Stryker Renaissance 2020: troubleshooting and error codes
The ServiceForge knowledge base holds 34 documented symptoms for the Stryker Renaissance 2020, with 70 ranked causes, 20 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Litter drifts after raised or lowered
Hydraulic litter drifts or continues to move after foot pedals are released
Ranked causes
- 1. Air in hydraulic system (common)
- 2. Valve control linkage misalignment (common)
- 3. Leaking hoses or fittings (occasional)
- 4. Faulty jack (occasional)
- 5. Faulty check valve (rare)
- 6. Faulty main valve (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Litter will not lift electrically with no motor sounds patient safety
No audible motor activity when attempting to lift; bed mattress platform cannot be raised
Ranked causes
- 1. Lift motor failure (common)
- 2. Motor capacitor connection failure (occasional)
- 3. Motor capacitor failure (occasional)
- 4. Voltage not reaching motors or boards (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Litter will not lower electrically patient safety
Bed mattress platform cannot be lowered using electric solenoid valves
Ranked causes
- 1. Blown fuse #3 at power board (common)
- 2. Insufficient voltage at solenoid valves (occasional)
- 3. Control switch board malfunction (occasional)
- 4. Insufficient voltage to boards (rare)
Measure voltage at solenoid terminals
- Test point
- Solenoid valve terminals on junction board
- Expected
- 99 V(AC) minimum, >= 99 V
- Tools
- Multimeter
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Weigh system not accurate
Weight measurement system provides inaccurate or unreliable measurements
Ranked causes
- 1. Physical binding on litter top (common)
- 2. Load cell malfunction (occasional)
- 3. Load cell calibration drift (occasional)
- 4. Scale CPU board failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed unresponsive to foot end keypad patient safety
Foot positioning commands not executed; other keypads respond normally
Ranked causes
- 1. Foot end switch board defect (common)
- 2. Foot end keypad cable disconnection (occasional)
- 3. Litter control board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed unresponsive to head end keypad patient safety
Head positioning commands not executed; other keypads respond normally
Ranked causes
- 1. Head end switch board defect (common)
- 2. Head end keypad cable disconnection (occasional)
- 3. Litter control board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed unresponsive to patient controls patient safety
Patient control commands not executed; other keypads respond normally
Ranked causes
- 1. Patient control disconnection or cable fault (common)
- 2. Patient control or cable malfunction (occasional)
- 3. Litter control board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Isolated problem with Knee Gatch or Fowler motor movement patient safety
Single positioning function (Knee Gatch or Fowler) not responding to commands
Ranked causes
- 1. Blown fuse #2 at power board (common)
- 2. Motor or connector issue (identify via isolation) (occasional)
- 3. Power board malfunction affecting that motor (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Litter will not lift when foot pedal pumped
Hydraulic litter does not lift when foot pedals are pumped
Ranked causes
- 1. Low hydraulic fluid level (common)
- 2. Air in hydraulic system (common)
- 3. Pump pedal cylinder malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scale system will not turn on caution
Integrated weighing scale system cannot be powered on
Ranked causes
- 1. Scale system lockout toggle switch activated (common)
- 2. Scale system switch board defect (occasional)
- 3. Scale CPU board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Weigh system shuts off prematurely
Weight measurement system stops functioning before expected or during measurement
Ranked causes
- 1. Weigh cable connection issue (common)
- 2. Load cell malfunction (occasional)
- 3. Scale CPU board failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed completely unresponsive with dead displays patient safety
No response to any commands and all displays are off
Ranked causes
- 1. Blown main fuse at junction board (common)
- 2. No power to litter control or power boards (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Faulty load cell suspected
Load cell output is jumpy, drifting, displays 000.0 or 999.9, or weight readings are inconsistent
Ranked causes
- 1. Load cell assembly malfunction (common)
- 2. Load cell cable fault (common)
Enter diagnostic mode
- Test point
- Scale monitor
- Expected
- DOWNLOAD BARCODE displayed
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Fowler and Knee Gatch both move when one activated patient safety
Dual movement when only one function is selected by operator
Ranked causes
- 1. Patient control malfunction or cross-connection (common)
- 2. Litter control board malfunction causing cross-switching (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fowler and/or Knee Gatch does not lower during Emergency Drop mode
Fowler and/or knee section of bed fails to lower during emergency drop operation
Ranked causes
- 1. Bellows cap switch malfunction or misposition (common)
- 2. Switch capability issue (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motor sounds slushy with non-smooth lift patient safety
Motor produces abnormal fluid noise and lift motion is jerky or uneven
Ranked causes
- 1. Air in hydraulic system (common)
- 2. Air bubbles in suction line causing motor damage (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scale system requires calibration
Load cell assembly or scale CPU replaced; scale readings need adjustment
Ranked causes
- 1. Multiple load cell assemblies replaced (occasional)
- 2. Scale CPU replaced (occasional)
Remove scale system power
- Test point
- Scale lockout switch
- Expected
- Scale lockout switch toward foot of bed
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Steer wheel does not engage or disengage properly
Steer wheel fails to engage or disengage correctly
Ranked causes
- 1. Steer wheel linkage misalignment (common)
- 2. Worn tension spring (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trendelenburg and/or Reverse Trendelenburg does not work in lowest position
Head-down or head-up positioning function fails when bed is in lowest position
Ranked causes
- 1. Bellows cap switch malfunction or misposition (common)
- 2. Solenoid malfunction (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Fowler display LED alignment
90 degree LED on Fowler display not properly lit or positioned
Ranked causes
- 1. Fowler potentiometer electrical alignment incorrect (common)
Adjust R8 potentiometer until 90 degree LED on Fowler display lights
- Test point
- R8 on litter control board, Fowler display LED
- Expected
- 90 degree LED lighted and centered in operating range
- Tools
- Jewelers Screwdriver, Red Lacquer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Fowler potentiometer alignment
LED alignment incorrect on Fowler display or after replacing litter control board
Ranked causes
- 1. Fowler potentiometer shaft misaligned mechanically (common)
Measure resistance between black wire and green wire on Fowler pot cable
- Test point
- Fowler pot cable, black and green wires
- Expected
- between 55 and 65 ohms, 55–65 ohms
- Tools
- Volt/Ohm Meter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Trendelenburg potentiometer alignment
Trendelenburg display alignment or potentiometer linkage misaligned
Ranked causes
- 1. Trendelenburg potentiometer shaft misaligned (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Weight system
Perform calibration procedure for the weight measurement system on the Renaissance 2020 hospital bed
Ranked causes
- 1. Weight system requires calibration (unknown)
Verify litter surface is flat before placing calibration weight
- Test point
- Litter surface
- Expected
- Flat
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiac Chair does not achieve proper positioning
Cardiac chair section fails to reach or maintain proper positioning
Ranked causes
- 1. Potentiometer adjustment incorrect (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Defective load cell identification
Isolate which load cell assembly is malfunctioning using corner weight display
Ranked causes
- 1. Individual load cell assembly failure (occasional)
Access Display Corner function
- Test point
- Diagnostic mode menu
- Expected
- Display Corner function available in diagnostic mode
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Emergency Drop delay longer than 3 seconds
Emergency drop function takes longer than 3 seconds to execute
Ranked causes
- 1. Litter control board malfunction (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LED position lights not accurate
LED indicators for bed position do not correctly reflect actual bed position
Ranked causes
- 1. Potentiometer position incorrect (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Litter will not lift electrically patient safety
Bed mattress platform cannot be raised using electric motor
Ranked causes
- 1. Motor capacitor failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Litter will not lower with foot pedals depressed
Hydraulic litter does not lower when foot pedals are depressed
Ranked causes
- 1. Valve control linkage misalignment (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Load cell assembly requires configuration
Single load cell assembly replaced; scale requires LBS/Volts configuration update
Ranked causes
- 1. Single load cell assembly replaced (common)
Enter diagnostic mode
- Test point
- Scale monitor
- Expected
- Scale monitor displays DOWNLOAD BARCODE
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No positioning functions with power lights still on patient safety
Fowler or Knee Gatch positioning not responding while main power is active
Ranked causes
- 1. Fowler or Gatch safety switch tape obstructed or damaged (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power board hums
Power board produces audible humming sound
Ranked causes
- 1. Power board failure or component issue (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Steering wheel engagement loss or excessive friction
Steer caster lever engagement not engaging or disengaging properly during steering operation
Ranked causes
- 1. Steer caster linkage lever engagement out of adjustment (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Weight system calibration not proceeding
Scale display does not show expected ADD LBS HIT ON message after correcting reference number
Ranked causes
- 1. Reference number not correctly entered or confirmed (common)
Verify reference number was correctly entered before pressing WEIGHT SYSTEM ON
- Test point
- Display screen
- Expected
- ADD LBS, HIT ON
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.