Smiths Medical Medfusion 3000: troubleshooting and error codes

The ServiceForge knowledge base holds 74 documented symptoms for the Smiths Medical Medfusion 3000, with 108 ranked causes, 152 diagnostic checks with test points and expected values, and 139 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Pump does not power on

Pump fails to turn on or respond to power button

Ranked causes

  1. 1. Dead battery (common)
  2. 2. Main board to interconnect cable disconnected (occasional)
  3. 3. AC power connection issue (occasional)
  4. 4. Battery connection issue (occasional)
  5. 5. Keypad cable disconnected (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump will not turn on or power up

Pump does not respond to power button; no display, no LED indicators, no startup sequence

Ranked causes

  1. 1. Dead or fully discharged battery; must charge before power-on (common)
  2. 2. Main PCB to Interconnect PCB ribbon cable loose or disconnected (occasional)
  3. 3. AC power supply connections loose or missing (occasional)
  4. 4. Battery connections loose or disconnected (occasional)
  5. 5. Keypad cable loose or disconnected (occasional)
First check

Connect AC power and charge battery

Test point
AC power connection; battery charge status (if readable)
Expected
AC power connected; allow 2+ minutes charge time

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery does not hold charge

Battery drains quickly when pump disconnected from AC power; pump shuts down during operation or cannot sustain operation on battery alone

Ranked causes

  1. 1. Battery ribbon cable disconnected or loose (common)
  2. 2. Battery charging circuit failure (occasional)
  3. 3. Battery chemistry degraded or cell failure (occasional)
  4. 4. Battery calibration required (occasional)
First check

Inspect and reseat battery ribbon cable

Test point
Battery ribbon cable connector on Interconnect PCB
Expected
Connector fully seated, no looseness

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Learn Error - Remote Not in Teach Mode

Learning pump displays error message 'Learn Error - Remote Not in Teach mode' when attempting configuration cloning from Teacher pump. Communication fails during infrared data transfer.

Ranked causes

  1. 1. Teacher pump not in Teach Mode or infrared ports misaligned (common)
  2. 2. Teacher and Learning pumps have different software versions (occasional)
  3. 3. Learning pump not properly configured in Learn Mode (occasional)
  4. 4. Pump hardware malfunction or damaged Interconnect PCB (rare)
First check

Verify Teacher pump in Teach Mode

Test point
Profile Settings menu
Expected
Teach Mode selected and active

2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms do not sound or sound faintly patient safety

Speaker produces no sound or very low volume for alarms; audio may be inaudible in clinical environment

Ranked causes

  1. 1. Alarm volume configured to lowest level (common)
  2. 2. Speaker grid holes obstructed (occasional)
  3. 3. Speaker failure (rare)
First check

Review alarm volume setting (V3 pumps)

Test point
Custom Program audio settings menu
Expected
Alarm volume set to medium or high level

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD has poor contrast or is unreadable

Display text is dim, hard to read, or has low contrast; visibility impaired regardless of ambient lighting

Ranked causes

  1. 1. LCD contrast setting at minimum or too low (common)
  2. 2. LCD negative supply voltage fault (rare)
  3. 3. LCD or Main PCB hardware failure (rare)
First check

Run Adjust Contrast utility

Test point
Biomed > Calibration > Adjust Contrast menu
Expected
Contrast adjustment slider available and responsive; display improves with slider adjustment

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No charge light illuminated

Battery charge indicator LED does not light when pump connected to AC power; may indicate battery not charging or power supply problem

Ranked causes

  1. 1. AC power not connected or pump not powered on (common)
  2. 2. Both AC and charge indicators off; indicator light or display issue (occasional)
  3. 3. Battery not charging or charger failure (occasional)
First check

Verify AC power connection

Test point
AC power inlet on pump rear
Expected
AC cord plugged into power outlet; AC indicator LED lit

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Occlusion alarms occur too frequently patient safety

Pump triggers occlusion alarm repeatedly during normal infusion; false alarms without actual line blockage

Ranked causes

  1. 1. Occlusion alarm limit set too low in Configuration (common)
  2. 2. High infusion line backpressure due to setup or medication (common)
  3. 3. Force sensor miscalibration or drift (occasional)
First check

Review occlusion alarm limit setting

Test point
Pump Configuration occlusion limit value
Expected
Occlusion limit set to appropriate level for line setup and medication

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms do not sound audibly caution

Alarms either do not sound or sound very faintly

Ranked causes

  1. 1. Alarm volume configured to lowest level (common)
  2. 2. Speaker failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery not working message

Pump displays battery failure message or no battery response

Ranked causes

  1. 1. Battery ribbon cable disconnected (common)
  2. 2. Battery pack failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery not working message displayed

Pump displays 'Battery Problem: Battery not working' message; battery pack is not communicating with pump electronics

Ranked causes

  1. 1. Battery ribbon cable disconnected or loose (common)
  2. 2. Battery pack defective (common)
First check

Inspect battery ribbon cable connection

Test point
Battery ribbon cable connector on Interconnect PCB (inside case bottom)
Expected
Connector fully seated and locked

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AC line leakage (Medfusion 3500BC) high voltage

Verify enclosure and earth leakage current on 3500BC model with functional earth ground

Ranked causes

  1. 1. Enclosure leakage exceeds 100 micro-amps (rare)
  2. 2. Earth leakage exceeds 100 micro-amps through AC power cord earth conductor (rare)
First check

Connect pump to Safety Analyzer

Test point
AC power inlet
Expected
Analyzer configured in Line Leakage mode

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Syringe size sensor

Load calibration slugs into barrel clamp and adjust position to find lowest size reading for both small and large standards, recording count values.

Ranked causes

  1. 1. Syringe size sensor requires calibration (common)
  2. 2. Large standard calibration needed (common)
First check

Verify small standard size reading

Test point
Barrel clamp with small slug
Expected
Lowest size reading achieved when clamp perpendicular and adjusted back/forth

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

General inspection caution

Annual maintenance visual and physical inspection of pump components, power cord, labels, and housing

Ranked causes

  1. 1. Physical damage or wear on pump components (occasional)
  2. 2. Loose internal or external parts detected during inspection (occasional)
First check

Inspect for physical damage to housing and keypads

Test point
Pump housing and keypad surfaces
Expected
No cracked housings or torn keypads

9 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Indicator lights do not flash

Status indicator lights do not illuminate during startup or function

Ranked causes

  1. 1. Keypad connection loose (common)
  2. 2. Keypad failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Indicator lights do not flash during startup or with alarms

LED indicator lights on pump do not illuminate during power-up or in response to alarms and function activation; display may be dark or unresponsive

Ranked causes

  1. 1. Keypad cable loose or disconnected from Main PCB (common)
  2. 2. Keypad or LED circuit failure (rare)
First check

Inspect keypad cable connection to Main PCB

Test point
Keypad ribbon cable connector on Main PCB
Expected
Connector fully seated and locked; no movement or looseness

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad key does not respond to presses

One or more keypad buttons do not register when pressed; no response or stuck key condition

Ranked causes

  1. 1. Keypad connection loose or disconnected (common)
  2. 2. Keypad hardware failure (rare)
First check

Run Keypad Test in Biomed diagnostics

Test point
Biomed > Diagnostics > Keypad Test menu
Expected
All keys pass diagnostic test; no failed keys reported

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD display poor contrast

Liquid Crystal Display text or graphics not clearly visible

Ranked causes

  1. 1. LCD contrast setting too low (common)
  2. 2. LCD negative supply or display failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Maximum infusion rate not available or set too low patient safety

Pump maximum delivery rate setting is lower than desired; maximum rate option in menu unavailable or locked

Ranked causes

  1. 1. Maximum rate configured low in pump Configuration (common)
  2. 2. Individual profile has lower maximum rate than global setting (occasional)
First check

Review global Max Delivery Rate setting

Test point
Profile Settings > View Settings > Max Delivery Rate
Expected
Max Delivery Rate set to desired value or higher

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No charge light indicator

Charge indicator light does not illuminate when battery charging or AC powered

Ranked causes

  1. 1. AC power not connected (common)
  2. 2. Battery charging issue (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

AC power indicator test high voltage

Verify AC power indicator LED lights when pump plugged into AC source and extinguishes when unplugged

Ranked causes

  1. 1. AC power indicator LED malfunction or power supply issue (rare)
First check

Verify AC indicator LED lights when connected to AC power

Test point
AC indicator LED on pump
Expected
AC indicator LED is lit

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Adjust LCD contrast voltage

LCD display contrast is too light or too dark. Factory default is 00BF but may vary by display manufacturer.

Ranked causes

  1. 1. LCD contrast voltage requires adjustment (occasional)
First check

Set contrast voltage

Test point
CONTRAST VOLTAGE
Expected
00BF (or adjusted value per lighting conditions)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Annual maintenance: General inspection caution

Visual and functional inspection of pump housing, mechanical components, power cord, labels, and rubber feet for damage, wear, or loose parts.

Ranked causes

  1. 1. Physical damage or normal wear (unknown)
First check

Inspect for obvious physical damage

Test point
pump housing, keypads, flippers, clamps
Expected
no cracked housings, torn keypads, or visible damage

8 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Annual maintenance: Power-up test caution

Functional verification of alarm sounds, LED indicators, battery capacity, self-tests, display legibility, and AC/battery switching upon power-on.

Ranked causes

  1. 1. Power supply or display malfunction (unknown)
First check

Plug AC power cord and verify AC/battery indicators light

Test point
AC power receptacle, AC indicator LED, battery charge indicator LED
Expected
both AC and battery indicators are lit

7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery calibration required

CPI (Capacity Inaccurate) counter exceeds threshold; pump requires recalibration cycle

Ranked causes

  1. 1. Shallow discharge cycles accumulated since last calibration (common)
First check

Monitor CPI value in Biomed > Diagnostics > Monitor Battery Status

Test point
Biomed > Diagnostics > Monitor Battery Status
Expected
CPI <= 80 Hex

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery depleted, pump loses calibration patient safety

Battery Depleted alarm occurs; pump loses all calibration information; approximately 5 minutes of operation remain before full disconnect; audible alarm continues from super-capacitor

Ranked causes

  1. 1. Battery capacity below 1600 mA-hours (common)
First check

Recharge pump and monitor battery status

Test point
Biomed > Diagnostics > Monitor Battery Status
Expected
After several minutes of recharging, gauge shows learned capacity; LMD value greater than 1600 mA-hours

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AC input assembly replacement high voltage

Steps to remove and install AC input module assembly with verification of AC power indicator function.

Ranked causes

  1. 1. AC Input Module Assembly failure or replacement required (unknown)
First check

Verify AC power indicator lights after replacement

Test point
AC power indicator light
Expected
AC power indicator lights on

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AC line leakage (Standard Medfusion 3000) high voltage

Verify enclosure leakage current and optional CF-rated patient contact leakage are within specification

Ranked causes

  1. 1. Leakage current exceeds Class II BF limit of 100 micro-amps (rare)
First check

Connect pump to Safety Analyzer

Test point
AC power inlet
Expected
Analyzer configured in Line Leakage mode

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AC line leakage (standard) high voltage

Verify electrical leakage current within acceptable limits for patient safety (BF or CF rating compliance)

Ranked causes

  1. 1. AC leakage current out of specification (unknown)
First check

Measure enclosure leakage current (BF rating)

Test point
Plunger driver tube or center post of DC input jack
Expected
Less than 100 micro-amps in normal setting, <100
Tools
Safety Analyzer in Line Leakage mode

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AC power supply replacement high voltage

Steps to remove and install AC power supply PCB with verification of AC power indicator light function.

Ranked causes

  1. 1. AC Power Supply PCB failure or replacement required (unknown)
First check

Verify AC power indicator lights after replacement

Test point
AC power indicator light
Expected
AC power indicator lights on

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Adjust contrast (voltage)

Adjust LCD display contrast voltage setting using up/down keys. Factory default is 00BF; adjust higher or lower based on lighting conditions.

Ranked causes

  1. 1. Display contrast requires adjustment for lighting conditions (occasional)
First check

Adjust contrast voltage setting

Test point
LCD display
Expected
Visual clarity appropriate for ambient lighting conditions

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: All sensors

Recalibration required after replacing case top to ensure proper sensor operation and pump accuracy.

Ranked causes

  1. 1. Case top replacement affects sensor alignment (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Battery charge and performance status

Monitor battery NiMH/NiCad cell capacity, charge level gauge accuracy, and shallow discharge cycle count to maintain performance

Ranked causes

  1. 1. Battery gauge accuracy degraded or capacity low (occasional)
First check

Monitor battery LMD (Last Measured Discharge) capacity

Test point
Biomed > Diagnostics > Monitor Battery Status menu
Expected
LMD ≥ 1600 mA-hours, ≥1600
Tools
Pump Biomed diagnostics menu

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Battery full discharge-recharge cycle

Execute complete battery discharge to depleted alarm, then recharge to recalibrate gauge and relearn true battery capacity

Ranked causes

  1. 1. Battery calibration required (CPI threshold exceeded or new battery installed) (occasional)
First check

Verify pump connected to AC and charge level reaches 99-100%

Test point
Biomed > Diagnostics > Monitor Battery Status gauge
Expected
Charge Level 99-100%

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Battery monitor accuracy

Battery gauge recalibration to ensure correct capacity reading and low battery alarm accuracy

Ranked causes

  1. 1. CPI (Capacity Inaccurate) exceeds threshold after shallow discharge cycles (common)
First check

Connect pump to AC power

Test point
AC power inlet
Expected
Pump powered from AC mains

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Bottom housing replacement

Steps to remove and install bottom housing assembly with serial number label configuration based on installed Interconnect PCB model.

Ranked causes

  1. 1. Bottom housing damage or replacement required (unknown)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Flow delivery accuracy

Verify syringe plunger travel distance matches expected range for infusion rate accuracy at 300 ml/hr with 50 ml volume limit

Ranked causes

  1. 1. Flow calibration out of specification (unknown)
First check

Measure syringe plunger travel distance for flow accuracy

Test point
Syringe flange to plunger distance
Expected
B-D 60cc: 3.473" to 3.615"; B Braun Perfusor/Omnifix 50cc: 3.146" to 3.275"; Fresenius 50cc: 2.904" to 3.023"; Monoject 60cc: 3.473" to 3.615"; Terumo 60cc: 2.901" to 3.020" inches, ±0.071" (B-D), ±0.065" (B Braun), ±0.060" (Fresenius), ±0.07
Tools
Calipers, 50/60 cc Luer-lock syringe, Safety Analyzer

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Force sensor verification caution

Verify force sensor readings are within acceptable range using calibration kit with force gauge; check at no load, 5 pounds, and 15 pounds applied force.

Ranked causes

  1. 1. Force sensor calibration drift or error (unknown)
First check

Verify force reading with no syringe loaded

Test point
screen force display
Expected
between –0.7 and +0.7 pounds, ±0.7 pounds

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Main PCB replacement

Steps to remove and install Main PCB with reconnection of all cable and sensor connections. Requires serial number and Configuration version for P.A.S.S. pumps.

Ranked causes

  1. 1. Main PCB failure or replacement required (unknown)
First check

Reconnect all cable connections on Main PCB

Test point
J10, J3, J6, J12, J2, J4, J11 connectors
Expected
All cables connected and secured per procedure

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Plunger position sensor

Plunger position sensor drift or misalignment. Requires calibration of position sensor reference points.

Ranked causes

  1. 1. Plunger position sensor requires recalibration (common)
First check

Verify plunger position sensor minimum and maximum values

Test point
PLUNGER POS DATA
Expected
151 to 901 COUNTS, to 901 COUNTS

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Plunger position sensor verification

Annual maintenance check to verify plunger position sensor readings are within specification using small and large calibration slugs

Ranked causes

  1. 1. Plunger position sensor out of calibration (unknown)
First check

Verify small calibration slug position reading with ~0.1 lbs force

Test point
Position reading on screen
Expected
between 0.680" and 0.720", inclusive inches, ±0.02

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Position and size

Sensor recalibration needed after case handle replacement to maintain accurate syringe position and size detection.

Ranked causes

  1. 1. Case handle replacement affects position/size sensor calibration (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pressure sensor

Pressure sensor calibration or offset adjustment after repair or drift detection.

Ranked causes

  1. 1. Pressure sensor requires recalibration (common)
First check

Verify pressure sensor values

Test point
PRESSURE DATA
Expected
0 COUNTS

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Save changes and exit

Save new calibration values and exit Biomed Calibration menu. Option to exit without saving if calibration error occurred.

Ranked causes

  1. 1. Need to save calibration changes after adjustment (common)
First check

Confirm calibration changes ready to save

Test point
Calibration menu
Expected
All required calibrations completed

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Size and position sensor combined

Combined recalibration of both syringe size and plunger position sensors when both require adjustment.

Ranked causes

  1. 1. Size and position sensors require recalibration (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Small syringe size and position sensor check caution

Verify position sensor reading is within acceptable range when plunger driver is loaded on calibration slug with light force applied.

Ranked causes

  1. 1. Position sensor calibration drift or error (unknown)
First check

Verify position sensor reading with calibration slug loaded

Test point
screen position display
Expected
between 0.660" and 0.740", inclusive inches, ±0.04 inches
Tools
Small calibration slug, calipers with 0.001" resolution

Calibration: Syringe size sensor verification

Annual maintenance check to verify syringe size sensor readings are within specification using small and large calibration slugs

Ranked causes

  1. 1. Syringe size sensor out of calibration (unknown)
First check

Verify small calibration slug size reading

Test point
Size reading on screen
Expected
between 0.256" and 0.272", inclusive inches, ±0.008

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Tubing holder size

Recalibration of size sensor needed after tubing holder replacement to maintain accurate syringe size detection.

Ranked causes

  1. 1. Tubing holder replacement affects size sensor calibration (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: View calibration data

Display pump calibration data on screen for factory service or troubleshooting. Shows Force, Plunger Position, Pressure, and Syringe Size count values.

Ranked causes

  1. 1. Need to verify calibration data for troubleshooting (occasional)
First check

Review Force calibration data

Test point
Calibration data screen
Expected
Force counts display (reference: 42-475 COUNTS range)

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Clutch or clutch cam replacement caution

Clutch halves or cam gear requires removal and reinstallation in carriage assembly during service.

Ranked causes

  1. 1. Worn or damaged clutch or cam requiring replacement (unknown)
First check

Remove e-ring from cam in carriage plate

Test point
Cam in carriage plate
Expected
E-ring removed
Tools
Needle nose pliers or hemostats

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow delivery rate outside acceptable tolerance patient safety

Measured syringe plunger travel does not fall within specified range for selected syringe type

Ranked causes

  1. 1. Pump delivery mechanism calibration error (unknown)
First check

Configure pump for flow delivery accuracy test

Test point
Begin Infusion screen
Expected
Rate of 300 ml/hr selected with volume limit of 50 ml

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Installation/Quick check-out: Force sensor verification with no load caution

Verify force reading is within nominal range when pump is powered on with no syringe loaded.

Ranked causes

  1. 1. Force sensor zero offset error (unknown)
First check

Verify force reading with no syringe loaded

Test point
screen force display
Expected
between –0.7 and +0.7 pounds, ±0.7 pounds

Keyclick beep is faint or absent

Button click sounds from keypad are faint or not heard

Ranked causes

  1. 1. Keyclick audio configured to lowest level or disabled (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keyclick beeps are faint or absent

Key press feedback sounds (beeps) are inaudible or very quiet; audio feedback disabled or configured to low level

Ranked causes

  1. 1. Keyclick volume configured to lowest level or disabled (common)
First check

Check keyclick loudness setting (V3 pumps)

Test point
Custom Program audio settings menu
Expected
Keyclick loudness set to medium or high level; not at minimum

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad button does not respond

Pressing keypad button does not register or control pump function

Ranked causes

  1. 1. Keypad mechanical or connection failure (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad replacement

Keypad requires removal from top housing and disconnection of interconnect and backlight cables; removal of ground tab.

Ranked causes

  1. 1. Damaged keypad or membrane switch requiring replacement (unknown)
First check

Disconnect Interconnect ribbon cable from Main PCB

Test point
J9 connector on Main PCB
Expected
Interconnect cable unplugged

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD and backlight assembly replacement

Old-style or new-style LCD display with backlight requires removal and reinstallation; backlight orientation critical for function.

Ranked causes

  1. 1. Damaged LCD display or backlight assembly requiring replacement (unknown)
First check

Disconnect display cable from Main PCB

Test point
J12 connector on Main PCB
Expected
Display cable disconnected

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Libraries not available in configuration

Drug infusion libraries do not appear in pump configuration menu

Ranked causes

  1. 1. Libraries disabled or cleared in configuration (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low battery capacity detected

LMD (Last Measured Discharge) below threshold; battery replacement needed

Ranked causes

  1. 1. Battery capacity degraded below 1600 mA-hours (common)
First check

Monitor LMD value in Biomed > Diagnostics > Monitor Battery Status

Test point
Biomed > Diagnostics > Monitor Battery Status
Expected
LMD >= 1600 mA-hours

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Maximum infusion rate not available

Maximum flow rate setting option does not appear or is limited

Ranked causes

  1. 1. Maximum rate configured too low (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • No drug libraries, profiles, or categories available
  • Occlusion alarm does not occur during test
  • Occlusion alarms occur frequently
  • Plunger cable replacement
  • Power-up test
  • Pump activation before manual procedures
  • Pump cleaning and disinfection
  • Pump option or feature is not available
  • Pump requires annual maintenance
  • Syringe manufacturer not available
  • Syringe manufacturer not available in pump Configuration
  • System Fault alarm activated
  • Watchdog Fail-Safe alarm triggered
  • Watchdog or system failure alarm

14 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
A2D Reference Voltage BGND TestAnalog-to-digital converter reference voltage invalid during background self-testBackground self-test measured invalid range on ADC reference voltagecritical
A2D Reference Voltage BGND TestInvalid range on ADC reference voltageBackground self-test executionerror
ACU Power Strobe FailureAuxiliary Control Unit processor not maintaining power strobe signalSelf-test found ACU processor not maintaining power strobecritical
ACU Watchdog FailureAuxiliary Control Unit watchdog alarm signal activeSelf-test found ACU watchdog alarm signal active, processor failed to pet watchdogcritical
Analog Supply BGND TestAnalog power supply voltage outside valid range during background self-testBackground self-test measured invalid range on analog supply voltagecritical
Aux Controller Unit PostACU processor failed motor current shutdown during watchdog alarm testPower-up self-test found ACU processor failed to shut down motor current during watchdog testcritical
Background CRC Test TimeoutSoftware timing failure in cyclic redundancy check during background testSoftware timing failure during background CRC testcritical
Background Self Test TimeoutSoftware timing failure in background self-testBackground self-test executionerror
Background Self Test TimeoutSoftware timing failure during background self-test operationSoftware timing failure during background self-testcritical
Backup Audible Alarm PostBackup audible alarm not working at correct power or frequencyPower-up self-testwarning
Backup Critical Data CorruptedBackup configuration data lost, primary data used to restoreBackup critical data in non-volatile EEPROM is lostinfo
Battery Communication TimeoutBattery gauge not communicating with pumpBattery current detected but gauge not respondingwarning
Battery Depleted alarmBattery is completely discharged and pump cannot operateBattery voltage drops to minimum usable level during unplugged operationcritical
Battery Not ChargingBattery is not charging despite external power presentAC or DC external power detected but battery not chargingwarning
Battery Not WorkingBattery very low while pump connected to external powerVery low battery signal detected during AC/DC power connectionwarning
Calculated Rate Out of RangePump calculated a rate that is not valid based on physical limitsInvalid combination of entered infusion parameters or syringe parametersunknown
Calibration RequiredSensor calibration data test failedPower-up or calibration Save operation showed calibration data failureerror
Calibration RequiredCalibration values in non-volatile memory invalid or missingCalibration values stored in EEPROM are invalid, typically after new main board installationcritical
Check Clutch / Plunger LeverPlunger driver motion is not correctPosition potentiometer senses plunger driver not moving correctlyunknown
Check Syringe Barrel ClampSyringe barrel clamp sensor has difficulty verifying clamp positionCertain syringe models/sizes; barrel clamp lifted during power-upunknown
Check Syringe Flange SensorSyringe not loaded correctly as sensed by flange sensorSyringe flange sensor detects improper loading or sensor failureunknown
Check Syringe Plunger SensorSyringe not loaded correctly as sensed by plunger sensorPlunger sensor detects incorrect loading or Syringe Plunger Driver moved during startupunknown
Check Syringe Plunger SensorPump detected plunger loaded when no syringe presentAt power-up when barrel clamp sensor indicates no syringecritical
Clear TVD to Start InfusionTotal volume delivered exceeds programmed limit and must be clearedTVD or TDD exceeds volume limit settinginfo
Completing System InitializationUser pressed start before background self-tests completedUser presses start button before pump self-testing completes at power-upinfo
Completing System InitializationSelf-tests still running when user attempted to start infusionUser presses Enter before background self-tests complete after power-upinfo
Control Key Switch BGND TestKeypad key detected stuck on during continuous background self-testContinuous background self-test found key stuck oncritical
Control Key Switch PostKeypad key detected stuck on during power-on self-testPower-on self-test found key stuck on keypadcritical
Control Key Switch PostSingle keypad key pressed for extended durationKey pressed for longer than 7 seconds; or locked keypad key pressed longer than 1 minuteerror
Critical Data Block BGND TestCritical delivery settings RAM corrupted as detected by background CRC testBackground self-test showed critical data block RAM corrupted via CRC test failurecritical
Critical Data Block PostConfiguration data backup and primary CRC mismatchBackup and primary configuration data fail CRC agreementcritical
Critical Data Block PostPrimary and backup configuration data in EEPROM do not agreePower-up self-test found primary and backup configuration/library data both good but not matchingcritical
Critical Data FailureSettings reverse calculation verification failed or critical flags invalidAfter programming settings, pump verification calculation failed or critical software flags invalidcritical
Critical Data FailureSettings validation or critical flag failureAfter programming settingserror
D2A Offset Voltage BGND TestForce sensor offset signal failure in DACBackground self-test found failure in force sensor offset signalcritical
Data Does Not Match Previous EntryUser confirmation value does not match original entryWhen confirmation of user-entered data is required but values don't matchinfo
Default Configuration RestoredAll configuration settings returned to factory defaultsBoth primary and backup configuration lost in EEPROMwarning
Depleted BatteryBattery fully discharged below low voltage thresholdBattery voltage measured by gauge falls below low voltage thresholdwarning
Depleted Battery (System Failure)Battery voltage critically low, delivery stoppedBattery voltage measured by pump gauge is too lowcritical
Depleted Battery (System Failure)Battery fully discharged below low voltage thresholdBattery voltage drops below fully depleted thresholdcritical
Display Controller PostLCD display test failure detected during power-up self-testPower-up self-test failure in LCD display controllercritical
External COM Task TimeoutExternal device sending commands too quickly or too many invalid commandsExternal computer/device sending excessive/invalid commands to pumperror
Flash Memory BGND TestProgram memory CRC failure detected during background self-testCRC failure found in program memory during background self-testscritical
Flash Memory PostProgram memory CRC failure detected during power-up self-testCRC failure found in program memory during power-up self-testscritical
Force Sensor BGND TestForce sensor signal out-of-range or not responding to bridge test during background testBackground self-test found force sensor signal out-of-range or no response to bridge test signalcritical
Force Sensor Bgnd TestForce sensor self-test failedSelf-test performed on force sensor and failed; plunger driver moved or received hard bump during startupunknown
Force Sensor Bridge TestForce sensor bridge shows no voltage change during testBackground self-test executionerror
Force Sensor Bridge TestForce sensor bridge strain gauge no voltage change when test signal assertedSoftware sensed no voltage change when sensor test signal was assertedcritical
Force Sensor TestForce sensor output did not change when DAC offset voltage changedPower-up self-test found force sensor output not responding to DAC offset changecritical
History Data CorruptedInfusion history log is corruptedSoftware unable to retrieve history data in 6 consecutive attemptsinfo
Improper ShutdownPrevious power-down did not occur through normal power key pressPump powered on; last shutdown was not via power button (e.g., power loss, crash)warning
Improper ShutdownPump powered off without proper shutdown procedurePower-down did not occur through pressing power buttoninfo
Infusion CompleteProgrammed infusion volume delivered successfullyInfusion reached the set volume limitwarning
Infusion CompleteInfusion has reached programmed volume limitDelivered volume reaches set volume limitinfo
Input Out of Range - Greater Than Max ValueUser-entered value exceeds maximum allowable limitUser attempts to enter number greater than maximum value allowedinfo
Input Out of Range - Less Than Min ValueUser-entered value below minimum allowable limitUser attempts to enter number less than minimum value allowedinfo
Input Task TimeoutSoftware timing failure in input processing taskSoftware timing failure in input taskcritical
Internal A2D Failure10-bit analog-to-digital converter internal failure detectedSoftware self-test identified failure in 10-bit ADCcritical
Invalid Infusion Parameter CombinationParameter combination does not allow valid flow rate calculationUser entered invalid parameter combination or wrong syringe size for Drug Programwarning
Invalid InterruptSoftware attempted to execute invalid interrupt vectorSoftware/hardware detected invalid interrupt executioncritical
Invalid Interrupt & System Advisory: Improper ShutdownCritical interrupt error coinciding with improper shutdownInvalid interrupt detected along with improper shutdown eventcritical
Invalid Interrupt & System Advisory: Improper ShutdownImproper shutdown alarm combined with invalid interrupt conditionSimultaneous improper shutdown and system interrupt detectederror
Invalid Library EntryLibrary entry was not defined correctly when createdPump attempts to use improperly configured library entryinfo
Invalid Library EntryDrug Program selected is invalid; pump cannot deliver itProblem detected with selected Drug Programerror
Invalid Rate for Syringe Size (V3 pumps only)Programmed infusion rate incompatible with selected syringe typeWhen using E-plates library entries, changing syringe size causes mismatchwarning
Invalid Syringe SizeSyringe size not recognized or improperly handled during deliveryUnrecognized syringe for model; barrel clamp lifted during deliveryunknown
Invalid Syringe SizeSyringe size sensor reading does not match selected syringe manufacturerBarrel clamp potentiometer sensor mismatch or barrel clamp lifted during deliverycritical
KVO in ProgressDelivery switched to KVO (keep vein open) rate after volume limit reachedInfusion reached set volume limit and automatically changed to KVO ratewarning
Learn Error - Remote Not in Teach ModeConfiguration cloning failed due to Teacher pump not ready or not communicatingLearning pump initiates learn process but cannot detect Teacher pump in active Teach Mode via infrarederror
LED Task TimeoutSoftware timing failure in LED control taskSoftware timing failure in LED taskcritical
Limits Adjusted - Cannot Delivery All DosesSoft limits adjusted to keep rate within allowable limitsPharmGuard soft limits and syringe parameters would result in rate outside allowable limitswarning
Logic Supply Voltage BGND TestLogic power supply voltage outside valid range during background self-testBackground self-test measured invalid range on logic supply voltagecritical
Low Battery (1% drop incremental)Battery capacity continues to drop below 10%Each 1% drop in battery capacity after initial signal when power cord unpluggedwarning
Low Battery alarmBattery charge level is lowBattery voltage drops below low threshold during unplugged operationwarning
Low Battery (each 1% drop in capacity following initial signBattery capacity drops below 10%; percentage alerts continue if unpluggedBattery capacity measured at each percent below 10%; continues while on battery powerinfo
Low Battery (initial signal)Battery capacity fallen below 10% thresholdInitial low battery detection when capacity drops below 10%warning
Maintenance Is RecommendedPeriodic maintenance reminder for annual maintenance testingMaintenance interval has elapsed since last scheduled preventive maintenanceinfo
MCU Power Strobe FailureMain processor power strobe output signal test failureSelf-test of power strobe output port signal from main processor failedcritical
Monitor Task TimeoutSoftware timing failure in monitoring taskSoftware timing failure in monitor taskcritical
Motor Failed to Stop ErrorStepper motor continues rotating after stop command issuedSoftware sensed motor rotation after stop pump motor command executedcritical
Motor Not RunningMotor is not runningErroneous sensor signal or mechanical/electrical failureunknown
Motor Not Running ErrorStepper motor does not rotate when rotation expectedMotor sensor did not detect stepper motor rotationcritical
Motor Not Running ErrorMotor sensor did not detect stepper motor rotationMotor sensor failed to detect stepper motor rotationcritical
Motor Rate ErrorStepper motor operating at incorrect speedStepper motor moving at wrong speed during operationcritical
Motor Rate ErrorMotor is not running at programmed rateMotor speed not matching programmed rateunknown
Motor Rate ErrorMotor rotation rate not matching expected valueMotor sensor did not measure correct stepper motor rotation ratecritical
Motor Supply Voltage BGND TestMotor power supply voltage outside valid range during background self-testBackground self-test measured invalid range on motor supply voltagecritical
Motor Volume Limit ErrorMotor has exceeded expected step count for programmed volume deliveryMotor runs more steps than required to deliver set volume limitcritical
Neglected Pump (User Call-back)Pump left in user input screen and inactive for extended timePump remains in user input screen (e.g., set rate, titrate rate) for more than 30 secondswarning
Non Critical Data CorruptedNon-critical settings data in EEPROM lost or corruptedNon-critical settings stored in EEPROM is lostinfo
Occlusion alarmLine occlusion detected during infusionPressure rise in infusion tubing indicates blocked lineerror
Occlusion - Check Infusion LineInfusion line is occludedForce sensor in plunger driver senses occlusion when force exceeds set limitunknown
Occlusion - Check Infusion LineHigh backpressure detected in infusion line due to blockageForce sensor in plunger driver sensed occlusion when force exceeds set limitcritical
Outside Range Limit - Silence Alarm to ContinueUser-entered infusion rate outside established library safety limitsUser enters rate which is outside of library limitswarning
Periodic Callback - Silence Alarm to ContinueUser-programmed periodic callback alarm activatedPeriodic alarm set by user reaches scheduled timewarning
PharmGuard Data CorruptedPharmGuard Safety Data in EEPROM lost or unreadablePharmGuard Safety Data stored in EEPROM is lost or corruptedinfo
PharmGuard Data Transfer is RecommendedPharmGuard Safety Data storage nearly fullPharmGuard Safety Data approaches storage capacitywarning
Plunger Sensor FailurePlunger position sensor voltage out of rangeSoftware sensed plunger sensor voltage out of rangecritical
Positive Supply BGND TestPower supply malfunction detectedBackground self-test executionerror
Positive Supply BGND TestPositive power supply voltage out of valid rangeBackground self-test measured invalid positive supply voltagecritical
Pressure Increasing - Check Infusion LineBackpressure in infusion line is gradually increasingForce sensor detects increasing pressure during infusionwarning
Primary Audible Alarm BGND TestNo current detected in primary speakerBackground self-test executionerror
Primary Audible Alarm PostNo current detected in primary speaker during power-up testPower-up self-testerror
Primary Critical Data CorruptedPrimary configuration data lost, backup data used to restorePrimary critical configuration data in EEPROM is lostinfo
Profile Does Not Match Last Settings ProfileCurrent pump profile does not match profile used for recalled settingsRecall Last Settings feature invoked for different profile than settings were programmed underinfo
Profile Does Not Match Last Settings ProfileRecall Last Settings incompatible with current delivery profileUser attempts Recall Last Settings but currently selected profile differs from saved profileinfo
Pump is LockedPump settings locked during active deliveryPump in delivery mode with settings lockedinfo
Pump Motor Drive Off PostNo current detected in stepper motor at power-upPower-up self-test sensed no current in stepper motorcritical
Pump Motor Drive Phase A PostStepper motor winding A open at power-upPower-up self-test found stepper motor phase A winding opencritical
Pump Motor Drive Phase B PostStepper motor winding B open at power-upPower-up self-test found stepper motor phase B winding opencritical
RAM BGND TestRAM failure on Main PCBSoftware test showed failure of RAM on Main PCBcritical
Rate Below Recommended Min for Syringe SizeProgrammed infusion rate below minimum recommended for syringe typeUser programs rate lower than manufacturer recommendation for selected syringeinfo
Restricted Flow - Bolus CancelledBolus dose delivery cancelled due to flow restrictionsDuring bolus rate reduction, rate fell below main infusion rate or time extended past 59 minutes maximumwarning
Restricted Flow - Bolus CancelledBolus rate reduction resulted in rate below main rate or max time exceeded; bolus cancelledDuring bolus dose rate reduction, calculated rate fell below main infusion rate or total time exceeded 59 minuteswarning
Restricted Flow - Loading CancelledLoading dose rate reduction resulted in rate below main rate or max time exceeded; loading dose cancelledDuring loading dose rate reduction, calculated rate fell below main rate or total time exceeded 59 minuteswarning
Restricted Flow - Loading CancelledLoading dose delivery cancelled due to flow restrictionsDuring loading dose rate reduction, rate fell below main infusion rate or time extended past 59 minuteswarning
Restricted Flow - Rate ReducedPump reduced infusion rate to overcome high backpressure during bolus or loading doseHigh backpressure detected during bolus or loading dose delivery on Medfusion 3500warning
Serial EEPROM TimeoutSerial EEPROM communication timeout on main boardSoftware test executionerror
Set Time and DateReal-time clock reset to defaultReal-time clock reset when battery depleted or main board supply dischargedwarning
Set Volume Limit Before KVOVolume limit must be set before KVO rate can be programmedUser attempts to set KVO rate without first setting volume limitwarning
SPI BUS TimeoutSPI bus component failure timeoutSoftware test executionerror
Supercap PostSuper-capacitor insufficient charge at power-upPower-up self-test sensed insufficient charge in super-capacitorwarning
Syringe Does Not Match Library EntryLoaded syringe size does not match library entry specificationWhen using E-plates library entries, loaded syringe does not match stored syringe sizeinfo
Syringe EmptyPlunger driven to end of syringe barrel, no fluid remainsPlunger driver position potentiometer senses syringe empty positioncritical
Syringe EmptySyringe has been completely emptiedPlunger driver position potentiometer reaches empty position for loaded syringe sizeerror
Syringe Empty - ManualSyringe reached empty position; user can resume to deliver residual fluid until occlusion limit reachedAfter Syringe Empty alert, user presses resume to extract remaining fluid in syringe barrelerror
Syringe Empty - ManualAfter reaching empty position, manual continuation option availableAfter Syringe Empty alarm, user presses Enter to resumecritical
Syringe Flange Not in PlaceSyringe flange is not properly insertedSyringe not loaded correctly or flange sensor failureunknown
Syringe Flange Not in PlacePlunger flange not detected in pump when syringe loadedWhen syringe is loaded but pump does not detect flange via barrel clamp sensorcritical
Syringe Flange Sensor FailureSyringe flange sensor voltage out of rangeSoftware sensed syringe sensor voltage out of rangecritical
Syringe Near EmptySyringe has only a few minutes of fluid remainingPlunger position indicates low fluid volumewarning
Syringe Plunger Not in PlacePump does not detect plunger in barrel when syringe loadedWhen syringe is loaded but pump does not detect plunger via barrel clamp sensorcritical
Syringe Plunger Not in PlaceSyringe plunger is not properly positionedSyringe not loaded correctly as sensed by barrel clamp size sensor or plunger sensor failureunknown
SYSTEM ADVISORY – BATTERY NOT WORKINGBattery malfunction or disconnection detectedBattery gauge detects failure or absence of battery powerwarning
System Advisory - DescriptionGeneric system advisory alert with additional descriptionVarious system-level advisory conditionsinfo
System Failure - DescriptionGeneric system failure alert with additional descriptionVarious critical system failure conditionscritical
Time Base BGND TestSystem time base does not agree with time of day clockBackground self-test found MCU time base did not match clock tickerror
Time of Day Clock PostSystem time base does not agree with time of day clock at power-upPower-up self-test found MCU time base did not match clock tickerror
Time of Day Clock TimeoutTime of day clock failed to communicateTime of day clock failed to communicate during software testerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.