Trophon EPR: troubleshooting and error codes
The ServiceForge knowledge base holds 24 documented symptoms for the Trophon EPR, with 25 ranked causes, 29 diagnostic checks with test points and expected values, and 173 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unable to fill nebuliser despite liquid present caution
Nebuliser level sensor cannot detect liquid during fill pump operation
Ranked causes
- 1. Tubing clogged or fill pump failure (common)
- 2. Reservoir assembly failure or low level sensor issue (occasional)
Inspect fill system tubing for blockage
- Test point
- Fill pump tubing from reservoir to nebuliser
- Expected
- Tubing clear, no blockage or kinks
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air path damaged or cracked caution
Disinfection air path shows cracks or structural damage
Ranked causes
- 1. Impact damage or stress crack in air path tubing or components (occasional)
Visually inspect air path for cracks or leaks
- Test point
- Air path tubing and connectors
- Expected
- No cracks, no visible leakage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Board communication error caution
Communication error between device PCBs; Bxxx board failure code displayed
Ranked causes
- 1. I²C connector loose or contact issue on control PCB (common)
Power cycle the device
- Test point
- Device power switch
- Expected
- Device powers down and back up cleanly
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chamber door micro switches misaligned or damaged caution
Chamber door fail error; micro switches not engaging properly or lever bent
Ranked causes
- 1. Door micro switch lever bent or switch contacts worn (occasional)
Inspect chamber door micro switch assembly
- Test point
- Micro switch lever and contacts
- Expected
- Lever straight, contacts clean and engaging
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chamber door unable to lock or unlock caution
Chamber door locking mechanism fails to engage or disengage
Ranked causes
- 1. Door not closed properly or locking mechanism stuck (occasional)
Verify chamber door is fully closed
- Test point
- Chamber door position
- Expected
- Door fully seated in frame
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chamber heater temperature out of range caution
Front or rear chamber wall not reaching required operating temperature
Ranked causes
- 1. Chamber heater element or control failure (rare)
Monitor chamber heater temperature during warm-up
- Test point
- Chamber temperature sensor
- Expected
- Temperature rises to set point within normal time
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chamber valve seal leaking caution
Chamber valve seal failure detected; valve not sealing properly
Ranked causes
- 1. Chamber valve seal worn or damaged (occasional)
Perform chamber pressure test
- Test point
- Chamber pressure sensor during cycle
- Expected
- Chamber maintains pressure without leakage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chamber valve unable to open caution
Chamber valve fails to open; device cannot execute cycle
Ranked causes
- 1. Chamber valve faulted or stuck (common)
Check for DGS04 or DFP014 errors
- Test point
- Techlog error list
- Expected
- No DGS04/DFP014 errors after valve replacement
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Damaged or missing cable ties caution
Internal cable ties broken, missing, or not secured tightly
Ranked causes
- 1. Normal wear or vibration damage to cable ties (occasional)
Inspect internal cable routing and cable tie condition
- Test point
- Cable tie locations throughout device
- Expected
- All cables secured with tight cable ties; no loose or broken ties
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Destruct valve seal leaking caution
Destruct valve seal failure detected; valve not sealing properly
Ranked causes
- 1. Destruct valve seal worn or damaged (occasional)
Perform destruct valve pressure test
- Test point
- Destruct valve during test cycle
- Expected
- Valve maintains seal; no leakage detected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Destruct valve unable to open caution
Destruct valve fails to open; disinfection cycle cannot complete
Ranked causes
- 1. Destruct valve faulted or stuck (common)
Check for DGS05 or DFP013 errors
- Test point
- Techlog error list
- Expected
- No DGS05/DFP013 errors after valve replacement
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
H2O2 mist dosage out of range caution
Dosage sensor reading above or below allowable maximum/minimum range
Ranked causes
- 1. Nebuliser assembly dosage sensor or heater malfunction (rare)
Run dosage test cycle and monitor sensor reading
- Test point
- Dosage sensor during test cycle
- Expected
- Dosage within min/max specification range
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hardware safety interlock engaged patient safety
Safety interlock activated; device has ceased current operation
Ranked causes
- 1. Customer interference with safety mechanism (common)
Verify no external interference with chamber or moving parts
- Test point
- Chamber and mechanical components
- Expected
- No obstruction; all guards in place
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nebuliser cup overheating or underheating caution
Nebuliser cup temperature outside normal operating range
Ranked causes
- 1. Nebuliser cup heating element failure (rare)
Monitor nebuliser cup temperature during idle and cycle
- Test point
- Temperature sensor reading
- Expected
- Temperature within normal operating range
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nebuliser fan speed out of range caution
Nebuliser fan RPM outside acceptable operating range
Ranked causes
- 1. Nebuliser fan motor wear or bearing failure (occasional)
Measure nebuliser fan RPM during operation
- Test point
- Fan motor speed sensor
- Expected
- RPM within specification range
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Needle stuck in dispense pump assembly
One or more dispense pump assembly needles stuck in extended position
Ranked causes
- 1. Dispense pump assembly failure (common)
Check for SS02 or PT0x errors in techlog
- Test point
- Techlog error display
- Expected
- No recurring SS02/PT0x errors
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Non-persistent error displayed on LCD
Error message on Trophon EPR LCD requests user to select OK, indicating a non-persistent error from operator error or temporary unexpected event
Ranked causes
- 1. Operator error or temporary unexpected event (common)
Review error message on LCD display
- Test point
- LCD display
- Expected
- Error message requesting OK selection
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Persistent error displayed on LCD patient safety
Error message on Trophon EPR LCD requests user to select PURGE, indicating a persistent error from component failure or serious system error
Ranked causes
- 1. Component failure or serious internal error (common)
Review error message on LCD display
- Test point
- LCD display
- Expected
- Error message requesting PURGE selection
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe clamp worn or damaged caution
Mechanical probe clamp assembly shows wear or structural damage
Ranked causes
- 1. Normal wear or impact damage to probe clamp (occasional)
Inspect probe clamp for cracks or deformation
- Test point
- Probe clamp assembly
- Expected
- Clamp intact, no cracks or severe wear
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe detect message constantly displayed caution
Insert/Remove Probe message remains on LCD display continuously
Ranked causes
- 1. Probe detect sensor failure or connector issue (occasional)
Verify probe connection status
- Test point
- Probe connector and insertion detection
- Expected
- Probe detected correctly when inserted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe detect sensor malfunction caution
Probe detect sensor not functioning or not detected by system during chamber operation
Ranked causes
- 1. Probe detect sensor assembly defective or failed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Purge pump replacement patient safety
Purge pump requires replacement; procedure for removal and installation of replacement pump A01017
Ranked causes
- 1. Purge pump failure or end-of-life (unknown)
Disinfectant reservoir empty
- Test point
- Disinfectant reservoir
- Expected
- Completely empty
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unknown error code 0xFFFF:FFFF caution
Unknown error displayed; control PCB battery voltage below 3V threshold
Ranked causes
- 1. Control PCB backup battery depleted (common)
Measure control PCB battery voltage
- Test point
- Battery connector on control PCB
- Expected
- 3.0V or higher, >= 3.0 V
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verification temperature out of range caution
Verification temperature sensor detects incorrect chamber temperature during validation cycle
Ranked causes
- 1. Temperature sensor reading error or actual chamber temp out of range (rare)
Monitor verification temperature reading across multiple cycles
- Test point
- Temperature sensor during validation
- Expected
- Temperature within specification for verification phase
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
00000001 | Chamber Door Open: Passed | Self-test executed | info |
00000002 | Chamber Door Open: Failed | Self-test executed | error |
00000004 | Insert Probe: Passed | Self-test executed | info |
00000008 | Insert Probe: Failed | Self-test executed | error |
00000010 | Remove Probe: Passed | Self-test executed | info |
00000020 | Remove Probe: Failed | Self-test executed | error |
00000040 | Chamber Door Closed: Passed | Self-test executed | info |
00000080 | Chamber Door Closed: Failed | Self-test executed | error |
00000100 | Valves and Fans: Passed | Self-test executed | info |
00000200 | Valves and Fans: Failed | Self-test executed | error |
00000400 | Button 1 Press: Failed | Self-test executed | error |
00000800 | Button 2 Press: Failed | Self-test executed | error |
00001000 | Button 3 Press: Failed | Self-test executed | error |
00002000 | Button 4 Press: Failed | Self-test executed | error |
00004000 | Start Button Press: Failed | Self-test executed | error |
00008000 | Beep: Passed | Self-test executed | info |
00010000 | Beep: Failed | Self-test executed | error |
00020000 | Waste Drawer Out: Passed | Self-test executed | info |
00040000 | Waste Drawer Out: Failed | Self-test executed | error |
00080000 | Waste Drawer In: Passed | Self-test executed | info |
00100000 | Waste Drawer In: Failed | Self-test executed | error |
00200000 | Cartridge Door Close: Passed | Self-test executed | info |
00400000 | Cartridge Door Close: Failed | Self-test executed | error |
00800000 | Needles Checked | Self-test executed | info |
01000000 | Rim Needle: Failed | Self-test executed | error |
02000000 | Neck Needle: Failed | Self-test executed | error |
04000000 | Nebulant Cartridge Out Detection: Passed | Self-test executed | info |
08000000 | Nebulant Cartridge Out Detection: Failed | Self-test executed | error |
0xFFFF:FFFF | Unknown error; control PCB battery critical | Control PCB backup battery voltage drops below 3.0V threshold | critical |
10000000 | Nebulant Cartridge In Detection: Passed | Self-test executed | info |
20000000 | Nebulant Cartridge In Detection: Failed | Self-test executed | error |
40000000 | Cartridge Door Open: Passed | Self-test executed | info |
80000000 | Cartridge Door Open: Failed | Self-test executed | error |
BF01 | I²C Bus Failure | Loss of communication on I²C bus | error |
BF02 | Missing Process Board U1 | Process board U1 not detected | error |
BF04 | Missing Process Board U2 | Process board U2 not detected | error |
BF08 | Missing Process Board U3 | Process board U3 not detected | error |
BF10 | Missing Nebuliser Board U1 | Nebuliser board U1 not detected | error |
BF20 | Missing PDS Board U1 | PDS board U1 not detected | error |
BF40 | Missing PDS Board U2 | PDS board U2 not detected | error |
BFxxx | Board Failure; PCB communication error | I²C bus communication timeout or integrity error between PCBs | error |
Chamber Door Fail | Chamber door micro switch failure | Door position micro switch fails to signal properly during open/close cycle | warning |
Chamber Valve: Leak Detected | Chamber valve seal leaking | Pressure decay test detects seal failure during or after valve cycle | error |
CL01 | Chamber door unable to lock | Door lock solenoid or mechanism fails to engage | warning |
CL02 | Chamber door unable to unlock | Door unlock solenoid or mechanism fails to disengage | warning |
Destruct Valve: Leak Sealing Detected | Destruct valve seal leaking | Pressure decay test detects seal failure during or after valve cycle | error |
DFP | Process Control Failure (While Disinfecting) | Process control failure during disinfection cycle | error |
DFP - 01 | Nebuliser Temperature: Over (Process control) | Process control failure during disinfection | error |
DFP013 | Destruct valve failure confirmed | Repeated destruct valve operation failures or seal detection | error |
DFP014 | Chamber valve failure confirmed | Repeated chamber valve operation failures or seal detection | error |
DFP - 02 | Nebuliser Temperature: Under (Process control) | Process control failure during disinfection | error |
DFP - 03 | Rear Chamber Heater: Over (Process control) | Process control failure during disinfection | error |
DFP - 04 | Rear Chamber Heater: Under (Process control) | Process control failure during disinfection | error |
DFP - 05 | Front Chamber Heater: Over (Process control) | Process control failure during disinfection | error |
DFP - 06 | Front Chamber Heater: Under (Process control) | Process control failure during disinfection | error |
DFP - 07 | Power Lost during Procedure (Process control) | Process control failure during disinfection | error |
DFP - 08 | Board Failure (Process control) | Process control failure during disinfection | critical |
DFP - 09 | Nebuliser Fan: Incorrect Speed (Process control) | Process control failure during disinfection | error |
DFP - 10 | Destruct Fan: Incorrect Speed (Process control) | Process control failure during disinfection | error |
DFP - 11 | Mist Strength: Inconsisent (Process control) | Process control failure during disinfection | error |
DFP - 12 | Mist Strength Send Failure (Process control) | Process control failure during disinfection | error |
DFP - 13 | Destruct Valve: Fail (Process control) | Process control failure during disinfection | error |
DFP - 14 | Chamber Valve: Fail (Process control) | Process control failure during disinfection | error |
DFP - 15 | Interlocks: Fail (Process control) | Process control failure during disinfection | error |
DFP - 16 | Invalid Stage Parameter (Process control) | Process control failure during disinfection | error |
DFV | Process Verification Failure (While Disinfecting) | Verification failure during disinfection cycle | error |
DFV - 0001 | Stage Timing: Over (Process verification) | Verification failure during disinfection | error |
DFV - 0002 | Stage Timing: Under (Process verification) | Verification failure during disinfection | error |
DFV0004 | Mist dosage over maximum allowable range | Dosage sensor reading exceeds maximum threshold during mist generation | warning |
DFV0008 | Mist dosage under minimum allowable range | Dosage sensor reading below minimum threshold during mist generation | warning |
DFV - 0010 | Verification Temperature: Over (Process verification) | Verification failure during disinfection | error |
DFV - 0020 | Verification Temperature: Under (Process verification) | Verification failure during disinfection | error |
DFV - 0040 | Mist Baseline: Over (Process verification) | Verification failure during disinfection | error |
DFV - 0080 | Mist Baseline: Under (Process verification) | Verification failure during disinfection | error |
DFV - 0100 | Mist Flowrate: Over (Process verification) | Verification failure during disinfection | error |
DFV - 0200 | Mist Flowrate: Under (Process verification) | Verification failure during disinfection | error |
DFV - 0400 | Invalid Dosage Parameter (Process verification) | Verification failure during disinfection | error |
DFV - 0800 | Destruct Valve Sealing: Leak Detected (Process verification) | Verification failure during disinfection | error |
DFV - 1000 | Chamber Valve Sealing: Leak Detected (Process verification) | Verification failure during disinfection | error |
DFV - 4000 | Nebuliser Liquid Refilled (Process verification) | Liquid level change during disinfection verification | info |
DGAPxx - 01 | Nebuliser Cup Temperature: Over (Priming process) | Diagnostic priming action process failure | error |
DGAPxx - 02 | Nebuliser Cup Temperature: Under (Priming process) | Diagnostic priming action process failure | error |
DGAPxx - 03 | Rear Chamber Heater: Over (Priming process) | Diagnostic priming action process failure | error |
DGAPxx - 04 | Rear Chamber Heater: Under (Priming process) | Diagnostic priming action process failure | error |
DGAPxx - 05 | Front Chamber Heater: Over (Priming process) | Diagnostic priming action process failure | error |
DGAPxx - 06 | Front Chamber Heater: Under (Priming process) | Diagnostic priming action process failure | error |
DGAPxx - 07 | Power Cycled during Process (Priming process) | Diagnostic priming action process failure | warning |
DGAPxx - 08 | Process Board Failure (Priming process) | Diagnostic priming action process failure | critical |
DGAPxx - 09 | Nebuliser Fan: Incorrect Speed (Priming process) | Diagnostic priming action process failure | error |
DGAPxx - 10 | Destruct Fan: Incorrect Speed (Priming process) | Diagnostic priming action process failure | error |
DGAPxx - 11 | Mist Strength: Inconsistent (Priming process) | Diagnostic priming action process failure | error |
