Smiths Medical Medfusion 4000: troubleshooting and error codes
The ServiceForge knowledge base holds 100 documented symptoms for the Smiths Medical Medfusion 4000, with 141 ranked causes, 111 diagnostic checks with test points and expected values, and 99 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Pump won't turn on
Pump fails to power on even after pressing power button
Ranked causes
- 1. Battery depleted (common)
- 2. Loose Main PCB to Interconnect PCB ribbon cable (occasional)
- 3. AC power connection issue (occasional)
- 4. Battery connections loose or defective (occasional)
- 5. Keypad cable loose or disconnected (occasional)
Verify AC power connection
- Test point
- AC outlet and pump inlet
- Expected
- Pump plugged into live AC outlet
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms do not sound
Alarms sound faintly or not at all during operation
Ranked causes
- 1. Alarm loudness configured to lowest level (common)
- 2. Speaker grid holes obstructed (occasional)
- 3. Speaker malfunction (occasional)
Verify alarm loudness setting
- Test point
- Profile Settings > View Settings > Alarm Loudness
- Expected
- Setting should not be at lowest level
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms do not sound or sound faintly
Alarm audio output is absent or at very low volume during alarm conditions
Ranked causes
- 1. Alarm loudness configured to lowest level (common)
- 2. Speaker grid holes obstructed (occasional)
- 3. Speaker malfunction (rare)
Verify alarm loudness setting
- Test point
- Profile Settings > View Settings > Alarm Loudness
- Expected
- Check current loudness level setting
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery does not hold charge
Battery charge capacity low or pump shuts down immediately when unplugged from AC power
Ranked causes
- 1. Battery ribbon cable disconnected (occasional)
- 2. Battery not charging adequately (occasional)
- 3. Battery pack defective (occasional)
Check battery ribbon cable connection
- Test point
- Battery door interior, Interconnect PCB connector
- Expected
- Battery ribbon cable firmly seated and fully connected
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Force sensor not responding to test signal
Force sensor does not change output voltage when test signal applied. Strain gauge may be unplugged or bridge open.
Ranked causes
- 1. Force sensor bridge unplugged or open connection (common)
- 2. Syringe loaded during power-up self-test (occasional)
- 3. Main PCB offset circuit, Plunger PCB, or cable failure (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motor does not rotate or stalls
Motor sensor detects no rotation of stepper motor. Pump cannot deliver medication.
Ranked causes
- 1. Motor sensor failure on Main PCB (common)
- 2. Worm/gear/leadscrew jamming or mechanical wear (occasional)
- 3. Drive train mechanical wear (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motor rotates at incorrect speed
Motor sensor measures incorrect rotation rate for stepper motor. Delivery rate inaccurate.
Ranked causes
- 1. Motor sensor failure on Main PCB (common)
- 2. Force sensor inoperative causing motor stall on occlusion (occasional)
- 3. Drive train mechanical wear (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Occlusion detected in infusion line patient safety
Force sensor sensed force exceeding occlusion threshold, indicating potential line blockage or tubing issue
Ranked causes
- 1. Actual occlusion in infusion line (common)
- 2. Infusion set tubing too small for rate or fluid viscosity (occasional)
- 3. Force sensor requires calibration (occasional)
Visually inspect infusion line for occlusion
- Expected
- No kinks, blockages, or disconnections visible
- Tools
- Visual inspection
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Plunger driver not moving correctly patient safety
Position potentiometer detects plunger driver motion anomaly during delivery, may be caused by slack in drive system, lever issues, clutch problems, or hardware failure
Ranked causes
- 1. Prime function not used before starting delivery (common)
- 2. Plunger levers moved or clutch not engaged (common)
- 3. Position potentiometer not calibrated or contaminated (occasional)
Verify plunger lever moves freely and captures syringe plunger
- Expected
- Lever moves without restriction and grips plunger
- Tools
- Visual/tactile inspection
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Backup audible alarm not functioning
Power-up self-testing found backup audible alarm not working at correct power or frequency. Alarm buzzer malfunction.
Ranked causes
- 1. Alarm buzzer failure on Main PCB (rare)
- 2. Main PCB ACU controller failure (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery fully depleted and pump non-functional
Battery voltage dropped below low voltage threshold. Pump stops delivery and watchdog generates alarm. May occur on battery power only.
Ranked causes
- 1. Battery fully discharged (common)
- 2. Chronic battery function issue (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not working message
Pump displays battery error message or battery status unavailable
Ranked causes
- 1. Battery ribbon cable disconnected (common)
- 2. Battery pack failure (occasional)
Check battery ribbon cable connection
- Test point
- Battery door interior, Interconnect PCB connector
- Expected
- Battery ribbon cable firmly seated and fully connected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery problem: Battery not working message
System displays battery fault or no battery detected error
Ranked causes
- 1. Battery ribbon cable disconnected or loose at Interconnect PCB (common)
- 2. Battery pack failure (occasional)
Inspect battery ribbon cable connection
- Test point
- Battery door, Interconnect PCB connector
- Expected
- Battery ribbon cable firmly seated and undamaged
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Main PCB replacement software installation
Load and verify correct software version on Main PCB after replacement; restore configuration from Interconnect PCB base board
Ranked causes
- 1. Main PCB replaced; software and configuration must be restored (common)
- 2. Configuration not loaded to Main PCB (common)
Verify software version after installation
- Test point
- Power-on screen
- Expected
- Software version matches expected pump software version
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Clock calendar not initialized
System date and time not set; clock defaults to incorrect date/time on startup or during operation
Ranked causes
- 1. Clock not initialized at startup (occasional)
- 2. Clock not initialized during operation (occasional)
Verify system date and time
- Test point
- Biomed > Utilities > View time and date
- Expected
- Correct date and time displayed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External device communication timeout
External computer or device connected to pump is sending too many commands rapidly or many invalid commands. Software cannot process requests within timeout window.
Ranked causes
- 1. External device command rate too high (common)
- 2. External device sending invalid commands (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Force sensor self-test failed patient safety
Self-test performed on force sensor failed, indicating sensor malfunction or plunger driver movement during startup
Ranked causes
- 1. Force sensor malfunction (common)
- 2. Plunger driver moved during pump startup (occasional)
Monitor force sensor operation using diagnostics
- Test point
- Biomed > Diagnostics > Monitor Analog Sensors
- Expected
- Force sensor responds to test signals
- Tools
- Biomed menu
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD display not responding or image corrupted
Display self-test failed. Display shows no image or incorrect display output.
Ranked causes
- 1. Display cable disconnected or loose (common)
- 2. LCD display hardware failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD has poor contrast
Liquid crystal display is difficult to read due to low or high contrast level
Ranked causes
- 1. LCD contrast requires adjustment (common)
- 2. LCD negative supply voltage out of range (rare)
Run LCD contrast adjustment
- Test point
- Biomed > Calibration > Adjust Contrast
- Expected
- Display contrast optimized and visible
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Liquid Crystal Display has poor contrast
LCD display has low contrast or is difficult to read
Ranked causes
- 1. LCD contrast setting too low (common)
- 2. LCD negative supply voltage issue (rare)
Adjust LCD contrast
- Test point
- Biomed > Calibration > Adjust Contrast
- Expected
- Contrast adjusted to acceptable level
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motor unit assembly and drive train installation caution
Reassembly of motor, worm, worm gear, and drive train after disassembly; includes lubrication and torque specifications.
Ranked causes
- 1. Motor unit disassembly requires reassembly with correct lubrication and torque (common)
- 2. Drive train assembly requires careful handling during installation (common)
Worm lubrication before motor installation
- Test point
- Worm hole
- Expected
- Filled with Dow 111 Silicone Grease Compound
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No charge light
Battery charge indicator light does not illuminate
Ranked causes
- 1. Pump not connected to AC power (common)
- 2. Battery charging issue (occasional)
Verify AC power connection
- Test point
- AC power outlet
- Expected
- Pump connected and AC indicator light on
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Occlusion alarms occur frequently
Pump generates occlusion alarms more often than expected during normal operation
Ranked causes
- 1. Occlusion alarm limit set too low (common)
- 2. High backpressure in infusion line (common)
Review occlusion alarm limit setting
- Test point
- Profile Settings or Biomed Configuration
- Expected
- Occlusion limit appropriate for infusion setup
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Primary speaker not producing audio alarm caution
Background or power-up self-test detects no current flowing in primary speaker. Alarm buzzer silent.
Ranked causes
- 1. Speaker disconnected or wire loose (common)
- 2. Speaker failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Program flash memory CRC failure
Flash memory contains program code with failed cyclic redundancy check. Program data corrupted.
Ranked causes
- 1. Flash memory corruption during software re-programming (occasional)
- 2. Flash memory chip failure on Main PCB (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump base not responding or shutdown failed
Pump base experienced failure during power down sequence. Base does not respond to commands.
Ranked causes
- 1. Power down sequence failure (occasional)
- 2. EEPROM read failure or disconnected Interconnect PCB (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Stepper motor not energized at startup
Power-up self-test senses no current in stepper motor drive circuit. Motor cannot operate.
Ranked causes
- 1. Pump operating on depleted battery power (common)
- 2. Motor supply or driver circuit failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Stepper motor phase A winding open circuit
Power-up self-test detected stepper motor phase A winding has opened connection. Motor cannot generate torque.
Ranked causes
- 1. Pump on depleted battery power during startup (common)
- 2. Motor winding disconnected or wire loose (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Stepper motor phase B winding open circuit
Power-up self-test detected stepper motor phase B winding has opened connection. Motor cannot generate torque.
Ranked causes
- 1. Pump on depleted battery power during startup (common)
- 2. Motor winding disconnected or wire loose (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Super-capacitor insufficient charge at startup
Power-up self-test detected super-capacitor voltage too low. Backup power supply insufficient.
Ranked causes
- 1. Super-capacitor low charge from new Main PCB or dead battery (common)
- 2. Super-capacitor hardware failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
A/D converter internal failure
Software self-test identified failure in 10-bit analog to digital converter. Sensor readings unreliable.
Ranked causes
- 1. A/D converter chip failure on Main PCB (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
A2D reference voltage out of range
Background self-test measured invalid range on analog-to-digital converter reference voltage. Voltage reading out of specification.
Ranked causes
- 1. A/D converter reference voltage regulator failure (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC power supply not functioning after installation high voltage
AC power indicator does not light after AC power supply replacement
Ranked causes
- 1. AC power supply not properly installed or defective (unknown)
Verify AC power indicator
- Test point
- AC power indicator LED
- Expected
- Indicator light illuminated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ACU power strobe not maintained
Self-test found ACU processor not maintaining its power strobe signal during test.
Ranked causes
- 1. ACU processor power strobe malfunction (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ACU watchdog alarm active
Self-test found ACU watchdog alarm signal is active. Processor may have failed to service watchdog within allotted time.
Ranked causes
- 1. Processor software failure or watchdog timeout (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Adjust display contrast voltage
Varies onscreen contrast of LCD display. Factory default is 00BF (may vary by LCD manufacturer). Adjustable higher or lower for comfortable viewing in different lighting conditions.
Ranked causes
- 1. Display contrast requires adjustment for optimal viewing (occasional)
Access contrast adjustment menu
- Test point
- Calibration menu
- Expected
- Adjust Contrast option selected
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Analog supply voltage invalid
Background self-test measured invalid range on analog supply voltage. Voltage reading out of specification.
Ranked causes
- 1. Analog supply voltage regulator failure (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Background CRC test timeout
Software CRC test failed to complete within allotted time. CRC calculation took too long.
Ranked causes
- 1. Software transient timing failure (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Background self-test timeout
Background self-test took longer than 8 seconds to complete, exceeding timeout threshold. Timing failure occurred.
Ranked causes
- 1. Software transient failure or timing issue (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Backlight connector polarity verification
Verify backlight wires are correctly connected to Main PCB; if reversed, backlight will not illuminate
Ranked causes
- 1. Backlight leads reversed during installation (occasional)
Verify backlight illuminates after power-up
- Test point
- LCD backlight
- Expected
- Backlight illuminated during power-up sequence
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not communicating or low
Battery gauge not communicating with pump, or very low battery signal detected while connected to AC power. Battery capacity shows as zero.
Ranked causes
- 1. Loose or disconnected battery connections (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery pack not charging caution
Battery charge indicator does not light when connected to AC power, or battery information absent from Monitor Battery Status screen
Ranked causes
- 1. Battery not properly connected to Interconnect PCB (common)
Verify battery charge indicator lights
- Test point
- Battery charge indicator LED
- Expected
- Indicator lights when AC power connected
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery pack replacement needed caution
Battery requires replacement or is not functioning after installation
Ranked causes
- 1. Exhausted or damaged battery pack (common)
Verify battery charge indicator lights after connecting pump to AC power
- Test point
- Battery charge indicator LED
- Expected
- Indicator lights
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery replacement needed - charge cycle limit reached patient safety
Battery charge cycle counter reaches or exceeds 400 cycles limit during Biomed diagnostics monitoring
Ranked causes
- 1. Battery at end-of-life charge cycle limit (common)
Access Biomed > Diagnostics > Monitor Battery Status and view Charge Cycles value
- Test point
- Charge Cycles field
- Expected
- Value below 400 cycles, maximum
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery replacement needed - low full capacity patient safety
Battery full-capacity indicator falls below minimum threshold during monitoring via Biomed diagnostics
Ranked causes
- 1. Battery aging or degradation reducing full capacity (common)
Access Biomed > Diagnostics > Monitor Battery Status and view Full-Capacity value
- Test point
- Full-Capacity field
- Expected
- Value above 3024 mAh for 22-4050-51 or above 4550 mAh for 22-4051-01, minimum threshold
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: AC power supply function verification high voltage
Verification that AC power supply is functioning correctly after replacement or service
Ranked causes
- 1. AC power supply replaced or serviced (unknown)
Verify AC power indicator lights when plugged in
- Test point
- AC power indicator LED
- Expected
- AC power indicator lights
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery function verification
Verify new battery is functioning nominally before returning pump to service
Ranked causes
- 1. New battery installation completed (common)
Connect pump to AC mains and observe battery charge indicator
- Test point
- Battery charge indicator light
- Expected
- Indicator lights on
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Bottom housing replacement caution
Installation and verification of replacement bottom housing assembly with all subcomponents
Ranked causes
- 1. Bottom housing replacement required (unknown)
Verify soft silicone gasket around lower housing lip is continuous and intact
- Test point
- Lower housing lip perimeter
- Expected
- Gasket continuous and intact with no damage
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Clutch or clutch cam replacement
After replacing clutch halves or cam, recalibrate and verify pump functionality per guidelines before returning to use.
Ranked causes
- 1. Clutch halves or cam replaced (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Display contrast adjustment
Procedure to adjust LCD display contrast within Biomed Calibration menu
Ranked causes
- 1. Display contrast requires adjustment for visibility (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Force sensor
Adjustment of force sensor settings. Sensor measures plunger force to ensure proper operation and safety limits.
Ranked causes
- 1. Initial force sensor calibration or recalibration required (common)
Access force sensor calibration
- Test point
- Calibration menu
- Expected
- Cal Force Sensor option selected
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Interconnect PCB reinstallation and verification
Procedure to reinstall Interconnect PCB with antenna, radio module, and clear shield; verify serial number and pump functionality before return to use.
Ranked causes
- 1. PCB reinstallation and verification procedure (common)
Verify serial number is correct following installation
- Test point
- pump startup screen and bottom serial number label
- Expected
- Serial number on startup screen matches label on pump bottom
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Leadscrew replacement
After replacing leadscrew, recalibrate and verify pump functionality per guidelines before returning to use.
Ranked causes
- 1. Leadscrew replaced (common)
Verify leadscrew Nylock nut torque
- Test point
- Nylock nut holding leadscrew on right end plate
- Expected
- 60 in.-oz (0.42 Nm) then loosen 1/4 turn, loosen 1/4 turn after initial torque
- Tools
- Torque wrench, 1/4 inch wrench
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Main PCB replacement and pump function verification
After replacing Main PCB, pump signals Configuration Required alarm and requires full recalibration, reconfiguration, and function verification before return to service.
Ranked causes
- 1. Main PCB replaced without configuration (common)
Verify Configuration correctly installed on Main PCB
- Test point
- Pump display after configuration load
- Expected
- Configuration matches facility-validated version
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Motor unit replacement
After replacing motor, worm, or worm gear, recalibrate and verify pump functionality per guidelines before returning to use.
Ranked causes
- 1. Motor, worm, or worm gear replaced (common)
Verify motor mounting screws are torqued correctly
- Test point
- 2 pan head screws holding motor
- Expected
- 100 in oz (0.71 Nm), as specified
- Tools
- Torque screwdriver with Phillips head
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Plunger position
Adjustment of plunger position sensor settings to establish empty position and plunger head travel limits.
Ranked causes
- 1. Initial plunger position calibration or recalibration required (common)
Access plunger position calibration menu
- Test point
- Calibration menu
- Expected
- Cal Plunger Position option selected
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Plunger position sensor
Calibration procedure to check or adjust plunger position sensor values within Biomed Calibration menu
Ranked causes
- 1. Plunger position sensor requires recalibration after maintenance or drift (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Position potentiometer replacement
After replacing position potentiometer, recalibrate and verify pump functionality per guidelines before returning to use.
Ranked causes
- 1. Position potentiometer replaced (common)
Verify position potentiometer mounting screw torque
- Test point
- Small flat head screw at right end plate
- Expected
- 16 in.-oz. (0.11 Nm), as specified
- Tools
- Torque screwdriver
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Size and position sensor
Calibration procedure to check or adjust syringe size and position sensor values within Biomed Calibration menu
Ranked causes
- 1. Sensor requires recalibration after maintenance or drift detected (unknown)
Select Calibration menu in Biomed
- Test point
- Biomed > Calibration
- Expected
- Calibration submenu options displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Syringe size and position
Concurrent adjustment of syringe size sensor (barrel clamp assembly, measures external diameter) and syringe position sensor (drive train assembly, measures plunger head position for empty position and travel determination).
Ranked causes
- 1. Initial calibration or recalibration required (common)
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Calibration: Syringe size sensor
- Clutch assembly replacement
- Configuration settings lost or invalid
- Critical data block RAM corrupted
- Firmware version mismatch between pump head and base
- Force sensor signal out of range
- Force sensor voltage invalid
- Indicator lights do not flash
- Key click or beep is faint or not present
- Keypad key does not work
- Keypad stuck or unresponsive
- Maximum rate not available
- MCU power strobe signal failure
- MCU time base synchronization failure
- Monitor software task timing failure
- Motor current not disabled during watchdog test
- Motor exceeded volume delivery limit
- Motor rotation sensor failure after stop command
- Motor supply voltage out of specification
- Network settings corrupted or missing
- No Profiles available
- OS timer failed during watchdog event
- PCB firmware recovery error: Firmware mismatch
- Plunger position sensor voltage out of range
- Positive supply voltage out of specification
- Pump fails to initialize sensors
- Pump option or feature not available
- Pump powers on with blank screen and buzzer sounds and will not power off after firmware upgrade
- Radio PCB removal
- Real time clock reset
- Serial EEPROM communication timeout
- Speaker does not produce power-on tones
- Speaker not functioning after installation
- SPI bus communication failure
- Syringe barrel clamp sensor difficulty
- Syringe flange sensor active
- Syringe flange sensor voltage out of range
- Syringe Manufacturer not available
- Syringe plunger sensor active
- System RAM memory corrupted
40 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
A2D REFERENCE VOLTAGE BGND TEST | ADC reference voltage invalid | Background self-test detects invalid voltage range | critical |
ACU POWER STROBE FAILURE | ACU processor power strobe not maintained | ACU processor fails to maintain power strobe signal | critical |
ACU WATCHDOG FAILURE | ACU watchdog alarm active | Processor failed to pet watchdog within allotted time | critical |
ANALOG SUPPLY BGND TEST | Analog supply voltage invalid | Background self-test detects invalid voltage range | critical |
AUX CONTROLLER UNIT POST | ACU POST motor current shutdown failure | ACU failed to shut down motor current during watchdog test | critical |
BACKGROUND CRC TEST TIMEOUT | Background CRC test timeout | Software timing failure | warning |
BACKGROUND SELF TEST TIMEOUT | Background self-test exceeds time limit | Background test takes more than 8 seconds | warning |
BACKUP AUDIBLE ALARM POST | Backup alarm power or frequency invalid | POST detects incorrect alarm power/frequency | critical |
Backup Critical Data Corrupted | Backup configuration data was lost; primary data restored backup | EEPROM redundancy check failed during initialization | warning |
BASE HARDWARE FAILURE | Base hardware self-test failed | Non-zero hardware test result or interface initialization failure | critical |
BASE NOT RESPONDING | Base power-down sequence failure | Failure during power down sequence | critical |
BASE UNKNOWN FAILURE | Read EEPROM failure | EEPROM read operation fails | critical |
BATTERY COMMUNICATION TIMEOUT | Battery gauge not communicating | Battery current detected but no gauge communication | warning |
BATTERY NOT WORKING | Battery malfunction | Very low battery signal detected while on AC power | warning |
CALCULATED RATE OUT OF RANGE | Pump calculated an invalid delivery rate based on entered values | Invalid combination of syringe parameters and desired rate | error |
CALIBRATION REQUIRED | Sensor calibration failed | Calibration data test fails during power-up or save | warning |
CHECK CLUTCH / PLUNGER LEVER | Plunger driver position not moving correctly during delivery | Position potentiometer detects abnormal plunger driver motion | error |
CHECK SYRINGE BARREL CLAMP | Syringe barrel clamp sensor verification failure | Certain syringe models cause barrel clamp sensor difficulty; clamp lifted during power-up | warning |
CHECK SYRINGE FLANGE SENSOR | Syringe not loaded correctly; flange sensor active | Syringe flange size sensor detects improper loading | error |
Check Syringe Plunger Sensor | Syringe plunger sensor indicates incorrect loading or movement during startup | Syringe not loaded correctly or plunger moved during pump startup | error |
CHECK SYRINGE PLUNGER SENSOR | Syringe not loaded correctly; plunger sensor active | Syringe not detected as properly loaded; plunger driver moved during startup | error |
Configuration Required | No Configuration file present on Main PCB | Pump powered up after Main PCB replacement without existing configuration | warning |
Configuration Required | Configuration file not present on Main PCB | Main PCB replaced or configuration lost; pump powered up without valid configuration | warning |
CONTROL KEY SWITCH BGND TEST | Keypad key stuck on | Key found stuck during background self-test | critical |
CONTROL KEY SWITCH POST | Keypad key pressed too long | Single key pressed >7sec or locked key >1min | critical |
CRITICAL DATA BLOCK BGND TEST | Critical delivery data corrupted | Background self-test shows data block corruption | critical |
CRITICAL DATA BLOCK POST | Critical configuration data CRC mismatch | Backup and primary data fail CRC check | critical |
CRITICAL DATA FAILURE | Incompatible pump settings or software data failure | Incompatible setting ranges or critical flags found invalid | critical |
Critical Data Failure | Programmed settings failed reverse-calculation validation | After settings programmed, pump verification failed or incompatible ranges detected | error |
D2A OFFSET VOLTAGE BGND TEST | Force sensor offset signal failure in background self-test | Background self-test detects failure in force sensor offset signal | critical |
Depleted Battery | Battery fully discharged below low voltage threshold | Battery voltage drops below minimum operating threshold | critical |
DISPLAY CONTROLLER POST | LCD display failure detected in power-up self-test | Power-up self-test fails in LCD display testing | error |
Display Controller Post | LCD display self-test failed during power-up | Power-up self-test detects display communication or response failure | error |
External COM Task Timeout | External device communication timeout | External computer sending too many commands or invalid commands too rapidly | warning |
Firmware Mismatch | Pump head and base firmware versions do not match | Active MCU and base ICB firmware versions or GUID are incompatible | error |
FIRMWARE MISMATCH | Firmware version mismatch between pump head and base | Active software in pump head (MCU) and base ICB differ or invalid GUID | error |
FLASH MEMORY BGND TEST | CRC failure in program memory during background self-test | Background self-tests detect CRC failure in program memory | critical |
FLASH MEMORY POST | CRC failure in program memory during power-up self-test | Power-up self-tests detect CRC failure in program memory | critical |
FORCE SENSOR BGND TEST | Force sensor signal out-of-range or bridge test signal not detected | Background self-test finds sensor out-of-range or no change with bridge-test signal | error |
Force Sensor Bgnd Test | Force sensor self-test failed indicating malfunction or disturbance during startup | Force sensor self-test failure during pump initialization | error |
FORCE SENSOR BGND TEST | Self-test failed on force sensor | Force sensor self-test failure during pump startup | error |
Force Sensor Bridge Test | Force sensor bridge circuit test failure | Software sensed no voltage change when sensor test signal asserted | error |
FORCE SENSOR TEST | Force sensor output unresponsive to DAC offset voltage change | Power-up self-test finds sensor output unchanged when DAC offset changed | error |
Force Sensor Test | Force sensor power-up test failure | Force sensor output does not change when DAC offset voltage changed | error |
History Data Corrupted | Infusion history log data lost | Software unable to retrieve history in 6 consecutive attempts | warning |
Improper Shutdown | Pump did not shut down through proper procedure | Pump powered on after improper power-down sequence | warning |
IMPROPER SHUTDOWN | Abnormal power-down detected | Last power-down not through proper shutdown procedure | info |
Infusion Complete | Programmed infusion has been successfully delivered | Infusion reaches programmed volume or time | info |
Internal A2D Failure | 10-bit A/D converter self-test failure | Software self-test identifies A/D converter hardware failure | error |
Invalid Entry | Problem with Drug Program selected; pump cannot deliver it | Invalid entry attempted on numerical entry screen | warning |
INVALID INFUSION PARAMETER COMBINATION | Infusion parameters do not allow valid flow rate calculation | User entered invalid parameter combination or wrong syringe size | warning |
INVALID SYRINGE SIZE | Syringe size not recognized or barrel clamp lifted during delivery | Unrecognized syringe size for selected model or barrel clamp lifted | error |
KB DATA INTEGRITY FAILURE | Network settings corrupted | Network configuration data validation fails | warning |
LIMITS ADJUSTED - CANNOT DELIVER ALL DOSES | Software adjusted limits to prevent out-of-range rates | Soft limits and syringe parameters conflict with allowable rate limits | warning |
Limits Adjusted - Cannot Deliver All Doses | Soft limit values adjusted to keep rate within allowable limits | Drug Program parameters would result in rate exceeding limits | warning |
Maintenance Is Recommended | Annual Maintenance Testing is due | Scheduled maintenance date reached | info |
MCU POWER STROBE FAILURE | Power strobe output signal from main processor failed | Self-test detects failure in power strobe output port signal | error |
MONITOR TASK TIMEOUT | Software timing failure in monitor task | Software timing failure in monitor task execution | error |
Motor Failed to Stop Error | Motor continues rotating after stop command | Software detects motor rotation after stop motor command executed | error |
MOTOR NOT RUNNING | Motor fails to operate during programmed infusion | Motor does not start when infusion programmed | error |
Motor Not Running Error | Motor sensor detects no rotation | Motor sensor does not detect stepper motor rotation | error |
MOTOR NOT RUNNING ERROR | Motor rotation not detected during operation | Motor sensor does not detect any rotation of stepper motor | error |
Motor Rate Error | Motor rotation rate out of specification | Motor sensor measures incorrect rotation rate for stepper motor | error |
MOTOR RATE ERROR | Motor running speed does not match programmed rate | Motor speed differs from programmed delivery rate | error |
Motor Supply Voltage BGND Test | Motor supply voltage out of specification (should be 17V) | Background self-test measures invalid range on motor supply voltage | error |
MOTOR VOLUME LIMIT ERROR | Motor exceeded step count for volume limit | Motor executes more steps than calculated for set volume limit | error |
NETWORK CONFIGURATION RESET | Network configuration reset executed | Network reset command executed in Biomed menu | info |
NON CRITICAL DATA CORRUPTED | Non-critical settings corrupted | Data corruption in non-volatile EEPROM | info |
Non Critical Data Corrupted | Other settings stored in EEPROM were lost | EEPROM data corruption detected | info |
OCCLUSION - CHECK INFUSION LINE | Infusion line occlusion detected; plunger force exceeds limit | Force sensor detects excessive pressure from occluded line | error |
OSTMR TASK TIMEOUT | OS Timer failed to respond during Watchdog event | OS Timer does not respond within specified time during Watchdog tickle | critical |
PharmGuard Data Corrupted | PharmGuard Safety Data buffer CRC error | CRC check failed when reading PharmGuard data | info |
Plunger Sensor Failure | Plunger position sensor voltage out of range | Plunger position potentiometer reads voltage outside valid range | error |
POSITIVE SUPPLY BGND TEST | Main PCB power supply malfunction detected | Power supply outputs incorrect voltage (should be 9.472V ±5%) | error |
Positive Supply BGND Test | Positive supply voltage out of specification (should be 9.472V ±5%) | Power supply voltage measured outside valid range during background test | error |
Primary Audible Alarm BGND Test | Primary speaker background self-test failure | Background self-test detects no current in primary speaker | unknown |
Primary Audible Alarm Post | Primary speaker power-up self-test failure | Power-up self-test detects no current in primary speaker | unknown |
Primary Critical Data Corrupted | Primary configuration data corrupted; backup restored | CRC mismatch between computed and saved values during startup | warning |
PUMP MOTOR DRIVE OFF POST | No current in stepper motor during power-up | Power-up self-test detects no current in stepper motor | error |
Pump Motor Drive Off Post | Motor not energized during power-up self-test | Power-up self-test detects no current in stepper motor | error |
PUMP MOTOR DRIVE PHASE A POST | Stepper motor phase A winding open during power-up | Power-up self-test finds stepper motor winding open | error |
PUMP MOTOR DRIVE PHASE B POST | Stepper motor phase B winding open during power-up | Power-up self-test finds stepper motor winding open | error |
RAM BGND Test | RAM cyclic redundancy test failure | Software test shows RAM failure during background self-test | error |
RAM BGND TEST | RAM failure on Main PCB during background test | Software test shows failure of RAM on Main PCB | critical |
SERIAL EEPROM TIMEOUT | Serial EEPROM failure on Main PCB | Software test shows failure of serial EEPROM on Main PCB | critical |
SET TIME AND DATE | Real time clock reset | Clock reset when battery depleted | info |
SPI BUS TIMEOUT | SPI bus component failure | Software test shows failure of component on SPI bus | critical |
SUPERCAP POST | Power-up self-test detected insufficient charge in super-capacitor | System power-up self-test | error |
SYRINGE FLANGE NOT IN PLACE | Syringe flange not properly inserted; flange clip detection failure | Barrel clamp size sensor detects flange not inserted or sensor failure | error |
SYRINGE FLANGE SENSOR FAILURE | Software detected syringe sensor voltage out of range | Syringe sensor electrical test during operation | error |
SYRINGE PLUNGER NOT IN PLACE | Syringe plunger not properly seated; plunger detection failure | Barrel clamp size sensor detects plunger not in holder or sensor failure | error |
System Advisory | System advisory alarm with specific condition description | System advisory condition detected | info |
System Failure | System failure with specific condition description | System failure condition detected | critical |
System Fault | System fault condition detected; fluid delivery stops immediately | One of system fault conditions occur | critical |
Task Stack Usage | Task stack usage exceeds 90 percent | Software monitoring detected stack usage above threshold | warning |
TIME BASE BGND TEST | Background self-test found system time base does not agree with time-of-day clock | Background diagnostic self-test | warning |
TIME OF DAY CLOCK POST | Power-up self-test found system time base does not agree with time-of-day clock | Power-up self-test | warning |
TIME OF DAY CLOCK TIMEOUT | Time of day clock failed to communicate | System attempts to communicate with time-of-day clock | error |
Watchdog Fail-Safe | System error handler of last resort indicating power, electronics, hardware or software failure | System enters inconsistent state | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI