Stryker 4701: troubleshooting and error codes
The ServiceForge knowledge base holds 17 documented symptoms for the Stryker 4701, with 39 ranked causes, 47 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No power caution
Bed does not receive power; no lights or control response
Ranked causes
- 1. Power cord not plugged into hospital grade outlet (common)
- 2. Circuit breaker is tripped (common)
- 3. Low voltage or no voltage at power supply or CPU board (occasional)
- 4. Check DC voltages at power supply (unknown)
- 5. Isolate fault to component or cable (unknown)
- 6. Verify AC power at CPU board (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot down motion not working caution
Foot section will not move downward
Ranked causes
- 1. Potentiometer limits were reset (common)
- 2. CPU board not providing control signal (occasional)
- 3. Siderail board or cabling fault (occasional)
- 4. Foot power board connection fault (occasional)
- 5. Foot power board failure (occasional)
Check 5 VDC at CPU board during foot down command
- Test point
- J32 (Z) on CPU board, pin 1 black to pin 5 orange
- Expected
- 5 VDC
- Tools
- DMM
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot up motion not working caution
Foot section will not move upward
Ranked causes
- 1. Potentiometer limits were reset (common)
- 2. CPU board not providing control signal (occasional)
- 3. Siderail board or cabling fault (occasional)
- 4. Foot power board connection fault (occasional)
- 5. Foot power board motor supply failure (occasional)
Check 5 VDC at CPU board during foot up command
- Test point
- J32 (Z) on CPU board, pin 1 black to pin 4 green
- Expected
- 5 VDC
- Tools
- DMM
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed down motion not working caution
Bed will not move in downward direction when button is pressed
Ranked causes
- 1. Two-pin shunt missing from connector Z (common)
- 2. Potentiometer limits were reset (occasional)
- 3. Low or no voltage to down motion circuit (occasional)
Enter Diagnostics and test bed down motion with shunt installed
- Test point
- Connector Z pins 1 and 2
- Expected
- Motion occurs
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low or no supply voltage high voltage
Insufficient AC or DC voltage detected at power supply or CPU board connectors
Ranked causes
- 1. Power supply fault or no voltage input (common)
- 2. CPU board fault or damaged connector W (occasional)
- 3. Load short circuit on power rails (occasional)
Verify AC input voltage at power supply
- Test point
- J1 pins 1 and 2 on power supply
- Expected
- 120 VAC
- Tools
- DMM
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed up motion not working caution
Bed will not move in upward direction when button is pressed
Ranked causes
- 1. Potentiometer limits were reset (common)
- 2. Low or no voltage to up motion circuit (occasional)
Check 5 VDC control signal during bed up command
- Test point
- TP 9 on CPU board, reference ground test point
- Expected
- 5 VDC
- Tools
- DMM
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fowler down motion not working caution
Fowler back rest will not move downward
Ranked causes
- 1. Potentiometer limits were reset (common)
- 2. Low or no voltage to Fowler down circuit (occasional)
Check 5 VDC control signal during Fowler down command
- Test point
- TP 5 on CPU board, reference ground test point
- Expected
- 5 VDC
- Tools
- DMM
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fowler up motion not working caution
Fowler back rest will not move upward
Ranked causes
- 1. Potentiometer limits were reset (common)
- 2. Low or no voltage to Fowler up circuit (occasional)
Check 5 VDC control signal during Fowler up command
- Test point
- TP 6 on CPU board, reference ground test point
- Expected
- 5 VDC
- Tools
- DMM
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lumbar does not get firm
Lumbar cushion firmness not increasing when firm button pressed
Ranked causes
- 1. Compressor not running (common)
- 2. Low or no voltage to lumbar firm circuit (occasional)
Check 5 VDC at CPU board during lumbar firm command
- Test point
- J32 (Z) on CPU board, pin 1 black to pin 8 red/black
- Expected
- 5 VDC
- Tools
- DMM
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on high voltage
No power to the bed unit
Ranked causes
- 1. Power cord not plugged into hospital grade wall outlet (common)
- 2. Circuit breaker is tripped (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Level/reverse Trendelenburg motion not working caution
Bed will not level or tilt with head higher than feet
Ranked causes
- 1. Low or no voltage to level motion circuit (common)
Check 5 VDC control signal during level command
- Test point
- TP 4 on CPU board, reference ground test point
- Expected
- 5 VDC
- Tools
- DMM
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lumbar does not get soft
Lumbar cushion firmness not decreasing when soft button pressed
Ranked causes
- 1. Low or no voltage to lumbar soft circuit (common)
Check 5 VDC at CPU board during lumbar soft command
- Test point
- J32 (Z) on CPU board, pin 1 black to pin 7 white/black
- Expected
- 5 VDC
- Tools
- DMM
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Seat does not get firm
Seat does not achieve firmness when seat firm button pressed; operator cannot get firm lumbar support.
Ranked causes
- 1. Low voltage or no voltage at control circuits (common)
Verify 5 VDC control signal present
- Test point
- J32 (Z) pin 1 black and pin 10 orange/black
- Expected
- 5 VDC when seat firm button pressed
- Tools
- DMM
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Seat does not get soft
Seat does not achieve softness when seat soft button pressed; operator cannot get soft lumbar support.
Ranked causes
- 1. Low voltage or no voltage at control circuits (common)
Verify 5 VDC control signal present
- Test point
- J32 (Z) pin 1 black and pin 9 green/black
- Expected
- 5 VDC when seat soft button pressed
- Tools
- DMM
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Skoocher in motion not working caution
Bed will not move toward head
Ranked causes
- 1. Low or no voltage to Skoocher in circuit (common)
Check 5 VDC control signal during Skoocher in command
- Test point
- TP 8 on CPU board, reference ground test point
- Expected
- 5 VDC
- Tools
- DMM
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Skoocher out motion not working caution
Bed will not move toward foot
Ranked causes
- 1. Low or no voltage to Skoocher out circuit (common)
Check 5 VDC control signal during Skoocher out command
- Test point
- TP 7 on CPU board, reference ground test point
- Expected
- 5 VDC
- Tools
- DMM
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trendelenburg motion not working caution
Bed will not tilt with head lower than feet
Ranked causes
- 1. Low or no voltage to Trendelenburg circuit (common)
Check 5 VDC control signal during Trendelenburg command
- Test point
- TP 4 on CPU board, reference ground test point
- Expected
- 5 VDC
- Tools
- DMM
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI