Smiths Medical FMS 2000: troubleshooting and error codes
The ServiceForge knowledge base holds 59 documented symptoms for the Smiths Medical FMS 2000, with 106 ranked causes, 83 diagnostic checks with test points and expected values, and 27 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Fluid out condition detected patient safety
System reports low or no fluid; stops pumping and heating; sounds alarm and displays message
Ranked causes
- 1. Infusion bag empty or depleted (common)
- 2. Bag clamps not fully opened (common)
- 3. Bag not fully spiked (occasional)
- 4. Tubing not seated firmly in fluid out sensor (occasional)
- 5. Tubing stretched in fluid out sensor (occasional)
- 6. Air vent filter clogged (occasional)
- 7. Coarse blood filter clogged (occasional)
- 8. Reservoir or recirculate line obstructed (rare)
Check fluid bag level and reprime function
- Test point
- Infusion bag fluid level; reprime volume count display
- Expected
- Bag contains fluid; REPRIME volume count decrements from 100 to 0 ml
2 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluid out alarm patient safety
Alarm message 'FLUID OUT, CHECK INLET TUBING AND FILTER. ADD MORE FLUID.' indicates the infusate reservoir is empty or fluid supply is depleted
Ranked causes
- 1. Out of infusate in fluid bag (common)
- 4. Fluid out sensor tubing not seated firmly in detector (occasional)
- 5. Fluid out sensor tubing stretched or misaligned (occasional)
- 6. Clogged air vent filter on fluid bag (occasional)
- 7. Clogged coarse blood filter in reservoir chamber (occasional)
- 8. Reservoir or recirculate line obstructed (occasional)
Verify fluid level in bag and reservoir
- Test point
- Fluid bag and reservoir chamber
- Expected
- Fluid present in bag with adequate volume; reservoir chamber has visible fluid level
6 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air detected in fluid line patient safety
Air bubble detected in air detection sensor; system alarms and stops pumping and heating
Ranked causes
- 1. Tubing not seated firmly in air detector (common)
- 2. Trapped air in air detection sensor (common)
- 3. Leak in disposable set (occasional)
- 4. Air detector sensor dirty or obstructed (occasional)
Verify tubing seating in air detector
- Test point
- Air detector sensor chamber
- Expected
- Tubing seated firmly with no gaps; fluid flows freely through sensor
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High pressure alarm patient safety
Alarm message 'HIGH PRESSURE DETECTED CHECK PATIENT LINE FOR BLOCKAGE.' indicates line pressure exceeds safe limits
Ranked causes
- 1. Patient line is blocked or obstructed (common)
- 2. Recirculate line is blocked (common)
- 3. Infusion site is not well placed or catheter bore is too small (common)
- 4. Pressure limit setting is set too low (occasional)
Inspect patient line for blockage
- Test point
- Patient line pathway from warmer to patient
- Expected
- Line is clear and unobstructed with no kinks, clots, or debris
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No heating during operation patient safety
System does not heat fluid to physiologic temperature; temperature display incorrect or heating fault message appears
Ranked causes
- 1. Windows on disposable or IR sensor wet or dirty (common)
- 2. Power module not calibrated properly (occasional)
- 3. Temperature probes out of calibration (occasional)
- 4. Power module malfunction or temperature probe failure (rare)
Examine windows on disposable set and IR sensor
- Test point
- Disposable set and IR sensor windows
- Expected
- Clean and dry
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air detection alarm patient safety
Air bubble detected in tubing; alarm sounds and message displays; pumping and heating stop; valve moves to recirculate position
Ranked causes
- 1. Air in the line (common)
- 2. Tubing in air detection sensor not seated firmly (common)
- 4. Air detector sensor dirty (occasional)
Open unit door
- Test point
- Door
- Expected
- Door opens
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not charging or no heating
Unit does not heat fluid; battery charge time excessive or run time too short
Ranked causes
- 1. Power cord not plugged into AC receptacle (common)
- 2. AC power source unavailable or cord disconnected (common)
- 3. Internal rechargeable battery depleted or failed (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
High pressure detected in line patient safety
System detects elevated line pressure; stops pumping and heating; sounds alarm
Ranked causes
- 1. Patient line or recirculate line blocked (common)
- 2. Infusion site poorly placed or catheter bore too small (occasional)
- 3. Pressure limit setting too low (occasional)
Inspect flow path for blockages
- Test point
- Patient line and recirculate line connections and tubing
- Expected
- Clear fluid flow with no visible kinks, clots, or obstructions
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No power or inadequate battery runtime
System not powered or battery discharges too quickly
Ranked causes
- 1. Power cord not plugged into AC mains (common)
- 2. Batteries discharged in DC operation (common)
- 3. Battery aged or failing (occasional)
Verify AC power source and cord connection
- Test point
- AC outlet, power cord plug
- Expected
- Power cord connected to live AC outlet
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Power module overtemp caution
Power driver module overheating; cabinet temperature reached 70C
Ranked causes
- 1. Blocked air intake vents (common)
- 2. Excessive flow rate or warm fluid (occasional)
- 3. Persistent overtemp condition (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not heat fluid to physiologic temperature patient safety
Warmer not reaching target temperature; fluid delivery cooler than expected
Ranked causes
- 1. Disposable set or IR sensor window wet or dirty (common)
- 2. Power module not calibrated properly (occasional)
- 3. Temperature probes out of calibration or power module malfunction (occasional)
Inspect windows on disposable set and IR sensor
- Expected
- Windows clean and dry, no condensation or contamination
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
System does not heat to physiologic temperature patient safety
Fluid output temperature below set physiologic level; heating ineffective
Ranked causes
- 1. Disposable set or IR sensor window wet or dirty (common)
- 2. Power module not calibrated properly or out of calibration (occasional)
- 3. Power module malfunction or temperature probe failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air detection alarm activated patient safety
Air bubble detected in air detection sensor above diversion valve; pump and heating stop; alarm displays and system enters recirculate mode.
Ranked causes
- 1. Air bubble trapped in air detection sensor (common)
- 2. Hydrophobic filter clogged by blood particulates (occasional)
Press MUTE button to silence alarm
- Test point
- Control panel MUTE button
- Expected
- Alarm sound stops; display continues to show air detection alert
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heater fault latch caution
Heater power control fault triggered by line noise or internal failure
Ranked causes
- 1. Excessive AC power line noise (common)
- 2. Internal heater failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Key pad does not accept input caution
Key pad continuously depressed or unresponsive to input; constant beeping alarm
Ranked causes
- 1. Key pad being continually depressed (common)
- 2. Key pad failure (rare)
Check key pad physical condition
- Test point
- Key pad buttons
- Expected
- Not depressed; free moving
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad does not accept input or constant beeping
Buttons unresponsive or system continuously sounds alarm
Ranked causes
- 1. Keypad held depressed or stuck mechanically (common)
- 2. Keypad failure (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad not accepting input
Keypad does not respond to key presses; constant beeping
Ranked causes
- 1. Key continuously depressed or stuck (common)
- 2. Keypad hardware failure (occasional)
Release all keys on keypad
- Expected
- No keys stuck or continuously pressed
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Missing disposable set caution
No disposable set installed in unit or disposable failed loading criteria; alarm sounds and message displays
Ranked causes
- 1. Disposable set not installed (common)
- 2. Disposable failed loading criteria check (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No power or battery run time too short caution
System unplugged or battery does not hold sufficient charge for adequate run time
Ranked causes
- 1. Power cord not plugged into AC power (common)
- 2. Rechargeable battery discharged in DC operation (occasional)
Verify power cord connection
- Test point
- AC power receptacle
- Expected
- Power cord firmly connected and AC power available
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No power or insufficient battery runtime
System does not power on or battery runtime is very short (less than 0.5 hour after full 8-hour charge).
Ranked causes
- 1. Power cord not plugged into AC power (common)
- 2. Battery discharged or battery degraded (occasional)
Verify AC power source and cord connection
- Test point
- AC outlet, power cord connectors
- Expected
- AC outlet live; power cord fully seated
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump fault caution
Pump failure with speed feedback encoder malfunction or blockage
Ranked causes
- 1. Pump blocked or dirty (common)
- 2. Pump encoder failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not prime fluid patient safety
System unable to pull fluid into lines during prime cycle; countdown does not go from 100 to 0 ml
Ranked causes
- 1. Recirculate line obstructed (common)
- 2. Fluid reservoir empty or absent (common)
Verify recirculate line not obstructed
- Expected
- Recirculate path clear
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Temperature probes fail to calibrate
Temperature probes do not calibrate successfully; calibration fails or produces errors
Ranked causes
- 1. Temperature probe hardware malfunction (occasional)
- 2. Incorrect fluid temperature used for calibration (common)
Verify calibration fluid temperature is correct
- Expected
- Fluid temperature matches calibration requirement
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Valve fault caution
Valve failure or position sensor malfunction; valve did not move to position
Ranked causes
- 1. Valve blocked or stuck (common)
- 2. Valve position sensor malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air detector fault
Air detector failure detected during system startup
Ranked causes
- 1. Air detector hardware failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm volume too low patient safety
Audible alarm volume cannot be heard adequately
Ranked causes
- 1. Alarm volume in Setup Routine turned down to lowest level (common)
Access System Setup menu
- Test point
- Setup Routine - Alarm Volume
- Expected
- Volume control accessible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dim display caution
Display brightness is too low
Ranked causes
- 1. Display brightness in Setup Routine turned to lowest level (common)
Access System Setup menu
- Test point
- Setup Routine - Display Brightness
- Expected
- Brightness control accessible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display brightness too dim
Screen display faint or difficult to read in clinical environment
Ranked causes
- 1. Display brightness set to minimum in Setup routine (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display brightness too low
Monitor display appears dim
Ranked causes
- 1. Display brightness reduced to minimum in Setup (common)
Access Calibration/Set-Up menu and increase display brightness
- Expected
- Display visible and readable
Display is dim
LCD or LED display brightness is very low or nearly unreadable.
Ranked causes
- 1. Display brightness setting turned down to lowest level (common)
Access System Setup brightness control
- Test point
- Setup menu display brightness slider/setting
- Expected
- Brightness control accessible and adjustable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Door is open caution
System detects open door; stops pumping and heating; sounds alarm
Ranked causes
- 1. Door not fully closed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Door open caution
System stops heating and pumping when door is opened or if opened during operation; audible alarm sounds; valve goes to recirculate position
Ranked causes
- 1. Door is open (common)
Door is fully closed
- Test point
- Door latch
- Expected
- Door closed without kinks in tubing
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Door open alarm patient safety
System detects open door; stops heating and pumping; valve goes to recirculate; audible alarm sounds
Ranked causes
- 1. Door physically open (common)
Close system door
- Test point
- Door position sensor
- Expected
- Door fully closed and latched
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow rate slower than set or will not reach set rate patient safety
Infusion pump output below commanded rate despite control settings
Ranked causes
- 1. Line pressure exceeds Pressure Limit setting; system reducing flow automatically (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow rate slowing or not reaching set rate patient safety
Infusion rate reduces below set rate; pressure reading flashes at pressure status line and system periodically beeps
Ranked causes
- 1. System reducing flow to maintain pressure under Pressure Limit (common)
Inspect tubing for kinks or obstructions
- Test point
- Infusion tubing
- Expected
- Clear without obstructions
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluid out caution
Fluid runs out during infusing or tubing in fluid out detector is pulled/kinked; pump and heating stop; Fluid Out message displays and audible alarm sounds; valve goes to recirculate position
Ranked causes
- 1. Empty fluid bags or kinked tubing in fluid detector (common)
Fluid bags are empty
- Test point
- Fluid bag level indicator
- Expected
- Bags contain adequate fluid
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluid out reprime failure patient safety
REPRIME volume count does not decrement from 100 to 0 ml during priming cycle
Ranked causes
- 1. Tubing blockage or disconnection (common)
Verify tubing connections
- Test point
- Fluid line connectors
- Expected
- All connections tight and secure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heater over power fault caution
Heater power consumption exceeds safe limit of 1500 watts
Ranked causes
- 1. Heater hardware fault (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heater power read back fault caution
Heater power feedback signal missing or open circuit
Ranked causes
- 1. Power feedback circuit open or malfunction (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal computer malfunction (Watch Dog alarm) patient safety
Hardware override circuit detects processor not operating properly. System stops pumping and heating; high pitch tone sounds.
Ranked causes
- 1. Processor operating improperly (unknown)
Attempt system restart
- Test point
- System power and processor
- Expected
- System restarts and operates normally without Watch Dog alarm
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Key pad sensitivity incorrect caution
Key pad is too sensitive or not responsive to input
Ranked causes
- 1. Key pad sensitivity in Setup set to Fast or Slow (common)
Access System Setup menu
- Test point
- Setup Routine - Key Pad Sensitivity
- Expected
- Sensitivity setting accessible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad is too sensitive or not responsive
Keys trigger actions too easily or do not respond when pressed; sensitivity is misadjusted.
Ranked causes
- 1. Keypad sensitivity setting at Fast or Slow setting (common)
Access System Setup keypad sensitivity control
- Test point
- Setup menu keypad sensitivity setting
- Expected
- Sensitivity control accessible with Fast/Slow/Normal options
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad sensitivity abnormal
Buttons unresponsive or overly sensitive; inconsistent input response
Ranked causes
- 1. Keypad sensitivity setting incorrect in Setup (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery
System flashes BATT LOW message and sounds audible alarm
Ranked causes
- 1. Battery depleted (common)
Verify BATT LOW message and alarm present
- Test point
- Display and audio output
- Expected
- BATT LOW message flashing with audible alarm
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery alarm
Alarm message 'LOW BATTERY' flashes and periodic beep sounds, indicating battery voltage is too low for continued operation
Ranked causes
- 1. Battery charge voltage is below safe operating threshold (common)
Verify AC power outlet availability
- Test point
- Power source for warmer unit
- Expected
- AC outlet is available and providing power to the system
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery condition caution
Battery voltage too low to sustain operation; LOW BATTERY message displayed and periodic beep sounds
Ranked causes
- 1. Battery charge depleted (common)
Measure battery voltage
- Test point
- Battery terminals or internal voltage measurement point
- Expected
- Battery charge voltage greater than 22.5 ± 0.3 volts for normal operation V, ±0.3 V
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery warning
System flashes BATT LOW message and sounds audible alarm
Ranked causes
- 1. Battery charge depleted (common)
Verify AC power connection
- Test point
- AC outlet
- Expected
- System powered from AC outlet
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing disposable
No disposable set installed in unit. Alarm sounds and message displays.
Ranked causes
- 1. Disposable set not installed or not properly loaded (common)
Verify disposable set is properly installed
- Test point
- Disposable load position
- Expected
- Disposable passes load criteria
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing disposable INOP alarm patient safety
Alarm message 'MISSING DISPOSABLE' is displayed; system indicates no disposable set is installed in the warmer unit
Ranked causes
- 1. No disposable set installed in the unit (common)
Verify disposable set is properly installed
- Test point
- Disposable set position in warmer
- Expected
- Disposable is fully installed and loads correctly per loaded disposal criteria
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No heating: battery not charging
Unit powered by battery but no heating function occurs. Battery does not charge when AC power applied.
Ranked causes
- 1. Power cord not plugged into AC receptacle (common)
Verify AC power cord connection
- Test point
- Power cord and AC receptacle
- Expected
- Power cord fully inserted in AC receptacle; indicator light on if present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Persistent Air Detection alarm patient safety
System repeatedly triggers air-in-line alarm during operation
Ranked causes
- 1. Actual air in fluid line or tubing (common)
See Air Detection in Alarm Message section of manual
- Expected
- Specific diagnosis and resolution per air detection section
Persistent Fluid Out alarm patient safety
System repeatedly triggers Fluid Out alarm; unable to complete prime countdown
Ranked causes
- 1. Fluid reservoir empty, disconnected, or blocked (common)
See Fluid Out in Alarm Message section of manual
- Expected
- Specific fluid supply diagnosis and resolution
Persistent High Pressure alarm patient safety
System repeatedly triggers high pressure alarm during operation
Ranked causes
- 1. Line or tubing obstructed; high viscosity fluid (common)
See High Pressure in Alarm Message section of manual
- Expected
- Specific cause diagnosis and resolution per High Pressure section
System does not prime patient safety
System will not count down from 100 to 0 ml in Prime mode
Ranked causes
- 1. Reservoir or recirculate line obstructed (common)
Verify reservoir and recirculate line clear
- Test point
- Reservoir and recirculate line
- Expected
- No obstructions present
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not prime or will not count down in prime cycle
Pump fails to draw fluid during priming; countdown from 100 to 0 ml does not complete
Ranked causes
- 1. Reservoir or recirculate line obstructed; fluid not available or bags not spiked (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature probes do not calibrate patient safety
Temperature probe calibration fails; incorrect fluid temperature or probe malfunction
Ranked causes
- 2. Temperature probe malfunction (rare)
Verify fluid temperature for calibration
- Test point
- Calibration fluid
- Expected
- Correct temperature per procedure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power up
Power switch pressed but unit does not turn on
Ranked causes
- 1. Power switch not completely depressed (common)
Power switch engagement
- Test point
- Power switch
- Expected
- Switch fully depressed and seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Valve does not move to desired position patient safety
Valve failure detected; system stops pumping and heating, alarm sounds and message displays. Valve did not reach target position within 2 seconds.
Ranked causes
- 1. Valve position sensor malfunction (common)
Observe valve position sensor response during valve operation
- Test point
- Valve mechanism
- Expected
- Valve moves to desired position within 2 seconds
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Watch dog alarm high voltage
Internal computer malfunction detected by hardware override circuit
Ranked causes
- 1. Processor malfunction (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
AC power | No AC power | Power cord not plugged in or AC outlet not powered | unknown |
AIR DETECTION | Air detected in the line | Air bubble detected in air detection sensor | warning |
AIR DETECTION | Air bubble detected in air detection sensor; fluid line contains air | Air bubble sensed in air detection sensor by optical detection | warning |
AIR DETECTOR FAULT | Air detector hardware failure | Air detector fails hardware test during system power-up | critical |
BATT LOW | Battery charge is low | Battery voltage drops below minimum threshold | warning |
Constant beep | Keypad continuously pressed | Key pad continuously depressed or stuck | unknown |
DOOR OPEN | Warmer door is open or door sensor indicates open status | Door sensor signals door is in open position; door is physically open or latch is not fully seated | error |
DOOR OPEN | Door is open | Door sensor signals door is open | warning |
FLUID OUT | Infusate supply is depleted or fluid flow is interrupted | Air bubble detected continuously in 1 ml of fluid pumped; tubing pulls away from sensor indicating vacuum in line | error |
FLUID OUT | Fluid depleted or fluid line pressure drop detected; system cannot pump | Air bubble sensed continuously in 1 ml of fluid pumped; tubing pulls away from sensor due to vacuum in line | error |
FLUID OUT | Air detected in top sensor indicating low or no fluid | Air bubble sensed continuously in 1 ml of fluid pumped | warning |
HEATER FAULT LATCH | Heater power control fault due to AC line noise or internal failure | Power control detects glitch on AC power line or internal failure | error |
HEATER FAULT LATCH | Heater power control failure at location A, B, C, or D | Power control detects glitch on AC power line or internal heater failure | error |
HEATER OVER POWER FAULT | Heater power consumption exceeds safe operating limit | Heater hardware power draw exceeds 1500 watts | error |
HEATER OVER POWER FAULT | Heater hardware fault with excessive power consumption | Power consumption exceeds 1500 watts | error |
HEATER POWER READ BACK FAULT | Heater power feedback sense coil is open or feedback circuit is malfunctioning | Zero feedback signal when power is applied to heater. Power feedback circuit malfunction. | error |
Heating fault message | System unable to heat fluid | Heater not reaching target temperature; sensor/power module failure | unknown |
HIGH PRESSURE | Line pressure exceeds safe limits; infusion blocked or restricted | Line pressure increase greater than 40 mm Hg/ml or absolute pressure exceeds 400 mm Hg; under pressure control, flow rate reduced to 0 ml/min | error |
IV:9 | Valve fault condition | Valve did not move to desired position within 2 seconds | error |
LOW BATTERY | Battery charge level is critically low; system cannot continue on battery power | Battery charge voltage drops below 22.5 ± 0.3 volts | warning |
MISSING DISPOSABLE | No disposable set is installed in the fluid warmer | Warmer fails to detect loaded disposable set; did not pass loaded disposal criteria when door is closed | error |
MISSING DISPOSABLE | Disposable set not installed or failed loading validation | No disposable set detected in unit or disposable did not pass loaded disposal criteria check | error |
POWER MODULE OVERTEMP | Power driver module overheating | Internal cabinet temperature reaches 70 degrees Celsius | warning |
PUMP FAULT | Pump malfunction due to encoder failure or mechanical jamming | Pump runs out of control or not at all. Difference between target and current speed > 15 ml/min for 1 second. Pump does not move when requested. | error |
Temp probe malfunction | Temperature probe not functioning | Probe fails during calibration or operation | unknown |
VALVE FAULT | Valve actuation failure or blockage | Valve position sensor malfunction; valve did not move to desired position within 2 seconds | error |
WATCH DOG | Internal computer malfunction or processor failure | Hardware override circuit detects processor not operating properly | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI