Sonosite C1.9 PLUS: troubleshooting and error codes
The ServiceForge knowledge base holds 12 documented symptoms for the Sonosite C1.9 PLUS, with 27 ranked causes, 17 diagnostic checks with test points and expected values, and 1 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
External monitor connection displays blank
External monitor shows no video output when connected to system external video connector
Ranked causes
- 1. Video format mismatch (common)
- 2. External monitor cable fault (occasional)
- 3. Main PCBA failure (occasional)
- 4. External monitor failure (occasional)
Verify external monitor cable is properly connected
- Test point
- external video connector and monitor input
- Expected
- Cable seated firmly on both ends
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not power on or shows no power indication
No beeps or video display when power button is pressed; no indication of power
Ranked causes
- 1. Battery not fully inserted (common)
- 2. Battery discharged or dead (common)
- 3. AC adapter failure (occasional)
- 4. Main PCBA failure (occasional)
Verify battery is fully inserted
- Test point
- battery compartment
- Expected
- Battery seated completely with no gaps
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trackball movement erratic or does not move cursor
Trackball exhibits intermittent function, loss of control, or does not move cursor properly on display
Ranked causes
- 1. Trackball contamination (common)
- 2. Trackball failure (occasional)
- 3. Control panel failure (occasional)
- 4. Main PCBA failure (rare)
Verify keyboard operation
- Test point
- control panel
- Expected
- Keyboard responds properly
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control panel key not responding
Individual keyboard keys do not respond when pressed; LED may be burned out
Ranked causes
- 1. Control panel failure (common)
- 2. LED burned out (occasional)
- 3. Main PCBA failure (rare)
Press each individual key on keyboard from patient information screen
- Test point
- control panel keys
- Expected
- Key responds and function executes
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System display works but external monitor shows blank
System display is functional but external monitor connected to external video connector displays nothing
Ranked causes
- 1. External monitor cable disconnected or faulty (common)
- 2. External monitor failure (occasional)
- 3. Main PCBA failure (rare)
Verify system display is operational
- Test point
- system display
- Expected
- System display shows video properly
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System display is blank
Display shows no video output; external monitor connection needed to verify if display or system is at fault
Ranked causes
- 1. System display failure (common)
- 2. Main PCBA failure (occasional)
Attach external monitor to external video connector
- Test point
- external video connector
- Expected
- External monitor displays properly
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not operate or runs for shorter than expected duration on battery
Battery fails to hold charge or system shuts down prematurely during operation
Ranked causes
- 1. Battery depleted (common)
- 2. AC power operation issue (occasional)
Test system on battery power only
- Test point
- battery
- Expected
- System operates for expected duration
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Artifacts visible on LCD screen patient safety
Artifacts appear on LCD screen either within the clinical image or in the area outside the clinical image
Ranked causes
- 1. Hardware and/or software errors (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Erratic or inconsistent system behavior patient safety
System exhibits erratic or inconsistent behavior during scanning
Ranked causes
- 1. Hardware failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incompatible transducer detected
System displays message 'Incompatible transducer, upgrade the system software' when required upgrade is cancelled or fails.
Ranked causes
- 1. Required transducer software upgrade rejected or failed (occasional)
Determine if upgrade is required or optional
- Test point
- Initial system message when transducer connected
- Expected
- Prompt asking 'Do you want to upgrade the system software?'
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Main PCBA assert code displayed
System displays assert code on screen indicating Main PCBA failure; maintenance icon appears
Ranked causes
- 1. Main PCBA failure with assert code (common)
Record the assert code displayed on system screen
- Test point
- system display
- Expected
- Assert code and maintenance icon visible
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System software upgrade in progress
After accepting a software upgrade from a connected transducer, system displays 'Upgrade in progress (20 minutes total)' and cannot be used during the process.
Ranked causes
- 1. Required or optional transducer software upgrade available (common)
Verify upgrade completion status
- Test point
- System display message
- Expected
- Successful upgrade or error code with contact information
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Incompatible transducer, upgrade the system software | System software is incompatible with connected transducer and cannot proceed | Required transducer upgrade is cancelled or software upgrade fails | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI