Trophon EPR: troubleshooting and error codes

The ServiceForge knowledge base holds 24 documented symptoms for the Trophon EPR, with 25 ranked causes, 29 diagnostic checks with test points and expected values, and 173 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Unable to fill nebuliser despite liquid present caution

Nebuliser level sensor cannot detect liquid during fill pump operation

Ranked causes

  1. 1. Tubing clogged or fill pump failure (common)
  2. 2. Reservoir assembly failure or low level sensor issue (occasional)
First check

Inspect fill system tubing for blockage

Test point
Fill pump tubing from reservoir to nebuliser
Expected
Tubing clear, no blockage or kinks

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air path damaged or cracked caution

Disinfection air path shows cracks or structural damage

Ranked causes

  1. 1. Impact damage or stress crack in air path tubing or components (occasional)
First check

Visually inspect air path for cracks or leaks

Test point
Air path tubing and connectors
Expected
No cracks, no visible leakage

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Board communication error caution

Communication error between device PCBs; Bxxx board failure code displayed

Ranked causes

  1. 1. I²C connector loose or contact issue on control PCB (common)
First check

Power cycle the device

Test point
Device power switch
Expected
Device powers down and back up cleanly

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Chamber door micro switches misaligned or damaged caution

Chamber door fail error; micro switches not engaging properly or lever bent

Ranked causes

  1. 1. Door micro switch lever bent or switch contacts worn (occasional)
First check

Inspect chamber door micro switch assembly

Test point
Micro switch lever and contacts
Expected
Lever straight, contacts clean and engaging

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Chamber door unable to lock or unlock caution

Chamber door locking mechanism fails to engage or disengage

Ranked causes

  1. 1. Door not closed properly or locking mechanism stuck (occasional)
First check

Verify chamber door is fully closed

Test point
Chamber door position
Expected
Door fully seated in frame

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Chamber heater temperature out of range caution

Front or rear chamber wall not reaching required operating temperature

Ranked causes

  1. 1. Chamber heater element or control failure (rare)
First check

Monitor chamber heater temperature during warm-up

Test point
Chamber temperature sensor
Expected
Temperature rises to set point within normal time

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Chamber valve seal leaking caution

Chamber valve seal failure detected; valve not sealing properly

Ranked causes

  1. 1. Chamber valve seal worn or damaged (occasional)
First check

Perform chamber pressure test

Test point
Chamber pressure sensor during cycle
Expected
Chamber maintains pressure without leakage

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Chamber valve unable to open caution

Chamber valve fails to open; device cannot execute cycle

Ranked causes

  1. 1. Chamber valve faulted or stuck (common)
First check

Check for DGS04 or DFP014 errors

Test point
Techlog error list
Expected
No DGS04/DFP014 errors after valve replacement

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Damaged or missing cable ties caution

Internal cable ties broken, missing, or not secured tightly

Ranked causes

  1. 1. Normal wear or vibration damage to cable ties (occasional)
First check

Inspect internal cable routing and cable tie condition

Test point
Cable tie locations throughout device
Expected
All cables secured with tight cable ties; no loose or broken ties

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Destruct valve seal leaking caution

Destruct valve seal failure detected; valve not sealing properly

Ranked causes

  1. 1. Destruct valve seal worn or damaged (occasional)
First check

Perform destruct valve pressure test

Test point
Destruct valve during test cycle
Expected
Valve maintains seal; no leakage detected

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Destruct valve unable to open caution

Destruct valve fails to open; disinfection cycle cannot complete

Ranked causes

  1. 1. Destruct valve faulted or stuck (common)
First check

Check for DGS05 or DFP013 errors

Test point
Techlog error list
Expected
No DGS05/DFP013 errors after valve replacement

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

H2O2 mist dosage out of range caution

Dosage sensor reading above or below allowable maximum/minimum range

Ranked causes

  1. 1. Nebuliser assembly dosage sensor or heater malfunction (rare)
First check

Run dosage test cycle and monitor sensor reading

Test point
Dosage sensor during test cycle
Expected
Dosage within min/max specification range

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Hardware safety interlock engaged patient safety

Safety interlock activated; device has ceased current operation

Ranked causes

  1. 1. Customer interference with safety mechanism (common)
First check

Verify no external interference with chamber or moving parts

Test point
Chamber and mechanical components
Expected
No obstruction; all guards in place

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nebuliser cup overheating or underheating caution

Nebuliser cup temperature outside normal operating range

Ranked causes

  1. 1. Nebuliser cup heating element failure (rare)
First check

Monitor nebuliser cup temperature during idle and cycle

Test point
Temperature sensor reading
Expected
Temperature within normal operating range

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nebuliser fan speed out of range caution

Nebuliser fan RPM outside acceptable operating range

Ranked causes

  1. 1. Nebuliser fan motor wear or bearing failure (occasional)
First check

Measure nebuliser fan RPM during operation

Test point
Fan motor speed sensor
Expected
RPM within specification range

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Needle stuck in dispense pump assembly

One or more dispense pump assembly needles stuck in extended position

Ranked causes

  1. 1. Dispense pump assembly failure (common)
First check

Check for SS02 or PT0x errors in techlog

Test point
Techlog error display
Expected
No recurring SS02/PT0x errors

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Non-persistent error displayed on LCD

Error message on Trophon EPR LCD requests user to select OK, indicating a non-persistent error from operator error or temporary unexpected event

Ranked causes

  1. 1. Operator error or temporary unexpected event (common)
First check

Review error message on LCD display

Test point
LCD display
Expected
Error message requesting OK selection

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Persistent error displayed on LCD patient safety

Error message on Trophon EPR LCD requests user to select PURGE, indicating a persistent error from component failure or serious system error

Ranked causes

  1. 1. Component failure or serious internal error (common)
First check

Review error message on LCD display

Test point
LCD display
Expected
Error message requesting PURGE selection

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Probe clamp worn or damaged caution

Mechanical probe clamp assembly shows wear or structural damage

Ranked causes

  1. 1. Normal wear or impact damage to probe clamp (occasional)
First check

Inspect probe clamp for cracks or deformation

Test point
Probe clamp assembly
Expected
Clamp intact, no cracks or severe wear

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Probe detect message constantly displayed caution

Insert/Remove Probe message remains on LCD display continuously

Ranked causes

  1. 1. Probe detect sensor failure or connector issue (occasional)
First check

Verify probe connection status

Test point
Probe connector and insertion detection
Expected
Probe detected correctly when inserted

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Probe detect sensor malfunction caution

Probe detect sensor not functioning or not detected by system during chamber operation

Ranked causes

  1. 1. Probe detect sensor assembly defective or failed (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Purge pump replacement patient safety

Purge pump requires replacement; procedure for removal and installation of replacement pump A01017

Ranked causes

  1. 1. Purge pump failure or end-of-life (unknown)
First check

Disinfectant reservoir empty

Test point
Disinfectant reservoir
Expected
Completely empty

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unknown error code 0xFFFF:FFFF caution

Unknown error displayed; control PCB battery voltage below 3V threshold

Ranked causes

  1. 1. Control PCB backup battery depleted (common)
First check

Measure control PCB battery voltage

Test point
Battery connector on control PCB
Expected
3.0V or higher, >= 3.0 V

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Verification temperature out of range caution

Verification temperature sensor detects incorrect chamber temperature during validation cycle

Ranked causes

  1. 1. Temperature sensor reading error or actual chamber temp out of range (rare)
First check

Monitor verification temperature reading across multiple cycles

Test point
Temperature sensor during validation
Expected
Temperature within specification for verification phase

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
00000001Chamber Door Open: PassedSelf-test executedinfo
00000002Chamber Door Open: FailedSelf-test executederror
00000004Insert Probe: PassedSelf-test executedinfo
00000008Insert Probe: FailedSelf-test executederror
00000010Remove Probe: PassedSelf-test executedinfo
00000020Remove Probe: FailedSelf-test executederror
00000040Chamber Door Closed: PassedSelf-test executedinfo
00000080Chamber Door Closed: FailedSelf-test executederror
00000100Valves and Fans: PassedSelf-test executedinfo
00000200Valves and Fans: FailedSelf-test executederror
00000400Button 1 Press: FailedSelf-test executederror
00000800Button 2 Press: FailedSelf-test executederror
00001000Button 3 Press: FailedSelf-test executederror
00002000Button 4 Press: FailedSelf-test executederror
00004000Start Button Press: FailedSelf-test executederror
00008000Beep: PassedSelf-test executedinfo
00010000Beep: FailedSelf-test executederror
00020000Waste Drawer Out: PassedSelf-test executedinfo
00040000Waste Drawer Out: FailedSelf-test executederror
00080000Waste Drawer In: PassedSelf-test executedinfo
00100000Waste Drawer In: FailedSelf-test executederror
00200000Cartridge Door Close: PassedSelf-test executedinfo
00400000Cartridge Door Close: FailedSelf-test executederror
00800000Needles CheckedSelf-test executedinfo
01000000Rim Needle: FailedSelf-test executederror
02000000Neck Needle: FailedSelf-test executederror
04000000Nebulant Cartridge Out Detection: PassedSelf-test executedinfo
08000000Nebulant Cartridge Out Detection: FailedSelf-test executederror
0xFFFF:FFFFUnknown error; control PCB battery criticalControl PCB backup battery voltage drops below 3.0V thresholdcritical
10000000Nebulant Cartridge In Detection: PassedSelf-test executedinfo
20000000Nebulant Cartridge In Detection: FailedSelf-test executederror
40000000Cartridge Door Open: PassedSelf-test executedinfo
80000000Cartridge Door Open: FailedSelf-test executederror
BF01I²C Bus FailureLoss of communication on I²C buserror
BF02Missing Process Board U1Process board U1 not detectederror
BF04Missing Process Board U2Process board U2 not detectederror
BF08Missing Process Board U3Process board U3 not detectederror
BF10Missing Nebuliser Board U1Nebuliser board U1 not detectederror
BF20Missing PDS Board U1PDS board U1 not detectederror
BF40Missing PDS Board U2PDS board U2 not detectederror
BFxxxBoard Failure; PCB communication errorI²C bus communication timeout or integrity error between PCBserror
Chamber Door FailChamber door micro switch failureDoor position micro switch fails to signal properly during open/close cyclewarning
Chamber Valve: Leak DetectedChamber valve seal leakingPressure decay test detects seal failure during or after valve cycleerror
CL01Chamber door unable to lockDoor lock solenoid or mechanism fails to engagewarning
CL02Chamber door unable to unlockDoor unlock solenoid or mechanism fails to disengagewarning
Destruct Valve: Leak Sealing DetectedDestruct valve seal leakingPressure decay test detects seal failure during or after valve cycleerror
DFPProcess Control Failure (While Disinfecting)Process control failure during disinfection cycleerror
DFP - 01Nebuliser Temperature: Over (Process control)Process control failure during disinfectionerror
DFP013Destruct valve failure confirmedRepeated destruct valve operation failures or seal detectionerror
DFP014Chamber valve failure confirmedRepeated chamber valve operation failures or seal detectionerror
DFP - 02Nebuliser Temperature: Under (Process control)Process control failure during disinfectionerror
DFP - 03Rear Chamber Heater: Over (Process control)Process control failure during disinfectionerror
DFP - 04Rear Chamber Heater: Under (Process control)Process control failure during disinfectionerror
DFP - 05Front Chamber Heater: Over (Process control)Process control failure during disinfectionerror
DFP - 06Front Chamber Heater: Under (Process control)Process control failure during disinfectionerror
DFP - 07Power Lost during Procedure (Process control)Process control failure during disinfectionerror
DFP - 08Board Failure (Process control)Process control failure during disinfectioncritical
DFP - 09Nebuliser Fan: Incorrect Speed (Process control)Process control failure during disinfectionerror
DFP - 10Destruct Fan: Incorrect Speed (Process control)Process control failure during disinfectionerror
DFP - 11Mist Strength: Inconsisent (Process control)Process control failure during disinfectionerror
DFP - 12Mist Strength Send Failure (Process control)Process control failure during disinfectionerror
DFP - 13Destruct Valve: Fail (Process control)Process control failure during disinfectionerror
DFP - 14Chamber Valve: Fail (Process control)Process control failure during disinfectionerror
DFP - 15Interlocks: Fail (Process control)Process control failure during disinfectionerror
DFP - 16Invalid Stage Parameter (Process control)Process control failure during disinfectionerror
DFVProcess Verification Failure (While Disinfecting)Verification failure during disinfection cycleerror
DFV - 0001Stage Timing: Over (Process verification)Verification failure during disinfectionerror
DFV - 0002Stage Timing: Under (Process verification)Verification failure during disinfectionerror
DFV0004Mist dosage over maximum allowable rangeDosage sensor reading exceeds maximum threshold during mist generationwarning
DFV0008Mist dosage under minimum allowable rangeDosage sensor reading below minimum threshold during mist generationwarning
DFV - 0010Verification Temperature: Over (Process verification)Verification failure during disinfectionerror
DFV - 0020Verification Temperature: Under (Process verification)Verification failure during disinfectionerror
DFV - 0040Mist Baseline: Over (Process verification)Verification failure during disinfectionerror
DFV - 0080Mist Baseline: Under (Process verification)Verification failure during disinfectionerror
DFV - 0100Mist Flowrate: Over (Process verification)Verification failure during disinfectionerror
DFV - 0200Mist Flowrate: Under (Process verification)Verification failure during disinfectionerror
DFV - 0400Invalid Dosage Parameter (Process verification)Verification failure during disinfectionerror
DFV - 0800Destruct Valve Sealing: Leak Detected (Process verification)Verification failure during disinfectionerror
DFV - 1000Chamber Valve Sealing: Leak Detected (Process verification)Verification failure during disinfectionerror
DFV - 4000Nebuliser Liquid Refilled (Process verification)Liquid level change during disinfection verificationinfo
DGAPxx - 01Nebuliser Cup Temperature: Over (Priming process)Diagnostic priming action process failureerror
DGAPxx - 02Nebuliser Cup Temperature: Under (Priming process)Diagnostic priming action process failureerror
DGAPxx - 03Rear Chamber Heater: Over (Priming process)Diagnostic priming action process failureerror
DGAPxx - 04Rear Chamber Heater: Under (Priming process)Diagnostic priming action process failureerror
DGAPxx - 05Front Chamber Heater: Over (Priming process)Diagnostic priming action process failureerror
DGAPxx - 06Front Chamber Heater: Under (Priming process)Diagnostic priming action process failureerror
DGAPxx - 07Power Cycled during Process (Priming process)Diagnostic priming action process failurewarning
DGAPxx - 08Process Board Failure (Priming process)Diagnostic priming action process failurecritical
DGAPxx - 09Nebuliser Fan: Incorrect Speed (Priming process)Diagnostic priming action process failureerror
DGAPxx - 10Destruct Fan: Incorrect Speed (Priming process)Diagnostic priming action process failureerror
DGAPxx - 11Mist Strength: Inconsistent (Priming process)Diagnostic priming action process failureerror
DGAPxx - 12Mist Strength Send Failure (Priming process)Diagnostic priming action process failureerror
DGAPxx - 13Destruct Valve: Fail (Priming process)Diagnostic priming action process failureerror
