Smiths Medical paraPAC P2: troubleshooting and error codes
The ServiceForge knowledge base holds 23 documented symptoms for the Smiths Medical paraPAC P2, with 45 ranked causes, 70 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Carry case damage or wear
Case lids torn at hinges, stiffeners degraded, straps loose or damaged, edges split, zips non-functional
Ranked causes
- 1. Case lid hinge tear from mishandling (occasional)
- 2. Wooden stiffener rigidity lost (occasional)
- 3. Elastic and Velcro strapping degraded (occasional)
- 4. Case edge splitting or wear (occasional)
- 5. Carry handle or strap damage (occasional)
- 6. Zip mechanism damage or non-functional (occasional)
Inspect case lids for hinge damage
- Test point
- Case lid hinge areas
- Expected
- No tears or cracks at hinges
- Tools
- Visual inspection
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Regulator assembly physical damage or wear caution
Visual inspection reveals cracks, bends, loose components, or rubber damage
Ranked causes
- 1. T-handle bent, cracked or stiff (occasional)
- 2. Gauge bent or damaged (occasional)
- 3. Gauge rubber shroud damaged or missing (occasional)
- 4. Secondary relief valve damaged or blocked (CE regulators) (rare)
Inspect T-handle condition and operation
- Test point
- T-handle assembly
- Expected
- Straight, undamaged, screws freely
- Tools
- Visual and manual inspection
10 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator casing or component physical damage caution
Damage to gas fittings, gauge, control knobs, case, labels, or model-specific components (nebuliser outlet, manometer connection)
Ranked causes
- 1. Physical damage to ventilator body or fittings (occasional)
- 2. Nebuliser outlet blocked or plug missing (P22D only) (occasional)
- 3. Manometer connection unsecured or blocked (EP option) (occasional)
- 4. Damaged labels on front panel (rare)
Inspect gas input and output fittings for damage
- Test point
- Gas connection fittings
- Expected
- No cracks, leaks or damage to connection threads
6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Constant Flow Selector Head damage or malfunction caution
Visible damage to selector head, poor mounting, or clickstop mechanism ineffective
Ranked causes
- 1. Flow head or outlet connector damaged (occasional)
- 2. Flow head loose mounting (occasional)
- 3. Clickstop mechanism worn or damaged (occasional)
Inspect flow head for visible damage
- Test point
- Constant Flow Selector Head body and outlet
- Expected
- No damage to head or outlet connector
- Tools
- Visual inspection
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control knob misalignment or slippage on spindle
Frequency or Tidal Volume knob indicator marks misaligned with calibration marks or knob has slipped on spindle
Ranked causes
- 1. Frequency knob misaligned or slipped on spindle (occasional)
- 2. Tidal Volume knob misaligned or slipped on spindle (occasional)
- 3. On/Off, Air Mix or Variable Relief knob end cap misaligned (occasional)
Rotate Frequency knob to both extreme limits
- Test point
- Frequency control knob
- Expected
- Indicator aligns with both 8 bpm and 40 bpm calibration marks and engages click at 12 bpm
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient hose damage or incomplete circuit patient safety
Patient hose or sensing line shows damage, splits, holes, or EP option circuit incomplete or defective
Ranked causes
- 1. Hose physical damage (splits, holes) (common)
- 2. EP option sensing line defective or missing filter (occasional)
- 3. Patient valve adaptor damaged or misfit (occasional)
Inspect patient hose for visible damage
- Test point
- Corrugated hose tube
- Expected
- No splits, cracks or holes visible
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient valve assembly damage or element defect patient safety
Patient valve body cracked, scored, or membrane element torn or split
Ranked causes
- 1. Patient valve body cracked, split or scored (occasional)
- 2. Valve element (lip membrane) torn or split (common)
- 3. Annular disc split or defective (2D/22D models only) (occasional)
Inspect patient valve assembly for cracks and scoring
- Test point
- Patient valve body
- Expected
- No cracks, splits or scoring visible on connection faces
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control knob stiff or rubbing on front panel
Control knob does not move smoothly, is loose on spindle, or rubs against front panel on one or both sides
Ranked causes
- 1. Bent spindle causing knob rub and possible needle cartridge damage (occasional)
- 2. Knob loose on spindle (common)
Rotate each control knob and observe for smooth movement
- Test point
- All control knobs
- Expected
- Knobs rotate smoothly without grinding, resistance or sticking
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cylinder key condition caution
Cylinder key shows cracking or cable damage
Ranked causes
- 1. Cylinder key cracked (occasional)
- 2. Attachment cable frayed or damaged (occasional)
Inspect cylinder key for cracks
- Test point
- Cylinder key body
- Expected
- No visible cracks or fractures
- Tools
- Visual inspection
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gauge surround or front panel label not secure or damaged
Gauge surround loose or front panel label punctured, torn or not securely fastened
Ranked causes
- 1. Gauge surround loosened or damaged (rare)
- 2. Front panel label punctured, torn or loose (occasional)
Verify gauge surround is firmly fastened
- Test point
- Gauge surround attachment
- Expected
- Surround firmly secured; no movement when gently tugged
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air entrainment NRV fault
Needle on inflation pressure monitor rises jerkily to 80 cmH2O instead of smoothly in Air Mix mode
Ranked causes
- 1. Air entrainment NRV internal fault (common)
Set Tidal Volume to 200 mL, Frequency to 25 bpm, Relief valve to maximum
- Test point
- Control knobs
- Expected
- Controls set as specified
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Periodic regulator service interval verification
Determine if regulator assembly requires 4-5 year complete disassembly service using datecode chart
Ranked causes
- 1. Regulator age exceeds 4-5 years (unknown)
Record regulator serial number and model
- Test point
- Regulator assembly label/marking
- Expected
- Serial number and model clearly readable
- Tools
- Visual inspection
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Constant Flow Selector Head performance out of specification high voltage
Actual flow rate does not match calibrated flowmeter reading at set values
Ranked causes
- 1. Flow head internal wear or damage (occasional)
Verify flow head setting at 2 L/min
- Test point
- Calibrated flowmeter outlet
- Expected
- 1.8 – 2.2 L/min, 1.8 – 2.2
- Tools
- Calibrated flowmeter (e.g. 500-82636)
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gauge rotated in shroud or mispositioned
Inflation pressure monitor gauge not in correct vertical position or has rotated in shroud
Ranked causes
- 1. Gauge rotated or shifted in shroud (rare)
Verify gauge vertical position in shroud
- Test point
- Inflation pressure monitor gauge
- Expected
- Gauge positioned vertically with correct orientation; has not rotated in shroud
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Input and output connectors unsecured or loose caution
Gas input or output connectors not firmly seated or tightened on ventilator
Ranked causes
- 1. Input or output connector loosened or not fully seated (common)
Inspect input connector for secure fit and leakage
- Test point
- Gas input connector
- Expected
- Connector hand-tight or wrench-tight with no gas leakage or audible hiss
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Input hose degradation caution
Hose shows age deterioration, internal damage, delamination, blisters or damage to union ends
Ranked causes
- 1. Hose age beyond 6 years or physical damage (common)
Check hose date code label
- Test point
- Hose label
- Expected
- Date of replacement not exceeded (within 6 years)
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Outlet connector leaking high voltage
Pressure decays at visible rate from contents gauge after cylinder valve closed
Ranked causes
- 1. Outlet connector loose or damaged (occasional)
Close cylinder valve and monitor contents gauge
- Test point
- Regulator contents gauge
- Expected
- Constant pressure reading, no visible decay
- Tools
- Visual inspection
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pin-index yoke movement when disconnected from cylinder caution
Visible movement or play between pin-index yoke and regulator assembly body
Ranked causes
- 1. Yoke mounting hardware loose (occasional)
Inspect pin-index yoke for movement with regulator disconnected
- Test point
- Pin-index yoke relative to regulator body
- Expected
- No visible movement or play when yoke manipulated
- Tools
- Visual and manual inspection
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Regulator dynamic output pressure too low high voltage
Dynamic output pressure drops below minimum specification when gas is flowing
Ranked causes
- 1. Regulator performance degraded (common)
Measure dynamic output pressure during flow
- Test point
- Test Unit pressure gauge with flow switch ON
- Expected
- no less than 45 psi (310 kPa), min 45 psi
- Tools
- Test Unit with pressure gauge and flow switch
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Regulator output leaking during static condition high voltage
Pressure decays from gauge after cylinder valve closed with regulator de-pressurized
Ranked causes
- 1. Internal regulator leak (occasional)
Monitor pressure stability after cylinder valve closure
- Test point
- Test Unit pressure gauge
- Expected
- Constant reading maintained
- Tools
- Test Unit with pressure gauge
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Regulator static output pressure out of range high voltage
Static output pressure reading from test unit falls outside 50-80 psi (345-552 kPa) specification
Ranked causes
- 1. Main valve seat leaking (common)
Measure static output pressure
- Test point
- Test Unit pressure gauge
- Expected
- 50 - 80 psi (345-552 kPa), 50-80 psi
- Tools
- Test Unit with pressure gauge
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sintered filter dirty or obscured
Input connector sintered filter is clogged, dirty or cannot be seen clearly
Ranked causes
- 1. Sintered filter clogged with debris (common)
Visually inspect sintered filter in input connector
- Test point
- Input connector filter
- Expected
- Filter clean and clearly visible; no debris or dust obscuring filter element
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator continues to cycle after switched off patient safety
Main pneumatic switch fails to stop cycling when turned to Off or Demand position
Ranked causes
- 1. Main Pneumatic switch leaking (common)
Select Off or Demand position and observe cycling
- Test point
- Main pneumatic switch
- Expected
- Ventilator should stop cycling
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.