Hospital beds

Stryker Bed Intouch Critical Care: troubleshooting and error codes

The ServiceForge knowledge base holds 131 documented symptoms for the Stryker Bed Intouch Critical Care, with 376 ranked causes, 231 diagnostic checks with test points and expected values, and 71 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Support surface has no power caution

Mattress/support surface unit is completely unpowered; no LED lights or alarm sounds when powered on

Ranked causes

  1. 1. Power cord disconnected or improperly seated (common)
  2. 2. Foot box connector disconnected from controller (common)
  3. 3. Fuses blown above foot box power cord connection (occasional)
  4. 4. Main fuse on power board open (occasional)
  5. 5. No 120 VAC at power inlet (occasional)
  6. 6. Power inlet filter failed (occasional)
  7. 7. Transformer failed (rare)
  8. 8. Power board failed (rare)
First check

Power cord connection

Test point
Power cord to wall outlet
Expected
Firmly seated connection

8 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Foot box overheated high voltage

Foot box temperature exceeds 140°F (60°C); unit auto-shuts down with message 'Auto. Overload Caused a Power Shutdown.'

Ranked causes

  1. 1. Kinked or restricted supply tubing (common)
  2. 2. Kinked low air loss tubing (common)
  3. 3. Plugged bulkhead fitting at foot box (occasional)
  4. 4. Kinked tubing from compressor to bulkhead (occasional)
  5. 5. Supply tubing too close or touching sensor U5 (occasional)
  6. 6. Blue cable white/green or orange/white wire intermittence (occasional)
  7. 7. Faulty transformer with high no-load current (rare)
First check

Allow unit to cool; reset service log and cycle power

Test point
Foot box temperature
Expected
Below 140 degrees F (60 degrees C)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Foot end controller overtemperature shutdown caution

Foot end control unit temperature exceeds 60 degrees Celsius; system displays auto-shutdown message and power is disabled.

Ranked causes

  1. 1. Kink or blockage in air supply hose to support surface (common)
  2. 2. Kink in low-pressure relief hose (occasional)
  3. 3. Blockage in guard fitting on foot end controller (occasional)
  4. 4. Kink in compressor discharge hose (occasional)
  5. 5. Supply hose too close to or contacting temperature sensor U5 (occasional)
  6. 6. Transformer no-load current exceeds specification (occasional)
  7. 7. Temperature sensor white/green or orange/white wire integrity issue (occasional)
First check

Allow unit to cool and reset service alarm

Test point
Foot end controller
Expected
Temperature drops below 60°C

7 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Foot-end box overheating caution

Excessive temperature in foot-end control box above 60°C; system displays auto-shutdown message and halts operation

Ranked causes

  1. 1. Blue cable white/green and orange/white wire short or fault (occasional)
  2. 2. Supply tubing kinked or twisted (common)
  3. 3. Low-loss air tubing kinked or twisted (occasional)
  4. 4. Bulkhead fitting blocked (occasional)
  5. 5. Compressor-to-fitting tubing kinked or twisted (occasional)
  6. 6. Supply tubing too close to or touching temperature sensor U5 (occasional)
  7. 7. Transformer no-load current excessive (rare)
First check

Allow system to cool and reset

Test point
Foot-end box temperature
Expected
Temperature returns below 60°C

7 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Foot-of-bed control box overtemperature caution

Temperature in foot-of-bed control box exceeds 60 °C; displays auto overload power shutdown message

Ranked causes

  1. 1. Blue cable white-green and orange-white conductors short or high resistance (occasional)
  2. 2. Kinked or blocked supply hose in mattress (occasional)
  3. 3. Kinked return hose causing low air loss (occasional)
  4. 4. Wall fitting at foot-of-bed control box blocked or restricted (occasional)
  5. 5. Kinked compressor-to-wall fitting hose (occasional)
  6. 6. Supply hose touching sensor U5 on power board (occasional)
  7. 7. Transformer no-load current exceeds 50 mA (occasional)
First check

Allow product to cool

Test point
Foot-of-bed control box
Expected
Temperature drop below 60 °C

7 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cell in diagnostic test does not inflate to 50 mmHg in 3 minutes

Production test code during diagnostic test: cell inflation fails to reach 50 mmHg pressure within 3 minutes

Ranked causes

  1. 1. Hose leaks at connections (common)
  2. 2. Leaks in cells (common)
  3. 3. CPR latch or O-ring defective (occasional)
  4. 4. Low compressor output during initial inflation (common)
  5. 5. Kinked supply hose in mattress deck (common)
  6. 6. Wall connector at foot-end control box clogged (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cell pressurization during diagnostic test exceeds 3 minutes

During diagnostic test, cell fails to reach 50 mmHg within 3 minutes; displayed only on service screen

Ranked causes

  1. 1. Leak in hose connections (common)
  2. 2. Leak in cells (common)
  3. 3. Resuscitation port not fully inserted (occasional)
  4. 4. Low compressor output (common)
  5. 5. Kinked supply hose (occasional)
  6. 6. Blockage in baffle fitting (occasional)
First check

Inspect hose connections for leaks

Test point
All hose fittings and connections
Expected
No leaks; connections secure

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Compressor runs continuously during rotation cell inflation caution

Compressor operates at 100% for extended period during rotation cell inflation. May also occur if rotation frequency parameters modified without stopping treatment first.

Ranked causes

  1. 1. Solenoid valve stuck or pressure sensor monitoring wrong cell (common)
  2. 2. Compressor output below specification (occasional)
  3. 3. Supply tubing kinked or obstructed (occasional)
  4. 4. Bulkhead fitting blocked at foot-end box (occasional)
  5. 5. Main valve 'M' is blocked (occasional)
  6. 6. CPR plug or o-ring not properly seated (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Diagnostic test cell fails to inflate during production test

During diagnostic test from service screen, cell does not reach 50 mmHg pressure within 3 minutes; alarm code 92 displayed.

Ranked causes

  1. 1. Tubing leaks at connections (common)
  2. 2. Leaks in test cell (common)
  3. 3. CPR engagement or O-ring failure (occasional)
  4. 4. Low compressor output flow (occasional)
  5. 5. Kinked supply tubing in support surface (occasional)
  6. 6. Plugged bulkhead fitting at foot box (occasional)
First check

Inspect tubing connections for leaks

Test point
All tubing junctions and fittings
Expected
No visible leaks or separation

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Diagnostic test cells fail to reach 50 mmHg pressure within 3 minutes

Cells do not inflate to 50 mmHg during production leak diagnostic test on service display

Ranked causes

  1. 1. Leak in supply tubing connections (common)
  2. 2. Leak in cells (common)
  3. 3. CPR not fully connected or O-rings damaged (occasional)
  4. 4. Compressor output below specification (common)
  5. 5. Support surface supply tube kinked or damaged (occasional)
  6. 6. Footend coupling blocked or defective (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Excessive temperature in foot end assembly caution

Foot end enclosure temperature abnormally high, triggering overheat alarm and automatic power shutdown. System displays auto overload shutdown message.

Ranked causes

  1. 1. Kinked or obstructed supply tubing (common)
  2. 2. Low-loss tubing kinked or blocked (common)
  3. 3. Connector coupling at foot end blocked or failed (occasional)
  4. 4. Compressor outlet tubing kinked (occasional)
  5. 5. Supply tubing too close to or in contact with temperature sensor U5 (occasional)
  6. 6. Transformer no-load current excessive (rare)
First check

Allow foot end assembly to cool and reset service alarm

Test point
Foot end enclosure temperature
Expected
Temperature drops below 60°C after cooling

7 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Large pressure leak detected patient safety

Support surface pressure setpoint not achieved within 10 minutes during inflation. System unable to complete Preparing Surface screen. Service Required alert displayed.

Ranked causes

  1. 1. Tubing leaks at connections (common)
  2. 2. Leakage from support surface cells (common)
  3. 3. CPR connection improperly connected or seals compromised (occasional)
  4. 4. Compressor output flow below specification during initial inflation (occasional)
  5. 5. Supply tubing kinked or obstructed (occasional)
  6. 6. Foot end connector coupling blocked or failed (occasional)
First check

Inspect all tubing connections for air leaks

Test point
All connection points between tubing and components
Expected
No air leaks, all connections sealed

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Percussion cell did not inflate to 50 mmHg within 3 minutes during diagnostic test caution

During diagnostic test from Service screen, the tested cell fails to reach 50 mmHg pressure within 3 minute timeout.

Ranked causes

  1. 1. Tubing leak at connections (common)
  2. 2. Air cell leak (common)
  3. 3. CPR plug or o-ring not properly seated (occasional)
  4. 4. Compressor output below specification (common)
  5. 5. Supply tubing kinked or obstructed (occasional)
  6. 6. Bulkhead fitting blocked at foot-end box (occasional)
First check

Inspect tubing connections for leaks

Test point
All tubing connections
Expected
No air leaks at connections

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Percussion motor does not run

Percussion motor fails to operate during therapy; no motor rotation or momentary spinning then stopping.

Ranked causes

  1. 1. Drive belt broken or worn (common)
  2. 2. Percussion shaft binding or bearing wear (common)
  3. 3. Loose connectors at head box control PC board (common)
  4. 4. Percussion motor failure or control board fault (occasional)
  5. 5. Optical sensor failure (occasional)
  6. 7. Misaligned optical sensor (occasional)
First check

Inspect drive belt condition

Test point
Percussion belt
Expected
Belt intact, no fraying or wear

6 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Percussion motor not working

Percussion motor does not operate; may fail to start or run intermittently

Ranked causes

  1. 1. Defective or worn drive belt (common)
  2. 2. Percussion shaft stuck or bearing wear (common)
  3. 3. Loose connectors at control board (occasional)
  4. 4. Motor not receiving power (common)
  5. 5. Optical sensor malfunction (occasional)
  6. 6. Sensor alignment incorrect (occasional)
First check

Inspect drive belt for defects or wear

Test point
Percussion motor drive belt
Expected
Belt intact and not worn

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressure setpoint not reached within time limit patient safety

Support surface pressure setpoint not attained within 10 minutes; system cannot complete Preparing Surface screen and displays service-required message

Ranked causes

  1. 1. Leak in tubing connections (common)
  2. 2. Leak in support surface bladder cells (common)
  3. 3. CPR button cap not fully seated or O-ring damaged (occasional)
  4. 4. Compressor output flow rate low (occasional)
  5. 5. Supply tubing kinked or twisted (occasional)
  6. 6. Bulkhead fitting blocked (occasional)
First check

Inspect all tubing connections for leaks

Test point
All tubing connection points
Expected
No air leaks detected during inflation
Tools
Soap solution or pressure gauge

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor select solenoid valve issue - compressor runs continuously

Compressor runs at 100% continuously while filling rotation cell; may occur when rotation speed parameters changed without stopping therapy first

Ranked causes

  1. 1. Sensor select solenoid valve stuck de-energized or energized; pressure transducer monitoring wrong cell (common)
  2. 2. Low compressor output during initial inflation (common)
  3. 3. Kinked supply hose in mattress deck (common)
  4. 4. Wall connector at foot-end control box clogged (occasional)
  5. 5. Main valve (marked M) stuck (occasional)
  6. 6. CPR latch or O-ring defective (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor select solenoid valve malfunction caution

Compressor running at 100% while filling rotation cell; also occurs when rotation parameters changed without stopping aid

Ranked causes

  1. 1. Sensor select solenoid valve stuck in energized state (occasional)
  2. 2. Low compressor output (common)
  3. 3. Kinked supply hose (occasional)
  4. 4. Blockage in baffle fitting (occasional)
  5. 5. Main valve stuck (rare)
  6. 6. Resuscitation port not fully inserted (occasional)
First check

Observe valve during diagnostic test rotation cell fill

Test point
Pressure display during left cell (patient perspective) inflation
Expected
Pressure display increases when left cell fills

1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor select solenoid valve malfunction - compressor at 100% during rotation

Compressor runs continuously at 100% while filling rotation cell, or error triggered by changing rotation rate without stopping therapy first

Ranked causes

  1. 1. Sensor select solenoid valve stuck or misaligned, monitoring wrong cell (common)
  2. 2. Compressor output below specification (common)
  3. 3. Support surface supply tube kinked or damaged (occasional)
  4. 4. Footend coupling blocked or defective (occasional)
  5. 5. Main valve M stuck, diverting airflow away from cells (occasional)
  6. 6. CPR not fully connected or O-rings damaged (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor select solenoid valve malfunction - compressor overrun caution

Compressor runs at 100% for extended time during rotation cell filling; error code 104-0. May occur if rotation rate parameters changed without stopping therapy.

Ranked causes

  1. 1. Sensor select solenoid valve stuck in wrong state (common)
  2. 2. Low compressor output flow (occasional)
  3. 3. Kinked supply tubing in support surface (occasional)
  4. 4. Plugged bulkhead fitting at foot box (occasional)
  5. 5. Main valve sticking or blocking air flow (occasional)
  6. 6. CPR engagement or O-ring seal failure (occasional)
First check

Run diagnostic test and observe left cell inflation

Test point
Pressure display during diagnostic test
Expected
Pressure display changes during left rotation cell fill

1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Surface pressure does not reach set peak within 10 minutes patient safety

Product fails to reach set peak pressure within 10 minutes; may not complete 'Preparing Surface' screen display

Ranked causes

  1. 1. Hose leaks at connections (common)
  2. 2. Leaks in mattress cells (common)
  3. 3. CPR latch and O-ring seals faulty (occasional)
  4. 4. Compressor output below 28-32 LPM during initial inflation (occasional)
  5. 5. Kinked supply hose in mattress (occasional)
  6. 6. Wall fitting blocked at foot-of-bed control box (occasional)
First check

Inspect hose connections for leaks

Test point
All hose connection points
Expected
No air leaks at connections

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Surface pressure setpoint not reached within time limit patient safety

Support surface failed to reach the target pressure setpoint within 10 minutes; 'Preparing Surface' screen did not complete; message 'Service Required by Qualified Service Personnel' is displayed.

Ranked causes

  1. 1. Air hose leak at connections (common)
  2. 2. Leak in support surface air cells (common)
  3. 3. CPR button not fully inserted or O-ring damaged (common)
  4. 4. Compressor output below specification (occasional)
  5. 5. Kink in air supply hose to support surface (occasional)
  6. 6. Blockage in guard fitting on foot end controller (occasional)
First check

Inspect all air hose connections for leaks

Test point
All air hose connection points
Expected
No air leakage at connections; all connections tight

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CPR pendant communication lost caution

No communication between touchscreen display and headboard control box. Headboard emits sound on disconnect. System displays pendant disconnected condition.

Ranked causes

  1. 1. Installation procedure not followed; pendant not connected before power-up (occasional)
  2. 2. Black connector not secured or torque-locked (common)
  3. 3. Wiring continuity loss in black cables from foot PCB to display PCB (common)
  4. 4. Continuity loss in blue, white/orange, white/blue, blue/white conductors between foot and head boxes (common)
  5. 5. Continuity loss in yellow and clear cables between foot and head end (occasional)
First check

Verify black connector fully seated and torque-locked

Test point
Black cable connector between pendant and headboard
Expected
Connector fully seated, no movement, secure torque-lock

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Huge pressure leak detected

Set point pressure not reached within 10 minutes; unit unable to complete 'Preparing Surface' screen; displays 'Service Required by Qualified Service Personnel' message.

Ranked causes

  1. 1. Tubing leaks at connections (common)
  2. 2. Cell leaks (common)
  3. 3. CPR engagement or O-ring failure (occasional)
  4. 4. Kinked supply tubing in support surface (occasional)
  5. 5. Plugged bulkhead fitting at foot box (occasional)
First check

Verify compressor output flow rate

Test point
Foot box compressor outlet
Expected
28 to 32 lpm, 28-32
Tools
Flow meter

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No communication between touchscreen and head-end controls caution

Touchscreen cannot communicate with head-end control box; head-end box beeps during disconnection

Ranked causes

  1. 1. Installation procedure not followed; bed powered on without dial connected (occasional)
  2. 2. Black cable connector loose or unlocked (common)
  3. 3. Black cable continuity fault (occasional)
  4. 4. Blue cable wire continuity fault (occasional)
  5. 5. Yellow cable wire continuity fault (occasional)
First check

Verify black cable connector is tight and locked

Test point
Black cable connector
Expected
Connector fully seated and locked

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pendant disconnected heartbeat lost

No communication between touchscreen and headbox control; headbox beeps during disconnect.

Ranked causes

  1. 1. Installation procedure not followed; powered without display connected (common)
  2. 2. Black cable connector not secure or twisted locked (common)
  3. 3. Black cable wire continuity failure (occasional)
  4. 4. Blue cable wire continuity failure (occasional)
  5. 5. Yellow cable clear wire continuity failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Stepper motor cannot find home position caution

Stepper motor/rotary valve unable to locate initial position within specified step count after power-up

Ranked causes

  1. 1. Stepper motor not completing home sequence correctly (occasional)
  2. 2. Stepper motor or valve mechanism binding or jamming (occasional)
  3. 3. Stepper motor control PCB connectors not fully engaged (occasional)
  4. 4. Optical sensor power supply voltage low or absent (occasional)
  5. 5. Optical sensor output signal incorrect (occasional)
First check

Verify stepper motor home position sequence

Test point
Stepper motor and rotary valve during power-up
Expected
Two complete rotations, then stops with flag in front of optical sensor

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Stepper motor could not find home position caution

Stepper motor unable to locate home within specified number of steps; error code 77 displayed.

Ranked causes

  1. 1. Stepper motor not completing revolutions or stopping at flag (common)
  2. 2. Valve bound or jammed (common)
  3. 3. Connectors not fully seated at TB9 and TB10 (occasional)
  4. 4. Optical sensor voltage missing (occasional)
  5. 5. Optical sensor output voltage incorrect (occasional)
First check

Measure optical sensor voltage

Test point
TB9 orange and green wires
Expected
1.2 volts V

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Stepper motor does not complete path retraction within step count limit

Stepper motor does not find return path in specified number of steps during startup

Ranked causes

  1. 1. Stepper motor or rotary valve does not complete full rotations during startup (common)
  2. 2. Rotary valve restriction or jamming (occasional)
  3. 3. TB9 or TB10 connectors not fully seated (occasional)
  4. 4. Optical sensor missing 1.2 V supply voltage (occasional)
  5. 5. Optical sensor output voltage out of range (occasional)
First check

Observe stepper motor and rotary valve operation during startup

Test point
Stepper motor and rotary valve assembly
Expected
Motor completes two full rotations, marker aligns with optical sensor at end

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Stepper motor Find Home position failure caution

Stepper motor unable to locate Find Home position within specified number of steps. Rotary valve cannot home properly. Alarm generated on stepper motor failure.

Ranked causes

  1. 1. Stepper motor or rotary valve does not complete two revolutions or flag not positioned in front of optical sensor (common)
  2. 2. Rotary valve stiff, limited, or stuck; not rotating freely by hand (common)
  3. 3. Stepper motor and valve connectors not fully seated on TB9 and TB10 PCB (occasional)
  4. 4. Optical sensor supply voltage out of specification (occasional)
  5. 5. Optical sensor output voltage out of specification when flag aligned or misaligned with sensor (occasional)
First check

Verify stepper motor and rotary valve complete two revolutions and flag homes

Test point
Motor and valve operation during unit power-on
Expected
Two complete revolutions with flag stopped in front of optical sensor

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Stepper motor home position not found caution

Stepper motor/rotary valve could not locate its home position within the specified number of steps during startup.

Ranked causes

  1. 1. Stepper motor not completing two full revolutions or stalled (common)
  2. 2. Valve stuck or binding mechanically (common)
  3. 3. Connector not fully seated on TB9 or TB10 (occasional)
  4. 4. Optical sensor power supply voltage out of range (occasional)
  5. 5. Optical sensor detection signal out of specification (occasional)
First check

Observe stepper motor startup operation

Test point
Stepper motor/rotary valve during system startup
Expected
Motor completes two full revolutions and stops with flag in front of optical sensor

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Left CPR button not detected patient safety

Left resuscitation button on patient side is not being detected by the system.

Ranked causes

  1. 1. CPR button not fully seated or pressed in (common)
  2. 2. Blue wires not connected to CPR switch common and normally open contacts (occasional)
  3. 3. No continuity through CPR switch (occasional)
  4. 4. Connector not fully seated on TB6 control board (occasional)
First check

Verify left CPR button is fully seated and pressed

Test point
Left CPR button on bed perimeter
Expected
Button fully inserted and engaged in seated position

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Left CPR connector not recognized patient safety

Left CPR button on patient side not detected by control system

Ranked causes

  1. 1. Left CPR connector not fully seated or snapped (common)
  2. 2. Blue wires not connected to CPR switch contacts (occasional)
  3. 3. CPR switch contact continuity broken (occasional)
  4. 4. TB6 connector not fully seated in control PCB (occasional)
First check

Visual inspection of left CPR connector

Test point
CPR connector on left patient side
Expected
Fully locked and snapped

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Left CPR plug not detected patient safety

Left side CPR emergency plug not being detected by system. Left CPR function unavailable or alarm triggered.

Ranked causes

  1. 1. Left CPR plug not fully connected or not seated with pressure (common)
  2. 2. Blue conductors not connected to CPR switch contacts on headboard (occasional)
  3. 3. CPR switch continuity failure between blue conductors (occasional)
  4. 4. Connector not fully connected to TB6 PCB on headboard (occasional)
First check

Verify left CPR plug fully connected and seated with pressure

Test point
Left side CPR plug connection on mattress
Expected
Plug fully inserted, seated with pressure, no movement

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pendant disconnected from head end controller caution

No heartbeat (communication loss) between touch screen pendant and head end control unit; head end controller emits beep during disconnection.

Ranked causes

  1. 1. Installation procedure not followed; support surface activated without pendant connected (common)
  2. 2. Black cable connector loose or not fully locked (common)
  3. 3. Continuity failure in black cable wires (occasional)
  4. 4. Continuity failure in colored cable wires between controllers (occasional)
First check

Verify black cable connector is seated and locked

Test point
Black cable connector
Expected
Connector fully seated and rotated to locked position

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pendant disconnected from head-of-bed control box

No connection between touchscreen pendant and head-of-bed control box; beep sounds when unplugged

Ranked causes

  1. 1. Black cable connector not fully seated or locked (common)
  2. 2. Wire continuity broken in black cable (occasional)
  3. 3. Continuity loss in blue cable signal wires (occasional)
  4. 4. Continuity loss in yellow cable wire (occasional)
First check

Visual inspection of black cable connector

Test point
Black cable at control boxes
Expected
Firmly seated and locked by turning

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Percussion cell inflation pressure not reached within 10 seconds

Percussion cells fail to inflate to pressure setpoint within 10 second window during therapy initiation

Ranked causes

  1. 1. Leak in percussion cells (common)
  2. 2. Compressor output below specification (common)
  3. 3. Support surface supply tube kinked or damaged (occasional)
  4. 4. Footend coupling blocked or defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Percussion cells do not inflate to set pressure

Percussion cells fail to inflate to programmed peak pressure within 10 seconds

Ranked causes

  1. 1. Leaks in percussion cells (common)
  2. 2. Low compressor output during initial inflation (common)
  3. 3. Kinked supply hose in mattress deck (common)
  4. 4. Wall connector at foot-end control box clogged (occasional)
First check

Test compressor output

Test point
Foot-end control box supply line
Expected
28–32 LPM during initial inflation, 28–32 LPM
Tools
Flow meter

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Percussion cells fail to inflate to pressure within 10 seconds caution

Percussion support surface cells do not reach configured pressure setpoint within 10 seconds of inflation attempt; alarm code 91 displayed.

Ranked causes

  1. 1. Leaks in percussion cells or tubing (common)
  2. 2. Low compressor output flow (common)
  3. 3. Kinked supply tubing in support surface (occasional)
  4. 4. Plugged bulkhead fitting at foot box (occasional)
First check

Inspect percussion cells for leaks

Test point
Percussion cell bladders
Expected
No visible leaks, tears, or punctures

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Percussion cells not inflated to target pressure within 10 seconds caution

Air cells for percussion therapy do not reach set pressure within 10 second timeout during initial inflation.

Ranked causes

  1. 1. Air cell leak (common)
  2. 2. Compressor output below specification (common)
  3. 3. Supply tubing kinked or obstructed (occasional)
  4. 4. Bulkhead fitting blocked at foot-end box (occasional)
First check

Inspect percussion cells for air leaks

Test point
Percussion cells
Expected
No visible leaks or air escape

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Percussion cells not pressurized within 10 seconds

Percussion air cells fail to reach set pressure within 10 seconds during inflation

Ranked causes

  1. 1. Leak in percussion cells (common)
  2. 2. Low compressor output (common)
  3. 3. Kinked supply hose (occasional)
  4. 4. Blockage in baffle fitting (occasional)
First check

Visually inspect percussion cells for leaks

Test point
Percussion cell assemblies
Expected
No visible leaks

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Right CPR button not detected patient safety

Right resuscitation button on patient side is not being detected by the system.

Ranked causes

  1. 1. CPR button not fully seated or pressed in (common)
  2. 2. Purple wires not connected to CPR switch common and normally open contacts (occasional)
  3. 3. No continuity through CPR switch (occasional)
  4. 4. Connector not fully seated on TB6 control board (occasional)
First check

Verify right CPR button is fully seated and pressed

Test point
Right CPR button on bed perimeter
Expected
Button fully inserted and engaged in seated position

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Right CPR connector not recognized patient safety

Right CPR button on patient side not detected by control system

Ranked causes

  1. 1. Right CPR connector not fully seated or snapped (common)
  2. 2. Purple wires not connected to CPR switch contacts (occasional)
  3. 3. CPR switch contact continuity broken (occasional)
  4. 4. TB6 connector not fully seated in control PCB (occasional)
First check

Visual inspection of right CPR connector

Test point
CPR connector on right patient side
Expected
Fully locked and snapped

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Right CPR plug not detected patient safety

Right side CPR emergency plug not being detected by system. Right CPR function unavailable or alarm triggered.

Ranked causes

  1. 1. Right CPR plug not fully connected or not seated with pressure (common)
  2. 2. Violet conductors not connected to CPR switch contacts (occasional)
  3. 3. CPR switch continuity failure between violet conductors (occasional)
  4. 4. Connector not fully connected to TB6 PCB on headboard (occasional)
First check

Verify right CPR plug fully connected and seated with pressure

Test point
Right side CPR plug connection on mattress
Expected
Plug fully inserted, seated with pressure, no movement

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Support surface will not inflate caution

Mattress does not inflate or inflate completely; user cannot achieve proper air support

Ranked causes

  1. 1. Air leak in system (common)
  2. 2. Unresolved error messages logged in system (occasional)
  3. 3. Pump not running (common)
  4. 4. Main air hose disconnected or kinked (occasional)
First check

Touch screen for alarm messages

Test point
Touch screen display
Expected
No critical errors

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

12 V power supply voltage below 10 V high voltage

12 V power supply voltage drops below 10 V.

Ranked causes

  1. 1. Loss of input supply voltage (unknown)
  2. 2. 12 V regulator failure (unknown)
  3. 3. Control board has failed (occasional)
First check

Verify primary supply voltage at control board

Test point
J4 pins 3 and 4, and TB1 pins 2 and 3
Expected
24–30 V DC unfiltered, 24–30

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

12V power supply below 10 volts high voltage

12-volt supply voltage drops below 10V threshold

Ranked causes

  1. 1. Insufficient unfiltered 24-30V input at power PCB or control PCB (common)
  2. 2. 12V regulator output low (common)
  3. 3. Control PCB failure (occasional)
First check

Measure unfiltered DC input voltage at power PCB

Test point
J4 pin 3 to pin 4
Expected
24-30 VCC, 24-30

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Excessive temperature in head end assembly caution

Head end enclosure temperature abnormally high above 60°C, triggering overheat alarm and automatic power shutdown.

Ranked causes

  1. 1. Percussion motor overheated or running excessively (common)
  2. 2. Percussion motor belt tension loose or improper (common)
  3. 3. Connecting rod bearings worn or seized (occasional)
First check

Allow head end assembly to cool

Test point
Head end enclosure temperature
Expected
Temperature drops below 60°C

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Footboard has no support surfaces display but support surface running and head box alarming caution

Touch screen displays no mattress surface information; mattress is inflating/deflating; head box is sounding alarm

Ranked causes

  1. 1. Foot box connector disconnected from bed controller (common)
  2. 2. Red connector improperly connected (occasional)
  3. 3. Controller module failed (rare)
First check

Black connector at foot box to bed controller

Test point
Black connector between foot box and bed frame controller
Expected
Firmly seated connection

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Head box overheated caution

Head box temperature exceeds 140°F (60°C); unit auto-shuts down with message 'Auto. Overload Caused a Power Shutdown.'

Ranked causes

  1. 1. Overheated percussion motor (common)
  2. 2. Incorrect percussion motor belt tension (common)
  3. 3. Worn connecting rod bearing (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Head end controller overtemperature shutdown caution

Head end control unit temperature exceeds 60 degrees Celsius; system displays auto-shutdown message and power is disabled.

Ranked causes

  1. 1. Percussion motor overheating (common)
  2. 2. Percussion motor belt tension too tight (common)
  3. 3. Connecting rod bearing failure (occasional)
First check

Inspect percussion motor for excessive heat

Test point
Percussion motor housing
Expected
Motor cool enough to touch

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Head-end box overheating caution

Excessive temperature in head-end control box above 60°C; system displays auto-shutdown message and halts operation

Ranked causes

  1. 1. Percussion motor overheated (common)
  2. 2. Percussion motor belt tension incorrect (occasional)
  3. 3. Connection rod bearing worn or dry (occasional)
First check

Verify percussion motor is not overheated

Test point
Percussion motor
Expected
Motor temperature normal, not excessively hot to touch

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Head-of-bed control box overtemperature caution

Temperature in head-of-bed control box exceeds 60 °C; displays auto overload power shutdown message

Ranked causes

  1. 1. Percussion motor overheating (common)
  2. 2. Percussion motor belt tension too high (occasional)
  3. 3. Connecting rod bearing seized or worn (occasional)
First check

Inspect percussion motor for overheating

Test point
Percussion motor assembly
Expected
Motor cool to touch, no visible damage or debris

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No pressure difference between rotation circuits

System detects no pressure difference between two rotation circuits; both rotation cells inflating simultaneously

Ranked causes

  1. 1. CPR latch or O-ring defective (common)
  2. 2. Leak path between rotation cells (common)
  3. 3. Alarm sensitivity issue at very low rotation speed or when patient moves or no patient on surface (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No pressure differential detected between rotation circuits caution

System failed to detect pressure difference between two rotation circuits; both rotation cells inflate simultaneously.

Ranked causes

  1. 1. CPR plug or o-ring not properly seated (occasional)
  2. 2. Leak path exists between rotation cells (occasional)
  3. 3. Alarm triggered by low rotation frequency, patient movement, or absent patient (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rotation cells both inflate simultaneously - no pressure differential

System detects no pressure difference between two rotation circuits; both rotation cells inflate at same time

Ranked causes

  1. 1. CPR not fully connected or O-rings damaged (common)
  2. 2. Leak between rotation cells (common)
  3. 3. Alarm sensitivity increased by low rotation rate, patient movement, or no patient on surface (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor select solenoid valve malfunction - dual cell filling caution

System fails to detect pressure difference between two rotation circuits; both rotation cells filling simultaneously; error code 104-1.

Ranked causes

  1. 1. CPR engagement or O-ring seal failure (common)
  2. 2. Leakage path between rotation cells (common)
  3. 3. Low rotation rate settings or patient movement interference (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

12-volt power supply below 10 volts caution

12 VDC power supply voltage measured below 10 volts

Ranked causes

  1. 1. Low or missing unfiltered DC supply at power board (common)
  2. 2. Defective 12V regulator on control board (common)
First check

Measure unfiltered DC voltage on power board

Test point
Power board J4 pins 3 and 4
Expected
24 to 30 VDC, 24-30 VDC
Tools
Voltmeter

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

12-volt power supply below minimum high voltage

12-volt power supply drops below 10 volts; alarm code 101 displayed.

Ranked causes

  1. 1. Low or failed 24-30 volt input supply (common)
  2. 2. Voltage regulator or control board failure (occasional)
First check

Measure unfiltered input voltage

Test point
Power PC board J4 pin 3 to pin 4
Expected
24 to 30 volts DC VDC, 24-30 VDC
Tools
DMM

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

12V power supply falls below 10V high voltage

12-volt power supply voltage drops below 10 volts

Ranked causes

  1. 1. Main power supply or regulator failure (common)
  2. 2. Control board 12V regulator defective (occasional)
First check

Measure unfiltered DC voltage at power board

Test point
J4 pins 3 and 4
Expected
24–30 V DC, 24–30 V
Tools
DMM

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Bladder will not inflate or deflate
  • Both CPR buttons not detected
  • Both CPR connectors not recognized
  • Both CPR plugs not detected
  • Diagnostic test cannot start
  • Electromagnetic valve over-current detected
  • Excessive pressure detected in unintended air cell
  • Excessive pressure in unintended air cell
  • Over-current detected in solenoid valve
  • Over-current detected on percussion motor
  • Over-current detected on stepper motor drive
  • Over-pressure detected in air cell
  • Over-pressure detected in non-filling cell
  • Percussion motor over-current alarm
  • Percussion motor over-current detected
  • Percussion motor overcurrent detected
  • Percussion motor overvoltage detected
  • Solenoid valve over-current detected
  • Solenoid valve overcurrent detected
  • Solenoid valve overvoltage detected
  • Stepper motor over-current detected
  • Stepper motor overcurrent detected
  • Stepper motor overvoltage detected
  • Surface diagnostic test fails or shows inaccurate results
  • Abnormal foot box temperature
  • Abnormal foot-of-bed control box temperature
  • Abnormal head box temperature
  • Compressor restart cycling too frequent
  • Compressor restarted 20 times with less than 20 seconds between restarts
  • Control board memory corrupted
  • Control panel memory corrupted
  • Controller has detected corrupted memory
  • Diagnostic test not completing or reporting inaccurate results
  • Diagnostic test produces inaccurate results
  • Excessive over-temperature at foot end within 24 hours
  • Excessive over-temperature at head end within 24 hours
  • Foot end controller abnormal temperature warning
  • Foot end high temperature alarm occurred twice in 24 hours
  • Foot end over-temperature service alert
  • Foot end over-temperature service required
  • Foot-end box low temperature warning
  • Footend temperature excessive, two or more events in 24 hours
  • Head elevation 60 degrees or greater
  • Head elevation exceeds 60 degrees
  • Head elevation exceeds 60 degrees during rotation treatment
  • Head elevation exceeds safe rotation therapy angle
  • Head end controller abnormal temperature warning
  • Head end high temperature alarm occurred twice in 24 hours
  • Head end over-temperature service alert
  • Head end over-temperature service required
  • Head-end box low temperature warning
  • Head-of-bed control box abnormal temperature
  • Headend temperature excessive, two or more events in 24 hours
  • Low pressure compressor restart cycling
  • Low pressure leak detected
  • Max Inflate function time limit exceeded
  • Max Inflate function timeout
  • Max Inflate time limit reached
  • Max Inflate timeout exceeded
  • Memory corruption detected
  • Overpressure reading in non-target cell
  • Small pressure leak or compressor cycling
  • System lost power and restarted
  • System power cycle occurred
  • System power loss and restart
  • System power reset occurred
  • System reset occurred
  • Turn Assist function time limit exceeded
  • Turn Assist function timeout
  • Turn Assist time limit reached
  • Turn Assist timeout exceeded

71 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
100Excessive pressure was sensed in an air cell that is not intended to be fillingOver-pressure detected in incorrect cell during operationerror
100Overpressure reading - excessive pressure in non-target cellExcessive pressure detected in air cell that should not be inflatingwarning
10112V Power Failure12-volt power supply voltage drops below 10 voltscritical
10112V power failure - 12V supply below 10 volts12-volt supply voltage drops below 10V thresholdcritical
103Memory corruption detected - controller memory errorController detects corrupted memory during operation or startuperror
104-0Sensor select solenoid valve error - compressor at 100% filling rotation cellCompressor runs continuously at 100% while filling rotation cell, or rotation rate parameters changed without stopping therapy firsterror
104-1Sensor Select Solenoid Valve - No Pressure DifferenceSystem detects no pressure difference between two rotation circuits; both rotation cells inflating simultaneouslyerror
104-1Sensor select solenoid valve error - both rotation cells inflating simultaneouslySystem detects no pressure difference between two rotation circuits; both cells inflate at same timewarning
134Reset occurred - system powered down and restartedSystem underwent power down and restart cycleinfo
134Reset OccurredSystem lost power and performed restartinfo
136Main circuit rotation cell leak during diagnostic testLeak detected during diagnostic test on main rotation circuiterror
136MAIN_LEAKDiagnostic test detected leak in rotation bladder circuiterror
136Main leak diagnostic test result - main leak circuit diagnosticRotation chamber leak diagnostic test conductedwarning
137Left rotation cell circuit leak during diagnostic testLeak detected during diagnostic test on left rotation circuiterror
137LEFT_SURFACE_LEAKDiagnostic test detected leak in rotation bladder circuiterror
137Left surface leak diagnostic test result - left rotation cell leak detectedLeft surface chamber leak diagnostic test conductedwarning
138RIGHT_SURFACE_LEAKDiagnostic test detected leak in rotation bladder circuiterror
138Right surface leak diagnostic test result - right rotation cell leak detectedRight surface chamber leak diagnostic test conductedwarning
138Right rotation cell circuit leak during diagnostic testLeak detected during diagnostic test on right rotation circuiterror
139Left percussion cell circuit leak during diagnostic testLeak detected during diagnostic test on left percussion circuiterror
139Left percussion leak diagnostic test result - left percussion cell leak detectedLeft percussion chamber leak diagnostic test conductedwarning
140Right percussion cell circuit leak during diagnostic testLeak detected during diagnostic test on right percussion circuiterror
140Right percussion leak diagnostic test result - right percussion cell leak detectedRight percussion chamber leak diagnostic test conductedwarning
146INVALID_COMMANDMattress control received an unrecognized or invalid commanderror
146An invalid command has been received by the support surface controllerController receives malformed or unrecognized commandwarning
65Pendant_Disconnected: No heartbeat between touch screen and head end controller of control cabinet. Head end gives beep during disconnection.Loss of communication between pendant and head end controllererror
66CPR_Left: The left resuscitation stop is not detected.Left CPR button pressed or system checks for left CPR button presence during startup or operationwarning
67CPR_Right: The right resuscitation stop is not detected.Right CPR button pressed or system checks for right CPR button presence during startup or operationwarning
68Both CPR plugs not detected by systemNeither left nor right CPR emergency plugs detected after initial checkwarning
68CPR_Both: None of the resuscitation stops are detected.Neither left nor right CPR button is detected during operation or startupwarning
69FootEndLow: Abnormal temperature in the foot end control unit. Temperature between 48.8 and 60 degrees C.Foot end controller temperature sensor reading between 48.8 and 60°C during operationinfo
69Abnormal temperature in foot end assemblyFoot end enclosure temperature between 48.8–60 °C during operationinfo
70Foot end enclosure overheat condition; automatic power shutdown triggeredFoot end temperature exceeds 60 °Ccritical
70FootEndHigh: Overtemperature in the foot end control unit. Temperature > 60 degrees C. Shows message 'Auto. Overload Caused a Power Shutdown'.Foot end controller temperature exceeds 60°Ccritical
70Foot-end overheatingTemperature in foot-end box exceeds 60°Ccritical
71HeadEndLow: Abnormal temperature in the head end control unit. Temperature between 48.8 and 60 degrees C.Head end controller temperature sensor reading between 48.8 and 60°C during operationinfo
71Abnormal temperature in head end assemblyHead end enclosure temperature between 48.8–60 °C during operationinfo
72Head-end overheatingTemperature in head-end box exceeds 60°Ccritical
72Head end enclosure overheat condition; automatic power shutdown triggeredHead end temperature exceeds 60 °Ccritical
72HeadEndHigh: Overtemperature in the head end control unit. Temperature > 60 degrees C. Shows message 'Auto. Overload Caused a Power Shutdown'.Head end controller temperature exceeds 60°Ccritical
73Turn Assist time limit exceededOperator selected Turn Assist function but did not return to therapy mode within time limitwarning
73Turn Assist nursing function operation exceeded time limitTurn Assist function remained active beyond allowed time without operator returning to therapy modewarning
74Max Inflate nursing function operation exceeded time limitMax Inflate function remained active beyond allowed time without operator returning to therapy modewarning
74Max Inflate time limit exceededOperator selected Max Inflate function but did not return to therapy mode within time limitwarning
75HugePressureLeak: The setpoint for pressure is not reached within 10 minutes. 'Preparing Surface' screen may not be completed. Message 'Service Required by Qualified Service Personnel' is displayed.Support surface unable to reach target pressure within 10 minutes of inflation starterror
75Large pressure leak detected; support surface cannot reach setpoint within 10 minutesInflation cycle unable to achieve target pressure setpoint within 10-minute windowcritical
76LowPressureLeak: The compressor has restarted 20 times with less than 20 seconds between each restart.Chronic low-level air leak detected by repeated compressor cyclingwarning
76Small pressure leak or compressor cycling; low-pressure leak event loggedCompressor restarted 20 or more times within 20-second intervalsinfo
76Compressor frequent restart detected (low pressure leak)Compressor restarted 20 times with less than 20 seconds between restartswarning
77Stepper motor Find Home failure; rotary valve unable to home to starting positionStepper motor unable to locate Find Home position within specified number of steps during initialization or operationerror
77Stepper motor path return failureStepper motor does not complete return path within specified step count during startuperror
77Stepper_Alarm: The stepper motor could not find 'Home' in the specified number of steps.Stepper motor/rotary valve unable to locate home position during startup sequenceerror
77Stepper motor home position search failed within specified step countStepper motor unable to locate optical sensor flag within expected number of revolutionserror
81Percussion motor speed alarm - motor not functioningMotor fails to operate during percussion therapy activationerror
85Headend elevation alarm - head of bed at 60 degrees or greaterRotation therapy selected with headend elevated more than 60 degreeswarning
91Main pressure leak detectedPercussion cells fail to pressurize to set point within 10 secondserror
91Main pressure leak - percussion cells fail to reach setpoint within 10 secondsPercussion cells do not inflate to pressure setpoint within 10-second window during therapy starterror
92Production test leak detectedDuring diagnostic test, cell fails to reach 50 mmHg within 3 minuteserror
92Production test leak - diagnostic cells fail to reach 50 mmHg within 3 minutesCells do not reach 50 mmHg pressure during production leak diagnostic test on service displaywarning
95Stepper Over CurrentStepper motor control circuit detects excessive currenterror
95An over-current has been detected on the stepper motor drive circuitStepper motor drive circuit over-current condition detectederror
96An over-current has been detected in one of the solenoid valvesSolenoid valve drive circuit over-current condition detectederror
96Solenoid valve overcurrent - overcurrent in solenoid circuitOvercurrent detected in one of the solenoid valve circuitserror
97Perc Over Current AlarmPercussion motor control circuit detects excessive currenterror
97Percussion motor overcurrent - overcurrent in percussion motor circuitOvercurrent detected in percussion motor drive circuiterror
98Footend temperature service required - multiple high temperature events in 24 hoursTwo or more code 70 (footend high temperature) alarms within 24-hour periodwarning
98Foot Over Temp ServiceTwo or more alarm code 70 (FootEndHigh) events within 24 hourserror
98Two or more FootEndHigh alarm code 70 occurred within 24 hoursMultiple over-temperature alarms at foot end within 24-hour periodwarning
99Headend temperature service required - multiple high temperature events in 24 hoursTwo or more code 72 (headend high temperature) alarms within 24-hour periodwarning
99Two or more high temperature alarms at head end within 24 hoursCode 72 alarm occurred twice or more in 24 hour periodwarning
99Head Over Temp ServiceTwo or more alarm code 72 (HeadEndHigh) events within 24 hourserror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI