Stryker Thermedx: troubleshooting and error codes

The ServiceForge knowledge base holds 6 documented symptoms for the Stryker Thermedx, with 6 ranked causes, 22 diagnostic checks with test points and expected values, and 3 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Calibration: Export Log File

Export calibration log data to USB drive for record keeping

Ranked causes

  1. 1. Log file export required after calibration (common)
First check

Navigate to export screen

Test point
Device display menu navigation
Expected
Export logs to USB drive button visible on calibration screen

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Fluid Bag Hook load cells

Upper load cells on left and right Fluid Bag Hooks require calibration verification with touchscreen interface.

Ranked causes

  1. 1. Load cell calibration procedure (common)
First check

Verify left Fluid Bag Hook calibration initiated

Test point
Touchscreen display
Expected
Instruction prompts to measure left Fluid Bag Hook

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Load Cell Canister Ring

Verification and calibration of canister ring load cells on single load cell configuration Thermedx fluid management systems

Ranked causes

  1. 1. Load cell calibration required during preventive maintenance (common)
First check

Verify device load cell configuration type

Test point
Device display configuration screen
Expected
Single load cell configuration (not four load cell)

6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Power Off Device caution

Proper shutdown procedure after calibration

Ranked causes

  1. 1. Device power-off required after calibration completion (common)
First check

Press Power off button first time

Test point
Device display Power off button location lower left corner
Expected
Button responds; confirmation prompt appears

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Sticker Placement

Apply calibration sticker to device after successful calibration completion

Ranked causes

  1. 1. Calibration sticker documentation required (common)
First check

Complete calibration sticker with required information

Test point
01278 Calibration Sticker form
Expected
Sticker filled with calibration date and due date (one year later)

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Upper Load Cells (Fluid Bag Hooks)

Verification and calibration of upper load cells (left and right fluid bag hooks)

Ranked causes

  1. 1. Upper load cell calibration required (common)
First check

Verify zero load reading on left hook

Test point
Diagnostic screen Wt2(L) field
Expected
Reading between -150 and -50 units, -150 to -50

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
Canister ring requires re-calibrationCanister ring requires re-calibrationDisplayed during calibrationwarning
Could not load calibration data. Call service for re-calibraUnable to load calibration dataDisplayed during calibrationerror
Sensor error, reboot system, call service if condition repeaSensor error detectedDisplayed during calibrationerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI