Stryker PneumoSure: troubleshooting and error codes

The ServiceForge knowledge base holds 18 documented symptoms for the Stryker PneumoSure, with 20 ranked causes, 12 diagnostic checks with test points and expected values, and 60 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Contamination alarm triggered patient safety

Device displays contamination alarm, may occur after device contamination, alarm testing, or manipulation during service or maintenance

Ranked causes

  1. 1. Erroneous manipulation or service task (e.g., pulled plug or cleaning) (common)
  2. 2. Device contamination during patient use (occasional)
First check

Determine trigger circumstance of contamination alarm

Test point
Review recent device operation history
Expected
Alarm occurred during service, maintenance, or testing (not during patient use)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SIDNE interface malfunction or speech control failure

SIDNE interface damaged or speech recognition/control not functioning

Ranked causes

  1. 1. SIDNE interface damaged or defective (common)
  2. 2. IFM board malfunction (occasional)
First check

Verify SIDNE interface connector and cable integrity

Test point
ST2 plug, capacitor connection (3)
Expected
All connectors seated and cable intact

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

12-hr Test performed

Extended test with device continuously connected to power supply for 12 hours without operation. Device not used or operated during test.

Ranked causes

  1. 1. 12-hr test required for service or validation (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Flow

Flow calibration procedure using 0-50 l/min flowmeter and calibration adapter.

Ranked causes

  1. 1. Flow calibration required (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: HIGH Pressure high voltage

HIGH pressure calibration procedure using 0-100 bar manometer and CO2 gas bottle with pressure greater than 50 bar.

Ranked causes

  1. 1. HIGH pressure calibration required (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: LOW Pressure

LOW pressure calibration procedure using 0-100 mm Hg manometer, calibration adapter, flowmeter, and LSV regulator.

Ranked causes

  1. 1. LOW pressure calibration required (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Low Pressure Regulator

Calibration of the Low Pressure Regulator (LPR) using calibration adapter, open tube end, LPR regulator, and LSV regulator.

Ranked causes

  1. 1. LPR calibration required (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Low Pressure Safety Valve

Calibration of the Low Pressure Safety Valve (LSV) using calibration adapter and LSV regulator.

Ranked causes

  1. 1. LSV calibration required (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Offset

Offset calibration procedure using flowmeter, small diameter tube, three-port valve, and calibration adapter.

Ranked causes

  1. 1. Offset calibration required during service (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Touch Screen

Touch screen calibration required after MIO board or LCD replacement; initiated by touching screen for 2 seconds during device check, then tapping black crosshairs.

Ranked causes

  1. 1. MIO board or MIO LCD replacement (unknown)
First check

Touch screen during device check for 2 seconds

Test point
Any location on touch screen
Expected
Calibration process initiates

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Dark touch screen on BAM board

Touch screen display is dark or not illuminated; green LED no longer lights up; no LEDs on board are illuminated

Ranked causes

  1. 1. BAM board malfunction or failure (common)
First check

Verify BAM board is installed and all connectors are seated

Test point
ST3 plug, ST1 plug, ST2 plug, ST4 plug
Expected
All plugs connected

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device front panel damaged or misaligned caution

Device front panel cracked, bent, or not seating properly on case

Ranked causes

  1. 1. Device front panel physical damage or fastener failure (occasional)
First check

Verify all fastening screws and cap nuts are present and tight

Test point
Bottom device screws (8), rear cap nuts (3)
Expected
All fasteners present and secure

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fluid sensor defective or malfunctioning

Fluid sensor not detecting or measuring fluid level correctly

Ranked causes

  1. 1. Fluid sensor failure (common)
First check

Verify fluid sensor connection and wiring

Test point
ST6 plug, PCU board connector
Expected
Plug seated, cable intact

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas tube heating malfunction or tube connection damaged caution

Gas heating element not functioning or tube connection physically damaged

Ranked causes

  1. 1. HPL board malfunction (common)
First check

Verify HPL board is installed and ST1 plug is connected

Test point
ST1 plug/cable, board fastening screws
Expected
Plug seated, board fastened securely

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Reset to Factory Defaults required caution

Device severely mis-calibrated and cannot be recalibrated normally. Restores all factory settings; device will not be calibrated and cannot be used until recalibrated.

Ranked causes

  1. 1. Device severely mis-calibrated beyond recovery (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Snap lock assembly damaged or misaligned caution

Snap lock not engaging properly or light barrier elements misaligned; tube detection not working

Ranked causes

  1. 1. Snap lock physical damage or misalignment (occasional)
First check

Verify snap lock positioning and light barrier element alignment

Test point
Snap lock (2), light barrier element A (6), light barrier element B
Expected
Snap lock seated, light barriers aligned per Fig. 6-38

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tube connection brass ring stuck or damaged caution

Tube connection brass ring cannot be turned or is physically damaged; tube connection leaking

Ranked causes

  1. 1. Tube connection brass ring stuck or damaged threads (occasional)
First check

Verify tube connection brass ring can turn freely with assembly key

Test point
Tube connection brass ring (5)
Expected
Ring turns smoothly counterclockwise

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tube detection board malfunction

Electronic tube detection not functioning; tube not detected when inserted or incorrectly detected

Ranked causes

  1. 1. Tube detection board failure or connector issue (occasional)
First check

Verify tube detection board connectors and wiring

Test point
Plug ST2 (4), plug ST1 (5), board fastening nuts (1)
Expected
All plugs seated, board fastened securely

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
0General errorUnknownerror
1TimeoutUnknownerror
10Offset I low pressure sensor too lowSensor calibration errorerror
11Offset II low pressure sensor too lowSensor calibration errorerror
110General calibration error, low pressure safety valveSafety valve calibration failurecritical
111Adjustment, low pressure safety valve not within rangeSafety valve adjustment out of specificationwarning
12Offset safety pressure sensor too lowSensor calibration errorerror
13Offset differential pressure sensor too lowSensor calibration errorerror
130General error, flowFlowmeter or sensor failureerror
131Flow resistance too highFlow measurement indicates obstructionerror
133Flow input value too lowFlow calibration measurement out of rangeerror
134Flow input value too highFlow calibration measurement out of rangeerror
135Modified flow sensor valueUnexpected flow sensor readingwarning
136Flow sensor amplification too lowSensor amplification calibration errorerror
137Flow sensor amplification too highSensor amplification calibration errorerror
138Offset proportional valve too lowProportional valve calibration errorerror
139Offset proportional valve too highProportional valve calibration errorerror
14Offset high pressure sensor too lowSensor calibration errorerror
140Hysteresis proportional valve too lowProportional valve calibration errorerror
141Hysteresis proportional valve too highProportional valve calibration errorerror
142Proportional valve amplification too lowProportional valve amplification calibration errorerror
143Proportional valve amplification too highProportional valve amplification calibration errorerror
144Sticky proportional valveProportional valve response lag detectederror
15Offset temperature sensor too lowSensor calibration errorerror
16Offset current monitor of high pressure gas heating too lowSensor calibration errorerror
17Offset gas heating too lowSensor calibration errorerror
2General software errorUnknownerror
20Offset I low pressure sensor too highSensor calibration errorerror
21Offset II low pressure sensor too highSensor calibration errorerror
22Offset safety pressure sensor too highSensor calibration errorerror
23Offset differential pressure sensor too highSensor calibration errorerror
24Offset high pressure sensor too highSensor calibration errorerror
25Offset temperature sensor too highSensor calibration errorerror
255Test time too longCalibration procedure exceeded time limitwarning
26Offset current monitor of high pressure gas heating too highSensor calibration errorerror
27Offset high pressure gas heating too highSensor calibration errorerror
3EEPROM write errorUnknownerror
4General gas supply errorUnknownwarning
50General calibration error, low pressureLow pressure calibration failureerror
51Low pressure – large leakDetected during calibrationwarning
52Low pressure – small leakDetected during calibrationwarning
53Low pressure input value too lowCalibration measurement out of rangeerror
54Low pressure input value too highCalibration measurement out of rangeerror
55Low pressure sensor amplification I too lowSensor amplification calibration errorerror
56Low pressure sensor amplification I too highSensor amplification calibration errorerror
57Low pressure sensor amplification II too lowSensor amplification calibration errorerror
58Low pressure sensor amplification II too highSensor amplification calibration errorerror
59Safety pressure sensor amplification too lowSensor amplification calibration errorerror
60Low pressure safety pressure sensor amplification too highSensor amplification calibration errorerror
61Proportional valve malfunctionUnknownerror
62Patient pressure sensor amplification too lowSensor amplification calibration errorerror
63Patient pressure sensor amplification too highSensor amplification calibration errorerror
70General calibration error, high pressureHigh pressure calibration failureerror
73High pressure input value too lowCalibration measurement out of rangeerror
74High pressure input value too highCalibration measurement out of rangeerror
75Modified high pressure sensor valueUnexpected sensor readingwarning
76High pressure sensor amplification too lowSensor amplification calibration errorerror
77High pressure sensor amplification too highSensor amplification calibration errorerror
90General calibration error, low pressure regulatorLPR calibration failureerror
91Adjustment, low pressure regulator not within rangeLPR adjustment out of specificationwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI