SunTech Medical Tango M2: troubleshooting and error codes
The ServiceForge knowledge base holds 7 documented symptoms for the SunTech Medical Tango M2, with 10 ranked causes, 14 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: K-sound detection window misconfigured patient safety
K-sound detection may be incorrect if Bin Skip or ECG Delay settings are changed from default values
Ranked causes
- 1. Bin Skip value not set correctly for ECG trigger type (occasional)
- 2. ECG Delay value not set correctly (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal contamination detected during inspection caution
Dust, debris, or liquid contamination found inside enclosure during internal inspection
Ranked causes
- 1. Dry contamination (dust or debris from grill) (occasional)
- 2. Liquid contamination from spillage or water ingress (occasional)
Visually inspect internal components for damage or contamination
- Test point
- all circuit boards and pneumatic tubing inside enclosure
- Expected
- No visible damage, contamination, or liquid
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pneumatic system leak or loss of pressure
Monitor loses pressure or shows abnormal leak rate during testing
Ranked causes
- 1. Damaged or kinked pneumatic tubing (common)
- 2. Liquid contamination in pneumatic system (occasional)
Perform leak rate test per Section 4 procedure
- Test point
- patient cable hose connection
- Expected
- Pressure loss within acceptable limits
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressure verification out of tolerance patient safety
Tango M2 displayed pressure readings do not match calibrated reference sphygmomanometer within specified tolerance at test pressures
Ranked causes
- 1. Pressure sensor out of calibration (common)
Verify pressure reading at 0 mmHg reference
- Test point
- patient hose connector
- Expected
- 0 - 1 mmHg, 0 - 1
- Tools
- calibrated reference sphygmomanometer
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: System leak detected patient safety
Pressure drops more than 6 mmHg in 60 seconds when system is pressurized to 250 mmHg, indicating pneumatic leak
Ranked causes
- 1. Pneumatic system leak in tubing, connectors or cuff (common)
Perform leak rate test
- Test point
- patient hose connector
- Expected
- pressure drop not to exceed 6 mmHg in 60 seconds at 250 mmHg, 0 to 6 mmHg per 60 sec
- Tools
- stop watch, 500ml volume, hand inflation bulb with bleed valve
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overpressure relief not functioning patient safety
Monitor fails to open relief valves at correct pressure during overpressure check
Ranked causes
- 1. Relief valve stuck or malfunction (unknown)
Set pressure to approximately 280mmHg using inflation bulb
- Test point
- patient cable hose connection
- Expected
- 280mmHg per monitor displayed value
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Safety timeout not functioning correctly patient safety
Monitor does not release pressure within specified time during safety timeout check
Ranked causes
- 1. Pneumatic system malfunction or valve failure (unknown)
Set pressure to approximately 200mmHg using inflation bulb
- Test point
- patient cable hose connection
- Expected
- 200mmHg per monitor displayed value
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.