Vantive Prismaflex: troubleshooting and error codes
The ServiceForge knowledge base holds 192 documented symptoms for the Vantive Prismaflex, with 406 ranked causes, 306 diagnostic checks with test points and expected values, and 85 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
TMP too high patient safety
Transmembrane pressure has reached user-set pressure limit. High pressure in main filter indicates restriction or clotting.
Ranked causes
- 1. Ultrafiltration rate too high for blood flow rate (common)
- 2. Inadequate anticoagulation of extracorporeal circuit (common)
- 3. Kinked lines in blood flowpath (occasional)
- 4. Syringe pump failed or incorrectly installed (occasional)
- 5. Air leak between deaeration chamber monitor line and return pressure sensor (occasional)
- 6. Filter or effluent pod failure (rare)
- 7. Filter or effluent pressure sensor failed (rare)
- 8. Wrong measurement of filter or effluent pressure (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filter clotted patient safety
Blood in filter is clotted, increasing pressure drop observed
Ranked causes
- 1. Inadequate anticoagulation of extracorporeal circuit (common)
- 2. Ultrafiltration too high (common)
- 3. Kinked lines in blood flowpath (occasional)
- 4. Syringe pump anticoagulation system failure (occasional)
- 5. Air leak in deaeration chamber monitor line (occasional)
- 6. Wrong measurement of filter or effluent pressure (rare)
- 7. Filter, effluent or return pressure sensor failed (rare)
Verify syringe pump plunger is moving upward during treatment
- Test point
- Syringe pump
- Expected
- Plunger moving upward
2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Plasmafilter clotting patient safety
Blood in plasmafilter is clotted, showing increasing pressure drop
Ranked causes
- 1. Inadequate anticoagulation of extracorporeal circuit (common)
- 2. Blood flow rate or plasmafiltration rate too high (common)
- 3. Kinked lines in blood flowpath (occasional)
- 4. Syringe pump anticoagulation system failure (occasional)
- 5. Air leak in deaeration chamber monitor line (occasional)
- 6. Filter or return pressure sensor failed (rare)
- 7. Wrong measurement of filter or effluent pressure (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Set disconnection alarm patient safety
Alarm triggered when filter pressure is ≤+10 mmHg but expected operating point is >+10 mmHg
Ranked causes
- 1. Filter pressure pod not installed or debris in sensor housing (common)
- 2. Line between blood pump and filter disconnected (common)
- 3. Blood flowpath obstructed before filter pressure pod (occasional)
- 4. Blood flow rate too low for access device (occasional)
- 5. Filter pressure sensor failed (rare)
- 6. Return line disconnection with return pressure alarm failure (rare)
- 7. Pressure measurement failure (rare)
Inspect filter pressure pod installation and cleanliness
- Test point
- Filter pressure pod and sensor housing
- Expected
- Pod installed, no debris
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ca Line Not Connected patient safety
Dedicated calcium infusion line is not properly connected to the calcium syringe or wrong line type is installed.
Ranked causes
- 1. Calcium line is not connected to the syringe (common)
- 2. Wrong line connected (common)
- 3. Unused syringe line on disposable set connected to calcium syringe (occasional)
- 4. Wrong syringe size installed (occasional)
- 5. Air in syringe (occasional)
- 6. Missing non-return valve on calcium infusion line (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Access patient safety
Access pressure deviation from operating point: ±50 mmHg (±70 mmHg if blood flow>200ml/min) depending on operating point range, or pressure above 0 mmHg when operating point below –10 mmHg.
Ranked causes
- 1. Possible leakage or disconnection of access line or catheter (common)
- 2. Possible kink or obstruction in access line or catheter (common)
- 3. Patient is coughing or being moved (occasional)
- 4. Catheter is clotted or out of position (occasional)
- 5. Blood flow rate is too high (occasional)
- 6. Blood flowpath is obstructed after access pressure pod (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow problem detected on fluid line patient safety
System detects abnormal flow rate on indicated fluid path; alarm triggered; operator or service action required
Ranked causes
- 1. Closed clamp or major leak on line or bag (common)
- 2. Effluent drain port not fully closed (occasional)
- 3. Bag connector not firmly tightened (Luer connection) (occasional)
- 4. Incorrect puncture of solution bag (spike connection) (occasional)
- 5. Air bubbles in solution line or effluent (occasional)
- 6. Non-occlusive pump or scale failure (rare)
Inspect for closed clamps or leaks on fluid lines and bags
- Test point
- All line clamps and bag connections
- Expected
- All clamps open, no visible leaks, bag hanging freely
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Return pressure dropping patient safety
Return pressure 50 mmHg or 70 mmHg (when blood flow >200 ml/min) below operating point
Ranked causes
- 1. Return line or catheter leak or disconnection (common)
- 2. Patient moving or being repositioned (common)
- 3. Blood flowpath obstructed or leaking before deaeration chamber (occasional)
- 4. Hydrophobic membrane wet or service line disconnected (occasional)
- 5. Return pressure sensor failed (rare)
- 6. Blood leak detected in MARS monitor (rare)
Verify return line and catheter connections
- Test point
- Return catheter to return line and patient connection points
- Expected
- Securely connected, no leakage
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood leak detected patient safety
Alarm triggered by blood leak detector sensing fluid in effluent line path
Ranked causes
- 1. Air bubble in effluent line (common)
- 2. Effluent line not properly installed in detector (common)
- 3. Liquid or debris in detector tubing (occasional)
- 4. Leak in filter membrane (occasional)
- 5. Formed elements or lipids in plasma (TPE) (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calcium line not connected to syringe patient safety
Calcium infusion line not properly connected or wrong line type installed
Ranked causes
- 1. Calcium infusion line not connected to syringe (common)
- 2. Wrong line connected (common)
- 3. Unused syringe line from set connected (occasional)
- 4. Wrong syringe size installed (occasional)
- 5. Air in syringe (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Prime Self-Test Code 20 failure
Pump occlusivity test failed during prime
Ranked causes
- 1. Return line misaligned or obstructed (common)
- 2. Monitor line connection issue (occasional)
- 3. Pump segments misloaded or obstructed (occasional)
- 4. Pump failure (rare)
- 5. Pressure sensor or ARPS circuit failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Prime Self-Test failure - Pressure Pod/Sensor patient safety
Post-prime repositioning of access, filter, or effluent pressure pods/sensors fails; alarm codes 1-7 depending on which pod(s) failed
Ranked causes
- 1. Pressure Pod/Sensor - Access failed (common)
- 2. Pressure Pod/Sensor - Filter failed (common)
- 3. Pressure Pod/Sensor - Effluent failed (common)
- 4. Pressure difference exceeds 20 mmHg for >5 sec during repositioning (occasional)
- 5. Repositioning timeout exceeded (>2 minutes) (occasional)
Monitor access and ARPS pressure sensor values during repositioning
- Test point
- Pressure sensors
- Expected
- Within ±20 mmHg of each other, ±20 mmHg
Prime self-test failure codes 17-18 caution
Blood leak detector normalization timeout or threshold error during prime test
Ranked causes
- 1. Effluent line incorrectly installed in detector (common)
- 2. Air bubble in effluent line at detector level (occasional)
- 3. Set not fully primed (occasional)
- 4. Liquid or debris in tubing path through detector (occasional)
- 5. Blood leak detector hardware failure (rare)
Verify effluent line installation
- Test point
- Blood leak detector inlet
- Expected
- Line inserted from bottom up through guides
- Tools
- Visual inspection
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replacement pump rate incorrect
Replacement pump flow rate outside normal range
Ranked causes
- 1. Momentary pump roller or segment issue (occasional)
- 2. Pump raceway obstruction or loose rotor screw (occasional)
- 3. Pump hardware failure (rare)
- 4. Rotor thumb screw loosened (occasional)
- 5. Pump failure confirmed (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
TMP access too high patient safety
Access transmembrane pressure has reached user-set pressure limit. High pressure alarm indicates restriction in plasmafiltration circuit.
Ranked causes
- 1. Inadequate anticoagulation of extracorporeal circuit (common)
- 2. Blood flow rate or plasmafiltration rate too high (occasional)
- 3. Kinked lines in blood flowpath (occasional)
- 4. Syringe pump failed or incorrectly installed (occasional)
- 5. Filter or effluent pressure sensor failed (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Access pressure extremely negative patient safety
Alarm indicates access pressure more negative than warning limit or 150 mmHg below operating point; patient moving, coughing, suctioning, or access line issues
Ranked causes
- 1. Patient movement or coughing (common)
- 2. Access line clamped, kinked or partially blocked (common)
- 3. Access catheter clotted, out of position or blood flow too high (common)
- 4. Access pressure sensor failed (occasional)
Inspect access line for clamps, kinks or blockages
- Test point
- Access line
- Expected
- Line unclamped, straight, unobstructed
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calcium Line Clamped patient safety
Calcium infusion line is clamped or obstructed, preventing calcium delivery during citrate-calcium treatment.
Ranked causes
- 1. Calcium line is clamped (common)
- 2. Central venous access on patient is clamped (common)
- 3. Central venous access obstructed by clots or catheter malposition (occasional)
- 4. Incorrect installation of syringe line (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot Detect Return patient safety
Return line disconnection detection disabled when return pressure operating point is more negative than +10 mmHg. Machine unable to detect return line and catheter disconnections.
Ranked causes
- 1. Return line or catheter is disconnected (common)
- 2. Catheter size too large or blood flow too low (occasional)
- 3. Chamber monitor line not securely connected to return pressure port (occasional)
- 4. Return pressure sensor failed (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Return patient safety
Return pressure exceeds operating point by ±50 mmHg (±70 mmHg if blood flow>200 ml/min); self-clears when condition resolves.
Ranked causes
- 1. Possible kink or obstruction in return line or catheter (common)
- 2. Patient is moving (occasional)
- 3. Catheter is clotted or out of position in vein (occasional)
- 4. Blood flow rate is too high (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dialysate pump rate incorrect patient safety
Dialysate (green) pump delivery rate does not match set parameters
Ranked causes
- 1. Momentary problem with pump roller or segment (common)
- 2. Impeding object, clamped or kinked line (common)
- 3. Thumb screw in rotor center loosened (occasional)
- 4. Clamped line in pump path (common)
Press CONTINUE to clear momentary pump issue
- Test point
- Alarm screen
- Expected
- Pump resumes normal rate; alarm clears
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Download Interrupted
Download of history data to technical data card has failed during data transfer.
