Syntec VitMan: troubleshooting and error codes
The ServiceForge knowledge base holds 164 documented symptoms for the Syntec VitMan, with 298 ranked causes, 248 diagnostic checks with test points and expected values, and 100 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
50cc sensor problem displayed patient safety
Status line message appears when 50cc vacuum sensor monitoring small chamber of aspiration cassette indicates a problem. Vacuum level reads zero when vacuum should be present, or pressure present when cassette missing.
Ranked causes
- 1. 50cc sensor output voltage low (common)
- 2. A/D converter malfunction (occasional)
- 3. 50cc sensor malfunction (occasional)
- 4. Leak in aspiration system (occasional)
- 5. Pressure present on sensor port (rare)
Read 50cc sensor output voltage using A/D converter
- Test point
- 50cc sensor port
- Expected
- Voltage greater than or equal to previously recorded atmospheric voltage
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bulb A missing
Status line message displayed if bulb A is not detected when VitMan is powered up or illumination is turned on while bulb A is selected.
Ranked causes
- 1. Bulb A not inserted (common)
- 2. Voltage across FET less than 2.5 volts (occasional)
- 3. Bulb power fuse open (occasional)
- 4. A/D converter malfunction (occasional)
- 5. +24 volt power supply failure (occasional)
Visually inspect bulb A socket
- Test point
- Bulb A socket
- Expected
- Bulb A installed
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bulb A missing message displayed
Status line message displayed when bulb A is not detected during VitMan power-up
Ranked causes
- 1. Bulb A not installed or disconnected (common)
- 2. FET voltage across FET less than 2.5 volts (occasional)
- 3. A/D converter malfunction reports low voltage (occasional)
- 4. Bulb power fuse open (occasional)
- 5. +24V power supply output not present (occasional)
Verify bulb A is installed and seated
- Test point
- Bulb A connector
- Expected
- Bulb A physically present and connected
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bulb A open
Status line message displayed when bulb A is not detected. Most likely solution is to replace the bulb.
Ranked causes
- 1. Bulb A failed or missing (common)
- 2. Voltage across FET less than 0.03 volts (occasional)
- 3. Bulb power fuse open (occasional)
- 4. A/D converter malfunction (occasional)
- 5. +24 volt power supply failure (occasional)
Visually inspect bulb A
- Test point
- Bulb A socket
- Expected
- Bulb present and intact
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bulb A open message
Status line displays 'Bulb A open' when bulb A is not detected
Ranked causes
- 1. Bulb failed or loose (common)
- 2. Voltage across FET less than 0.03 volts (occasional)
- 3. A/D converter reports low voltage (occasional)
- 4. Bulb power fuse open (occasional)
- 5. +24 volt power supply failure (occasional)
Measure voltage across FET
- Test point
- FET terminals
- Expected
- 0.03 volts or greater V
- Tools
- A/D converter
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bulb A open message displayed
Status line message displayed when bulb A is not detected, indicating bulb failure
Ranked causes
- 1. Bulb A filament open or failed (common)
- 2. FET voltage less than 0.03 volts (occasional)
- 3. A/D converter malfunction reports low voltage (occasional)
- 4. Bulb power fuse open (occasional)
- 5. +24V power supply output absent or low (occasional)
Visually inspect bulb A
- Test point
- Bulb A element
- Expected
- Filament intact, no visible damage
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bulb B open
Status line message displayed indicating bulb B is not detected or has failed
Ranked causes
- 1. Bulb B failed or not installed (common)
- 2. FET voltage less than 0.03 volts (occasional)
- 3. Bulb power fuse open (occasional)
- 4. A/D converter reports low voltage (rare)
- 5. +24 volt power supply output missing (occasional)
Verify bulb B is installed and functional
- Test point
- Bulb B socket
- Expected
- Bulb present and not burnt out
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low system pressure patient safety
Status line message displayed when main pressure pump pressure falls below acceptable levels. Vitrector, ultrasound and aspiration surgical functions are disabled when detected.
Ranked causes
- 1. Vitrector operating at high cut rate consuming excessive air (occasional)
- 2. Vitrector pressure regulator improperly adjusted too high (occasional)
- 3. Air leaks in system (occasional)
- 4. Clogged air filter (occasional)
- 5. Poor compressor performance (occasional)
Measure main pump pressure
- Test point
- Main pressure pump
- Expected
- 2585 mmHg (50 psi) nominal
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Low system pressure message patient safety
Main pressure pump below acceptable levels; vitrector, ultrasound and aspiration functions disabled
Ranked causes
- 1. Vitrector high cut rate air consumption (occasional)
- 2. Improper vitrector pressure regulator adjustment (occasional)
- 3. Air leaks in system (occasional)
- 4. Clogged air filter (occasional)
- 5. Poor compressor performance (occasional)
Measure main pump pressure
- Test point
- Main pressure pump outlet
- Expected
- 2585 mmHg (50 psi) nominal; alert threshold 2172 mmHg (42 psi), Alert if below 2172 mmHg
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low system pressure message displayed patient safety
Status line displays 'Low system pressure' message; vitrector, ultrasound and aspiration surgical functions are disabled
Ranked causes
- 1. Vitrector operating at high cut rates consuming excessive air (occasional)
- 2. Improper vitrector pressure regulator adjustment (occasional)
- 3. Air leaks in system (occasional)
- 4. Clogged air filter (occasional)
- 5. Poor compressor performance (occasional)
Verify main pump pressure reading
- Test point
- System pressure display
- Expected
- 2585 mmHg (50 psi) nominal; warning at 2172 mmHg (42 psi)
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bulb A missing message
Status line shows 'Bulb A missing' if bulb A is not detected when VitMan is powered up
Ranked causes
- 1. Voltage across FET less than 2.5 volts (common)
- 2. A/D converter reports low voltage (common)
- 3. Bulb power fuse open (common)
- 4. +24 volt output not present (common)
Measure voltage across FET
- Test point
- FET terminals
- Expected
- 2.5 volts or greater V
- Tools
- A/D converter
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bulb B extremely low voltage across FET
Status line message displayed when voltage across FET is less than 0.03 volts
Ranked causes
- 1. FET voltage less than 0.03 volts (occasional)
- 2. A/D converter reports low voltage (occasional)
- 3. Bulb power fuse open (common)
- 4. Power supply +24V output failure (occasional)
Measure voltage across FET
- Test point
- FET
- Expected
- 0.03 volts or less V
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bulb B missing message displayed
Status line message appears when bulb B not detected at power-up
Ranked causes
- 1. Voltage across FET less than 2.5 volts (common)
- 2. Bulb power fuse open (common)
- 3. +24 volt output of power supply not present (occasional)
- 4. A/D converter malfunction (rare)
Measure voltage across FET at bulb circuit
- Test point
- Bulb FET drain-source
- Expected
- less than 2.5 volts if bulb missing
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bulb B open message displayed
Status line message when bulb B not detected during operation
Ranked causes
- 1. Bulb B failed or missing (common)
- 2. Voltage across FET less than 0.03 volts (occasional)
- 3. Bulb power fuse open (occasional)
- 4. +24V power supply output malfunction (occasional)
Measure voltage across bulb FET
- Test point
- Bulb FET
- Expected
- greater than 0.03 volts when bulb functional
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Diskette read problem
Message displayed when boot block, file allocation tables, root directory cannot be read, or data cable faulty, or disk drive power issue
Ranked causes
- 1. Floppy disk corruption or format issue (common)
- 2. Data cable between CPU board and disk drive is faulty (occasional)
- 3. No power present on disk drive (occasional)
- 4. Disk drive jumper missing or set to drive 0 (occasional)
Attempt to reformat floppy disk
- Test point
- Disk drive
- Expected
- Format operation succeeds and disk is readable
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Diskette read problem displayed
Status line shows read problem message, indicating inability to access disk boot block, file allocation tables or root directory
Ranked causes
- 1. Floppy disk requires formatting (occasional)
- 2. Faulty data cable between CPU board and disk drive (occasional)
- 3. No power to disk drive (occasional)
- 4. Disk drive jumper missing or selecting drive 0 (occasional)
Diskette read problem message displayed
Status line displays 'Diskette read problem' message; cannot read boot block, file allocation tables or root directory
Ranked causes
- 1. Data cable between CPU board and disk drive is faulty (occasional)
- 2. Power not present on disk drive (occasional)
- 3. Disk drive select jumper missing or set to drive 0 (occasional)
- 4. Floppy disk needs formatting (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Eye sensor problem patient safety
Eye pressure sensor indicates problem; vacuum present when off or no pressure when on; sensor output voltage low
Ranked causes
- 1. Vacuum present on sensor when air exchange function off (occasional)
- 2. No pressure present when air exchange function on (occasional)
- 5. Pressure sensor malfunction (rare)
- 6. Leak in system pressure (occasional)
Verify air exchange function status
- Test point
- air exchange surgical function
- Expected
- off
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
500cc sensor problem
Vacuum sensor monitoring vacuum pump pressure indicates a problem. Message displayed when aspiration cassette present with zero vacuum, or cassette missing with pressure present. May appear with Low vacuum level, Check cassette tubing, or Please lock cassette messages.
Ranked causes
- 2. Pressure present on sensor (occasional)
- 3. A/D converter problem (occasional)
- 4. Sensor malfunction (occasional)
Measure 500cc sensor output voltage
- Test point
- Sensor port
- Expected
- Voltage reduced by atmospheric reference should match vacuum level present
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
50cc sensor problem
Vacuum sensor monitoring small chamber of aspiration cassette indicates problem. Message displayed when cassette present with zero vacuum when vacuum should be present, or cassette missing with pressure present.
Ranked causes
- 2. Pressure present on sensor (occasional)
- 3. A/D converter problem (occasional)
- 4. Sensor malfunction (occasional)
Measure 50cc sensor output voltage
- Test point
- Sensor port
- Expected
- Voltage reduced by atmospheric reference should match vacuum level present
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air sensor problem
Pressure sensor indicates problem; vacuum present when function off or no pressure when on; sensor output voltage appears low
Ranked causes
- 1. Air sensor output voltage low (common)
- 2. Vacuum present when function disabled (occasional)
- 3. No pressure when function enabled (occasional)
Measure sensor output voltage at A/D converter
- Test point
- A/D converter input for air sensor
- Expected
- Voltage above baseline atmospheric reading
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Analog electronics board failure
Analog electronics board not functioning; signals lost or distorted from sensor inputs
Ranked causes
- 1. Loose or disconnected connector on analog electronics board (common)
- 2. PEEP EEPROM memory failure on analog electronics board (occasional)
- 3. Defective analog electronics board requiring replacement (rare)
Run Cassette test
- Test point
- System test menu
- Expected
- Test passes without errors
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bulb A FET problem
Bulb A is present and enabled but not illuminated; status line displays error message
Ranked causes
- 1. Field effect transistor (FET) failure or malfunction (common)
- 2. A/D converter reporting incorrect voltage reading (occasional)
- 3. Voltage across FET exceeds 2.5 volts (occasional)
Verify bulb A is physically present and enabled in configuration
- Test point
- Bulb A status in system menu
- Expected
- Enabled and detected
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bulb A not illuminated
Bulb A is present and enabled but does not light up; status line displays 'Bulb A FET problem' message
Ranked causes
- 1. Field effect transistor (FET) failure (common)
- 2. Voltage across FET exceeds 2.5 volts (common)
- 3. A/D converter fault (occasional)
Verify bulb A is present and enabled
- Test point
- Status line display
- Expected
- Bulb A present and enabled state confirmed
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive sensor problem
Pressure sensor for main compressor indicates system pressure is zero or output voltage is abnormally low
Ranked causes
- 1. Pressure sensor output voltage too low (common)
- 2. Pressure sensor malfunction (occasional)
- 3. Vacuum present at sensor port causing voltage offset error (rare)
Measure pressure sensor output voltage at A/D converter input
- Test point
- Pressure sensor output line
- Expected
- Voltage above zero and greater than atmospheric baseline
- Tools
- Voltmeter or multimeter
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Floppy fuse problem caution
Disk drive fuse on analog board has opened or is faulty. Disk drive power not present. Power cycle may clear if fuse is resettable.
Ranked causes
- 1. Floppy fuse open (common)
- 2. Power supply +12V output problem (occasional)
- 3. Disk drive +5V regulator failure (occasional)
Verify floppy fuse is intact
- Test point
- Analog board floppy fuse
- Expected
- Fuse intact with continuity
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot pedal problem patient safety
Status line message indicates foot pedal down switch malfunction while foot pedal position shows depression of 10% or more
Ranked causes
- 1. Foot pedal down switch internal failure (unknown)
- 2. Foot pedal cable damage or disconnection (unknown)
- 3. Foot pedal down switch sensor hardware failure (unknown)
View foot pedal condition using calibration routine
- Test point
- Service Menu > Calibration Routines > Calibrate Foot Pedal
- Expected
- Foot pedal position and switch status displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ground fault detected high voltage
Status line message indicates ultrasound handpiece is not properly grounded; no energy delivered to ungrounded handpiece
Ranked causes
- 1. Ultrasound handpiece not grounded (common)
- 2. Ground path interruption in output cable (occasional)
- 3. Ultrasound electronics board mounting hardware not grounded (occasional)
Inspect front panel ultrasound output cable ground connection
- Test point
- Front panel ultrasound output cable connector
- Expected
- Connection secure and free of corrosion
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Main compressor not running or weak output
Compressor fails to start, runs intermittently, or delivers insufficient pump pressure when configured to run
Ranked causes
- 1. Fuse F120 open or blown on main compressor board (common)
- 2. +12V supply voltage not present to compressor (UMC not configured) (occasional)
- 3. Compressor mechanical failure or blockage (rare)
Inspect fuse F120 on main compressor board
- Test point
- Fuse F120
- Expected
- Fuse element intact and not discolored
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ultrasound present message displayed
Status line message appears when ultrasound on/off button is pressed but ultrasound option is not present or not functioning
Ranked causes
- 1. Ultrasound data cable disconnected (common)
- 2. No power to ultrasound board (occasional)
- 3. Serial EEPROM problem or not detected (occasional)
Verify ultrasound data cable connection
- Test point
- Ultrasound board connector
- Expected
- Cable fully seated
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor or A/D converter problem
Status line message indicating sensor malfunction or A/D converter failure; may also indicate air exchange system leak
Ranked causes
- 1. Sensor malfunction (unknown)
- 2. A/D converter malfunction (unknown)
- 3. Air exchange system leak (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Service required message displayed patient safety
Status line message indicates serious problem detected; accompanied by other status messages identifying problem area
Ranked causes
- 1. Air exchange valve problem with excessive pressure difference (common)
- 2. Air pump unable to generate pressure (common)
- 3. Temperature sensor communication failure (occasional)
Measure pressure difference between eye pressure sensor and air exchange pump pressure sensor
- Test point
- eye pressure sensor and air exchange pump pressure sensor
- Expected
- less than 15 mmHg (close to zero mmHg for low flow rate conditions), < 15 mmHg
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vacuum pump does not operate
Vacuum pump fails to function during system operation or test procedures.
Ranked causes
- 1. Missing or open connector on pump (common)
- 2. Electronic control module disconnected or loose connection (common)
- 3. Defective pump motor (occasional)
Verify +12V CDV supply voltage presence
- Test point
- I+12V CDV connector
- Expected
- 12V present
- Tools
- DMM
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vacuum pump runs but insufficient pressure
Pump operates but fails to generate adequate vacuum pressure for system operation.
Ranked causes
- 1. Loose or disconnected inlet line on pump (common)
- 2. Clogged or kinked inlet hose (occasional)
- 3. Internal pump wear or seal degradation (rare)
Verify inlet line connections
- Test point
- Pump inlet connection and hose
- Expected
- Connection secure, no leaks
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
-12 volt fuse problem high voltage
Status line message indicates -12 volt power not present on analog board; resettable fuse has opened
Ranked causes
- 1. Fuse F (-12V) open on analog board (common)
- 2. DC-DC converter failure (-12V supply) (occasional)
Power cycle device
- Test point
- Main power switch
- Expected
- Device restarts normally without -12V fuse message
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
-22 volt fuse problem high voltage
Status line message indicates -22 volt power not present on analog board; resettable fuse opened; LCD contrast affected
Ranked causes
- 1. Fuse F (-22V) open on analog board (common)
- 2. DC-DC converter failure (-22V supply) (occasional)
Power cycle device
- Test point
- Main power switch
- Expected
- Device restarts; LCD contrast restored; no -22V fuse message
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
+12 volt fuse problem high voltage
Status line message indicates +12 volt power not present on analog board; fuse has opened (resettable)
Ranked causes
- 1. Fuse F (12V) open on analog board (common)
- 2. DC-DC converter failure (12V supply) (occasional)
Power cycle device
- Test point
- Main power switch
- Expected
- Device restarts normally without +12V fuse message
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
A/D reference voltage out of range
A/D converter reference voltage not at specified +5V (±5%)
Ranked causes
- 1. A/D reference potentiometer R32 incorrectly adjusted (common)
- 2. Fuse F111 open or blown (occasional)
Measure A/D reference voltage
- Test point
- TP101 (GND)
- Expected
- +5V ±5%, ±0.25 V
- Tools
- DMM
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air pump fuse open
Air exchange pump fuse on analog board has opened; speaker will not operate; +24V power not present on pump control circuitry
Ranked causes
- 1. Fuse F (air pump) open on analog board (common)
- 2. +24V power supply output failure (occasional)
Visually inspect air pump fuse on analog board
- Test point
- analog board fuse
- Expected
- fuse intact, not burned or open
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air pump fuse problem
Air exchange pump fuse on analog board has opened, +24V power not present on pump control circuitry. Speaker may not operate.
Ranked causes
- 1. Air pump fuse F open (analog board) (common)
- 2. +24V power supply output failure (occasional)
Visual inspection of air pump fuse on analog board
- Test point
- Analog board fuse
- Expected
- Fuse intact, not open
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bulb A not illuminating caution
Bulb A is present and enabled but does not illuminate; status line displays message indicating hardware problem.
Ranked causes
- 1. Field effect transistor (FET) failure or excessive voltage across FET (common)
- 2. A/D converter malfunction reporting false high voltage (occasional)
Measure voltage across FET
- Test point
- FET terminals
- Expected
- Less than 2.5 volts V, < 2.5 V
- Tools
- Multimeter
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bulb A problem
Status line displays 'Bulb A problem' when bulb detection logic indicates bulb A is on but illumination surgical function is off
Ranked causes
- 1. FET voltage between 0.03 and 2.5 volts with illumination off (common)
- 2. A/D converter problem reporting intermediate voltage (occasional)
Measure voltage across FET
- Test point
- FET pins
- Expected
- Between 0.03 and 2.5 volts with illumination off V, to 2.5 V
- Tools
- Multimeter
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bulb A problem message
Status line displays 'Bulb A problem' if bulb detection logic indicates bulb A is on while illumination surgical function is off
Ranked causes
- 1. FET voltage between 0.03 and 2.5 volts when illumination off (common)
- 2. A/D converter reports voltage in problem range (occasional)
Verify illumination surgical function is off
- Test point
- Illumination control
- Expected
- Illumination off
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Bulb A problem message displayed
Status line message displayed when bulb detection logic indicates bulb A is on while illumination surgical function is off
Ranked causes
- 1. FET voltage between 0.03 and 2.5 volts with illumination off (occasional)
- 2. A/D converter malfunction reports intermediate voltage (occasional)
Verify illumination function is off
- Test point
- Illumination control
- Expected
- Illumination function disabled
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bulb B excessive voltage across FET
Status line message displayed when voltage across FET exceeds 2.5 volts while bulb is enabled
Ranked causes
- 1. FET voltage greater than 2.5 volts with bulb enabled (occasional)
- 2. A/D converter reports high voltage (occasional)
Measure voltage across FET
- Test point
- FET
- Expected
- Greater than 2.5 volts
Bulb B missing
Status line displays 'Bulb B missing' when bulb is not detected at power-up
Ranked causes
- 2. A/D converter reporting low voltage (occasional)
- 4. +24 volt output of power supply not present (occasional)
Inspect bulb power fuse
- Test point
- Bulb power fuse
- Expected
- Fuse intact, continuity present
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bulb B problem message displayed
Status line message when bulb B detected as on while illumination surgical function is off
Ranked causes
- 1. Bulb detection logic false positive (unknown)
- 2. A/D converter malfunction reporting high voltage (rare)
Verify illumination surgical function switch state
- Test point
- Illumination control
- Expected
- Function off
Bulb fuse problem caution
Status line message indicates illumination fuse on analog board has opened; bulb power not present
Ranked causes
- 1. Illumination fuse opened (common)
- 2. Power supply +24V output failure (occasional)
Verify bulb power presence and fuse continuity
- Test point
- illumination fuse on analog board
- Expected
- Fuse intact and conductive
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device disabled due to high temperature patient safety
Status line displays 'Device disabled' when internal temperature reaches 55°C or higher; surgical functions become affected after ten seconds
Ranked causes
- 1. Internal device temperature at or above 55°C (unknown)
- 2. Temperature sensor malfunction or communication failure (unknown)
Monitor internal device temperature
- Test point
- Temperature sensor on analog board
- Expected
- Below 55°C
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device disabled message appears patient safety
Status line displays 'Device disabled' message; after 10 seconds surgical functions affected
Ranked causes
- 1. Internal device temperature 55 °C or higher (common)
- 2. Temperature sensor malfunction or communication failure (occasional)
Measure internal device temperature
- Test point
- Temperature sensor on analog board
- Expected
- Below 55 °C
- Tools
- Thermometer or temperature probe
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device disabled message displayed patient safety
Status line shows 'Device disabled' message; after 10 seconds surgical functions are affected. Illumination turns off, system pressure compressor and vacuum pump turn off, disabling aspiration, irrigation, vitrector and ultrasound. Air exchange remains unaffected.
Ranked causes
- 1. Internal device temperature 55 °C or higher (common)
- 2. Temperature sensor malfunction or communication failure (occasional)
Verify internal device temperature
- Test point
- Temperature sensor on analog board
- Expected
- Below 55 °C
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive fuse open
Status line message 'Drive fuse problem' displayed; main compressor (vitrector drive pressure) fuse on analog board has opened
Ranked causes
- 1. Fuse opened (common)
- 2. +12 volt power supply output problem (occasional)
Measure +12 volt output at power supply
- Test point
- +12V rail
- Expected
- +12 volts
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive fuse problem caution
Main compressor vitrector drive pressure fuse on analog board has opened; message displayed in status line
Ranked causes
- 1. Main compressor fuse on analog board is open (common)
- 2. Problem with +12 volt output of power supply (occasional)
Inspect main compressor fuse on analog board
- Test point
- Fuse on analog board
- Expected
- Fuse intact and properly seated
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive fuse problem displayed caution
Status line shows fuse problem message indicating main compressor fuse has opened
Ranked causes
- 1. Main compressor fuse opened on analog board (common)
- 2. Main compressor power not present (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive fuse problem message displayed
Status line displays 'Drive fuse problem'; main compressor fuse on analog board has opened
Ranked causes
- 1. Main compressor fuse opened (common)
- 2. +12 volt power supply output problem (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive sensor problem displayed
Status line shows sensor problem message indicating main compressor pressure sensor reading zero or low voltage
Ranked causes
- 1. System pressure sensor voltage appears low (common)
- 2. Vacuum present at sensor port (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive sensor problem message displayed
Status line displays 'Drive sensor problem'; pressure sensor indicates system pressure is zero or output voltage is low
Ranked causes
- 1. System pressure sensor output voltage is low (common)
- 2. Vacuum present on sensor port (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan fuse problem caution
Fan fuse on analog board has opened or is faulty. Fan power not present. Power cycle may clear if fuse is resettable.
Ranked causes
- 1. Fan fuse open (common)
- 2. Power supply +12V output problem (occasional)
Verify fan fuse is intact
- Test point
- Analog board fan fuse
- Expected
- Fuse intact with continuity
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
I+12V CDV present but pump fails to start
Power supply to pump is correct but motor does not respond or activate.
Ranked causes
- 1. Open or broken pump motor winding (occasional)
- 2. Control relay or driver circuit malfunction (occasional)
Verify pump motor power and ground
- Test point
- Pump motor connectors
- Expected
- I+12V CDV present on one connector, ground on other
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low system pressure warning displayed caution
Status line message 'Low system pressure' shown when main pressure pump pressure drops below acceptable level; vitrector, ultrasound and aspiration functions disabled
Ranked causes
- 1. Main pump pressure below 2172 mmHg (42 psi) (occasional)
- 2. High vitrector cut rate consuming excessive air (occasional)
Measure main pressure pump pressure
- Test point
- Main pump output
- Expected
- 2585 mmHg (50 psi) nominal; must be ≥ 2172 mmHg (42 psi), ≥ 2172 mmHg
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Low temperature
- Low temperature message displayed
- Low temperature warning message displayed
- Over temperature message displayed
- Over temperature warning
- Over temperature warning displayed
- Over temperature warning message displayed
- Pedal fuse problem detected
- Pedal fuse problem message displayed
- Shutdown at 55°C message displayed
- Unit does not power on
- Vacuum fuse problem
- Air exchange check
- Air exchange excess
- Air exchange pressure display
- Air exchange problem
- Air exchange problem detected
- Air exchange problem message displayed
- Air pump hours exceeded
- Air pump hours excess
- Air pump pressure generation failure
- Air pump problem
- Air pump problem detected
- Air pump problem detected at power-up
- Air pump problem message displayed
- Air sensor problem detected
- Anterior Frag mode active
- Aspiration threshold blocking valve
- Aspiration threshold limit message displayed
- Aspiration threshold prevents aspiration
- Aspiration threshold prevents operation
- Bulb A large hours
- Bulb A large hours message
- Bulb A large hours message displayed
- Bulb B FET problem
- Bulb B FET problem message
- Bulb B FET problem message displayed
- Bulb B large hours
- Bulb B large hours alert (voltage high)
- Bulb B large hours message
- Bulb B large hours message displayed
- Bulb B problem
- Calibrate air pump
- Calibrate foot pedal
- Calibration: Aspiration Function
- Calibration: Pressure Regulators
- Calibration: Vacuum Regulator
- Cassette full
- Cassette over full
- Cassette position bad
- Contrast adjust message displayed
- Contrast adjustment mode active
- CPU electronics board failure
- Device disabled due to high internal temperature
- Disk space warning
- Disk space warning message displayed
- Disk write protected
- Disk write protected message displayed
- Diskette is full
- Diskette is full message displayed
- Eye pressure alarm
- Floppy disk drive checkout
- High cassette level
- High pump pressure
- High system pressure
- High temperature
- LED/backlight adjust
- LED/backlight adjust message displayed
- Lift pedal to prime
- Lift pedal to prime message displayed
- Lift pedal to reflux
- Low air pump pressure
- Low air pump pressure message displayed
- Low air pump pressure warning displayed
- Low temperature warning displayed
- Low vacuum level
- Low vacuum level message displayed
- Low vacuum level warning displayed
- No handpiece present message displayed
- No keyboard heartbeat message displayed
- No stage selected message displayed
- No stages enabled message displayed
- Over temperature
- Overpressure detected
- Overpressure detected message displayed
- Overpressure detected warning
- Overpressure detected warning message displayed
- Overpressure warning displayed
- Pedal fuse problem
- Pedal fuse problem warning
- Service required (air-related)
- Shutdown at 55°C threshold
- Stabilizing message displayed
- Stabilizing message displayed at startup
- Stage indicator message displayed
- System fuse blown or open
- Temp sensor problem message displayed
- Temperature sensor problem message displayed
- Temperature sensor undetectable
- Threshold is xx mmHg message displayed
- Unit powers on but display is blank
- Vacuum pump hours exceeded
- Valve fuse failure checkout
- Valve fuse problem
104 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
+12 volt fuse problem | +12V power supply not present; fuse opened or DC-DC converter failed | +12V fuse opens or +12V DC-DC converter malfunction | error |
-12 volt fuse problem | -12V fuse on analog board opened; -12V power supply failure | -12V supply lost or fuse opened | error |
+12 volt fuse problem | 12V fuse on analog board opened; 12V power supply failure | +12V supply lost or fuse opened | error |
-12 volt fuse problem | -12V power supply not present; fuse opened or DC-DC converter failed | -12V fuse opens or -12V DC-DC converter malfunction | error |
+12 volt fuse problem | +12 volt power not available on analog board | +12V fuse opens or DC-DC converter fails | error |
-12 volt fuse problem | -12 volt power not available on analog board | -12V fuse opens or DC-DC converter fails | error |
-22 volt fuse problem | -22V power supply not present; fuse opened or DC-DC converter failed; LCD contrast affected | -22V fuse opens or -22V DC-DC converter malfunction | error |
-22 volt fuse problem | -22 volt power not available; LCD contrast lost or display unreadable | -22V fuse opens or DC-DC converter fails | error |
Air exchange problem | Pressure difference between eye pressure sensor and air exchange pump pressure sensor exceeds safe limit | Pressure difference of 15 mmHg or greater detected when air exchange valve is open | error |
Air exchange set: xxx | Air exchange surgical function is active; pressure setting is displayed | Air exchange surgical function enabled | info |
Air pump fuse problem | Air exchange pump fuse on analog board has opened; fuse must be replaced to clear error | Non-resettable fuse fails; also occurs if +24V output of power supply fails | error |
Air pump hours excess | Air pump runtime has exceeded maximum limit of 2000 hours; pneumatics module requires service | Air pump cumulative run time exceeds 2000 hours | warning |
Air pump problem | Air exchange pump unable to generate pressure during power-up self-test | Power-up test detects no pressure generation from air exchange pump when surgical function enabled | error |
Air sensor problem | Pressure sensor indicates problem; low output voltage detected | Function off and vacuum present, or function on and no pressure present; sensor output voltage less than baseline atmospheric voltage | warning |
Bulb A FET problem | Bulb A is not illuminating despite being enabled and present | Bulb A enabled but not lit or FET voltage exceeds 2.5V | error |
Bulb A FET problem | Bulb A illumination failure due to FET or A/D converter fault | Bulb A enabled but not illuminated; FET voltage exceeds 2.5 V or A/D converter malfunction detected | error |
Bulb A hours xxx | Bulb A in-use time display in hours | Illumination surgical function turned on with bulb A selected | info |
Bulb A hours xxx | Operating hours status for bulb A | When illumination surgical function is turned on with bulb A selected | info |
Bulb A large hours | Bulb A operation time exceeds safe usage limit | VitMan powered up or illumination turned on with bulb A selected and hours > 40 | warning |
Bulb A large hours | Bulb A has exceeded 40 operating hours | VitMan powered up or illumination turned on with bulb A selected and hours exceed 40 | warning |
Bulb A missing | Bulb A not detected by the system | VitMan powered up or illumination turned on with bulb A selected; bulb detection voltage below 2.5V threshold | warning |
Bulb A open | Bulb A filament detected as open or defective | Bulb A not detected after installation; FET voltage below 0.03V or fuse open or +24V supply failed | error |
Bulb A open | Bulb A not detected | Bulb A cannot be detected by monitoring circuit | error |
Bulb A problem | Bulb A appears on while illumination function is disabled | FET voltage between 0.03V and 2.5V when illumination is off; A/D converter reads erroneous voltage in this range | warning |
Bulb B FET problem | Bulb B FET circuit failure preventing bulb illumination | Bulb B installed and enabled but does not illuminate; FET not switching properly | error |
Bulb B FET problem | Bulb B present and enabled but not illuminating due to FET hardware failure | Bulb B enabled and installed but no illumination occurs | error |
Bulb B hours xxx | In-use time display for bulb B in hours | Illumination surgical function turned on with bulb B selected | info |
Bulb "B" hours xxx | Displays current operating hours of bulb B (xxx = hours) | Illumination surgical function is turned on with bulb B selected | info |
Bulb B large hours | Bulb B operating time has exceeded design life threshold | Power-up or when illumination turned on with bulb B selected and hours exceed 40 | info |
Bulb "B" large hours | Bulb B operating hours exceed safe service life | Bulb B hours exceed 40 hours at power-up or when illumination is turned on | warning |
Bulb B missing | Bulb B not detected at power-up or when illumination turned on | FET voltage below 2.5V or A/D converter fault or fuse open or +24V absent | warning |
Bulb "B" missing | Bulb B is not detected at power-up | FET voltage drops below 2.5 volts at startup, bulb not installed, fuse open, or +24V missing | warning |
Bulb B open | Bulb B not detected after power-up | FET voltage less than 0.03 volts | error |
Bulb B open | Bulb B filament open or failed | When bulb B not detected during operation | warning |
Bulb "B" open | Bulb B is not detected or has failed open circuit | FET voltage across bulb circuit drops below 0.03 volts, or bulb power fuse opens, or +24V supply fails | warning |
Bulb B problem | Bulb detection logic indicates bulb B is on while illumination surgical function is off | Bulb B enabled state does not match illumination function state | warning |
Bulb "B" problem | Bulb B detection logic indicates bulb is on while illumination function is off | Bulb detection logic detects illumination signal when surgical function is disabled | error |
Bulb fuse problem | Illumination fuse on analog board has opened or bulb power not present | Fuse opens or power supply +24V output fails | warning |
Calibrate air pump | Air exchange calibration tables are invalid | Invalid calibration data detected | warning |
Calibrate aspiration | Aspiration operation calibration tables are invalid | Invalid calibration data detected | warning |
Calibrate foot pedal | Foot pedal calibration tables are invalid | Invalid calibration data detected | warning |
Cassette full | Aspiration cassette has reached first level fullness warning threshold | Approximately 300 cc of fluid aspirated | info |
Cassette over full | Aspiration cassette has exceeded maximum capacity; vacuum pump is disabled | Approximately 325 cc of fluid aspirated | critical |
Cassette position bad | Aspiration cassette is not properly aligned in cassette chamber | Cassette inserted without proper lock knob rotation | error |
Contrast adjust | LCD contrast adjustment in progress | LEFT ARROW or RIGHT ARROW pressed while not in menu mode | info |
Currently in stage 1 | Status line message indicating operation in stage 1 mode | Stage button pressed and released when in stage 1 | info |
Currently in stage 2 | Status line message indicating operation in stage 2 mode | Stage button pressed and released when in stage 2 | info |
Currently in stage 3 | Status line message indicating operation in stage 3 mode | Stage button pressed and released when in stage 3 | info |
Currently in stage 4 | Status line message indicating operation in stage 4 mode | Stage button pressed and released when in stage 4 | info |
Currently in stage 5 | Status line message indicating operation in stage 5 mode | Stage button pressed and released when in stage 5 | info |
Currently in stage 6 | Status line message indicating operation in stage 6 mode | Stage button pressed and released when in stage 6 | info |
Currently in stage 7 | Status line message indicating operation in stage 7 mode | Stage button pressed and released when in stage 7 | info |
Device disabled | System has disabled surgical functions due to overtemperature condition | Internal device temperature reaches 55°C or higher | critical |
Device disabled | Internal device temperature is 55 °C or higher and surgical functions are being shut down | Internal temperature reaches or exceeds 55 °C | warning |
Disk available hh:mm | Minimum time remaining to record data on floppy disk in hours and minutes format | Displayed after data from previous minute is saved to disk | info |
Disk available hh:mm | Floppy disk space status message showing minimum time remaining to record data | Displayed in status line after each minute of data is saved to disk | info |
Diskette is full | Floppy disk has no remaining free space for data recording | Disk space exhausted during data save operation | error |
Diskette is full | No free space remains on floppy disk for data storage | When all disk capacity is consumed | error |
Diskette read problem | Floppy disk structure is corrupted or disk drive connection is faulty | Boot block, file allocation tables, or root directory cannot be read from disk | error |
Diskette read problem | Unable to read boot block, file allocation tables or root directory on floppy disk | Boot block, file allocation tables or root directory cannot be accessed | error |
Disk space warning | Disk available time below threshold | Minimum time remaining less than 30 minutes | warning |
Disk space warning | Available disk space is critically low | When minimum time remaining is less than 30 minutes | warning |
Disk space warning | Minimum time remaining on floppy disk is less than 30 minutes | Insufficient free space on floppy disk | warning |
Disk space warning | Warning message displayed when floppy disk free space is low | Minimum time remaining is less than 30 minutes | warning |
Disk write protected | Cannot save data to floppy disk due to write protection | Checked when attempting to save data to disk | error |
Drive fuse problem | Main compressor fuse on analog board has opened; no power to compressor | When main compressor fuse opens due to overcurrent or power surge | error |
Drive sensor problem | Pressure sensor monitoring main compressor indicates fault condition or zero pressure reading | Sensor output voltage indicates zero pressure or voltage appears low | warning |
Eye pressure alarm | Current eye pressure not within 5 mmHg of set point during air exchange surgical function | Air exchange surgical function active and eye pressure deviates from set point | warning |
Eye sensor problem | Eye pressure sensor is indicating a malfunction or vacuum/pressure condition error | Air exchange function off with vacuum present, or on with no pressure present; sensor voltage low | error |
Fan fuse problem | Fan fuse on analog board has opened or fan power is not present | Fan fuse opens; +12V power supply output fails | warning |
Floppy fuse problem | Disk drive fuse on analog board has opened or disk drive power is not present | Floppy fuse opens; +12V or +5V power fails | warning |
Foot pedal is active | Foot pedal is active; several user actions are prohibited | Foot pedal engaged during operation | info |
Foot pedal problem | Foot pedal down switch indicates up position while foot pedal position shows ≥10% depression | Foot pedal down switch malfunction or cable/connector failure | warning |
Ground fault detected | Ultrasound handpiece is not properly grounded | Ground path interruption or handpiece not seated | error |
High cassette level | Aspiration cassette fluid level is high (first warning level) | Approximately 250 cc of fluid aspirated into cassette | warning |
High pump pressure | Air exchange pump pressure exceeds set point by more than 8 mmHg | Air exchange surgical function active and pump pressure high | warning |
High system pressure | Main pressure pump pressure exceeds upper limit (2844 mmHg / 55 psi) | Main pump pressure above 2844 mmHg (55 psi) | warning |
High temperature | Internal device temperature is 50 °C or higher | Internal device temperature reaches or exceeds 50 °C | warning |
Hold to enter stage 4 | Stage four is ready to be entered | Stage button pressed and held when stage four will be entered upon release | info |
Now saving stage 3 | System is saving current machine state to stage three memory | Stage button is pressed and held while in stage three save mode | info |
Now saving stage 4 | System is saving current machine state to stage four memory | Stage button is pressed and held while in stage four save mode | info |
Now saving stage 5 | System is saving current machine state to stage five memory | Stage button is pressed and held while in stage five save mode | info |
Now saving stage 6 | System is saving current machine state to stage six memory | Stage button is pressed and held while in stage six save mode | info |
Now saving stage 7 | System is saving current machine state to stage seven memory | Stage button is pressed and held while in stage seven save mode | info |
Overpressure detected | Eye pressure exceeds the configured warning pressure limit | Current eye pressure exceeds warning pressure limit | warning |
Over temperature | Internal device temperature has reached 53°C or higher; warning condition only | Internal temperature sensor reading >= 53°C | warning |
Pedal fuse problem | Foot pedal fuse on CPU board has opened; foot pedal power supply failure | Foot pedal fuse opens due to overcurrent or fault condition | warning |
Service required | Serious problem detected requiring VitMan repair | Various hardware problems detected during operation or startup | critical |
Shutdown at 55°C | Internal device temperature at or above 53°C but below 55°C | Temperature sensor on analog board reports internal temperature >= 53°C | warning |
Stabilizing... | Hardware problem detected during pressure sensor thermal stabilization phase; device non-functional | Hardware malfunction detected while pressure sensors are thermally stabilizing at power-up | critical |
Stage 3 selected | Stage 3 is selected and ready to enter | Stage button pressed and held, then released | info |
Stage 4 selected | Stage 4 is selected and ready to enter | Stage button pressed and held, then released | info |
Stage 5 selected | Stage 5 is selected and ready to enter | Stage button pressed and held, then released | info |
Stage 6 selected | Stage 6 is selected and ready to enter | Stage button pressed and held, then released | info |
Stage 7 selected | Stage 7 is selected and ready to enter | Stage button pressed and held, then released | info |
Temp sensor problem | Temperature sensor is undetectable or not communicating properly | Temperature sensor fails to establish communication with system | warning |
Threshold is xx mmHg | Aspiration operation prevented by threshold setting; xx shows current threshold value in mmHg | Aspiration valve cannot open because vacuum level does not meet threshold requirement | warning |
Vacuum fuse problem | Vacuum pump fuse on analog board has opened | Vacuum pump fuse opens; vacuum pump power not present | warning |
Vacuum pump hours | Vacuum pump runtime has exceeded maximum service interval | Pump operating hours exceed threshold | info |
Valve fuse problem | Valve fuse on analog board has opened | Valve fuse opens due to overcurrent or age | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI