Valleylab Force FX: troubleshooting and error codes

The ServiceForge knowledge base holds 40 documented symptoms for the Valleylab Force FX, with 92 ranked causes, 36 diagnostic checks with test points and expected values, and 110 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Generator on and accessory activated but no output delivered high voltage

Electrosurgical output not generated despite power on and activation signal

Ranked causes

  1. 1. Malfunctioning footswitch or handswitching instrument (common)
  2. 2. Incompatible footswitch (occasional)
  3. 3. Footswitch connected to wrong receptacle for instrument connection (occasional)
  4. 4. Footswitch connected to Monopolar 2 for Monopolar 1 instrument (occasional)
  5. 5. Power set too low (common)
  6. 6. Alarm condition exists (occasional)
  7. 7. Blown fuse on Power Supply/RF board (occasional)
  8. 8. Control board malfunction (occasional)
  9. 9. High voltage power supply malfunction (occasional)
  10. 10. RF output stage malfunction (occasional)
First check

Verify accessory connections

Test point
Footswitch and instrument receptacles
Expected
Secure connections to appropriate receptacles

1 more check with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator does not respond when turned on high voltage

Device fails to power up or initialize when power switch is activated

Ranked causes

  1. 1. Disconnected power cord, faulty wall receptacle, or faulty power cord (common)
  2. 2. Fuse drawer open or fuses blown (common)
  3. 3. Loose or disconnected internal cables (occasional)
  4. 4. Faulty power entry module or connections (occasional)
  5. 5. Faulty low voltage power supply (occasional)
  6. 6. Damaged Control board connectors and/or malfunctioning Control board (occasional)
  7. 7. Shorts or disconnects on Power Supply/RF board (occasional)
  8. 8. Faulty power switch (rare)
  9. 9. Malfunctioning front panel components (rare)
First check

Verify power cord connections

Test point
Generator and wall receptacle
Expected
Secure connections

1 more check with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator is on but did not complete self-test

Device powers on but fails to pass initialization self-test routine

Ranked causes

  1. 1. Alarm condition exists (common)
  2. 2. Software malfunction (occasional)
  3. 3. Loose or disconnected internal cables (occasional)
  4. 4. Faulty low voltage power supply (occasional)
  5. 5. Damaged Control board connectors and/or malfunctioning Control board (occasional)
  6. 6. Shorts or disconnects on Power Supply/RF board (occasional)
  7. 7. Faulty power switch (rare)
  8. 8. Malfunctioning front panel components (rare)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator malfunction high voltage

Generator stops functioning or fails self-test; visible physical damage, loose cords, or damaged casing suspected

Ranked causes

  1. 1. Visible physical damage to generator housing (occasional)
  2. 2. Accessory cords not properly connected (common)
  3. 3. Power cord damage or exposed wires (occasional)
  4. 4. Fuses damaged or loosely seated (common)
First check

Inspect generator for visible physical damage

Test point
Entire chassis exterior
Expected
No cracks, dents, burn marks, or visible damage

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Interference with other devices when generator is activated patient safety

Other equipment experiences electromagnetic interference only during electrosurgical activation

Ranked causes

  1. 1. Metal-to-metal sparking (common)
  2. 2. High settings used for fulguration (common)
  3. 3. Electrically inconsistent ground wires in operating room (occasional)
  4. 4. Monitor responding to radiated frequencies (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Internal component damage on circuit boards high voltage

Visible damage, cracks, corrosion, or broken wires detected on internal circuit boards after cover removal

Ranked causes

  1. 1. Damaged Control board (rare)
  2. 2. Damaged Display board (rare)
  3. 3. Damaged Footswitch board (internal) (rare)
  4. 4. Severe damage on Power Supply/RF board (rare)
First check

Remove cover and inspect all boards for damage

Test point
All internal circuit boards (Control, Display, Footswitch, Power Supply/RF)
Expected
No damaged components, no broken wires, no cracks, no corrosion visible on any board

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Mode buttons do not function correctly when pressed

Front panel mode buttons fail to change operating modes or show no response

Ranked causes

  1. 1. Faulty ribbon cable between Control board and Display board (common)
  2. 2. Incorrect modes communicated through Control board (occasional)
  3. 3. Faulty ribbon cable between front panel and Display board (occasional)
  4. 4. Incorrect modes communicated from front panel (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Abnormal neuromuscular stimulation during surgery patient safety

Patient displays involuntary muscle contractions or abnormal stimulation during procedure; requires immediate cessation of surgery

Ranked causes

  1. 1. Metal-to-metal sparking (common)
  2. 2. High power setting during coagulation (common)
  3. 3. Abnormal 50-60 Hz leakage currents (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Activation and alarm tones do not sound

Speaker produces no audio output for activation tones or alarm alerts

Ranked causes

  1. 1. Poor connection or damaged Footswitch board ribbon cable (common)
  2. 2. Faulty connections or speaker on Footswitch board (occasional)
  3. 3. Audio signal malfunction on Control board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm 212 appears in Cut display during self-test

Specific error code 212 displayed during power-on self-test sequence

Ranked causes

  1. 1. Battery removed or replaced without generator calibration (common)
  2. 2. Faulty battery on Control board causing loss of battery-backed memory (occasional)
  3. 3. Loss of battery-backed RAM due to faulty component on Control board (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Blank or confusing LED display

Display board shows garbled, missing, or incorrect information

Ranked causes

  1. 1. Faulty ribbon cable between Control board and Display board (common)
  2. 2. Incorrect display modes communicated through Control board (occasional)
  3. 3. Display board malfunction (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Continuous monitor interference patient safety

Cardiac monitor displays noise or artifacts continuously regardless of generator activation

Ranked causes

  1. 1. Faulty chassis-to-ground connections (common)
  2. 2. Electrical equipment grounded to different objects (occasional)
  3. 3. Malfunctioning monitor (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

T_ON ASIC malfunction alarm high voltage

Generator displays an alarm number related to T_ON ASIC malfunction during or after self-test

Ranked causes

  1. 1. Programmable IC U9 pins not properly seated in socket (common)
  2. 2. Programmable IC U9 defective or failed (occasional)
  3. 3. Control board defective or failed (rare)
First check

Verify IC U9 pins seated properly in socket on Control board

Test point
IC U9 socket
Expected
All pins fully inserted and aligned

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery-backed RAM malfunction indicated high voltage

Alarm number appears, indicating memory corruption or loss in battery-backed RAM; requires battery replacement and recalibration

Ranked causes

  1. 1. Battery depleted or failed (common)
  2. 2. Control board defective (occasional)
First check

Turn off generator and disconnect power cord

Test point
Main power connector
Expected
Power disconnected, wait 5 minutes minimum

10 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Footswitch will not activate output

Footswitch fails to trigger electrosurgical output generation

Ranked causes

  1. 1. Malfunctioning or damaged footswitch receptacle (common)
  2. 2. Footswitch activation signal lost on Control board (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pacemaker interference patient safety

Pacemaker malfunction or inappropriate sensing triggered by electrosurgery

Ranked causes

  1. 1. Intermittent connections or metal-to-metal sparking (common)
  2. 2. Current path too close to pacemaker during monopolar electrosurgery (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

REM indicator illuminates red patient safety

REM Polyhesive Patient Return Electrode not properly applied or not connected to generator

Ranked causes

  1. 1. REM Patient Return Electrode not properly applied to patient (common)
  2. 2. REM Patient Return Electrode not connected to generator (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration procedure notice: Minimum load resistor thresholds caution

Equipment damage will result if generator activated with load resistor below specified threshold during calibration

Ranked causes

  1. 1. Exceeding maximum output power into low impedance load during calibration (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration value out of acceptable range

Alarm 120 displayed; calibration value exceeds acceptable limits

Ranked causes

  1. 1. Calibration procedure failed or sensor malfunction (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cover removal and reinstallation high voltage

Procedure for accessing internal generator components by removing and reinstalling the generator cover

Ranked causes

  1. 1. Access required for internal component inspection (common)
First check

Cover removal

Test point
Cover fastening screws
Expected
All five screws removed and cover lifts clear of chassis

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display board seven-segment LED failure

Seven-segment LED on display board is non-functional, dim, or discolored

Ranked causes

  1. 1. LED worn out or defective (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Dosage error during diagnostics caution

Alarm 160 displayed; internal dosage test failed with ADC values shown

Ranked causes

  1. 1. ECON factor out of calibration or ADC malfunction (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fan replacement caution

Fan requires replacement on Power Supply/RF board assembly

Ranked causes

  1. 1. Fan failure or wear (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flash memory CRC test failed caution

Alarm 183-185 displayed; Flash memory CRC test failed in main or feedback microcontroller

Ranked causes

  1. 1. Flash memory corruption or microcontroller failure (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fuse blown in fuse drawer caution

Generator loses power or protection circuit trips; fuses in the drawer below power cord receptacle are open or damaged.

Ranked causes

  1. 1. Blown fuse in fuse drawer (120 V or 240 V operation) (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fuse blown on Power Supply/RF board high voltage

Generator fails to operate; fuse on internal Power Supply/RF board is blown or open.

Ranked causes

  1. 1. Blown fuse on Power Supply/RF board (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator fails internal diagnostics at startup caution

Alarm 4 displayed; main microcontroller CPU test failed during power-up

Ranked causes

  1. 1. Control board CPU malfunction (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loose bipolar instrument receptacle connection caution

Bipolar instrument connector fits loosely into front panel receptacle; instrument operation may be unreliable

Ranked causes

  1. 1. Worn or damaged bipolar instrument receptacle (occasional)
First check

Test bipolar instrument connector fit in front panel receptacle

Test point
Bipolar Instrument receptacle on front panel
Expected
Footswitching and handswitching instrument connectors insert securely with no movement

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loose footswitch receptacle connection caution

Footswitch connector fits loosely into rear panel receptacles; footswitching may be unreliable

Ranked causes

  1. 1. Worn or damaged footswitch receptacle contacts (occasional)
First check

Test bipolar footswitch receptacle fit

Test point
Three footswitch receptacles on rear panel
Expected
Bipolar footswitch connector inserts securely with no movement or looseness

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loose monopolar instrument receptacle connection caution

Monopolar instrument connector fits loosely into front panel receptacles; instrument operation may be unreliable

Ranked causes

  1. 1. Worn or damaged monopolar instrument receptacle (occasional)
First check

Test monopolar 1/CEM connector fit in front panel receptacle

Test point
Monopolar 1/CEM Instrument receptacle on front panel
Expected
Footswitching and handswitching instrument connectors insert securely with no movement

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Microcontroller RAM check failed caution

Alarm 180/181 displayed; internal RAM diagnostic failed on main or feedback microcontroller

Ranked causes

  1. 1. RAM memory malfunction in microcontroller (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Optoisolator internal component failed caution

Alarm 81 displayed; optoisolator component test failed

Ranked causes

  1. 1. Optoisolator component failure (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Patient Return Electrode receptacle damaged patient safety

Receptacle shows visible damage, broken pin, or obstruction; patient return connection compromised

Ranked causes

  1. 1. Damaged or obstructed Patient Return Electrode receptacle (rare)
First check

Inspect Patient Return Electrode receptacle

Test point
Patient Return Electrode receptacle on front panel
Expected
No broken pins, no obstructions, clear unobstructed opening

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

REM alarm indicator does not illuminate patient safety

No illumination (red or green) from REM alarm indicator light

Ranked causes

  1. 1. Return electrode without REM safety feature connected (common)
First check

Verify REM Polyhesive Patient Return Electrode is connected

Test point
Return electrode type
Expected
REM Polyhesive Patient Return Electrode installed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

REM alarm indicator flashing red patient safety

REM Polyhesive Patient Return Electrode alarm condition detected; indicator flashes red until corrected

Ranked causes

  1. 1. Patient return electrode not properly applied or connected (common)
First check

Verify REM Patient Return Electrode is properly applied to patient

Test point
Patient contact
Expected
Electrode making full contact with patient skin

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

REM alarm indicator illuminates red (steady) patient safety

Indicator shows red light (steady, not flashing) when REM Patient Return Electrode first connected or system alarm present

Ranked causes

  1. 1. REM Patient Return Electrode not yet applied to patient (common)
First check

Confirm REM Patient Return Electrode applied and connected

Test point
Electrode connection
Expected
Electrode properly applied to patient and connected to generator

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

REM indicator does not illuminate patient safety

Return electrode connected without REM safety feature

Ranked causes

  1. 1. Return electrode without REM safety feature connected (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

REM indicator flashes red patient safety

REM system has detected an alarm condition that must be corrected

Ranked causes

  1. 1. REM system alarm condition detected (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

REM sense circuit open or failed

Alarm 122 displayed; open circuit detected in REM sense line during calibration

Ranked causes

  1. 1. Open circuit in REM sense line (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

T_ON test fails during diagnostics caution

Alarm 12 or 16 displayed; T_ON timing or average test failed

Ranked causes

  1. 1. T_ON ASIC malfunction or calibration error (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
0Main microcontroller failed to hold feedback microcontroller in resetAt startuperror
1Main microcontroller failed to power up feedback microcontrollerAt startuperror
10Software malfunction - invalid alarm number on serial portDuring operationerror
100Software malfunction - rem_update_state outside rangeDuring operationerror
101Software malfunction - rem_pad_state outside rangeDuring operationerror
102Software malfunction - rem_state in re_rem_control outside rangeDuring operationerror
103Software malfunction - rem_state in re_non_rem_control outside rangeDuring operationerror
104Software malfunction - rem_flash_state outside rangeDuring operationerror
105Software malfunction - rem_led_state outside rangeDuring operationerror
11Invalid activated power, function, mode, or key request mode echoedAt startup or during operationerror
110Software malfunction - cal_state outside range in ca_generator_setupDuring calibrationerror
111Software malfunction - cal_state outside range in ca_clock_date_setupDuring calibrationerror
112Software malfunction - cal_mode outside range in ca_clock_date_setupDuring calibrationerror
113Software malfunction - cal_state outside range in ca_clock_time_setupDuring calibrationerror
114Software malfunction - cal_mode outside range in ca_clock_time_setupDuring calibrationerror
115Software malfunction - cal_state outside range in ca_coag_leakage_calDuring calibrationerror
116Software malfunction - cal_state outside range in ca_econ_calDuring calibrationerror
117Software malfunction - econ_cal_state outside rangeDuring calibrationerror
118Software malfunction - cal_state outside range in ca_rem_calDuring calibrationerror
119Software malfunction - rem_cal_state outside rangeDuring calibrationerror
12T_ON_ERR test failedDuring startup diagnosticserror
120Calibration value outside acceptable rangeDuring calibration procedureerror
121Software malfunction - cal_mode outside range in ca_generator_setupDuring calibrationerror
122Open circuit REM sense failureDuring calibrationerror
123Main microcontroller unable to access GEN_SEM semaphoreAt startup or during operationerror
124Main microcontroller unable to access FEEDBACK_SEM semaphoreAt startup or during operationerror
125Main microcontroller unable to access ECON_SEM semaphoreAt startup or during operationerror
126Main microcontroller unable to access KEY_ACTIVE_SEM semaphoreAt startup or during operationerror
13Main microcontroller unable to access GEN_SEM semaphoreAt startuperror
14Calibration data checksum error on main microcontrollerAt startupcritical
151Main microcontroller unable to access GEN_SEM semaphoreAt startup or during operationerror
152Software malfunction - cal_state outside range in cs_i_sns_calDuring calibrationerror
154Main microcontroller unable to access FEEDBACK_SEM semaphoreAt startup or during operationerror
16T_ON average test failedDuring diagnosticserror
160Dosage error test failed with ADC values displayedDuring startup diagnosticserror
170Watchdog malfunction - correct value not sent from feedback microcontrollerAt startup or during operationcritical
171Watchdog malfunction - correct value not sent from main microcontrollerAt startup or during operationcritical
172Watchdog malfunction - main microcontroller unable to access watchdog semaphoreAt startup or during operationerror
173Watchdog malfunction - feedback microcontroller unable to access watchdog semaphoreAt startup or during operationerror
174Software malfunction - which_errors value outside state rangeDuring operationerror
180Main microcontroller ST Microelectronics RAM check failedDuring startup diagnosticscritical
181Feedback microcontroller ST Microelectronics RAM check failedDuring startup diagnosticscritical
182Main microcontroller dual-port RAM check failedDuring startup diagnosticscritical
183Main microcontroller Page 0 Flash Memory CRC test failedDuring startup diagnosticscritical
184Main microcontroller Page F Flash Memory CRC test failedDuring startup diagnosticscritical
185Feedback microcontroller Flash Memory CRC test failedDuring startup diagnosticscritical
189Software malfunction - display_state outside state rangeDuring operationerror
190Bipolar buttons may be stuckInternal diagnostics detected stuck bipolar button (up/down arrow, Precise, Standard, or Macro)warning
191Cut buttons may be stuckInternal diagnostics detected stuck cut button (up/down arrow, Low, Pure, or Blend)warning
192Coag buttons may be stuckInternal diagnostics detected stuck coag button (up/down arrow, Desiccate, Fulgurate, or Spray)warning
193Recall button may be stuckInternal diagnostics detected stuck recall buttonwarning
194Handswitch or Monopolar 1 Footswitch cut pedal stuckInternal diagnostics detected stuck cut pedal on handswitch or monopolar 1 footswitchwarning
195Handswitch or Monopolar 1 Footswitch coag pedal stuckInternal diagnostics detected stuck coag pedal on handswitch or monopolar 1 footswitchwarning
196Handswitch or Monopolar 2 Footswitch cut pedal stuckInternal diagnostics detected stuck cut pedal on handswitch or monopolar 2 footswitchwarning
197Handswitch or Monopolar 2 Footswitch coag pedal stuckInternal diagnostics detected stuck coag pedal on handswitch or monopolar 2 footswitchwarning
198Handswitch or Bipolar Footswitch pedal stuckInternal diagnostics detected stuck pedal on handswitch or bipolar footswitchwarning
199Main and feedback microcontrollers not compatibleInternal diagnostics detected incompatible microcontrollerserror
2Feedback microcontroller failed to power up and initialize RAMAt startuperror
200Main microcontroller SEML line may be stuckInternal diagnostics detected stuck SEML line on main microcontrollererror
201Feedback microcontroller SEMR line may be stuckInternal diagnostics detected stuck SEMR line on feedback microcontrollererror
202Main microcontroller watchdog test failedInternal diagnostics watchdog test failed on main microcontrollererror
203Feedback microcontroller watchdog test failedInternal diagnostics watchdog test failed on feedback microcontrollererror
206Software malfunction in doserr_test_state main valueSoftware malfunction with doserr_test_state value in st_m_doserr_test outside rangeerror
207Software malfunction in doserr_test_state feedback valueSoftware malfunction with doserr_test_state value in st_fb_doserr_test outside rangeerror
208Main microcontroller timer interrupt failedInternal diagnostics timer interrupt failed on main microcontrollercritical
209Feedback microcontroller timer interrupt failedInternal diagnostics timer interrupt failed on feedback microcontrollercritical
210Software malfunction in data_type valueSoftware malfunction with data_type value outside state rangeerror
211Software malfunction in which_data valueSoftware malfunction with which_data value outside state rangeerror
212Battery removed or replacedFirst power-on after battery removal or replacementinfo
212Generator not calibratedGenerator lacks calibration datawarning
213Firmware not compatible with hardwareInternal diagnostics detected firmware/hardware incompatibilityerror
215Real-time clock chip (U1) not compatible with firmwareInternal diagnostics detected RTC chip incompatibilityerror
216Internal memory battery exhaustedInternal battery backing up memory has been depletedwarning
220Feedback microcontroller unable to access KEY_REQ_SEM semaphoreSemaphore access failure on feedback microcontrollererror
221Feedback microcontroller unable to access GEN_SEM semaphoreSemaphore access failure on feedback microcontrollererror
222Main microcontroller failed to copy battery-backed RAM to dual-port RAMMain microcontroller failed to complete RAM copy in time allottederror
223Main microcontroller failed to complete initialization in timeMain microcontroller initialization timeouterror
224Calibration data checksum error on feedback microcontrollerInternal diagnostics detected checksum error in calibration dataerror
225Feedback microcontroller CPU test failedInternal diagnostics CPU test failed on feedback microcontrollercritical
226Feedback microcontroller unable to access FEEDBACK_SEM semaphoreSemaphore access failure on feedback microcontrollererror
230Software malfunction in keyed_state valueSoftware malfunction with keyed_state value outside state rangeerror
231Software malfunction in key_req_mode valueSoftware malfunction with key_req_mode value outside state rangeerror
232Feedback microcontroller unable to access KEY_ACTIVE_SEM semaphoreSemaphore access failure on feedback microcontrollererror
240Software malfunction in activation_seq valueSoftware malfunction with activation_seq value outside state rangeerror
241Software malfunction in feedback value in do_feedback()Software malfunction with feedback value in do_feedback() outside rangeerror
242Software malfunction in feedback value in do_cal_feedback()Software malfunction with feedback value in do_cal_feedback() outside rangeerror
244Software malfunction in disable_seq valueSoftware malfunction with disable_seq value outside state rangeerror
245Software malfunction in spark_state valueSoftware malfunction with spark_state value outside state rangeerror
246Feedback microcontroller unable to access FEEDBACK_SEM semaphoreSemaphore access failure on feedback microcontrollererror
247Feedback microcontroller unable to access ECON_SEM semaphoreSemaphore access failure on feedback microcontrollererror
260A/D conversion did not complete in allowed timeA/D conversion timeout during operationcritical
261Delay time out of bounds on main microcontrollerSoftware malfunction with delay time out of boundserror
262Delay time out of bounds on feedback microcontrollerSoftware malfunction with delay time out of boundserror
270Software malfunction in me_rtc_get_() data typeSoftware malfunction with data type in me_rtc_get_() outside rangeerror
271Software malfunction in me_rtc_put_() data typeSoftware malfunction with data type in me_rtc_put_() outside rangeerror
3Feedback microcontroller failed checksum battery-backed RAM dataAt startuperror
4Main microcontroller CPU test failedDuring startup diagnosticscritical
451Internal temperature limit exceededLength of activation time caused internal temperature to exceed limitcritical
5Main microcontroller unable to access FEEDBACK_SEM semaphoreAt startup or during operationerror
6Main microcontroller unable to access ECON_SEM semaphoreAt startuperror
68Main microcontroller unable to access GEN_SEM semaphoreAt startup or during operationerror
69Software malfunction - flash_power_state value outside rangeDuring operationerror
7Main microcontroller unable to access KEY_ACTIVE_SEM semaphoreAt startuperror
70Software malfunction - key_req_mode value outside rangeDuring operationerror
71Software malfunction - request+1 value outside rangeDuring operationerror
80Software malfunction - *input[i].p_state value outside rangeDuring operationerror
81Optoisolator component test failedDuring startup diagnosticscritical
90Generator model number in Flash Memory incorrectAt startuperror
95Generator model number in Flash Memory incorrectAt startuperror
ECON display errorECON calibration display did not change as expected after footswitch activationAfter activating bipolar, cut, or coag output with footswitcherror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.