Stryker Neptune: troubleshooting and error codes
The ServiceForge knowledge base holds 19 documented symptoms for the Stryker Neptune, with 32 ranked causes, 28 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Inappropriate device use for closed drainage or post-operative applications patient safety
Device used in applications for which it is not intended, such as closed wound drainage systems or post-operative suctioning
Ranked causes
- 1. Neptune used with closed wound drainage systems (rare)
- 2. Neptune used for post-operative suction (rare)
- 3. Neptune used to power other suction-powered accessories (rare)
Verify application is appropriate for Neptune device
- Test point
- Procedure type and suction requirements
- Expected
- Procedure requires high-level suction to tissue (not closed drainage, not post-operative)
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inappropriate use of Neptune with incompatible accessories or drainage systems patient safety
Application of Neptune high level suction to Pleur Evac devices, closed wound drainage systems, or post-operative settings where such use is contraindicated
Ranked causes
- 1. Neptune used with Pleur Evac or similar powered suction accessories (occasional)
- 2. Neptune applied to closed wound drainage systems (occasional)
- 3. Neptune used in post-operative setting (occasional)
Verify intended application is not powered suction accessory
- Test point
- Procedure specification and accessory documentation
- Expected
- Neptune to be used only for direct tissue or cavity suction, not powering other devices
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tissue damage from high suction application patient safety
Injury to vital anatomical structures or hemorrhage resulting from inappropriate high flow suction exposure during procedures requiring low or no suction
Ranked causes
- 1. High flow suction applied to procedures requiring low or no suction (e.g., passive chest drainage) (common)
- 2. Operator unfamiliar with tissue type and appropriate suction level for procedure (common)
- 3. Suction level not verified or checked prior to use (common)
Identify tissue type for planned procedure
- Test point
- Procedure documentation
- Expected
- Tissue type documented and reviewed
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tissue trauma or hemorrhage during suction use patient safety
Injury to vital anatomical structures, hemorrhage during surgical procedures using Neptune high level suction devices
Ranked causes
- 1. High flow suction applied to procedures requiring low or no suction (common)
- 2. Device used with inappropriate applications (wound drainage, post-operative use) (occasional)
- 3. Inadequate user training or awareness of device capabilities (common)
Verify suction level is appropriate for planned procedure before activation
- Test point
- Device control panel
- Expected
- Suction level matches procedure requirements (high level for appropriate procedures, low level for sensitive tissue areas)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tissue trauma or hemorrhage from inappropriate suction application patient safety
Injury to vital anatomical structures or hemorrhage during surgical procedures, may result in serious injury or death
Ranked causes
- 1. High flow suction applied to tissue requiring low or no suction (common)
- 2. Inadequate operator training on proper device use (common)
- 3. Operator unable to see or control suction level during procedure (occasional)
Verify suction level is appropriate for procedure type
- Test point
- Device display
- Expected
- mm-Hg setting appropriate for procedure
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High flow suction applied to low-suction procedures patient safety
Operator applies high flow suction or extends suction exposure to tissue during procedures requiring no, low, or minimal vacuum/flow (e.g., passive chest drainage), risking injury to vital structures.
Ranked causes
- 1. Operator misunderstanding of procedure suction requirements (common)
- 2. Inadequate operator training on Neptune device and applications (common)
Verify procedure suction level requirement before device use
- Test point
- Surgical procedure protocol documentation
- Expected
- Procedure requires no suction, low vacuum, or low flow suction (e.g., passive chest drainage)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inadequate operator training on Neptune device patient safety
Users unfamiliar with unit measurements, proper procedure selection, or safe operation of Neptune system
Ranked causes
- 1. User not trained on Neptune device operation and safety (common)
- 2. User unaware of unit of measure displayed on device (occasional)
Verify user has received Neptune training and competency certification
- Test point
- Facility training documentation and personnel master list
- Expected
- User name listed on approved training roster with documented competency
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Neptune used with incompatible suction systems patient safety
Device connected to other suction powered accessories or closed wound drainage systems; risk of device malfunction or patient harm
Ranked causes
- 1. Connection to incompatible suction powered accessories (occasional)
- 2. Use with closed wound drainage systems (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device connected to powered suction accessories or closed wound drainage patient safety
Neptune high-level suction used to power other suction devices or connected to closed wound drainage systems
Ranked causes
- 1. Unintended device application or configuration (occasional)
Verify Neptune is connected only to patient tissue, not to other suction devices or drainage systems
- Test point
- Suction tubing and connections
- Expected
- Direct connection to surgical field or cannula only
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device control panel obscured or not visible patient safety
Control panel covered by surgical drapes or other objects preventing operator visibility of suction level settings
Ranked causes
- 1. Inadequate setup or draping during procedure (occasional)
Confirm device control panel is unobstructed and clearly visible before procedure starts
- Test point
- Device control panel view from operator position
- Expected
- Panel fully visible and readable without moving drapes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device used for post-operative suction patient safety
Neptune high-level suction device applied to post-operative wound drainage
Ranked causes
- 1. User unclear on indicated applications for Neptune device (common)
Confirm procedure is intra-operative and not post-operative before using Neptune
- Test point
- Procedure type classification
- Expected
- Active intra-operative suction application only, not post-operative drainage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
High suction applied to low-suction procedures patient safety
Excessive suction applied to procedures requiring no suction, low vacuum or low flow (e.g., passive chest drainage); risk of injury to vital structures
Ranked causes
- 1. Inappropriate suction level selection for procedure type (common)
Verify suction level is appropriate for planned procedure
- Test point
- Device control panel display
- Expected
- mm-Hg setting visible and suction level matches procedure requirements
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
High-flow suction applied to unsuitable tissue patient safety
High flow suction or extended suction exposure applied to tissue in procedures requiring no, low vacuum, or low flow suction such as passive chest drainage
Ranked causes
- 1. Inadequate user training on appropriate suction levels for procedure type (common)
Verify appropriate suction level is selected for planned procedure
- Test point
- Device control panel
- Expected
- Suction level matches procedure requirements (no, low vacuum, or low flow as needed)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inadequate user training on Neptune operation patient safety
Users unfamiliar with device operation, applications, and risks; potential for misuse and patient injury
Ranked causes
- 1. Insufficient training of surgical personnel (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing or illegible warning labels on Neptune device patient safety
Warning labels not present or not visible on device, preventing users from seeing critical safety information
Ranked causes
- 1. Warning labels missing, damaged, or obscured (occasional)
Inspect device for warning labels
- Test point
- All surfaces of Neptune device
- Expected
- All warning labels present, legible, and not obscured by drapes or objects
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Neptune device used for post-operative wound suction patient safety
Operator applies Neptune high-level suction device for post-operative wound drainage or suctioning, which is not an approved application.
Ranked causes
- 1. Operator attempts to use Neptune for post-operative wound management (occasional)
Confirm procedure phase before selecting suction device
- Test point
- Surgical timeline and procedure phase documentation
- Expected
- Procedure is intra-operative; not post-operative wound management or drainage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Neptune device used with closed wound drainage systems patient safety
Operator connects Neptune to closed wound drainage system (e.g., PleurEvac), which is not an approved application for this high-level suction device.
Ranked causes
- 1. Operator applies Neptune to unapproved drainage system applications (occasional)
Verify intended suction application before connection
- Test point
- Procedure order and suction requirement specification
- Expected
- Application is high-level surgical suction only; not wound drainage, PleurEvac connection, or post-operative use
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Neptune used for post-operative suction patient safety
Device applied to post-operative patient care; risk of adverse outcomes and patient injury
Ranked causes
- 1. Inappropriate application to post-operative phase (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unclear understanding of displayed units of measure patient safety
User unsure of pressure units displayed on Neptune control panel, potentially selecting incorrect suction levels
Ranked causes
- 1. Insufficient user training on unit selection and display (occasional)
Verify displayed pressure unit is set to mm-Hg
- Test point
- Device control panel unit indicator
- Expected
- mm-Hg displayed as active unit
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI