Stryker SwitchPoint Infinity 3: troubleshooting and error codes
The ServiceForge knowledge base holds 48 documented symptoms for the Stryker SwitchPoint Infinity 3, with 86 ranked causes, 90 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No video on monitor
The monitor displays no video image; screen is black or shows no signal.
Ranked causes
- 1. Monitor is powered off or in standby (common)
- 2. Video router main power switch is OFF (common)
- 3. Monitor is not set to DVI/Digital RGB input (common)
- 4. Video is not properly routed to the monitor (occasional)
Verify monitor power status
- Test point
- Monitor power switch and power LED
- Expected
- Monitor powered ON with power indicator lit
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch panel does not respond after system power on
Touch panel is unresponsive to user input after the system is turned on.
Ranked causes
- 1. Touch panel power switch is OFF (common)
- 2. Loose or disconnected power cord or signal cable (common)
- 3. Touch panel monitor is OFF (common)
- 4. Touch input is disabled on AIO system (occasional)
Verify touch panel power switch position
- Test point
- Touch panel front power switch
- Expected
- ON position
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch panel screen is blank
The touch panel display shows no video output or image.
Ranked causes
- 1. Touch panel is powered off (common)
- 2. SPI3 router is powered off (common)
- 3. Video cables are loose or disconnected (common)
- 4. Remote computer power saving mode activated in KVM (occasional)
Verify touch panel is powered on
- Test point
- Touch panel power switch
- Expected
- ON position; power LED illuminated
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio not leaving the room
The audio from the wireless microphone is not being transmitted or heard in other locations.
Ranked causes
- 1. Wireless microphone batteries are depleted (common)
- 2. Wireless receiver is not functioning (occasional)
- 3. Microphone frequency does not match receiver frequency (occasional)
Check microphone battery LED indicator
- Test point
- LED on top of wireless microphone belt pack
- Expected
- LED lights momentarily on power-up, then extinguishes; does not stay on
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing images in PIP/Quad view
One or more of the image windows in the picture-in-picture or quad view display are blank or missing.
Ranked causes
- 1. Video source not routed to PIP/Quad window (common)
- 2. Routed video source is powered off (common)
- 3. PIP/Quad needs to be turned OFF and ON (occasional)
Verify sources are routed to all PIP/Quad windows
- Test point
- PIP/Quad page source routing configuration
- Expected
- Each image window position has a source assigned
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No endoscopic camera video is displayed on flat panel monitor
The endoscopic camera video feed does not appear on the display screen.
Ranked causes
- 1. Light source, camera, or monitor is powered off (common)
- 2. Endoscopic camera output not routed to monitor (common)
- 3. Digital capture device has no video output (occasional)
Verify endoscopic equipment power status
- Test point
- Light source, camera, and monitor power switches and indicators
- Expected
- All devices powered ON and operational
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No image in preview when source is selected
The preview window on the control system does not display the video source selected.
Ranked causes
- 1. S-video or DVI cable is loosely connected (common)
- 2. Video source is powered off (common)
- 3. AIO DVI cable not fully connected to touch panel (occasional)
Verify S-video or DVI preview cable connections
- Test point
- Back of control section and router Output 2 preview connectors
- Expected
- Cable firmly seated at both connection points
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIP/Quad display shows only internal color bars
The Picture-In-Picture (PIP) or quad-view display only shows test color bar pattern instead of video sources.
Ranked causes
- 1. No source is routed to PIP/Quad on control page (common)
- 2. Routed video sources are powered off (common)
- 3. Video sources themselves are generating color bars (occasional)
Verify video source assignment to PIP/Quad windows
- Test point
- PIP/Quad page source routing settings
- Expected
- Video sources assigned to each PIP or quad window
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SDC touch screen does not respond
The SDC (Surgical Display Console) touch screen is unresponsive to touch input.
Ranked causes
- 1. SDC is not powered up (common)
- 2. SDC is running unsupported software version (common)
- 3. RS-232 or video cables are loose or disconnected (common)
Verify SDC is powered up
- Test point
- SDC power switch and status indicator
- Expected
- SDC powered ON and operational
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sidne or Device Control not connected message appears
A message appears on the control panel stating 'Sidne not connected' or 'Device Control not Connected'.
Ranked causes
- 1. Sidne or SDC3 device is powered off (common)
- 2. Sidne is running unsupported software version (common)
- 3. Power or video cables are loose or disconnected (common)
Verify Sidne or SDC3 is powered on
- Test point
- Sidne or SDC3 power switch and LED
- Expected
- Device is powered ON with operational LEDs
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch panel will not power up
The touch panel does not power on when the power switch is activated.
Ranked causes
- 1. Power cable is not firmly connected (common)
- 2. Power cable needs reseating after disconnect (common)
- 3. Monitor is powered but not displaying (LED off) (occasional)
Inspect touch panel power cable connection
- Test point
- Touch panel power inlet connector
- Expected
- Power cable firmly seated and fully inserted
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio is distorted
Audio output from the speakers exhibits distortion, clipping, or noise.
Ranked causes
- 1. Input level is too high (common)
- 2. Speaker output level is too high (common)
Measure audio input level
- Test point
- Audio input level indicator or control
- Expected
- Input level within normal range without peaks causing distortion
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Color bars appear on touch panel
The touch panel displays internal color bar test pattern instead of normal video content.
Ranked causes
- 1. Control system is powered off (common)
- 2. Power or video cables loose or disconnected (common)
Verify control system power status
- Test point
- Control system power switch and power LED
- Expected
- Control system powered ON with LED illuminated
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control system will not power up
The control system does not power on when the power switch is activated.
Ranked causes
- 1. Power cable is not firmly connected (common)
- 2. Front power switch is not in ON position (common)
Inspect control system power cable connection
- Test point
- Control system power inlet connector
- Expected
- Power cable firmly seated and fully inserted
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mouse and keyboard do not function on home screen
The mouse and keyboard inputs do not work when the system is displaying the home screen.
Ranked causes
- 1. Router requires reboot (common)
- 2. USB connection on control system is loose or disconnected (common)
Reboot the router to restore USB functionality
- Test point
- Router power cycle
- Expected
- After reboot, mouse and keyboard are responsive on home screen
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mouse and keyboard do not function on PC screen
The mouse and keyboard inputs do not work when the system is displaying the PC screen view.
Ranked causes
- 1. Router requires reboot (common)
- 2. USB connection on control system is loose or disconnected (common)
Reboot the router to restore USB in PC mode
- Test point
- Router power cycle
- Expected
- After reboot, mouse and keyboard are responsive in PC screen view
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audio from surgeon's microphone
The wireless microphone worn by the surgeon produces no audible output.
Ranked causes
- 1. Microphone is turned OFF (common)
- 2. Microphone batteries are depleted (common)
Verify microphone power status
- Test point
- Wireless microphone power button
- Expected
- Microphone powered ON
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No room-to-room audio connection
Audio communication between rooms is not working during a call.
Ranked causes
- 1. Speaker controls volume is muted in one or both rooms (common)
- 2. Speaker controls volume level is too low (common)
Verify speaker controls are not muted
- Test point
- Speaker controls mute setting in both rooms
- Expected
- Mute is OFF in both rooms
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Observation room PC is not responding
The observation room PC is not communicating with or responding to the SwitchPoint system.
Ranked causes
- 1. Observation room PC is powered off (common)
- 2. PC requires reboot to restore communication (occasional)
Verify observation room PC is powered up
- Test point
- PC power switch and power LED
- Expected
- PC is powered ON and operational
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Room camera video works but control does not
The video feed from the BRC-z330 room camera is displayed but the pan/tilt/zoom or other camera controls are not responding.
Ranked causes
- 1. Room camera requires reset (common)
- 2. RS-232 connection to camera is loose or disconnected (common)
Press Reset button on Room Camera page
- Test point
- Room Camera page reset function
- Expected
- Camera control responds after reset
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Router will not power up
The SPI3 video router does not power on when the power switch is activated.
Ranked causes
- 1. Power cable is not firmly connected (common)
- 2. Front power switch is not in ON position (common)
Inspect router power cable connection
- Test point
- Router power inlet connector
- Expected
- Power cable firmly seated and fully inserted
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch input does not work
The touch screen does not respond to touch input from the user.
Ranked causes
- 1. Serial or USB cable disconnected or loose (common)
- 2. Touch panel requires calibration (common)
Verify serial or USB cable connection security
- Test point
- Serial or USB connector at control section and touch panel
- Expected
- Cable firmly seated at both connection points
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch panel not switching between PC view and normal view
The touch panel fails to switch between PC view and home/normal operating view when selected.
Ranked causes
- 1. External PC is powered off (common)
- 2. RS-232 or USB cable connection is loose or disconnected (common)
Verify external PC is powered on
- Test point
- PC power switch and power LED
- Expected
- PC is powered ON and operational
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Video is distorted or noisy
The video display exhibits noise, lines, ghosting, or other visual distortion.
Ranked causes
- 1. Video cables are not securely connected or routed poorly (common)
- 2. Input video source device is malfunctioning (occasional)
Inspect video cable connections and routing
- Test point
- All video cable connections between sources and router
- Expected
- Cables firmly seated; routed away from power lines
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable connection inspection
Inspect all cable connections on control system front panel for secure seating
Ranked causes
- 1. Loose cable connections (occasional)
Visually inspect cable connections
- Test point
- Front of Control System
- Expected
- All cables fully seated and secure
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Touch Panel
Touch screen requires calibration for accurate input detection and responsiveness
Ranked causes
- 1. Touch panel calibration drift (unknown)
Connect USB keyboard to expansion module
- Test point
- Keyboard USB Port
- Expected
- Keyboard connected and recognized
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot calibrate touch panel
The touch panel calibration routine fails or cannot be initiated.
Ranked causes
- 1. Serial or USB cable is disconnected or loose (common)
Verify serial or USB cable connection prior to calibration
- Test point
- Serial or USB connector at touch panel and control system
- Expected
- Cable firmly seated at both ends
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot control remote room Pan/Tilt/Zoom during call
The pan, tilt, and zoom controls for the far-end camera are not responding during a video call.
Ranked causes
- 1. Farside camera control is not enabled in Calls page (common)
Verify farside camera control is enabled in Calls page
- Test point
- Calls page farside camera control setting
- Expected
- Farside camera control is turned ON
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Characters on screen are dim
Text and graphics on the touch panel display appear faint or with low brightness.
Ranked causes
- 1. Display brightness setting too low (common)
Adjust brightness via OSD menu
- Test point
- ON-Screen Display brightness control
- Expected
- Brightness increased to readable level
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Clean Touch Screen caution
Periodic cleaning of touch screen and monitor housing to maintain clarity and prevent damage
Ranked causes
- 1. Touch screen soiling or contamination (unknown)
Unplug Touch Panel from power
- Test point
- Power outlet
- Expected
- Touch Panel powered down
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Equipment malfunction due to electromagnetic interference caution
Device performance affected by portable or mobile RF communications equipment in the vicinity
Ranked causes
- 1. Improper use of RF communications equipment near device (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Equipment performance degradation when stacked or adjacent to other equipment caution
Malfunction or erratic behavior when device is placed adjacent to or stacked with other equipment
Ranked causes
- 1. Equipment stacking or adjacent placement without verification (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan status check (SPI3)
Verify video router fan operation by checking LED indicators for green illumination
Ranked causes
- 1. Fan malfunction or LED failure (unknown)
Power on video router
- Test point
- Video router power switch
- Expected
- Video router operational
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image on monitor does not change when new video is routed
The video displayed on the monitor does not switch to the newly selected video source.
Ranked causes
- 1. Monitor is set to incorrect input mode (not DVI/Digital RGB) (common)
Verify monitor input source selection
- Test point
- Monitor input/source selection menu
- Expected
- Monitor set to DVI/Digital RGB input
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Increased electromagnetic emissions or decreased immunity from non-specified accessories caution
Equipment EMC performance compromised when using accessories, transducers, or cables not specified by manufacturer
Ranked causes
- 1. Use of non-Stryker accessories, transducers, or cables (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keep equipment area dust-free
Maintain clean environment around SPI3 to prevent dust accumulation affecting performance
Ranked causes
- 1. Dust accumulation in system (common)
Inspect surrounding area for dust
- Test point
- SPI3 equipment vicinity
- Expected
- Area free of dust and debris
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audio in room
The audio system produces no sound output through the speakers.
Ranked causes
- 1. Speaker controls volume is turned down or muted (common)
Verify speaker control volume level and mute status
- Test point
- Speaker control interface or panel
- Expected
- Volume level at audible setting; mute function OFF
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Picture color is unusual or rolling on touch panel display
Video display on touch panel shows color distortion, color shift, or rolling/vertical sync issues.
Ranked causes
- 1. Loose video cable connections (common)
Verify video cable connections are tight
- Test point
- Touch panel rear video connectors and router Output1
- Expected
- All video cables firmly seated and tightened
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIP/Quad inset does not show up
The picture-in-picture inset window or quad view does not appear on the monitor.
Ranked causes
- 1. PIP/Quad option is not enabled (common)
Verify PIP/Quad option is enabled
- Test point
- PIP/Quad enable setting on control interface
- Expected
- PIP/Quad option is turned ON
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen flashes when touch panel is first turned on
The touch panel display exhibits flickering or flashing immediately after power on.
Ranked causes
- 1. Power supply stabilization required (common)
Perform power cycle on touch panel
- Test point
- Touch panel power cable connector
- Expected
- Display stable after power cycle without flashing
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SDC touch screen is black with no display
The SDC touch screen displays a black screen with no image or text output.
Ranked causes
- 1. SDC is configured for SwitchPoint Infinity 3 SDC remote configuration (common)
SDC touchscreen active during remote control patient safety
SDC touchscreen remains active and responsive even when being controlled remotely by SwitchPoint Infinity Touch Panel, despite display appearing black. Touching the SDC touchscreen during remote operation may cause inadvertent device use.
Ranked causes
- 1. SDC touchscreen hardware design maintains active state during remote control mode (common)
Verify SDC touchscreen state during remote control operation
- Test point
- SDC display and touch panel
- Expected
- Touchscreen display is black but remains responsive to touch input when SwitchPoint Infinity software is actively controlling device
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SDC touchscreen active during SwitchPoint control patient safety
SDC touchscreen remains active and responsive even when controlled by SwitchPoint Infinity software, despite display appearing black, risking inadvertent commands
Ranked causes
- 1. SDC touchscreen not disabled during SwitchPoint control (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System is slow to respond
The touch panel and control system exhibit sluggish response times to user input and commands.
Ranked causes
- 1. System requires reboot (common)
Perform system reboot
- Test point
- System power cycle and startup
- Expected
- System boots normally and responds quickly to input
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch does not work when controlling SDC
Touch input on the touch panel fails to control the Surgical Display Console (SDC).
Ranked causes
- 1. SDC is not properly calibrated (unknown)
Verify SDC is operating normally and calibrated
- Test point
- SDC touch responsiveness and software status
- Expected
- SDC operational with current calibration
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch panel not responding to touch or responding incorrectly
Monitor does not respond accurately when touched, or cursor does not follow finger correctly on screen
Ranked causes
- 1. Touch panel requires calibration (common)
Launch calibration session
- Test point
- Keyboard
- Expected
- DOS window appears and calibration targets display
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Voltage level indicator check (SPI3)
Verify video router voltage regulation by checking LED indicators for green illumination
Ranked causes
- 1. Voltage supply malfunction or LED failure (unknown)
Power on video router
- Test point
- Video router power switch
- Expected
- Video router operational
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Weekly system restart
Restart SPI3 control system weekly for optimal performance
Ranked causes
- 1. System performance optimization (common)
Execute system shutdown
- Test point
- Control system power
- Expected
- All systems gracefully shut down
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI