Stryker TP700: troubleshooting and error codes
The ServiceForge knowledge base holds 12 documented symptoms for the Stryker TP700, with 33 ranked causes, 3 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Warning indicator and audible alarm (flash and beep) patient safety
Warning light flashing and audible beep sounding; indicates alarm condition
Ranked causes
- 1. High heat setpoint (107°F / 42°C) or cooling setpoint (50°F / 10°C) selected (common)
- 2. Loss of power while unit was in therapy mode (possible power fail) (occasional)
- 3. Unit is running ON cycle period after reaching desired setpoint (common)
- 4. Unit just entered or exited lock mode (occasional)
- 5. Setpoint outside body temperature range selected (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow indicator light is ON
Flow warning light illuminated; indicates water flow obstruction or low level
Ranked causes
- 1. Water flow to pad or hose is restricted (common)
- 2. Clamp is closed (common)
- 3. Water level is low or reservoir is empty (common)
- 4. T/Pump is filled with water that is too hot (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not heat
T/Pump will not heat water to desired temperature; heating function inoperative
Ranked causes
- 1. Reservoir is empty (common)
- 2. Flow is blocked (common)
- 3. Therapy cycle is OFF (common)
- 4. Heater is not receiving power or is defective (occasional)
Measure voltage to heater at J2 on PCB
- Test point
- J2 on PCB
- Expected
- 120V ± 10%, ±10%
- Tools
- DMM
1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on high voltage
T/Pump will not turn on; no operation when power button pressed
Ranked causes
- 1. Electrical cord not plugged into properly grounded Hospital Grade receptacle (common)
- 2. Damaged power cord or plug (common)
- 3. Defective PC Board (occasional)
- 4. Defective Membrane Panel (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not cool
T/Pump will not cool water to desired temperature; cooling function inoperative
Ranked causes
- 1. Reservoir is empty (common)
- 2. Flow is blocked (common)
- 3. Ice is depleted (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not pump
T/Pump will not pump water; no fluid flow despite power
Ranked causes
- 1. Water level is low or reservoir is empty (common)
- 2. Loss of power to motor (occasional)
- 3. Defective motor, impeller magnet, or magnet driver assembly (occasional)
Measure voltage across motor connections
- Test point
- motor connections
- Expected
- 120V ± 10%, ±10%
- Tools
- DMM
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water leaks from hose connectors caution
Water dripping or leaking from connector points; indicates seal failure or improper connection
Ranked causes
- 1. Damaged O-ring (occasional)
- 2. Locking ring on Clik-Tite connector not snapped into place (common)
- 3. Quick disconnect not seated properly (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature or Therapy Time buttons do not work
Control buttons on membrane panel unresponsive or non-functional
Ranked causes
- 1. Buttons have been locked (common)
- 2. Defective Membrane Panel (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature Setpoint light blinking
Temperature indicator light flashing; indicates warming or cooling activity
Ranked causes
- 1. Unit is warming up to selected setpoint (common)
- 2. Unit is in cooling mode for longer than 40 minutes (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Both Temperature and Therapy Cycle Setpoint lights blinking
Both indicator lights flashing simultaneously; indicates OFF therapy cycle timing
Ranked causes
- 1. Unit is in OFF therapy cycle time (common)
Flow indicator and Standby indicator are ON with unit not pumping
Both flow and standby lights illuminated; unit does not pump despite appearing to run
Ranked causes
- 1. Unit detected flow warning for more than 5 minutes and entered standby (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Warning indicator on with unit in Standby Mode patient safety
Warning light illuminates when unit is in Standby; indicates over-temperature shutdown
Ranked causes
- 1. Unit shut down in over-temperature condition (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.