Smiths Medical HL-90: troubleshooting and error codes
The ServiceForge knowledge base holds 35 documented symptoms for the Smiths Medical HL-90, with 74 ranked causes, 83 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Over Temperature alarm triggers continuously patient safety
Over Temperature alarm sounds and remains on during operation
Ranked causes
- 1. Insufficient recirculating solution flow due to kinked warming set (common)
- 2. Airlock in Fluid Warming Set (common)
- 3. Pump seized (occasional)
- 4. Triac short circuit (occasional)
- 5. Opto defective (occasional)
- 6. Thermistor defective (occasional)
- 7. Over Temperature alarm set point too low (occasional)
- 8. Unit accidentally filled with hot recirculating solution (rare)
- 9. Ambient temperature too high (rare)
10 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Over temperature alarm continuously on caution
Alarm condition triggered; fluid warming system is overheating or alarm setpoint is too low.
Ranked causes
- 1. Insufficient recirculating solution flow (common)
- 2. Triac short circuit (occasional)
- 3. Opto defective (occasional)
- 4. Thermistor defective (occasional)
- 5. Over temperature alarm setpoint too low (rare)
- 6. Unit filled with hot recirculating solution (rare)
- 7. Ambient temperature too high (rare)
Visually inspect Warming Set for kinks
- Test point
- Warming Set tubing
- Expected
- straight with free flow
2 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluid Warming Set Interlock alarm triggers continuously caution
Fluid Warming Set Interlock alarm sounds and remains on
Ranked causes
- 1. Microswitch not being made; set not fitted correctly (common)
- 2. Microswitch lever out of shape (occasional)
- 3. Microswitch is faulty (occasional)
- 4. Wiring or PCB fault (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluid Warming Set interlock alarm continuously on caution
Alarm indicating Warming Set is detected as not properly installed or microswitch is stuck; Warming Set may not be fitted correctly.
Ranked causes
- 1. Fluid Warming Set not fitted correctly (common)
- 2. Microswitch lever out of shape or faulty microswitch (occasional)
- 3. Wiring or PCB fault at J3 connector (rare)
Verify Warming Set is firmly seated
- Test point
- Warming Set connector
- Expected
- set seated fully and firmly
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluid Warming Set Interlock alarm not working caution
Fluid Warming Set Interlock alarm does not trigger when it should
Ranked causes
- 1. Microswitch lever stuck (common)
- 2. Defective microswitch (occasional)
- 3. Wiring or PCB fault (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No electrical power to unit
Unit is not receiving mains power; all indicators are off
Ranked causes
- 1. Power cord unplugged or unit switched off (common)
- 2. Fuse F1 or F2 blown (common)
- 3. Transformer open circuit or F3 blown (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recirculating Solution Level alarm triggers continuously caution
Recirculating Solution Level alarm sounds and remains on
Ranked causes
- 1. Recirculating solution is actually low (common)
- 2. Sticky float switch (occasional)
- 3. Wiring or PCB fault (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit has no power high voltage
Unit does not power on; mains supply or internal power distribution failure.
Ranked causes
- 1. Unit not plugged in or socket not live (common)
- 2. Fuse F1 or F2 blown (common)
- 3. Transformer open circuit or fuse F3 blown (occasional)
Verify unit is plugged in and turned on
- Test point
- power switch and mains connector
- Expected
- unit switched on, plugged into live socket
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not heat up caution
No heat output or insufficient heating; thermostat, triac, heater, or setpoint may be faulty.
Ranked causes
- 1. Bad thermostat, triac or heater element (common)
- 2. Idle temperature setpoint too low (occasional)
- 3. Ambient temperature too high (rare)
Test thermostat, triac and heater element
- Test point
- heating circuit components
- Expected
- all components functional and properly connected
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not heat up caution
Unit does not reach desired temperature or provide heat
Ranked causes
- 1. Bad T-Stat, Triac or Heater (common)
- 2. Idle temperature set point too low (occasional)
- 3. Ambient temperature too high (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Temperature display accuracy
Verify displayed recirculating solution temperature matches actual temperature measured at heater element outlet.
Ranked causes
- 1. Temperature calibration drift or offset (common)
- 2. Over-temperature alarm malfunction (occasional)
Verify ambient room temperature within operating range
- Test point
- Room environment
- Expected
- 22.0°C ± 2.0°C, ±2.0°C
- Tools
- Room thermometer
8 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Floating or erratic LCD reading caution
LCD display shows unstable or drifting numerical values
Ranked causes
- 1. Bad LCD unit (common)
- 2. Bad 5V regulator (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluid Warming Set interlock alarm not working caution
Alarm not triggering when expected; microswitch stuck or not operating, or wiring fault.
Ranked causes
- 1. Microswitch lever stuck (occasional)
- 2. Defective microswitch or wiring fault (occasional)
Check microswitch lever operation
- Test point
- microswitch lever
- Expected
- lever moves freely, makes and breaks contact
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD display reading floating or erratic
LCD display shows unstable or changing values; LCD unit or 5V regulator may be faulty.
Ranked causes
- 1. Bad LCD unit (occasional)
- 2. Bad 5V regulator (occasional)
Test LCD display by substitution
- Test point
- LCD display unit
- Expected
- stable, consistent reading with correct value
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Over temperature alarm not working caution
Display reads correctly but alarm is not triggering; thermistor or alarm setpoint may be at fault.
Ranked causes
- 1. Over temperature alarm setpoint too high (common)
- 2. Thermistor defective or display not reading correctly (occasional)
Verify over temperature alarm setpoint
- Test point
- alarm setpoint setting
- Expected
- appropriate threshold for application
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Over Temperature alarm not working patient safety
Over Temperature alarm does not trigger when it should
Ranked causes
- 1. Display reading correctly but alarm set point too high (common)
- 2. Display not reading correctly; thermistor defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recirculating solution level alarm continuously on caution
Low solution level alarm triggered; actual solution level may be low or float switch may be stuck.
Ranked causes
- 1. Recirculating solution is actually low (common)
- 2. Sticky float switch or wiring fault (occasional)
Visually inspect recirculating solution level
- Test point
- solution reservoir level
- Expected
- above minimum level mark
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recirculating Solution Level alarm not working caution
Recirculating Solution Level alarm does not trigger when solution is low
Ranked causes
- 1. Defective float switch (common)
- 2. Wiring or PCB fault (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: nominal temperature set point patient safety
Adjustment of water temperature nominal set point to 40°C or 42°C depending on territory
Ranked causes
- 1. Territory-specific nominal temperature configuration required (common)
Verify water temperature at 40°C nominal set point
- Test point
- Water temperature sensor
- Expected
- 39.9°C to 41.1°C, -0.1 to +1.1°C
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Temperature verification
Verify temperature calibration of the HOTLINE blood warmer unit meets specifications
Ranked causes
- 1. Annual calibration maintenance requirement (common)
Allow unit to stabilize temperature during normal operation
- Test point
- Display
- Expected
- Temperature gradually rising towards 41°C
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disposable set interlock alarm test failure patient safety
Unit does not trigger interlock alarm and associated indicators when disposable set is disconnected
Ranked causes
- 1. Disposable set interlock alarm system malfunction (unknown)
Gently ease disposable set away from interlock block
- Test point
- Disposable set connector
- Expected
- Alarm triggers before full disconnection
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
General alarm test failure patient safety
Unit does not trigger all alarm indicators and audible alert when general alarm test button pressed
Ranked causes
- 1. Alarm system malfunction (unknown)
Verify GREEN LED extinguishes
- Test point
- GREEN LED indicator
- Expected
- LED extinguished
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Maintenance: O-Ring seal replacement
Replace O-Ring seals in disposable interface block connector as part of annual maintenance
Ranked causes
- 1. Scheduled annual O-Ring seal replacement (common)
Remove socket head screws
- Test point
- Disposable interface block
- Expected
- All screws removed
- Tools
- 1/8-inch Allen key
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Maintenance: Recirculating solution change (distilled water)
Drain and replace distilled water recirculating solution at 30-day intervals
Ranked causes
- 1. Scheduled 30-day recirculating solution change (common)
Unplug the unit from power supply
- Test point
- Mains power
- Expected
- Unit disconnected from electricity
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Maintenance: Recirculating solution change (isopropyl alcohol) caution
Drain and replace 35% isopropyl alcohol recirculating solution at 30-day intervals
Ranked causes
- 1. Scheduled 30-day recirculating solution change (common)
Prepare two batches of 1.4 litres solution per Formula 2
- Test point
- Solution preparation
- Expected
- Formula 2: 700ml 70% isopropyl alcohol PLUS 700ml distilled water, prepared twice litres
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Over temperature alarm test failure patient safety
Unit does not trigger over temperature alarm and associated indicators when over temperature alarm test button is held
Ranked causes
- 1. Over temperature alarm system malfunction (unknown)
Verify display temperature rises
- Test point
- Temperature display
- Expected
- 43°C - 44°C, to 44°C
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recirculating solution leaks at Warming Set connector caution
Fluid leaking from connection between unit and Warming Set; O-rings worn or degraded.
Ranked causes
- 1. Worn O-rings at connector (common)
Inspect O-rings for wear or degradation
- Test point
- O-rings on unit and Warming Set connector
- Expected
- O-rings intact and not cracked or deformed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recirculating solution leaks at Warming Set connector caution
Fluid leaks from the connection point between the warming set and the unit
Ranked causes
- 1. Worn O-rings (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recirculating solution level alarm not working caution
Alarm not triggering when solution level is low; float switch defective or wiring fault.
Ranked causes
- 1. Defective float switch or wiring fault (occasional)
Test float switch operation at low level
- Test point
- float switch
- Expected
- makes contact when solution is low
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recirculating solution level alarm test failure patient safety
Unit does not trigger low level alarm and associated indicators when float switch is depressed
Ranked causes
- 1. Recirculating solution level alarm system malfunction (unknown)
Depress float switch arm using non-magnetic tool
- Test point
- Float switch
- Expected
- Switch depressed without magnetic attraction
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recirculating solution replacement (Hydrogen Peroxide) caution
Replace 0.3% Hydrogen Peroxide recirculating solution at annual service interval
Ranked causes
- 1. Hydrogen Peroxide solution degradation at 12-month interval (common)
Prepare two 1.4 litre batches of 0.3% Hydrogen Peroxide solution per formula
- Test point
- Solution mixing area
- Expected
- Two batches of 1.4 litres each, 0.3% Hydrogen Peroxide concentration
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recirculating solution replacement (Isopropyl Alcohol) caution
Replace Isopropyl Alcohol disinfectant solution at 30-day intervals if Hydrogen Peroxide unavailable
Ranked causes
- 1. Isopropyl Alcohol solution potency loss at 30 days (common)
Prepare two 1.4 litre batches of Isopropyl Alcohol solution (Formula 2) immediately before use
- Test point
- Solution mixing area
- Expected
- Two fresh batches prepared per Formula 2, used same day
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature verification check
Verify displayed temperature matches actual recirculating solution temperature without full calibration
Ranked causes
- 1. Display temperature does not match actual temperature (occasional)
Attach TEMP CHECK thermometer to interlock block and Disposable Set
- Test point
- Interlock block connector
- Expected
- TEMP CHECK properly connected to measure solution flow temperature
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Warming Set hard to install
Warming Set connector difficult to insert into unit socket; O-rings may be dry or damaged.
Ranked causes
- 1. Dry O-rings on connector (common)
Inspect O-rings for dryness or deterioration
- Test point
- O-rings on Warming Set connector
- Expected
- flexible and lubricated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Warming Set hard to install caution
Fluid Warming Set is difficult to connect to the unit
Ranked causes
- 1. Dry O-rings (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.