DGAPxx - 12 | Mist Strength Send Failure (Priming process) | Diagnostic priming action process failure | error |
DGAPxx - 13 | Destruct Valve: Fail (Priming process) | Diagnostic priming action process failure | error |
DGAPxx - 14 | Chamber Valve: Fail (Priming process) | Diagnostic priming action process failure | error |
DGAPxx - 15 | Interlocks Fail (Priming process) | Diagnostic priming action process failure | error |
DGAPxx - 16 | Invalid Stage Parameter (Priming process) | Diagnostic priming action process failure | error |
DGAVxxxx - 0001 | Stage Timing: Over (Priming verification) | Diagnostic priming verification failure | error |
DGAVxxxx - 0002 | Stage Timing: Under (Priming verification) | Diagnostic priming verification failure | error |
DGAVxxxx - 0004 | Insufficient Liquid During Priming | Diagnostic priming verification failure | error |
DGAVxxxx - 0008 | Priming Terminated – Reset Required | Diagnostic priming verification failure | warning |
DGAVxxxx - 0010 | Verification Temperature: Over (Priming verification) | Diagnostic priming verification failure | error |
DGAVxxxx - 0020 | Verification Temperature: Under (Priming verification) | Diagnostic priming verification failure | error |
DGAVxxxx - 0040 | Mist Baseline: Over (Priming verification) | Diagnostic priming verification failure | error |
DGAVxxxx - 0080 | Mist Baseline: Under (Priming verification) | Diagnostic priming verification failure | error |
DGAVxxxx - 0100 | Mist Flow Rate: Over (Priming verification) | Diagnostic priming verification failure | error |
DGAVxxxx - 0200 | Mist Flow Rate: Under (Priming verification) | Diagnostic priming verification failure | error |
DGAVxxxx - 0400 | Invalid Dosage Parameter (Priming verification) | Diagnostic priming verification failure | error |
DGAVxxxx - 0800 | Destruct Valve Sealing: Leak Detected (Priming verification) | Diagnostic priming verification failure | error |
DGAVxxxx - 1000 | Chamber Valve Sealing: Leak Detected (Priming verification) | Diagnostic priming verification failure | error |
DGAVxxxx - 8000 | User Exited Priming Cycle | Diagnostic priming verification failure | info |
DGPxxxx - 0001 | Nebuliser Cup Temperature: Over (Purge failure) | Purge diagnostics failure | error |
DGPxxxx - 0002 | Nebuliser Cup Temperature: Under (Purge failure) | Purge diagnostics failure | error |
DGPxxxx - 0004 | Rear Chamber Heater: Over (Purge failure) | Purge diagnostics failure | error |
DGPxxxx - 0008 | Rear Chamber Heater: Under (Purge failure) | Purge diagnostics failure | error |
DGPxxxx - 0010 | Front Chamber Heater: Over (Purge failure) | Purge diagnostics failure | error |
DGPxxxx - 0020 | Front Chamber Heater: Under (Purge failure) | Purge diagnostics failure | error |
DGPxxxx - 0040 | Power Cycled during Process - Will Resume when Warm | Purge diagnostics failure | warning |
DGPxxxx - 0080 | Process Board Failure (Purge failure) | Purge diagnostics failure | critical |
DGPxxxx - 0100 | Nebuliser Fan: Incorrect Speed (Purge failure) | Purge diagnostics failure | error |
DGPxxxx - 0200 | Destruct Fan: Incorrect Speed (Purge failure) | Purge diagnostics failure | error |
DGPxxxx - 0400 | Warning: Mist Strength Send Failure (Purge failure, may recover) | Purge diagnostics failure | warning |
DGPxxxx - 0800 | Error: Mist Strength Send Failure (Purge failure, may not recover) | Purge diagnostics failure | error |
DGPxxxx - 1000 | Destruct Valve: Fail (Purge failure) | Purge diagnostics failure | error |
DGPxxxx - 2000 | Chamber Valve: Fail (Purge failure) | Purge diagnostics failure | error |
DGPxxxx - 8000 | Verification Temperature: Out of Range (Purge failure) | Purge diagnostics failure | error |
DGS04 | Chamber valve unable to open | Chamber valve fails to open signal during cycle initiation | error |
DGS05 | Destruct valve unable to open | Destruct valve fails to open signal during disinfection cycle | error |
DGSxx - 02 | Nebuliser Fan: Unable to Turn On | System diagnostics test failure | error |
DGSxx - 03 | Exhaust Fan: Unable to Turn On | System diagnostics test failure | error |
DGSxx - 04 | Chamber Valve Unable to Open | System diagnostics test failure | error |
DGSxx - 05 | Destruct Valve Unable to Open | System diagnostics test failure | error |
DGSxx - 06 | Destruct Valve Unable to Close | System diagnostics test failure | error |
DGSxx - 07 | Chamber Valve Unable to Close | System diagnostics test failure | error |
DGSxx - 08 | Exhaust Fan: Unable to Turn Off | System diagnostics test failure | error |
