Smiths Medical Medfusion 4000: troubleshooting and error codes

The ServiceForge knowledge base holds 100 documented symptoms for the Smiths Medical Medfusion 4000, with 141 ranked causes, 111 diagnostic checks with test points and expected values, and 99 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Pump won't turn on

Pump fails to power on even after pressing power button

Ranked causes

  1. 1. Battery depleted (common)
  2. 2. Loose Main PCB to Interconnect PCB ribbon cable (occasional)
  3. 3. AC power connection issue (occasional)
  4. 4. Battery connections loose or defective (occasional)
  5. 5. Keypad cable loose or disconnected (occasional)
First check

Verify AC power connection

Test point
AC outlet and pump inlet
Expected
Pump plugged into live AC outlet

6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms do not sound

Alarms sound faintly or not at all during operation

Ranked causes

  1. 1. Alarm loudness configured to lowest level (common)
  2. 2. Speaker grid holes obstructed (occasional)
  3. 3. Speaker malfunction (occasional)
First check

Verify alarm loudness setting

Test point
Profile Settings > View Settings > Alarm Loudness
Expected
Setting should not be at lowest level

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms do not sound or sound faintly

Alarm audio output is absent or at very low volume during alarm conditions

Ranked causes

  1. 1. Alarm loudness configured to lowest level (common)
  2. 2. Speaker grid holes obstructed (occasional)
  3. 3. Speaker malfunction (rare)
First check

Verify alarm loudness setting

Test point
Profile Settings > View Settings > Alarm Loudness
Expected
Check current loudness level setting

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery does not hold charge

Battery charge capacity low or pump shuts down immediately when unplugged from AC power

Ranked causes

  1. 1. Battery ribbon cable disconnected (occasional)
  2. 2. Battery not charging adequately (occasional)
  3. 3. Battery pack defective (occasional)
First check

Check battery ribbon cable connection

Test point
Battery door interior, Interconnect PCB connector
Expected
Battery ribbon cable firmly seated and fully connected

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Force sensor not responding to test signal

Force sensor does not change output voltage when test signal applied. Strain gauge may be unplugged or bridge open.

Ranked causes

  1. 1. Force sensor bridge unplugged or open connection (common)
  2. 2. Syringe loaded during power-up self-test (occasional)
  3. 3. Main PCB offset circuit, Plunger PCB, or cable failure (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Motor does not rotate or stalls

Motor sensor detects no rotation of stepper motor. Pump cannot deliver medication.

Ranked causes

  1. 1. Motor sensor failure on Main PCB (common)
  2. 2. Worm/gear/leadscrew jamming or mechanical wear (occasional)
  3. 3. Drive train mechanical wear (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Motor rotates at incorrect speed

Motor sensor measures incorrect rotation rate for stepper motor. Delivery rate inaccurate.

Ranked causes

  1. 1. Motor sensor failure on Main PCB (common)
  2. 2. Force sensor inoperative causing motor stall on occlusion (occasional)
  3. 3. Drive train mechanical wear (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Occlusion detected in infusion line patient safety

Force sensor sensed force exceeding occlusion threshold, indicating potential line blockage or tubing issue

Ranked causes

  1. 1. Actual occlusion in infusion line (common)
  2. 2. Infusion set tubing too small for rate or fluid viscosity (occasional)
  3. 3. Force sensor requires calibration (occasional)
First check

Visually inspect infusion line for occlusion

Expected
No kinks, blockages, or disconnections visible
Tools
Visual inspection

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Plunger driver not moving correctly patient safety

Position potentiometer detects plunger driver motion anomaly during delivery, may be caused by slack in drive system, lever issues, clutch problems, or hardware failure

Ranked causes

  1. 1. Prime function not used before starting delivery (common)
  2. 2. Plunger levers moved or clutch not engaged (common)
  3. 3. Position potentiometer not calibrated or contaminated (occasional)
First check

Verify plunger lever moves freely and captures syringe plunger

Expected
Lever moves without restriction and grips plunger
Tools
Visual/tactile inspection

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Backup audible alarm not functioning

Power-up self-testing found backup audible alarm not working at correct power or frequency. Alarm buzzer malfunction.

Ranked causes

  1. 1. Alarm buzzer failure on Main PCB (rare)
  2. 2. Main PCB ACU controller failure (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery fully depleted and pump non-functional

Battery voltage dropped below low voltage threshold. Pump stops delivery and watchdog generates alarm. May occur on battery power only.

Ranked causes

  1. 1. Battery fully discharged (common)
  2. 2. Chronic battery function issue (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery not working message

Pump displays battery error message or battery status unavailable

Ranked causes

  1. 1. Battery ribbon cable disconnected (common)
  2. 2. Battery pack failure (occasional)
First check

Check battery ribbon cable connection

Test point
Battery door interior, Interconnect PCB connector
Expected
Battery ribbon cable firmly seated and fully connected

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery problem: Battery not working message

System displays battery fault or no battery detected error

Ranked causes

  1. 1. Battery ribbon cable disconnected or loose at Interconnect PCB (common)
  2. 2. Battery pack failure (occasional)
First check

Inspect battery ribbon cable connection

Test point
Battery door, Interconnect PCB connector
Expected
Battery ribbon cable firmly seated and undamaged

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Main PCB replacement software installation

Load and verify correct software version on Main PCB after replacement; restore configuration from Interconnect PCB base board

Ranked causes

  1. 1. Main PCB replaced; software and configuration must be restored (common)
  2. 2. Configuration not loaded to Main PCB (common)
First check

Verify software version after installation

Test point
Power-on screen
Expected
Software version matches expected pump software version

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Clock calendar not initialized

System date and time not set; clock defaults to incorrect date/time on startup or during operation

Ranked causes

  1. 1. Clock not initialized at startup (occasional)
  2. 2. Clock not initialized during operation (occasional)
First check

Verify system date and time

Test point
Biomed > Utilities > View time and date
Expected
Correct date and time displayed

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

External device communication timeout

External computer or device connected to pump is sending too many commands rapidly or many invalid commands. Software cannot process requests within timeout window.

Ranked causes

  1. 1. External device command rate too high (common)
  2. 2. External device sending invalid commands (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Force sensor self-test failed patient safety

Self-test performed on force sensor failed, indicating sensor malfunction or plunger driver movement during startup

Ranked causes

  1. 1. Force sensor malfunction (common)
  2. 2. Plunger driver moved during pump startup (occasional)
First check

Monitor force sensor operation using diagnostics

Test point
Biomed > Diagnostics > Monitor Analog Sensors
Expected
Force sensor responds to test signals
Tools
Biomed menu

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD display not responding or image corrupted

Display self-test failed. Display shows no image or incorrect display output.

Ranked causes

  1. 1. Display cable disconnected or loose (common)
  2. 2. LCD display hardware failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD has poor contrast

Liquid crystal display is difficult to read due to low or high contrast level

Ranked causes

  1. 1. LCD contrast requires adjustment (common)
  2. 2. LCD negative supply voltage out of range (rare)
First check

Run LCD contrast adjustment

Test point
Biomed > Calibration > Adjust Contrast
Expected
Display contrast optimized and visible

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Liquid Crystal Display has poor contrast

LCD display has low contrast or is difficult to read

Ranked causes

  1. 1. LCD contrast setting too low (common)
  2. 2. LCD negative supply voltage issue (rare)
First check

Adjust LCD contrast

Test point
Biomed > Calibration > Adjust Contrast
Expected
Contrast adjusted to acceptable level

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Motor unit assembly and drive train installation caution

Reassembly of motor, worm, worm gear, and drive train after disassembly; includes lubrication and torque specifications.

Ranked causes

  1. 1. Motor unit disassembly requires reassembly with correct lubrication and torque (common)
  2. 2. Drive train assembly requires careful handling during installation (common)
First check

Worm lubrication before motor installation

Test point
Worm hole
Expected
Filled with Dow 111 Silicone Grease Compound

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No charge light

Battery charge indicator light does not illuminate

Ranked causes

  1. 1. Pump not connected to AC power (common)
  2. 2. Battery charging issue (occasional)
First check

Verify AC power connection

Test point
AC power outlet
Expected
Pump connected and AC indicator light on

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Occlusion alarms occur frequently

Pump generates occlusion alarms more often than expected during normal operation

Ranked causes

  1. 1. Occlusion alarm limit set too low (common)
  2. 2. High backpressure in infusion line (common)
First check

Review occlusion alarm limit setting

Test point
Profile Settings or Biomed Configuration
Expected
Occlusion limit appropriate for infusion setup

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Primary speaker not producing audio alarm caution

Background or power-up self-test detects no current flowing in primary speaker. Alarm buzzer silent.

Ranked causes

  1. 1. Speaker disconnected or wire loose (common)
  2. 2. Speaker failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Program flash memory CRC failure

Flash memory contains program code with failed cyclic redundancy check. Program data corrupted.

Ranked causes

  1. 1. Flash memory corruption during software re-programming (occasional)
  2. 2. Flash memory chip failure on Main PCB (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump base not responding or shutdown failed

Pump base experienced failure during power down sequence. Base does not respond to commands.

Ranked causes

  1. 1. Power down sequence failure (occasional)
  2. 2. EEPROM read failure or disconnected Interconnect PCB (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Stepper motor not energized at startup

Power-up self-test senses no current in stepper motor drive circuit. Motor cannot operate.

Ranked causes

  1. 1. Pump operating on depleted battery power (common)
  2. 2. Motor supply or driver circuit failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Stepper motor phase A winding open circuit

Power-up self-test detected stepper motor phase A winding has opened connection. Motor cannot generate torque.

Ranked causes

  1. 1. Pump on depleted battery power during startup (common)
  2. 2. Motor winding disconnected or wire loose (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Stepper motor phase B winding open circuit

Power-up self-test detected stepper motor phase B winding has opened connection. Motor cannot generate torque.

Ranked causes

  1. 1. Pump on depleted battery power during startup (common)
  2. 2. Motor winding disconnected or wire loose (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Super-capacitor insufficient charge at startup

Power-up self-test detected super-capacitor voltage too low. Backup power supply insufficient.

Ranked causes

  1. 1. Super-capacitor low charge from new Main PCB or dead battery (common)
  2. 2. Super-capacitor hardware failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

A/D converter internal failure

Software self-test identified failure in 10-bit analog to digital converter. Sensor readings unreliable.

Ranked causes

  1. 1. A/D converter chip failure on Main PCB (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

A2D reference voltage out of range

Background self-test measured invalid range on analog-to-digital converter reference voltage. Voltage reading out of specification.

Ranked causes

  1. 1. A/D converter reference voltage regulator failure (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

AC power supply not functioning after installation high voltage

AC power indicator does not light after AC power supply replacement

Ranked causes

  1. 1. AC power supply not properly installed or defective (unknown)
First check

Verify AC power indicator

Test point
AC power indicator LED
Expected
Indicator light illuminated

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

ACU power strobe not maintained

Self-test found ACU processor not maintaining its power strobe signal during test.

Ranked causes

  1. 1. ACU processor power strobe malfunction (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ACU watchdog alarm active

Self-test found ACU watchdog alarm signal is active. Processor may have failed to service watchdog within allotted time.

Ranked causes

  1. 1. Processor software failure or watchdog timeout (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Adjust display contrast voltage

Varies onscreen contrast of LCD display. Factory default is 00BF (may vary by LCD manufacturer). Adjustable higher or lower for comfortable viewing in different lighting conditions.

Ranked causes

  1. 1. Display contrast requires adjustment for optimal viewing (occasional)
First check

Access contrast adjustment menu

Test point
Calibration menu
Expected
Adjust Contrast option selected

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Analog supply voltage invalid

Background self-test measured invalid range on analog supply voltage. Voltage reading out of specification.

Ranked causes

  1. 1. Analog supply voltage regulator failure (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Background CRC test timeout

Software CRC test failed to complete within allotted time. CRC calculation took too long.

Ranked causes

  1. 1. Software transient timing failure (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Background self-test timeout

Background self-test took longer than 8 seconds to complete, exceeding timeout threshold. Timing failure occurred.

Ranked causes

  1. 1. Software transient failure or timing issue (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Backlight connector polarity verification

Verify backlight wires are correctly connected to Main PCB; if reversed, backlight will not illuminate

Ranked causes

  1. 1. Backlight leads reversed during installation (occasional)
First check

Verify backlight illuminates after power-up

Test point
LCD backlight
Expected
Backlight illuminated during power-up sequence

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery not communicating or low

Battery gauge not communicating with pump, or very low battery signal detected while connected to AC power. Battery capacity shows as zero.

Ranked causes

  1. 1. Loose or disconnected battery connections (common)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery pack not charging caution

Battery charge indicator does not light when connected to AC power, or battery information absent from Monitor Battery Status screen

Ranked causes

  1. 1. Battery not properly connected to Interconnect PCB (common)
First check

Verify battery charge indicator lights

Test point
Battery charge indicator LED
Expected
Indicator lights when AC power connected

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery pack replacement needed caution

Battery requires replacement or is not functioning after installation

Ranked causes

  1. 1. Exhausted or damaged battery pack (common)
First check

Verify battery charge indicator lights after connecting pump to AC power

Test point
Battery charge indicator LED
Expected
Indicator lights

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery replacement needed - charge cycle limit reached patient safety

Battery charge cycle counter reaches or exceeds 400 cycles limit during Biomed diagnostics monitoring

Ranked causes

  1. 1. Battery at end-of-life charge cycle limit (common)
First check

Access Biomed > Diagnostics > Monitor Battery Status and view Charge Cycles value

Test point
Charge Cycles field
Expected
Value below 400 cycles, maximum

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery replacement needed - low full capacity patient safety

Battery full-capacity indicator falls below minimum threshold during monitoring via Biomed diagnostics

Ranked causes

  1. 1. Battery aging or degradation reducing full capacity (common)
First check

Access Biomed > Diagnostics > Monitor Battery Status and view Full-Capacity value

Test point
Full-Capacity field
Expected
Value above 3024 mAh for 22-4050-51 or above 4550 mAh for 22-4051-01, minimum threshold

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AC power supply function verification high voltage

Verification that AC power supply is functioning correctly after replacement or service

Ranked causes

  1. 1. AC power supply replaced or serviced (unknown)
First check

Verify AC power indicator lights when plugged in

Test point
AC power indicator LED
Expected
AC power indicator lights

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Battery function verification

Verify new battery is functioning nominally before returning pump to service

Ranked causes

  1. 1. New battery installation completed (common)
First check

Connect pump to AC mains and observe battery charge indicator

Test point
Battery charge indicator light
Expected
Indicator lights on

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Bottom housing replacement caution

Installation and verification of replacement bottom housing assembly with all subcomponents

Ranked causes

  1. 1. Bottom housing replacement required (unknown)
First check

Verify soft silicone gasket around lower housing lip is continuous and intact

Test point
Lower housing lip perimeter
Expected
Gasket continuous and intact with no damage

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Clutch or clutch cam replacement

After replacing clutch halves or cam, recalibrate and verify pump functionality per guidelines before returning to use.

Ranked causes

  1. 1. Clutch halves or cam replaced (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Display contrast adjustment

Procedure to adjust LCD display contrast within Biomed Calibration menu

Ranked causes

  1. 1. Display contrast requires adjustment for visibility (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Force sensor

Adjustment of force sensor settings. Sensor measures plunger force to ensure proper operation and safety limits.

Ranked causes

  1. 1. Initial force sensor calibration or recalibration required (common)
First check

Access force sensor calibration

Test point
Calibration menu
Expected
Cal Force Sensor option selected

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Interconnect PCB reinstallation and verification

Procedure to reinstall Interconnect PCB with antenna, radio module, and clear shield; verify serial number and pump functionality before return to use.

Ranked causes

  1. 1. PCB reinstallation and verification procedure (common)
First check

Verify serial number is correct following installation

Test point
pump startup screen and bottom serial number label
Expected
Serial number on startup screen matches label on pump bottom

1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Leadscrew replacement

After replacing leadscrew, recalibrate and verify pump functionality per guidelines before returning to use.

Ranked causes

  1. 1. Leadscrew replaced (common)
First check

Verify leadscrew Nylock nut torque

Test point
Nylock nut holding leadscrew on right end plate
Expected
60 in.-oz (0.42 Nm) then loosen 1/4 turn, loosen 1/4 turn after initial torque
Tools
Torque wrench, 1/4 inch wrench

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Main PCB replacement and pump function verification

After replacing Main PCB, pump signals Configuration Required alarm and requires full recalibration, reconfiguration, and function verification before return to service.

Ranked causes

  1. 1. Main PCB replaced without configuration (common)
First check

Verify Configuration correctly installed on Main PCB

Test point
Pump display after configuration load
Expected
Configuration matches facility-validated version

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Motor unit replacement

After replacing motor, worm, or worm gear, recalibrate and verify pump functionality per guidelines before returning to use.

Ranked causes

  1. 1. Motor, worm, or worm gear replaced (common)
First check

Verify motor mounting screws are torqued correctly

Test point
2 pan head screws holding motor
Expected
100 in oz (0.71 Nm), as specified
Tools
Torque screwdriver with Phillips head

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Plunger position

Adjustment of plunger position sensor settings to establish empty position and plunger head travel limits.

Ranked causes

  1. 1. Initial plunger position calibration or recalibration required (common)
First check

Access plunger position calibration menu

Test point
Calibration menu
Expected
Cal Plunger Position option selected

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Plunger position sensor

Calibration procedure to check or adjust plunger position sensor values within Biomed Calibration menu

Ranked causes

  1. 1. Plunger position sensor requires recalibration after maintenance or drift (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Position potentiometer replacement

After replacing position potentiometer, recalibrate and verify pump functionality per guidelines before returning to use.

Ranked causes

  1. 1. Position potentiometer replaced (common)
First check

Verify position potentiometer mounting screw torque

Test point
Small flat head screw at right end plate
Expected
16 in.-oz. (0.11 Nm), as specified
Tools
Torque screwdriver

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Size and position sensor

Calibration procedure to check or adjust syringe size and position sensor values within Biomed Calibration menu

Ranked causes

  1. 1. Sensor requires recalibration after maintenance or drift detected (unknown)
First check

Select Calibration menu in Biomed

Test point
Biomed > Calibration
Expected
Calibration submenu options displayed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Syringe size and position

Concurrent adjustment of syringe size sensor (barrel clamp assembly, measures external diameter) and syringe position sensor (drive train assembly, measures plunger head position for empty position and travel determination).

Ranked causes

  1. 1. Initial calibration or recalibration required (common)

5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Calibration: Syringe size sensor
  • Clutch assembly replacement
  • Configuration settings lost or invalid
  • Critical data block RAM corrupted
  • Firmware version mismatch between pump head and base
  • Force sensor signal out of range
  • Force sensor voltage invalid
  • Indicator lights do not flash
  • Key click or beep is faint or not present
  • Keypad key does not work
  • Keypad stuck or unresponsive
  • Maximum rate not available
  • MCU power strobe signal failure
  • MCU time base synchronization failure
  • Monitor software task timing failure
  • Motor current not disabled during watchdog test
  • Motor exceeded volume delivery limit
  • Motor rotation sensor failure after stop command
  • Motor supply voltage out of specification
  • Network settings corrupted or missing
  • No Profiles available
  • OS timer failed during watchdog event
  • PCB firmware recovery error: Firmware mismatch
  • Plunger position sensor voltage out of range
  • Positive supply voltage out of specification
  • Pump fails to initialize sensors
  • Pump option or feature not available
  • Pump powers on with blank screen and buzzer sounds and will not power off after firmware upgrade
  • Radio PCB removal
  • Real time clock reset
  • Serial EEPROM communication timeout
  • Speaker does not produce power-on tones
  • Speaker not functioning after installation
  • SPI bus communication failure
  • Syringe barrel clamp sensor difficulty
  • Syringe flange sensor active
  • Syringe flange sensor voltage out of range
  • Syringe Manufacturer not available
  • Syringe plunger sensor active
  • System RAM memory corrupted

40 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
A2D REFERENCE VOLTAGE BGND TESTADC reference voltage invalidBackground self-test detects invalid voltage rangecritical
ACU POWER STROBE FAILUREACU processor power strobe not maintainedACU processor fails to maintain power strobe signalcritical
ACU WATCHDOG FAILUREACU watchdog alarm activeProcessor failed to pet watchdog within allotted timecritical
ANALOG SUPPLY BGND TESTAnalog supply voltage invalidBackground self-test detects invalid voltage rangecritical
AUX CONTROLLER UNIT POSTACU POST motor current shutdown failureACU failed to shut down motor current during watchdog testcritical
BACKGROUND CRC TEST TIMEOUTBackground CRC test timeoutSoftware timing failurewarning
BACKGROUND SELF TEST TIMEOUTBackground self-test exceeds time limitBackground test takes more than 8 secondswarning
BACKUP AUDIBLE ALARM POSTBackup alarm power or frequency invalidPOST detects incorrect alarm power/frequencycritical
Backup Critical Data CorruptedBackup configuration data was lost; primary data restored backupEEPROM redundancy check failed during initializationwarning
BASE HARDWARE FAILUREBase hardware self-test failedNon-zero hardware test result or interface initialization failurecritical
BASE NOT RESPONDINGBase power-down sequence failureFailure during power down sequencecritical
BASE UNKNOWN FAILURERead EEPROM failureEEPROM read operation failscritical
BATTERY COMMUNICATION TIMEOUTBattery gauge not communicatingBattery current detected but no gauge communicationwarning
BATTERY NOT WORKINGBattery malfunctionVery low battery signal detected while on AC powerwarning
CALCULATED RATE OUT OF RANGEPump calculated an invalid delivery rate based on entered valuesInvalid combination of syringe parameters and desired rateerror
CALIBRATION REQUIREDSensor calibration failedCalibration data test fails during power-up or savewarning
CHECK CLUTCH / PLUNGER LEVERPlunger driver position not moving correctly during deliveryPosition potentiometer detects abnormal plunger driver motionerror
CHECK SYRINGE BARREL CLAMPSyringe barrel clamp sensor verification failureCertain syringe models cause barrel clamp sensor difficulty; clamp lifted during power-upwarning
CHECK SYRINGE FLANGE SENSORSyringe not loaded correctly; flange sensor activeSyringe flange size sensor detects improper loadingerror
Check Syringe Plunger SensorSyringe plunger sensor indicates incorrect loading or movement during startupSyringe not loaded correctly or plunger moved during pump startuperror
CHECK SYRINGE PLUNGER SENSORSyringe not loaded correctly; plunger sensor activeSyringe not detected as properly loaded; plunger driver moved during startuperror
Configuration RequiredNo Configuration file present on Main PCBPump powered up after Main PCB replacement without existing configurationwarning
Configuration RequiredConfiguration file not present on Main PCBMain PCB replaced or configuration lost; pump powered up without valid configurationwarning
CONTROL KEY SWITCH BGND TESTKeypad key stuck onKey found stuck during background self-testcritical
CONTROL KEY SWITCH POSTKeypad key pressed too longSingle key pressed >7sec or locked key >1mincritical
CRITICAL DATA BLOCK BGND TESTCritical delivery data corruptedBackground self-test shows data block corruptioncritical
CRITICAL DATA BLOCK POSTCritical configuration data CRC mismatchBackup and primary data fail CRC checkcritical
CRITICAL DATA FAILUREIncompatible pump settings or software data failureIncompatible setting ranges or critical flags found invalidcritical
Critical Data FailureProgrammed settings failed reverse-calculation validationAfter settings programmed, pump verification failed or incompatible ranges detectederror
D2A OFFSET VOLTAGE BGND TESTForce sensor offset signal failure in background self-testBackground self-test detects failure in force sensor offset signalcritical
Depleted BatteryBattery fully discharged below low voltage thresholdBattery voltage drops below minimum operating thresholdcritical
DISPLAY CONTROLLER POSTLCD display failure detected in power-up self-testPower-up self-test fails in LCD display testingerror
Display Controller PostLCD display self-test failed during power-upPower-up self-test detects display communication or response failureerror
External COM Task TimeoutExternal device communication timeoutExternal computer sending too many commands or invalid commands too rapidlywarning
Firmware MismatchPump head and base firmware versions do not matchActive MCU and base ICB firmware versions or GUID are incompatibleerror
FIRMWARE MISMATCHFirmware version mismatch between pump head and baseActive software in pump head (MCU) and base ICB differ or invalid GUIDerror
FLASH MEMORY BGND TESTCRC failure in program memory during background self-testBackground self-tests detect CRC failure in program memorycritical
FLASH MEMORY POSTCRC failure in program memory during power-up self-testPower-up self-tests detect CRC failure in program memorycritical
FORCE SENSOR BGND TESTForce sensor signal out-of-range or bridge test signal not detectedBackground self-test finds sensor out-of-range or no change with bridge-test signalerror
Force Sensor Bgnd TestForce sensor self-test failed indicating malfunction or disturbance during startupForce sensor self-test failure during pump initializationerror
FORCE SENSOR BGND TESTSelf-test failed on force sensorForce sensor self-test failure during pump startuperror
Force Sensor Bridge TestForce sensor bridge circuit test failureSoftware sensed no voltage change when sensor test signal assertederror
FORCE SENSOR TESTForce sensor output unresponsive to DAC offset voltage changePower-up self-test finds sensor output unchanged when DAC offset changederror
Force Sensor TestForce sensor power-up test failureForce sensor output does not change when DAC offset voltage changederror
History Data CorruptedInfusion history log data lostSoftware unable to retrieve history in 6 consecutive attemptswarning
Improper ShutdownPump did not shut down through proper procedurePump powered on after improper power-down sequencewarning
IMPROPER SHUTDOWNAbnormal power-down detectedLast power-down not through proper shutdown procedureinfo
Infusion CompleteProgrammed infusion has been successfully deliveredInfusion reaches programmed volume or timeinfo
Internal A2D Failure10-bit A/D converter self-test failureSoftware self-test identifies A/D converter hardware failureerror
Invalid EntryProblem with Drug Program selected; pump cannot deliver itInvalid entry attempted on numerical entry screenwarning
INVALID INFUSION PARAMETER COMBINATIONInfusion parameters do not allow valid flow rate calculationUser entered invalid parameter combination or wrong syringe sizewarning
INVALID SYRINGE SIZESyringe size not recognized or barrel clamp lifted during deliveryUnrecognized syringe size for selected model or barrel clamp liftederror
KB DATA INTEGRITY FAILURENetwork settings corruptedNetwork configuration data validation failswarning
LIMITS ADJUSTED - CANNOT DELIVER ALL DOSESSoftware adjusted limits to prevent out-of-range ratesSoft limits and syringe parameters conflict with allowable rate limitswarning
Limits Adjusted - Cannot Deliver All DosesSoft limit values adjusted to keep rate within allowable limitsDrug Program parameters would result in rate exceeding limitswarning
Maintenance Is RecommendedAnnual Maintenance Testing is dueScheduled maintenance date reachedinfo
MCU POWER STROBE FAILUREPower strobe output signal from main processor failedSelf-test detects failure in power strobe output port signalerror
MONITOR TASK TIMEOUTSoftware timing failure in monitor taskSoftware timing failure in monitor task executionerror
Motor Failed to Stop ErrorMotor continues rotating after stop commandSoftware detects motor rotation after stop motor command executederror
MOTOR NOT RUNNINGMotor fails to operate during programmed infusionMotor does not start when infusion programmederror
Motor Not Running ErrorMotor sensor detects no rotationMotor sensor does not detect stepper motor rotationerror
MOTOR NOT RUNNING ERRORMotor rotation not detected during operationMotor sensor does not detect any rotation of stepper motorerror
Motor Rate ErrorMotor rotation rate out of specificationMotor sensor measures incorrect rotation rate for stepper motorerror
MOTOR RATE ERRORMotor running speed does not match programmed rateMotor speed differs from programmed delivery rateerror
Motor Supply Voltage BGND TestMotor supply voltage out of specification (should be 17V)Background self-test measures invalid range on motor supply voltageerror
MOTOR VOLUME LIMIT ERRORMotor exceeded step count for volume limitMotor executes more steps than calculated for set volume limiterror
NETWORK CONFIGURATION RESETNetwork configuration reset executedNetwork reset command executed in Biomed menuinfo
NON CRITICAL DATA CORRUPTEDNon-critical settings corruptedData corruption in non-volatile EEPROMinfo
Non Critical Data CorruptedOther settings stored in EEPROM were lostEEPROM data corruption detectedinfo
OCCLUSION - CHECK INFUSION LINEInfusion line occlusion detected; plunger force exceeds limitForce sensor detects excessive pressure from occluded lineerror
OSTMR TASK TIMEOUTOS Timer failed to respond during Watchdog eventOS Timer does not respond within specified time during Watchdog ticklecritical
PharmGuard Data CorruptedPharmGuard Safety Data buffer CRC errorCRC check failed when reading PharmGuard datainfo
Plunger Sensor FailurePlunger position sensor voltage out of rangePlunger position potentiometer reads voltage outside valid rangeerror
POSITIVE SUPPLY BGND TESTMain PCB power supply malfunction detectedPower supply outputs incorrect voltage (should be 9.472V ±5%)error
Positive Supply BGND TestPositive supply voltage out of specification (should be 9.472V ±5%)Power supply voltage measured outside valid range during background testerror
Primary Audible Alarm BGND TestPrimary speaker background self-test failureBackground self-test detects no current in primary speakerunknown
Primary Audible Alarm PostPrimary speaker power-up self-test failurePower-up self-test detects no current in primary speakerunknown
Primary Critical Data CorruptedPrimary configuration data corrupted; backup restoredCRC mismatch between computed and saved values during startupwarning
PUMP MOTOR DRIVE OFF POSTNo current in stepper motor during power-upPower-up self-test detects no current in stepper motorerror
Pump Motor Drive Off PostMotor not energized during power-up self-testPower-up self-test detects no current in stepper motorerror
PUMP MOTOR DRIVE PHASE A POSTStepper motor phase A winding open during power-upPower-up self-test finds stepper motor winding openerror
PUMP MOTOR DRIVE PHASE B POSTStepper motor phase B winding open during power-upPower-up self-test finds stepper motor winding openerror
RAM BGND TestRAM cyclic redundancy test failureSoftware test shows RAM failure during background self-testerror
RAM BGND TESTRAM failure on Main PCB during background testSoftware test shows failure of RAM on Main PCBcritical
SERIAL EEPROM TIMEOUTSerial EEPROM failure on Main PCBSoftware test shows failure of serial EEPROM on Main PCBcritical
SET TIME AND DATEReal time clock resetClock reset when battery depletedinfo
SPI BUS TIMEOUTSPI bus component failureSoftware test shows failure of component on SPI buscritical
SUPERCAP POSTPower-up self-test detected insufficient charge in super-capacitorSystem power-up self-testerror
SYRINGE FLANGE NOT IN PLACESyringe flange not properly inserted; flange clip detection failureBarrel clamp size sensor detects flange not inserted or sensor failureerror
SYRINGE FLANGE SENSOR FAILURESoftware detected syringe sensor voltage out of rangeSyringe sensor electrical test during operationerror
SYRINGE PLUNGER NOT IN PLACESyringe plunger not properly seated; plunger detection failureBarrel clamp size sensor detects plunger not in holder or sensor failureerror
System AdvisorySystem advisory alarm with specific condition descriptionSystem advisory condition detectedinfo
System FailureSystem failure with specific condition descriptionSystem failure condition detectedcritical
System FaultSystem fault condition detected; fluid delivery stops immediatelyOne of system fault conditions occurcritical
Task Stack UsageTask stack usage exceeds 90 percentSoftware monitoring detected stack usage above thresholdwarning
TIME BASE BGND TESTBackground self-test found system time base does not agree with time-of-day clockBackground diagnostic self-testwarning
TIME OF DAY CLOCK POSTPower-up self-test found system time base does not agree with time-of-day clockPower-up self-testwarning
TIME OF DAY CLOCK TIMEOUTTime of day clock failed to communicateSystem attempts to communicate with time-of-day clockerror
Watchdog Fail-SafeSystem error handler of last resort indicating power, electronics, hardware or software failureSystem enters inconsistent statecritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.