Stryker Bed Intouch Critical Care: troubleshooting and error codes
The ServiceForge knowledge base holds 131 documented symptoms for the Stryker Bed Intouch Critical Care, with 376 ranked causes, 231 diagnostic checks with test points and expected values, and 71 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Support surface has no power caution
Mattress/support surface unit is completely unpowered; no LED lights or alarm sounds when powered on
Ranked causes
- 1. Power cord disconnected or improperly seated (common)
- 2. Foot box connector disconnected from controller (common)
- 3. Fuses blown above foot box power cord connection (occasional)
- 4. Main fuse on power board open (occasional)
- 5. No 120 VAC at power inlet (occasional)
- 6. Power inlet filter failed (occasional)
- 7. Transformer failed (rare)
- 8. Power board failed (rare)
Power cord connection
- Test point
- Power cord to wall outlet
- Expected
- Firmly seated connection
8 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot box overheated high voltage
Foot box temperature exceeds 140°F (60°C); unit auto-shuts down with message 'Auto. Overload Caused a Power Shutdown.'
Ranked causes
- 1. Kinked or restricted supply tubing (common)
- 2. Kinked low air loss tubing (common)
- 3. Plugged bulkhead fitting at foot box (occasional)
- 4. Kinked tubing from compressor to bulkhead (occasional)
- 5. Supply tubing too close or touching sensor U5 (occasional)
- 6. Blue cable white/green or orange/white wire intermittence (occasional)
- 7. Faulty transformer with high no-load current (rare)
Allow unit to cool; reset service log and cycle power
- Test point
- Foot box temperature
- Expected
- Below 140 degrees F (60 degrees C)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot end controller overtemperature shutdown caution
Foot end control unit temperature exceeds 60 degrees Celsius; system displays auto-shutdown message and power is disabled.
Ranked causes
- 1. Kink or blockage in air supply hose to support surface (common)
- 2. Kink in low-pressure relief hose (occasional)
- 3. Blockage in guard fitting on foot end controller (occasional)
- 4. Kink in compressor discharge hose (occasional)
- 5. Supply hose too close to or contacting temperature sensor U5 (occasional)
- 6. Transformer no-load current exceeds specification (occasional)
- 7. Temperature sensor white/green or orange/white wire integrity issue (occasional)
Allow unit to cool and reset service alarm
- Test point
- Foot end controller
- Expected
- Temperature drops below 60°C
7 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot-end box overheating caution
Excessive temperature in foot-end control box above 60°C; system displays auto-shutdown message and halts operation
Ranked causes
- 1. Blue cable white/green and orange/white wire short or fault (occasional)
- 2. Supply tubing kinked or twisted (common)
- 3. Low-loss air tubing kinked or twisted (occasional)
- 4. Bulkhead fitting blocked (occasional)
- 5. Compressor-to-fitting tubing kinked or twisted (occasional)
- 6. Supply tubing too close to or touching temperature sensor U5 (occasional)
- 7. Transformer no-load current excessive (rare)
Allow system to cool and reset
- Test point
- Foot-end box temperature
- Expected
- Temperature returns below 60°C
7 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot-of-bed control box overtemperature caution
Temperature in foot-of-bed control box exceeds 60 °C; displays auto overload power shutdown message
Ranked causes
- 1. Blue cable white-green and orange-white conductors short or high resistance (occasional)
- 2. Kinked or blocked supply hose in mattress (occasional)
- 3. Kinked return hose causing low air loss (occasional)
- 4. Wall fitting at foot-of-bed control box blocked or restricted (occasional)
- 5. Kinked compressor-to-wall fitting hose (occasional)
- 6. Supply hose touching sensor U5 on power board (occasional)
- 7. Transformer no-load current exceeds 50 mA (occasional)
Allow product to cool
- Test point
- Foot-of-bed control box
- Expected
- Temperature drop below 60 °C
7 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cell in diagnostic test does not inflate to 50 mmHg in 3 minutes
Production test code during diagnostic test: cell inflation fails to reach 50 mmHg pressure within 3 minutes
Ranked causes
- 1. Hose leaks at connections (common)
- 2. Leaks in cells (common)
- 3. CPR latch or O-ring defective (occasional)
- 4. Low compressor output during initial inflation (common)
- 5. Kinked supply hose in mattress deck (common)
- 6. Wall connector at foot-end control box clogged (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cell pressurization during diagnostic test exceeds 3 minutes
During diagnostic test, cell fails to reach 50 mmHg within 3 minutes; displayed only on service screen
Ranked causes
- 1. Leak in hose connections (common)
- 2. Leak in cells (common)
- 3. Resuscitation port not fully inserted (occasional)
- 4. Low compressor output (common)
- 5. Kinked supply hose (occasional)
- 6. Blockage in baffle fitting (occasional)
Inspect hose connections for leaks
- Test point
- All hose fittings and connections
- Expected
- No leaks; connections secure
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor runs continuously during rotation cell inflation caution
Compressor operates at 100% for extended period during rotation cell inflation. May also occur if rotation frequency parameters modified without stopping treatment first.
Ranked causes
- 1. Solenoid valve stuck or pressure sensor monitoring wrong cell (common)
- 2. Compressor output below specification (occasional)
- 3. Supply tubing kinked or obstructed (occasional)
- 4. Bulkhead fitting blocked at foot-end box (occasional)
- 5. Main valve 'M' is blocked (occasional)
- 6. CPR plug or o-ring not properly seated (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Diagnostic test cell fails to inflate during production test
During diagnostic test from service screen, cell does not reach 50 mmHg pressure within 3 minutes; alarm code 92 displayed.
Ranked causes
- 1. Tubing leaks at connections (common)
- 2. Leaks in test cell (common)
- 3. CPR engagement or O-ring failure (occasional)
- 4. Low compressor output flow (occasional)
- 5. Kinked supply tubing in support surface (occasional)
- 6. Plugged bulkhead fitting at foot box (occasional)
Inspect tubing connections for leaks
- Test point
- All tubing junctions and fittings
- Expected
- No visible leaks or separation
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Diagnostic test cells fail to reach 50 mmHg pressure within 3 minutes
Cells do not inflate to 50 mmHg during production leak diagnostic test on service display
Ranked causes
- 1. Leak in supply tubing connections (common)
- 2. Leak in cells (common)
- 3. CPR not fully connected or O-rings damaged (occasional)
- 4. Compressor output below specification (common)
- 5. Support surface supply tube kinked or damaged (occasional)
- 6. Footend coupling blocked or defective (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive temperature in foot end assembly caution
Foot end enclosure temperature abnormally high, triggering overheat alarm and automatic power shutdown. System displays auto overload shutdown message.
Ranked causes
- 1. Kinked or obstructed supply tubing (common)
- 2. Low-loss tubing kinked or blocked (common)
- 3. Connector coupling at foot end blocked or failed (occasional)
- 4. Compressor outlet tubing kinked (occasional)
- 5. Supply tubing too close to or in contact with temperature sensor U5 (occasional)
- 6. Transformer no-load current excessive (rare)
Allow foot end assembly to cool and reset service alarm
- Test point
- Foot end enclosure temperature
- Expected
- Temperature drops below 60°C after cooling
7 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Large pressure leak detected patient safety
Support surface pressure setpoint not achieved within 10 minutes during inflation. System unable to complete Preparing Surface screen. Service Required alert displayed.
Ranked causes
- 1. Tubing leaks at connections (common)
- 2. Leakage from support surface cells (common)
- 3. CPR connection improperly connected or seals compromised (occasional)
- 4. Compressor output flow below specification during initial inflation (occasional)
- 5. Supply tubing kinked or obstructed (occasional)
- 6. Foot end connector coupling blocked or failed (occasional)
Inspect all tubing connections for air leaks
- Test point
- All connection points between tubing and components
- Expected
- No air leaks, all connections sealed
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Percussion cell did not inflate to 50 mmHg within 3 minutes during diagnostic test caution
During diagnostic test from Service screen, the tested cell fails to reach 50 mmHg pressure within 3 minute timeout.
Ranked causes
- 1. Tubing leak at connections (common)
- 2. Air cell leak (common)
- 3. CPR plug or o-ring not properly seated (occasional)
- 4. Compressor output below specification (common)
- 5. Supply tubing kinked or obstructed (occasional)
- 6. Bulkhead fitting blocked at foot-end box (occasional)
Inspect tubing connections for leaks
- Test point
- All tubing connections
- Expected
- No air leaks at connections
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Percussion motor does not run
Percussion motor fails to operate during therapy; no motor rotation or momentary spinning then stopping.
Ranked causes
- 1. Drive belt broken or worn (common)
- 2. Percussion shaft binding or bearing wear (common)
- 3. Loose connectors at head box control PC board (common)
- 4. Percussion motor failure or control board fault (occasional)
- 5. Optical sensor failure (occasional)
- 7. Misaligned optical sensor (occasional)
Inspect drive belt condition
- Test point
- Percussion belt
- Expected
- Belt intact, no fraying or wear
6 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Percussion motor not working
Percussion motor does not operate; may fail to start or run intermittently
Ranked causes
- 1. Defective or worn drive belt (common)
- 2. Percussion shaft stuck or bearing wear (common)
- 3. Loose connectors at control board (occasional)
- 4. Motor not receiving power (common)
- 5. Optical sensor malfunction (occasional)
- 6. Sensor alignment incorrect (occasional)
Inspect drive belt for defects or wear
- Test point
- Percussion motor drive belt
- Expected
- Belt intact and not worn
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure setpoint not reached within time limit patient safety
Support surface pressure setpoint not attained within 10 minutes; system cannot complete Preparing Surface screen and displays service-required message
Ranked causes
- 1. Leak in tubing connections (common)
- 2. Leak in support surface bladder cells (common)
- 3. CPR button cap not fully seated or O-ring damaged (occasional)
- 4. Compressor output flow rate low (occasional)
- 5. Supply tubing kinked or twisted (occasional)
- 6. Bulkhead fitting blocked (occasional)
Inspect all tubing connections for leaks
- Test point
- All tubing connection points
- Expected
- No air leaks detected during inflation
- Tools
- Soap solution or pressure gauge
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor select solenoid valve issue - compressor runs continuously
Compressor runs at 100% continuously while filling rotation cell; may occur when rotation speed parameters changed without stopping therapy first
Ranked causes
- 1. Sensor select solenoid valve stuck de-energized or energized; pressure transducer monitoring wrong cell (common)
- 2. Low compressor output during initial inflation (common)
- 3. Kinked supply hose in mattress deck (common)
- 4. Wall connector at foot-end control box clogged (occasional)
- 5. Main valve (marked M) stuck (occasional)
- 6. CPR latch or O-ring defective (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor select solenoid valve malfunction caution
Compressor running at 100% while filling rotation cell; also occurs when rotation parameters changed without stopping aid
Ranked causes
- 1. Sensor select solenoid valve stuck in energized state (occasional)
- 2. Low compressor output (common)
- 3. Kinked supply hose (occasional)
- 4. Blockage in baffle fitting (occasional)
- 5. Main valve stuck (rare)
- 6. Resuscitation port not fully inserted (occasional)
Observe valve during diagnostic test rotation cell fill
- Test point
- Pressure display during left cell (patient perspective) inflation
- Expected
- Pressure display increases when left cell fills
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor select solenoid valve malfunction - compressor at 100% during rotation
Compressor runs continuously at 100% while filling rotation cell, or error triggered by changing rotation rate without stopping therapy first
Ranked causes
- 1. Sensor select solenoid valve stuck or misaligned, monitoring wrong cell (common)
- 2. Compressor output below specification (common)
- 3. Support surface supply tube kinked or damaged (occasional)
- 4. Footend coupling blocked or defective (occasional)
- 5. Main valve M stuck, diverting airflow away from cells (occasional)
- 6. CPR not fully connected or O-rings damaged (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor select solenoid valve malfunction - compressor overrun caution
Compressor runs at 100% for extended time during rotation cell filling; error code 104-0. May occur if rotation rate parameters changed without stopping therapy.
Ranked causes
- 1. Sensor select solenoid valve stuck in wrong state (common)
- 2. Low compressor output flow (occasional)
- 3. Kinked supply tubing in support surface (occasional)
- 4. Plugged bulkhead fitting at foot box (occasional)
- 5. Main valve sticking or blocking air flow (occasional)
- 6. CPR engagement or O-ring seal failure (occasional)
Run diagnostic test and observe left cell inflation
- Test point
- Pressure display during diagnostic test
- Expected
- Pressure display changes during left rotation cell fill
1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Surface pressure does not reach set peak within 10 minutes patient safety
Product fails to reach set peak pressure within 10 minutes; may not complete 'Preparing Surface' screen display
Ranked causes
- 1. Hose leaks at connections (common)
- 2. Leaks in mattress cells (common)
- 3. CPR latch and O-ring seals faulty (occasional)
- 4. Compressor output below 28-32 LPM during initial inflation (occasional)
- 5. Kinked supply hose in mattress (occasional)
- 6. Wall fitting blocked at foot-of-bed control box (occasional)
Inspect hose connections for leaks
- Test point
- All hose connection points
- Expected
- No air leaks at connections
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Surface pressure setpoint not reached within time limit patient safety
Support surface failed to reach the target pressure setpoint within 10 minutes; 'Preparing Surface' screen did not complete; message 'Service Required by Qualified Service Personnel' is displayed.
Ranked causes
- 1. Air hose leak at connections (common)
- 2. Leak in support surface air cells (common)
- 3. CPR button not fully inserted or O-ring damaged (common)
- 4. Compressor output below specification (occasional)
- 5. Kink in air supply hose to support surface (occasional)
- 6. Blockage in guard fitting on foot end controller (occasional)
Inspect all air hose connections for leaks
- Test point
- All air hose connection points
- Expected
- No air leakage at connections; all connections tight
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPR pendant communication lost caution
No communication between touchscreen display and headboard control box. Headboard emits sound on disconnect. System displays pendant disconnected condition.
Ranked causes
- 1. Installation procedure not followed; pendant not connected before power-up (occasional)
- 2. Black connector not secured or torque-locked (common)
- 3. Wiring continuity loss in black cables from foot PCB to display PCB (common)
- 4. Continuity loss in blue, white/orange, white/blue, blue/white conductors between foot and head boxes (common)
- 5. Continuity loss in yellow and clear cables between foot and head end (occasional)
Verify black connector fully seated and torque-locked
- Test point
- Black cable connector between pendant and headboard
- Expected
- Connector fully seated, no movement, secure torque-lock
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Huge pressure leak detected
Set point pressure not reached within 10 minutes; unit unable to complete 'Preparing Surface' screen; displays 'Service Required by Qualified Service Personnel' message.
Ranked causes
- 1. Tubing leaks at connections (common)
- 2. Cell leaks (common)
- 3. CPR engagement or O-ring failure (occasional)
- 4. Kinked supply tubing in support surface (occasional)
- 5. Plugged bulkhead fitting at foot box (occasional)
Verify compressor output flow rate
- Test point
- Foot box compressor outlet
- Expected
- 28 to 32 lpm, 28-32
- Tools
- Flow meter
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No communication between touchscreen and head-end controls caution
Touchscreen cannot communicate with head-end control box; head-end box beeps during disconnection
Ranked causes
- 1. Installation procedure not followed; bed powered on without dial connected (occasional)
- 2. Black cable connector loose or unlocked (common)
- 3. Black cable continuity fault (occasional)
- 4. Blue cable wire continuity fault (occasional)
- 5. Yellow cable wire continuity fault (occasional)
Verify black cable connector is tight and locked
- Test point
- Black cable connector
- Expected
- Connector fully seated and locked
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pendant disconnected heartbeat lost
No communication between touchscreen and headbox control; headbox beeps during disconnect.
Ranked causes
- 1. Installation procedure not followed; powered without display connected (common)
- 2. Black cable connector not secure or twisted locked (common)
- 3. Black cable wire continuity failure (occasional)
- 4. Blue cable wire continuity failure (occasional)
- 5. Yellow cable clear wire continuity failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Stepper motor cannot find home position caution
Stepper motor/rotary valve unable to locate initial position within specified step count after power-up
Ranked causes
- 1. Stepper motor not completing home sequence correctly (occasional)
- 2. Stepper motor or valve mechanism binding or jamming (occasional)
- 3. Stepper motor control PCB connectors not fully engaged (occasional)
- 4. Optical sensor power supply voltage low or absent (occasional)
- 5. Optical sensor output signal incorrect (occasional)
Verify stepper motor home position sequence
- Test point
- Stepper motor and rotary valve during power-up
- Expected
- Two complete rotations, then stops with flag in front of optical sensor
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Stepper motor could not find home position caution
Stepper motor unable to locate home within specified number of steps; error code 77 displayed.
Ranked causes
- 1. Stepper motor not completing revolutions or stopping at flag (common)
- 2. Valve bound or jammed (common)
- 3. Connectors not fully seated at TB9 and TB10 (occasional)
- 4. Optical sensor voltage missing (occasional)
- 5. Optical sensor output voltage incorrect (occasional)
Measure optical sensor voltage
- Test point
- TB9 orange and green wires
- Expected
- 1.2 volts V
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Stepper motor does not complete path retraction within step count limit
Stepper motor does not find return path in specified number of steps during startup
Ranked causes
- 1. Stepper motor or rotary valve does not complete full rotations during startup (common)
- 2. Rotary valve restriction or jamming (occasional)
- 3. TB9 or TB10 connectors not fully seated (occasional)
- 4. Optical sensor missing 1.2 V supply voltage (occasional)
- 5. Optical sensor output voltage out of range (occasional)
Observe stepper motor and rotary valve operation during startup
- Test point
- Stepper motor and rotary valve assembly
- Expected
- Motor completes two full rotations, marker aligns with optical sensor at end
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Stepper motor Find Home position failure caution
Stepper motor unable to locate Find Home position within specified number of steps. Rotary valve cannot home properly. Alarm generated on stepper motor failure.
Ranked causes
- 1. Stepper motor or rotary valve does not complete two revolutions or flag not positioned in front of optical sensor (common)
- 2. Rotary valve stiff, limited, or stuck; not rotating freely by hand (common)
- 3. Stepper motor and valve connectors not fully seated on TB9 and TB10 PCB (occasional)
- 4. Optical sensor supply voltage out of specification (occasional)
- 5. Optical sensor output voltage out of specification when flag aligned or misaligned with sensor (occasional)
Verify stepper motor and rotary valve complete two revolutions and flag homes
- Test point
- Motor and valve operation during unit power-on
- Expected
- Two complete revolutions with flag stopped in front of optical sensor
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Stepper motor home position not found caution
Stepper motor/rotary valve could not locate its home position within the specified number of steps during startup.
Ranked causes
- 1. Stepper motor not completing two full revolutions or stalled (common)
- 2. Valve stuck or binding mechanically (common)
- 3. Connector not fully seated on TB9 or TB10 (occasional)
- 4. Optical sensor power supply voltage out of range (occasional)
- 5. Optical sensor detection signal out of specification (occasional)
Observe stepper motor startup operation
- Test point
- Stepper motor/rotary valve during system startup
- Expected
- Motor completes two full revolutions and stops with flag in front of optical sensor
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Left CPR button not detected patient safety
Left resuscitation button on patient side is not being detected by the system.
Ranked causes
- 1. CPR button not fully seated or pressed in (common)
- 2. Blue wires not connected to CPR switch common and normally open contacts (occasional)
- 3. No continuity through CPR switch (occasional)
- 4. Connector not fully seated on TB6 control board (occasional)
Verify left CPR button is fully seated and pressed
- Test point
- Left CPR button on bed perimeter
- Expected
- Button fully inserted and engaged in seated position
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Left CPR connector not recognized patient safety
Left CPR button on patient side not detected by control system
Ranked causes
- 1. Left CPR connector not fully seated or snapped (common)
- 2. Blue wires not connected to CPR switch contacts (occasional)
- 3. CPR switch contact continuity broken (occasional)
- 4. TB6 connector not fully seated in control PCB (occasional)
Visual inspection of left CPR connector
- Test point
- CPR connector on left patient side
- Expected
- Fully locked and snapped
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Left CPR plug not detected patient safety
Left side CPR emergency plug not being detected by system. Left CPR function unavailable or alarm triggered.
Ranked causes
- 1. Left CPR plug not fully connected or not seated with pressure (common)
- 2. Blue conductors not connected to CPR switch contacts on headboard (occasional)
- 3. CPR switch continuity failure between blue conductors (occasional)
- 4. Connector not fully connected to TB6 PCB on headboard (occasional)
Verify left CPR plug fully connected and seated with pressure
- Test point
- Left side CPR plug connection on mattress
- Expected
- Plug fully inserted, seated with pressure, no movement
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pendant disconnected from head end controller caution
No heartbeat (communication loss) between touch screen pendant and head end control unit; head end controller emits beep during disconnection.
Ranked causes
- 1. Installation procedure not followed; support surface activated without pendant connected (common)
- 2. Black cable connector loose or not fully locked (common)
- 3. Continuity failure in black cable wires (occasional)
- 4. Continuity failure in colored cable wires between controllers (occasional)
Verify black cable connector is seated and locked
- Test point
- Black cable connector
- Expected
- Connector fully seated and rotated to locked position
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pendant disconnected from head-of-bed control box
No connection between touchscreen pendant and head-of-bed control box; beep sounds when unplugged
Ranked causes
- 1. Black cable connector not fully seated or locked (common)
- 2. Wire continuity broken in black cable (occasional)
- 3. Continuity loss in blue cable signal wires (occasional)
- 4. Continuity loss in yellow cable wire (occasional)
Visual inspection of black cable connector
- Test point
- Black cable at control boxes
- Expected
- Firmly seated and locked by turning
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Percussion cell inflation pressure not reached within 10 seconds
Percussion cells fail to inflate to pressure setpoint within 10 second window during therapy initiation
Ranked causes
- 1. Leak in percussion cells (common)
- 2. Compressor output below specification (common)
- 3. Support surface supply tube kinked or damaged (occasional)
- 4. Footend coupling blocked or defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Percussion cells do not inflate to set pressure
Percussion cells fail to inflate to programmed peak pressure within 10 seconds
Ranked causes
- 1. Leaks in percussion cells (common)
- 2. Low compressor output during initial inflation (common)
- 3. Kinked supply hose in mattress deck (common)
- 4. Wall connector at foot-end control box clogged (occasional)
Test compressor output
- Test point
- Foot-end control box supply line
- Expected
- 28–32 LPM during initial inflation, 28–32 LPM
- Tools
- Flow meter
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Percussion cells fail to inflate to pressure within 10 seconds caution
Percussion support surface cells do not reach configured pressure setpoint within 10 seconds of inflation attempt; alarm code 91 displayed.
Ranked causes
- 1. Leaks in percussion cells or tubing (common)
- 2. Low compressor output flow (common)
- 3. Kinked supply tubing in support surface (occasional)
- 4. Plugged bulkhead fitting at foot box (occasional)
Inspect percussion cells for leaks
- Test point
- Percussion cell bladders
- Expected
- No visible leaks, tears, or punctures
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Percussion cells not inflated to target pressure within 10 seconds caution
Air cells for percussion therapy do not reach set pressure within 10 second timeout during initial inflation.
Ranked causes
- 1. Air cell leak (common)
- 2. Compressor output below specification (common)
- 3. Supply tubing kinked or obstructed (occasional)
- 4. Bulkhead fitting blocked at foot-end box (occasional)
Inspect percussion cells for air leaks
- Test point
- Percussion cells
- Expected
- No visible leaks or air escape
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Percussion cells not pressurized within 10 seconds
Percussion air cells fail to reach set pressure within 10 seconds during inflation
Ranked causes
- 1. Leak in percussion cells (common)
- 2. Low compressor output (common)
- 3. Kinked supply hose (occasional)
- 4. Blockage in baffle fitting (occasional)
Visually inspect percussion cells for leaks
- Test point
- Percussion cell assemblies
- Expected
- No visible leaks
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Right CPR button not detected patient safety
Right resuscitation button on patient side is not being detected by the system.
Ranked causes
- 1. CPR button not fully seated or pressed in (common)
- 2. Purple wires not connected to CPR switch common and normally open contacts (occasional)
- 3. No continuity through CPR switch (occasional)
- 4. Connector not fully seated on TB6 control board (occasional)
Verify right CPR button is fully seated and pressed
- Test point
- Right CPR button on bed perimeter
- Expected
- Button fully inserted and engaged in seated position
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Right CPR connector not recognized patient safety
Right CPR button on patient side not detected by control system
Ranked causes
- 1. Right CPR connector not fully seated or snapped (common)
- 2. Purple wires not connected to CPR switch contacts (occasional)
- 3. CPR switch contact continuity broken (occasional)
- 4. TB6 connector not fully seated in control PCB (occasional)
Visual inspection of right CPR connector
- Test point
- CPR connector on right patient side
- Expected
- Fully locked and snapped
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Right CPR plug not detected patient safety
Right side CPR emergency plug not being detected by system. Right CPR function unavailable or alarm triggered.
Ranked causes
- 1. Right CPR plug not fully connected or not seated with pressure (common)
- 2. Violet conductors not connected to CPR switch contacts (occasional)
- 3. CPR switch continuity failure between violet conductors (occasional)
- 4. Connector not fully connected to TB6 PCB on headboard (occasional)
Verify right CPR plug fully connected and seated with pressure
- Test point
- Right side CPR plug connection on mattress
- Expected
- Plug fully inserted, seated with pressure, no movement
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Support surface will not inflate caution
Mattress does not inflate or inflate completely; user cannot achieve proper air support
Ranked causes
- 1. Air leak in system (common)
- 2. Unresolved error messages logged in system (occasional)
- 3. Pump not running (common)
- 4. Main air hose disconnected or kinked (occasional)
Touch screen for alarm messages
- Test point
- Touch screen display
- Expected
- No critical errors
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
12 V power supply voltage below 10 V high voltage
12 V power supply voltage drops below 10 V.
Ranked causes
- 1. Loss of input supply voltage (unknown)
- 2. 12 V regulator failure (unknown)
- 3. Control board has failed (occasional)
Verify primary supply voltage at control board
- Test point
- J4 pins 3 and 4, and TB1 pins 2 and 3
- Expected
- 24–30 V DC unfiltered, 24–30
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
12V power supply below 10 volts high voltage
12-volt supply voltage drops below 10V threshold
Ranked causes
- 1. Insufficient unfiltered 24-30V input at power PCB or control PCB (common)
- 2. 12V regulator output low (common)
- 3. Control PCB failure (occasional)
Measure unfiltered DC input voltage at power PCB
- Test point
- J4 pin 3 to pin 4
- Expected
- 24-30 VCC, 24-30
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive temperature in head end assembly caution
Head end enclosure temperature abnormally high above 60°C, triggering overheat alarm and automatic power shutdown.
Ranked causes
- 1. Percussion motor overheated or running excessively (common)
- 2. Percussion motor belt tension loose or improper (common)
- 3. Connecting rod bearings worn or seized (occasional)
Allow head end assembly to cool
- Test point
- Head end enclosure temperature
- Expected
- Temperature drops below 60°C
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Footboard has no support surfaces display but support surface running and head box alarming caution
Touch screen displays no mattress surface information; mattress is inflating/deflating; head box is sounding alarm
Ranked causes
- 1. Foot box connector disconnected from bed controller (common)
- 2. Red connector improperly connected (occasional)
- 3. Controller module failed (rare)
Black connector at foot box to bed controller
- Test point
- Black connector between foot box and bed frame controller
- Expected
- Firmly seated connection
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head box overheated caution
Head box temperature exceeds 140°F (60°C); unit auto-shuts down with message 'Auto. Overload Caused a Power Shutdown.'
Ranked causes
- 1. Overheated percussion motor (common)
- 2. Incorrect percussion motor belt tension (common)
- 3. Worn connecting rod bearing (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head end controller overtemperature shutdown caution
Head end control unit temperature exceeds 60 degrees Celsius; system displays auto-shutdown message and power is disabled.
Ranked causes
- 1. Percussion motor overheating (common)
- 2. Percussion motor belt tension too tight (common)
- 3. Connecting rod bearing failure (occasional)
Inspect percussion motor for excessive heat
- Test point
- Percussion motor housing
- Expected
- Motor cool enough to touch
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head-end box overheating caution
Excessive temperature in head-end control box above 60°C; system displays auto-shutdown message and halts operation
Ranked causes
- 1. Percussion motor overheated (common)
- 2. Percussion motor belt tension incorrect (occasional)
- 3. Connection rod bearing worn or dry (occasional)
Verify percussion motor is not overheated
- Test point
- Percussion motor
- Expected
- Motor temperature normal, not excessively hot to touch
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head-of-bed control box overtemperature caution
Temperature in head-of-bed control box exceeds 60 °C; displays auto overload power shutdown message
Ranked causes
- 1. Percussion motor overheating (common)
- 2. Percussion motor belt tension too high (occasional)
- 3. Connecting rod bearing seized or worn (occasional)
Inspect percussion motor for overheating
- Test point
- Percussion motor assembly
- Expected
- Motor cool to touch, no visible damage or debris
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No pressure difference between rotation circuits
System detects no pressure difference between two rotation circuits; both rotation cells inflating simultaneously
Ranked causes
- 1. CPR latch or O-ring defective (common)
- 2. Leak path between rotation cells (common)
- 3. Alarm sensitivity issue at very low rotation speed or when patient moves or no patient on surface (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No pressure differential detected between rotation circuits caution
System failed to detect pressure difference between two rotation circuits; both rotation cells inflate simultaneously.
Ranked causes
- 1. CPR plug or o-ring not properly seated (occasional)
- 2. Leak path exists between rotation cells (occasional)
- 3. Alarm triggered by low rotation frequency, patient movement, or absent patient (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rotation cells both inflate simultaneously - no pressure differential
System detects no pressure difference between two rotation circuits; both rotation cells inflate at same time
Ranked causes
- 1. CPR not fully connected or O-rings damaged (common)
- 2. Leak between rotation cells (common)
- 3. Alarm sensitivity increased by low rotation rate, patient movement, or no patient on surface (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor select solenoid valve malfunction - dual cell filling caution
System fails to detect pressure difference between two rotation circuits; both rotation cells filling simultaneously; error code 104-1.
Ranked causes
- 1. CPR engagement or O-ring seal failure (common)
- 2. Leakage path between rotation cells (common)
- 3. Low rotation rate settings or patient movement interference (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
12-volt power supply below 10 volts caution
12 VDC power supply voltage measured below 10 volts
Ranked causes
- 1. Low or missing unfiltered DC supply at power board (common)
- 2. Defective 12V regulator on control board (common)
Measure unfiltered DC voltage on power board
- Test point
- Power board J4 pins 3 and 4
- Expected
- 24 to 30 VDC, 24-30 VDC
- Tools
- Voltmeter
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
12-volt power supply below minimum high voltage
12-volt power supply drops below 10 volts; alarm code 101 displayed.
Ranked causes
- 1. Low or failed 24-30 volt input supply (common)
- 2. Voltage regulator or control board failure (occasional)
Measure unfiltered input voltage
- Test point
- Power PC board J4 pin 3 to pin 4
- Expected
- 24 to 30 volts DC VDC, 24-30 VDC
- Tools
- DMM
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
12V power supply falls below 10V high voltage
12-volt power supply voltage drops below 10 volts
Ranked causes
- 1. Main power supply or regulator failure (common)
- 2. Control board 12V regulator defective (occasional)
Measure unfiltered DC voltage at power board
- Test point
- J4 pins 3 and 4
- Expected
- 24–30 V DC, 24–30 V
- Tools
- DMM
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Bladder will not inflate or deflate
- Both CPR buttons not detected
- Both CPR connectors not recognized
- Both CPR plugs not detected
- Diagnostic test cannot start
- Electromagnetic valve over-current detected
- Excessive pressure detected in unintended air cell
- Excessive pressure in unintended air cell
- Over-current detected in solenoid valve
- Over-current detected on percussion motor
- Over-current detected on stepper motor drive
- Over-pressure detected in air cell
- Over-pressure detected in non-filling cell
- Percussion motor over-current alarm
- Percussion motor over-current detected
- Percussion motor overcurrent detected
- Percussion motor overvoltage detected
- Solenoid valve over-current detected
- Solenoid valve overcurrent detected
- Solenoid valve overvoltage detected
- Stepper motor over-current detected
- Stepper motor overcurrent detected
- Stepper motor overvoltage detected
- Surface diagnostic test fails or shows inaccurate results
- Abnormal foot box temperature
- Abnormal foot-of-bed control box temperature
- Abnormal head box temperature
- Compressor restart cycling too frequent
- Compressor restarted 20 times with less than 20 seconds between restarts
- Control board memory corrupted
- Control panel memory corrupted
- Controller has detected corrupted memory
- Diagnostic test not completing or reporting inaccurate results
- Diagnostic test produces inaccurate results
- Excessive over-temperature at foot end within 24 hours
- Excessive over-temperature at head end within 24 hours
- Foot end controller abnormal temperature warning
- Foot end high temperature alarm occurred twice in 24 hours
- Foot end over-temperature service alert
- Foot end over-temperature service required
- Foot-end box low temperature warning
- Footend temperature excessive, two or more events in 24 hours
- Head elevation 60 degrees or greater
- Head elevation exceeds 60 degrees
- Head elevation exceeds 60 degrees during rotation treatment
- Head elevation exceeds safe rotation therapy angle
- Head end controller abnormal temperature warning
- Head end high temperature alarm occurred twice in 24 hours
- Head end over-temperature service alert
- Head end over-temperature service required
- Head-end box low temperature warning
- Head-of-bed control box abnormal temperature
- Headend temperature excessive, two or more events in 24 hours
- Low pressure compressor restart cycling
- Low pressure leak detected
- Max Inflate function time limit exceeded
- Max Inflate function timeout
- Max Inflate time limit reached
- Max Inflate timeout exceeded
- Memory corruption detected
- Overpressure reading in non-target cell
- Small pressure leak or compressor cycling
- System lost power and restarted
- System power cycle occurred
- System power loss and restart
- System power reset occurred
- System reset occurred
- Turn Assist function time limit exceeded
- Turn Assist function timeout
- Turn Assist time limit reached
- Turn Assist timeout exceeded
71 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
100 | Excessive pressure was sensed in an air cell that is not intended to be filling | Over-pressure detected in incorrect cell during operation | error |
100 | Overpressure reading - excessive pressure in non-target cell | Excessive pressure detected in air cell that should not be inflating | warning |
101 | 12V Power Failure | 12-volt power supply voltage drops below 10 volts | critical |
101 | 12V power failure - 12V supply below 10 volts | 12-volt supply voltage drops below 10V threshold | critical |
103 | Memory corruption detected - controller memory error | Controller detects corrupted memory during operation or startup | error |
104-0 | Sensor select solenoid valve error - compressor at 100% filling rotation cell | Compressor runs continuously at 100% while filling rotation cell, or rotation rate parameters changed without stopping therapy first | error |
104-1 | Sensor Select Solenoid Valve - No Pressure Difference | System detects no pressure difference between two rotation circuits; both rotation cells inflating simultaneously | error |
104-1 | Sensor select solenoid valve error - both rotation cells inflating simultaneously | System detects no pressure difference between two rotation circuits; both cells inflate at same time | warning |
134 | Reset occurred - system powered down and restarted | System underwent power down and restart cycle | info |
134 | Reset Occurred | System lost power and performed restart | info |
136 | Main circuit rotation cell leak during diagnostic test | Leak detected during diagnostic test on main rotation circuit | error |
136 | MAIN_LEAK | Diagnostic test detected leak in rotation bladder circuit | error |
136 | Main leak diagnostic test result - main leak circuit diagnostic | Rotation chamber leak diagnostic test conducted | warning |
137 | Left rotation cell circuit leak during diagnostic test | Leak detected during diagnostic test on left rotation circuit | error |
137 | LEFT_SURFACE_LEAK | Diagnostic test detected leak in rotation bladder circuit | error |
137 | Left surface leak diagnostic test result - left rotation cell leak detected | Left surface chamber leak diagnostic test conducted | warning |
138 | RIGHT_SURFACE_LEAK | Diagnostic test detected leak in rotation bladder circuit | error |
138 | Right surface leak diagnostic test result - right rotation cell leak detected | Right surface chamber leak diagnostic test conducted | warning |
138 | Right rotation cell circuit leak during diagnostic test | Leak detected during diagnostic test on right rotation circuit | error |
139 | Left percussion cell circuit leak during diagnostic test | Leak detected during diagnostic test on left percussion circuit | error |
139 | Left percussion leak diagnostic test result - left percussion cell leak detected | Left percussion chamber leak diagnostic test conducted | warning |
140 | Right percussion cell circuit leak during diagnostic test | Leak detected during diagnostic test on right percussion circuit | error |
140 | Right percussion leak diagnostic test result - right percussion cell leak detected | Right percussion chamber leak diagnostic test conducted | warning |
146 | INVALID_COMMAND | Mattress control received an unrecognized or invalid command | error |
146 | An invalid command has been received by the support surface controller | Controller receives malformed or unrecognized command | warning |
65 | Pendant_Disconnected: No heartbeat between touch screen and head end controller of control cabinet. Head end gives beep during disconnection. | Loss of communication between pendant and head end controller | error |
66 | CPR_Left: The left resuscitation stop is not detected. | Left CPR button pressed or system checks for left CPR button presence during startup or operation | warning |
67 | CPR_Right: The right resuscitation stop is not detected. | Right CPR button pressed or system checks for right CPR button presence during startup or operation | warning |
68 | Both CPR plugs not detected by system | Neither left nor right CPR emergency plugs detected after initial check | warning |
68 | CPR_Both: None of the resuscitation stops are detected. | Neither left nor right CPR button is detected during operation or startup | warning |
69 | FootEndLow: Abnormal temperature in the foot end control unit. Temperature between 48.8 and 60 degrees C. | Foot end controller temperature sensor reading between 48.8 and 60°C during operation | info |
69 | Abnormal temperature in foot end assembly | Foot end enclosure temperature between 48.8–60 °C during operation | info |
70 | Foot end enclosure overheat condition; automatic power shutdown triggered | Foot end temperature exceeds 60 °C | critical |
70 | FootEndHigh: Overtemperature in the foot end control unit. Temperature > 60 degrees C. Shows message 'Auto. Overload Caused a Power Shutdown'. | Foot end controller temperature exceeds 60°C | critical |
70 | Foot-end overheating | Temperature in foot-end box exceeds 60°C | critical |
71 | HeadEndLow: Abnormal temperature in the head end control unit. Temperature between 48.8 and 60 degrees C. | Head end controller temperature sensor reading between 48.8 and 60°C during operation | info |
71 | Abnormal temperature in head end assembly | Head end enclosure temperature between 48.8–60 °C during operation | info |
72 | Head-end overheating | Temperature in head-end box exceeds 60°C | critical |
72 | Head end enclosure overheat condition; automatic power shutdown triggered | Head end temperature exceeds 60 °C | critical |
72 | HeadEndHigh: Overtemperature in the head end control unit. Temperature > 60 degrees C. Shows message 'Auto. Overload Caused a Power Shutdown'. | Head end controller temperature exceeds 60°C | critical |
73 | Turn Assist time limit exceeded | Operator selected Turn Assist function but did not return to therapy mode within time limit | warning |
73 | Turn Assist nursing function operation exceeded time limit | Turn Assist function remained active beyond allowed time without operator returning to therapy mode | warning |
74 | Max Inflate nursing function operation exceeded time limit | Max Inflate function remained active beyond allowed time without operator returning to therapy mode | warning |
74 | Max Inflate time limit exceeded | Operator selected Max Inflate function but did not return to therapy mode within time limit | warning |
75 | HugePressureLeak: The setpoint for pressure is not reached within 10 minutes. 'Preparing Surface' screen may not be completed. Message 'Service Required by Qualified Service Personnel' is displayed. | Support surface unable to reach target pressure within 10 minutes of inflation start | error |
75 | Large pressure leak detected; support surface cannot reach setpoint within 10 minutes | Inflation cycle unable to achieve target pressure setpoint within 10-minute window | critical |
76 | LowPressureLeak: The compressor has restarted 20 times with less than 20 seconds between each restart. | Chronic low-level air leak detected by repeated compressor cycling | warning |
76 | Small pressure leak or compressor cycling; low-pressure leak event logged | Compressor restarted 20 or more times within 20-second intervals | info |
76 | Compressor frequent restart detected (low pressure leak) | Compressor restarted 20 times with less than 20 seconds between restarts | warning |
77 | Stepper motor Find Home failure; rotary valve unable to home to starting position | Stepper motor unable to locate Find Home position within specified number of steps during initialization or operation | error |
77 | Stepper motor path return failure | Stepper motor does not complete return path within specified step count during startup | error |
77 | Stepper_Alarm: The stepper motor could not find 'Home' in the specified number of steps. | Stepper motor/rotary valve unable to locate home position during startup sequence | error |
77 | Stepper motor home position search failed within specified step count | Stepper motor unable to locate optical sensor flag within expected number of revolutions | error |
81 | Percussion motor speed alarm - motor not functioning | Motor fails to operate during percussion therapy activation | error |
85 | Headend elevation alarm - head of bed at 60 degrees or greater | Rotation therapy selected with headend elevated more than 60 degrees | warning |
91 | Main pressure leak detected | Percussion cells fail to pressurize to set point within 10 seconds | error |
91 | Main pressure leak - percussion cells fail to reach setpoint within 10 seconds | Percussion cells do not inflate to pressure setpoint within 10-second window during therapy start | error |
92 | Production test leak detected | During diagnostic test, cell fails to reach 50 mmHg within 3 minutes | error |
92 | Production test leak - diagnostic cells fail to reach 50 mmHg within 3 minutes | Cells do not reach 50 mmHg pressure during production leak diagnostic test on service display | warning |
95 | Stepper Over Current | Stepper motor control circuit detects excessive current | error |
95 | An over-current has been detected on the stepper motor drive circuit | Stepper motor drive circuit over-current condition detected | error |
96 | An over-current has been detected in one of the solenoid valves | Solenoid valve drive circuit over-current condition detected | error |
96 | Solenoid valve overcurrent - overcurrent in solenoid circuit | Overcurrent detected in one of the solenoid valve circuits | error |
97 | Perc Over Current Alarm | Percussion motor control circuit detects excessive current | error |
97 | Percussion motor overcurrent - overcurrent in percussion motor circuit | Overcurrent detected in percussion motor drive circuit | error |
98 | Footend temperature service required - multiple high temperature events in 24 hours | Two or more code 70 (footend high temperature) alarms within 24-hour period | warning |
98 | Foot Over Temp Service | Two or more alarm code 70 (FootEndHigh) events within 24 hours | error |
98 | Two or more FootEndHigh alarm code 70 occurred within 24 hours | Multiple over-temperature alarms at foot end within 24-hour period | warning |
99 | Headend temperature service required - multiple high temperature events in 24 hours | Two or more code 72 (headend high temperature) alarms within 24-hour period | warning |
99 | Two or more high temperature alarms at head end within 24 hours | Code 72 alarm occurred twice or more in 24 hour period | warning |
99 | Head Over Temp Service | Two or more alarm code 72 (HeadEndHigh) events within 24 hours | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.