Ranked causes
- 1. Technical data card is full (common)
- 2. No technical data card present or card is damaged (common)
- 3. Internal malfunction related to technical data card holder/reader (operator level) (rare)
- 4. Internal malfunction related to technical data card holder/reader (service level) (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Effluent bag tinged pink or red patient safety
Effluent collection bag shows pink or red discoloration
Ranked causes
- 1. Patient's disease state - removed free hemoglobin (occasional)
- 2. Effluent contains red blood cells below detection limit (occasional)
- 3. Hemolysis due to occlusion (common)
- 4. Hemolysis during TPE therapy (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filter extremely positive pressure caution
Filter pod pressure exceeds safe limit
Ranked causes
- 1. Line clamped or kinked between filter pressure pod and filter or between filter and deaeration chamber (common)
- 2. Machine operating at high return pressure with filter clotting beginning (occasional)
- 3. Excessive pressure in return line (occasional)
- 4. Filter pressure sensor failed (rare)
Inspect and unclamp lines if clamped or kinked
- Test point
- Filter pressure line path
- Expected
- Lines unclamped and straight
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
HP cartridge clotting patient safety
Hemofilter cartridge showing increasing pressure drop and clotting
Ranked causes
- 1. Inadequate anticoagulation of extracorporeal circuit (common)
- 2. Kinked lines in blood flowpath (occasional)
- 3. Syringe pump anticoagulation system failure (occasional)
- 4. Air leak in deaeration chamber monitor line (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Normalization Failed alarm patient safety
Blood leak detector normalization attempted 3 times and failed; may indicate filter blockage, defective line, air bubble, or detector failure
Ranked causes
- 1. Filter blood leak or effluent line issue (common)
- 2. Air bubble in effluent line (occasional)
- 3. Effluent line not correctly installed (occasional)
- 4. Blood leak detector failed (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
PBP pump rate incorrect caution
Rate of pre-blood pump delivers incorrect flow rate
Ranked causes
- 1. Momentary pump roller or segment problem (occasional)
- 2. Impeding object, clamped or kinked line (common)
- 3. Rotor center thumbscrew loosened (occasional)
- 4. PBP pump hardware failure (rare)
Verify pump operation
- Test point
- PBP pump in operation
- Expected
- Normal flow rate
- Tools
- Visual observation
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Prime Self-Test Code 27 failure
Pressure monitoring (TMPa) test failed
Ranked causes
- 1. Return line misplacement (common)
- 2. Pressure pod missing or sensor debris (occasional)
- 3. Set not fully primed (common)
- 4. Pressure sensor or ARPS failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Prime self-test failure code 16 caution
Return pressure sensor test failed during prime sequence
Ranked causes
- 1. Clamped lines in set (occasional)
- 2. Chamber monitor line not securely connected (common)
- 3. Air in set or bad priming quality (occasional)
- 4. Return pressure sensor hardware failure (rare)
Verify lines unclamped
- Test point
- Set tubing
- Expected
- All lines open and flowing
- Tools
- Visual inspection
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Return disconnection detected patient safety
Return line or catheter disconnected; return pressure lower than +10 mmHg when operating point is higher
Ranked causes
- 1. Return line or catheter disconnected (common)
- 2. Chamber monitor line not properly connected or fluid barrier wet (occasional)
- 3. Blood flowpath obstructed before deaeration chamber (occasional)
- 4. Return pressure sensor failed (rare)
Verify return line and catheter connection
- Test point
- Return line connections
- Expected
- Catheter securely connected to both return line and patient
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scale zero test failed caution
One or more scales failed the zero calibration test
Ranked causes
- 1. Unexpected presence of bag on scale (common)
- 2. Carrying bar missing from scale (occasional)
- 3. Foreign objects on or touching scales (common)
- 4. Scale hardware failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
TMP Excessive patient safety
Access transmembrane pressure or filter transmembrane pressure exceeds safe limit
Ranked causes
- 1. Effluent rate too high (common)
- 2. Plasmafilter pressure drop increasing; insufficient anticoagulation (occasional)
- 3. Ultrafiltration rate (UFR) too high; wrong pressure measurement; inadequate anticoagulation (common)
- 4. During CRRT MARS: MARSFLUX and diaFLUX combined pressure exceeds limit (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
24 V power supply relay failure or return clamp sensor malfunction patient safety
24 V relay fails to disable or re-enable; return clamp does not close/open on command; or clamp position sensor does not detect movement
Ranked causes
- 1. 24 V power relay failure or stuck contacts (occasional)
- 2. Return clamp solenoid or motor drive circuit failure (occasional)
- 3. Return clamp position sensor failure or misalignment (occasional)
Command return clamp to close and verify closed state
- Test point
- Return clamp position sensor output
- Expected
- Clamp closed on command
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Access pressure extremely positive patient safety
Alarm indicates access pressure more positive than warning limit; external device delivering high pressure or blood flow rate set too low
Ranked causes
- 1. External device delivering blood at high pressure (common)
- 2. Blood flow rate set too low for external device pressure (common)
- 3. Access pressure sensor failed (occasional)
Verify external device pressure setting
- Test point
- External device controls
- Expected
- Device pressure within specification for current blood flow
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air bubble detector or pump control failure during simulated air bubble test patient safety
During air bubble security test, control system fails to activate safe state (stop all pumps and close return clamp) or pump stop command does not execute when air bubble is simulated
Ranked causes
- 1. ABD transmitter power insufficient to simulate air bubble condition (occasional)
- 2. Blood pump stop command circuit failure (occasional)
- 3. Control system software or CCA module malfunction (occasional)
Simulate air bubble and verify alarm signal activation
- Test point
- ABD receiver output; air in blood alarm signal
- Expected
- ABD alarm signal activated when transmitter reduced
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air in blood line patient safety
Alarm indicates air or foam detected in blood tubing or return line; patient disconnected, set not fully primed, or air trapped in deaeration chamber
Ranked causes
- 1. Disconnected line or leaking connection (common)
- 2. Set not fully primed (common)
- 3. Air or foam in tubing (occasional)
Visually inspect all blood line connections
- Test point
- Patient access, return line, set connections
- Expected
- All connections firmly seated and dry
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ARPS pump power relay failure or pump malfunction patient safety
12 V relay powering Automatic Repositioning System motor disabled or ARPS pump fails to start/stop; no pressure change detected during pump operation
Ranked causes
- 1. 12 V relay failure or open circuit (occasional)
- 2. ARPS pump mechanical failure or blockage (occasional)
- 3. ARPS tubing leak or pressure sensor malfunction (common)
Verify 12 V relay disable function
- Test point
- 12 V relay control signal
- Expected
- Relay disables on command
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bag volume incorrect patient safety
Variable Empty Bag method active; fluid amount in bag does not match Allowed Volume setting; alarm triggers
Ranked causes
- 1. Loaded solution bag volume does not match Allowed Volume setting (common)
- 2. No bag on scale or foreign object present (occasional)
- 3. Bag partially supported or not hanging freely (occasional)
Verify bag volume matches Allowed Volume
- Test point
- Solution bag and scale weight reading
- Expected
- Bag volume matches or exceeds Allowed Volume
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pump rate incorrect patient safety
Blood pump delivery rate does not match set parameters during operation
Ranked causes
- 1. Momentary problem with pump roller or segment (common)
- 2. Impeding object, clamped or kinked line in pump raceway (common)
- 3. Thumb screw in rotor center has loosened (occasional)
Press CONTINUE to clear momentary pump issue
- Test point
- Alarm screen
- Expected
- Alarm clears and pump resumes normal rate
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication error
Internal communication link error between protective task, protective slave, or control system modules
Ranked causes
- 1. Software download error or I2C bus issue (occasional)
- 2. Component connection failure or voltage loss (occasional)
- 3. Component hardware failure (rare)
Power cycle and restart machine
- Test point
- Power switch and startup sequence
- Expected
- Machine restarts without communication error
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Effluent bag incorrect patient safety
Effluent bag volume capacity does not match Allowed Volume setting; e.g. 5000 ml bag hung when 9000 ml expected
Ranked causes
- 1. Wrong effluent bag size hung on scale versus Allowed Volume (common)
- 2. No bag on scale (occasional)
- 3. Effluent bag partially supported or not hanging freely (occasional)
Verify effluent bag size matches Allowed Volume
- Test point
- Effluent bag capacity, Allowed Volume setting
- Expected
- Bag capacity matches Allowed Volume
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Effluent bag volume mismatch patient safety
Bag size on scale does not match configured allowed volume during priming
Ranked causes
- 1. Bag size smaller than allowed volume setting (common)
- 2. No bag on scale (common)
- 3. Effluent bag partially supported (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Effluent line not in blood leak detector patient safety
Effluent line of new set not properly seated in blood leak detector during priming
Ranked causes
- 1. Effluent line not installed in detector (common)
- 2. Blood leak detector failed (occasional)
- 3. Line in detector not empty (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Effluent pump rate incorrect patient safety
Effluent (yellow) pump delivery rate does not match set parameters
Ranked causes
- 1. Momentary problem with pump roller or segment (common)
- 2. Impeding object, clamped or kinked line (common)
- 3. Thumb screw in rotor center loosened (occasional)
Press CONTINUE to clear momentary pump issue
- Test point
- Alarm screen
- Expected
- Pump resumes normal rate; alarm clears
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lines clamped or occluded patient safety
One or more fluid lines blocked, restricting flow and causing pressure alarm
Ranked causes
- 1. One of the lines is clamped (common)
- 2. Occluded disposable set (common)
- 3. One or more pressure sensors failed (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lines crossed or tangled patient safety
Patient lines or bag lines are improperly routed, causing blockage or scale misreading
Ranked causes
- 1. Lines are crossed or tangled (common)
- 2. Foreign object on scale (occasional)
- 3. One or more scales failed (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Plasmafilter clotted patient safety
Filter pressure drop exceeds limit for plasmafilter; clots formed or excessive TMP reached
Ranked causes
- 1. Inadequate anticoagulation of blood flowpath (common)
- 2. Clamped lines in blood flowpath (occasional)
- 3. Ultrafiltration rate too high for filter (occasional)
Inspect blood flowpath for clamped lines
- Test point
- Entire blood flowpath
- Expected
- All lines unclamped
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure pod repositioning failure or sensor verification out of range patient safety
Effluent, Access, or Filter pressure pods fail to reposition; sensor verification shows pressure outside ±20 mmHg tolerance during repositioning sequence; repositioning sequence takes too long (>5 seconds deviation) or falls outside final tolerance after completion
Ranked causes
- 1. ARPS pump unable to achieve target pressure for pod repositioning (occasional)
- 2. Pressure sensor offset or calibration drift (common)
- 3. Pod membrane rupture or air leak in pressure line (occasional)
Measure initial Effluent pressure and store value
- Test point
- Effluent pressure sensor
- Expected
- Pressure value recorded within normal operating range
12 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Prime self-test failure (codes 1-7) caution
Prime self-test failed due to pressure pod or sensor issues affecting Access, Filter, or Effluent
Ranked causes
- 1. Pressure pod not installed (common)
- 2. Clamped lines in set (occasional)
- 3. Pressure sensor hardware failure (rare)
Verify pod installation
- Test point
- Access/Filter/Effluent sensor pods
- Expected
- All pods correctly installed
- Tools
- Visual inspection
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump occlusion test failure or external leakage in disposable set patient safety
Peristaltic pump rollers cannot completely occlude tubing within pump raceway, or excessive pressure drop indicates leak in disposable set; number of rotations exceeds limit or pressure drop exceeds allowed tolerance
Ranked causes
- 1. Disposable set damaged, kinked, or improperly seated in pump raceway (common)
- 2. Peristaltic pump roller wear or mechanical failure (occasional)
- 3. Pressure sensor calibration drift or sensor malfunction (occasional)
Verify blood pump can build return pressure to set-dependent threshold
- Test point
- Return pressure sensor; blood pump encoder rotation count
- Expected
- Pressure reaches target without exceeding maximum rotation count for selected set
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replacement Container Empty patient safety
Replacement solution container is empty, partially supported, or has fallen
Ranked causes
- 1. Replacement container empty (common)
- 2. Replacement container partially supported (occasional)
- 3. Replacement container has fallen (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replacement pump 2 rate incorrect patient safety
Rate of replacement pump 2 (green) pump is not at the correct setting
Ranked causes
- 1. Momentary pump roller or segment issue (occasional)
- 2. Impeding object or line obstruction in pump (common)
- 3. Pump failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Return pressure sensor out of range or ARPS-to-return pressure mismatch patient safety
Return pressure below 250 mmHg after occlusion test, or return and ARPS pressure sensors differ by more than 20 mmHg after ARPS pump equalization
Ranked causes
- 1. Return pressure sensor reading low or misaligned (occasional)
- 2. ARPS pump unable to reach target pressure or has regulator leak (occasional)
- 3. ARPS return valve stuck or leaking (occasional)
Verify return pressure above minimum threshold after occlusion test
- Test point
- Return pressure sensor
- Expected
- Greater than 250 mmHg, >250
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self-test overdue caution
Periodic self-test failed to complete within last 150 minutes of treatment
Ranked causes
- 1. Self-test interrupted by secondary alarms (common)
- 2. Self-test interrupted by operator interventions (common)
- 3. Self-test repeatedly overridden by operator (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Syringe line clamped
Syringe line blocked by clamp, kink, or obstruction; anticoagulant flow interrupted
Ranked causes
- 1. Syringe line clamped, kinked, or obstructed (common)
- 2. Incorrect installation of syringe line (occasional)
- 3. Alarm is recurring (occasional)
Inspect syringe line for clamps and kinks
- Test point
- Syringe line full length
- Expected
- No clamps, kinks, or obstructions visible
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
24V and Return Clamp Test failure patient safety
Periodic self-test failure; 24V relay, +24V supply, or return clamp sensor malfunction; alarm codes 24, 25, or 26
Ranked causes
- 1. 24V relay malfunction or +24V supply lost (occasional)
- 2. Return clamp sensor malfunction or did not detect position change (occasional)
Close return clamp
- Test point
- Return clamp control
- Expected
- Clamp closed
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Air detector malfunction patient safety
Air bubble detector failed self-tests or return line not installed/improperly installed. System detects air detector circuit issue.