DGAPxx - 14Chamber Valve: Fail (Priming process)Diagnostic priming action process failureerror
DGAPxx - 15Interlocks Fail (Priming process)Diagnostic priming action process failureerror
DGAPxx - 16Invalid Stage Parameter (Priming process)Diagnostic priming action process failureerror
DGAVxxxx - 0001Stage Timing: Over (Priming verification)Diagnostic priming verification failureerror
DGAVxxxx - 0002Stage Timing: Under (Priming verification)Diagnostic priming verification failureerror
DGAVxxxx - 0004Insufficient Liquid During PrimingDiagnostic priming verification failureerror
DGAVxxxx - 0008Priming Terminated – Reset RequiredDiagnostic priming verification failurewarning
DGAVxxxx - 0010Verification Temperature: Over (Priming verification)Diagnostic priming verification failureerror
DGAVxxxx - 0020Verification Temperature: Under (Priming verification)Diagnostic priming verification failureerror
DGAVxxxx - 0040Mist Baseline: Over (Priming verification)Diagnostic priming verification failureerror
DGAVxxxx - 0080Mist Baseline: Under (Priming verification)Diagnostic priming verification failureerror
DGAVxxxx - 0100Mist Flow Rate: Over (Priming verification)Diagnostic priming verification failureerror
DGAVxxxx - 0200Mist Flow Rate: Under (Priming verification)Diagnostic priming verification failureerror
DGAVxxxx - 0400Invalid Dosage Parameter (Priming verification)Diagnostic priming verification failureerror
DGAVxxxx - 0800Destruct Valve Sealing: Leak Detected (Priming verification)Diagnostic priming verification failureerror
DGAVxxxx - 1000Chamber Valve Sealing: Leak Detected (Priming verification)Diagnostic priming verification failureerror
DGAVxxxx - 8000User Exited Priming CycleDiagnostic priming verification failureinfo
DGPxxxx - 0001Nebuliser Cup Temperature: Over (Purge failure)Purge diagnostics failureerror
DGPxxxx - 0002Nebuliser Cup Temperature: Under (Purge failure)Purge diagnostics failureerror
DGPxxxx - 0004Rear Chamber Heater: Over (Purge failure)Purge diagnostics failureerror
DGPxxxx - 0008Rear Chamber Heater: Under (Purge failure)Purge diagnostics failureerror
DGPxxxx - 0010Front Chamber Heater: Over (Purge failure)Purge diagnostics failureerror
DGPxxxx - 0020Front Chamber Heater: Under (Purge failure)Purge diagnostics failureerror
DGPxxxx - 0040Power Cycled during Process - Will Resume when WarmPurge diagnostics failurewarning
DGPxxxx - 0080Process Board Failure (Purge failure)Purge diagnostics failurecritical
DGPxxxx - 0100Nebuliser Fan: Incorrect Speed (Purge failure)Purge diagnostics failureerror
DGPxxxx - 0200Destruct Fan: Incorrect Speed (Purge failure)Purge diagnostics failureerror
DGPxxxx - 0400Warning: Mist Strength Send Failure (Purge failure, may recover)Purge diagnostics failurewarning
DGPxxxx - 0800Error: Mist Strength Send Failure (Purge failure, may not recover)Purge diagnostics failureerror
DGPxxxx - 1000Destruct Valve: Fail (Purge failure)Purge diagnostics failureerror
DGPxxxx - 2000Chamber Valve: Fail (Purge failure)Purge diagnostics failureerror
DGPxxxx - 8000Verification Temperature: Out of Range (Purge failure)Purge diagnostics failureerror
DGS04Chamber valve unable to openChamber valve fails to open signal during cycle initiationerror
DGS05Destruct valve unable to openDestruct valve fails to open signal during disinfection cycleerror
DGSxx - 02Nebuliser Fan: Unable to Turn OnSystem diagnostics test failureerror
DGSxx - 03Exhaust Fan: Unable to Turn OnSystem diagnostics test failureerror
DGSxx - 04Chamber Valve Unable to OpenSystem diagnostics test failureerror
DGSxx - 05Destruct Valve Unable to OpenSystem diagnostics test failureerror
DGSxx - 06Destruct Valve Unable to CloseSystem diagnostics test failureerror
DGSxx - 07Chamber Valve Unable to CloseSystem diagnostics test failureerror
DGSxx - 08Exhaust Fan: Unable to Turn OffSystem diagnostics test failureerror
DGSxx - 09Nebuliser Fan: Unable to Turn OffSystem diagnostics test failureerror
DGW0001Nebuliser cup temperature over rangeTemperature sensor detects nebuliser cup exceeding max setpoint during idle or cyclewarning
DGW0002Nebuliser cup temperature under rangeTemperature sensor detects nebuliser cup below min setpoint during operationwarning
DGWxxxx - 0001Nebuliser Cup Temperature: OverWarm-up diagnostics failureerror
DGWxxxx - 0002Nebuliser Cup Temperature: UnderWarm-up diagnostics failureerror
DGWxxxx - 0004Rear Chamber Heater: OverWarm-up diagnostics failureerror
DGWxxxx - 0008Rear Chamber Heater: UnderWarm-up diagnostics failureerror
DGWxxxx - 0010Front Chamber Heater: OverWarm-up diagnostics failureerror
DGWxxxx - 0020Front Chamber Heater: UnderWarm-up diagnostics failureerror
DGWxxxx - 0080Process Board FailureWarm-up diagnostics failurecritical
DGWxxxx - 0100Nebuliser Fan: Incorrect SpeedWarm-up diagnostics failureerror
DGWxxxx - 0200Destruct Fan: Incorrect SpeedWarm-up diagnostics failureerror
DGWxxxx - 0400Warning: Mist Strength Send Failure (Warning: May Recover)Warm-up diagnostics failurewarning
DGWxxxx - 0800Error: Mist Strength Send Failure (May Not Recover)Warm-up diagnostics failureerror
DGWxxxx - 1000Destruct Valve: FailWarm-up diagnostics failureerror
DGWxxxx - 2000Chamber Valve: FailWarm-up diagnostics failureerror
DGWxxxx - 8000Verification Temperature: Out of RangeWarm-up diagnostics failureerror
DR - 01PDS Door failed to open after being unlockedDoor unlock command issued but door did not openerror
DR - 02PDS Door unable to open due to liquid detectedDoor open command issued but liquid detected insidewarning
EFPExhaust System Failure – ProcessExhaust system failure during disinfection processerror
EFVExhaust System Failure – VerificationExhaust system verification failure during disinfectionerror
Insert/Remove Probe constantly displayedProbe detection failure; message remains on LCD continuouslyProbe detect sensor fails to signal probe removal or remains activated continuouslywarning
Interlocks: FailHardware safety interlock engaged and operation haltedSafety interlock sensor detects interference or abnormal access during operationerror
Nebuliser Fan Incorrect SpeedNebuliser fan RPM outside operating rangeFan speed sensor detects RPM below minimum or above maximum during operationwarning
NL - 01Unable to fill nebuliser, even though liquid appears to be presentNebuliser level failure during fill operationerror
PF - 01No Liquid in CartridgePDS liquid failure when cartridge supposed to contain liquiderror
PF - 02Too many fills on a single CartridgePDS liquid failure after many cycleserror
PF - 03Liquid from cartridge disappeared – It didn't appear in the ReservoirPDS liquid failure after cartridge piercingerror
PT - 03Neck Needle StuckPDS timeout during liquid transfererror
PT - 04Rim Needle StuckPDS timeout during liquid transfererror
PT - 06Liquid left in cartridge reservoir after purge (Retry Available)PDS timeout during purge cyclewarning
PT - 07Liquid left in nebuliser cup after purge (Retry Available)PDS timeout during purge cyclewarning
PT - 08Too many repeated tries to empty reservoir during a purge (Retry Available)PDS timeout with repeated purge attemptswarning
PT - 09Too many repeated tries to empty nebuliser cup during a purge (Retry Available)PDS timeout with repeated purge attemptswarning
PT0xNeedle Stuck error in pump assemblyPump needle position sensor detects needle in extended position longer than timeout periodwarning
SS01Cartridge Liquid Sensor inconsistent resultsSensor indicates no liquid then liquid detected after piercingwarning
SS02Service Suggested error; needle stuck in PDSDispense pump assembly needle fails to retractwarning
SS02Reservoir Level Sensor inconsistent resultsSensor indicates liquid when previously indicated nonewarning
Verification Temperature: Out of RangeVerification cycle temperature out of specificationTemperature sensor reading outside acceptable range during verification phasewarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.