Spacelabs Qube Mini: troubleshooting and error codes

The ServiceForge knowledge base holds 24 documented symptoms for the Spacelabs Qube Mini, with 72 ranked causes, 48 diagnostic checks with test points and expected values, and 50 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Power-ON diagnostic test failure

One or more diagnostic tests fail during system startup; errors logged in System Startup screen upper portion

Ranked causes

  1. 1. CPU malfunction or communication error (unknown)
  2. 2. DRAM read/write failure (unknown)
  3. 3. Real-time clock not running (unknown)
  4. 4. GDS SRAM read/write failure (unknown)
  5. 5. Video memory or controller failure (unknown)
  6. 6. Ethernet register communication failure (unknown)
  7. 7. Flash memory checksum error (unknown)
  8. 8. Touchscreen controller failure (unknown)
  9. 9. Audio codec register failure (unknown)
First check

Access Boot menu during system startup

Test point
System startup sequence
Expected
Boot menu accessible before 4-3-2-1 countdown ends

1 more check with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Barcode scanner, keyboard, printer, or mouse not functioning

USB peripherals (scanner, keyboard, printer, mouse) do not operate

Ranked causes

  1. 1. Barcode scanner not in keyboard entry mode (common)
  2. 2. USB peripheral defective (occasional)
  3. 3. USB connector on Interconnect PCBA damaged (occasional)
  4. 4. Interconnect PCBA failure (occasional)
  5. 5. CPU PCBA failure (rare)
  6. 6. SDLC PCBA failure (rare)
First check

Configure barcode scanner to keyboard entry mode

Test point
Barcode scanner settings
Expected
Scanner set to keyboard entry mode

2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ethernet network link not functional

Monitor does not communicate on Ethernet network; green link LED not lit or not blinking

Ranked causes

  1. 1. Ethernet cabling or network switch malfunction (common)
  2. 2. Monitor Ethernet network configuration incorrect (occasional)
  3. 3. Interconnect PCBA Ethernet interface failure (occasional)
  4. 4. CPU PCBA failure (rare)
  5. 5. SDLC PCBA failure (rare)
First check

Verify green link LED on Ethernet connector

Test point
RJ45 connector LED
Expected
Green LED lit (link established) and blinking (network activity)

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

POD does not sign on when installed in monitor

Patient observation device (POD) fails to establish communication with monitor

Ranked causes

  1. 1. +5 VDC supply to POD is absent (common)
  2. 2. SDLC communication link failure (occasional)
  3. 3. POD Interface Cable damaged or disconnected (occasional)
  4. 4. POD Interface PCBA failure (occasional)
  5. 5. CPU PCBA failure (rare)
First check

Connect known good POD to monitor and test sign-on

Test point
POD module bay
Expected
POD signs on and communicates normally

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touchscreen unresponsive

Display present but touchscreen does not respond to input

Ranked causes

  1. 1. Touchscreen requires calibration (common)
  2. 2. Internal touchscreen cable loose or disconnected (occasional)
  3. 3. Front-End LCD/Touchscreen assembly failure (occasional)
  4. 4. Wireless/Touch PCBA failure (rare)
  5. 5. CPU PCBA failure (rare)
First check

Calibrate touchscreen via Diagnostic or Biomed menu using mouse

Test point
Touchscreen calibration interface
Expected
Touchscreen accepts calibration input and responds normally after completion

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit powers on but fails system startup

ON/OFF switch illuminates and system begins startup but does not complete boot or reach user interface; display may be blank

Ranked causes

  1. 1. Flash ROM boot sector corruption (occasional)
  2. 2. NVRAM configuration corrupted (occasional)
  3. 3. System voltage out of tolerance (occasional)
  4. 4. Power-on diagnostics failure (occasional)
  5. 5. CPU PCBA failure (rare)
First check

Access Extended Diagnostics via touchscreen or mouse

Test point
Touchscreen/mouse input
Expected
Extended Diagnostics menu appears

6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery not recognized or fails to charge

Monitor does not detect battery presence or battery does not charge despite recognition

Ranked causes

  1. 1. Battery Connector PCBA physically damaged (occasional)
  2. 2. Battery defective (common)
  3. 3. SDLC PCBA failure (occasional)
  4. 4. CPU PCBA failure (rare)
First check

Inspect Battery Connector PCBA for physical damage

Test point
Battery Connector PCBA
Expected
No physical damage visible

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

High-pitched tones heard and no display present patient safety

System emits two-frequency high-pitch tone (System Failure Alarm) with no visible output on display; alarm indicates loss of CPU communication or power supply fault

Ranked causes

  1. 1. Serial communication loss between main CPU and Always-On CPU (occasional)
  2. 2. Main CPU detected fault and triggered safety circuit (occasional)
  3. 3. Always-On CPU (power supply) detected fault and triggered safety circuit (occasional)
  4. 4. AC mains loss or battery depleted (common)
First check

Access error log for alarm trigger cause

Test point
Error log via Boot Menu
Expected
Error entry identifying CPU, Always-On CPU, or battery failure

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not power on

AC-mains indicator does not illuminate when external power connected; ON/OFF switch does not illuminate when pressed; no startup sequence begins

Ranked causes

  1. 1. External power supply not functioning (common)
  2. 2. Power DC-IN cable failure (common)
  3. 3. SDLC PCBA input power circuits failure (occasional)
  4. 4. Power PCBA or CPU PCBA failure (occasional)
First check

Verify external power supply output

Test point
Power supply green LED
Expected
Green LED illuminates

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power-ON self-test failure or incorrect LED behavior

Power button does not illuminate, AC mains LED not lit, battery LED off when battery present, or battery LED blinking at high rate

Ranked causes

  1. 1. Power supply not connected or connector loose (common)
  2. 2. Power jack defective or connector loose (occasional)
  3. 3. Battery cannot complete full charge cycle (occasional)
First check

Verify AC mains LED illuminates

Test point
Front bezel AC mains LED
Expected
Lit when plugged into AC power

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System software failure window shown after startup

Monitor completes system startup sequence but fails to enter user interface and displays System Software Failure window with error details

Ranked causes

  1. 1. Internal USB Flash drive corrupted or unformatted (common)
  2. 2. USB Flash drive PCBA hardware failure (occasional)
  3. 3. CPU PCBA failure (occasional)
First check

Check error log for directory not found entry

Test point
Error log via Boot Menu, Directory /slh0/usr/apps/
Expected
Error entry '0x00fe0400: MonInit System software failure: Directory /slh0/usr/apps/ not found'

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Boot menu access failure or display malfunction

Cannot access Boot menu during startup or display is non-functional; diagnostic information cannot be retrieved

Ranked causes

  1. 1. Countdown 4-3-2-1 completed before Boot menu key sequence entered (common)
  2. 2. Touchscreen display non-functional or unresponsive (occasional)
First check

Monitor display functional during power-ON countdown

Test point
Video output
Expected
4-3-2-1 countdown visible on screen

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG module not recognized or LEADS OFF displayed without cause patient safety

ECG parameter key appears with ??? symbol and LEADS OFF message when no patient cable connected, or ECG count and lead information not displayed when patient simulator connected

Ranked causes

  1. 1. ECG module inserted without patient cable (normal during module insertion verification) (common)
  2. 2. ECG module not communicating with monitor after patient cable connection (occasional)
First check

Verify ECG module insertion

Test point
ECG parameter display area
Expected
ECG parameter key with ??? symbol and LEADS OFF message

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Module does not sign on when installed in monitor bay

Patient module fails to establish communication with monitor when inserted into module bay

Ranked causes

  1. 1. SDLC PCBA failure (common)
  2. 2. CPU PCBA failure (occasional)
First check

Verify module is fully seated in monitor module bay

Test point
Module Bay connection
Expected
Module fully inserted and mechanically locked

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Module signs on but parameters do not function patient safety

Module communicates with monitor but parameter data (e.g. ECG) is not transmitted or displayed

Ranked causes

  1. 1. SDLC PCBA failure preventing data transmission (common)
  2. 2. CPU PCBA failure (occasional)
First check

Verify module cable connections between module bay and CPU PCBA

Test point
Module bay SDLC connectors and internal data cables
Expected
All connections fully seated and undamaged

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor fails power-on diagnostics

Monitor does not pass self-test at startup; requires manual diagnostic execution

Ranked causes

  1. 1. Software or hardware fault identified in self-test (unknown)
  2. 2. Display failure preventing diagnostic menu access (occasional)
First check

Verify serial port is configured for DIAGNOSTICS mode

Test point
Serial port setting
Expected
DIAGNOSTICS

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Xhibit Central Monitor displays busy message patient safety

Xhibit Central Monitor shows 'Data not available - [bedname] busy' instead of monitor data

Ranked causes

  1. 1. Monitor has reached maximum of eight simultaneous client connections (common)
  2. 2. Xhibit Central Monitor is on different subnet with lower priority (occasional)
First check

Review Linked Network Devices in System Info

Test point
Biomed menu System Info > Links
Expected
Eight or fewer active client connections listed

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Touchscreen

Recalibrate touchscreen controller and reset monitor

Ranked causes

  1. 1. Touchscreen calibration drift or inaccuracy (occasional)
First check

Access Boot Menu

Test point
Boot Menu main screen
Expected
Boot Menu displayed

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cybersecurity logs not received by log server

Monitor fails to transmit cybersecurity event logs to network log server

Ranked causes

  1. 1. Log Server configuration in Biomed menu is incorrect or outdated (common)
First check

Verify Log Server Setup configuration in Biomed menu

Test point
Biomed menu Log Server Setup
Expected
IP address, Port Number, SSL Certificate, SSL Key match IT-provided values

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Enter Boot Menu during startup

Access Boot menu to run extended diagnostics or view error logs during 4-3-2-1 countdown

Ranked causes

  1. 1. Need to access Boot menu diagnostics or configuration (common)
First check

Turn on monitor and watch for 4-3-2-1 countdown

Test point
display during startup
Expected
countdown visible

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Extended Diagnostics access

Need to run extended diagnostic tests beyond Power-ON tests to isolate hardware issues

Ranked causes

  1. 1. Power-ON diagnostic inconclusive or needs detailed analysis (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor startup sequence incomplete or diagnostic menu not displayed

Diagnostic menu fails to display after power ON or monitor keys do not appear on right side of screen

Ranked causes

  1. 1. Internal firmware or initialization error detected during power-ON (rare)
First check

Verify Diagnostic menu displays after power ON

Test point
Monitor display
Expected
Diagnostic menu appears after power ON button press

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power cycling results in loss of patient data patient safety

Patient data is lost when monitor is powered off and on

Ranked causes

  1. 1. Patient data not backed up to persistent storage before power loss (unknown)

System software version mismatch failure

Monitor fails power-on diagnostics; error log shows app.bin build ID, version, or checksum does not match fd_image_c.bin

Ranked causes

  1. 1. app.bin, bootrom.bin, and fd_image_c.bin are different versions (common)
First check

Access Boot Menu and review error log for software failure entry

Test point
Error Log E- functions
Expected
Entry showing app.bin mismatch with fd_image_c.bin

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
01030000Diagnostics passedAll diagnostic tests passed successfullyinfo
01030001Diagnostics failed; degraded performanceDiagnostic test failure detectederror
01030002Critical failureCritical diagnostic test failurecritical
01030003Cannot diagnose at this boot stageDiagnostics unable to run at bootunknown
01030100Diagnostics Port Test failedDiagnostic port communication failurecritical
010302008260 Test failed8260 device test failurecritical
010302018260 SCC Transmit failedSerial Controller Chip transmit failurecritical
010302028260 SCC configuration not recognizedSerial Controller Chip configuration errorcritical
01030300DRAM Test failedDynamic RAM test failurecritical
01030301DRAM Test cannot allocate memoryDRAM memory allocation failurecritical
01030400GDS RAM Test failedGraphics Display System RAM test failurecritical
01030500FLASH ROM Test failedFlash ROM test failurecritical
01030501FLASH ROM boot larger than FlashBoot code size exceeds Flash memorycritical
01030502FLASH ROM app larger than FlashApplication code size exceeds Flash memorycritical
01030503FLASH ROM boot checksum errorBoot code checksum verification failedcritical
01030504FLASH ROM app checksum errorApplication code checksum verification failedcritical
01030505FLASH ROM read/write memory test errorFlash memory read/write test failurecritical
01030600Power Subsystem Test failedPower supply test failurecritical
01030601VIN out of specInput voltage exceeds acceptable rangecritical
01030602VCOM out of specNegative voltage supply out of rangecritical
0103060318V out of spec18 V supply voltage out of rangecritical
0103060412V out of spec12 V supply voltage out of rangecritical
010306055V out of spec5 V supply voltage out of rangecritical
010306063.3V out of spec3.3 V supply voltage out of rangecritical
010306073.3V always on supply out of spec3.3 V always-on supply voltage out of rangecritical
01030608-12V out of specNegative 12 V supply voltage out of rangecritical
01030609Battery A temp out of specBattery A temperature exceeds acceptable rangewarning
0103060BPower PCB temp out of specPower board temperature exceeds acceptable rangeunknown
0103060CCannot communicate with battery ABattery communication failure detectedcritical
0103060EWrong chemistry for battery ABattery chemistry mismatch detectedwarning
01030610Wrong voltage for battery ABattery voltage out of specificationwarning
01030A00Video Test failedVideo system test failurecritical
01030A03Video configuration not recognizedVideo hardware configuration errorcritical
01030A04Video DRAM failureVideo memory test failurecritical
01030A05Video could not detect H-sync signalHorizontal sync signal not detectedcritical
01030A06Video could not detect blue video signalBlue video signal not detectedcritical
01030D00Cannot access AUDIO device registersAudio device register access failurecritical
01030D01Could not open audio deviceAudio device initialization failurecritical
01030D02Audio device did not respond correctlyAudio device communication failurecritical
01030D03Audio Loopback test failedAudio loopback test failurecritical
01030E00NVRAM Test failedNon-volatile RAM test failurecritical
01030E01NVRAM Clock not runningNVRAM clock stopped or defectivecritical
01030E02NVRAM checksum errorNVRAM data checksum verification failedcritical
01030E03NVRAM read/write memory test failedNVRAM read/write test failurecritical
0x00fe014bCMOS Error Log initializedCMOS error log initializationinfo
0x0100006fUser reset monitorManual reset initiated by userinfo
0x010000adInterrupt Broadcast Storm overNetwork broadcast storm condition clearedwarning
0x010001a6Local clock adjusted to time masterTime synchronization drift detectedinfo
0x01000212Touchscreen calibration completeTouchscreen calibration procedure completedinfo
Power-ON diagnostic failure (unspecified code)One or more hardware subsystems failed self-test during system startupSystem power-ON or cold boot; automatic diagnostic executionerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI