Syntec VitMan: troubleshooting and error codes

The ServiceForge knowledge base holds 164 documented symptoms for the Syntec VitMan, with 298 ranked causes, 248 diagnostic checks with test points and expected values, and 100 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

50cc sensor problem displayed patient safety

Status line message appears when 50cc vacuum sensor monitoring small chamber of aspiration cassette indicates a problem. Vacuum level reads zero when vacuum should be present, or pressure present when cassette missing.

Ranked causes

  1. 1. 50cc sensor output voltage low (common)
  2. 2. A/D converter malfunction (occasional)
  3. 3. 50cc sensor malfunction (occasional)
  4. 4. Leak in aspiration system (occasional)
  5. 5. Pressure present on sensor port (rare)
First check

Read 50cc sensor output voltage using A/D converter

Test point
50cc sensor port
Expected
Voltage greater than or equal to previously recorded atmospheric voltage

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bulb A missing

Status line message displayed if bulb A is not detected when VitMan is powered up or illumination is turned on while bulb A is selected.

Ranked causes

  1. 1. Bulb A not inserted (common)
  2. 2. Voltage across FET less than 2.5 volts (occasional)
  3. 3. Bulb power fuse open (occasional)
  4. 4. A/D converter malfunction (occasional)
  5. 5. +24 volt power supply failure (occasional)
First check

Visually inspect bulb A socket

Test point
Bulb A socket
Expected
Bulb A installed

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bulb A missing message displayed

Status line message displayed when bulb A is not detected during VitMan power-up

Ranked causes

  1. 1. Bulb A not installed or disconnected (common)
  2. 2. FET voltage across FET less than 2.5 volts (occasional)
  3. 3. A/D converter malfunction reports low voltage (occasional)
  4. 4. Bulb power fuse open (occasional)
  5. 5. +24V power supply output not present (occasional)
First check

Verify bulb A is installed and seated

Test point
Bulb A connector
Expected
Bulb A physically present and connected

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bulb A open

Status line message displayed when bulb A is not detected. Most likely solution is to replace the bulb.

Ranked causes

  1. 1. Bulb A failed or missing (common)
  2. 2. Voltage across FET less than 0.03 volts (occasional)
  3. 3. Bulb power fuse open (occasional)
  4. 4. A/D converter malfunction (occasional)
  5. 5. +24 volt power supply failure (occasional)
First check

Visually inspect bulb A

Test point
Bulb A socket
Expected
Bulb present and intact

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bulb A open message

Status line displays 'Bulb A open' when bulb A is not detected

Ranked causes

  1. 1. Bulb failed or loose (common)
  2. 2. Voltage across FET less than 0.03 volts (occasional)
  3. 3. A/D converter reports low voltage (occasional)
  4. 4. Bulb power fuse open (occasional)
  5. 5. +24 volt power supply failure (occasional)
First check

Measure voltage across FET

Test point
FET terminals
Expected
0.03 volts or greater V
Tools
A/D converter

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bulb A open message displayed

Status line message displayed when bulb A is not detected, indicating bulb failure

Ranked causes

  1. 1. Bulb A filament open or failed (common)
  2. 2. FET voltage less than 0.03 volts (occasional)
  3. 3. A/D converter malfunction reports low voltage (occasional)
  4. 4. Bulb power fuse open (occasional)
  5. 5. +24V power supply output absent or low (occasional)
First check

Visually inspect bulb A

Test point
Bulb A element
Expected
Filament intact, no visible damage

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bulb B open

Status line message displayed indicating bulb B is not detected or has failed

Ranked causes

  1. 1. Bulb B failed or not installed (common)
  2. 2. FET voltage less than 0.03 volts (occasional)
  3. 3. Bulb power fuse open (occasional)
  4. 4. A/D converter reports low voltage (rare)
  5. 5. +24 volt power supply output missing (occasional)
First check

Verify bulb B is installed and functional

Test point
Bulb B socket
Expected
Bulb present and not burnt out

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low system pressure patient safety

Status line message displayed when main pressure pump pressure falls below acceptable levels. Vitrector, ultrasound and aspiration surgical functions are disabled when detected.

Ranked causes

  1. 1. Vitrector operating at high cut rate consuming excessive air (occasional)
  2. 2. Vitrector pressure regulator improperly adjusted too high (occasional)
  3. 3. Air leaks in system (occasional)
  4. 4. Clogged air filter (occasional)
  5. 5. Poor compressor performance (occasional)
First check

Measure main pump pressure

Test point
Main pressure pump
Expected
2585 mmHg (50 psi) nominal

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Low system pressure message patient safety

Main pressure pump below acceptable levels; vitrector, ultrasound and aspiration functions disabled

Ranked causes

  1. 1. Vitrector high cut rate air consumption (occasional)
  2. 2. Improper vitrector pressure regulator adjustment (occasional)
  3. 3. Air leaks in system (occasional)
  4. 4. Clogged air filter (occasional)
  5. 5. Poor compressor performance (occasional)
First check

Measure main pump pressure

Test point
Main pressure pump outlet
Expected
2585 mmHg (50 psi) nominal; alert threshold 2172 mmHg (42 psi), Alert if below 2172 mmHg

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low system pressure message displayed patient safety

Status line displays 'Low system pressure' message; vitrector, ultrasound and aspiration surgical functions are disabled

Ranked causes

  1. 1. Vitrector operating at high cut rates consuming excessive air (occasional)
  2. 2. Improper vitrector pressure regulator adjustment (occasional)
  3. 3. Air leaks in system (occasional)
  4. 4. Clogged air filter (occasional)
  5. 5. Poor compressor performance (occasional)
First check

Verify main pump pressure reading

Test point
System pressure display
Expected
2585 mmHg (50 psi) nominal; warning at 2172 mmHg (42 psi)

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bulb A missing message

Status line shows 'Bulb A missing' if bulb A is not detected when VitMan is powered up

Ranked causes

  1. 1. Voltage across FET less than 2.5 volts (common)
  2. 2. A/D converter reports low voltage (common)
  3. 3. Bulb power fuse open (common)
  4. 4. +24 volt output not present (common)
First check

Measure voltage across FET

Test point
FET terminals
Expected
2.5 volts or greater V
Tools
A/D converter

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bulb B extremely low voltage across FET

Status line message displayed when voltage across FET is less than 0.03 volts

Ranked causes

  1. 1. FET voltage less than 0.03 volts (occasional)
  2. 2. A/D converter reports low voltage (occasional)
  3. 3. Bulb power fuse open (common)
  4. 4. Power supply +24V output failure (occasional)
First check

Measure voltage across FET

Test point
FET
Expected
0.03 volts or less V

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bulb B missing message displayed

Status line message appears when bulb B not detected at power-up

Ranked causes

  1. 1. Voltage across FET less than 2.5 volts (common)
  2. 2. Bulb power fuse open (common)
  3. 3. +24 volt output of power supply not present (occasional)
  4. 4. A/D converter malfunction (rare)
First check

Measure voltage across FET at bulb circuit

Test point
Bulb FET drain-source
Expected
less than 2.5 volts if bulb missing

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bulb B open message displayed

Status line message when bulb B not detected during operation

Ranked causes

  1. 1. Bulb B failed or missing (common)
  2. 2. Voltage across FET less than 0.03 volts (occasional)
  3. 3. Bulb power fuse open (occasional)
  4. 4. +24V power supply output malfunction (occasional)
First check

Measure voltage across bulb FET

Test point
Bulb FET
Expected
greater than 0.03 volts when bulb functional

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Diskette read problem

Message displayed when boot block, file allocation tables, root directory cannot be read, or data cable faulty, or disk drive power issue

Ranked causes

  1. 1. Floppy disk corruption or format issue (common)
  2. 2. Data cable between CPU board and disk drive is faulty (occasional)
  3. 3. No power present on disk drive (occasional)
  4. 4. Disk drive jumper missing or set to drive 0 (occasional)
First check

Attempt to reformat floppy disk

Test point
Disk drive
Expected
Format operation succeeds and disk is readable

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Diskette read problem displayed

Status line shows read problem message, indicating inability to access disk boot block, file allocation tables or root directory

Ranked causes

  1. 1. Floppy disk requires formatting (occasional)
  2. 2. Faulty data cable between CPU board and disk drive (occasional)
  3. 3. No power to disk drive (occasional)
  4. 4. Disk drive jumper missing or selecting drive 0 (occasional)

Diskette read problem message displayed

Status line displays 'Diskette read problem' message; cannot read boot block, file allocation tables or root directory

Ranked causes

  1. 1. Data cable between CPU board and disk drive is faulty (occasional)
  2. 2. Power not present on disk drive (occasional)
  3. 3. Disk drive select jumper missing or set to drive 0 (occasional)
  4. 4. Floppy disk needs formatting (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Eye sensor problem patient safety

Eye pressure sensor indicates problem; vacuum present when off or no pressure when on; sensor output voltage low

Ranked causes

  1. 1. Vacuum present on sensor when air exchange function off (occasional)
  2. 2. No pressure present when air exchange function on (occasional)
  3. 5. Pressure sensor malfunction (rare)
  4. 6. Leak in system pressure (occasional)
First check

Verify air exchange function status

Test point
air exchange surgical function
Expected
off

4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

500cc sensor problem

Vacuum sensor monitoring vacuum pump pressure indicates a problem. Message displayed when aspiration cassette present with zero vacuum, or cassette missing with pressure present. May appear with Low vacuum level, Check cassette tubing, or Please lock cassette messages.

Ranked causes

  1. 2. Pressure present on sensor (occasional)
  2. 3. A/D converter problem (occasional)
  3. 4. Sensor malfunction (occasional)
First check

Measure 500cc sensor output voltage

Test point
Sensor port
Expected
Voltage reduced by atmospheric reference should match vacuum level present

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

50cc sensor problem

Vacuum sensor monitoring small chamber of aspiration cassette indicates problem. Message displayed when cassette present with zero vacuum when vacuum should be present, or cassette missing with pressure present.

Ranked causes

  1. 2. Pressure present on sensor (occasional)
  2. 3. A/D converter problem (occasional)
  3. 4. Sensor malfunction (occasional)
First check

Measure 50cc sensor output voltage

Test point
Sensor port
Expected
Voltage reduced by atmospheric reference should match vacuum level present

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air sensor problem

Pressure sensor indicates problem; vacuum present when function off or no pressure when on; sensor output voltage appears low

Ranked causes

  1. 1. Air sensor output voltage low (common)
  2. 2. Vacuum present when function disabled (occasional)
  3. 3. No pressure when function enabled (occasional)
First check

Measure sensor output voltage at A/D converter

Test point
A/D converter input for air sensor
Expected
Voltage above baseline atmospheric reading

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Analog electronics board failure

Analog electronics board not functioning; signals lost or distorted from sensor inputs

Ranked causes

  1. 1. Loose or disconnected connector on analog electronics board (common)
  2. 2. PEEP EEPROM memory failure on analog electronics board (occasional)
  3. 3. Defective analog electronics board requiring replacement (rare)
First check

Run Cassette test

Test point
System test menu
Expected
Test passes without errors

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bulb A FET problem

Bulb A is present and enabled but not illuminated; status line displays error message

Ranked causes

  1. 1. Field effect transistor (FET) failure or malfunction (common)
  2. 2. A/D converter reporting incorrect voltage reading (occasional)
  3. 3. Voltage across FET exceeds 2.5 volts (occasional)
First check

Verify bulb A is physically present and enabled in configuration

Test point
Bulb A status in system menu
Expected
Enabled and detected

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bulb A not illuminated

Bulb A is present and enabled but does not light up; status line displays 'Bulb A FET problem' message

Ranked causes

  1. 1. Field effect transistor (FET) failure (common)
  2. 2. Voltage across FET exceeds 2.5 volts (common)
  3. 3. A/D converter fault (occasional)
First check

Verify bulb A is present and enabled

Test point
Status line display
Expected
Bulb A present and enabled state confirmed

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive sensor problem

Pressure sensor for main compressor indicates system pressure is zero or output voltage is abnormally low

Ranked causes

  1. 1. Pressure sensor output voltage too low (common)
  2. 2. Pressure sensor malfunction (occasional)
  3. 3. Vacuum present at sensor port causing voltage offset error (rare)
First check

Measure pressure sensor output voltage at A/D converter input

Test point
Pressure sensor output line
Expected
Voltage above zero and greater than atmospheric baseline
Tools
Voltmeter or multimeter

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Floppy fuse problem caution

Disk drive fuse on analog board has opened or is faulty. Disk drive power not present. Power cycle may clear if fuse is resettable.

Ranked causes

  1. 1. Floppy fuse open (common)
  2. 2. Power supply +12V output problem (occasional)
  3. 3. Disk drive +5V regulator failure (occasional)
First check

Verify floppy fuse is intact

Test point
Analog board floppy fuse
Expected
Fuse intact with continuity

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Foot pedal problem patient safety

Status line message indicates foot pedal down switch malfunction while foot pedal position shows depression of 10% or more

Ranked causes

  1. 1. Foot pedal down switch internal failure (unknown)
  2. 2. Foot pedal cable damage or disconnection (unknown)
  3. 3. Foot pedal down switch sensor hardware failure (unknown)
First check

View foot pedal condition using calibration routine

Test point
Service Menu > Calibration Routines > Calibrate Foot Pedal
Expected
Foot pedal position and switch status displayed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ground fault detected high voltage

Status line message indicates ultrasound handpiece is not properly grounded; no energy delivered to ungrounded handpiece

Ranked causes

  1. 1. Ultrasound handpiece not grounded (common)
  2. 2. Ground path interruption in output cable (occasional)
  3. 3. Ultrasound electronics board mounting hardware not grounded (occasional)
First check

Inspect front panel ultrasound output cable ground connection

Test point
Front panel ultrasound output cable connector
Expected
Connection secure and free of corrosion

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Main compressor not running or weak output

Compressor fails to start, runs intermittently, or delivers insufficient pump pressure when configured to run

Ranked causes

  1. 1. Fuse F120 open or blown on main compressor board (common)
  2. 2. +12V supply voltage not present to compressor (UMC not configured) (occasional)
  3. 3. Compressor mechanical failure or blockage (rare)
First check

Inspect fuse F120 on main compressor board

Test point
Fuse F120
Expected
Fuse element intact and not discolored

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No ultrasound present message displayed

Status line message appears when ultrasound on/off button is pressed but ultrasound option is not present or not functioning

Ranked causes

  1. 1. Ultrasound data cable disconnected (common)
  2. 2. No power to ultrasound board (occasional)
  3. 3. Serial EEPROM problem or not detected (occasional)
First check

Verify ultrasound data cable connection

Test point
Ultrasound board connector
Expected
Cable fully seated

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor or A/D converter problem

Status line message indicating sensor malfunction or A/D converter failure; may also indicate air exchange system leak

Ranked causes

  1. 1. Sensor malfunction (unknown)
  2. 2. A/D converter malfunction (unknown)
  3. 3. Air exchange system leak (unknown)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Service required message displayed patient safety

Status line message indicates serious problem detected; accompanied by other status messages identifying problem area

Ranked causes

  1. 1. Air exchange valve problem with excessive pressure difference (common)
  2. 2. Air pump unable to generate pressure (common)
  3. 3. Temperature sensor communication failure (occasional)
First check

Measure pressure difference between eye pressure sensor and air exchange pump pressure sensor

Test point
eye pressure sensor and air exchange pump pressure sensor
Expected
less than 15 mmHg (close to zero mmHg for low flow rate conditions), < 15 mmHg

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Vacuum pump does not operate

Vacuum pump fails to function during system operation or test procedures.

Ranked causes

  1. 1. Missing or open connector on pump (common)
  2. 2. Electronic control module disconnected or loose connection (common)
  3. 3. Defective pump motor (occasional)
First check

Verify +12V CDV supply voltage presence

Test point
I+12V CDV connector
Expected
12V present
Tools
DMM

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Vacuum pump runs but insufficient pressure

Pump operates but fails to generate adequate vacuum pressure for system operation.

Ranked causes

  1. 1. Loose or disconnected inlet line on pump (common)
  2. 2. Clogged or kinked inlet hose (occasional)
  3. 3. Internal pump wear or seal degradation (rare)
First check

Verify inlet line connections

Test point
Pump inlet connection and hose
Expected
Connection secure, no leaks

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

-12 volt fuse problem high voltage

Status line message indicates -12 volt power not present on analog board; resettable fuse has opened

Ranked causes

  1. 1. Fuse F (-12V) open on analog board (common)
  2. 2. DC-DC converter failure (-12V supply) (occasional)
First check

Power cycle device

Test point
Main power switch
Expected
Device restarts normally without -12V fuse message

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

-22 volt fuse problem high voltage

Status line message indicates -22 volt power not present on analog board; resettable fuse opened; LCD contrast affected

Ranked causes

  1. 1. Fuse F (-22V) open on analog board (common)
  2. 2. DC-DC converter failure (-22V supply) (occasional)
First check

Power cycle device

Test point
Main power switch
Expected
Device restarts; LCD contrast restored; no -22V fuse message

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

+12 volt fuse problem high voltage

Status line message indicates +12 volt power not present on analog board; fuse has opened (resettable)

Ranked causes

  1. 1. Fuse F (12V) open on analog board (common)
  2. 2. DC-DC converter failure (12V supply) (occasional)
First check

Power cycle device

Test point
Main power switch
Expected
Device restarts normally without +12V fuse message

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

A/D reference voltage out of range

A/D converter reference voltage not at specified +5V (±5%)

Ranked causes

  1. 1. A/D reference potentiometer R32 incorrectly adjusted (common)
  2. 2. Fuse F111 open or blown (occasional)
First check

Measure A/D reference voltage

Test point
TP101 (GND)
Expected
+5V ±5%, ±0.25 V
Tools
DMM

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air pump fuse open

Air exchange pump fuse on analog board has opened; speaker will not operate; +24V power not present on pump control circuitry

Ranked causes

  1. 1. Fuse F (air pump) open on analog board (common)
  2. 2. +24V power supply output failure (occasional)
First check

Visually inspect air pump fuse on analog board

Test point
analog board fuse
Expected
fuse intact, not burned or open

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air pump fuse problem

Air exchange pump fuse on analog board has opened, +24V power not present on pump control circuitry. Speaker may not operate.

Ranked causes

  1. 1. Air pump fuse F open (analog board) (common)
  2. 2. +24V power supply output failure (occasional)
First check

Visual inspection of air pump fuse on analog board

Test point
Analog board fuse
Expected
Fuse intact, not open

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bulb A not illuminating caution

Bulb A is present and enabled but does not illuminate; status line displays message indicating hardware problem.

Ranked causes

  1. 1. Field effect transistor (FET) failure or excessive voltage across FET (common)
  2. 2. A/D converter malfunction reporting false high voltage (occasional)
First check

Measure voltage across FET

Test point
FET terminals
Expected
Less than 2.5 volts V, < 2.5 V
Tools
Multimeter

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bulb A problem

Status line displays 'Bulb A problem' when bulb detection logic indicates bulb A is on but illumination surgical function is off

Ranked causes

  1. 1. FET voltage between 0.03 and 2.5 volts with illumination off (common)
  2. 2. A/D converter problem reporting intermediate voltage (occasional)
First check

Measure voltage across FET

Test point
FET pins
Expected
Between 0.03 and 2.5 volts with illumination off V, to 2.5 V
Tools
Multimeter

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bulb A problem message

Status line displays 'Bulb A problem' if bulb detection logic indicates bulb A is on while illumination surgical function is off

Ranked causes

  1. 1. FET voltage between 0.03 and 2.5 volts when illumination off (common)
  2. 2. A/D converter reports voltage in problem range (occasional)
First check

Verify illumination surgical function is off

Test point
Illumination control
Expected
Illumination off

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Bulb A problem message displayed

Status line message displayed when bulb detection logic indicates bulb A is on while illumination surgical function is off

Ranked causes

  1. 1. FET voltage between 0.03 and 2.5 volts with illumination off (occasional)
  2. 2. A/D converter malfunction reports intermediate voltage (occasional)
First check

Verify illumination function is off

Test point
Illumination control
Expected
Illumination function disabled

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bulb B excessive voltage across FET

Status line message displayed when voltage across FET exceeds 2.5 volts while bulb is enabled

Ranked causes

  1. 1. FET voltage greater than 2.5 volts with bulb enabled (occasional)
  2. 2. A/D converter reports high voltage (occasional)
First check

Measure voltage across FET

Test point
FET
Expected
Greater than 2.5 volts

Bulb B missing

Status line displays 'Bulb B missing' when bulb is not detected at power-up

Ranked causes

  1. 2. A/D converter reporting low voltage (occasional)
  2. 4. +24 volt output of power supply not present (occasional)
First check

Inspect bulb power fuse

Test point
Bulb power fuse
Expected
Fuse intact, continuity present

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bulb B problem message displayed

Status line message when bulb B detected as on while illumination surgical function is off

Ranked causes

  1. 1. Bulb detection logic false positive (unknown)
  2. 2. A/D converter malfunction reporting high voltage (rare)
First check

Verify illumination surgical function switch state

Test point
Illumination control
Expected
Function off

Bulb fuse problem caution

Status line message indicates illumination fuse on analog board has opened; bulb power not present

Ranked causes

  1. 1. Illumination fuse opened (common)
  2. 2. Power supply +24V output failure (occasional)
First check

Verify bulb power presence and fuse continuity

Test point
illumination fuse on analog board
Expected
Fuse intact and conductive

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device disabled due to high temperature patient safety

Status line displays 'Device disabled' when internal temperature reaches 55°C or higher; surgical functions become affected after ten seconds

Ranked causes

  1. 1. Internal device temperature at or above 55°C (unknown)
  2. 2. Temperature sensor malfunction or communication failure (unknown)
First check

Monitor internal device temperature

Test point
Temperature sensor on analog board
Expected
Below 55°C

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device disabled message appears patient safety

Status line displays 'Device disabled' message; after 10 seconds surgical functions affected

Ranked causes

  1. 1. Internal device temperature 55 °C or higher (common)
  2. 2. Temperature sensor malfunction or communication failure (occasional)
First check

Measure internal device temperature

Test point
Temperature sensor on analog board
Expected
Below 55 °C
Tools
Thermometer or temperature probe

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device disabled message displayed patient safety

Status line shows 'Device disabled' message; after 10 seconds surgical functions are affected. Illumination turns off, system pressure compressor and vacuum pump turn off, disabling aspiration, irrigation, vitrector and ultrasound. Air exchange remains unaffected.

Ranked causes

  1. 1. Internal device temperature 55 °C or higher (common)
  2. 2. Temperature sensor malfunction or communication failure (occasional)
First check

Verify internal device temperature

Test point
Temperature sensor on analog board
Expected
Below 55 °C

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive fuse open

Status line message 'Drive fuse problem' displayed; main compressor (vitrector drive pressure) fuse on analog board has opened

Ranked causes

  1. 1. Fuse opened (common)
  2. 2. +12 volt power supply output problem (occasional)
First check

Measure +12 volt output at power supply

Test point
+12V rail
Expected
+12 volts

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive fuse problem caution

Main compressor vitrector drive pressure fuse on analog board has opened; message displayed in status line

Ranked causes

  1. 1. Main compressor fuse on analog board is open (common)
  2. 2. Problem with +12 volt output of power supply (occasional)
First check

Inspect main compressor fuse on analog board

Test point
Fuse on analog board
Expected
Fuse intact and properly seated

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive fuse problem displayed caution

Status line shows fuse problem message indicating main compressor fuse has opened

Ranked causes

  1. 1. Main compressor fuse opened on analog board (common)
  2. 2. Main compressor power not present (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive fuse problem message displayed

Status line displays 'Drive fuse problem'; main compressor fuse on analog board has opened

Ranked causes

  1. 1. Main compressor fuse opened (common)
  2. 2. +12 volt power supply output problem (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive sensor problem displayed

Status line shows sensor problem message indicating main compressor pressure sensor reading zero or low voltage

Ranked causes

  1. 1. System pressure sensor voltage appears low (common)
  2. 2. Vacuum present at sensor port (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive sensor problem message displayed

Status line displays 'Drive sensor problem'; pressure sensor indicates system pressure is zero or output voltage is low

Ranked causes

  1. 1. System pressure sensor output voltage is low (common)
  2. 2. Vacuum present on sensor port (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fan fuse problem caution

Fan fuse on analog board has opened or is faulty. Fan power not present. Power cycle may clear if fuse is resettable.

Ranked causes

  1. 1. Fan fuse open (common)
  2. 2. Power supply +12V output problem (occasional)
First check

Verify fan fuse is intact

Test point
Analog board fan fuse
Expected
Fuse intact with continuity

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

I+12V CDV present but pump fails to start

Power supply to pump is correct but motor does not respond or activate.

Ranked causes

  1. 1. Open or broken pump motor winding (occasional)
  2. 2. Control relay or driver circuit malfunction (occasional)
First check

Verify pump motor power and ground

Test point
Pump motor connectors
Expected
I+12V CDV present on one connector, ground on other

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low system pressure warning displayed caution

Status line message 'Low system pressure' shown when main pressure pump pressure drops below acceptable level; vitrector, ultrasound and aspiration functions disabled

Ranked causes

  1. 1. Main pump pressure below 2172 mmHg (42 psi) (occasional)
  2. 2. High vitrector cut rate consuming excessive air (occasional)
First check

Measure main pressure pump pressure

Test point
Main pump output
Expected
2585 mmHg (50 psi) nominal; must be ≥ 2172 mmHg (42 psi), ≥ 2172 mmHg

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Low temperature
  • Low temperature message displayed
  • Low temperature warning message displayed
  • Over temperature message displayed
  • Over temperature warning
  • Over temperature warning displayed
  • Over temperature warning message displayed
  • Pedal fuse problem detected
  • Pedal fuse problem message displayed
  • Shutdown at 55°C message displayed
  • Unit does not power on
  • Vacuum fuse problem
  • Air exchange check
  • Air exchange excess
  • Air exchange pressure display
  • Air exchange problem
  • Air exchange problem detected
  • Air exchange problem message displayed
  • Air pump hours exceeded
  • Air pump hours excess
  • Air pump pressure generation failure
  • Air pump problem
  • Air pump problem detected
  • Air pump problem detected at power-up
  • Air pump problem message displayed
  • Air sensor problem detected
  • Anterior Frag mode active
  • Aspiration threshold blocking valve
  • Aspiration threshold limit message displayed
  • Aspiration threshold prevents aspiration
  • Aspiration threshold prevents operation
  • Bulb A large hours
  • Bulb A large hours message
  • Bulb A large hours message displayed
  • Bulb B FET problem
  • Bulb B FET problem message
  • Bulb B FET problem message displayed
  • Bulb B large hours
  • Bulb B large hours alert (voltage high)
  • Bulb B large hours message
  • Bulb B large hours message displayed
  • Bulb B problem
  • Calibrate air pump
  • Calibrate foot pedal
  • Calibration: Aspiration Function
  • Calibration: Pressure Regulators
  • Calibration: Vacuum Regulator
  • Cassette full
  • Cassette over full
  • Cassette position bad
  • Contrast adjust message displayed
  • Contrast adjustment mode active
  • CPU electronics board failure
  • Device disabled due to high internal temperature
  • Disk space warning
  • Disk space warning message displayed
  • Disk write protected
  • Disk write protected message displayed
  • Diskette is full
  • Diskette is full message displayed
  • Eye pressure alarm
  • Floppy disk drive checkout
  • High cassette level
  • High pump pressure
  • High system pressure
  • High temperature
  • LED/backlight adjust
  • LED/backlight adjust message displayed
  • Lift pedal to prime
  • Lift pedal to prime message displayed
  • Lift pedal to reflux
  • Low air pump pressure
  • Low air pump pressure message displayed
  • Low air pump pressure warning displayed
  • Low temperature warning displayed
  • Low vacuum level
  • Low vacuum level message displayed
  • Low vacuum level warning displayed
  • No handpiece present message displayed
  • No keyboard heartbeat message displayed
  • No stage selected message displayed
  • No stages enabled message displayed
  • Over temperature
  • Overpressure detected
  • Overpressure detected message displayed
  • Overpressure detected warning
  • Overpressure detected warning message displayed
  • Overpressure warning displayed
  • Pedal fuse problem
  • Pedal fuse problem warning
  • Service required (air-related)
  • Shutdown at 55°C threshold
  • Stabilizing message displayed
  • Stabilizing message displayed at startup
  • Stage indicator message displayed
  • System fuse blown or open
  • Temp sensor problem message displayed
  • Temperature sensor problem message displayed
  • Temperature sensor undetectable
  • Threshold is xx mmHg message displayed
  • Unit powers on but display is blank
  • Vacuum pump hours exceeded
  • Valve fuse failure checkout
  • Valve fuse problem

104 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
+12 volt fuse problem+12V power supply not present; fuse opened or DC-DC converter failed+12V fuse opens or +12V DC-DC converter malfunctionerror
-12 volt fuse problem-12V fuse on analog board opened; -12V power supply failure-12V supply lost or fuse openederror
+12 volt fuse problem12V fuse on analog board opened; 12V power supply failure+12V supply lost or fuse openederror
-12 volt fuse problem-12V power supply not present; fuse opened or DC-DC converter failed-12V fuse opens or -12V DC-DC converter malfunctionerror
+12 volt fuse problem+12 volt power not available on analog board+12V fuse opens or DC-DC converter failserror
-12 volt fuse problem-12 volt power not available on analog board-12V fuse opens or DC-DC converter failserror
-22 volt fuse problem-22V power supply not present; fuse opened or DC-DC converter failed; LCD contrast affected-22V fuse opens or -22V DC-DC converter malfunctionerror
-22 volt fuse problem-22 volt power not available; LCD contrast lost or display unreadable-22V fuse opens or DC-DC converter failserror
Air exchange problemPressure difference between eye pressure sensor and air exchange pump pressure sensor exceeds safe limitPressure difference of 15 mmHg or greater detected when air exchange valve is openerror
Air exchange set: xxxAir exchange surgical function is active; pressure setting is displayedAir exchange surgical function enabledinfo
Air pump fuse problemAir exchange pump fuse on analog board has opened; fuse must be replaced to clear errorNon-resettable fuse fails; also occurs if +24V output of power supply failserror
Air pump hours excessAir pump runtime has exceeded maximum limit of 2000 hours; pneumatics module requires serviceAir pump cumulative run time exceeds 2000 hourswarning
Air pump problemAir exchange pump unable to generate pressure during power-up self-testPower-up test detects no pressure generation from air exchange pump when surgical function enablederror
Air sensor problemPressure sensor indicates problem; low output voltage detectedFunction off and vacuum present, or function on and no pressure present; sensor output voltage less than baseline atmospheric voltagewarning
Bulb A FET problemBulb A is not illuminating despite being enabled and presentBulb A enabled but not lit or FET voltage exceeds 2.5Verror
Bulb A FET problemBulb A illumination failure due to FET or A/D converter faultBulb A enabled but not illuminated; FET voltage exceeds 2.5 V or A/D converter malfunction detectederror
Bulb A hours xxxBulb A in-use time display in hoursIllumination surgical function turned on with bulb A selectedinfo
Bulb A hours xxxOperating hours status for bulb AWhen illumination surgical function is turned on with bulb A selectedinfo
Bulb A large hoursBulb A operation time exceeds safe usage limitVitMan powered up or illumination turned on with bulb A selected and hours > 40warning
Bulb A large hoursBulb A has exceeded 40 operating hoursVitMan powered up or illumination turned on with bulb A selected and hours exceed 40warning
Bulb A missingBulb A not detected by the systemVitMan powered up or illumination turned on with bulb A selected; bulb detection voltage below 2.5V thresholdwarning
Bulb A openBulb A filament detected as open or defectiveBulb A not detected after installation; FET voltage below 0.03V or fuse open or +24V supply failederror
Bulb A openBulb A not detectedBulb A cannot be detected by monitoring circuiterror
Bulb A problemBulb A appears on while illumination function is disabledFET voltage between 0.03V and 2.5V when illumination is off; A/D converter reads erroneous voltage in this rangewarning
Bulb B FET problemBulb B FET circuit failure preventing bulb illuminationBulb B installed and enabled but does not illuminate; FET not switching properlyerror
Bulb B FET problemBulb B present and enabled but not illuminating due to FET hardware failureBulb B enabled and installed but no illumination occurserror
Bulb B hours xxxIn-use time display for bulb B in hoursIllumination surgical function turned on with bulb B selectedinfo
Bulb "B" hours xxxDisplays current operating hours of bulb B (xxx = hours)Illumination surgical function is turned on with bulb B selectedinfo
Bulb B large hoursBulb B operating time has exceeded design life thresholdPower-up or when illumination turned on with bulb B selected and hours exceed 40info
Bulb "B" large hoursBulb B operating hours exceed safe service lifeBulb B hours exceed 40 hours at power-up or when illumination is turned onwarning
Bulb B missingBulb B not detected at power-up or when illumination turned onFET voltage below 2.5V or A/D converter fault or fuse open or +24V absentwarning
Bulb "B" missingBulb B is not detected at power-upFET voltage drops below 2.5 volts at startup, bulb not installed, fuse open, or +24V missingwarning
Bulb B openBulb B not detected after power-upFET voltage less than 0.03 voltserror
Bulb B openBulb B filament open or failedWhen bulb B not detected during operationwarning
Bulb "B" openBulb B is not detected or has failed open circuitFET voltage across bulb circuit drops below 0.03 volts, or bulb power fuse opens, or +24V supply failswarning
Bulb B problemBulb detection logic indicates bulb B is on while illumination surgical function is offBulb B enabled state does not match illumination function statewarning
Bulb "B" problemBulb B detection logic indicates bulb is on while illumination function is offBulb detection logic detects illumination signal when surgical function is disablederror
Bulb fuse problemIllumination fuse on analog board has opened or bulb power not presentFuse opens or power supply +24V output failswarning
Calibrate air pumpAir exchange calibration tables are invalidInvalid calibration data detectedwarning
Calibrate aspirationAspiration operation calibration tables are invalidInvalid calibration data detectedwarning
Calibrate foot pedalFoot pedal calibration tables are invalidInvalid calibration data detectedwarning
Cassette fullAspiration cassette has reached first level fullness warning thresholdApproximately 300 cc of fluid aspiratedinfo
Cassette over fullAspiration cassette has exceeded maximum capacity; vacuum pump is disabledApproximately 325 cc of fluid aspiratedcritical
Cassette position badAspiration cassette is not properly aligned in cassette chamberCassette inserted without proper lock knob rotationerror
Contrast adjustLCD contrast adjustment in progressLEFT ARROW or RIGHT ARROW pressed while not in menu modeinfo
Currently in stage 1Status line message indicating operation in stage 1 modeStage button pressed and released when in stage 1info
Currently in stage 2Status line message indicating operation in stage 2 modeStage button pressed and released when in stage 2info
Currently in stage 3Status line message indicating operation in stage 3 modeStage button pressed and released when in stage 3info
Currently in stage 4Status line message indicating operation in stage 4 modeStage button pressed and released when in stage 4info
Currently in stage 5Status line message indicating operation in stage 5 modeStage button pressed and released when in stage 5info
Currently in stage 6Status line message indicating operation in stage 6 modeStage button pressed and released when in stage 6info
Currently in stage 7Status line message indicating operation in stage 7 modeStage button pressed and released when in stage 7info
Device disabledSystem has disabled surgical functions due to overtemperature conditionInternal device temperature reaches 55°C or highercritical
Device disabledInternal device temperature is 55 °C or higher and surgical functions are being shut downInternal temperature reaches or exceeds 55 °Cwarning
Disk available hh:mmMinimum time remaining to record data on floppy disk in hours and minutes formatDisplayed after data from previous minute is saved to diskinfo
Disk available hh:mmFloppy disk space status message showing minimum time remaining to record dataDisplayed in status line after each minute of data is saved to diskinfo
Diskette is fullFloppy disk has no remaining free space for data recordingDisk space exhausted during data save operationerror
Diskette is fullNo free space remains on floppy disk for data storageWhen all disk capacity is consumederror
Diskette read problemFloppy disk structure is corrupted or disk drive connection is faultyBoot block, file allocation tables, or root directory cannot be read from diskerror
Diskette read problemUnable to read boot block, file allocation tables or root directory on floppy diskBoot block, file allocation tables or root directory cannot be accessederror
Disk space warningDisk available time below thresholdMinimum time remaining less than 30 minuteswarning
Disk space warningAvailable disk space is critically lowWhen minimum time remaining is less than 30 minuteswarning
Disk space warningMinimum time remaining on floppy disk is less than 30 minutesInsufficient free space on floppy diskwarning
Disk space warningWarning message displayed when floppy disk free space is lowMinimum time remaining is less than 30 minuteswarning
Disk write protectedCannot save data to floppy disk due to write protectionChecked when attempting to save data to diskerror
Drive fuse problemMain compressor fuse on analog board has opened; no power to compressorWhen main compressor fuse opens due to overcurrent or power surgeerror
Drive sensor problemPressure sensor monitoring main compressor indicates fault condition or zero pressure readingSensor output voltage indicates zero pressure or voltage appears lowwarning
Eye pressure alarmCurrent eye pressure not within 5 mmHg of set point during air exchange surgical functionAir exchange surgical function active and eye pressure deviates from set pointwarning
Eye sensor problemEye pressure sensor is indicating a malfunction or vacuum/pressure condition errorAir exchange function off with vacuum present, or on with no pressure present; sensor voltage lowerror
Fan fuse problemFan fuse on analog board has opened or fan power is not presentFan fuse opens; +12V power supply output failswarning
Floppy fuse problemDisk drive fuse on analog board has opened or disk drive power is not presentFloppy fuse opens; +12V or +5V power failswarning
Foot pedal is activeFoot pedal is active; several user actions are prohibitedFoot pedal engaged during operationinfo
Foot pedal problemFoot pedal down switch indicates up position while foot pedal position shows ≥10% depressionFoot pedal down switch malfunction or cable/connector failurewarning
Ground fault detectedUltrasound handpiece is not properly groundedGround path interruption or handpiece not seatederror
High cassette levelAspiration cassette fluid level is high (first warning level)Approximately 250 cc of fluid aspirated into cassettewarning
High pump pressureAir exchange pump pressure exceeds set point by more than 8 mmHgAir exchange surgical function active and pump pressure highwarning
High system pressureMain pressure pump pressure exceeds upper limit (2844 mmHg / 55 psi)Main pump pressure above 2844 mmHg (55 psi)warning
High temperatureInternal device temperature is 50 °C or higherInternal device temperature reaches or exceeds 50 °Cwarning
Hold to enter stage 4Stage four is ready to be enteredStage button pressed and held when stage four will be entered upon releaseinfo
Now saving stage 3System is saving current machine state to stage three memoryStage button is pressed and held while in stage three save modeinfo
Now saving stage 4System is saving current machine state to stage four memoryStage button is pressed and held while in stage four save modeinfo
Now saving stage 5System is saving current machine state to stage five memoryStage button is pressed and held while in stage five save modeinfo
Now saving stage 6System is saving current machine state to stage six memoryStage button is pressed and held while in stage six save modeinfo
Now saving stage 7System is saving current machine state to stage seven memoryStage button is pressed and held while in stage seven save modeinfo
Overpressure detectedEye pressure exceeds the configured warning pressure limitCurrent eye pressure exceeds warning pressure limitwarning
Over temperatureInternal device temperature has reached 53°C or higher; warning condition onlyInternal temperature sensor reading >= 53°Cwarning
Pedal fuse problemFoot pedal fuse on CPU board has opened; foot pedal power supply failureFoot pedal fuse opens due to overcurrent or fault conditionwarning
Service requiredSerious problem detected requiring VitMan repairVarious hardware problems detected during operation or startupcritical
Shutdown at 55°CInternal device temperature at or above 53°C but below 55°CTemperature sensor on analog board reports internal temperature >= 53°Cwarning
Stabilizing...Hardware problem detected during pressure sensor thermal stabilization phase; device non-functionalHardware malfunction detected while pressure sensors are thermally stabilizing at power-upcritical
Stage 3 selectedStage 3 is selected and ready to enterStage button pressed and held, then releasedinfo
Stage 4 selectedStage 4 is selected and ready to enterStage button pressed and held, then releasedinfo
Stage 5 selectedStage 5 is selected and ready to enterStage button pressed and held, then releasedinfo
Stage 6 selectedStage 6 is selected and ready to enterStage button pressed and held, then releasedinfo
Stage 7 selectedStage 7 is selected and ready to enterStage button pressed and held, then releasedinfo
Temp sensor problemTemperature sensor is undetectable or not communicating properlyTemperature sensor fails to establish communication with systemwarning
Threshold is xx mmHgAspiration operation prevented by threshold setting; xx shows current threshold value in mmHgAspiration valve cannot open because vacuum level does not meet threshold requirementwarning
Vacuum fuse problemVacuum pump fuse on analog board has openedVacuum pump fuse opens; vacuum pump power not presentwarning
Vacuum pump hoursVacuum pump runtime has exceeded maximum service intervalPump operating hours exceed thresholdinfo
Valve fuse problemValve fuse on analog board has openedValve fuse opens due to overcurrent or agewarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.