Stryker FL27: troubleshooting and error codes
The ServiceForge knowledge base holds 49 documented symptoms for the Stryker FL27, with 89 ranked causes, 134 diagnostic checks with test points and expected values, and 210 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unit does not power on high voltage
Bed is plugged in but receives no power; wall outlet has correct country voltage
Ranked causes
- 1. Wall outlet lacks correct country voltage (occasional)
- 2. Main power fuses open (common)
- 3. No voltage at transformer connection J11 (occasional)
- 4. No 30VAC output from transformer (occasional)
- 5. CPU/Power board fuse F1 open or board faulty (rare)
Verify correct voltage at wall outlet
- Test point
- Wall outlet
- Expected
- Country voltage option present per CV table
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on high voltage
Bed is plugged in but has no power; wall outlet has mains voltage
Ranked causes
- 1. Wall outlet has no correct country voltage (common)
- 2. Main power fuse open (occasional)
- 3. Transformer connection cable issue or 25A fuse failure (occasional)
- 4. Transformer assembly failure (rare)
- 5. Fuse F1 or CPU/Power board failure (rare)
Verify bed is plugged in and outlet has correct country voltage
- Test point
- wall outlet
- Expected
- Country voltage option present at outlet
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiac chair positioning fails
Bed fails to achieve cardiac chair position (Fowler and foot positioning required)
Ranked causes
- 1. Fowler angle sensor miscalibration (common)
- 2. Foot section angle sensor miscalibration (common)
- 3. CPU/Power board failure (occasional)
Verify fowler and foot section angles on display
Electric brake does not release caution
Brake motor fails to unlock bed when brake button pressed
Ranked causes
- 1. Manual brake set in brake position (common)
- 2. Brake motor failure (occasional)
- 3. Brake switch malfunction (occasional)
Brake Not Set LED flashing and Brake Set LED OFF
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Foot extension does not lower caution
Foot section fails to move downward when control pressed; debris may block movement
Ranked causes
- 1. Debris in foot end traction sockets (common)
- 2. Motor stalled or needs recalibration (occasional)
- 3. Motor or CPU/Power board failure (occasional)
Inspect foot end traction sockets for debris
- Test point
- foot end traction sockets
- Expected
- No debris present
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot section does not raise
Foot lift motor fails to extend when up motion commanded in bed calibration mode
Ranked causes
- 1. Foot lift motor failure (common)
- 2. CPU/Power board failure (occasional)
- 3. Litter angle sensor miscalibration (occasional)
Enter Bed Calibration menu via head siderail trend button
4 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Head section does not raise
Head lift motor fails to extend when up motion commanded in calibration mode
Ranked causes
- 1. Head lift motor failure (common)
- 2. CPU/Power board failure (occasional)
- 3. Litter angle sensor miscalibration (occasional)
Enter Bed Calibration menu via head siderail
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Electric brake does not release caution
Brake pedal engaged but electric brake motor does not disengage when brake button pressed
Ranked causes
- 1. Brake pedal in brake position (correct operation) (common)
- 2. Brake motor or CPU/Power board failure (occasional)
Verify Brake Not Set LED is flashing and Brake Set LED is OFF
- Test point
- bed control panel LEDs
- Expected
- Brake Not Set LED flashing; Brake Set LED OFF
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot extension does not raise caution
Foot section fails to move upward when control pressed
Ranked causes
- 1. Motor stalled or needs recalibration (common)
- 2. Motor or CPU/Power board failure (occasional)
Enter Bed Calibration menu and press foot up button; observe motor
- Test point
- foot extension motor
- Expected
- Motor runs upward
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot lift section does not lower caution
Foot section fails to move downward when control pressed
Ranked causes
- 1. Motor stalled or needs recalibration (common)
- 2. Motor or CPU/Power board failure (occasional)
Enter Bed Calibration menu and press reverse trend button; observe motor
- Test point
- foot lift motor
- Expected
- Motor runs downward
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot lift section does not raise caution
Foot section fails to move upward when control pressed; may occur after other motion failures
Ranked causes
- 1. Motor stalled or needs recalibration (common)
- 2. Motor or CPU/Power board failure (occasional)
Enter Bed Calibration menu and press trend button; observe motor
- Test point
- foot lift motor
- Expected
- Motor runs upward
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot positioning does not lower
Foot positioning motor fails to retract when down motion commanded
Ranked causes
- 1. Foot motor failure (common)
- 2. CPU/Power board failure (occasional)
24VDC at J1 (black pin 1, red pin 2) when down pressed
Foot positioning does not raise
Foot positioning motor fails to extend when up motion commanded
Ranked causes
- 1. Foot motor failure (common)
- 2. CPU/Power board failure (occasional)
Motor runs in calibration mode
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Foot section does not lower
Foot lift motor fails to retract when down motion commanded
Ranked causes
- 1. Foot lift motor failure (common)
- 2. CPU/Power board failure (occasional)
24VDC at J6 (red pin 1, black pin 2) when reverse trend button pressed
Fowler section does not lower caution
Fowler section fails to move downward when control pressed
Ranked causes
- 1. Motor stalled or needs recalibration (common)
- 2. Motor or CPU/Power board failure (occasional)
Enter Bed Calibration menu and press fowler down button; observe motor
- Test point
- fowler motor
- Expected
- Motor runs downward
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fowler section does not lower
Fowler motor fails to retract when down motion commanded
Ranked causes
- 1. Fowler motor failure (common)
- 2. CPU/Power board failure (occasional)
24VDC at J5 (red pin 1, black pin 2) when down button pressed
Fowler section does not raise caution
Fowler section fails to move upward when control pressed
Ranked causes
- 1. Motor stalled or needs recalibration (common)
- 2. Motor or CPU/Power board failure (occasional)
Enter Bed Calibration menu and press fowler up button; observe motor
- Test point
- fowler motor
- Expected
- Motor runs upward
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fowler section does not raise
Fowler motor fails to extend when up motion commanded
Ranked causes
- 1. Fowler motor failure (common)
- 2. CPU/Power board failure (occasional)
Motor runs in calibration mode
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Gatch section does not lower
Gatch motor fails to retract when down motion commanded
Ranked causes
- 1. Gatch motor failure (common)
- 2. CPU/Power board failure (occasional)
24VDC at J3 (black pin 1, red pin 2) when down pressed
Gatch section does not lower caution
Gatch section fails to move downward when control pressed
Ranked causes
- 1. Motor stalled or needs recalibration (common)
- 2. Motor or CPU/Power board failure (occasional)
Enter Bed Calibration menu and press gatch down button; observe motor
- Test point
- gatch motor
- Expected
- Motor runs downward
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gatch section does not raise
Gatch motor fails to extend when up motion commanded
Ranked causes
- 1. Gatch motor failure (common)
- 2. CPU/Power board failure (occasional)
Motor runs in calibration mode
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Gatch section does not raise caution
Gatch section fails to move upward when control pressed
Ranked causes
- 1. Motor stalled or needs recalibration (common)
- 2. Motor or CPU/Power board failure (occasional)
Enter Bed Calibration menu and press gatch up button; observe motor
- Test point
- gatch motor
- Expected
- Motor runs upward
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head lift section does not lower caution
Head section fails to move downward when control pressed
Ranked causes
- 1. Motor stalled or needs recalibration (common)
- 2. Motor or CPU/Power board failure (occasional)
Enter Bed Calibration menu and press bed down button; observe motor
- Test point
- head lift motor
- Expected
- Motor runs downward
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head lift section does not raise caution
Head section fails to move upward when control pressed
Ranked causes
- 1. Motor stalled or needs recalibration (common)
- 2. Motor or CPU/Power board failure (occasional)
Enter Bed Calibration menu and press bed up button; observe motor
- Test point
- head lift motor
- Expected
- Motor runs upward
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head section does not lower
Head lift motor fails to retract when down motion commanded
Ranked causes
- 1. Head lift motor failure (common)
- 2. CPU/Power board failure (occasional)
24VDC at J4 (red pin 1, black pin 2) when bed down button pressed
HOB 300 position not achieved
Head of bed angle does not position correctly to 30 degrees or display inaccurate
Ranked causes
- 1. Fowler angle sensor miscalibration (common)
- 2. CPU/Power board failure (occasional)
Verify fowler angle display accuracy
Reverse Trendelenburg positioning fails
Bed fails to tilt into Reverse Trendelenburg position or angle sensor inaccurate
Ranked causes
- 1. Trend angle sensor miscalibration (common)
- 2. CPU/Power board failure (occasional)
Verify touch screen angle display accuracy
Trendelenburg positioning fails
Bed fails to tilt into Trendelenburg position or angle sensor inaccurate
Ranked causes
- 1. Trend angle sensor miscalibration (common)
- 2. CPU/Power board failure (occasional)
Check touch screen trend angle display accuracy
Vascular positioning fails
Bed fails to achieve vascular position or multiple angle sensors inaccurate
Ranked causes
- 1. Multiple angle sensors miscalibrated (common)
- 2. CPU/Power board failure (occasional)
Verify fowler, gatch, foot, trend display accuracy
Bed display shows non-zero angle readings at 16 inch height
Display readings fluctuate or do not maintain zero degrees when bed is at lowest position
Ranked causes
- 1. Incomplete or failed initial calibration requiring full recalibration cycle (common)
Return to Configuration Screen
- Test point
- Footboard control panel Main Menu button
- Expected
- Configuration Screen displayed with Bed Calibration option
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed does not raise in Reverse Trendelenburg position caution
Bed fails to raise in Reverse Trendelenburg (head-up) position when control pressed
Ranked causes
- 1. Bed needs recalibration or trend angle sensor failure (common)
Verify trend angle display accuracy on touch screen
- Test point
- touch screen trend angle display
- Expected
- Accurate display of litter level
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed does not raise in Trendelenburg position caution
Bed fails to raise in Trendelenburg (head-down) position when control pressed
Ranked causes
- 1. Bed needs recalibration or trend angle sensor failure (common)
Verify trend angle display accuracy on touch screen
- Test point
- touch screen trend angle display
- Expected
- Accurate display of litter level
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed Options configuration not applied
Selected bed options such as iBed, iAudio, iDoc are not properly configured or saved on the bed system.
Ranked causes
- 1. Bed Options not confirmed or saved (common)
Verify Bed Options screen displays correct options
- Test point
- Bed Options screen
- Expected
- All applicable bed options visible and selectable
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed options not saved or incorrect after configuration
Selected bed options (iBed, iAudio, iDoc, Wi-Fi) not matching actual configuration or not persisting after setup
Ranked causes
- 1. Bed options not confirmed or power not cycled after selection (common)
Press and hold HOB 30 button and BRAKE button simultaneously for 5 seconds
- Test point
- Head end control buttons
- Expected
- Bed Options screen displays with list of available options
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Bed position and level sensors patient safety
Six-step bed calibration procedure to establish reference points for height, tilt, and levelness across all bed articulation positions
Ranked causes
- 1. Initial setup or sensor drift requiring recalibration (common)
Flatten all bed articulations to reference position
- Test point
- Litter surface position
- Expected
- Completely flat
10 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Touch screen
Validation and adjustment of touch screen coordinate calibration to ensure accurate input detection
Ranked causes
- 1. Touch screen calibration drift (occasional)
Press and hold stylus or finger on center of cross hair
- Test point
- Center of screen
- Expected
- Cross hair registers touch and moves to next location
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Touch screen caution
Procedure to recalibrate touch screen coordinates to ensure accurate touch recognition at all screen locations.
Ranked causes
- 1. Touch screen calibration initialization (unknown)
Press and hold on upper center target
- Test point
- Upper center of screen
- Expected
- Target cross-hair responds to touch
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiac chair position not achievable caution
Bed fails to reach cardiac chair configuration where fowler and foot sections coordinate
Ranked causes
- 1. Fowler or foot angle sensor needs recalibration or replacement (common)
Verify fowler and foot section angle display accuracy on touch screen
- Test point
- touch screen fowler and foot angle displays
- Expected
- Accurate angle displays for both sections
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Full diagnostic menu not accessible or shows incomplete information
Full Diagnostic button from Configuration Screen does not open menu or displays incomplete data in Boards, Error Codes, Input States, Motors, Buttons Pressed, or Signal Values screens
Ranked causes
- 1. CPU board or touch screen communication fault (occasional)
Press Full Diagnostic button from Configuration Screen
- Test point
- Configuration Screen
- Expected
- Full Diagnostic screen with menu items appears (Boards, Error Codes, Input States, Motors, Buttons Pressed, Signal Values)
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head of bed position angle inaccurate caution
Bed fails to achieve or maintain proper head of bed angle position
Ranked causes
- 1. Fowler angle sensor needs recalibration or replacement (common)
Verify fowler angle display accuracy on touch screen
- Test point
- touch screen fowler angle display
- Expected
- Accurate angle display
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Isolibrium active error displayed on support surface patient safety
Support surface showing active error code in Isolibrium diagnostic menu; up to five errors displayed simultaneously ordered by priority
Ranked causes
- 1. Support surface sensor, pressure or mechanical fault (common)
Access Isolibrium diagnostic menu and tap Active Errors tab
- Test point
- Service Diagnostic screen
- Expected
- List of up to five active errors displayed in priority order
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Isolibrium error history not clearing or showing incorrect past errors
Error History tab shows old errors not being removed, or errors remaining after Clear Error History button pressed, or errors appearing that should not be logged
Ranked causes
- 1. Active error preventing history clear or active error re-logged on clear (occasional)
Access Isolibrium diagnostic menu and tap Error History tab
- Test point
- Service Diagnostic screen
- Expected
- List of up to 10 errors displayed from oldest to newest
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Isolibrium support surface diagnostic test fails or shows incorrect results patient safety
Diagnostic test displays red X on one or more zones (Seat Left/Right, Back Left/Right, Turn Left/Right, Low Air Loss) indicating failure
Ranked causes
- 1. Support surface pressure system fault in specific zone (common)
Review diagnostic test results for failed zones
- Test point
- Diagnostic Test Result screen
- Expected
- Green check marks for all zones; no red X marks
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Serial number not entered or incorrect on bed
Bed serial number is missing, incomplete, or does not match the actual unit serial number.
Ranked causes
- 1. Serial number not saved during setup (occasional)
Access Serial Number screen
- Test point
- Configuration menu
- Expected
- Bed Configuration - Serial Number screen displayed
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Serial number not saved or incorrect in configuration
Bed serial number not recording correctly or reverting after entry or confirmation
Ranked causes
- 1. Serial number not confirmed or power not cycled after entry (common)
Press and hold HOB 30 button and Vascular Position button for 5 seconds
- Test point
- Head end control buttons
- Expected
- Bed Configuration - Serial Number screen displays
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch screen does not respond to input
Touch screen is unresponsive to finger or stylus contact during operation.
Ranked causes
- 1. Touch screen calibration out of adjustment (common)
Verify touch screen responds to stylus or finger at calibration target location
- Test point
- Center of target cross-hair on screen
- Expected
- Target cross-hair displayed and responsive to touch input
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch screen not responding to input
Touch screen does not respond to finger or stylus touches, requiring alternative access method to calibration
Ranked causes
- 1. Touch screen calibration out of alignment (common)
Attempt to press HOB 300+, Brake and Drive buttons simultaneously
- Test point
- Footboard buttons
- Expected
- Touch screen calibration screen appears
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vascular position not achievable caution
Bed fails to reach vascular configuration requiring coordinated movement of multiple sections
Ranked causes
- 1. Fowler, gatch, foot, or trend angle sensor needs recalibration or replacement (common)
Verify fowler, gatch, foot, and trend angle display accuracy on touch screen
- Test point
- touch screen angle displays for all sections
- Expected
- Accurate angle displays on all sections
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wi-Fi configuration not updating
Wireless network settings not saving or wireless feature disabled after configuration changes
Ranked causes
- 1. Wi-Fi settings not saved and power not cycled (common)
Verify network settings entered in Wi-Fi Configuration
- Test point
- Wi-Fi Configuration tab
- Expected
- All network parameters, authentication and security settings correctly entered
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0100 | Isolibrium sensor board communication failure | Sensor board not responding | error |
0101 | Isolibrium sensor board EEPROM corrupted | EEPROM memory corruption detected | critical |
0102 | Isolibrium sensor board not calibrated | Sensor board calibration missing or lost | error |
0200 | Isolibrium GUI CANopen heartbeat lost | Communication heartbeat timeout from GUI | error |
0400 | Isolibrium EEPROM corrupted | System EEPROM memory corruption | critical |
0500 | Isolibrium ambient temperature sensor failure | Temperature sensor not responding | error |
0501 | Isolibrium ambient temperature out of operating range | Temperature reading exceeds limits | warning |
0600 | Isolibrium pressure sensor failure | Pressure sensor not responding | error |
0601 | Isolibrium seat left pressure sensor failure | Left seat pressure sensor not responding | error |
0602 | Isolibrium seat right pressure sensor failure | Right seat pressure sensor not responding | error |
0603 | Isolibrium back left pressure sensor failure | Left back pressure sensor not responding | error |
0604 | Isolibrium back right pressure sensor failure | Right back pressure sensor not responding | error |
0605 | Isolibrium turn left pressure sensor failure | Left turn pressure sensor not responding | error |
0606 | Isolibrium turn right pressure sensor failure | Right turn pressure sensor not responding | error |
0700 | Isolibrium mattress zone overpressure | Mattress pressure exceeds maximum safe level | warning |
0701 | Isolibrium seat left zone overpressure | Left seat zone pressure exceeds maximum | warning |
0702 | Isolibrium seat right zone overpressure | Right seat zone pressure exceeds maximum | warning |
0703 | Isolibrium back left zone overpressure | Left back zone pressure exceeds maximum | warning |
0704 | Isolibrium back right zone overpressure | Right back zone pressure exceeds maximum | warning |
0800 | Isolibrium turn bladder overpressure | Turn bladder pressure exceeds maximum | warning |
0801 | Isolibrium turn left bladder overpressure | Left turn bladder pressure exceeds maximum | warning |
0802 | Isolibrium turn right bladder overpressure | Right turn bladder pressure exceeds maximum | warning |
0900 | Isolibrium compressor current not detected | Compressor not drawing current when on | error |
1000 | Isolibrium fan current not detected | Fan not drawing current when on | error |
1001 | Isolibrium fan 1 current not detected | Fan 1 not drawing current when on | error |
1002 | Isolibrium fan 2 current not detected | Fan 2 not drawing current when on | error |
1100 | Isolibrium valve not detected | Valve solenoid not responding | error |
1101 | Isolibrium seat left inflate valve not detected | Seat left inflate valve not responding | error |
1102 | Isolibrium seat left deflate valve not detected | Seat left deflate valve not responding | error |
1103 | Isolibrium seat right inflate valve not detected | Seat right inflate valve not responding | error |
1104 | Isolibrium seat right deflate valve not detected | Seat right deflate valve not responding | error |
1105 | Isolibrium back left inflate valve not detected | Back left inflate valve not responding | error |
1106 | Isolibrium back left deflate valve not detected | Back left deflate valve not responding | error |
1107 | Isolibrium back right inflate valve not detected | Back right inflate valve not responding | error |
1108 | Isolibrium back right deflate valve not detected | Back right deflate valve not responding | error |
1109 | Isolibrium turn left inflate valve not detected | Turn left inflate valve not responding | error |
1110 | Isolibrium turn left deflate valve not detected | Turn left deflate valve not responding | error |
1111 | Isolibrium turn right inflate valve not detected | Turn right inflate valve not responding | error |
1112 | Isolibrium turn right deflate valve not detected | Turn right deflate valve not responding | error |
1113 | Isolibrium intake valve not detected | Intake valve not responding | error |
1114 | Isolibrium exhaust valve not detected | Exhaust valve not responding | error |
1200 | Isolibrium compressor current when off | Compressor drawing current when system off | error |
1300 | Isolibrium fan current when off | Fan drawing current when system off | error |
1301 | Isolibrium fan 1 current when off | Fan 1 drawing current when system off | error |
1302 | Isolibrium fan 2 current when off | Fan 2 drawing current when system off | error |
1400 | Isolibrium valve current when off | Valve drawing current when system off | error |
1401 | Isolibrium seat left inflate valve current when off | Seat left inflate valve drawing current when off | error |
1402 | Isolibrium seat left deflate valve current when off | Seat left deflate valve drawing current when off | error |
1403 | Isolibrium seat right inflate valve current when off | Seat right inflate valve drawing current when off | error |
1404 | Isolibrium seat right deflate valve current when off | Seat right deflate valve drawing current when off | error |
1405 | Isolibrium back left inflate valve current when off | Back left inflate valve drawing current when off | error |
1406 | Isolibrium back left deflate valve current when off | Back left deflate valve drawing current when off | error |
1407 | Isolibrium back right inflate valve current when off | Back right inflate valve drawing current when off | error |
1408 | Isolibrium back right deflate valve current when off | Back right deflate valve drawing current when off | error |
1409 | Isolibrium turn left inflate valve current when off | Turn left inflate valve drawing current when off | error |
1410 | Isolibrium turn left deflate valve current when off | Turn left deflate valve drawing current when off | error |
1411 | Isolibrium turn right inflate valve current when off | Turn right inflate valve drawing current when off | error |
1412 | Isolibrium turn right deflate valve current when off | Turn right deflate valve drawing current when off | error |
1413 | Isolibrium intake valve current when off | Intake valve drawing current when off | error |
1414 | Isolibrium exhaust valve current when off | Exhaust valve drawing current when off | error |
1500 | Isolibrium mattress leak detected | Pressure loss detected in mattress | warning |
1501 | Isolibrium seat left zone leak detected | Pressure loss detected in left seat zone | warning |
1502 | Isolibrium seat right zone leak detected | Pressure loss detected in right seat zone | warning |
1503 | Isolibrium back left zone leak detected | Pressure loss detected in left back zone | warning |
1504 | Isolibrium back right zone leak detected | Pressure loss detected in right back zone | warning |
1505 | Isolibrium turn left bladder leak detected | Pressure loss detected in left turn bladder | warning |
1506 | Isolibrium turn right bladder leak detected | Pressure loss detected in right turn bladder | warning |
1600 | Isolibrium mattress deflate timeout | Mattress failed to deflate within time limit | error |
1601 | Isolibrium seat left deflate timeout | Left seat zone failed to deflate within time limit | error |
1602 | Isolibrium seat right deflate timeout | Right seat zone failed to deflate within time limit | error |
1603 | Isolibrium back left deflate timeout | Left back zone failed to deflate within time limit | error |
1604 | Isolibrium back right deflate timeout | Right back zone failed to deflate within time limit | error |
1605 | Isolibrium turn left deflate timeout | Left turn bladder failed to deflate within time limit | error |
1606 | Isolibrium turn right deflate timeout | Right turn bladder failed to deflate within time limit | error |
1700 | Isolibrium reference voltage hardware failure | Reference voltage circuit failure detected | critical |
1800 | Isolibrium weight input out of range | Weight sensor reading exceeds limits | warning |
1900 | Isolibrium HOB angle out of range for lateral rotation | Head of bed angle exceeds limits for rotation | error |
1901 | Isolibrium HOB angle out of range for turn assist | Head of bed angle exceeds limits for turn assist | error |
Atd Invalid Values | Digital to analog converter is damaged | Power sensor board detecting invalid ADC readings | error |
Atd Invalid Values | Digital to analog converter is damaged | error | |
Bed Serial Number not available | Bed serial number not configured | Serial number missing from device configuration | warning |
Brake motor time out | Brake motor operation exceeded time limit | Motor running too long or wiring fault | error |
Brake motor time out | Brake motor control timeout | error | |
Brake pot bad range | Potentiometer may be out of place or defective | error | |
Brake pot bad range | Brake potentiometer reading outside acceptable range | Potentiometer moved or disconnected | error |
Brake pot Disconnected or short | Brake potentiometer circuit open or shorted | Potentiometer wiring fault or component failure | error |
Brake pot Disconnected or short | Potentiometer wiring problem or component failure | error | |
Brake unable to elec ctl | Brake motor control electrical issue | error | |
Brake unable to elec ctl | Brake motor cannot be electronically controlled | Wiring or limit switch failure | error |
Calibration Error | Previous calibration procedure failed | Incorrect calibration step execution | error |
Calibration Error | Previous calibration step performed incorrectly | error | |
Cmd WO safe from nurse SRH | Head end board network communication error | error | |
Cmd WO safe from nurse SRH | Head end board network communication error | Control command without safe signal from head end board | warning |
Cmd WO safe from nurse SRL | Left siderail outside board network communication error | Control command without safe signal from left siderail | warning |
Cmd WO safe from nurse SRL | Left siderail outside board network communication error | error | |
Cmd WO safe from nurse SRR | Right siderail outside board network communication error | error | |
Cmd WO safe from nurse SRR | Right siderail outside board network communication error | Control command without safe signal from right siderail | warning |
Cmd WO safe from pat pend H | Head optional pendant control network communication error | Control command without safe signal from pendant | warning |
Cmd WO safe from pat pend H | Optional pendant control head network communication error | error | |
Cmd WO safe from pat pend L | Optional pendant control left network communication error | error | |
Cmd WO safe from pat pend L | Left optional pendant control network communication error | Control command without safe signal from pendant | warning |
Cmd WO safe from pat pend R | Right optional pendant control network communication error | Control command without safe signal from pendant | warning |
Cmd WO safe from pat pend R | Optional pendant control right network communication error | error | |
Cmd WO safe from room | Communication board network communication error | error | |
Cmd WO safe from room | Communication board network communication error | Control command without safe signal from room board | warning |
Cmd WO safe from TS | Touch screen network communication error | Control command without safe signal from touch screen | warning |
Cmd WO safe from TS | Touch screen network communication error | error | |
DHCP Error: Unable to obtain IP Address | Bed unable to obtain IP address from DHCP server | Network connection failure or DHCP misconfiguration | error |
Fixed IP Address duplication | Static IP address conflict with another device on network | IP address already in use on network | warning |
Foot Left Load cell over range | Left foot load cell reading exceeds maximum | Load cell receiving signal out of range | error |
Foot Left Load cell over range | Foot left load cell or cabling damaged | error | |
Foot Right Load cell over range | Foot right load cell or cabling damaged | error | |
Foot Right Load cell over range | Right foot load cell reading exceeds maximum | Load cell receiving signal out of range | error |
GPIO Failure Init | GPIO expansion chip initialization failed | PCA9555 chip not responding during startup | critical |
GPIO Failure Init | PCA9555 GPIO expansion chip initialization failed | error | |
GPIO Failure Read | PCA9555 GPIO expansion chip read failure | error | |
GPIO Failure Read | GPIO expansion chip read operation failed | PCA9555 chip not responding to read command | critical |
GPIO Failure Write | GPIO expansion chip write operation failed | PCA9555 chip not responding to write command | critical |
GPIO Failure Write | PCA9555 GPIO expansion chip write failure | error | |
Head Left Load cell over range | Head left load cell or cabling damaged | error | |
Head Left Load cell over range | Left head load cell reading exceeds maximum | Load cell receiving signal out of range | error |
Head Right cell over range | Right head load cell reading exceeds maximum | Load cell receiving signal out of range | error |
Head Right cell over range | Head right load cell or cabling damaged | error | |
Internal Communication Error | Internal software communication failure | Software malfunction or data corruption | error |
IR Crosstalk detected | Infrared crosstalk detected with iBed Locator system | IR signal interference from other modules | warning |
IR Interface Board not detected | Room interface board not detected by system | Interface board hardware failure or disconnection | error |
Isolibrium active error (unspecified code) | Support surface has detected a fault condition requiring attention | Support surface sensor, pressure system, mechanical or electrical fault | error |
Left IR Module not detected | Left infrared module not detected by system | IR module hardware failure or disconnection | error |
Limit switch foot side rail left | Left foot side rail limit switch failure | Limit switch not responding or wiring fault | error |
Limit switch foot side rail left | Foot side rail left limit switch failure | error | |
Limit switch foot side rail right | Foot side rail right limit switch failure | error | |
Limit switch foot side rail right | Right foot side rail limit switch failure | Limit switch not responding or wiring fault | error |
Limit switch head side rail left | Left head side rail limit switch failure | Limit switch not responding or wiring fault | error |
Limit switch head side rail left | Head side rail left limit switch failure | error | |
Limit switch head side rail right | Head side rail right limit switch failure | error | |
Limit switch head side rail right | Right head side rail limit switch failure | Limit switch not responding or wiring fault | error |
Motor Brake Overheat | Brake motor thermal overload condition | Motor running continuously without adequate cool-down | warning |
Motor Brake Overheat | Brake motor has ran too long | warning | |
Motor Brake Overload | Brake motor drawing too many amps | error | |
Motor Brake Overload | Brake motor current exceeds maximum | Motor drawing excessive current | error |
Motor Foot Overheat | Foot motor thermal overload condition | Motor running continuously without adequate cool-down | warning |
Motor Foot Overheat | Foot motor has ran too long | warning | |
Motor Foot Overload | Foot motor drawing too many amps | error | |
Motor Foot Overload | Foot motor current exceeds maximum | Motor drawing excessive current | error |
Motor Gatch Overheat | Gatch motor thermal overload condition | Motor running continuously without adequate cool-down | warning |
Motor Gatch Overheat | Gatch motor has ran too long | warning | |
Motor Gatch Overload | Gatch motor drawing too many amps | error | |
Motor Gatch Overload | Gatch motor current exceeds maximum | Motor drawing excessive current | error |
Motor Head Overheat | Head (Fowler) motor thermal overload condition | Motor running continuously without adequate cool-down | warning |
Motor Head Overheat | Fowler motor has ran too long | warning | |
Motor Head Overload | Fowler motor drawing too many amps | error | |
Motor Head Overload | Head (Fowler) motor current exceeds maximum | Motor drawing excessive current | error |
Motor HL Foot Overheat | Head-lift foot motor thermal overload condition | Motor running continuously without adequate cool-down | warning |
Motor HL Foot Overheat | Foot lift motor has ran too long | warning | |
Motor HL Foot Overload | Foot lift motor drawing too many amps | error | |
Motor HL Foot Overload | Head-lift foot motor current exceeds maximum | Motor drawing excessive current | error |
Motor HL Head Overheat | Head-lift head motor thermal overload condition | Motor running continuously without adequate cool-down | warning |
Motor HL Head Overheat | Head lift motor has ran too long | warning | |
Motor HL Head Overload | Head lift motor drawing too many amps | error | |
Motor HL Head Overload | Head-lift head motor current exceeds maximum | Motor drawing excessive current | error |
Motor Zoom Overheat | Zoom motor thermal overload condition | Motor running continuously without adequate cool-down | warning |
Motor Zoom Overheat | Zoom motor has ran too long | warning | |
Motor Zoom Overload | Zoom motor drawing too many amps | error | |
Motor Zoom Overload | Zoom motor current exceeds maximum | Motor drawing excessive current | error |
No Error | Angle sensor failure or calibration issue | Angle sensor malfunction or improper calibration | warning |
No Error | Angle sensor failure or calibration issue | error | |
One Motor Drive Short | Short on drive motor detected | error | |
One Motor Drive Short | Short circuit detected on motor drive | Electrical short on drive motor circuit | error |
Right IR Module not detected | Right infrared module not detected by system | IR module hardware failure or disconnection | error |
Safe WO Cmd from A1 | Network A1 safe signal without command wiring issue | Wiring defect on network A1 | warning |
Safe WO Cmd from A1 | Network A1 has a wiring issue | error | |
Safe WO Cmd from A2 | Network A2 has a wiring issue | error | |
Safe WO Cmd from A2 | Network A2 safe signal without command wiring issue | Wiring defect on network A2 | warning |
Safe WO Cmd from B1 | Network B1 safe signal without command wiring issue | Wiring defect on network B1 | warning |
Safe WO Cmd from B1 | Network B1 has a wiring issue | error | |
Safe WO Cmd from B2 | Network B2 has a wiring issue | error | |
Safe WO Cmd from B2 | Network B2 safe signal without command wiring issue | Wiring defect on network B2 | warning |
Scale ADC Error | Scale analog-to-digital converter not calibrated | Scale chip ADC calibration missing | error |
Scale ADC Error | Scale chip not calibrated | error | |
Scale Chip Failure | Control board is bad | error | |
Scale Chip Failure | Scale control chip failure | Scale chip not responding | error |
Tilt base over range | Tilt sensor base reading out of acceptable range | Tilt sensor signal exceeds limits | error |
Tilt base over range | Tilt sensor out of place or improperly assembled | error | |
Tilt Error Base | Base angle sensor is damaged | error | |
Tilt Error Base | Tilt sensor base angle measurement failure | Tilt sensor not responding | error |
Tilt Error Foot | Tilt sensor foot angle measurement failure | Tilt sensor not responding | error |
Tilt Error Foot | Foot angle sensor is damaged | error | |
Tilt Error Gatch | Gatch angle sensor is damaged | error | |
Tilt Error Gatch | Tilt sensor gatch angle measurement failure | Tilt sensor not responding | error |
Tilt Error Head | Tilt sensor head angle measurement failure | Tilt sensor not responding | error |
Tilt Error Head | Head angle sensor is damaged | error | |
Tilt Error HiLo Foot | HiLo foot angle sensor is damaged | error | |
Tilt Error HiLo Foot | Tilt sensor head-lift foot angle measurement failure | Tilt sensor not responding | error |
Tilt Error Trend | Tilt sensor trend angle measurement failure | Tilt sensor not responding | error |
Tilt Error Trend | Trend angle sensor is damaged | error | |
Tilt foot over range | Foot tilt sensor out of place or improperly assembled | error | |
Tilt foot over range | Tilt sensor foot reading out of acceptable range | Tilt sensor signal exceeds limits | error |
Tilt gatch over range | Tilt sensor gatch reading out of acceptable range | Tilt sensor signal exceeds limits | error |
Tilt gatch over range | Gatch tilt sensor out of place or improperly assembled | error | |
Tilt head over range | Head tilt sensor out of place or improperly assembled | error | |
Tilt head over range | Tilt sensor head reading out of acceptable range | Tilt sensor signal exceeds limits | error |
Tilt hiLo foot over range | Tilt sensor head-lift foot reading out of acceptable range | Tilt sensor signal exceeds limits | error |
Tilt hiLo foot over range | HiLo foot tilt sensor out of place or improperly assembled | error | |
Tilt trend over range | Trend tilt sensor out of place or improperly assembled | error | |
Tilt trend over range | Tilt sensor trend reading out of acceptable range | Tilt sensor signal exceeds limits | error |
Unable to connect to network | Bed unable to connect to wireless network | Network connection failure | error |
Wi-Fi Settings corrupted | Wi-Fi configuration data corrupted | Wi-Fi settings data corruption | error |
Wireless Adapter Failure | Wi-Fi radio adapter not detected or not communicating | Wi-Fi adapter hardware failure or disconnection | error |
Zoom time out for switch | Zoom motor switch response timeout | Zoom motor not responding to switch command | error |
Zoom time out for switch | Zoom motor control timeout | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.