Ranked causes
- 1. Air bubble detector failed self-tests (rare)
- 2. Return line not installed or improperly installed in air bubble detector (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bag empty
Indicated solution or effluent bag is empty; alarm alerts operator to change bag; system prevents flow
Ranked causes
- 1. Bag volume depleted during treatment (common)
- 2. Bag partially supported or has fallen (occasional)
Verify bag is empty per weight on scale
- Test point
- Scale reading for indicated bag
- Expected
- Weight at or below tare weight
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery depleted during power loss patient safety
Main power is lost and backup battery has no energy; machine cannot operate
Ranked causes
- 1. Main power lost and battery exhausted (occasional)
- 2. Machine unplugged with battery discharged (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Exhausted caution
Back-up battery power level too low; machine needs charging to restore battery capacity
Ranked causes
- 1. Back-up battery depleted (common)
- 2. Old battery or broken internal wiring (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood leak detected in MARS treatment patient safety
MARS monitor detected blood leak during CRRT MARS treatment
Ranked causes
- 1. Actual blood leak in circuit (occasional)
- 2. MARS monitor false alarm (occasional)
Confirm blood leak presence
- Test point
- Circuit visual inspection
- Expected
- Blood leak confirmed or not confirmed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Blood leak detector failed
- Blood leak detector normalization timeout or out-of-range values
- Blood leak detector threshold detection failure
- Blood leakage from pressure pod diaphragm
- Blood pump failed
- Calcium infusion line clamped
- Calcium syringe not loaded
- Cannot save custom data
- Dialysate pump failed
- Dialysate scale sensor failed
- Display blank or logo screen stuck
- Effluent bag full
- Effluent bag full during priming
- Effluent pump noise or poor function
- Effluent Scale Sensor alarm during priming
- HP cartridge clotted
- Line in Air Detector alarm
- Line in Clamp alarm
- Line presence sensor malfunction in ABD or return clamp
- Lower Pinch Valve alarm
- Memory backup battery depleted
- No Line in Air Detector alarm
- No Line in Clamp alarm
- PBP scale sensor failed
- Pressure zero test failed
- Prime Self-Test Code 19 failure
- Prime Self-Test Code 21-23 failure
- Prime Self-Test Code 25 failure
- Replacement scale sensor failure
- Return extremely positive pressure
- Return line clamp stuck closed
- Return pressure sensor malfunction or miscalibration
- Return pressure too high
- Scale Open
- Scale open alarm during priming
- Syringe empty
- Syringe not loaded
- Syringe pump malfunction
- Unsuitable calcium solution
- Upper pinch valve malfunction
- Voltage out of range
- Air bubble not detected or false macro-bubble detection
- Alarm indicator light not illuminating or incorrect color/pattern
- Alarm tone not sounding or incorrect tone volume/pattern
- Anticoagulation Check Points
- Anticoagulation delivery failed - syringe pump
- Anticoagulation suspended
- ARPS repositioning valve not opening or pump not tracking pressure
- Auto blood return malfunction
- Bar code reading failure
- Battery alarm recurring
- Blood Flow Stopped
- Blood leak detector malfunction
- Blood Leak Detector Test failure
- Blood pump brake not engaging or disengaging
- Blood pump relay not responding
- Ca Syringe Almost Empty
- Ca Syringe Not Loaded
- Calcium syringe empty
- Calibration: Air Detector
- Calibration: Clock and Date
- Calibration: Display
- Calibration: External Communication Interface
- Calibration: Filter Clotting Limits
- Calibration: Pitch and Volume
- Calibration: Pressure Sensors
- Calibration: Preventive Maintenance
- Calibration: Scales
- Calibration: Screen Brightness
- Calibration: Syringe Pump
- Change of syringe clip
- Checksum interrupted
- Citrate delivery failed - PBP pump
- Custom data access error
- Display goes blank momentarily then reappears
- Display goes blank with buzzer
- Display image floats around
- Effluent pump failed
- Filter measurement failure
- Fluid pumps stopped
- Functional Test: Expected Warning Alarms During Water-Based Test
- Gain Limit Reached
- General System Failure
- General System Failure Code 1
- General System Failure Code 2
- General System Failure Code 3
- General System Failure Code 4
- General System Failure Code 5
- General System Failure Code 6
- Incomplete bolus
- Installation test
- Leakage from set connections
- Library Data error
- Loader already in loaded position
- Loading error - pinch valves
- Loss Limit Reached
- Main power lost
- MARS treatment not started
- Memory Error Code 1
- Memory Error Code 3
- Memory Error Code 4
- Memory Error Code 5
- Memory Error Code 6
- Memory Error Code 7
- Patient Fluid Gain Excessive
- Power failure recovery
- Pressure sensor reading outside accepted range
- Pressures circuit board hardware failure
- Prime Self-Test Code 24 failure
- Prime Self-Test Code 26 failure
- Prime Self-Test Code 28 failure
- Pump does not respond to 24 VDC control
- Pump fails to start or stalls at high speed
- Pump speed does not match set value
- Recirculation time exceeded
- Return Pressure Sensor Test failure
- Scale circuit board hardware failure
- Scale failed
- Scale out of calibration
- Scale out of tolerance error during calibration
- Scale reading outside accepted range
- Set-up error alarm during pre-prime self-test
- Softkeys non-responsive
- Software update
- Syringe almost empty
- Syringe pump ADC value out of tolerance
- Syringe pump force or position measurement out of range
- Time to change set
- TMPa Calibration failure
- Total Plasma Exchange (TPE) Prescription Delivered
- Unresolved Flow Problems
- Wrong set loaded
132 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1 | Blood pump unexpected movement during Safe state | Blood pump motion detected when machine in Safe state | error |
102 | Filter is Clotting | Filter pressure rise detected above threshold | warning |
105 | TMP Too High | Transmembrane pressure exceeds user limit | warning |
114 | Battery Exhausted | System battery voltage falls below safe operating level | error |
115 | Main Power Lost | AC mains supply interrupted | error |
18 | Prime Self Test malfunction | Detected during blood leak detector test with empty tube in holder | error |
19 | Air/pumps security test failure | Air bubble or pump test malfunction during prime | error |
2 | Fluid pump unexpected movement during Safe state | Fluid pump motion detected when machine in Safe state | error |
20 | Pump occlusivity test failure | Return line obstruction or pump pressure anomaly | error |
21-23 | Pinch valve failure | Pinch valve actuator malfunction during prime | error |
24 | 24V/12V power supply test failure | Voltage regulation or ARPS circuit malfunction | error |
25 | Return clamp sensor failure | Return clamp sensor malfunction or obstruction | error |
26 | 24V and return clamp sensor failure | Simultaneous voltage and clamp sensor failures | critical |
27 | Pressure monitoring (TMPa) failure | Pressure sensor or monitoring system malfunction | error |
28 | Syringe pump hardware test failure | Syringe pump test not completed within 600 seconds | error |
3 | Return clamp not properly closed | Return clamp position check fails during initialization | error |
4 | Hall sensor mismatch in Protective Slave | Criteria_Counter vs Criteria_Exhausted values mismatched | error |
5 | I2C communication error during startup | I2C bus communication fails during initialization | error |
6 | Syringe pump unexpected movement | Syringe pump motion detected during Safe state | error |
90 | Check Access | Access line pressure or condition detected as abnormal | warning |
91 | Check Return | Return line pressure detected as abnormal | warning |
92 | Blood Flow Stopped | Blood pump or circuit flow has ceased | error |
93 | Syringe not loaded / Ca Syringe Not Loaded | Syringe detection mechanism indicates syringe missing | warning |
96 | Syringe Empty | Syringe volume sensor detects empty condition | warning |
Air Detector | Air bubble detector failed self-tests or line improperly installed | Detector self-test fails or return line not detected in position | error |
Air in Blood | Air bubbles detected in return line | Air bubbles present in patient blood return line | warning |
Auto Blood Return | Blood return volume incongruence detected | Discrepancy in auto blood return during treatment | warning |
Battery Low | Battery power is low | Battery charge level drops below operating threshold | warning |
Blood Leak Detector | Effluent line not installed in detector or detector failed self-tests | Effluent line missing/improperly installed or detector circuit failure | warning |
Blood leak detector failed | Blood leak detector malfunction detected; unable to continue treatment safely | Blood leak detector hardware fails or cannot communicate with system | error |
Blood Pump – Rate incorrect | Blood pump not delivering correct flow rate during treatment | Pump output does not match programmed rate; flowmeter detects deviation | warning |
Cannot Save Custom Data | System unable to store customized therapy parameter settings | Write error to Compact Flash Card during custom mode save operation | warning |
Ca Syringe Not Loaded | Calcium syringe not detected in loading mechanism at start of priming | Syringe detection sensor does not register loaded syringe during pre-treatment checks | error |
Checksum Interrupted | Internal firmware checksum update was interrupted; system state and settings may be corrupted | Power loss or system reset occurs during checksum verification or update process | warning |
Clamp Stuck Closed – Return line | Return line clamp cannot open; patient blood cannot be returned | Clamp motor stalled or mechanical jam prevents opening during treatment transition | critical |
Communication Error – Code 2 | No communication with protective task microcontroller | Protective task processor not responding to status queries from main control system | critical |
Communication Error – Code 3 | Communication link error on protective slave module | Protective slave processor detects parity/CRC error or timeout in data link | critical |
Communication Error – Code 4 | Communication link error on control system processor | Main control processor detects data link error with peripheral modules | critical |
Communication Error – Code 5 | Missing status command response from protective slave | Protective slave does not respond to periodic status query from main controller | critical |
Communication Error – Code 6 | Missing alarm command response from protective slave | Protective slave cannot send alarm status back to main controller | critical |
Communication Error – Code 7 | Protective task unable to send message to slave module | Master processor cannot deliver commands to protective slave over I2C bus | critical |
Custom Data – Access error | System unable to access stored custom therapy values for selected configuration | Flash card read error when loading custom mode data; data file missing or corrupted | error |
Dialysate Pump – Pump failed | Dialysate pump unable to operate; hardware failure detected | Pump function diagnostic fails; pump unable to achieve minimum dialysate output | critical |
Dialysate Pump – Rate incorrect | Dialysate (green) pump not delivering correct flow rate | Pump output does not match programmed dialysate flow rate; flowmeter detects deviation | warning |
Dialysate Scale Sensor – Failed | Scale position sensor unable to detect dialysate bag tray open/closed state during priming | Sensor does not register scale tray position change; unable to verify bag attachment | error |
Effluent Pump – Pump failed | Effluent pump unable to operate; hardware failure detected | Pump function diagnostic fails; pump unable to achieve minimum effluent output | critical |
Effluent Pump – Rate incorrect | Effluent (yellow) pump not delivering correct drain flow rate | Pump output does not match programmed effluent flow rate; flowmeter detects deviation | warning |
General System Failure | Patient safety cannot be monitored due to system failure | Failure during self-tests, software errors, or hardware failure | critical |
MALFUNCTION: No Line in Air Detector | Line presence sensor in Air Bubble Detector failed or line not loaded | Micro switch in ABD does not detect line after set loading during post-prime | error |
MALFUNCTION: No Line in Clamp | Line presence sensor in return line clamp failed or line not loaded | Micro switch in return line clamp does not detect line after set loading during post-prime | error |
MALFUNCTION:Prime Self-Test Code: 1 | Prime Self-Test failed due to Access pressure pod/sensor | Access pod repositioning failed during post-prime | error |
MALFUNCTION: Prime Self-Test Code 16 | Return pressure sensor out of range or ARPS-return pressure mismatch exceeds tolerance | Return pressure below 250 mmHg after occlusion; or ARPS and return pressure differ >±20 mmHg after equalization | error |
MALFUNCTION: Prime Self-Test Code 17 | Blood Leak Detector normalization timeout or normalization value out of range | BLD normalization does not complete within 30 seconds or receiver/transmitter PWM values outside spec | error |
MALFUNCTION: Prime Self-Test Code 18 | Blood Leak Detector threshold detection failure during leak simulation test | BLD receiver value does not drop below 15000 within 5 seconds when transmitter PWM lowered to simulate leak | error |
MALFUNCTION: Prime Self-Test Code 19 | Air/Pump security test failure during simulated air bubble condition | Blood pump fails to stop on air bubble alarm or fluid pumps fail to respond to start/stop commands | critical |
MALFUNCTION:Prime Self-Test Code: 2 | Prime Self-Test failed due to Filter pressure pod/sensor | Filter pod repositioning failed during post-prime | error |
MALFUNCTION: Prime Self-Test Code 20 | Pump occlusion test failure or external leakage in disposable set | Blood pump unable to reach target return pressure in allowed rotations; excessive pressure drop after 10 seconds | error |
MALFUNCTION: Prime Self-Test Code 24 | 12 V or 24 V power relay failure or ARPS pump malfunction | 12 V relay fails to disable/enable or ARPS pump does not respond; OR 24 V relay fails to disable/enable | error |
MALFUNCTION: Prime Self-Test Code 25 | Return clamp position sensor failure or malfunction | Return clamp does not move on command or position sensor does not detect open/close state | error |
MALFUNCTION: Prime Self-Test Code 26 | 24 V power and return clamp sensor combined failure | 24 V relay malfunction AND return clamp position sensor failure detected together | critical |
MALFUNCTION:Prime Self-Test Code: 27 | TMPa calibration failed | Pressure sensor characteristics out of 20 percent tolerance during TPE calibration | error |
MALFUNCTION:Prime Self-Test Code: 3 | Prime Self-Test failed due to Access and Filter pods/sensors | Both access and filter pod repositioning failed during post-prime | error |
MALFUNCTION:Prime Self-Test Code: 4 | Prime Self-Test failed due to Effluent pressure pod/sensor | Effluent pod repositioning failed during post-prime | error |
MALFUNCTION:Prime Self-Test Code: 5 | Prime Self-Test failed due to Access and Effluent pods/sensors | Both access and effluent pod repositioning failed during post-prime | error |
MALFUNCTION:Prime Self-Test Code: 6 | Prime Self-Test failed due to Filter and Effluent pods/sensors | Both filter and effluent pod repositioning failed during post-prime | error |
MALFUNCTION:Prime Self-Test Code: 7 | Prime Self-Test failed due to Access, Filter, and Effluent pods/sensors | All three pod repositioning failed during post-prime | critical |
MALFUNCTION:Self-Test Failure Code: 16 | Return pressure sensor failure | Return and ARPS pressure readings differ by more than ±20 mmHg | error |
MALFUNCTION:Self-Test Failure Code: 18 | Blood leak detector threshold error | BLD receiver value does not fall below 15000 within 5 seconds of transmitter PWM reduction | error |
MALFUNCTION:Self-Test Failure Code: 24 | 24V / 12V supply failure | +24V relay malfunction or supply does not restore after being disabled | error |
MALFUNCTION:Self-Test Failure Code: 25 | Return clamp sensor failure | Return clamp position sensor does not detect open or closed position correctly | error |
MALFUNCTION:Self-Test Failure Code: 26 | 24V and Return clamp sensor failure | Both +24V relay and return clamp position sensor failures detected | critical |
Memory Error 1 | Memory error on Protective task | Protective task memory corruption detected | critical |
Memory Error 3 | Wrong CRC of a set value | Checksum validation fails for set value | critical |
Memory Error 4 | Set value incongruence between Protective slave and task | Protective and control task parameters don't match | critical |
Memory Error 5 | Incongruence on alarm structure of control system | Alarm structure validation fails at startup | critical |
Memory Error 6 | Set value incongruence between protective and control | Parameter mismatch between protective and control systems | critical |
Memory Error 7 | Backup memory failure | Backup memory read/write error | critical |
Power Failure | Mains or backup power lost to control unit | Loss of electrical power to Prismaflex control unit | critical |
Pump failed – Blood | Blood pump unable to operate; hardware failure detected | Pump function diagnostic fails; pump unable to achieve minimum output | critical |
Syringe Empty | Syringe pump is in end-of-travel position indicating empty syringe | Syringe depleted during treatment or not loaded | warning |
Syringe Line Clamped | Syringe delivery line is obstructed or kinked | Line clamped, kinked, or obstructed during treatment | warning |
Syringe Not Loaded | Syringe not loaded after syringe test performed | Syringe test fails to detect loaded syringe; missing or loose installation | error |
Time to Change Set | Hours of use have reached operator-set time limit for therapy/set | Operating time exceeds set limit for therapy/set combination | info |
TMP Access Too High | Access transmembrane pressure exceeded user-set limit | Pressure in plasmafiltration circuit exceeds threshold | error |
TMP Too High | Main filter transmembrane pressure exceeded user-set limit | Pressure in filter circuit exceeds threshold during CRRT | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community