Stryker 2950: troubleshooting and error codes
The ServiceForge knowledge base holds 38 documented symptoms for the Stryker 2950, with 108 ranked causes, 78 diagnostic checks with test points and expected values, and 35 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Compressor on 100% while filling rotation cell; sensor select solenoid stuck or failing caution
Compressor runs continuously; pressure transducer may be monitoring wrong cell during rotation
Ranked causes
- 1. Sensor select solenoid valve stuck in energized state (occasional)
- 2. Sensor select solenoid valve stuck in de-energized state (occasional)
- 3. Compressor output insufficient (occasional)
- 4. Kinked supply tubing in mattress (common)
- 5. Plugged bulkhead fitting at foot box (occasional)
- 6. Main valve sticking or failing (occasional)
- 7. CPR engagement or O-ring issue (occasional)
Run diagnostic test and observe pressure display during rotation cell fill
- Test point
- service screen diagnostic test
- Expected
- Pressure display changes when cell inflates
5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Percussion motor does not run
Percussion motor fails to operate or starts then stops
Ranked causes
- 1. Drive belt broken or worn (common)
- 2. Percussion shaft binding or bearing wear (common)
- 3. Shaft pulley has loose set screws (occasional)
- 4. Percussion motor connectors not fully seated at TB2 or TB8 (occasional)
- 5. Percussion motor voltage not present or incorrect (occasional)
- 6. Optical sensor voltage missing or out of specification (occasional)
- 7. Optical sensor blue/white voltage out of specification or flag misaligned (occasional)
Inspect percussion drive belt
- Test point
- drive belt
- Expected
- Belt intact and not excessively worn
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Diagnostic test cell did not inflate to 50 mmHg within 3 minutes caution
Production or diagnostic test failed; cell inflation insufficient during service screen testing
Ranked causes
- 1. Tubing leaks at connections (common)
- 2. Leaks in cells (common)
- 3. CPR engagement or O-ring issue (occasional)
- 4. Compressor output insufficient (occasional)
- 5. Kinked supply tubing in mattress (common)
- 6. Plugged bulkhead fitting at foot box (occasional)
Inspect tubing connections for leaks
- Test point
- all tubing connections
- Expected
- No visible leaks or moisture
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mattress has no power caution
Mattress is not receiving or supplying electrical power
Ranked causes
- 1. Power cord disconnected or improperly connected (common)
- 2. Head box alarm sounding, foot box connector disconnected (common)
- 3. Fuse blown or defective (common)
- 4. Power inlet filter defective (occasional)
- 5. Power board DC voltage output defective (occasional)
- 6. Transformer defective (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mattress unable to reach set point pressure within 10 minutes patient safety
Unit may not complete Preparing Surface screen; displays Service Required message
Ranked causes
- 1. Tubing leaks at connections (common)
- 2. Leaks in air cells (common)
- 3. CPR engagement or O-ring issue (occasional)
- 4. Compressor output insufficient (occasional)
- 5. Kinked supply tubing in mattress (common)
- 6. Plugged bulkhead fitting at foot box (occasional)
Inspect tubing connections for leaks
- Test point
- all tubing connections
- Expected
- No visible leaks or moisture
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Over temperature in foot box caution
Foot box temperature exceeds 140°F (60°C); displays auto overload shutdown message
Ranked causes
- 1. Blue cable white/green or orange/white conductor intermittent or shorted (occasional)
- 2. Kinked supply tubing in mattress (common)
- 3. Kinked Low Air Loss tubing (common)
- 4. Plugged bulkhead fitting at foot box (occasional)
- 5. Kinked tubing from compressor to bulkhead fitting (occasional)
- 6. Supply tubing too close to or touching sensor U5 (rare)
Let unit cool and reset service alarm
- Test point
- mattress power and service screen
- Expected
- Temperature returns below 140°F
7 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch screen does not respond caution
Display will not accept input or commands
Ranked causes
- 1. No power to display backlight (common)
- 2. Display power supply voltage missing (occasional)
- 3. Controller or display defective (occasional)
- 4. Communication cable defective between foot and head box (occasional)
- 5. CPU board defective (occasional)
- 6. Controller connector contact problems (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No heartbeat between touchscreen and head box control
Pendant disconnected or communication lost between control units
Ranked causes
- 1. Pendant not connected during power-up (common)
- 2. Black cable connector loose or not twist locked (common)
- 3. Wire continuity break in black cable from foot box to pendant PCB (occasional)
- 4. Wire continuity break in blue cable between foot and head box (occasional)
- 5. Wire continuity break in yellow cable between foot and head box (occasional)
Verify black cable connector is secure and twist locked
- Test point
- black cable connector at head box
- Expected
- Connector fully seated and twist locked
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Stepper motor cannot find home in specified steps
Stepper motor/rotary valve does not complete home finding sequence during power-up
Ranked causes
- 1. Stepper motor or rotary valve not reaching home position correctly (occasional)
- 2. Rotary valve binding or jammed (occasional)
- 3. Connectors not fully seated in head box control PC board TB9 or TB10 (occasional)
- 4. Optical sensor voltage not present or out of range (occasional)
- 5. Optical sensor blue/white wire voltage out of specification (occasional)
Observe stepper motor home finding sequence during power-up
- Test point
- stepper motor and rotary valve
- Expected
- Motor completes two revolutions and stops with flag in front of optical sensor
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Left CPR plug not detected
CPR plug on patient left side not engaged or switch not detecting contact
Ranked causes
- 1. CPR plug not fully engaged or not snapped on patient left side (common)
- 2. Blue wires not connected to CPR switch contacts (occasional)
- 3. No continuity between blue wires when CPR plug engaged (occasional)
- 4. Connector not fully engaged to head box control PC board TB6 (occasional)
Verify left CPR plug is fully engaged and snapped
- Test point
- CPR plug on patient left side
- Expected
- Plug fully engaged and snapped into socket
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Percussion cells did not inflate to pressure within 10 seconds patient safety
Main percussion cell pressure failed to reach set point during inflation cycle
Ranked causes
- 1. Leaks in percussion cells (common)
- 2. Compressor output insufficient (occasional)
- 3. Kinked supply tubing in mattress (common)
- 4. Plugged bulkhead fitting at foot box (occasional)
Inspect percussion cells for leaks
- Test point
- all percussion cells
- Expected
- No visible leaks or air escape
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Right CPR plug not detected
CPR plug on patient right side not engaged or switch not detecting contact
Ranked causes
- 1. CPR plug not fully engaged or not snapped on patient right side (common)
- 2. Violet wires not connected to CPR switch contacts (occasional)
- 3. No continuity between violet wires when CPR plug engaged (occasional)
- 4. Connector not fully engaged to head box control PC board TB6 (occasional)
Verify right CPR plug is fully engaged and snapped
- Test point
- CPR plug on patient right side
- Expected
- Plug fully engaged and snapped into socket
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
12 volt supply drops below 10 volts
Power supply voltage insufficient; control system may malfunction
Ranked causes
- 1. 24-30 volt input power not present at power PC board or control PC board (occasional)
- 2. 12 volt output voltage not present at TP9 (occasional)
- 3. Control PC board failure or regulation circuit issue (occasional)
Measure input voltage at power PC board
- Test point
- power PC board J4 pins 3 and 4
- Expected
- 24 to 30 volts DC unfiltered VDC, 24-30 VDC
- Tools
- DMM
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mattress will not inflate patient safety
Mattress chambers do not inflate when powered on
Ranked causes
- 1. Alarm condition indicated in error log (common)
- 2. Pump not running or CPR plugs not inserted (common)
- 3. Main air hose disconnected or kinked (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No pressure difference detected between two rotation circuits caution
Both rotation cells filling simultaneously instead of alternating; system cannot track individual cell pressure
Ranked causes
- 1. CPR engagement or O-ring issue (common)
- 2. Leakage path between rotation cells (common)
- 3. Low rotation rate setting combined with patient movement or empty mattress (common)
Verify CPR engagement and O-ring condition
- Test point
- CPR plugs and O-ring seals
- Expected
- CPR fully engaged, O-rings properly seated
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Over temperature in head box caution
Head box temperature exceeds 140°F (60°C); displays auto overload shutdown message
Ranked causes
- 1. Overheated percussion motor (common)
- 2. Percussion motor belt tension incorrect (common)
- 3. Connecting rod bearing worn or damaged (occasional)
Verify percussion motor is not overheated
- Test point
- percussion motor housing
- Expected
- Motor cooled to safe temperature
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bladder will not inflate or deflate patient safety
Mattress chambers remain static and do not respond to control inputs
Ranked causes
- 1. Solenoid resistance out of specification (common)
- 2. Selector valve resistance out of specification (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Both CPR plugs not detected
Neither left nor right CPR plug is engaged or detected by the system
Ranked causes
- 1. CPR plugs not fully engaged or snapped (common)
- 2. Connector not fully engaged to head box control PC board TB6 (occasional)
Verify both CPR plugs are fully engaged and snapped
- Test point
- CPR plugs on both patient left and right sides
- Expected
- Both plugs fully engaged and snapped into sockets
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive pressure sensed in unintended air cell patient safety
Cell not intended to be filled is at high pressure; rotary valve may be stuck
Ranked causes
- 1. Rotary valve stopped at wrong position filling wrong cell (common)
- 2. Tubing connected incorrectly (occasional)
Verify rotary valve position during therapy
- Test point
- rotary valve flag position at optical sensor
- Expected
- Flag aligned to correct cell being filled
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Over-current detected in solenoid valve
Solenoid valve drive shows excessive current draw
Ranked causes
- 1. Cut or shorted wires between control PC board TB12 or TB14 and solenoid valve (occasional)
- 2. Solenoid valve failure (occasional)
Inspect wiring to solenoid valve
- Test point
- wires from control PC board TB12 or TB14 to solenoid valve
- Expected
- No cuts, nicks or short circuits
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Over-current detected on percussion motor
Percussion motor drive shows excessive current draw
Ranked causes
- 1. Cut or shorted wires between control PC board TB2 and percussion motor (occasional)
- 2. Percussion motor failure (occasional)
Inspect wiring between TB2 and percussion motor
- Test point
- wires from control PC board TB2 to percussion motor
- Expected
- No cuts, nicks or short circuits
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Over-current detected on stepper motor drive circuit
Stepper motor drive shows excessive current draw
Ranked causes
- 1. Cut or shorted wires between control PC board TB10 and stepper motor (occasional)
- 2. Rotary valve assembly failure (occasional)
Inspect wiring between TB10 and stepper motor
- Test point
- wires from control PC board TB10 to stepper motor
- Expected
- No cuts, nicks or short circuits
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pendant has no power while mattress operates caution
Wireless controller not receiving power while system is running and head box is alarming
Ranked causes
- 1. Pendant connector disconnected from foot box controller (common)
- 2. Pendant or red connector defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scale system performance degraded during dynamic therapy patient safety
Patient motion and position from active Percussion, Vibration, Rotation or Turn-Assist therapy affects scale accuracy and weight measurement reliability
Ranked causes
- 1. Scale zeroing or weighing attempted with active dynamic therapy modes (common)
- 2. Scale system not confirmed operational after mattress installation (occasional)
Verify dynamic therapy modes are inactive
- Test point
- Mattress control panel
- Expected
- All Percussion, Vibration, Rotation and Turn-Assist indicators off
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal temperature in foot box caution
Foot box temperature between 120-140°F (48.8-60°C), logged as information for trending
Ranked causes
- 1. Normal operating condition or early warning of temperature issue (unknown)
Abnormal temperature in head box caution
Head box temperature between 120-140°F (48.8-60°C), logged as information for trending
Ranked causes
- 1. Normal operating condition or early warning of temperature issue (unknown)
Alarm: Auto Overload Power Shut Down caution
Screen displays 'ALARM Auto. Overload Caused a Power Shut Down'; mattress automatically shuts down; requires immediate unplugging
Ranked causes
- 1. Power overload condition (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm: Service Required by Qualified Service Personnel patient safety
Screen displays 'ALARM Service Required by Qualified Service Personnel'; mattress cannot provide therapy; occurs during power-up or therapy selection
Ranked causes
- 1. Unspecified fault preventing therapy delivery (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor restarted 20 times with less than 20 seconds between restarts caution
Low pressure leak detected; logged as information for trending purposes
Ranked causes
- 1. Small air leak in system (common)
Controller detected corrupted memory
Control system memory error; controller requires replacement
Ranked causes
- 1. Control PC board memory failure or corruption (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head elevation at 60 degrees or greater during rotation therapy patient safety
Operator selected rotation therapy with head elevated beyond safe angle
Ranked causes
- 1. Operator error - rotation therapy selected at high elevation (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mattress diagnostic test needed patient safety
Service technician needs to run mattress diagnostic to verify bladder integrity and leak status
Ranked causes
- 1. Routine service and verification of mattress condition (common)
Verify mattress bladders are fully deflated before starting diagnostic test
- Test point
- Mattress cells
- Expected
- All cells completely deflated, no patient or external pressure on mattress
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Max Inflate time limit reached patient safety
Operator selected Max Inflate nursing feature and did not return to therapy mode
Ranked causes
- 1. Operator did not return to therapy mode after max inflate (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power cord interference with scale system caution
XPRT mattress power cord not secured, creating risk of physical interference with bed frame scale system components
Ranked causes
- 1. Power cord loose or unsecured during installation (common)
Inspect power cord routing and security
- Test point
- Mattress power cord path from outlet to mattress
- Expected
- Cord secured away from scale system and bed frame moving components
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch screen displays 'Preparing Surface' and will not proceed patient safety
Display shows preparing surface message and prevents normal operation
Ranked causes
- 1. Air leak in system (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Turn-Assist time limit reached patient safety
Operator selected turn-assist nursing feature and did not return to therapy mode
Ranked causes
- 1. Operator did not return to therapy mode after turn-assist (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Two or more foot box over-temperature alarms within 24 hours caution
Multiple FootEndHigh (code 70) alarms logged; thermal trend indicates persistent issue
Ranked causes
- 1. Recurring foot box temperature exceeding 140°F (common)
Reset service log and monitor temperature
- Test point
- service screen
- Expected
- Service log cleared; foot box temperature remains below 140°F
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Two or more head box over-temperature alarms within 24 hours caution
Multiple HeadEndHigh (code 72) alarms logged; thermal trend indicates persistent issue
Ranked causes
- 1. Recurring head box temperature exceeding 140°F (common)
Reset service log via service screen
- Test point
- service screen and power cycling
- Expected
- Service log cleared; head box temperature remains below 140°F
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
100 | Over pressure reading; excessive pressure in unintended air cell | High pressure detected in cell not intended to be filled | warning |
101 | 12V power failure; 12 volt supply drops below 10 volts | 12 volt supply voltage falls below 10 volts DC | error |
103 | Corrupted memory detected by controller | Controller detects memory corruption | critical |
104-0 | Sensor select solenoid valve issue; compressor on 100% while filling rotation cell | Compressor runs at 100% for extended time during rotation cell fill; error occurs if rotation rate parameters changed without stopping therapy | error |
104-1 | Sensor select solenoid valve issue; no pressure difference between rotation circuits | System detects both rotation cells filling simultaneously; pressure difference not detected between circuits | warning |
136 | Diagnostic test detected leak in rotation bladder circuit | Leak detection during main circuit diagnostic test | warning |
137 | Diagnostic test detected leak in left surface rotation bladder circuit | Leak detection in left surface circuit during diagnostic | warning |
138 | Diagnostic test detected leak in right surface rotation bladder circuit | Leak detection in right surface circuit during diagnostic | warning |
139 | Diagnostic test detected leak in left percussion bladder circuit | Leak detection in left percussion circuit during diagnostic | warning |
140 | Diagnostic test detected leak in right percussion bladder circuit | Leak detection in right percussion circuit during diagnostic | warning |
146 | Invalid command received by mattress controller | Malformed or unrecognized command sent to controller | error |
30 | Stepper motor did not reach home during initial rotation | Initial stepper motor rotation sequence failure | error |
6 | Failed first attempt to position the stepper motor | Stepper motor positioning error during operation | error |
65 | No heartbeat between touchscreen and head box control (pendant disconnected) | Loss of communication between control units or pendant not connected | warning |
66 | Left CPR plug not detected | System cannot detect engagement of left CPR plug | warning |
67 | Right CPR plug not detected | System cannot detect engagement of right CPR plug | warning |
68 | Both CPR plugs not detected | System cannot detect engagement of both CPR plugs | warning |
69 | Abnormal temperature in foot box | Foot box temperature between 120-140°F (48.8-60°C) | info |
70 | Over temperature in foot box; auto overload caused power shutdown | Foot box temperature exceeds 140°F (60°C) | error |
71 | Abnormal temperature in head box | Head box temperature between 120-140°F (48.8-60°C) | info |
72 | Over temperature in head box; auto overload caused power shutdown | Head box temperature exceeds 140°F (60°C) | error |
73 | Turn-Assist time limit reached | Turn-Assist feature selected but operator did not return to therapy mode | warning |
74 | Max Inflate time limit reached | Max Inflate feature selected but operator did not return to therapy mode | warning |
75 | Huge pressure leak; set point not reached within 10 minutes | System pressure fails to reach set point during Preparing Surface; unit may not complete screen | error |
76 | Low pressure leak; compressor restarted 20+ times with <20 sec intervals | Compressor restarted 20 times with less than 20 seconds between restarts | info |
77 | Stepper motor alarm; motor could not find home in specified steps | Stepper motor/rotary valve does not complete home finding sequence during power-up | error |
81 | Percussion motor speed alarm; percussion motor does not run | Percussion motor fails to start or run | error |
85 | Head elevation 60 degrees or greater during rotation therapy | Rotation therapy selected with head elevation ≥60 degrees | warning |
91 | Main pressure leak; percussion cells did not inflate within 10 seconds | Percussion cell pressure fails to reach set point during inflation cycle | error |
92 | Production test leak; diagnostic test cell did not inflate to 50 mmHg within 3 minutes | Cell under diagnostic test from service screen fails to inflate adequately | warning |
95 | Stepper motor over-current detected | Over-current detected on stepper motor drive circuit | error |
96 | Solenoid valve over-current detected | Over-current detected in solenoid valve drive circuit | error |
97 | Percussion motor over-current detected | Over-current detected on percussion motor drive circuit | error |
98 | Foot over-temperature service alarm; 2+ code 70 alarms within 24 hours | Multiple FootEndHigh (code 70) alarms logged within 24 hour period | error |
99 | Head over-temperature service alarm; 2+ code 72 alarms within 24 hours | Multiple HeadEndHigh (code 72) alarms logged within 24 hour period | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.