Stryker FL27: troubleshooting and error codes

The ServiceForge knowledge base holds 49 documented symptoms for the Stryker FL27, with 89 ranked causes, 134 diagnostic checks with test points and expected values, and 210 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Unit does not power on high voltage

Bed is plugged in but receives no power; wall outlet has correct country voltage

Ranked causes

  1. 1. Wall outlet lacks correct country voltage (occasional)
  2. 2. Main power fuses open (common)
  3. 3. No voltage at transformer connection J11 (occasional)
  4. 4. No 30VAC output from transformer (occasional)
  5. 5. CPU/Power board fuse F1 open or board faulty (rare)
First check

Verify correct voltage at wall outlet

Test point
Wall outlet
Expected
Country voltage option present per CV table

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not power on high voltage

Bed is plugged in but has no power; wall outlet has mains voltage

Ranked causes

  1. 1. Wall outlet has no correct country voltage (common)
  2. 2. Main power fuse open (occasional)
  3. 3. Transformer connection cable issue or 25A fuse failure (occasional)
  4. 4. Transformer assembly failure (rare)
  5. 5. Fuse F1 or CPU/Power board failure (rare)
First check

Verify bed is plugged in and outlet has correct country voltage

Test point
wall outlet
Expected
Country voltage option present at outlet

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cardiac chair positioning fails

Bed fails to achieve cardiac chair position (Fowler and foot positioning required)

Ranked causes

  1. 1. Fowler angle sensor miscalibration (common)
  2. 2. Foot section angle sensor miscalibration (common)
  3. 3. CPU/Power board failure (occasional)
First check

Verify fowler and foot section angles on display

Electric brake does not release caution

Brake motor fails to unlock bed when brake button pressed

Ranked causes

  1. 1. Manual brake set in brake position (common)
  2. 2. Brake motor failure (occasional)
  3. 3. Brake switch malfunction (occasional)
First check

Brake Not Set LED flashing and Brake Set LED OFF

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Foot extension does not lower caution

Foot section fails to move downward when control pressed; debris may block movement

Ranked causes

  1. 1. Debris in foot end traction sockets (common)
  2. 2. Motor stalled or needs recalibration (occasional)
  3. 3. Motor or CPU/Power board failure (occasional)
First check

Inspect foot end traction sockets for debris

Test point
foot end traction sockets
Expected
No debris present

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Foot section does not raise

Foot lift motor fails to extend when up motion commanded in bed calibration mode

Ranked causes

  1. 1. Foot lift motor failure (common)
  2. 2. CPU/Power board failure (occasional)
  3. 3. Litter angle sensor miscalibration (occasional)
First check

Enter Bed Calibration menu via head siderail trend button

4 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Head section does not raise

Head lift motor fails to extend when up motion commanded in calibration mode

Ranked causes

  1. 1. Head lift motor failure (common)
  2. 2. CPU/Power board failure (occasional)
  3. 3. Litter angle sensor miscalibration (occasional)
First check

Enter Bed Calibration menu via head siderail

3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Electric brake does not release caution

Brake pedal engaged but electric brake motor does not disengage when brake button pressed

Ranked causes

  1. 1. Brake pedal in brake position (correct operation) (common)
  2. 2. Brake motor or CPU/Power board failure (occasional)
First check

Verify Brake Not Set LED is flashing and Brake Set LED is OFF

Test point
bed control panel LEDs
Expected
Brake Not Set LED flashing; Brake Set LED OFF

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Foot extension does not raise caution

Foot section fails to move upward when control pressed

Ranked causes

  1. 1. Motor stalled or needs recalibration (common)
  2. 2. Motor or CPU/Power board failure (occasional)
First check

Enter Bed Calibration menu and press foot up button; observe motor

Test point
foot extension motor
Expected
Motor runs upward

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Foot lift section does not lower caution

Foot section fails to move downward when control pressed

Ranked causes

  1. 1. Motor stalled or needs recalibration (common)
  2. 2. Motor or CPU/Power board failure (occasional)
First check

Enter Bed Calibration menu and press reverse trend button; observe motor

Test point
foot lift motor
Expected
Motor runs downward

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Foot lift section does not raise caution

Foot section fails to move upward when control pressed; may occur after other motion failures

Ranked causes

  1. 1. Motor stalled or needs recalibration (common)
  2. 2. Motor or CPU/Power board failure (occasional)
First check

Enter Bed Calibration menu and press trend button; observe motor

Test point
foot lift motor
Expected
Motor runs upward

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Foot positioning does not lower

Foot positioning motor fails to retract when down motion commanded

Ranked causes

  1. 1. Foot motor failure (common)
  2. 2. CPU/Power board failure (occasional)
First check

24VDC at J1 (black pin 1, red pin 2) when down pressed

Foot positioning does not raise

Foot positioning motor fails to extend when up motion commanded

Ranked causes

  1. 1. Foot motor failure (common)
  2. 2. CPU/Power board failure (occasional)
First check

Motor runs in calibration mode

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Foot section does not lower

Foot lift motor fails to retract when down motion commanded

Ranked causes

  1. 1. Foot lift motor failure (common)
  2. 2. CPU/Power board failure (occasional)
First check

24VDC at J6 (red pin 1, black pin 2) when reverse trend button pressed

Fowler section does not lower caution

Fowler section fails to move downward when control pressed

Ranked causes

  1. 1. Motor stalled or needs recalibration (common)
  2. 2. Motor or CPU/Power board failure (occasional)
First check

Enter Bed Calibration menu and press fowler down button; observe motor

Test point
fowler motor
Expected
Motor runs downward

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fowler section does not lower

Fowler motor fails to retract when down motion commanded

Ranked causes

  1. 1. Fowler motor failure (common)
  2. 2. CPU/Power board failure (occasional)
First check

24VDC at J5 (red pin 1, black pin 2) when down button pressed

Fowler section does not raise caution

Fowler section fails to move upward when control pressed

Ranked causes

  1. 1. Motor stalled or needs recalibration (common)
  2. 2. Motor or CPU/Power board failure (occasional)
First check

Enter Bed Calibration menu and press fowler up button; observe motor

Test point
fowler motor
Expected
Motor runs upward

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fowler section does not raise

Fowler motor fails to extend when up motion commanded

Ranked causes

  1. 1. Fowler motor failure (common)
  2. 2. CPU/Power board failure (occasional)
First check

Motor runs in calibration mode

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Gatch section does not lower

Gatch motor fails to retract when down motion commanded

Ranked causes

  1. 1. Gatch motor failure (common)
  2. 2. CPU/Power board failure (occasional)
First check

24VDC at J3 (black pin 1, red pin 2) when down pressed

Gatch section does not lower caution

Gatch section fails to move downward when control pressed

Ranked causes

  1. 1. Motor stalled or needs recalibration (common)
  2. 2. Motor or CPU/Power board failure (occasional)
First check

Enter Bed Calibration menu and press gatch down button; observe motor

Test point
gatch motor
Expected
Motor runs downward

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gatch section does not raise

Gatch motor fails to extend when up motion commanded

Ranked causes

  1. 1. Gatch motor failure (common)
  2. 2. CPU/Power board failure (occasional)
First check

Motor runs in calibration mode

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Gatch section does not raise caution

Gatch section fails to move upward when control pressed

Ranked causes

  1. 1. Motor stalled or needs recalibration (common)
  2. 2. Motor or CPU/Power board failure (occasional)
First check

Enter Bed Calibration menu and press gatch up button; observe motor

Test point
gatch motor
Expected
Motor runs upward

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Head lift section does not lower caution

Head section fails to move downward when control pressed

Ranked causes

  1. 1. Motor stalled or needs recalibration (common)
  2. 2. Motor or CPU/Power board failure (occasional)
First check

Enter Bed Calibration menu and press bed down button; observe motor

Test point
head lift motor
Expected
Motor runs downward

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Head lift section does not raise caution

Head section fails to move upward when control pressed

Ranked causes

  1. 1. Motor stalled or needs recalibration (common)
  2. 2. Motor or CPU/Power board failure (occasional)
First check

Enter Bed Calibration menu and press bed up button; observe motor

Test point
head lift motor
Expected
Motor runs upward

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Head section does not lower

Head lift motor fails to retract when down motion commanded

Ranked causes

  1. 1. Head lift motor failure (common)
  2. 2. CPU/Power board failure (occasional)
First check

24VDC at J4 (red pin 1, black pin 2) when bed down button pressed

HOB 300 position not achieved

Head of bed angle does not position correctly to 30 degrees or display inaccurate

Ranked causes

  1. 1. Fowler angle sensor miscalibration (common)
  2. 2. CPU/Power board failure (occasional)
First check

Verify fowler angle display accuracy

Reverse Trendelenburg positioning fails

Bed fails to tilt into Reverse Trendelenburg position or angle sensor inaccurate

Ranked causes

  1. 1. Trend angle sensor miscalibration (common)
  2. 2. CPU/Power board failure (occasional)
First check

Verify touch screen angle display accuracy

Trendelenburg positioning fails

Bed fails to tilt into Trendelenburg position or angle sensor inaccurate

Ranked causes

  1. 1. Trend angle sensor miscalibration (common)
  2. 2. CPU/Power board failure (occasional)
First check

Check touch screen trend angle display accuracy

Vascular positioning fails

Bed fails to achieve vascular position or multiple angle sensors inaccurate

Ranked causes

  1. 1. Multiple angle sensors miscalibrated (common)
  2. 2. CPU/Power board failure (occasional)
First check

Verify fowler, gatch, foot, trend display accuracy

Bed display shows non-zero angle readings at 16 inch height

Display readings fluctuate or do not maintain zero degrees when bed is at lowest position

Ranked causes

  1. 1. Incomplete or failed initial calibration requiring full recalibration cycle (common)
First check

Return to Configuration Screen

Test point
Footboard control panel Main Menu button
Expected
Configuration Screen displayed with Bed Calibration option

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed does not raise in Reverse Trendelenburg position caution

Bed fails to raise in Reverse Trendelenburg (head-up) position when control pressed

Ranked causes

  1. 1. Bed needs recalibration or trend angle sensor failure (common)
First check

Verify trend angle display accuracy on touch screen

Test point
touch screen trend angle display
Expected
Accurate display of litter level

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed does not raise in Trendelenburg position caution

Bed fails to raise in Trendelenburg (head-down) position when control pressed

Ranked causes

  1. 1. Bed needs recalibration or trend angle sensor failure (common)
First check

Verify trend angle display accuracy on touch screen

Test point
touch screen trend angle display
Expected
Accurate display of litter level

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed Options configuration not applied

Selected bed options such as iBed, iAudio, iDoc are not properly configured or saved on the bed system.

Ranked causes

  1. 1. Bed Options not confirmed or saved (common)
First check

Verify Bed Options screen displays correct options

Test point
Bed Options screen
Expected
All applicable bed options visible and selectable

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed options not saved or incorrect after configuration

Selected bed options (iBed, iAudio, iDoc, Wi-Fi) not matching actual configuration or not persisting after setup

Ranked causes

  1. 1. Bed options not confirmed or power not cycled after selection (common)
First check

Press and hold HOB 30 button and BRAKE button simultaneously for 5 seconds

Test point
Head end control buttons
Expected
Bed Options screen displays with list of available options

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Bed position and level sensors patient safety

Six-step bed calibration procedure to establish reference points for height, tilt, and levelness across all bed articulation positions

Ranked causes

  1. 1. Initial setup or sensor drift requiring recalibration (common)
First check

Flatten all bed articulations to reference position

Test point
Litter surface position
Expected
Completely flat

10 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Touch screen

Validation and adjustment of touch screen coordinate calibration to ensure accurate input detection

Ranked causes

  1. 1. Touch screen calibration drift (occasional)
First check

Press and hold stylus or finger on center of cross hair

Test point
Center of screen
Expected
Cross hair registers touch and moves to next location

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Touch screen caution

Procedure to recalibrate touch screen coordinates to ensure accurate touch recognition at all screen locations.

Ranked causes

  1. 1. Touch screen calibration initialization (unknown)
First check

Press and hold on upper center target

Test point
Upper center of screen
Expected
Target cross-hair responds to touch

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cardiac chair position not achievable caution

Bed fails to reach cardiac chair configuration where fowler and foot sections coordinate

Ranked causes

  1. 1. Fowler or foot angle sensor needs recalibration or replacement (common)
First check

Verify fowler and foot section angle display accuracy on touch screen

Test point
touch screen fowler and foot angle displays
Expected
Accurate angle displays for both sections

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Full diagnostic menu not accessible or shows incomplete information

Full Diagnostic button from Configuration Screen does not open menu or displays incomplete data in Boards, Error Codes, Input States, Motors, Buttons Pressed, or Signal Values screens

Ranked causes

  1. 1. CPU board or touch screen communication fault (occasional)
First check

Press Full Diagnostic button from Configuration Screen

Test point
Configuration Screen
Expected
Full Diagnostic screen with menu items appears (Boards, Error Codes, Input States, Motors, Buttons Pressed, Signal Values)

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Head of bed position angle inaccurate caution

Bed fails to achieve or maintain proper head of bed angle position

Ranked causes

  1. 1. Fowler angle sensor needs recalibration or replacement (common)
First check

Verify fowler angle display accuracy on touch screen

Test point
touch screen fowler angle display
Expected
Accurate angle display

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Isolibrium active error displayed on support surface patient safety

Support surface showing active error code in Isolibrium diagnostic menu; up to five errors displayed simultaneously ordered by priority

Ranked causes

  1. 1. Support surface sensor, pressure or mechanical fault (common)
First check

Access Isolibrium diagnostic menu and tap Active Errors tab

Test point
Service Diagnostic screen
Expected
List of up to five active errors displayed in priority order

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Isolibrium error history not clearing or showing incorrect past errors

Error History tab shows old errors not being removed, or errors remaining after Clear Error History button pressed, or errors appearing that should not be logged

Ranked causes

  1. 1. Active error preventing history clear or active error re-logged on clear (occasional)
First check

Access Isolibrium diagnostic menu and tap Error History tab

Test point
Service Diagnostic screen
Expected
List of up to 10 errors displayed from oldest to newest

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Isolibrium support surface diagnostic test fails or shows incorrect results patient safety

Diagnostic test displays red X on one or more zones (Seat Left/Right, Back Left/Right, Turn Left/Right, Low Air Loss) indicating failure

Ranked causes

  1. 1. Support surface pressure system fault in specific zone (common)
First check

Review diagnostic test results for failed zones

Test point
Diagnostic Test Result screen
Expected
Green check marks for all zones; no red X marks

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Serial number not entered or incorrect on bed

Bed serial number is missing, incomplete, or does not match the actual unit serial number.

Ranked causes

  1. 1. Serial number not saved during setup (occasional)
First check

Access Serial Number screen

Test point
Configuration menu
Expected
Bed Configuration - Serial Number screen displayed

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Serial number not saved or incorrect in configuration

Bed serial number not recording correctly or reverting after entry or confirmation

Ranked causes

  1. 1. Serial number not confirmed or power not cycled after entry (common)
First check

Press and hold HOB 30 button and Vascular Position button for 5 seconds

Test point
Head end control buttons
Expected
Bed Configuration - Serial Number screen displays

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touch screen does not respond to input

Touch screen is unresponsive to finger or stylus contact during operation.

Ranked causes

  1. 1. Touch screen calibration out of adjustment (common)
First check

Verify touch screen responds to stylus or finger at calibration target location

Test point
Center of target cross-hair on screen
Expected
Target cross-hair displayed and responsive to touch input

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touch screen not responding to input

Touch screen does not respond to finger or stylus touches, requiring alternative access method to calibration

Ranked causes

  1. 1. Touch screen calibration out of alignment (common)
First check

Attempt to press HOB 300+, Brake and Drive buttons simultaneously

Test point
Footboard buttons
Expected
Touch screen calibration screen appears

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Vascular position not achievable caution

Bed fails to reach vascular configuration requiring coordinated movement of multiple sections

Ranked causes

  1. 1. Fowler, gatch, foot, or trend angle sensor needs recalibration or replacement (common)
First check

Verify fowler, gatch, foot, and trend angle display accuracy on touch screen

Test point
touch screen angle displays for all sections
Expected
Accurate angle displays on all sections

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wi-Fi configuration not updating

Wireless network settings not saving or wireless feature disabled after configuration changes

Ranked causes

  1. 1. Wi-Fi settings not saved and power not cycled (common)
First check

Verify network settings entered in Wi-Fi Configuration

Test point
Wi-Fi Configuration tab
Expected
All network parameters, authentication and security settings correctly entered

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
0100Isolibrium sensor board communication failureSensor board not respondingerror
0101Isolibrium sensor board EEPROM corruptedEEPROM memory corruption detectedcritical
0102Isolibrium sensor board not calibratedSensor board calibration missing or losterror
0200Isolibrium GUI CANopen heartbeat lostCommunication heartbeat timeout from GUIerror
0400Isolibrium EEPROM corruptedSystem EEPROM memory corruptioncritical
0500Isolibrium ambient temperature sensor failureTemperature sensor not respondingerror
0501Isolibrium ambient temperature out of operating rangeTemperature reading exceeds limitswarning
0600Isolibrium pressure sensor failurePressure sensor not respondingerror
0601Isolibrium seat left pressure sensor failureLeft seat pressure sensor not respondingerror
0602Isolibrium seat right pressure sensor failureRight seat pressure sensor not respondingerror
0603Isolibrium back left pressure sensor failureLeft back pressure sensor not respondingerror
0604Isolibrium back right pressure sensor failureRight back pressure sensor not respondingerror
0605Isolibrium turn left pressure sensor failureLeft turn pressure sensor not respondingerror
0606Isolibrium turn right pressure sensor failureRight turn pressure sensor not respondingerror
0700Isolibrium mattress zone overpressureMattress pressure exceeds maximum safe levelwarning
0701Isolibrium seat left zone overpressureLeft seat zone pressure exceeds maximumwarning
0702Isolibrium seat right zone overpressureRight seat zone pressure exceeds maximumwarning
0703Isolibrium back left zone overpressureLeft back zone pressure exceeds maximumwarning
0704Isolibrium back right zone overpressureRight back zone pressure exceeds maximumwarning
0800Isolibrium turn bladder overpressureTurn bladder pressure exceeds maximumwarning
0801Isolibrium turn left bladder overpressureLeft turn bladder pressure exceeds maximumwarning
0802Isolibrium turn right bladder overpressureRight turn bladder pressure exceeds maximumwarning
0900Isolibrium compressor current not detectedCompressor not drawing current when onerror
1000Isolibrium fan current not detectedFan not drawing current when onerror
1001Isolibrium fan 1 current not detectedFan 1 not drawing current when onerror
1002Isolibrium fan 2 current not detectedFan 2 not drawing current when onerror
1100Isolibrium valve not detectedValve solenoid not respondingerror
1101Isolibrium seat left inflate valve not detectedSeat left inflate valve not respondingerror
1102Isolibrium seat left deflate valve not detectedSeat left deflate valve not respondingerror
1103Isolibrium seat right inflate valve not detectedSeat right inflate valve not respondingerror
1104Isolibrium seat right deflate valve not detectedSeat right deflate valve not respondingerror
1105Isolibrium back left inflate valve not detectedBack left inflate valve not respondingerror
1106Isolibrium back left deflate valve not detectedBack left deflate valve not respondingerror
1107Isolibrium back right inflate valve not detectedBack right inflate valve not respondingerror
1108Isolibrium back right deflate valve not detectedBack right deflate valve not respondingerror
1109Isolibrium turn left inflate valve not detectedTurn left inflate valve not respondingerror
1110Isolibrium turn left deflate valve not detectedTurn left deflate valve not respondingerror
1111Isolibrium turn right inflate valve not detectedTurn right inflate valve not respondingerror
1112Isolibrium turn right deflate valve not detectedTurn right deflate valve not respondingerror
1113Isolibrium intake valve not detectedIntake valve not respondingerror
1114Isolibrium exhaust valve not detectedExhaust valve not respondingerror
1200Isolibrium compressor current when offCompressor drawing current when system offerror
1300Isolibrium fan current when offFan drawing current when system offerror
1301Isolibrium fan 1 current when offFan 1 drawing current when system offerror
1302Isolibrium fan 2 current when offFan 2 drawing current when system offerror
1400Isolibrium valve current when offValve drawing current when system offerror
1401Isolibrium seat left inflate valve current when offSeat left inflate valve drawing current when offerror
1402Isolibrium seat left deflate valve current when offSeat left deflate valve drawing current when offerror
1403Isolibrium seat right inflate valve current when offSeat right inflate valve drawing current when offerror
1404Isolibrium seat right deflate valve current when offSeat right deflate valve drawing current when offerror
1405Isolibrium back left inflate valve current when offBack left inflate valve drawing current when offerror
1406Isolibrium back left deflate valve current when offBack left deflate valve drawing current when offerror
1407Isolibrium back right inflate valve current when offBack right inflate valve drawing current when offerror
1408Isolibrium back right deflate valve current when offBack right deflate valve drawing current when offerror
1409Isolibrium turn left inflate valve current when offTurn left inflate valve drawing current when offerror
1410Isolibrium turn left deflate valve current when offTurn left deflate valve drawing current when offerror
1411Isolibrium turn right inflate valve current when offTurn right inflate valve drawing current when offerror
1412Isolibrium turn right deflate valve current when offTurn right deflate valve drawing current when offerror
1413Isolibrium intake valve current when offIntake valve drawing current when offerror
1414Isolibrium exhaust valve current when offExhaust valve drawing current when offerror
1500Isolibrium mattress leak detectedPressure loss detected in mattresswarning
1501Isolibrium seat left zone leak detectedPressure loss detected in left seat zonewarning
1502Isolibrium seat right zone leak detectedPressure loss detected in right seat zonewarning
1503Isolibrium back left zone leak detectedPressure loss detected in left back zonewarning
1504Isolibrium back right zone leak detectedPressure loss detected in right back zonewarning
1505Isolibrium turn left bladder leak detectedPressure loss detected in left turn bladderwarning
1506Isolibrium turn right bladder leak detectedPressure loss detected in right turn bladderwarning
1600Isolibrium mattress deflate timeoutMattress failed to deflate within time limiterror
1601Isolibrium seat left deflate timeoutLeft seat zone failed to deflate within time limiterror
1602Isolibrium seat right deflate timeoutRight seat zone failed to deflate within time limiterror
1603Isolibrium back left deflate timeoutLeft back zone failed to deflate within time limiterror
1604Isolibrium back right deflate timeoutRight back zone failed to deflate within time limiterror
1605Isolibrium turn left deflate timeoutLeft turn bladder failed to deflate within time limiterror
1606Isolibrium turn right deflate timeoutRight turn bladder failed to deflate within time limiterror
1700Isolibrium reference voltage hardware failureReference voltage circuit failure detectedcritical
1800Isolibrium weight input out of rangeWeight sensor reading exceeds limitswarning
1900Isolibrium HOB angle out of range for lateral rotationHead of bed angle exceeds limits for rotationerror
1901Isolibrium HOB angle out of range for turn assistHead of bed angle exceeds limits for turn assisterror
Atd Invalid ValuesDigital to analog converter is damagedPower sensor board detecting invalid ADC readingserror
Atd Invalid ValuesDigital to analog converter is damagederror
Bed Serial Number not availableBed serial number not configuredSerial number missing from device configurationwarning
Brake motor time outBrake motor operation exceeded time limitMotor running too long or wiring faulterror
Brake motor time outBrake motor control timeouterror
Brake pot bad rangePotentiometer may be out of place or defectiveerror
Brake pot bad rangeBrake potentiometer reading outside acceptable rangePotentiometer moved or disconnectederror
Brake pot Disconnected or shortBrake potentiometer circuit open or shortedPotentiometer wiring fault or component failureerror
Brake pot Disconnected or shortPotentiometer wiring problem or component failureerror
Brake unable to elec ctlBrake motor control electrical issueerror
Brake unable to elec ctlBrake motor cannot be electronically controlledWiring or limit switch failureerror
Calibration ErrorPrevious calibration procedure failedIncorrect calibration step executionerror
Calibration ErrorPrevious calibration step performed incorrectlyerror
Cmd WO safe from nurse SRHHead end board network communication errorerror
Cmd WO safe from nurse SRHHead end board network communication errorControl command without safe signal from head end boardwarning
Cmd WO safe from nurse SRLLeft siderail outside board network communication errorControl command without safe signal from left siderailwarning
Cmd WO safe from nurse SRLLeft siderail outside board network communication errorerror
Cmd WO safe from nurse SRRRight siderail outside board network communication errorerror
Cmd WO safe from nurse SRRRight siderail outside board network communication errorControl command without safe signal from right siderailwarning
Cmd WO safe from pat pend HHead optional pendant control network communication errorControl command without safe signal from pendantwarning
Cmd WO safe from pat pend HOptional pendant control head network communication errorerror
Cmd WO safe from pat pend LOptional pendant control left network communication errorerror
Cmd WO safe from pat pend LLeft optional pendant control network communication errorControl command without safe signal from pendantwarning
Cmd WO safe from pat pend RRight optional pendant control network communication errorControl command without safe signal from pendantwarning
Cmd WO safe from pat pend ROptional pendant control right network communication errorerror
Cmd WO safe from roomCommunication board network communication errorerror
Cmd WO safe from roomCommunication board network communication errorControl command without safe signal from room boardwarning
Cmd WO safe from TSTouch screen network communication errorControl command without safe signal from touch screenwarning
Cmd WO safe from TSTouch screen network communication errorerror
DHCP Error: Unable to obtain IP AddressBed unable to obtain IP address from DHCP serverNetwork connection failure or DHCP misconfigurationerror
Fixed IP Address duplicationStatic IP address conflict with another device on networkIP address already in use on networkwarning
Foot Left Load cell over rangeLeft foot load cell reading exceeds maximumLoad cell receiving signal out of rangeerror
Foot Left Load cell over rangeFoot left load cell or cabling damagederror
Foot Right Load cell over rangeFoot right load cell or cabling damagederror
Foot Right Load cell over rangeRight foot load cell reading exceeds maximumLoad cell receiving signal out of rangeerror
GPIO Failure InitGPIO expansion chip initialization failedPCA9555 chip not responding during startupcritical
GPIO Failure InitPCA9555 GPIO expansion chip initialization failederror
GPIO Failure ReadPCA9555 GPIO expansion chip read failureerror
GPIO Failure ReadGPIO expansion chip read operation failedPCA9555 chip not responding to read commandcritical
GPIO Failure WriteGPIO expansion chip write operation failedPCA9555 chip not responding to write commandcritical
GPIO Failure WritePCA9555 GPIO expansion chip write failureerror
Head Left Load cell over rangeHead left load cell or cabling damagederror
Head Left Load cell over rangeLeft head load cell reading exceeds maximumLoad cell receiving signal out of rangeerror
Head Right cell over rangeRight head load cell reading exceeds maximumLoad cell receiving signal out of rangeerror
Head Right cell over rangeHead right load cell or cabling damagederror
Internal Communication ErrorInternal software communication failureSoftware malfunction or data corruptionerror
IR Crosstalk detectedInfrared crosstalk detected with iBed Locator systemIR signal interference from other moduleswarning
IR Interface Board not detectedRoom interface board not detected by systemInterface board hardware failure or disconnectionerror
Isolibrium active error (unspecified code)Support surface has detected a fault condition requiring attentionSupport surface sensor, pressure system, mechanical or electrical faulterror
Left IR Module not detectedLeft infrared module not detected by systemIR module hardware failure or disconnectionerror
Limit switch foot side rail leftLeft foot side rail limit switch failureLimit switch not responding or wiring faulterror
Limit switch foot side rail leftFoot side rail left limit switch failureerror
Limit switch foot side rail rightFoot side rail right limit switch failureerror
Limit switch foot side rail rightRight foot side rail limit switch failureLimit switch not responding or wiring faulterror
Limit switch head side rail leftLeft head side rail limit switch failureLimit switch not responding or wiring faulterror
Limit switch head side rail leftHead side rail left limit switch failureerror
Limit switch head side rail rightHead side rail right limit switch failureerror
Limit switch head side rail rightRight head side rail limit switch failureLimit switch not responding or wiring faulterror
Motor Brake OverheatBrake motor thermal overload conditionMotor running continuously without adequate cool-downwarning
Motor Brake OverheatBrake motor has ran too longwarning
Motor Brake OverloadBrake motor drawing too many ampserror
Motor Brake OverloadBrake motor current exceeds maximumMotor drawing excessive currenterror
Motor Foot OverheatFoot motor thermal overload conditionMotor running continuously without adequate cool-downwarning
Motor Foot OverheatFoot motor has ran too longwarning
Motor Foot OverloadFoot motor drawing too many ampserror
Motor Foot OverloadFoot motor current exceeds maximumMotor drawing excessive currenterror
Motor Gatch OverheatGatch motor thermal overload conditionMotor running continuously without adequate cool-downwarning
Motor Gatch OverheatGatch motor has ran too longwarning
Motor Gatch OverloadGatch motor drawing too many ampserror
Motor Gatch OverloadGatch motor current exceeds maximumMotor drawing excessive currenterror
Motor Head OverheatHead (Fowler) motor thermal overload conditionMotor running continuously without adequate cool-downwarning
Motor Head OverheatFowler motor has ran too longwarning
Motor Head OverloadFowler motor drawing too many ampserror
Motor Head OverloadHead (Fowler) motor current exceeds maximumMotor drawing excessive currenterror
Motor HL Foot OverheatHead-lift foot motor thermal overload conditionMotor running continuously without adequate cool-downwarning
Motor HL Foot OverheatFoot lift motor has ran too longwarning
Motor HL Foot OverloadFoot lift motor drawing too many ampserror
Motor HL Foot OverloadHead-lift foot motor current exceeds maximumMotor drawing excessive currenterror
Motor HL Head OverheatHead-lift head motor thermal overload conditionMotor running continuously without adequate cool-downwarning
Motor HL Head OverheatHead lift motor has ran too longwarning
Motor HL Head OverloadHead lift motor drawing too many ampserror
Motor HL Head OverloadHead-lift head motor current exceeds maximumMotor drawing excessive currenterror
Motor Zoom OverheatZoom motor thermal overload conditionMotor running continuously without adequate cool-downwarning
Motor Zoom OverheatZoom motor has ran too longwarning
Motor Zoom OverloadZoom motor drawing too many ampserror
Motor Zoom OverloadZoom motor current exceeds maximumMotor drawing excessive currenterror
No ErrorAngle sensor failure or calibration issueAngle sensor malfunction or improper calibrationwarning
No ErrorAngle sensor failure or calibration issueerror
One Motor Drive ShortShort on drive motor detectederror
One Motor Drive ShortShort circuit detected on motor driveElectrical short on drive motor circuiterror
Right IR Module not detectedRight infrared module not detected by systemIR module hardware failure or disconnectionerror
Safe WO Cmd from A1Network A1 safe signal without command wiring issueWiring defect on network A1warning
Safe WO Cmd from A1Network A1 has a wiring issueerror
Safe WO Cmd from A2Network A2 has a wiring issueerror
Safe WO Cmd from A2Network A2 safe signal without command wiring issueWiring defect on network A2warning
Safe WO Cmd from B1Network B1 safe signal without command wiring issueWiring defect on network B1warning
Safe WO Cmd from B1Network B1 has a wiring issueerror
Safe WO Cmd from B2Network B2 has a wiring issueerror
Safe WO Cmd from B2Network B2 safe signal without command wiring issueWiring defect on network B2warning
Scale ADC ErrorScale analog-to-digital converter not calibratedScale chip ADC calibration missingerror
Scale ADC ErrorScale chip not calibratederror
Scale Chip FailureControl board is baderror
Scale Chip FailureScale control chip failureScale chip not respondingerror
Tilt base over rangeTilt sensor base reading out of acceptable rangeTilt sensor signal exceeds limitserror
Tilt base over rangeTilt sensor out of place or improperly assemblederror
Tilt Error BaseBase angle sensor is damagederror
Tilt Error BaseTilt sensor base angle measurement failureTilt sensor not respondingerror
Tilt Error FootTilt sensor foot angle measurement failureTilt sensor not respondingerror
Tilt Error FootFoot angle sensor is damagederror
Tilt Error GatchGatch angle sensor is damagederror
Tilt Error GatchTilt sensor gatch angle measurement failureTilt sensor not respondingerror
Tilt Error HeadTilt sensor head angle measurement failureTilt sensor not respondingerror
Tilt Error HeadHead angle sensor is damagederror
Tilt Error HiLo FootHiLo foot angle sensor is damagederror
Tilt Error HiLo FootTilt sensor head-lift foot angle measurement failureTilt sensor not respondingerror
Tilt Error TrendTilt sensor trend angle measurement failureTilt sensor not respondingerror
Tilt Error TrendTrend angle sensor is damagederror
Tilt foot over rangeFoot tilt sensor out of place or improperly assemblederror
Tilt foot over rangeTilt sensor foot reading out of acceptable rangeTilt sensor signal exceeds limitserror
Tilt gatch over rangeTilt sensor gatch reading out of acceptable rangeTilt sensor signal exceeds limitserror
Tilt gatch over rangeGatch tilt sensor out of place or improperly assemblederror
Tilt head over rangeHead tilt sensor out of place or improperly assemblederror
Tilt head over rangeTilt sensor head reading out of acceptable rangeTilt sensor signal exceeds limitserror
Tilt hiLo foot over rangeTilt sensor head-lift foot reading out of acceptable rangeTilt sensor signal exceeds limitserror
Tilt hiLo foot over rangeHiLo foot tilt sensor out of place or improperly assemblederror
Tilt trend over rangeTrend tilt sensor out of place or improperly assemblederror
Tilt trend over rangeTilt sensor trend reading out of acceptable rangeTilt sensor signal exceeds limitserror
Unable to connect to networkBed unable to connect to wireless networkNetwork connection failureerror
Wi-Fi Settings corruptedWi-Fi configuration data corruptedWi-Fi settings data corruptionerror
Wireless Adapter FailureWi-Fi radio adapter not detected or not communicatingWi-Fi adapter hardware failure or disconnectionerror
Zoom time out for switchZoom motor switch response timeoutZoom motor not responding to switch commanderror
Zoom time out for switchZoom motor control timeouterror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI