Valleylab Force FX: troubleshooting and error codes
The ServiceForge knowledge base holds 40 documented symptoms for the Valleylab Force FX, with 92 ranked causes, 36 diagnostic checks with test points and expected values, and 110 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Generator on and accessory activated but no output delivered high voltage
Electrosurgical output not generated despite power on and activation signal
Ranked causes
- 1. Malfunctioning footswitch or handswitching instrument (common)
- 2. Incompatible footswitch (occasional)
- 3. Footswitch connected to wrong receptacle for instrument connection (occasional)
- 4. Footswitch connected to Monopolar 2 for Monopolar 1 instrument (occasional)
- 5. Power set too low (common)
- 6. Alarm condition exists (occasional)
- 7. Blown fuse on Power Supply/RF board (occasional)
- 8. Control board malfunction (occasional)
- 9. High voltage power supply malfunction (occasional)
- 10. RF output stage malfunction (occasional)
Verify accessory connections
- Test point
- Footswitch and instrument receptacles
- Expected
- Secure connections to appropriate receptacles
1 more check with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator does not respond when turned on high voltage
Device fails to power up or initialize when power switch is activated
Ranked causes
- 1. Disconnected power cord, faulty wall receptacle, or faulty power cord (common)
- 2. Fuse drawer open or fuses blown (common)
- 3. Loose or disconnected internal cables (occasional)
- 4. Faulty power entry module or connections (occasional)
- 5. Faulty low voltage power supply (occasional)
- 6. Damaged Control board connectors and/or malfunctioning Control board (occasional)
- 7. Shorts or disconnects on Power Supply/RF board (occasional)
- 8. Faulty power switch (rare)
- 9. Malfunctioning front panel components (rare)
Verify power cord connections
- Test point
- Generator and wall receptacle
- Expected
- Secure connections
1 more check with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator is on but did not complete self-test
Device powers on but fails to pass initialization self-test routine
Ranked causes
- 1. Alarm condition exists (common)
- 2. Software malfunction (occasional)
- 3. Loose or disconnected internal cables (occasional)
- 4. Faulty low voltage power supply (occasional)
- 5. Damaged Control board connectors and/or malfunctioning Control board (occasional)
- 6. Shorts or disconnects on Power Supply/RF board (occasional)
- 7. Faulty power switch (rare)
- 8. Malfunctioning front panel components (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator malfunction high voltage
Generator stops functioning or fails self-test; visible physical damage, loose cords, or damaged casing suspected
Ranked causes
- 1. Visible physical damage to generator housing (occasional)
- 2. Accessory cords not properly connected (common)
- 3. Power cord damage or exposed wires (occasional)
- 4. Fuses damaged or loosely seated (common)
Inspect generator for visible physical damage
- Test point
- Entire chassis exterior
- Expected
- No cracks, dents, burn marks, or visible damage
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Interference with other devices when generator is activated patient safety
Other equipment experiences electromagnetic interference only during electrosurgical activation
Ranked causes
- 1. Metal-to-metal sparking (common)
- 2. High settings used for fulguration (common)
- 3. Electrically inconsistent ground wires in operating room (occasional)
- 4. Monitor responding to radiated frequencies (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal component damage on circuit boards high voltage
Visible damage, cracks, corrosion, or broken wires detected on internal circuit boards after cover removal
Ranked causes
- 1. Damaged Control board (rare)
- 2. Damaged Display board (rare)
- 3. Damaged Footswitch board (internal) (rare)
- 4. Severe damage on Power Supply/RF board (rare)
Remove cover and inspect all boards for damage
- Test point
- All internal circuit boards (Control, Display, Footswitch, Power Supply/RF)
- Expected
- No damaged components, no broken wires, no cracks, no corrosion visible on any board
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mode buttons do not function correctly when pressed
Front panel mode buttons fail to change operating modes or show no response
Ranked causes
- 1. Faulty ribbon cable between Control board and Display board (common)
- 2. Incorrect modes communicated through Control board (occasional)
- 3. Faulty ribbon cable between front panel and Display board (occasional)
- 4. Incorrect modes communicated from front panel (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal neuromuscular stimulation during surgery patient safety
Patient displays involuntary muscle contractions or abnormal stimulation during procedure; requires immediate cessation of surgery
Ranked causes
- 1. Metal-to-metal sparking (common)
- 2. High power setting during coagulation (common)
- 3. Abnormal 50-60 Hz leakage currents (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Activation and alarm tones do not sound
Speaker produces no audio output for activation tones or alarm alerts
Ranked causes
- 1. Poor connection or damaged Footswitch board ribbon cable (common)
- 2. Faulty connections or speaker on Footswitch board (occasional)
- 3. Audio signal malfunction on Control board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm 212 appears in Cut display during self-test
Specific error code 212 displayed during power-on self-test sequence
Ranked causes
- 1. Battery removed or replaced without generator calibration (common)
- 2. Faulty battery on Control board causing loss of battery-backed memory (occasional)
- 3. Loss of battery-backed RAM due to faulty component on Control board (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blank or confusing LED display
Display board shows garbled, missing, or incorrect information
Ranked causes
- 1. Faulty ribbon cable between Control board and Display board (common)
- 2. Incorrect display modes communicated through Control board (occasional)
- 3. Display board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Continuous monitor interference patient safety
Cardiac monitor displays noise or artifacts continuously regardless of generator activation
Ranked causes
- 1. Faulty chassis-to-ground connections (common)
- 2. Electrical equipment grounded to different objects (occasional)
- 3. Malfunctioning monitor (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
T_ON ASIC malfunction alarm high voltage
Generator displays an alarm number related to T_ON ASIC malfunction during or after self-test
Ranked causes
- 1. Programmable IC U9 pins not properly seated in socket (common)
- 2. Programmable IC U9 defective or failed (occasional)
- 3. Control board defective or failed (rare)
Verify IC U9 pins seated properly in socket on Control board
- Test point
- IC U9 socket
- Expected
- All pins fully inserted and aligned
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery-backed RAM malfunction indicated high voltage
Alarm number appears, indicating memory corruption or loss in battery-backed RAM; requires battery replacement and recalibration
Ranked causes
- 1. Battery depleted or failed (common)
- 2. Control board defective (occasional)
Turn off generator and disconnect power cord
- Test point
- Main power connector
- Expected
- Power disconnected, wait 5 minutes minimum
10 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Footswitch will not activate output
Footswitch fails to trigger electrosurgical output generation
Ranked causes
- 1. Malfunctioning or damaged footswitch receptacle (common)
- 2. Footswitch activation signal lost on Control board (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacemaker interference patient safety
Pacemaker malfunction or inappropriate sensing triggered by electrosurgery
Ranked causes
- 1. Intermittent connections or metal-to-metal sparking (common)
- 2. Current path too close to pacemaker during monopolar electrosurgery (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
REM indicator illuminates red patient safety
REM Polyhesive Patient Return Electrode not properly applied or not connected to generator
Ranked causes
- 1. REM Patient Return Electrode not properly applied to patient (common)
- 2. REM Patient Return Electrode not connected to generator (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration procedure notice: Minimum load resistor thresholds caution
Equipment damage will result if generator activated with load resistor below specified threshold during calibration
Ranked causes
- 1. Exceeding maximum output power into low impedance load during calibration (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration value out of acceptable range
Alarm 120 displayed; calibration value exceeds acceptable limits
Ranked causes
- 1. Calibration procedure failed or sensor malfunction (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cover removal and reinstallation high voltage
Procedure for accessing internal generator components by removing and reinstalling the generator cover
Ranked causes
- 1. Access required for internal component inspection (common)
Cover removal
- Test point
- Cover fastening screws
- Expected
- All five screws removed and cover lifts clear of chassis
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display board seven-segment LED failure
Seven-segment LED on display board is non-functional, dim, or discolored
Ranked causes
- 1. LED worn out or defective (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dosage error during diagnostics caution
Alarm 160 displayed; internal dosage test failed with ADC values shown
Ranked causes
- 1. ECON factor out of calibration or ADC malfunction (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan replacement caution
Fan requires replacement on Power Supply/RF board assembly
Ranked causes
- 1. Fan failure or wear (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flash memory CRC test failed caution
Alarm 183-185 displayed; Flash memory CRC test failed in main or feedback microcontroller
Ranked causes
- 1. Flash memory corruption or microcontroller failure (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fuse blown in fuse drawer caution
Generator loses power or protection circuit trips; fuses in the drawer below power cord receptacle are open or damaged.
Ranked causes
- 1. Blown fuse in fuse drawer (120 V or 240 V operation) (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fuse blown on Power Supply/RF board high voltage
Generator fails to operate; fuse on internal Power Supply/RF board is blown or open.
Ranked causes
- 1. Blown fuse on Power Supply/RF board (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator fails internal diagnostics at startup caution
Alarm 4 displayed; main microcontroller CPU test failed during power-up
Ranked causes
- 1. Control board CPU malfunction (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loose bipolar instrument receptacle connection caution
Bipolar instrument connector fits loosely into front panel receptacle; instrument operation may be unreliable
Ranked causes
- 1. Worn or damaged bipolar instrument receptacle (occasional)
Test bipolar instrument connector fit in front panel receptacle
- Test point
- Bipolar Instrument receptacle on front panel
- Expected
- Footswitching and handswitching instrument connectors insert securely with no movement
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loose footswitch receptacle connection caution
Footswitch connector fits loosely into rear panel receptacles; footswitching may be unreliable
Ranked causes
- 1. Worn or damaged footswitch receptacle contacts (occasional)
Test bipolar footswitch receptacle fit
- Test point
- Three footswitch receptacles on rear panel
- Expected
- Bipolar footswitch connector inserts securely with no movement or looseness
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loose monopolar instrument receptacle connection caution
Monopolar instrument connector fits loosely into front panel receptacles; instrument operation may be unreliable
Ranked causes
- 1. Worn or damaged monopolar instrument receptacle (occasional)
Test monopolar 1/CEM connector fit in front panel receptacle
- Test point
- Monopolar 1/CEM Instrument receptacle on front panel
- Expected
- Footswitching and handswitching instrument connectors insert securely with no movement
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Microcontroller RAM check failed caution
Alarm 180/181 displayed; internal RAM diagnostic failed on main or feedback microcontroller
Ranked causes
- 1. RAM memory malfunction in microcontroller (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Optoisolator internal component failed caution
Alarm 81 displayed; optoisolator component test failed
Ranked causes
- 1. Optoisolator component failure (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Return Electrode receptacle damaged patient safety
Receptacle shows visible damage, broken pin, or obstruction; patient return connection compromised
Ranked causes
- 1. Damaged or obstructed Patient Return Electrode receptacle (rare)
Inspect Patient Return Electrode receptacle
- Test point
- Patient Return Electrode receptacle on front panel
- Expected
- No broken pins, no obstructions, clear unobstructed opening
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
REM alarm indicator does not illuminate patient safety
No illumination (red or green) from REM alarm indicator light
Ranked causes
- 1. Return electrode without REM safety feature connected (common)
Verify REM Polyhesive Patient Return Electrode is connected
- Test point
- Return electrode type
- Expected
- REM Polyhesive Patient Return Electrode installed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
REM alarm indicator flashing red patient safety
REM Polyhesive Patient Return Electrode alarm condition detected; indicator flashes red until corrected
Ranked causes
- 1. Patient return electrode not properly applied or connected (common)
Verify REM Patient Return Electrode is properly applied to patient
- Test point
- Patient contact
- Expected
- Electrode making full contact with patient skin
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
REM alarm indicator illuminates red (steady) patient safety
Indicator shows red light (steady, not flashing) when REM Patient Return Electrode first connected or system alarm present
Ranked causes
- 1. REM Patient Return Electrode not yet applied to patient (common)
Confirm REM Patient Return Electrode applied and connected
- Test point
- Electrode connection
- Expected
- Electrode properly applied to patient and connected to generator
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
REM indicator does not illuminate patient safety
Return electrode connected without REM safety feature
Ranked causes
- 1. Return electrode without REM safety feature connected (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
REM indicator flashes red patient safety
REM system has detected an alarm condition that must be corrected
Ranked causes
- 1. REM system alarm condition detected (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
REM sense circuit open or failed
Alarm 122 displayed; open circuit detected in REM sense line during calibration
Ranked causes
- 1. Open circuit in REM sense line (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
T_ON test fails during diagnostics caution
Alarm 12 or 16 displayed; T_ON timing or average test failed
Ranked causes
- 1. T_ON ASIC malfunction or calibration error (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0 | Main microcontroller failed to hold feedback microcontroller in reset | At startup | error |
1 | Main microcontroller failed to power up feedback microcontroller | At startup | error |
10 | Software malfunction - invalid alarm number on serial port | During operation | error |
100 | Software malfunction - rem_update_state outside range | During operation | error |
101 | Software malfunction - rem_pad_state outside range | During operation | error |
102 | Software malfunction - rem_state in re_rem_control outside range | During operation | error |
103 | Software malfunction - rem_state in re_non_rem_control outside range | During operation | error |
104 | Software malfunction - rem_flash_state outside range | During operation | error |
105 | Software malfunction - rem_led_state outside range | During operation | error |
11 | Invalid activated power, function, mode, or key request mode echoed | At startup or during operation | error |
110 | Software malfunction - cal_state outside range in ca_generator_setup | During calibration | error |
111 | Software malfunction - cal_state outside range in ca_clock_date_setup | During calibration | error |
112 | Software malfunction - cal_mode outside range in ca_clock_date_setup | During calibration | error |
113 | Software malfunction - cal_state outside range in ca_clock_time_setup | During calibration | error |
114 | Software malfunction - cal_mode outside range in ca_clock_time_setup | During calibration | error |
115 | Software malfunction - cal_state outside range in ca_coag_leakage_cal | During calibration | error |
116 | Software malfunction - cal_state outside range in ca_econ_cal | During calibration | error |
117 | Software malfunction - econ_cal_state outside range | During calibration | error |
118 | Software malfunction - cal_state outside range in ca_rem_cal | During calibration | error |
119 | Software malfunction - rem_cal_state outside range | During calibration | error |
12 | T_ON_ERR test failed | During startup diagnostics | error |
120 | Calibration value outside acceptable range | During calibration procedure | error |
121 | Software malfunction - cal_mode outside range in ca_generator_setup | During calibration | error |
122 | Open circuit REM sense failure | During calibration | error |
123 | Main microcontroller unable to access GEN_SEM semaphore | At startup or during operation | error |
124 | Main microcontroller unable to access FEEDBACK_SEM semaphore | At startup or during operation | error |
125 | Main microcontroller unable to access ECON_SEM semaphore | At startup or during operation | error |
126 | Main microcontroller unable to access KEY_ACTIVE_SEM semaphore | At startup or during operation | error |
13 | Main microcontroller unable to access GEN_SEM semaphore | At startup | error |
14 | Calibration data checksum error on main microcontroller | At startup | critical |
151 | Main microcontroller unable to access GEN_SEM semaphore | At startup or during operation | error |
152 | Software malfunction - cal_state outside range in cs_i_sns_cal | During calibration | error |
154 | Main microcontroller unable to access FEEDBACK_SEM semaphore | At startup or during operation | error |
16 | T_ON average test failed | During diagnostics | error |
160 | Dosage error test failed with ADC values displayed | During startup diagnostics | error |
170 | Watchdog malfunction - correct value not sent from feedback microcontroller | At startup or during operation | critical |
171 | Watchdog malfunction - correct value not sent from main microcontroller | At startup or during operation | critical |
172 | Watchdog malfunction - main microcontroller unable to access watchdog semaphore | At startup or during operation | error |
173 | Watchdog malfunction - feedback microcontroller unable to access watchdog semaphore | At startup or during operation | error |
174 | Software malfunction - which_errors value outside state range | During operation | error |
180 | Main microcontroller ST Microelectronics RAM check failed | During startup diagnostics | critical |
181 | Feedback microcontroller ST Microelectronics RAM check failed | During startup diagnostics | critical |
182 | Main microcontroller dual-port RAM check failed | During startup diagnostics | critical |
183 | Main microcontroller Page 0 Flash Memory CRC test failed | During startup diagnostics | critical |
184 | Main microcontroller Page F Flash Memory CRC test failed | During startup diagnostics | critical |
185 | Feedback microcontroller Flash Memory CRC test failed | During startup diagnostics | critical |
189 | Software malfunction - display_state outside state range | During operation | error |
190 | Bipolar buttons may be stuck | Internal diagnostics detected stuck bipolar button (up/down arrow, Precise, Standard, or Macro) | warning |
191 | Cut buttons may be stuck | Internal diagnostics detected stuck cut button (up/down arrow, Low, Pure, or Blend) | warning |
192 | Coag buttons may be stuck | Internal diagnostics detected stuck coag button (up/down arrow, Desiccate, Fulgurate, or Spray) | warning |
193 | Recall button may be stuck | Internal diagnostics detected stuck recall button | warning |
194 | Handswitch or Monopolar 1 Footswitch cut pedal stuck | Internal diagnostics detected stuck cut pedal on handswitch or monopolar 1 footswitch | warning |
195 | Handswitch or Monopolar 1 Footswitch coag pedal stuck | Internal diagnostics detected stuck coag pedal on handswitch or monopolar 1 footswitch | warning |
196 | Handswitch or Monopolar 2 Footswitch cut pedal stuck | Internal diagnostics detected stuck cut pedal on handswitch or monopolar 2 footswitch | warning |
197 | Handswitch or Monopolar 2 Footswitch coag pedal stuck | Internal diagnostics detected stuck coag pedal on handswitch or monopolar 2 footswitch | warning |
198 | Handswitch or Bipolar Footswitch pedal stuck | Internal diagnostics detected stuck pedal on handswitch or bipolar footswitch | warning |
199 | Main and feedback microcontrollers not compatible | Internal diagnostics detected incompatible microcontrollers | error |
2 | Feedback microcontroller failed to power up and initialize RAM | At startup | error |
200 | Main microcontroller SEML line may be stuck | Internal diagnostics detected stuck SEML line on main microcontroller | error |
201 | Feedback microcontroller SEMR line may be stuck | Internal diagnostics detected stuck SEMR line on feedback microcontroller | error |
202 | Main microcontroller watchdog test failed | Internal diagnostics watchdog test failed on main microcontroller | error |
203 | Feedback microcontroller watchdog test failed | Internal diagnostics watchdog test failed on feedback microcontroller | error |
206 | Software malfunction in doserr_test_state main value | Software malfunction with doserr_test_state value in st_m_doserr_test outside range | error |
207 | Software malfunction in doserr_test_state feedback value | Software malfunction with doserr_test_state value in st_fb_doserr_test outside range | error |
208 | Main microcontroller timer interrupt failed | Internal diagnostics timer interrupt failed on main microcontroller | critical |
209 | Feedback microcontroller timer interrupt failed | Internal diagnostics timer interrupt failed on feedback microcontroller | critical |
210 | Software malfunction in data_type value | Software malfunction with data_type value outside state range | error |
211 | Software malfunction in which_data value | Software malfunction with which_data value outside state range | error |
212 | Battery removed or replaced | First power-on after battery removal or replacement | info |
212 | Generator not calibrated | Generator lacks calibration data | warning |
213 | Firmware not compatible with hardware | Internal diagnostics detected firmware/hardware incompatibility | error |
215 | Real-time clock chip (U1) not compatible with firmware | Internal diagnostics detected RTC chip incompatibility | error |
216 | Internal memory battery exhausted | Internal battery backing up memory has been depleted | warning |
220 | Feedback microcontroller unable to access KEY_REQ_SEM semaphore | Semaphore access failure on feedback microcontroller | error |
221 | Feedback microcontroller unable to access GEN_SEM semaphore | Semaphore access failure on feedback microcontroller | error |
222 | Main microcontroller failed to copy battery-backed RAM to dual-port RAM | Main microcontroller failed to complete RAM copy in time allotted | error |
223 | Main microcontroller failed to complete initialization in time | Main microcontroller initialization timeout | error |
224 | Calibration data checksum error on feedback microcontroller | Internal diagnostics detected checksum error in calibration data | error |
225 | Feedback microcontroller CPU test failed | Internal diagnostics CPU test failed on feedback microcontroller | critical |
226 | Feedback microcontroller unable to access FEEDBACK_SEM semaphore | Semaphore access failure on feedback microcontroller | error |
230 | Software malfunction in keyed_state value | Software malfunction with keyed_state value outside state range | error |
231 | Software malfunction in key_req_mode value | Software malfunction with key_req_mode value outside state range | error |
232 | Feedback microcontroller unable to access KEY_ACTIVE_SEM semaphore | Semaphore access failure on feedback microcontroller | error |
240 | Software malfunction in activation_seq value | Software malfunction with activation_seq value outside state range | error |
241 | Software malfunction in feedback value in do_feedback() | Software malfunction with feedback value in do_feedback() outside range | error |
242 | Software malfunction in feedback value in do_cal_feedback() | Software malfunction with feedback value in do_cal_feedback() outside range | error |
244 | Software malfunction in disable_seq value | Software malfunction with disable_seq value outside state range | error |
245 | Software malfunction in spark_state value | Software malfunction with spark_state value outside state range | error |
246 | Feedback microcontroller unable to access FEEDBACK_SEM semaphore | Semaphore access failure on feedback microcontroller | error |
247 | Feedback microcontroller unable to access ECON_SEM semaphore | Semaphore access failure on feedback microcontroller | error |
260 | A/D conversion did not complete in allowed time | A/D conversion timeout during operation | critical |
261 | Delay time out of bounds on main microcontroller | Software malfunction with delay time out of bounds | error |
262 | Delay time out of bounds on feedback microcontroller | Software malfunction with delay time out of bounds | error |
270 | Software malfunction in me_rtc_get_() data type | Software malfunction with data type in me_rtc_get_() outside range | error |
271 | Software malfunction in me_rtc_put_() data type | Software malfunction with data type in me_rtc_put_() outside range | error |
3 | Feedback microcontroller failed checksum battery-backed RAM data | At startup | error |
4 | Main microcontroller CPU test failed | During startup diagnostics | critical |
451 | Internal temperature limit exceeded | Length of activation time caused internal temperature to exceed limit | critical |
5 | Main microcontroller unable to access FEEDBACK_SEM semaphore | At startup or during operation | error |
6 | Main microcontroller unable to access ECON_SEM semaphore | At startup | error |
68 | Main microcontroller unable to access GEN_SEM semaphore | At startup or during operation | error |
69 | Software malfunction - flash_power_state value outside range | During operation | error |
7 | Main microcontroller unable to access KEY_ACTIVE_SEM semaphore | At startup | error |
70 | Software malfunction - key_req_mode value outside range | During operation | error |
71 | Software malfunction - request+1 value outside range | During operation | error |
80 | Software malfunction - *input[i].p_state value outside range | During operation | error |
81 | Optoisolator component test failed | During startup diagnostics | critical |
90 | Generator model number in Flash Memory incorrect | At startup | error |
95 | Generator model number in Flash Memory incorrect | At startup | error |
ECON display error | ECON calibration display did not change as expected after footswitch activation | After activating bipolar, cut, or coag output with footswitch | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.