DGSxx - 09 | Nebuliser Fan: Unable to Turn Off | System diagnostics test failure | error |
DGW0001 | Nebuliser cup temperature over range | Temperature sensor detects nebuliser cup exceeding max setpoint during idle or cycle | warning |
DGW0002 | Nebuliser cup temperature under range | Temperature sensor detects nebuliser cup below min setpoint during operation | warning |
DGWxxxx - 0001 | Nebuliser Cup Temperature: Over | Warm-up diagnostics failure | error |
DGWxxxx - 0002 | Nebuliser Cup Temperature: Under | Warm-up diagnostics failure | error |
DGWxxxx - 0004 | Rear Chamber Heater: Over | Warm-up diagnostics failure | error |
DGWxxxx - 0008 | Rear Chamber Heater: Under | Warm-up diagnostics failure | error |
DGWxxxx - 0010 | Front Chamber Heater: Over | Warm-up diagnostics failure | error |
DGWxxxx - 0020 | Front Chamber Heater: Under | Warm-up diagnostics failure | error |
DGWxxxx - 0080 | Process Board Failure | Warm-up diagnostics failure | critical |
DGWxxxx - 0100 | Nebuliser Fan: Incorrect Speed | Warm-up diagnostics failure | error |
DGWxxxx - 0200 | Destruct Fan: Incorrect Speed | Warm-up diagnostics failure | error |
DGWxxxx - 0400 | Warning: Mist Strength Send Failure (Warning: May Recover) | Warm-up diagnostics failure | warning |
DGWxxxx - 0800 | Error: Mist Strength Send Failure (May Not Recover) | Warm-up diagnostics failure | error |
DGWxxxx - 1000 | Destruct Valve: Fail | Warm-up diagnostics failure | error |
DGWxxxx - 2000 | Chamber Valve: Fail | Warm-up diagnostics failure | error |
DGWxxxx - 8000 | Verification Temperature: Out of Range | Warm-up diagnostics failure | error |
DR - 01 | PDS Door failed to open after being unlocked | Door unlock command issued but door did not open | error |
DR - 02 | PDS Door unable to open due to liquid detected | Door open command issued but liquid detected inside | warning |
EFP | Exhaust System Failure – Process | Exhaust system failure during disinfection process | error |
EFV | Exhaust System Failure – Verification | Exhaust system verification failure during disinfection | error |
Insert/Remove Probe constantly displayed | Probe detection failure; message remains on LCD continuously | Probe detect sensor fails to signal probe removal or remains activated continuously | warning |
Interlocks: Fail | Hardware safety interlock engaged and operation halted | Safety interlock sensor detects interference or abnormal access during operation | error |
Nebuliser Fan Incorrect Speed | Nebuliser fan RPM outside operating range | Fan speed sensor detects RPM below minimum or above maximum during operation | warning |
NL - 01 | Unable to fill nebuliser, even though liquid appears to be present | Nebuliser level failure during fill operation | error |
PF - 01 | No Liquid in Cartridge | PDS liquid failure when cartridge supposed to contain liquid | error |
PF - 02 | Too many fills on a single Cartridge | PDS liquid failure after many cycles | error |
PF - 03 | Liquid from cartridge disappeared – It didn't appear in the Reservoir | PDS liquid failure after cartridge piercing | error |
PT - 03 | Neck Needle Stuck | PDS timeout during liquid transfer | error |
PT - 04 | Rim Needle Stuck | PDS timeout during liquid transfer | error |
PT - 06 | Liquid left in cartridge reservoir after purge (Retry Available) | PDS timeout during purge cycle | warning |
PT - 07 | Liquid left in nebuliser cup after purge (Retry Available) | PDS timeout during purge cycle | warning |
PT - 08 | Too many repeated tries to empty reservoir during a purge (Retry Available) | PDS timeout with repeated purge attempts | warning |
PT - 09 | Too many repeated tries to empty nebuliser cup during a purge (Retry Available) | PDS timeout with repeated purge attempts | warning |
PT0x | Needle Stuck error in pump assembly | Pump needle position sensor detects needle in extended position longer than timeout period | warning |
SS01 | Cartridge Liquid Sensor inconsistent results | Sensor indicates no liquid then liquid detected after piercing | warning |
SS02 | Service Suggested error; needle stuck in PDS | Dispense pump assembly needle fails to retract | warning |
SS02 | Reservoir Level Sensor inconsistent results | Sensor indicates liquid when previously indicated none | warning |
Verification Temperature: Out of Range | Verification cycle temperature out of specification | Temperature sensor reading outside acceptable range during verification phase | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI