Stryker PneumoSure: troubleshooting and error codes
The ServiceForge knowledge base holds 18 documented symptoms for the Stryker PneumoSure, with 20 ranked causes, 12 diagnostic checks with test points and expected values, and 60 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Contamination alarm triggered patient safety
Device displays contamination alarm, may occur after device contamination, alarm testing, or manipulation during service or maintenance
Ranked causes
- 1. Erroneous manipulation or service task (e.g., pulled plug or cleaning) (common)
- 2. Device contamination during patient use (occasional)
Determine trigger circumstance of contamination alarm
- Test point
- Review recent device operation history
- Expected
- Alarm occurred during service, maintenance, or testing (not during patient use)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SIDNE interface malfunction or speech control failure
SIDNE interface damaged or speech recognition/control not functioning
Ranked causes
- 1. SIDNE interface damaged or defective (common)
- 2. IFM board malfunction (occasional)
Verify SIDNE interface connector and cable integrity
- Test point
- ST2 plug, capacitor connection (3)
- Expected
- All connectors seated and cable intact
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
12-hr Test performed
Extended test with device continuously connected to power supply for 12 hours without operation. Device not used or operated during test.
Ranked causes
- 1. 12-hr test required for service or validation (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Flow
Flow calibration procedure using 0-50 l/min flowmeter and calibration adapter.
Ranked causes
- 1. Flow calibration required (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: HIGH Pressure high voltage
HIGH pressure calibration procedure using 0-100 bar manometer and CO2 gas bottle with pressure greater than 50 bar.
Ranked causes
- 1. HIGH pressure calibration required (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: LOW Pressure
LOW pressure calibration procedure using 0-100 mm Hg manometer, calibration adapter, flowmeter, and LSV regulator.
Ranked causes
- 1. LOW pressure calibration required (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Low Pressure Regulator
Calibration of the Low Pressure Regulator (LPR) using calibration adapter, open tube end, LPR regulator, and LSV regulator.
Ranked causes
- 1. LPR calibration required (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Low Pressure Safety Valve
Calibration of the Low Pressure Safety Valve (LSV) using calibration adapter and LSV regulator.
Ranked causes
- 1. LSV calibration required (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Offset
Offset calibration procedure using flowmeter, small diameter tube, three-port valve, and calibration adapter.
Ranked causes
- 1. Offset calibration required during service (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Touch Screen
Touch screen calibration required after MIO board or LCD replacement; initiated by touching screen for 2 seconds during device check, then tapping black crosshairs.
Ranked causes
- 1. MIO board or MIO LCD replacement (unknown)
Touch screen during device check for 2 seconds
- Test point
- Any location on touch screen
- Expected
- Calibration process initiates
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dark touch screen on BAM board
Touch screen display is dark or not illuminated; green LED no longer lights up; no LEDs on board are illuminated
Ranked causes
- 1. BAM board malfunction or failure (common)
Verify BAM board is installed and all connectors are seated
- Test point
- ST3 plug, ST1 plug, ST2 plug, ST4 plug
- Expected
- All plugs connected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device front panel damaged or misaligned caution
Device front panel cracked, bent, or not seating properly on case
Ranked causes
- 1. Device front panel physical damage or fastener failure (occasional)
Verify all fastening screws and cap nuts are present and tight
- Test point
- Bottom device screws (8), rear cap nuts (3)
- Expected
- All fasteners present and secure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluid sensor defective or malfunctioning
Fluid sensor not detecting or measuring fluid level correctly
Ranked causes
- 1. Fluid sensor failure (common)
Verify fluid sensor connection and wiring
- Test point
- ST6 plug, PCU board connector
- Expected
- Plug seated, cable intact
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas tube heating malfunction or tube connection damaged caution
Gas heating element not functioning or tube connection physically damaged
Ranked causes
- 1. HPL board malfunction (common)
Verify HPL board is installed and ST1 plug is connected
- Test point
- ST1 plug/cable, board fastening screws
- Expected
- Plug seated, board fastened securely
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reset to Factory Defaults required caution
Device severely mis-calibrated and cannot be recalibrated normally. Restores all factory settings; device will not be calibrated and cannot be used until recalibrated.
Ranked causes
- 1. Device severely mis-calibrated beyond recovery (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Snap lock assembly damaged or misaligned caution
Snap lock not engaging properly or light barrier elements misaligned; tube detection not working
Ranked causes
- 1. Snap lock physical damage or misalignment (occasional)
Verify snap lock positioning and light barrier element alignment
- Test point
- Snap lock (2), light barrier element A (6), light barrier element B
- Expected
- Snap lock seated, light barriers aligned per Fig. 6-38
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tube connection brass ring stuck or damaged caution
Tube connection brass ring cannot be turned or is physically damaged; tube connection leaking
Ranked causes
- 1. Tube connection brass ring stuck or damaged threads (occasional)
Verify tube connection brass ring can turn freely with assembly key
- Test point
- Tube connection brass ring (5)
- Expected
- Ring turns smoothly counterclockwise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tube detection board malfunction
Electronic tube detection not functioning; tube not detected when inserted or incorrectly detected
Ranked causes
- 1. Tube detection board failure or connector issue (occasional)
Verify tube detection board connectors and wiring
- Test point
- Plug ST2 (4), plug ST1 (5), board fastening nuts (1)
- Expected
- All plugs seated, board fastened securely
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0 | General error | Unknown | error |
1 | Timeout | Unknown | error |
10 | Offset I low pressure sensor too low | Sensor calibration error | error |
11 | Offset II low pressure sensor too low | Sensor calibration error | error |
110 | General calibration error, low pressure safety valve | Safety valve calibration failure | critical |
111 | Adjustment, low pressure safety valve not within range | Safety valve adjustment out of specification | warning |
12 | Offset safety pressure sensor too low | Sensor calibration error | error |
13 | Offset differential pressure sensor too low | Sensor calibration error | error |
130 | General error, flow | Flowmeter or sensor failure | error |
131 | Flow resistance too high | Flow measurement indicates obstruction | error |
133 | Flow input value too low | Flow calibration measurement out of range | error |
134 | Flow input value too high | Flow calibration measurement out of range | error |
135 | Modified flow sensor value | Unexpected flow sensor reading | warning |
136 | Flow sensor amplification too low | Sensor amplification calibration error | error |
137 | Flow sensor amplification too high | Sensor amplification calibration error | error |
138 | Offset proportional valve too low | Proportional valve calibration error | error |
139 | Offset proportional valve too high | Proportional valve calibration error | error |
14 | Offset high pressure sensor too low | Sensor calibration error | error |
140 | Hysteresis proportional valve too low | Proportional valve calibration error | error |
141 | Hysteresis proportional valve too high | Proportional valve calibration error | error |
142 | Proportional valve amplification too low | Proportional valve amplification calibration error | error |
143 | Proportional valve amplification too high | Proportional valve amplification calibration error | error |
144 | Sticky proportional valve | Proportional valve response lag detected | error |
15 | Offset temperature sensor too low | Sensor calibration error | error |
16 | Offset current monitor of high pressure gas heating too low | Sensor calibration error | error |
17 | Offset gas heating too low | Sensor calibration error | error |
2 | General software error | Unknown | error |
20 | Offset I low pressure sensor too high | Sensor calibration error | error |
21 | Offset II low pressure sensor too high | Sensor calibration error | error |
22 | Offset safety pressure sensor too high | Sensor calibration error | error |
23 | Offset differential pressure sensor too high | Sensor calibration error | error |
24 | Offset high pressure sensor too high | Sensor calibration error | error |
25 | Offset temperature sensor too high | Sensor calibration error | error |
255 | Test time too long | Calibration procedure exceeded time limit | warning |
26 | Offset current monitor of high pressure gas heating too high | Sensor calibration error | error |
27 | Offset high pressure gas heating too high | Sensor calibration error | error |
3 | EEPROM write error | Unknown | error |
4 | General gas supply error | Unknown | warning |
50 | General calibration error, low pressure | Low pressure calibration failure | error |
51 | Low pressure – large leak | Detected during calibration | warning |
52 | Low pressure – small leak | Detected during calibration | warning |
53 | Low pressure input value too low | Calibration measurement out of range | error |
54 | Low pressure input value too high | Calibration measurement out of range | error |
55 | Low pressure sensor amplification I too low | Sensor amplification calibration error | error |
56 | Low pressure sensor amplification I too high | Sensor amplification calibration error | error |
57 | Low pressure sensor amplification II too low | Sensor amplification calibration error | error |
58 | Low pressure sensor amplification II too high | Sensor amplification calibration error | error |
59 | Safety pressure sensor amplification too low | Sensor amplification calibration error | error |
60 | Low pressure safety pressure sensor amplification too high | Sensor amplification calibration error | error |
61 | Proportional valve malfunction | Unknown | error |
62 | Patient pressure sensor amplification too low | Sensor amplification calibration error | error |
63 | Patient pressure sensor amplification too high | Sensor amplification calibration error | error |
70 | General calibration error, high pressure | High pressure calibration failure | error |
73 | High pressure input value too low | Calibration measurement out of range | error |
74 | High pressure input value too high | Calibration measurement out of range | error |
75 | Modified high pressure sensor value | Unexpected sensor reading | warning |
76 | High pressure sensor amplification too low | Sensor amplification calibration error | error |
77 | High pressure sensor amplification too high | Sensor amplification calibration error | error |
90 | General calibration error, low pressure regulator | LPR calibration failure | error |
91 | Adjustment, low pressure regulator not within range | LPR adjustment out of specification | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.