self-test failureerror
0x1106xMains instant voltage BIOT failureMains instantaneous voltage measurement self-test failureerror
0x1107xMains power BIOT failureMains power monitoring self-test failureerror
0x15Front blood leak sensor BIOT failureFront leak sensor built-in self-test failureerror
0x1500xCurrent leakage reduction self-test failureLeakage current monitoring self-test failurecritical
0x16Rear blood leak sensor BIOT failureRear leak sensor built-in self-test failureerror
0x17Bag detect sensor 1 BIOT failureBag detector 1 built-in self-test failureerror
0x18Bag detect sensor 2 BIOT failureBag detector 2 built-in self-test failureerror
0x1DxxxxI2C error for accelerometer sensorI2C communication failure with accelerometererror
0x1EAll on-board HCB I2C devices failedAll I2C devices on heater control board unresponsivecritical
0x1FInlet, outlet, and compensation I2C device failuresAll three temperature sensor I2C devices failedcritical
0x23.3V power supply out of range3.3V rail voltage outside specificationserror
0x2000xPlate temperature sensor POST failurePlate temperature sensor power-on self-test failureerror
0x35.0V power supply out of range5.0V rail voltage outside specificationserror
0x32Temperature sensors not in range or not consistentTemperature sensor reading out of range or inconsistenterror
0x33Power applied to heating elements when plate temps too highOver-temperature condition during heatingcritical
0x34Power applied when outlet temp above 43COutlet temperature exceeds 43 degrees Celsiuscritical
0x35Power applied when outlet temp above setpoint by 1COutlet temperature exceeds setpoint plus 1 degreecritical
0x38Controller failed to reach setpoint temperatureTemperature not reaching set value during normal operationerror
0x3CInlet zone power mismatch between commanded and measuredInlet heating element power does not match expected valueerror
0x3DOutlet zone power mismatch between commanded and measuredOutlet heating element power does not match expected valueerror
0x3EBoth inlet and outlet zone power mismatchBoth heating zones power mismatch detectedcritical
0x41xxxxI2C error for I/O bus extenderI2C communication failure with I/O expandererror
0x48xxxxI2C error for inlet temperature sensorI2C communication failure with inlet temp sensorerror
0x49xxxxI2C error for outlet temperature sensorI2C communication failure with outlet temp sensorerror
0x4AxxxxI2C error for compensation temperature sensorI2C communication failure with compensation sensorerror
0x5Preventive maintenance past dueMaintenance interval exceededwarning
0x50xxxxI2C error for digital potentiometerI2C communication failure with digital potentiometererror
0x6LRC shutdown protection circuit failureLeakage reduction circuit failurecritical
0x8FPGA CRC errorFPGA firmware CRC check failurecritical
0x9FPGA diagnosed malfunctionFPGA self-test failurecritical
0x9000Command to switch on membrane power switch failureMembrane power switch activation failedcritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI